(314 days)
Fibroid Mapping Reviewer Application (FMRA) is intended to be used by physicians in the clinic or hospital to generate a 3-D model from ultrasound images of the uterus of women with uterine fibroids. The model represents clinically relevant dimensions, including the location and dimensions of the fibroid (maximum length, width and depth).
Fibroid Mapping Reviewer Application (FMRA) is a software that allows the user to load an existing ultrasound data into the application for calibration, review, annotation and generation of a 3D rendered uterus model. It aids in the measurement of fibroid dimensions and mapping the location of fibroids in the uterus.
The ultrasound data of the patient's uterus, exported from ultrasound equipment via storage media such as a USB drive, is loaded into the proposed software. The user marks the anatomical features of the uterus and the fibroid. The software post-processes the 2D volume ultrasound images of the uterus to obtain the mapping & measurement of fibroids which are represented in a 3D rendered uterus model. It allows the clinician to know the parameters of each fibroid (like fibroid size and location) by just selecting a fibroid in the model.
The proposed device is intended to assist qualified physicians in accurate diagnosis and planning of image guided interventional of uterine fibroids.
Here's a breakdown of the acceptance criteria and study details for the Fibroid Mapping Reviewer Application (FMRA) based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria | Reported Device Performance (FMRA vs. MRI) |
|---|---|
| Location | Variation of spatial coordinate within +/- 2 mm |
| Dimensions | Variation of fibroid dimensions within +/- 2 mm |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: The document does not explicitly state the total number of cases or patients used in the test set. It mentions "phantom models (i.e., incorporating simulated uterus and fibroids)" and that "input images for FMRA were also collected from 3 different ultrasound machines." This implies the test data was generated using phantom models.
- Data Provenance: The test data was generated using phantom models (simulated uterus and fibroids). The images for FMRA input were collected from 3 different ultrasound machines.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
- Number of Experts: 3 certified medical professionals.
- Qualifications of Experts: They are referred to as "certified medical professionals." Specific qualifications (e.g., radiologist, years of experience) are not provided in this document.
4. Adjudication Method for the Test Set
The document does not explicitly state an adjudication method (like 2+1, 3+1). It states "Validation testing was executed by 3 certified medical professionals," implying that these professionals were involved in the validation process, likely comparing FMRA output to MRI values and assessing if the acceptance criteria were met. There's no mention of a consensus or tie-breaking mechanism.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance
- MRMC Study: No, a multi-reader multi-case (MRMC) comparative effectiveness study comparing human readers with and without AI assistance was not explicitly mentioned or described. The study focused on validating the FMRA's accuracy against MRI, not on human reader performance improvement.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
- Standalone Performance: Yes, the described validation is essentially a standalone (algorithm-only) performance study. The FMRA's output (spatial coordinates and fibroid dimensions) was directly compared against ground truth established by MRI values, with the "certified medical professionals" acting as evaluators of this comparison rather than integrating FMRA into a human workflow.
7. The Type of Ground Truth Used
- Ground Truth Type: MRI (Magnetic Resonance Imaging) values are considered the "gold standard for imaging" and were used as the ground truth.
8. The Sample Size for the Training Set
The document does not provide information regarding the sample size used for the training set. It focuses on the validation of the device after its development.
9. How the Ground Truth for the Training Set Was Established
The document does not provide information on how the ground truth for the training set was established. It only describes the ground truth for the validation (test) set.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a larger font size than the rest of the name. The full name of the agency is "U.S. Food & Drug Administration".
Nesa Medtech Private Limited % Rob Packard President/Consultant Medical Device Academy, Inc. 345 Lincoln Hill Road SHREWSBURY VT 05738
July 17, 2023
Re: K222683
Trade/Device Name: Fibroid Mapping Reviewer Application (FMRA) Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: LLZ Dated: September 6, 2022 Received: June 9, 2023
Dear Rob Packard:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
{1}------------------------------------------------
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica Lamb
Jessica Lamb. Ph.D. Assistant Director Imaging Software Team DHT8B: Division of Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K222683
Device Name
Fibroid Mapping Reviewer Application (FMRA)
Indications for Use (Describe)
Fibroid Mapping Reviewer Application (FMRA) is intended to be used by physicians in the clinic or hospital to generate a 3-D model from ultrasound images of the uterus of women with uterine fibroids. The model represents clinically relevant dimensions, including the location and dimensions of the fibroid (maximum length, width and depth).
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | Over The Counter Use (21 CFR 801 Subpart C) |
|---|---|
| ---------------------------------------------- | --------------------------------------------- |
X Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) SUMMARY
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR $807.92:
I. SUBMITTER Nesa Medtech Private Limited 15, 33rd Main 7th Block Bhavani Colony, Banashankari 3rd Stage,
Banashankari, Bengaluru, Karnataka 560085, India
Contact Person: Sreekar Kothamachu Phone Number: +91 9900166988 Email: sreekar@nesamedtech.com
Date Prepared: October 25, 2022
| II.DEVICE | |
|---|---|
| Name of Device: | Fibroid Mapping Reviewer Application (FMRA) |
| Classification Name: | Medical Image Management and Processing System |
| Regulation: | 21 CFR §892.2050 |
| Regulatory Class: | Class II |
| Product Classification Code: | LLZ |
III. PREDICATE DEVICE
| Predicate Manufacturer: | Eigen LLC |
|---|---|
| Predicate Trade Name: | 3-D Imaging Workstation |
| Predicate 510(k): | K081093 |
IV. DEVICE DESCRIPTION
Fibroid Mapping Reviewer Application (FMRA) is a software that allows the user to load an existing ultrasound data into the application for calibration, review, annotation and generation of a 3D rendered uterus model. It aids in the measurement of fibroid dimensions and mapping the location of fibroids in the uterus.
The ultrasound data of the patient's uterus, exported from ultrasound equipment via storage media such as a USB drive, is loaded into the proposed software. The user marks the anatomical features of the uterus and the fibroid. The software post-processes the 2D volume ultrasound images of the uterus to obtain the mapping & measurement of fibroids which are represented in a 3D rendered uterus model. It allows the clinician to know the parameters of each fibroid (like fibroid size and location) by just selecting a fibroid in the model.
The proposed device is intended to assist qualified physicians in accurate diagnosis and planning of image guided interventional of uterine fibroids.
V. INDICATIONS FOR USE
Fibroid Mapping Reviewer Application (FMRA) is intended to be used by physicians in the clinic or hospital to generate a 3-D model from ultrasound images of the uterus of women with uterine fibroids. The model represents clinically relevant dimensions, including the location of the fibroid (maximum length, width and depth).
{4}------------------------------------------------
COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE VI.
The following characteristics were compared between the subject device in order to demonstrate substantial equivalence:
| Feature | Subject Device | Predicate Device | Comparison |
|---|---|---|---|
| Name | Fibroid Mapping Reviewer Application(FMRA) | 3-D Imaging Workstation | N/A |
| 510(k) | K081093 | N/A | |
| Indications forUse | Fibroid Mapping Reviewer Application(FMRA) is intended to be used byphysicians in the clinic or hospital togenerate a 3-D model from ultrasoundimages of the uterus of women withuterine fibroids. The model representsclinically relevant dimensions, includingthe location and dimensions of thefibroid (maximum length, width anddepth). | The 3-D Imaging Workstation isintended to be used by physicians inthe clinic or hospital for 2-D and 3-Dvisualization of ultrasound images ofthe prostate gland. Additional softwarefeatures include patient datamanagement, multi-planarreconstruction, segmentation, imagemeasurement and 3-D imageregistration. | The subject devices are indicated fordifferent targeted anatomies. Theintended use of visualizingultrasound images is the same. |
| Product Code | LLZ | LLZ | Same |
| Class | II | II | Same |
| Windows OS | Windows OS (10) | Windows OS | Same |
| Medical imagingsoftware | Yes | Yes | Same |
| General Image2DReview | 2D ultrasound frames can be reviewed | 2D ultrasound frames can be reviewed | Same |
| 3D RenderingView | 2D ultrasound data is converted into 3Drendered model | 2D ultrasound data is converted into3D rendered model | Same |
| Import/Export | Exported from commercially available ultrasound machines via storage media such as a USB drive | Live video received from commercially available ultrasound machines | Different. The difference in the method of import/export of data does not raise questions of safety and effectiveness. |
| Review Tools | Yes | Yes | Same |
| Measurement Tools | Yes | Yes | Same |
| Annotation Tools | Yes | Yes | Same |
| Patient data management | Yes | Yes | Same |
| Image Storage and Communication | .mov, .avi, .mp4 | DICOM, jpeg | Different. The difference in the format of data does not raise questions of safety and effectiveness. |
{5}------------------------------------------------
{6}------------------------------------------------
VII. PERFORMANCE DATA
The overall verification & validation activities performed for FMRA includes the following:
- Product & Engineering specifications were verified & validated. ●
- Validation of FMRA included a comparison of the accuracy of FMRA output parameters on phantom models (i.e., incorporating simulated uterus and fibroids) with MRI values.
- o MRI Imaging is considered to be the gold standard for imaging.
- Acceptance criteria between FMRA Output parameters & MRI Values - O
- . Location - Variation of spatial co-ordinate should be within +/- 2 MM
- . Dimensions - Variation of fibroid dimensions should be within +/- 2 MM
- o Validation testing was executed by 3 certified medical professionals, and the input images for FMRA were also collected from 3 different ultrasound machines.
The following standards were followed in the development of the FMRA:
- ISO 14971: 2019 Medical Devices - Application of Risk Management to Medical Devices
- IEC 62304: 2015 Medical Device Software - Software Life Cycle Processes
FMRA passed all applicable testing in accordance with acceptance criteria and also relevant standards. The results demonstrate that the FMRA satisfies the performance, functional, and safety requirements relative to the product specifications, risk analyses, and Instructions for Use, and does not raise different questions of safety and effectiveness than the predicate device.
VIII. CONCLUSIONS
The Fibroid Mapping Reviewer Application i.e. FMRA has the same intended use as the predicate devices. Intended use of both the systems are to process the targeted features/ anatomies in 2-D ultrasound images deriving 3D representative models. The subject device is using this technology to visualize and measure fibroid dimension and location relative to other anatomy. The predicate device uses this technology to visualize the prostate gland. For both, ultrasound is the traditional method for evaluating the respective indicated uses, and both the subject device and predicate devices are software applications that improve the provider's ability to use the ultrasound information. Performance tests demonstrate that the FMRA has been tested to prove the same. The biological sex of the patient does not impact the performance of the ultrasound relating to its ability to capture images of targeted tissues and therefore does not impact the equivalence of the technology's performance in the respective sex. The method of acquisition (transvaginal vs. transrectal) is also not a substantial difference because the method and means of acquiring the image is not important, but what is important for each system is that the ultrasound views captured are within the required specifications for each software. This relates more to the specific indications as opposed to the general intended use
N/A