K Number
K222739

Validate with FDA (Live)

Date Cleared
2023-04-06

(209 days)

Product Code
Regulation Number
880.5570
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Disposable Insulin Pen Needle is intended for use with pen injector device for subcutaneous injection of insulin. It is sterile and for single use only. It can be used by the patient at home or healthcare professionals at medical/health care centers.

Device Description

The Disposable Insulin Pen Needle consists of a needle tube, a needle hub, a needle container, a needle shield and a seal. The needle tube is a double-ended needle that can be assembled into the needle hub using UV glue. The needle hub has the means of needle assembly attachment to allow it to be screwed onto the pen-injector device. This allows the Non-Patient (NP) end of the needle to penetrate through the rubber septum of the pen injector cartridge. The Patient and NP ends of the needle are lubricated using silicon oil for ease of injection and rubber septum penetration. The needle shield is assembled over the Patient end of the needle to protect the point from damage and accidental needle sticks. This needle assembly is inserted into a needle container and sealed with a peel-away label to provide a sterile barrier and tamper evidence. The peelaway label is pre-printed with information, which includes the lot number and needle gauge / length. It is supplied with several models. Different models are distinguished by needle gauge and length. The Disposable Insulin Pen Needle is a single-use disposable device and is provided sterile (EO sterilization). It is non-toxic and nonpyrogenic.

AI/ML Overview

Here's a breakdown of the requested information based on the provided FDA 510(k) summary for the Hantech Medical Device Co., Ltd. Disposable Insulin Pen Needle (K222739).

It's important to note that this document is a 510(k) premarket notification, which demonstrates substantial equivalence to a predicate device, rather than a full study proving a device meets novel acceptance criteria. Therefore, many of the typical AI/ML study components (like expert consensus, adjudication, MRMC studies, training set details) are not applicable here. This submission focuses on demonstrating that the new device performs similarly to an already approved device through established engineering and biocompatibility testing.


1. Table of Acceptance Criteria and Reported Device Performance

For this type of device, "acceptance criteria" are generally derived from recognized international and national standards, and "reported device performance" indicates conformity to these standards.

Acceptance Criteria CategorySpecific Criteria (from referenced standards)Reported Device Performance
BiocompatibilityAbsence of cytotoxicity (ISO 10993-5:2009)No cytotoxicity
Absence of skin sensitization (ISO 10993-10:2010)No evidence of sensitization
Absence of hemolysis (ISO 10993-4:2017)No evidence of hemolysis
Acceptable intradermal reactivity (ISO 10993-10:2010)Not explicitly detailed as "acceptable" but implied by conformity. "No evidence of skin irritation" is noted.
Absence of acute systemic toxicity (ISO 10993-11:2017)No systemic toxicity
Non-pyrogenic (ISO 10993-11:2017)Non-pyrogenic
Absence of subacute systemic toxicity (ISO 10993-11:2017)No Subacute Systemic Toxicity
Acceptable particulate levels (USP 788)Implied by conformity, no specific value given.
SterilizationSterility Assurance Level (SAL) of $10^{-6}$ (ISO 11135:2014)Sterilized by ethylene oxide gas SAL = $10^{-6}$
Compliance with ethylene oxide sterilization residuals (ISO 10993-7:2008)Implied by conformity to ISO 10993-7.
Packaging integrity and sterile barrier system requirements (ISO 11607-1:2019, ISO 11607-2:2019)Implied by conformity to these standards.
Shelf life stability (based on stability study and accelerated aging - ASTM F1980-16)5 years shelf life
PerformanceRequirements for hypodermic needles (ISO 7864:2016)Complied with ISO 7864
Requirements for stainless steel needle tubing (ISO 9626:2016)Complied with ISO 9626
Requirements for needle-based injection systems (ISO 11608-2:2012)Complied with ISO 11608-2
LabelingCompliance with 21 CFR Part 801Meet the requirements of 21 CFR Part 801

2. Sample Size Used for the Test Set and the Data Provenance

The document does not specify a "test set" in the context of clinical data or AI model evaluation. Instead, testing refers to product compliance with international standards for medical devices. The sample sizes for these specific engineering and biocompatibility tests are not detailed in this summary, but would typically be defined by the relevant ISO standards (e.g., a certain number of units for sterility testing or mechanical strength tests).

  • Data Provenance: The tests were conducted by the manufacturer, Hantech Medical Device Co., Ltd., which is located in Ningbo, Zhejiang, China. The testing would be prospective in the sense that physical devices were manufactured and then tested to demonstrate compliance.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

Not applicable. For this type of medical device 510(k) submission, "ground truth" is not established by human experts interpreting clinical data. Instead, it's established by objective measurements against predefined engineering standards and biocompatibility requirements. The "experts" involved would be qualified laboratory technicians, engineers, and biocompatibility specialists carrying out the specified tests following established protocols.

4. Adjudication Method for the Test Set

Not applicable. There is no qualitative assessment by human readers that would require an adjudication method. Test results are quantitative measurements or assessments against scientific criteria.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

Not applicable. This device is a physical medical product (insulin pen needle), not an AI/ML diagnostic or assistive technology. Therefore, no MRMC study or AI assistance evaluation was performed.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Not applicable. As noted above, this is a physical medical device, not an algorithm or software.

7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

The "ground truth" for this substantially equivalent device is the demonstrated compliance with recognized international and national standards for mechanical performance, sterility, and biocompatibility, as well as comparison to the characteristics of a legally marketed predicate device. This is based on objective laboratory and quality control testing, not clinical outcomes data, pathology, or expert consensus on interpretations.

8. The Sample Size for the Training Set

Not applicable. This is not an AI/ML device that requires a training set.

9. How the Ground Truth for the Training Set Was Established

Not applicable. This is not an AI/ML device, so no training set or its associated ground truth establishment is relevant.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. FOOD & DRUG ADMINISTRATION".

April 6, 2023

Hantech Medical Device Co., Ltd. Rachel Jin Official Correspondent No 288, Sanheng Road Changhe Industridal Park, Cixi Ningbo, Zhejiang 315326 China

Re: K222739

Trade/Device Name: Disposable Insulin Pen Needle Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: Class II Product Code: FMI Dated: March 7, 2023 Received: March 7, 2023

Dear Rachel Jin:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Alan Sth

CAPT Alan Stevens Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K222739

Device Name Disposable Insulin Pen Needle

Indications for Use (Describe)

The Disposable Insulin Pen Needle is intended for use with pen injector device for subcutaneous injection of insulin. It is sterile and for single use only. It can be used by the patient at home or healthcare professionals at medical/health care centers.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K22Z739 510(k) summary

I Submitter

Device submitter: Hantech Medical Device Co., Ltd. No 288, Sanheng Road Changhe Industridal Park, Cixi 315326, Ningbo PEOPLE'S REPUBLIC OF CHINA

Contact person: Name: Arnold Yang Title: Regulatory Affairs Manager Phone: +86 18917368988 Fax: +86 574 5899 5557 E-mail: rachel@hantechmedical.com

Date: 03/07/2023

II Device

Trade Name of Device: Disposable Insulin Pen Needle Common Name: Disposable Insulin Pen Needle Regulation Number: 21 CFR 880.5570 Classification: II Classification Name: Needle, Hypodermic, Single Lumen Product code: FMI Review Panel: General Hospital

III Predicate Devices

Trade name:Promisemed Insulin Pen Needle
Common name:Insulin Pen Needle
Classification:Class II, 21 CFR 880.5570
Product Code:FMI
Premarket Notification:K210059
Manufacturer:Promisemed Hangzhou Meditech Co., Ltd.

IV Device description

The Disposable Insulin Pen Needle consists of a needle tube, a needle hub, a needle container, a needle shield and a seal. The needle tube is a double-ended needle that can be assembled into the needle hub using UV glue. The needle hub has the means of needle assembly attachment to allow it to be screwed onto the pen-injector device. This allows the Non-Patient (NP) end of the needle to penetrate through the rubber septum of the pen injector

{4}------------------------------------------------

cartridge. The Patient and NP ends of the needle are lubricated using silicon oil for ease of injection and rubber septum penetration. The needle shield is assembled over the Patient end of the needle to protect the point from damage and accidental needle sticks. This needle assembly is inserted into a needle container and sealed with a peel-away label to provide a sterile barrier and tamper evidence. The peelaway label is pre-printed with information, which includes the lot number and needle gauge / length. It is supplied with several models. Different models are distinguished by needle gauge and length. The Disposable Insulin Pen Needle is a single-use disposable device and is provided sterile (EO sterilization). It is non-toxic and nonpyrogenic.

DeviceNeedle lengthNeedle gauge
Disposable Insulin Pen Needle4mm, 5mm, 6mm, 8mm, 12mm,34G, 33G, 32G, 31G, 30G, 29G,

V Indications for use

The Disposable Insulin Pen Needle is intended for use with pen injector device for subcutaneous injection of insulin. It is sterile and for single use only. It can be used by the patient at home or healthcare professionals at medical/health care centers.

VI Comparison of technological characteristics with the predicate devices

The Disposable Insulin Pen Needle has the same intended use, technology, and design as the predicate device and performance specifications are either identical or substantially equivalent to existing legally marketed predicate device. The differences between the Disposable Insulin Pen Needle and predicate device do not alter suitability of the proposed device for its intended use.

DevicefeatureSubject DevicePredicate Device K210059Comments
Indicationsfor useThe Disposable Insulin PenNeedle is intended for usewith pen injector device forsubcutaneous injection ofinsulin. It is sterile and forsingle use only. It can beused by the patient at homeor healthcare professionalsat medical/health carecenters.Promisemed Insulin PenNeedle is intended for usewith pen injector device forsubcutaneous injection ofinsulin. It is sterile and forsingle use only. It is suitablefor all age groups includingneonate, infant, children andadult, and can be used by thepatient at home or healthcareprofessionals at medical/health care centers.Same
ProductcodeFMIFMISame
Regulation21 CFR 880.557021 CFR 880.5570Same
DevicefeatureSubject DevicePredicate Device K210059Comments
number
ClassCLASS IICLASS IISame
Type of UseOver-The-Counter UseOver-The-Counter UseSame
Classification NameNeedle, Hypodermic, Single LumenNeedle, Hypodermic, Single LumenSame
Principle of operationThe user proceeds withinserting the needle into theskin manually. The patientend and the cartridge end ofthe tube are lubricated usinga silicone based lubricant forease of injection and rubberseptum penetration.The user proceeds withinserting the needle into theskin manually. The patientend and the cartridge end ofthe tube are lubricated usinga silicone based lubricant forease of injection and rubberseptum penetration.Same
Specific drug useInsulinInsulinSame
Needlegauge29G, 30G, 31G, 32G, 33G,34G29G, 30G, 31G, 32G, 33G,34G
NeedleLength4mm, 5mm, 6mm, 8mm,12mm$3.5mm \pm 0.4mm$(4mm, 5mm, 6mm, 8mm,12mm) $\pm 1.2mm$Substantially equivalentComment 1
LubricantSilicon oilSilicon oilSame
Single UseYesYesSame
MaterialsNeedle tube Stainless Steel(SUS304)Needle tube X5CrNi18-10Substantially equivalentComment 2
Needle Hub PolypropyleneNeedle Hub Polypropylene
Needle containerPolypropyleneNeedle containerPolypropylene
Needle ShieldPolypropyleneNeedle ShieldPolyethylene
Joint mediumUV glueJoint mediumUV glue
SealDialyzer paperSealDialyzer paper
SterilizationSterilized by ethylene oxidegas SAL = 10-6Sterilized by ethylene oxidegas SAL = 10-6Same
Shelf Life5 years5 yearsSame
PerformanceComplied with ISO 7864, ISO9626, ISO 11608-2Complied with ISO 7864, ISO9626, ISO 11608-2Same
BiocompatibilityComplied with ISO10993series standards, and thefollowing tests areperformedCytotoxicity: No cytotoxicityComplied with ISO10993series standards, and thefollowing tests areperformedCytotoxicity: No cytotoxicitySame
DevicefeatureSubject DevicePredicate Device K210059Comments
- Skin Irritation: No evidenceof skin irritation- Skin Sensitization: Noevidence of sensitization-Acute Systemic Toxicity:Nosystemic toxicity-Hemolysis: No evidence ofhemolysis-Pyrogen: Non-pyrogenic-Subacute Systemic Toxicity:No Subacute SystemicToxicity- Skin Irritation: No evidenceof skin irritation- Skin Sensitization: Noevidence of sensitization-Acute Systemic Toxicity:Nosystemic toxicity-Hemolysis: No evidence ofhemolysis-Pyrogen: Non-pyrogenic
LabelingMeet the requirements of 21CFR Part 801Meet the requirements of 21CFR Part 801Same

{5}------------------------------------------------

{6}------------------------------------------------

Discussion:

Comment 1

The subject device's needle length is covered by the range of lengths in the predicate device.

Comment 2

The materials of needle tube is different between the subject device and predicate device. The biocompatibility test of the subject device was conducted to demonstrate that the subject device met the biocompatibility requirements. So this difference does not raise any safety and effectiveness problems.

VII Performance data

The following performance data were provided in support of the substantial equivalence determination.

Biocompatibility testing

{7}------------------------------------------------

Biocompatibility of the Disposable Insulin Pen Needle was evaluated in accordance with ISO 10993-1:2018 for the body contact category of "External communication device - Blood path indirect" with a contact duration of "Prolonged (24 hours-30days )". The following tests were performed, as recommended:

CytotoxicityISO 10993-5: 2009
Skin sensitizationISO 10993-10: 2010
HemolysisISO 10993-4: 2017
Intradermal reactivityISO 10993-10: 2010
Acute systemic toxicityISO 10993-11: 2017
PyrogenicityISO 10993-11: 2017
Subacute Systemic ToxicityISO 10993-11: 2017
ParticulatesUSP 788

Sterilization and shelf life testing

The sterilization method has been validated to ISO11135, which has thereby determined the routine control and monitoring parameters. The shelf life of the Disposable Insulin Pen Needle is determined based on stability study which includes ageing test.

The testing is performed according to the following standards:

  • ISO 11135:2014 Sterilization of health-care products Ethylene oxide A Requirements for the development, validation and routine control of a sterilization process for medical devices
  • ISO 10993-7:2008 Biological evaluation of medical devices Part 7: Ethylene oxide A sterilization residuals
  • ISO 11607-1: 2019 Packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems and packaging systems

  • ISO 11607-2: 2019 Packaging for terminally sterilized medical devices - Part 2: Validation requirements for forming, sealing and assembly processes

  • ASTM F1980-16 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices

Performance testing

Performance testing is performed according to the following standards:

  • ISO 7864:2016 Disposable Medical Safety Hypodermic Needles — Requirements and test methods

  • ISO 9626:2016, Stainless Steel Needle Tubing For The Manufacture of Medical Devices

  • ISO ISO 11608-2:2012, Needle-based injection systems for medical use -Requirements and test methods - Part 2: Needles

VIII Conclusion

{8}------------------------------------------------

The Disposable Insulin Pen Needle is substantially equivalent to its predicate device (Promisemed Insulin Pen Needle). The non-clinical testing demonstrates that the device is as safe, as effective and performs as well as the legally marketed device.

§ 880.5570 Hypodermic single lumen needle.

(a)
Identification. A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.(b)
Classification. Class II (performance standards).