K Number
K222698

Validate with FDA (Live)

Device Name
CUVIS-spine
Manufacturer
Date Cleared
2022-10-07

(30 days)

Product Code
Regulation Number
882.4560
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

CUVIS-spine is intended for use as an aid for precisely locating anatomical structures and for the spatial positioning and orientation of guide bush to be used by surgeons for navigating compatible surgical instruments in open or percutaneous surgical procedures provided that the required markers and rigid patient anatomy can be identified on O-arm or C-arm.

Device Description

The CUVIS-spine is a mobile system mainly comprising the robotic arm, the main console and the staff console as an option. The robotic arm is positioned on the floor near the side of the surgical table. The location of the main console or the staff console is appropriately determined considering the user preference and the environments. CUVIS-spine is a pedicle screw guide system which consists of Robotic Arm, Main Console, Staff Console, Guide bush, Source Calibrator, Registration Tool, Registration Tool Adapter, Robotic Arm Drape, Tool Drape, Detector Drape, Marker Ball, Patient Marker, Marker Driver, Detector Calibrator, Dilator, Serration-tip Dilator, Drill Bit, Tapper, Stylet Tapper, Screwdriver, Instrument Container.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the CUVIS-spine device, based on the provided document:

1. Table of Acceptance Criteria and Reported Device Performance

The document focuses on claiming substantial equivalence to a predicate device (CUVIS-spine K201569) rather than explicitly stating new, quantifiable acceptance criteria for performance metrics. However, it does mention that "Pose accuracy and Repeatability of the CUVIS-spine were tested and validated." Without specific numerical targets for these, a direct table of acceptance criteria vs. performance is not fully possible.

However, based on the non-clinical tests mentioned, we can infer performance was evaluated against general medical device safety and efficacy standards. The "Accuracy" row in the comparison table with the predicate device also indicates that the proposed device maintains accuracy.

Feature / Performance AspectAcceptance Criteria (Implied/General)Reported Device Performance
Pose AccuracyTo be within acceptable limits for a stereotaxic instrument for spinal surgery, comparable to predicate."Tested and validated." (Specific numerical results not provided in this summary.)
RepeatabilityTo be within acceptable limits for a stereotaxic instrument for spinal surgery, comparable to predicate."Tested and validated." (Specific numerical results not provided in this summary.)
BiocompatibilityCompliance with ISO 10993 standards (cytotoxicity, irritation, skin sensitization, intracutaneous reactivity)."Evaluated according to ISO 10993-5" and "evaluated according to ISO 10993-10."
Electrical SafetyCompliance with ES 60601-1."Tested and evaluated according to ... ES 60601-1."
Electromagnetic CompatibilityCompliance with IEC 60601-1-2."Tested and evaluated according to ... IEC 60601-1-2."
Risk ManagementCompliance with ISO 14971."Recorded by referring to ISO 14971."
UsabilityCompliance with IEC 60601-1-6."Documented by referring to IEC 60601-1-6."

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: The document mentions a cadaveric study was performed for the robotic-assisted pedicle screw placement. However, the sample size (number of cadavers or placement procedures) is not specified in the provided text.
  • Data Provenance: The document does not explicitly state the country of origin for the cadaveric study data. It is a non-clinical, prospective study (as it was performed specifically for testing the proposed device).

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not provided in the document. For a cadaveric study of pedicle screw placement, ground truth would typically be established by imaging (e.g., post-procedure CT scans) reviewed by qualified medical professionals (e.g., orthopedic surgeons or radiologists). The number and qualifications of such experts are not detailed here.

4. Adjudication Method for the Test Set

This information is not provided in the document.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

No, an MRMC comparative effectiveness study involving human readers with and without AI assistance is not mentioned in the document. The study described is a cadaveric study focused on physical performance and accuracy, not reader performance.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

The device itself (CUVIS-spine) is a robotic-assisted surgical system, implying human interaction. The non-clinical tests focused on the device's "Pose accuracy and Repeatability," which are standalone measures of the robot's mechanical performance. However, there isn't a specific mention of an "algorithm only" study in the context of image interpretation or diagnosis separate from the physical robotic guidance. The cadaveric study inherently involves the algorithm driving the robot.

7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)

For the cadaveric study on robotic-assisted pedicle screw placement, an implied ground truth would likely be established through post-procedure imaging (e.g., CT scans) to assess the accuracy of screw placement relative to the planned trajectory and anatomical landmarks. This assessment would typically be performed by experts, but the specifics are not detailed.

8. The Sample Size for the Training Set

This device is a robotic surgical system, not an AI model that learns from large datasets in the traditional sense (e.g., for image classification). Therefore, a "training set" as understood in machine learning for diagnostic AI is not applicable in the context described in this document. The robotic system's "training" would involve mechanical calibration, software development, and validation against engineering specifications, not data-driven learning from patient cases in the same way.

9. How the Ground Truth for the Training Set Was Established

As explained in point 8, a "training set" for an AI model learning from data is not described. The ground truth for the robotic system's development would be based on engineering specifications, physical measurements, and validation against known anatomical models and surgical planning parameters.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 7, 2022

Curexo, Inc. % Do Kim CEO BT Solutions, Inc. Unit 904, Eonju-ro 68-gil 5, Gangnam-gu Seoul, 06210 Korea, South

Re: K222698

Trade/Device Name: CUVIS-spine Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: September 5, 2022 Received: September 7, 2022

Dear Do Kim:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrl/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation -emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K222698

Device Name

CUVIS-spine (Model name: CS100)

Indications for Use (Describe)

CUVIS-spine is intended for use as an aid for precisely locating anatomical structures and for the spatial positioning and orientation of guide bush to be used by surgeons for navigating compatible surgical instruments in open or percutaneous surgical procedures provided that the required markers and rigid patient anatomy can be identified on O-arm or C-arm.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.

{3}------------------------------------------------

CUVIS-spine 510(k) Summary K222698

5. 510(k) Summary

5.1. General Information

Applicant/Submitter:CUREXO, INC.
Address:577, Gangnam-Daero, Seocho-GuSeoul, 06530, Republic of KoreaTel: +82-31-788-7935
Contact Person:Do Hyun Kim
Address:Unit 904, Eonju-ro 86gil 5, Gangnam-guSeoul 06210, Korea.Tel: +82-2-538-9140Email: ceo@btsolutions.co.kr
Preparation Date:September 5, 2022

5.2. Device Name and Code

Device Trade Name:CUVIS-spine
Model Name:CS100
Common Name:Pedicle Screw Guide System
Classification Name:Stereotaxic instrument
Product Code:OLO
Regulation Number:21 CFR 882.4560
Classification:II
Review Panel:Orthopedic

5.3. Predicate Device

CUVIS-spine is substantially equivalent to the following legally marketed predicate devices

Table 1 Primary Predicate device

ApplicantDevice Name510(k) Number
CUREXO, INCCUVIS-spineK201569

5.4. Device Description

The CUVIS-spine is a mobile system mainly comprising the robotic arm, the main console and

{4}------------------------------------------------

CUVIS-spine K22698

the staff console as an option. The robotic arm is positioned on the floor near the side of the surgical table. The location of the main console or the staff console is appropriately determined considering the user preference and the environments.

CUVIS-spine is a pedicle screw guide system which consists of Robotic Arm, Main Console, Staff Console, Guide bush, Source Calibrator, Registration Tool, Registration Tool Adapter, Robotic Arm Drape, Tool Drape, Detector Drape, Marker Ball, Patient Marker, Marker Driver, Detector Calibrator, Dilator, Serration-tip Dilator, Drill Bit, Tapper, Stylet Tapper, Screwdriver, Instrument Container.

5.5. Indications / Intended Use

CUVIS-spine is intended for use as an aid for precisely locating anatomical structures and for the spatial positioning and orientation of guide bush to be used by surgeons for navigating and/or guiding compatible surgical instruments in open or percutaneous surgical procedures provided that the required markers and rigid patient anatomy can be identified on O-arm or Carm.

5.6. Technical Characteristics in Comparison to Predicate Devices

CUVIS-spine is substantially equivalent to the following legally marketed predicate devices

Proposed DevicePredicate Device
DeviceCUVIS-spineCUVIS-spine
(K201569)
Device description and indications for use
ManufacturerCUREXO, INCCUREXO, INC
ClassificationClass IIClass II
Product CodeOLOOLO
Regulation No.882.4560882.4560
General Device descriptionThe CUVIS-spine is a mobilesystem mainly comprising therobotic arm, the main console andthe staff console as an option. Therobotic arm is positioned on thefloor near the side of the surgicalThe CUVIS-spine is a mobilesystem mainly comprising therobotic arm, the main console andthe staff console as an option. Therobotic arm is positioned on thefloor near the side of the surgical
Proposed DevicePredicate Device
DeviceCUVIS-spineCUVIS-spine(K201569)
table. The location of the mainconsole or the staff console isappropriately determinedconsidering the user preferenceand the environments.table. The location of the mainconsole or the staff console isappropriately determinedconsidering the user preferenceand the environments
CUVIS-spine is intended for useas an aid for precisely locatinganatomical structures and for thespatial positioning and orientationCUVIS-spine is intended for useas an aid for precisely locatinganatomical structures and for thespatial positioning and orientation
of guide bush to be used bysurgeons for navigating and/orof guide bush to be used bysurgeons for navigating and/or
Indicationsfor useguiding compatible surgicalinstruments in open orguiding compatible surgicalinstruments in open or
percutaneous surgical proceduresif the required markers and rigidpercutaneous surgical proceduresif the required markers and rigid
patient anatomy can be identifiedon O-arm or C-arm.patient anatomy can be identifiedon O-arm or C-arm.
Technical Characteristics
- Intraoperative images- Intraoperative images
- Patient registration- Patient registration
Principle ofoperation- Surgical planning- Surgical planning
- Real-time tracking of navigatedinstruments- Real-time tracking ofnavigated instruments
- Guidance of instruments- Guidance of instruments
ImageO-arm, C-armO-arm, C-arm
Input Images- 3D intra-operative exam- 3D intra-operative exam
- 2D intra-operative exam- 2D intra-operative exam
TrackerOptical Tracking SystemOptical Tracking System
GuideDilator and TapperDilator and Tapper
TargetTrackingYESYES
SoftwareInformation- SRC (Control Software)- SPN (Planner Software)- SRC (Control Software)- SPN (Planner Software)
Save/loadPlanningYESYES
Merge imagesYESYES
Proposed DevicePredicate Device
DeviceCUVIS-spineCUVIS-spine(K201569)
functionality
Trajectory planningparameters- Entry point- Entry point
- Target point- Target point
- Instrument length/diameter- Instrument length/diameter
Localization meansOptical system (infrared camera)Optical system (infrared camera)
Image-guidedYESYES
ControllerForced-controlled movementForced-controlled movement
allowing robot arm positioningallowing robot arm positioning
(called hand guide function)(called hand guide function)
Patient registrationmethod- Intra-op CT : Registration tool- Fluoroscopy : Source calibrator- Intra-op CT : Registration tool- Fluoroscopy : Source calibrator
Real time display ofinstrument positionYESYES
Accessories- Registration instruments(Guide bush, Source calibrator,Registration tool, Registrationtool adapter, Detectorcalibrator)- Registration instruments(Guide bush, Source calibrator,Registration tool, Registrationtool adapter, Detectorcalibrator)
- Patient reference instruments(Patient marker, Clamp, Pin,Marker ball, Adapter, Pin Vise,Slide Hammer, Probe)- Patient reference instruments(Patient marker, Clamp,Marker ball)
- Surgical instruments(Marker Driver, Dilator,Serration-tip dilator, Bur, Stylettapper, Tapper, Screwdriver,Robotic arm drape, Tool drape,Detector drape, InstrumentContainer)- Surgical instruments(Marker Driver, Dilator,Serration-tip dilator, Bur,Stylet tapper, Tapper,Screwdriver, Robotic armdrape, Tool drape, Detectordrape, Instrument Container)
Electrical Safety andElectromagenticcompatibility- IEC60601-1- IEC60601-1-2- IEC60601-1- IEC60601-1-2
AccuracyYesYes
Proposed DevicePredicate Device
DeviceCUVIS-spineCUVIS-spine(K201569)
verification onanatomicallandmarks
Bio-compatibilityThe biocompatibility evaluationfor the patient marker has beenconducted in accordance with ENISO 10993 standards.The biocompatibility evaluationfor the patient marker has beenconducted in accordance with ENISO 10993 standards.
Parts ofContact withPatientPatient markerPatient marker, Clamp*
Nature ofBody ContactImplant device / boneImplant device / bone
Patient FixationReference is fixed to patient'sbony structure for tracking systemReference is fixed to patient'sbony structure for tracking system

{5}------------------------------------------------

CUVIS-spine K222698

{6}------------------------------------------------

CUVIS-spine K222696

{7}------------------------------------------------

*For Clamp corresponds to the Body of the Patient Marker of the initial approved device. The structure of Patient Marker, the initial approved device, of the Maker Frame and Body were disassembled, and the name was changed for each structure. This change was documented on August 14, 2021.

5.7. Performance Data

Non-clinical tests: Pose accuracy and Repeatability of the CUVIS-spine were tested and validated. A cadaveric study of the proposed device had been performed for the robotic-assisted pedicle screw placement.

Biocompatibility were tested using following consensus standards:

  • Tests for in vitro cytotoxicity were tested and evaluated according to the FDA-recognized consensus standard, ISO 10993-5.
  • -Tests for irritation, skin sensitization and intracutaneous reactivity were tested and evaluated according to the FDA-recognized consensus standard, ISO 10993-10.

Electromagnetic compatibility and electrical safety, etc, were tested using following consensus standards:

  • Basic safety and essential performance of the CUVIS-spine is tested and evaluated -

{8}------------------------------------------------

according to the FDA-recognized consensus standard, ES 60601-1.

  • Effect to the device by electromagnetic disturbances were tested and evaluated according to the FDA-recognized consensus standard IEC 60601-1-2.
  • Risk management was recorded by referring to ISO 14971. -
  • Usability was documented by referring to IEC 60601-1-6. -

5.8. Substantial Equivalence

CUVIS-spine is not based on a new technology. The predicate device is already available in the US market. There is a slight difference between CUVIS-spine and the predicate device. However, it does not affect the technical, clinical and biological aspects. It can be claimed that CUVIS-spine is substantially equivalent to the predicate device.

5.9. Conclusions

On the basis of the information provided in this Summary, CUREXO, INC. believes that CUVIS-spine is substantially equivalent to legally commercialized predicate device for the purposes of this 510 (k) submission.

§ 882.4560 Stereotaxic instrument.

(a)
Identification. A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.(b)
Classification. Class II (performance standards).