K Number
K222711

Validate with FDA (Live)

Device Name
HS186B Scooter
Date Cleared
2023-01-09

(123 days)

Product Code
Regulation Number
890.3800
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

It is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.

Device Description

The subject device, Model HS186B Mobility Scooter, is an indoor/outdoor electric scooter that is intended to be used by individuals that are able to walk but suffer from mobility limitations. It has a base with frame, two front wheels, two rear wheels, two anti-tip wheels, a seat, a tiller, a control panel, an electric motor, an electromagnetic brake, a rechargeable Lithium-lon Battery with a charger. The movement of the scooter is controlled by the rider who operates the throttle Control Lever, speed knob and hand grips. The device is installed with an electromagnetic brake that will engage automatically when the scooter is not in use and the brake cannot be used manually. The Scooter is suitable for use indoors and on flat roads near buildings.

AI/ML Overview

The provided text is a 510(k) summary for a medical device (HS186B Scooter) and an FDA clearance letter. It details a comparison between the proposed device and a predicate device, focusing on non-clinical testing and safety features to demonstrate substantial equivalence. However, it does not describe acceptance criteria for an AI device or a study involving AI performance metrics.

Therefore, I cannot fulfill your request for:

  • A table of acceptance criteria and reported device performance for an AI device.
  • Sample size used for the test set and data provenance for an AI device.
  • Number of experts and their qualifications for establishing ground truth for an AI test set.
  • Adjudication method for an AI test set.
  • MRMC comparative effectiveness study results (AI improvement vs. without AI assistance).
  • Standalone AI performance.
  • Type of ground truth used for AI.
  • Sample size for AI training set.
  • How ground truth for AI training set was established.

This document pertains to a physical medical device (a motorized scooter) and its performance is evaluated against ISO standards and through comparison with a predicate device, not through AI-specific metrics. The "Non-Clinical Testing" section describes adherence to various ISO standards relevant to wheelchairs and scooters, and a "Biocompatibility" assessment. The document explicitly states "No clinical study implemented for the scooter." It does not mention any AI components or related studies.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

January 9, 2023

Suzhou Master Machinery Manufacturing Co.,Ltd. % Boyle Wang Official Correspondent Shanghai Truthful Information Technology Co., Ltd. RM. 1801, No. 161 East Lujiazui Rd.,Pudong Shanghai, Shanghai 200120 China

Re: K222711

Trade/Device Name: HS186B Scooter Regulation Number: 21 CFR 890.3800 Regulation Name: Motorized three-wheeled vehicle Regulatory Class: Class II Product Code: INI Dated: October 10, 2022 Received: October 11, 2022

Dear Boyle Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Tushar Bansal -S

for Heather Dean, PhD Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K222711

Device Name HS186B Scooter

Indications for Use (Describe)

It is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR 807.92.

1.0 Submitter's information

Name: Suzhou Master Machinery Manufacturing Co.,Ltd. Address: No.48,Binjiang Road,Zhangpu Town, Kunshan,Jiangsu, China 215321 Tel: 86-512-57452882 Fax: 86-512 - 57452992 Contact: Siqi.Shangguan Date of Preparation: Jul.08,2022

Designated Submission Correspondent

Mr. Boyle Wang Shanghai Truthful Information Technology Co., Ltd. Room 1801, No. 161 East Lujiazui Rd., Pudong Shanghai, 200120 China Tel: +86-21-50313932 Email: Info@truthful.com.cn

2.0 Device information

HS186B Scooter Trade name: Common name: Scooter Classification name: Motorized Three-Wheeled Vehicle Model(s): HS186B

3.0 Classification

Production code: INI Regulation number: 21 CFR 890.3800 Classification: Class II Panel: Physical Medicine

4.0 Predicate device information

Nanjing Jin Bai He Medical Apparatus Co., Ltd. Manufacturer: Trade/Device: Scooter (Model: FDB01) 510(k) number: K201196

5.0 Indication for Use Statement

{4}------------------------------------------------

It is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.

6.0 Device description

The subject device, Model HS186B Mobility Scooter, is an indoor/outdoor electric scooter that is intended to be used by individuals that are able to walk but suffer from mobility limitations. It has a base with frame, two front wheels, two rear wheels, two anti-tip wheels, a seat, a tiller, a control panel, an electric motor, an electromagnetic brake, a rechargeable Lithium-lon Battery with a charger. The movement of the scooter is controlled by the rider who operates the throttle Control Lever, speed knob and hand grips.

The device is installed with an electromagnetic brake that will engage automatically when the scooter is not in use and the brake cannot be used manually. The Scooter is suitable for use indoors and on flat roads near buildings.

Folding design: To fold the scooter pull out the fold knob, at the same time gently lower the seat back down into its folded position, Now with the seat folded, loosen off the tiller adjuster knob and gently lower the tiller into the folded position. The tiller handles should be able to rest between the chassis. Re-tighten the tiller knob to hold the tiller into position. This will then allow you to pull the scooter on its rear tip wheels, the folding is completed.

Hold the straight handle, push the right throttle control lever forward with the right thumb, and the scooter will move forward. Push the left throttle control lever forward with the left thumb, the scooter will move backwards. The horn will sound a warning

sound when the scooter is moving backwards.

When the throttle control lever is fully released, it will automatically stop in the "center" position, and the electromagnetic brake will automatically brake the scooter.

The motor of the Scooter is DC24V 120W; the battery is 24V 10AH, Li-ion battery; the charger is 24V/2A.

Max. loading can not be over than 120Kgs.

Max. distance of travel on the fully charged battery is 15.8km and Max. speed forward is 7km/h.

The braking time is about 1.6s, and the braking distance is ≤1m.

7.0 Summary of Non-Clinical Testing

Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:

{5}------------------------------------------------

ISO 7176-1:2014 Wheelchairs - Part 1: Determination of static stability

ISO 7176-2:2017 Wheelchairs - Part 2: Determination of dynamic stability of electrically powered wheelchairs

ISO 7176-3:2012 Wheelchairs - Part 3: Determination of effectiveness of brakes

ISO 7176-4:2008 Wheelchairs - Part 4: Energy consumption of electric wheelchairs and scooters for determination of theoretical distance range

ISO 7176-5:2008 Wheelchairs - Part 5: Determination of overall dimensions, mass and manoeuvring space

ISO 7176-6:2018 Wheelchairs - Part 6: Determination of maximum speed, acceleration and deceleration of electric wheelchairs

ISO 7176-7:1998 Wheelchairs - Part 7: Measurement of seating and wheel dimensions

ISO 7176-8:2014 Wheelchairs - Part 8: Requirements and test methods for static, impact and fatigue strengths

ISO 7176-9:2009 Wheelchairs - Part 9: Climatic tests for electric wheelchairs

ISO 7176-10:2008 Wheelchairs - Part 10: Determination of obstacle-climbing ability of electrically powered wheelchairs

ISO 7176-11: 2012 Wheelchairs - Part 11: Test dummies

ISO 7176-13:1989 Wheelchairs - Part 13: Determination of coefficient of friction of test surfaces

ISO 7176-14:2008 Wheelchairs - Part 14: Power and control systems for electrically powered wheelchairs and scooters - Requirements and test methods

ISO 7176-15:1996 Wheelchairs - Part 15: Requirements for information disclosure, documentation and labeling

ISO 7176-16:2012 Wheelchairs -- Part 16: Resistance to ignition of postural support devices

ISO 7176-21:2009 Wheelchairs - Part 21: Requirements and test methods for electromagnetic compatibility of electrically powered wheelchairs and scooters, and battery chargers

ISO 7176-22:2014 Wheelchairs - Part 22: Set-up procedures

ISO 7176-25:2013 Wheelchairs - Part 25: Batteries and chargers for powered wheelchairs

IEC 60601-1-2:2014 Medical electrical equipment - Part 1-2: General requirements for safety - Collateral standard: Electromagnetic compatibility - Requirements and tests EN 12184 :2014 Electrically powered wheelchairs, scooters and their chargers-Requirements and test methods

Biocompatibility of patient-contacting parts

Patient-contacting material are carried out biocompatibility assessment in accordance with ISO 10993-1: 2018, including:

Cytotoxicity per ISO 10993-5:2009 Biological Evaluation of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity

{6}------------------------------------------------

Irritation and Skin Sensitization per ISO 10993-10:2010 Biological Evaluation of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization.

8.0 _Summary of Clinical Testing

No clinical study implemented for the scooter.

9.0 Technological Characteristic Comparison Table

{7}------------------------------------------------

ItemSubject devicePredicate deviceRemark
Product CodeINIINISame
Regulation No.21 CFR 890.380021 CFR 890.3800Same
ClassIIIISame
Product nameScooter (Model: HS186B)Scooter (Model: FDB01)-
510(k) No.K222711K201196-
IntendedUse/Indication forUseIt is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.It is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.Same
Use environmentIndoor and outdoor useIndoor and outdoor useSame
PatientPopulationThis product is suitable for disabled people with mobility difficulties and elderly people.This product is suitable for disabled people with mobility difficulties and elderly people.Same
Product structureIt has a base with frame, two front wheels, two rear wheels, two anti-tip wheels, a seat, a tiller, a control panel, an electric motor, an electromagnetic brake, a rechargeable Lithium-Ion Battery with a charger.It has a base with metal alloy frame, two front wheels, two rear wheels, two anti-tip wheels, a seat, an adjustable steering column, a tiller console, an electric motor, an electromagnetic brake, 2 rechargeable Lithium-Ion Battery with an off-board charger.Similar
Driving systemDirect drive on the rear wheelsDirect drive on the rear wheelsSame
Number ofwheels44Same
Main framematerialAluminium alloyAluminium alloySame
MotorBrushless motor, DC24V* 120W*1 pcsbrush differential rear axle. 24V *180WMinor differences in the
BatteryDC 24V 10Ah Lithium-ion, 1 pcs24V 6AH Lithium-ion, 2 pcsdimensions will not
Battery chargerOff-board charger Input: 100-240 VAC Output: DC 24V, 2AOff-board charger Input: 100-240 VAC Output: DC 24V, 6 Ampimpact the safety and effectiveness of the substantial equivalence.
ItemSubject DevicePredicate DeviceK201196Remark
Dimensions(mm)1016x560x9141050 X 550 X 870Minor differences in thedimensions will notimpact the safety andeffectiveness of thesubstantial equivalence.
Foldeddimensions(mm)950x560x460480 x550 x790Minor differences in thefolded dimensions will notimpact the safety andeffectiveness of thesubstantial equivalence.
Weight, w/Battery42.1lbs. /19.1kg63.1lbs. /29 kgThe difference willnot raise any new safetyand effectivenessconcerns.
Frame designFoldable/The scooter is driven by electricityscooter, is a transportation tool andauxiliary tools for the purpose ofwalking, without removing thebattery can be folded, easy to carryand storage.Foldable/Electric scooter is driven byelectricity scooter, is atransportationtoolandauxiliary tools for thepurpose of walking, withoutremoving the battery can befolded, easy to carry andstorage.Same
SeatingDesignTo fold the scooter pull out the foldknob, at the same time gently lowerthe seat back down into its foldedposition, Now with the seat folded,loosen off the tiller adjuster knoband gently lower the tiller into thefolded position. The tiller handlesshould be able to rest between thechassis. Re-tighten the tiller knob tohold the tiller into position. This willthen allow you to pull the scooter onits rear tip wheels, the folding iscompleted.One-click foldingintelligent electric scooter,with a sensitive remotecontrol, gently press, thecar automaticallycompleted foldingcontraction, folding volumesmall, electromagneticautomatic brake, the overallsimple design, convenientand practical, Travel does nothave to consume physicalforce to do folding, poorphysical strength and theelderly the best walkingtool.The difference willnot raise any new safetyand effectivenessconcerns.
SeatingAttachmentWithout any seataccessories (integrated, powerWithout any seataccessories (integrated,Same
integrated,powerbase, specialtypowerbase, dedicated power supply)power base, dedicatedpower supply)
FoldingmechanismA foldable seat frames(The backrest could be folded toseat)Seat can be folded; batterycan be dismantled; Theframe can be folded backand forthSame
Frontwheel(inch)6(Solid tire)7(Solid tire)Smaller size of the frontwheel.The difference will not raiseany new safety andeffectiveness concerns.
Rear tire (inch)10(Solid tire)8(Solid tire)Larger size of rear wheelsbring steadier pivotingfunction than predicatedevice.
Anti-tip Wheels(inch)2.52.5Same
CruisingRange(km)15.8Not Publicly AvailableThe difference will not raiseany new safety andeffectiveness concerns.
Obstacleclimbing(mm)5060The smaller height in theobstacle climbing will notimpact the safety andeffectiveness of the subjectdevice.
Groundclearance75mm50mmThe device has been testedaccording to ISO7176 seriesstandards and the testrecords support its safety andeffectiveness.
Static stabilityforward25°30°Both of the devices areevaluated according tostandard ISO 7176-1:2014,
Static stabilityrearward14.2°20.1°standard ISO 7176-1:2014,so the different static stabilitywill not impact the safety andeffectiveness
Static stabilitysideways12.6°15.3°will not impact the safety andeffectiveness
Max. loading(kg)265lbs(120kg)265lbs (120kg)Same
Maximum safeoperationalinclineMaximum safe operational inclinefor posteriorly is >25Maximum safe operationalincline for posteriorly is 20.1°Larger safe operationalincline of subject brings moreconvenient for the use
for anteriorly is 14.2°Maximum safe operational inclinefor sideways is 14.2°incline for anteriorly is 30°Maximum safe operationalincline for sideways is 15.3°environment
Min. Turningradium1225mm1200mmThe little difference in theturning radius will bring moreconvenience when it turns.The difference will not raiseany new safety andeffectiveness concerns.
Minimumbrakingdistance1.0m for the speed of 7km/h1.6s1.1 m for the speed of 6km/h1sSimilar
MaxSpeedForwards1.88m/s (7 km/h)1.67m/s (6 km/h)The devices are evaluatedaccording to standard ISO7176-6:2018, so the differentwill not impact the safetyand effectiveness
Max.SpeedBackward0.58 m/s (2 km/h)0.83m/s (3 km/h)The devices are evaluatedaccording to standard ISO7176-6:2018, so the differentwill not impact the safety andeffectiveness
ControllerYangzhou Feya ElectronicTechnology Co., Ltd.FY0-7 FULLBritish PG Controller PG45AAlthough different controlleris used, both the controlsystem, including theelectromagnetic brakes andthe user interface are similar.Both devices have the sameuser interface, the patientuses the foldable tillerhandle/handlebar for steeringand a thumb operatedpotentiometer throttle controllever located at the top of thetiller to engage anddisengage the scooter motionin both the forward andreverse directions. When thethrottle control lever isreleased, the electromagneticbrake will be actuated andthe scooter is slow to stop.The speed mode of the bothdevices was single mode and
the speed control dial cancontrol the maximum speed.Both of the control systemsare evaluated according tostandard ISO 7176-14:2008and software validationrequirement and there are nonew safety and effectivenessconcerns due to thedifference.
Wheel(type)LockElectromagnetic brakeSame
Electromagnetic brake

Table1-General Comparison

{8}------------------------------------------------

Table2 Performance Comparison

{9}------------------------------------------------

{10}------------------------------------------------

{11}------------------------------------------------

Table3 Safetv Comparison

ItemProposed DevicePredicate DeviceRemark
MaterialscontactinguserArmrest: TRPBackrest: TeryleneSeat: LeatherArmrest: PU Foamed;Backrest/seat: Leather packageFootplates: ABS plasticsBiocompatibilityevaluation hasbeen carried outper ISO 10993-1.There are no newsafety andeffectivenessconcerns due to thedifference.
Biocompatibility of materialscontactinguserComply with ISO 10993-1, FDAGuidance, Tests includedCytotoxicity (ISO 10993-5:2009),Sensitization and IntracutaneousReactivity (ISO 10993-10:2010)Comply with ISO 10993-1, FDAGuidance, Tests includedCytotoxicity (ISO 10993-5:2009),Sensitization and IntracutaneousReactivity (ISO 10993-10:2010)Same

Summary of substantial equivalence discussion:

Despite of the above differences, the two devices all completed the performance tests in accordance with ISO 7176 series standards. There are no safety and effectiveness aspects concerned. In conclusion, the technological characteristics, features, specifications, materials, mode of operation, and intended use of the device substantially equivalent to the predicate devices quoted above. The differences between the subject device and predicate devices do not raise new issues of safety or effectiveness. The differences between the subject device and predicate devices do not raise new issues of safety or effectiveness.

Conclusion 10.0

The conclusions drawn from the comparison and analysis above demonstrate that the

{12}------------------------------------------------

subject device is as safe, as effective, and performs as well as the legally marketed predicated device in K201196 and raises no new questions of safety or effectiveness. The differences between both devices are insignificant in terms of safety and effectiveness.

§ 890.3800 Motorized three-wheeled vehicle.

(a)
Identification. A motorized three-wheeled vehicle is a gasoline-fueled or battery-powered device intended for medical purposes that is used for outside transportation by disabled persons.(b)
Classification. Class II (performance standards).