K Number
K222692

Validate with FDA (Live)

Device Name
BriefCase
Date Cleared
2022-12-05

(90 days)

Product Code
Regulation Number
892.2080
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

BriefCase is a radiological computer aided triage and notification software indicated for use in the analysis of chest and abdominal CT images. The device is intended to assist hospital networks and appropriately trained medical specialists within the standard-of-care bone health setting in workflow triage by flagging and communication of suspected positive cases of Vertebral Compression Fractures (VCFx) findings.

BriefCase uses an artificial intelligence algorithm to analyze images and highlight cases with detected findings on a standalone application in parallel to the ongoing standard of care image interpretation. The device does not alter the original medical image and is not intended to be used as a diagnosis device.

The results of BriefCase are intended to be used in conjunction with other patient information and based on their professional judgment, to assist with triage/prioritization of medical images. Notified clinicians are responsible for viewing full images per the standard of care.

Device Description

BriefCase is a radiological computer-assisted triage and notification software device. The software system is based on an algorithm programmed component and consists of a standard off-the-shelf operating system, the Microsoft Windows server 2012 64bit, and additional applications, which include PostgreSQL, DICOM module and the BriefCase Image Processing Application. The device consists of the following three modules: (1) Aidoc Hospital Server (AHS/Orchestrator) for image acquisition: (2) Aidoc Cloud Server (ACS) for image processing; and (3) Aidoc Desktop Application for workflow integration.

DICOM images are received, saved, filtered and de-identified before processing. Filtration matches metadata fields with keywords. Series are processed chronologically by running the algorithms on each series to detect suspected cases.

The desktop application feed displays all incoming suspect cases, each notified case in a line. Hovering over a line in the feed pops up a compressed, low-quality, grayscale, unannotated image that is captioned "not for diagnostic use", "for prioritization only" and is displayed as a preview function. This compressed preview is meant for informational purposes only, does not contain any marking of the findings, and is not intended for primary diagnosis beyond notification.

Presenting the users with worklist prioritization facilitates earlier triage by prompting the user to assess the relevant original images in the PACS. Thus, the suspect case receives attention earlier than would have been the case in the standard of care practice alone.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study proving the device meets those criteria, based on the provided FDA 510(k) summary for Aidoc Medical, Ltd.'s BriefCase:

1. Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Performance Goals)Reported Device Performance
Area Under the Curve (AUC) > 0.95AUC = 0.976 (95% CI: 0.961%, 0.991%)
Sensitivity > 80%Sensitivity = 95.11% (95% CI: 90.92%, 97.74%)
Specificity > 80%Specificity = 93.28% (95% CI: 87.63%, 96.88%)
Additional Metrics Reported:
Negative Predictive Value (NPV)99.5% (95% CI: 99.1%- 99.8%)
Positive Predictive Value (PPV)55.2% (95% CI: 39.6%)
Positive Likelihood Ratio (PLR)14.1606 (95% CI: 7.528-26.637)
Negative Likelihood Ratio (NLR)0.0524 (95% CI: 0.028- 0.099)
Time-to-Notification (Primary Goal: Comparable to Predicate)Mean = 117.2 seconds (95% CI: 98.64-135.85)

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: 318 cases
  • Data Provenance: Retrospective, from 5 US-based clinical sites. The data was distinct in time or center from the cases used to train the algorithm.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

  • Number of Experts: Three (3)
  • Qualifications: Senior board-certified radiologists.

4. Adjudication Method for the Test Set

The document explicitly states that the ground truth was "determined by three senior board-certified radiologists." While it doesn't detail the exact adjudication method (e.g., majority vote, consensus after discussion), the phrasing suggests a consensual or majority-based approach among the three experts. It is not explicitly stated as 2+1 or 3+1, but the involvement of three implies an adjudication process to establish the single ground truth for each case.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not conducted to specifically evaluate how much human readers improve with AI vs. without AI assistance. The study focused on the standalone performance of the AI algorithm and compared its time-to-notification to a predicate device.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

Yes, a standalone performance study of the algorithm (BriefCase software) was performed. The primary endpoints (AUC, Sensitivity, Specificity) were evaluated for the algorithm's performance in identifying Vertebral Compression Fractures.

7. The Type of Ground Truth Used

The ground truth used was expert consensus based on the determination by three senior board-certified radiologists.

8. The Sample Size for the Training Set

The sample size for the training set is not explicitly stated in the provided document. It only mentions that the test set cases were "distinct in time or center from the cases used to train the algorithm."

9. How the Ground Truth for the Training Set Was Established

The document does not provide details on how the ground truth for the training set was established. It only indicates that the training data was separate from the test data.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Aidoc Medical, Ltd. % John J. Smith, M.D., J.D. Partner Hogan Lovells U.S. LLP 555 Thirteenth Street NW WASHINGTON DC 20004

Re: K222692

December 5, 2022

Trade/Device Name: BriefCase Regulation Number: 21 CFR 892.2080 Regulation Name: Radiological computer aided triage and notification software Regulatory Class: Class II Product Code: QFM Dated: November 10, 2022 Received: November 10, 2022

Dear John Smith:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

{1}------------------------------------------------

https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jessica Lamb

Jessica Lamb, PhD Assistant Director Imaging Software Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known)

K222692

Device Name

BriefCase

Indications for Use (Describe)

BriefCase is a radiological computer aided triage and notification software indicated for use in the analysis of chest and abdominal CT images. The device is intended to assist hospital networks and appropriately trained medical specialists within the standard-of-care bone health setting in workflow triage by flagging and communication of suspected positive cases of Vertebral Compression Fractures (VCFx) findings.

BriefCase uses an artificial intelligence algorithm to analyze images and highlight cases with detected findings on a standalone application in parallel to the ongoing standard of care interpretation. The device does not alter the original medical image and is not intended to be used as a diagnosis device.

The results of BriefCase are intended to be used in conjunction with other patient information and based on their professional judgment, to assist with triage/prioritization of medical images. Notified clinicians are responsible for viewing full images per the standard of care.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including
the time to review instructions, search existing data sources, gather and maintain the data needed and
complete and review the collection of information. Send comments regarding this burden estimate or any
other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human

Services Food and Drug Administration

Office of Chief Information Officer

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Paperwork Reduction Act (PRA) Staff

PRAStaff@fda.hhs.qov

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Aidoc. The logo is in blue, with the letters "aidoc" in a sans-serif font. There is a small orange circle to the right of the "c" in "aidoc". The logo is simple and modern.

K222692

510(k) Summary Aidoc Medical, Ltd.'s BriefCase

Submitter:

Aidoc Medical, Ltd.3 Aminadav St.Tel-Aviv, IsraelPhone:+972-73-7946870
Contact Person:Amalia Schreier, LL.M.
Date Prepared:September 6, 2022
Name of Device:BriefCase
Classification Name:Radiological computer-assisted triage and notification softwaredevice
Requlatory Class:Class II
Product Code:QFM (21 C.F.R. 892.2080)
Primary Predicate Device:HealthVCF (K192901)

Device Description

BriefCase is a radiological computer-assisted triage and notification software device. The software system is based on an algorithm programmed component and consists of a standard off-the-shelf operating system, the Microsoft Windows server 2012 64bit, and additional applications, which include PostgreSQL, DICOM module and the BriefCase Image Processing Application. The device consists of the following three modules: (1) Aidoc Hospital Server (AHS/Orchestrator) for image acquisition: (2) Aidoc Cloud Server (ACS) for image processing; and (3) Aidoc Desktop Application for workflow integration.

DICOM images are received, saved, filtered and de-identified before processing. Filtration matches metadata fields with keywords. Series are processed chronologically by running the algorithms on each series to detect suspected cases.

The desktop application feed displays all incoming suspect cases, each notified case in a line. Hovering over a line in the feed pops up a compressed, low-quality, grayscale, unannotated image that is captioned "not for diagnostic use", "for prioritization only" and is displayed as a preview function. This compressed preview is meant for informational purposes only, does not contain any marking of the findings, and is not intended for primary diagnosis beyond notification.

Presenting the users with worklist prioritization facilitates earlier triage by prompting the user to assess the relevant original images in the PACS. Thus, the suspect case receives attention earlier than would have been the case in the standard of care practice alone.

{4}------------------------------------------------

Intended Use / Indications for Use

BriefCase is a radiological computer aided triage and notification software indicated for use in the analysis of chest and abdominal CT images. The device is intended to assist hospital networks and appropriately trained medical specialists within the standard-of-care bone health setting in workflow triage by flaqging and communication of suspected positive cases of Vertebral Compression Fractures (VCFx) findings.

BriefCase uses an artificial intelligence algorithm to analyze images and highlight cases with detected findings on a standalone application in parallel to the ongoing standard of care image interpretation. The device does not alter the original medical image and is not intended to be used as a diagnosis device.

The results of BriefCase are intended to be used in conjunction with other patient information and based on their professional judgment, to assist with triage/prioritization of medical images. Notified clinicians are responsible for viewing full images per the standard of care.

Comparison of Technological Characteristics

The subject BriefCase for Vertebral Compression Fractures (VCFx) triage and primary predicate HealthVCF triage (K192901) are very similar, as explained below.

Both devices are radiological computer-aided triage and notification software programs. Both devices are artificial intelligence, deep-learning algorithms incorporating software packages for use with PACS and radiology workstations. The predicate HealthVCF evaluates images from CT scanners as does the proposed Briefcase for Vertebral Compression Fractures (VCFx). The predicate and subject devices run on chest and abdomen CTs. The predicate and subject devices both present nondiagnostic slices preview images.

The proposed device for Vertebral Compression Fractures has similar technology and design as the predicate device, and similar indications for use, i.e., both devices are intended to aid in prioritization and triage of time-critical radiological images. The subject and predicate HealthVCF devices raise the same types of safety and effectiveness questions, namely, accurate detection of findings within the processed study. It is important to note that, like the predicate, the subject device neither removes cases from the standard of care reading queue nor deprioritizes cases. Both devices operate in parallel with the standard of care, which remains the default option for all cases. A table comparing the key features of the subject and the predicate devices is provided below.

{5}------------------------------------------------

Predicate Device HealthVCF(K192901)Subject DeviceAidoc Briefcase forVertebral CompressionFractures (VCFx) Triage
Intended Use / Indications forUseHealthVCF is a passivenotification for prioritization-only, parallel-workflow softwaretool used by clinicians toprioritize specific patientswithin the standard-of-carebone health setting forsuspected vertebralcompression fractures.HealthVCF uses an artificialintelligence algorithm toanalyze chest and abdominalCT scans and flags those thatare suggestive of the presenceof at least one vertebralcompression at the exam level.These flags are viewed by theclinician in Bone Health andFracture Liaison Serviceprograms in the medical settingvia a worklist application ontheir Picture Archiving andCommunication System(PACS). HealthVCF does notsend a proactive alert directlyto the user.BriefCase is a radiologicalcomputer aided triage andnotification software indicatedfor use in the analysis of chestand abdominal CT images. Thedevice is intended to assisthospital networks andappropriately trained medicalspecialists within the standard-of-care bone health setting inworkflow triage by flagging andcommunication of suspectedpositive cases of VertebralCompression Fractures (VCFx)findings.BriefCase uses an artificialintelligence algorithm toanalyze images and highlightcases with detected findings ona standalone application inparallel to the ongoing standardof care image interpretation.The device does not alter theoriginal medical image and isnot intended to be used as adiagnosis device.
HealthVCF does not providediagnostic information beyondtriage and prioritization, it doesnot remove cases from theradiology worklist, and shouldnot be used in place of fullpatient evaluation, or reliedupon to make or confirmdiagnosis.The results of BriefCase areintended to be used inconjunction with other patientinformation and based on theirprofessional judgment, toassist with triage/prioritizationof medical images. Notifiedclinicians are responsible forviewing full images per thestandard of care.
Predicate Device HealthVCF(K192901)Subject DeviceAidoc Briefcase forVertebral CompressionFractures (VCFx) Triage
User populationBone Health CliniciansHospital networks andappropriately trained medicalspecialists within the standard-of-care bone health setting.
Anatomical region of interestChest and abdomenChest and abdomen
Data acquisition protocolChest and abdominal CT scansChest and abdominal CT scans
Notification-only, parallelworkflow toolYesYes
Interference with standardworkflowNoNo
AlgorithmArtificial intelligence algorithmArtificial intelligence algorithm
StructureData input andvalidation (to ensurecompatibility forprocessing by thealgorithm)HealthVCF algorithm.Zebra Worklist.AHS module (orchestrator,image acquisition);ACS module (imageprocessing));Aidoc Desktop applicationfor workflow integration(feed and non-diagnosticImage Viewer)

Table 1. Key Feature Comparison

{6}------------------------------------------------

Performance Data

Pivotal Study Summary

Aidoc conducted a retrospective, blinded, multicenter, study with the BriefCase software to evaluate the software's performance in identifying Vertebral Compression Fractures in chest and abdominal CT images in 318 cases from 5 US-based clinical sites, compared to the ground truth as determined by three senior board-certified radiologists. The cases collected for the pivotal dataset were all distinct in time or center from the cases used to train the algorithm.

Primary endpoints were pre-specified standalone performance goal (PG) of area under the curve (AUC) > 0.95 for the study level receiver operating characteristic (ROC) curve. Secondary endpoints were BriefCase time-to-notification compared to the predicate device. Positive Value (PPV), Negative Predictive Value (NPV), Positive Likelihood Ratio (PLR), and Negative Likelihood Ratio (NLR) will also be assessed.

{7}------------------------------------------------

Primary Endpoint

AUC was 97.6% (95% Cl: 96.1%, 99.1%), Sensitivity was 95.11% (95% Cl: 90.92%, 97.74%) and Specificity was 93.28% (95% Cl: 87.63%, 96.88%). As the AUC exceeded 0.95 and sensitivity and specificity both exceeded 80%, the study's primary endpoints were met.

Lower confidence limits for AUC, sensitivity and specificity were all above the pre-specified performance goals, demonstrating that the pre-specified performance goals were met.

Secondary Endpoint

In addition, the time-to-notification metric observed for the BriefCase software in the five medical centers was compared to the equivalent metric of the predicate devices.

  • -The BriefCase time-to-notification includes the time to get the DICOM exam, de-identify it, upload it to the cloud, analyze and send a notification on a positive suspect case back to the desktop application.
    The BriefCase time-to-notification was measured for all True Positive cases (i.e., identified as positive both by the reviewers as well as the BriefCase device) and is given in Table 2 below. The table also displays the same metric reported for the predicate HealthVCF.

The time-to-notification results obtained for the subject BriefCase device show comparability with the primary predicate with regard to the standard of care review. The BriefCase mean time-to-notification for Vertebral Compression Fractures was 117.2 seconds (95% Cl: 98.64-135.85). The time-to-notification for the HealthVCF triage was 61.36 seconds.

Time -to-notificationNMean Estimate95% Lower CL95% Upper CLMedianIQR
Predicate K192901 Processing Timeunknown61.36unknownunknownunknownunknown
BriefCase Time-to-notification184117.298.64135.8571.599.5

Table 2. Time-to- Notification Comparison for BriefCase and Predicate Devices

NPV was 99.5% (95% CI: 99.1%- 99.8%) and PPV was 55.2% (95% CI: 39.6%).

PLR was 14.1606 (95% Cl: 7.528-26.637) and NLR was 0.0524 (95% Cl: 0.028- 0.099).

Thus, the reported similar time-to-notification data demonstrates that when using the subject BriefCase for Vertebral Compression Fractures triage the radiologists may have the same benefit in time saving as with the HealthVCF triage.

{8}------------------------------------------------

As can be seen in Table 3 the mean age of patients whose scans were reviewed for VCFx was 65.7 years, with a standard deviation of 16.9 years. Gender distribution was 46.5% male, and 53.5% female (Table 4). Scanner distribution can also be found in Table 5 below.

Table 3. Descriptive Statistics for Age

MeanStdMinMedianMaxN
Age (Years)65.716.9186890318

Table 4. Frequency Distribution of Gender

GroundTruthResultsGenderAll
MaleFemaleN%
N%N%
Positive9228.9%9228.9%18457.9%
Negative5617.6%7824.5%13442.1%
All14846.5%17053.5%318100.0%

Table 5. Frequency Distribution of Manufacturer

ManufacturerN%
GE MEDICALSYSTEMS11435.9%
SIEMENS9128.6%
TOSHIBA5717.9%
Philips5617.6%
Total318100%

{9}------------------------------------------------

Conclusions

The subject BriefCase for Vertebral Compression Fractures and the predicate HealthVCF triage are intended to aid in prioritization and triage of radiological images for the indications of vertebral compression fractures. Both devices are software packages with substantially equivalent technological characteristics and principles of operation, incorporating Al alqorithms that process images, and software to send notifications and display unannotated compressed low-quality preview images. In both devices, the labeling clearly states that the devices are not for diagnostic use and for prioritization only and instructs the user to further evaluate and diagnose based only on the original images in the local PACS.

Both devices operate in parallel to the standard of care workflow in the sense that they do not change the original image, do not provide any marking on the output preview, do not remove images from the standard of care FIFO queue and do not de-prioritize cases, thus not disturbing standard interpretation of the images. Both devices notify the radiologist of time-sensitive critical cases within the range of several minutes, and thus contribute similarly to the standard of care workflow turnaround time reduction through preemptive triage.

The subject BriefCase device for Vertebral Compression Fractures triage is thus substantially equivalent to the primary predicate HealthVCF triage.

§ 892.2080 Radiological computer aided triage and notification software.

(a)
Identification. Radiological computer aided triage and notification software is an image processing prescription device intended to aid in prioritization and triage of radiological medical images. The device notifies a designated list of clinicians of the availability of time sensitive radiological medical images for review based on computer aided image analysis of those images performed by the device. The device does not mark, highlight, or direct users' attention to a specific location in the original image. The device does not remove cases from a reading queue. The device operates in parallel with the standard of care, which remains the default option for all cases.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) A detailed description of the notification and triage algorithms and all underlying image analysis algorithms including, but not limited to, a detailed description of the algorithm inputs and outputs, each major component or block, how the algorithm affects or relates to clinical practice or patient care, and any algorithm limitations.
(ii) A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide effective triage (
e.g., improved time to review of prioritized images for pre-specified clinicians).(iii) Results from performance testing that demonstrate that the device will provide effective triage. The performance assessment must be based on an appropriate measure to estimate the clinical effectiveness. The test dataset must contain sufficient numbers of cases from important cohorts (
e.g., subsets defined by clinically relevant confounders, effect modifiers, associated diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals for these individual subsets can be characterized with the device for the intended use population and imaging equipment.(iv) Stand-alone performance testing protocols and results of the device.
(v) Appropriate software documentation (
e.g., device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results).(2) Labeling must include the following:
(i) A detailed description of the patient population for which the device is indicated for use;
(ii) A detailed description of the intended user and user training that addresses appropriate use protocols for the device;
(iii) Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (
e.g., poor image quality for certain subpopulations), as applicable;(iv) A detailed description of compatible imaging hardware, imaging protocols, and requirements for input images;
(v) Device operating instructions; and
(vi) A detailed summary of the performance testing, including: test methods, dataset characteristics, triage effectiveness (
e.g., improved time to review of prioritized images for pre-specified clinicians), diagnostic accuracy of algorithms informing triage decision, and results with associated statistical uncertainty (e.g., confidence intervals), including a summary of subanalyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.