K Number
K222480

Validate with FDA (Live)

Date Cleared
2022-11-14

(89 days)

Product Code
Regulation Number
878.4040
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Standard Surgical Gown (AAMI Level 2) and Reinforced Surgical Gown (AAMI Level 2) are sterile, single use surgical apparel intended to be worn by healthcare professionals to help protect both the patient and the healthcare worker from the transfer of microorganisms, body fluids, and particulate material.

The Standard Surgical Gown (AAMI Level 2) and Reinforced Surgical Gown (AAMI Level 2) meets the level 2 requirements of ANSI/AAMI PB70:2012 Liquid barrier performance and classification of protective apparel and drapes intended for use in healthcare facilities.

The Standard Surgical Gown (AAMI Level 2) and Reinforced Surgical Gown (AAMI Level 2) have been validated using an ethylene oxide sterilization process. The Standard Surgical Gown (AAMI Level 2) and Reinforced Surgical Gown (AAMI Level 2) are also sold as bulk single-use, non-sterile, to repackager/relabeler establishments for further packaging and sterilization using the validated EtO sterilization method according to ISO 11135-1 prior to being provided to the end user.

Device Description

Not Found

AI/ML Overview

This document is a 510(k) summary for a medical device, specifically a Standard Surgical Gown (AAMI Level 2) and Reinforced Surgical Gown (AAMI Level 2). It outlines the FDA's determination of substantial equivalence. However, it does not contain information related to software algorithm performance, clinical study acceptance criteria, or the details of a study proving a device meets such criteria.

The document primarily focuses on:

  • Regulatory Clearance: The FDA's decision that the device is substantially equivalent to legally marketed predicate devices.
  • Indications for Use: The intended purpose of the surgical gowns (protection from microorganisms, body fluids, particulate matter) and their compliance with ANSI/AAMI PB70:2012 for liquid barrier performance (Level 2).
  • Sterilization Method: Validation of ethylene oxide (EtO) sterilization.
  • Device Models: A list of different models based on size, sterility, and whether they include hand towels and wrap.

Therefore, I cannot provide the requested information about acceptance criteria and study details from the provided text, as this document is a regulatory approval letter, not a performance study report.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

November 14, 2022

Foshan Nanhai Plus Medical Co., Ltd. % Olivia Meng Regulatory Affairs Manager Guangzhou Osmunda Medical Device Technical Service Co., Ltd. 14F, Building C, Ancillary Project of Phase IV Standard Industrial Park Guangzhou International Bio-Island Guangzhou, Guangdong 510320 China

Re: K222480

Trade/Device Name: Standard Surgical Gown (AAMI Level 2): Reinforced Surgical Gown (AAMI Level 2) Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FYA Dated: August 17, 2022 Received: August 17, 2022

Dear Olivia Meng:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of

{1}------------------------------------------------

Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Bifeng Qian -S

Bifeng Qian, M.D., Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

Submission Number (if known)

K222480

Device Name

Standard Surgical Gown (AAMI Level 2); Reinforced Surgical Gown (AAMI Level 2)

Indications for Use (Describe)

The Standard Surgical Gown (AAMI Level 2) and Reinforced Surgical Gown (AAMI Level 2) are sterile, single use surgical apparel intended to be worn by healthcare professionals to help protect both the patient and the healthcare worker from the transfer of microorganisms, body fluids, and particulate material.

The Standard Surgical Gown (AAMI Level 2) and Reinforced Surgical Gown (AAMI Level 2) meets the level 2 requirements of ANSI/AAMI PB70:2012 Liquid barrier performance and classification of protective apparel and drapes intended for use in healthcare facilities.

The Standard Surgical Gown (AAMI Level 2) and Reinforced Surgical Gown (AAMI Level 2) have been validated using an ethylene oxide sterilization process. The Standard Surgical Gown (AAMI Level 2) and Reinforced Surgical Gown (AAMI Level 2) are also sold as bulk single-use, non-sterile, to repackager/relabeler establishments for further packaging and sterilization using the validated EtO sterilization method according to ISO 11135-1 prior to being provided to the end user.

ile

Models without Hand Towels and Wrap:
Standard Surgical Gown (AAMI Level 2)Reinforced Surgical Gown (AAMI Level 2)
PM3501SGN M Non-sterilePM3501RGN M Non-sterile
PM3502SGN L Non-sterilePM3502RGN L Non-sterile
PM3503SGN XL Non-sterilePM3503RGN XL Non-sterile
PM3504SGN 2XL Non-sterilePM3504RGN 2XL Non-sterile
PM3505SGN 3XL Non-sterilePM3505RGN 3XL Non-sterile
PM3501SGS M SterilePM3501RGS M Sterile
PM3502SGS L SterilePM3502RGS L Sterile
PM3503SGS XL SterilePM3503RGS XL Sterile
PM3504SGS 2XL SterilePM3504RGS 2XL Sterile
PM3505SGS 3XL SterilePM3505RGS 3XL Sterile
Models with Hand Towels and Wrap:
Standard Surgical Gown (AAMI Level 2)Reinforced Surgical Gown (AAMI Level 2)
PM3511SGN M Non-sterilePM3511RGN M Non-sterile
PM3512SGN L Non-sterilePM3512RGN L Non-sterile
1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 111 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1I will a wall would with them it it was a firms of the1991 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 -
PM3513SGNXLNon-sterilePM3513RGNXLNon-ster
PM3514SGN2XLNon-sterilePM3514RGN2XLNon-ster
PM3515SGN3XLNon-sterilePM3515RGN3XLNon-ster
PM3511SGSMSterilePM3511RGSMSterile
PM3512SGSSterilePM3512RGSSterile
PM3513SGSXLSterilePM3513RGSXLSterile
PM3514SGS2XLSterilePM3514RGS2XLSterile
PM3515SGS3XISterilePM3515RGS3XISterile

{3}------------------------------------------------

] Prescription Use (Part 21 CFR 801 Subpart D)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid 0MB number."

§ 878.4040 Surgical apparel.

(a)
Identification. Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.(b)
Classification. (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.