The Lansinoh® Smartpump 3.0 Double Electric Breast Pump is intended to express and collect the breastmilk of a nursing woman for the purpose of feeding collected breastmilk to a baby. The pump is intended for a single user.

Device Description

The Lansinoh Smartpump 3.0 Double Electric Breast Pump is a double electric breast pump intended to express the breast milk of a nursing woman. The pumping can be performed on one breast or on both breasts at the same time. The pumps can be powered by a rechargeable 7.4 V, 1,500 mAh lithium-ion battery or by an AC adapter that is provided with each pump. The pumping system consists of a diaphragm-type vacuum pump which is driven by a microprocessor-controlled DC electric motor. The user interface consists of a front panel keypad and LCD display. The user is able to adjust cycle mode and vacuum level based on personal comfort and preference.

AI/ML Overview

The provided text describes the acceptance criteria and the studies conducted for the Smartpump 3.0 Double Electric Breast Pump (K222726) to demonstrate its substantial equivalence to a predicate device (Lansinoh® Smartpump™ Double Electric Breast Pump (DEBP 2.2) [K182749]).

Here's a breakdown of the requested information based on the document:

1. A table of acceptance criteria and the reported device performance

The document doesn't explicitly present a table of "acceptance criteria" in the way one might expect for an AI diagnostic device (e.g., sensitivity, specificity thresholds). Instead, the performance studies are focused on demonstrating that the device meets its design specifications and relevant safety standards, and functions equivalently to the predicate. The "acceptance criteria" are implied by the standards and specifications the device passed.

Category of TestingAcceptance Criteria (Implied)Reported Device Performance
BiocompatibilityPatient-contacting components must be identical in formulation, processing, sterilization, and geometry to those cleared in K182749, and contain no new chemicals. Milk-contacting components must meet FDA's Food Additive criteria (21 CFR 175-179).Patient-contacting components are identical to those in K182749; no additional biocompatibility testing was needed. All milk-contacting materials meet FDA's Food Additive criteria.
Electrical Safety (IEC 60601-1)Device must satisfy general requirements for basic safety and essential performance, including US deviations.The device passed all tests in accordance with IEC 60601-1:2005 (3rd ed) + CORR. 1:2006 + CORR.2:2007+A1:2012 (except Clause 11.7, biocompatibility, which was addressed separately).
EMC (IEC 60601-1-2)Device must satisfy requirements for electromagnetic disturbances.The device passed all tests in accordance with IEC 60601-1-2:2014+A1:2021.
Home Healthcare (IEC 60601-1-11)Device must satisfy requirements for medical electrical equipment used in the home healthcare environment.The device passed all tests in accordance with IEC 60601-1-11:2015.
Battery Safety (IEC 62133-2)Secondary cells and batteries must meet safety requirements for portable sealed secondary cells.The device's battery complied with all relevant sections of IEC 62133-2:2017.
Software V&VSoftware should be verified and validated as recommended by FDA guidance and in accordance with IEC 62304:2016 for a Moderate level of concern. Cybersecurity documentation provided per FDA guidance.Software verification and validation testing was conducted and documentation provided; the software was considered a Moderate level of concern. Cybersecurity documentation was also provided.
Shelf-Life / Use-LifeExpected use-life of 500 hours demonstrated to operate within specifications.Devices demonstrated to operate within specifications for up to 500 hours of continuous operation at the maximum setting (Expression mode 3, Level 8).
Battery Life TestingBattery life must be approximately 2 hours at highest level (Expression Mode 3, Level 8).Testing supported that the battery life, as noted in the labeling, is approximately 2 hours at the highest level (Expression Mode 3, Level 8).
Suction Pressure Stability TestingDevice must meet its specifications and perform within the specified working ranges of pressure and cycle speed for each mode/level.The suction curves for each cycle mode and suction level demonstrated that the device meets its specifications and performs within the specified working ranges of pressure and cycle speed for each mode/level.
Backflow TestingThe design must prevent breastmilk from flowing into the tubing or pump body, especially at maximum pressure/cycle settings and various orientations. No milk should be present in the tubing during or after testing. Leverage data from K182749 for backflow protection mechanism materials if identical.The diaphragm was demonstrated to prevent backflow of milk into the tubing and pump. No milk was present in test devices' tubing during and following the test at maximum pressure/cycle settings (Expression Mode 3, Level 8) and various orientations. Backflow data from K182749 was leveraged.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Sample Size for Test Set: The document does not specify a numerical sample size for "test sets" in the context of human subjects or a dataset. The testing described (biocompatibility, electrical safety, software V&V, bench testing like battery life, suction pressure, backflow) appears to be primarily laboratory-based device testing, likely involving a limited number of device units.
  • Data Provenance: Not applicable in the context of this device's type of testing. The data originates from laboratory and bench testing of the device itself.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not applicable as the device is a breast pump, not an AI diagnostic device requiring expert interpretation of medical images or data to establish ground truth. The "ground truth" for this device's performance is objective measurements against engineering specifications and industry standards.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not applicable. Adjudication methods are typically used in clinical studies or expert review processes for diagnostic devices, not for basic performance testing of a breast pump.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This information is not applicable. This is not an AI diagnostic device that assists human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This information is not applicable. The Smartpump 3.0 is a physical medical device, not an AI algorithm. Its "standalone" performance refers to its functional operation as an electric breast pump, which is what the bench tests evaluate.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The "ground truth" for this device's performance is established by:

  • Compliance with recognized industry standards (e.g., IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 62133-2, IEC 62304).
  • Demonstration of performance against engineering specifications for parameters like battery life, suction pressure range and stability, and backflow prevention.
  • Meeting FDA Food Additive criteria for milk-contacting materials (21 CFR 175-179).

8. The sample size for the training set

This information is not applicable. This is not an AI/ML device that requires a training set. The "software" mentioned has a "Moderate level of concern," indicating it's likely firmware or control software, not an AI model trained on data.

9. How the ground truth for the training set was established

This information is not applicable as there is no training set for an AI/ML model for this device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

December 23, 2022

Lansinoh Laboratories Lindsay Ewers Director of Quality Assurance 99 Canal Center Plaza. Suite 550 Alexandria, VA 22314

Re: K222726

Trade/Device Name: Smartpump 3.0 Double Electric Breast Pump Regulation Number: 21 CFR§ 884.5160 Regulation Name: Powered Breast Pump Regulatory Class: II Product Code: HGX Dated: December 2, 2022 Received: December 2, 2022

Dear Lindsay Ewers:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Reginald K. Avery -S

for

Monica D. Garcia, Ph.D. Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K222726

Device Name

Smartpump 3.0 Double Electric Breast Pump

Indications for Use (Describe)

The Lansinoh® Smartpump 3.0 Double Electric Breast Pump is intended to express and collect the breastmilk of a nursing woman for the purpose of feeding collected breastmilk to a baby. The pump is intended for a single user.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

| | Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

1. SUBMITTER

Applicant:Lansinoh Laboratories, Inc.99 Canal Center Plaza, Suite 550Alexandria, VA 22314
Contact:Lindsay EwersDirector of Quality AssuranceLansinoh Laboratories, Inc.(727)-542-3743lewers@lansinoh.com
Submission Correspondent:Lindsay EwersDirector of Quality AssuranceLansinoh Laboratories, Inc.-(727)-542-3743lewers@lansinoh.com
Date Prepared:December 22, 2022

2. DEVICE

Device Trade Name:Smartpump 3.0 Double Electric Breast Pump
Device Common Name:Powered breast pump
Regulation Name:Powered breast pump
Regulation Number:21 CFR 884.5160
Regulatory Class:Class II
Product Code:HGX

3. PREDICATE DEVICE

Predicate Device: Lansinoh® Smartpump™ Double Electric Breast Pump (DEBP 2.2) [K182749]

The predicate device has not been subject to a design-related recall.

DEVICE DESCRIPTION 4.

The Lansinoh Smartpump 3.0 Double Electric Breast Pump is a double electric breast pump intended to express the breast milk of a nursing woman. The pumping can be performed on one breast or on both breasts at the same time. The pumps can be powered by a rechargeable 7.4 V, 1,500 mAh lithium-ion battery or by an AC adapter that is provided with each pump. The pumping system consists of a diaphragm-type vacuum pump which is driven by a microprocessor-controlled DC electric motor. The user interface consists of a front panel keypad

{4}------------------------------------------------

and LCD display. The user is able to adjust cycle mode and vacuum level based on personal comfort and preference.

INTENDED USE/INDICATIONS FOR USE ર.

The Lansinoh® Smartpump 3.0 Double Electric Breast Pump is intended to express and collect the breastmilk of a nursing woman for the purpose of feeding collected breastmilk to a baby. The pump is intended for a single user.

SUBSTANTIAL EQUIVALENCE 6.

Technological Comparisons

The table below compares the key technological feature of the subject devices to the predicate device (Lansinoh® Smartpump™ Double Electric Breast Pump (DEBP 2.2) [K182749]).

Subject DeviceLansinoh® Smartpump 3.0 DoubleElectric Breast PumpPredicate DeviceLansinoh® Smartpump™ DoubleElectric Breast Pump (DEBP 2.2)
510(k) NumberK222726K182749
Indications For UseThe Lansinoh® Smartpump 3.0Double Electric Breast Pump isintended to express and collect thebreastmilk of a nursing woman for thepurpose of feeding collected breastmilkto a baby. The pump is intended for asingle user.The Lansinoh® Smartpump™ DoubleElectric Breast Pump is intended toexpress and collect the breastmilk of anursing woman for the purpose offeeding collected breastmilk to a baby.The pump is intended for multipleusers, and single users.
ApplicantLansinoh LaboratoriesLansinoh Laboratories
ClassificationRegulation884.5160884.5160
Product CodeHGXHGX
Regulatory ClassClass IIClass II
Patient PopulationLactating WomenLactating Women
Suction ModesStimulation and ExpressionStimulation and Expression
Suction Levels(stimulation)59-236 mmHg46-140 mmHg
Cycles per Second(stimulation)1.36-2.691.61-2.33
Suction Levels(expression)89 mmHg - 280 mmHg95 mmHg - 280 mmHg
Cycles per Second(expression)0.53-1.830.58 - 1.69

Table 1: Technological Comparison

{5}------------------------------------------------

Subject DeviceLansinoh® Smartpump 3.0 DoubleElectric Breast PumpPredicate DeviceLansinoh® Smartpump™ DoubleElectric Breast Pump (DEBP 2.2)
Suction Settings88
Expression modes33
Power SupplyAC AdapterReplaceable, rechargeable 7.4 V, 1,500mAh lithium-ion polymer batteryAC Adapter6 AA alkaline batteries
Pumping OptionSingle or DoubleSingle or Double
Back Flow ProtectionYesYes
Let Down FunctionYesYes
Cycling/Suction ControlMechanismMicroprocessor with internal programmemoryMicroprocessor with internal programmemory
Program Memory32K flash ROM, 2K RAM memory32K flash ROM, 2K RAM memory
Communication withmobile appBluetooth BLE 4.2Bluetooth version 4.1
Expected Use Life500 hours500 hours
AccessoriesTubing with Velcro Strap; Flange Cap;Diaphragm; Breast Flange Body;Comfort Fit Breast Flange (Cone);Cone Soft Edge; White Valve; 5 ozContainer; Container Ring; SealingDisk; Storage LidTubing with Velcro Strap; Flange Cap;Diaphragm; Breast Flange Body;Comfort Fit Breast Flange (Cone);Cone Soft Edge; White Valve; 5 ozContainer; Container Ring; SealingDisk; Storage Lid
Wireless ConnectivityBluetooth BLE 4.2 with an optionalmobile appBluetooth with an optional mobile app

The subject device has the same intended use, but different technological characteristics compared to the predicate. The subject and predicate device operate at different cycle speeds, levels, and suction strengths for stimulation and expression modes, and power supply. The predicate device is also indicated for multiple and single users while the subject device is indicated for single users only. The differences in the indications and technological characteristics do not raise different questions of safety and effectiveness.

7. PERFORMANCE DATA

Biocompatibility Testing

The patient contacting components, in their final finished form, are identical to the Smartpump 1.0 Double Electric Breast Pump cleared in K182749 in formulation, processing, sterilization, and geometry and no other chemicals have been added (e.g., plasticizers, fillers, additives, cleaning agents, mold release agents). Therefore, no additional biocompatibility testing was conducted.

{6}------------------------------------------------

The milk contacting components are those components that are provided in the pump kit. All materials in contact with milk have been tested to meet FDA's Food Additive criteria (21 CFR 175-179)

Electrical safety and electromagnetic compatibility (EMC)

The Smartpump 3.0 Double Electric Breast Pump was tested in accordance with IEC 60601-1:2005 (3rd ed) + CORR. 1:2006 + CORR.2:2007+A1:2012 Medical electrical equipment: Part 1: General requirements for basic safety and essential performance including US deviations, with the exception of Clause 11.7 regarding biocompatibility. The device passed all tests.

The Smartpump 3.0 Double Electric Breast Pump was tested in accordance with the FDArecognized standard IEC 60601-1-2:2014+A1:2021, Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance – Collateral Standard: electromagnetic disturbances – Requirements and tests. The device passed all tests.

The Smartpump 3.0 Double Electric Breast Pump was tested in accordance with IEC 60601-1-11:2015, Medical electrical equipment-Part 1-11: General requirements for basic safety and essential performance – Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment. The device passed all tests

The Smartpump 3.0 Double Electric Breast Pump was tested in accordance with the FDArecognized standard, IEC 62133-2:2017, Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications – Part 2: Lithium systems, and found to comply with all relevant sections.

Software Verification and Validation Testing

Software verification and validation testing was conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." and in accordance with IEC 62304:2016, Medical device software - Software life cycle processes. The software for this device was considered as a Moderate level of concern.

Cybersecurity documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices"

Sterilization, Cleaning, and Shelf-Life

Sterilization and Cleaning

The device is reusable, provided non-sterile, and is not sterile when used. Cleaning instructions are provided in the labeling.

Shelf-Life

Shelf-life is not applicable due to the low likelihood of time-dependent product degradation. However, in accordance with IEC 60601-1:2005/(R)2012 the subject devices expected use-life is 500 hours. In testing, the devices were demonstrated to operate within specifications for up to 500 hours of continuous operation at the maximum setting (Expression mode 3, Level 8).

{7}------------------------------------------------

Bench Testing

The following bench testing was performed to demonstrate substantial equivalence:

  • Battery life Testing .
    Battery life was measured with the pumps at the highest level ((Expression Mode 3, Level 8). The "fail time" was defined as the number of minutes the pump could run at this level before dropping below specifications. Testing supports that the battery life of the device, as noted in the labeling, is approximately 2 hours.

  • Suction Pressure Stability Testing ●
    Devices were evaluated for suction performance at each of the available settings. The suction curves for each cvcle mode and suction level demonstrated that the device meets its specifications and performs within the specified working ranges of pressure and cycle speed for each mode/level.

  • Backflow Testing ●
    The Lansinoh pumps are designed as a closed milk collection system. The diaphragm provides a physical barrier, preventing breastmilk from flowing into the tubing or pump body. There have been no changes to the Backflow Protection mechanism materials used in the subject device since its previous clearance in K182749. Therefore, Backflow Protection data from K182749 can be leveraged to support the biocompatibility of the subject devices. The purpose of the backflow test is to demonstrate that the design prevents backflow into the tubing and pump. Devices were tested at maximum pressure/cycle settings (Expression Mode 3, Level 8) at various orientations to simulate worst-case conditions. The testing demonstrated that no milk was present in test devices' tubing during and following the test; therefore, the diaphragm was demonstrated to prevent back flow of milk into the tubing and pump.

CONCLUSION 8.

The results of the testing described above demonstrate that the Smartpump 3.0 Double Electric Breast Pump is as safe and effective as the predicate device and supports a determination of substantial equivalence.

§ 884.5160 Powered breast pump.

(a)
Identification. A powered breast pump in an electrically powered suction device used to express milk from the breast.(b)
Classification. Class II (performance standards).