K Number
K222440
Manufacturer
Date Cleared
2022-09-09

(28 days)

Product Code
Regulation Number
880.6850
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Vis-U-All Low Temperature Sterilization Pouches/Tubing are sterilization containment pouches for use by health care providers to enclose:

  • medical devices in a single or double pouch configuration
  • trays* containing medical devices in a single or double pouch configuration
  • small items requiring surface sterilization in a single pouch configuration within a tray*
    NOTE: Trays must be legally marketed for use in the V PRO Low Temperature or STERRAD Sterilization Systems and contain a vent surface area to tray volume ratio ≥ 0.135 in-1 with the maximum number of instrument organizers installed.

to be sterilized in the:

  • Lumen. Non Lumen. Flexible. Fast Non Lumen and Fast Cvcles of the V-PRO® Low Temperature Sterilization Systems
  • Default Cycle of the STERRAD 100S Sterilizer
  • Standard and Advanced Cycles of the STERRAD NX and NX with ALLClear Technology Sterilizers
  • Express, Standard, Flex Scope and DUO Cycles of the STERRAD 100NX with ALLClear Technology Sterilizers

*STERRAD and ALLClear are trademarks of Advanced Sterilization Products

The pouches maintain the sterility of the enclosed devices until used.

When used to enclose medical devices, the pouches are intended to contain the devices in such a manner as to leave a minimum of one inch between the devices and seal on all sides. When used to enclose a tray, the tray must fit loosely within the pouch.

Intended Sterilization Cycles and Intended Pouch Loads when Medical Devices are:

  • Directly pouched
  • Placed inside of a tray and the tray pouched

The PRO-LITE Sterilization Trays are used to contain medical devices for sterilization in the following Cycles:

  • Lumen, Non Lumen, Flexible, Fast Non Lumen and Fast Cycles of the V-PRO Low Temperature Sterilization Systems
  • Default Cycle of the STERRAD®* 100S Sterilizer
  • Standard and Advanced Cycles of the STERRAD NX and NX with ALLClear Technology Sterilizers
  • Standard, Flex Scope, Express and DUO Cycles of the STERRAD 100NX with ALLClear Technology Sterilizers

*STERRAD and ALLClear are trademarks of Advanced Sterilization Products

Prior to placing in the Sterilizer, the trays must either be:

  • wrapped with a legally marketed sterilization wrap for use in the Sterilizers listed above
    Or
  • placed inside a legally marketed pouch for enclosing trays in the Sterilizers listed above.

The PRO-LITE Sterilization Trays are not intended to maintain sterility; they are intended to be used in conjunction with a validated, FDA-cleared sterilization wrap or pouch in order to maintain sterility of the enclosed medical instruments.

Device Description

The proposed Vis-U-All Low Temperature Sterilization Pouches/Tubing is identical to the predicate and is a Tyvek/plastic film sterilization containment pouch designed for devices to be sterilized by the health care provider in V-PRO or STERRAD Low Temperature Sterilization Systems. As is the predicate device, the proposed device is available as a self-seal pouch, a heat-seal pouch, or heat-seal tubing.

The PRO-LITE Sterilization Trays contain medical devices for sterilization in the V-PRO Low Temperature Sterilization Systems and STERRAD Sterilizers. The trays are available in various sizes to accommodate the loads to be processed. The proposed trays are identical in design to the predicate Sterilization Tray (K183402) and are composed of a base and a lid. The lid includes clamping mechanisms designed to secure the lid onto the base. There are numerous holes in the base and lid for sterilant penetration. The tray is categorized as a cassette and requires complete enclosure in a legally-marketed sterilization wrap or pouch to maintain sterility of the enclosed devices. Both the base and the lid for the proposed tray are made of a mineral-filled polypropylene material. Optional instrument organizers and sterilization mats are provided as accessories.

AI/ML Overview

This document, a 510(k) Premarket Notification from the FDA, describes the acceptance criteria and the study proving the device meets these criteria for two products: "Vis-U-All Low Temperature Sterilization Pouches/Tubing" and "PRO-LITE Sterilization Trays."

Since the request is about "acceptance criteria and the study that proves the device meets the acceptance criteria" specifically for a device (implying a medical device with an associated study), and the document describes the performance testing for these sterilization products, I will extract and synthesize the information relevant to these performance tests as evidence of meeting acceptance criteria.

Note: The provided document is a 510(k) summary for medical devices (sterilization pouches and trays), not an AI/ML powered device. Therefore, typical AI/ML study components like "number of experts used to establish ground truth," "adjudication method," "MRMC study," "effect size of human readers improving with AI," "standalone performance," "training set size," and "ground truth for training set" are not applicable to this type of submission. The document focuses on demonstrating substantial equivalence through non-clinical performance testing (e.g., sterilization efficacy).

Here's an analysis of the acceptance criteria and proof, as presented for these sterilization devices:

1. A table of acceptance criteria and the reported device performance

For both devices, the acceptance criteria implicitly align with demonstrating "Sterile efficacy" and the "ability to sterilize medical devices" under various conditions and cycles. The performance is reported as meeting these implicit criteria, indicated by "PASS" results for all tests.

For "Vis-U-All Low Temperature Sterilization Pouches/Tubing" (Table 3, Page 18):

Acceptance Criteria (Implicit)Reported Device Performance (Result)Conclusion
Sterile Efficacy for Mated Surfaces under Worst-Case Conditions in V-PRO Sterilizer CyclesSterile efficacy was demonstrated for mated surfaces under worst case conditions in the V-PRO Sterilizer cycles.PASS
Sterile Efficacy with Standard Injection WeightThe standard injection weight resulted in all sterile results within the validation load used to qualify each sterilizer cycle.PASS
Ability to Sterilize Medical Devices under Worst-Case Processing Conditions (Simulated Use)Simulated use testing verified the ability of the sterilizer cycles to sterilize medical devices under worst-case processing conditions.PASS
Ability to Sterilize Patient-Soiled, Clinically-Cleaned Medical Instruments (In Use)The in use investigation demonstrated the ability of the V-PRO Sterilizer cycles to sterilize patient-soiled, clinically-cleaned, medical instruments.PASS

For "PRO-LITE Sterilization Tray" (Table 8, Page 29-30):

Acceptance Criteria (Implicit)Reported Device Performance (Result)Conclusion
Sterile Efficacy for Mated Surfaces under Worst-Case Conditions in V-PRO Sterilizer CyclesSterile efficacy was demonstrated for mated surfaces under worst case conditions in the V-PRO Sterilizer cycles.PASS
Sterile Efficacy with Standard Injection WeightThe standard injection weight resulted in all sterile results within the validation load used to qualify each sterilizer cycle.PASS
Ability to Sterilize Medical Devices under Worst-Case Processing Conditions (Simulated Use)Simulated use testing verified the ability of the sterilizer cycles to sterilize medical devices under worst-case processing conditions.PASS
Ability to Sterilize Patient-Soiled, Clinically-Cleaned Medical Instruments (In Use)The in use investigation demonstrated the ability of the V-PRO Sterilizer cycles to sterilize patient-soiled, clinically-cleaned, medical instruments.PASS

2. Sample sized used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

The document does not explicitly state the sample sizes (e.g., number of pouches/trays, number of sterilization cycles) used for each test. It refers to "worst-case conditions" and "validation load." Similarly, data provenance (country of origin, retrospective/prospective) is not specified, but the testing would typically be prospective and conducted in a controlled laboratory or simulated clinical environment.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

This is not applicable to the type of device and study presented. Ground truth for sterilization efficacy is typically established through scientific methods like sterility testing (e.g., microbial challenge tests, biological indicators) and physical/chemical monitoring, not human expert interpretation of images or data.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

Not applicable, as this is related to human interpretation of data, which is not the primary method for evaluating sterilization efficacy.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. These are physical devices for sterilization, not AI-powered diagnostic tools.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This concept pertains to AI algorithms, not sterilization instruments.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The ground truth for sterilization efficacy is established through direct measurements of sterility (e.g., absence of viable microorganisms after sterilization) and functional performance under simulated and actual use conditions as described in the summary:

  • Sterile efficacy: Demonstrated through microbial challenge and physical/chemical indicators.
  • Simulated Use Test: Verifies the ability of the sterilizer cycles to sterilize medical devices under worst-case processing conditions.
  • In Use Test: Demonstrates the ability to sterilize patient-soiled, clinically-cleaned medical instruments.

This involves direct, objective testing for sterility, rather than interpretation by human experts.

8. The sample size for the training set

Not applicable. These devices do not use "training sets" as they are not AI/ML models.

9. How the ground truth for the training set was established

Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services - USA. To the right of this symbol, there is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".

September 9, 2022

Steris Anthony Piotrkowski Director Regulatory Affairs 5960 Heisey Road Mentor, Ohio 44060

Re: K222440

Trade/Device Name: Vis-U-All Low Temperature Sterilization Pouches / Tubing, PRO-LITE Sterilization Trays Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: FRG, KCT Dated: August 11, 2022 Received: August 12, 2022

Dear Anthony Piotrkowski:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K222440

Device Name

Vis-U-All Low Temperature Sterilization Pouches/Tubing

Indications for Use (Describe)

The Vis-U-All Low Temperature Sterilization Pouches/Tubing are sterilization containment pouches for use by health care providers to enclose:

  • · medical devices in a single or double pouch configuration
  • · trays* containing medical devices in a single or double pouch configuration
  • · small items requiring surface sterilization in a single pouch configuration within a tray*

NOTE: Trays must be legally marketed for use in the V PRO Low Temperature or STERRAD Sterilization Systems and contain a vent surface area to tray volume ratio ≥ 0.135 in-1 with the maximum number of instrument organizers installed.

to be sterilized in the:

  • · Lumen. Non Lumen. Flexible. Fast Non Lumen and Fast Cvcles of the V-PRO® Low Temperature Sterilization Systems
  • · Default Cycle of the STERRAD 100S Sterilizer
  • Standard and Advanced Cycles of the STERRAD NX and NX with ALLClear Technology Sterilizers

· Express, Standard, Flex Scope and DUO Cycles of the STERRAD 100NX with ALLClear Technology Sterilizers

*STERRAD and ALLClear are trademarks of Advanced Sterilization Products

The pouches maintain the sterility of the enclosed devices until used.

When used to enclose medical devices, the pouches are intended to contain the devices in such a manner as to leave a minimum of one inch between the devices and seal on all sides. When used to enclose a tray, the tray must fit loosely within the pouch.

Intended Sterilization Cycles and Intended Pouch Loads when Medical Devices are:

  • Directly pouched
  • · Placed inside of a tray and the tray pouched

V-PRO 60 & s2 Lumen Cycle

· Non-lumened devices with diffusion-restricted spaces such as the hinged portion of forceps and scissors.

  • · Non-lumened devices including non-lumened rigid and semi-rigid endoscopes
  • · Medical devices, including single, dual and triple channeled rigid endoscopes, with the following configurations:
  • o single or dual lumen devices
  • ≥ 0.77 mm internal diameter (ID) and < 410 mm in length
  • ≥ 1.8 mm ID x ≤ 542 mm in length
  • o triple lumen devices
  • ≥1.2 mm ID and ≤ 275 mm in length
  • ≥1.8 mm ID and ≤ 310 mm in length

Or

≥2.8 mm ID and ≤ 317 mm in length

V-PRO 60 & s2 Non Lumen Cycle Non-lumened devices including non-lumened rigid, semi-rigid and flexible endoscopes and non-lumened devices with

{3}------------------------------------------------

diffusion-restricted spaces such as the hinged portion of forceps and scissors.

V-PRO 60 & s2 Flexible Cycle

Load 1: One flexible surgical endoscope with a light cord (if not integral to endoscope) and mat without any additional load. The flexible endoscope may be a:

· single or dual lumen device with lumens that are ≥ 1 mm ID and < 990 mm in length

Load 2: Non-lumened devices including non-lumened rigid, and flexible endoscopes and non-lumened devices with diffusion-restricted areas such as the hinged portion of forceps or scissors. Medical devices, including rigid and semirigid endoscopes, with the following configurations:

≥ 0.76 mm ID and ≤ 233 mm in length

≥ 1.0 mm ID and ≤ 254 mm in length

≥ 1.8 mm ID and ≤ 542 mm in length

V-PRO s2 Fast Cycle

· Non-lumened devices including non-lumened rigid and flexible endoscopes and non-lumened devices with diffusion-restricted spaces such as the hinged portion of forceps and scissors.

· Medical devices (including single, dual and triple channeled rigid endoscopes) with the following configurations:

o single or dual lumen devices

≥ 0.77 mm ID and ≤ 410 mm in length

≥ 1.8 mm ID x ≤ 542 mm in length

· Triple channeled devices with stainless steel lumens that are either:

≥1.2 mm ID and ≤ 275 mm in length

≥1.8 mm ID and < 310 mm in length

Or

≥2.8 mm ID and ≤ 317 mm in length

V-PRO 1, 1 Plus, maX & maX 2 Lumen Cycle

· Non-lumened devices with diffusion-restricted spaces such as the hinged portion of forceps and scissors.

· Non-lumened devices including non-lumened rigid and semi-rigid endoscopes

· Medical devices, including single, dual or triple channeled stainless steel lumens that are:

≥ 0.77 mm ID and ≤ 527 mm in length

≥ 0.8 mm ID and ≤ 542 mm in length

≥ 0.48 mm ID and < 100 mm in length

• Medical devices with Dead end lumens that are ≥ 1.3 mm ID and ≤ 73 mm in length

· Devices with rigid non-metallic lumens (such as those used in endoscope sheaths, take-apart forceps and trocars) that are:

≥ 3 mm ID and ≤ 298 mm in length

4 mm ID and < 424 mm in length

V-PRO 1, 1 Plus, maX & maX2 Non Lumen Cycle

Non-lumened devices including non-lumened rigid, semi-rigid and flexible endossopes and non-lumened devices with diffusion-restricted spaces such as the hinged portion of forceps and scissors.

V-PRO maX and maX 2 Flexible Cycle

Load 1: Single or dual lumen surgical flexible endoscopes (such as those used in ENT, Urology and Surgical Care) and bronchoscopes with a light cord (if not integral to endoscope) and mat with no additional load.

The flexible endoscopes may contain either a single or dual channel lumen that is > 1 mm ID and < 1050 mm in length Load 2:

• Non-lumened devices including non-lumened rigid, semi-rigid and flexible endoscopes and non-lumened devices with diffusion-restricted spaces such as the hinged portion of forceps and scissors

· Single, dual or triple channel stainless steel lumen that is ≥ 0.48 mm ID and ≤ 100 mm in length.

V-PRO maX 2 Fast Non Lumen Cycle

{4}------------------------------------------------

Non-lumened devices including non-lumened rigid, semi-rigid and flexible endoscopes and non-lumened devices with diffusion-restricted spaces such as the hinged portion of forceps and scissors.

STERRAD 100S Default Cycle

Metal and nonmetal medical devices including instruments which have diffusion-restricted spaces, such as the hinged portion of forceps and scissors.

Metal and nonmetal lumened instruments with:

  • ≥ 6 mm ID and ≤ 310 mm in length
    Medical devices with a single stainless steel lumen with:

≥ 1 mm ID and ≤ 125 mm in length

≥ 2 mm ID and ≤ 250 mm in length

≥ 3 mm ID and ≤ 400 mm in length

STERRAD NX and NX with ALLClear Technology Standard Cycle

Metal and nonmetal medical devices including instruments which have diffusion-restricted spaces, such as the hinged portion of forceps and scissors.

Medical devices with a single stainless steel lumen with:

≥ 1 mm ID and ≤ 150 mm in length

≥2 mm ID and < 400 mm in length

STERRAD NX and NX with ALLClear Technology Advanced Cycle

Metal and nonmetal medical devices including instruments which have diffusion-restricted spaces, such as the hinged portion of forceps and scissors.

Medical Devices, including most flexible endoscopes, with:

a single stainless steel lumen with:

≥ 1 mm ID and ≤ 500 mm in length

Single channel polyethylene and Teflon (polytetrafluoroethylene)

≥1 mm ID and ≤ 850 mm in length

STERRAD 100NX and 100NX with ALLClear Technology Standard Cycle

Metal and nonmetal medical devices including instruments which have diffusion-restricted spaces, such as the hinged portion of forceps and scissors.

Medical devices with a single stainless steel lumen with:

0.7 mm ID and < 500 mm in length

STERRAD 100NX and 100NX with ALLClear Technology Flex Scope Cycle

Metal and nonmetal medical devices including instruments which have diffusion-restricted spaces, such as the hinged portion of forceps and scissors.

Medical Devices, including most flexible endoscopes, with:

□ Single channel polyethylene and Teflon (polytetrafluoroethylene)

≥ 1 mm ID and ≤ 850 mm in length

STERRAD 100NX and 100NX with ALLClear Technology Express Cycle

Metal and nonmetal medical devices (surfaces sterilization only) and instruments which have diffusionrestricted spaces, such as the hinged portion of forceps and scissors.

STERRAD 100NX and 100NX with ALLClear Technology Duo Cycle

Medical devices including:

most flexible endoscopes with a single channel of polyethylene and Teflon (polytetrafluoroethylene) with > 1 mm ID and < 875 mm in length

□ accessory devices that are normally connected to a flexible endoscope during use

□ flexible endoscopes without lumens

Type of Use (Select one or both, as applicable)

] Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

{5}------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{6}------------------------------------------------

Indications for Use

510(k) Number (if known) K222440

Device Name PRO-LITE Sterilization Tray

  • The PRO-LITE Sterilization Trays are used to contain medical devices for sterilization in the following Cycles:
  • · Lumen, Non Lumen, Flexible, Fast Non Lumen and Fast Cycles of the V-PRO Low Temperature Sterilization Systems
  • · Default Cycle of the STERRAD®* 100S Sterilizer
  • · Standard and Advanced Cycles of the STERRAD NX and NX with ALLClear Technology Sterilizers

• Standard, Flex Scope, Express and DUO Cycles of the STERRAD 100NX with ALLClear Technology Sterilizers

*STERRAD and ALLClear are trademarks of Advanced Sterilization Products

Prior to placing in the Sterilizer, the trays must either be:

· wrapped with a legally marketed sterilization wrap for use in the Sterilizers listed above

Or

· placed inside a legally marketed pouch for enclosing trays in the Sterilizers listed above.

The PRO-LITE Sterilization Trays are not intended to maintain sterility; they are intended to be used in conjunction with a validated, FDA-cleared sterilization wrap or pouch in order to maintain sterility of the enclosed medical instruments.

Intended Sterilization Cycles and Intended Tray Load for Tray Models: VP0045, VP0047, VP0048, VP0048, VP0048, VP0050. VP0051. VP0052

V-PRO 60 and s2 Lumen Cycle:

  • · Non-lumened devices with diffusion-restricted spaces such as the hinged portion of forceps and scissors
  • · Non-lumened devices including non-lumened rigid and semi-rigid endoscopes

· Medical devices, including single, dual and triple channeled rigid endoscopes with the following configurations:

o Single or dual channeled devices with stainless steel lumens that are:

≥ 0.77 mm ID and ≤ 410 mm in length

≥ 1.8 mm ID x ≤ 542 mm in length

o Triple channeled devices with stainless steel lumens that are:

≥ 1.2 mm ID and ≤ 275 mm in length

≥ 1.8 mm ID and ≤ 310 mm in length

Or

≥ 2.8 mm ID and ≤ 317 mm in length

V-PRO 60 and s2 Non Lumen Cycle:

Non-lumened devices including non-lumened rigid, semi-rigid and flexible endossopes and non-lumened devices with diffusion-restricted spaces such as the hinged portion of forceps and scissors.

V-PRO 60 and s2 Flexible Cycle:

Load 1: One flexible surgical endoscope with a light cord (if not integral to endoscope) and mat without any additional load. The flexible endoscope may be a:

· Single or dual lumen device with lumens that are ≥ 1 mm ID and ≤ 990 mm in length

Load 2: Non-lumened devices including non-lumened rigid, semi-rigid, and flexible endoscopes and non-lumened

{7}------------------------------------------------

devices with diffusion-restricted areas such as the hinged portion of forceps or scissors. Medical devices, including rigid and semi-rigid endoscopes with the following configurations:

  • ≥ 0.76 mm ID and ≤ 233 mm in length
  • ≥ 1.0 mm ID and < 254 mm in length
  • ≥ 1.8 mm ID and ≤ 542 mm in length

Intended Sterilization Cycle and Intended Tray Load for Tray Models: VP0045, VP0047, VP0049, VP0049, VP0049

V-PRO s2 Fast Cycle:

· Non-lumened devices including non-lumened rigid, semi-rigid and flexible endossopes, and non-lumened devices with diffusion-restricted areas such as the hinged portion of forceps and scissors.

• Medical devices (including single, dual and triple channeled rigid endoscopes) with the following configurations:

o Single or dual channeled devices with stainless steel lumens

≥ 0.77 mm ID and ≤ 410 mm in length

≥ 1.8 mm ID and ≤ 542 mm in length

o Triple channeled devices with stainless steel lumens that are either

≥ 1.2 mm ID and ≤ 275 mm in length

≥ 1.8 mm ID and < 310 mm in length

≥ 2.8 mm ID and ≤ 317 mm in length

Intended Sterilization Cycles and Intended Tray Load for Tray Models: VP0045, VP0047, VP0048, VP0049, VP0050, VP0051, VP0052, VP0053

V-PRO 1, 1 Plus, maX, and maX 2 Lumen Cycle:

· Non-lumened devices with diffusion-restricted spaces such as the hinged portion of forceps and scissors

· Medical devices, including single, dual or triple channeled stainless steel lumens that are:

≥ 0.77 mm ID and ≤ 527 mm in length

≥ 0.8 mm ID and < 542 mm in length

≥ 0.48 mm ID and ≤ 100 mm in length

• Medical devices with dead end stainless steel lumens that are ≥ 1.3 mm ID and ≤ 73 mm in length

· Devices with rigid non-metallic lumens (such as those used in endoscope sheaths, take-apart forceps and trocars) that are:

3 mm ID and < 298 mm in length

≥ 4 mm ID and ≤ 424 mm in length

V-PRO 1 Plus, maX, and maX 2 Non Lumen Cycle:

Non-lumened devices including non-lumened rigid, semi-rigid and flexible endoscopes and non-lumened devices with diffusion-restricted spaces such as the hinged portion of forceps and scissors

V-PRO maX, and maX 2 Flexible Cycle:

Load 1: Single or dual lumen surgical flexible endoscopes (such as those used in ENT, Urology and Surgical Care) and bronchoscopes with a light cord (if not integral to endoscope) and mat with no additional load. The flexible endoscopes may contain either a single or dual lumen that is > 1 mm ID and ≤ 1050 mm in length.

Load 2:

• Non-lumened devices including non-lumened rigid, semi-rigid and flexible endoscopes and non-lumened devices with diffusion-restricted spaces such as the hinged portion of forceps and scissors

• Single, dual or triple channel stainless steel lumens that are > 0.48 mm ID and ≤ 100 mm in length

V-PRO maX 2 Fast Non Lumen Cycle:

Non-lumened devices including non-lumened rigid, semi-rigid and flexible endossopes and non-lumened devices with diffusion-restricted spaces such as the hinged portion of forceps and scissors.

Intended Sterilization Cycles and Intended Tray Load for Tray Models: VP0045, VP0047, VP0048, VP0049,

{8}------------------------------------------------

VP0050, VP0051, VP0052

STERRAD 100S Default Cycle:

Metal and nonmetal medical devices including instruments which have diffusion-restricted spaces, such as the hinged portion of forceps and scissors. Metal and nonmetal lumened instruments with

≥ 6 mm ID and ≤ 310 mm in length

Medical devices with a single stainless steel lumen with:

≥ 1 mm ID and ≤ 125 mm in length

2 mm ID and < 250 mm in length

≥3 mm ID and ≤ 400 mm in length

Intended Sterilization Cycles and Intended Tray Load for Tray Models: VP0045, VP0046, VP0049

STERRAD NX and NX with ALLClear Technology Standard Cycle:

Metal and non-metal medical devices including instruments which have diffusion-restricted spaces, such as the hinged portion of forceps and scissors.

Medical devices with a single stainless steel lumen with:

≥ 1 mm ID and ≤ 150 mm in length

≥2 mm ID and < 400 mm in length

STERRAND NX and NX with ALLClear Technology Advanced Cycle:

Metal and non-metal medical devices including instruments which have diffusion-restricted spaces, such as the hinged portion of forceps and scissors

Medical devices, including most flexible endoscopes, with:

  • o a single stainless steel lumen with:
    ≥ 1 mm ID and ≤ 500 mm in length

  • o single channel polyethylene and Teflon (polytetrafluoroethylene)
    ≥ 1mm ID and ≤ 850 mm in length

Intended Sterilization Cycles and Intended Tray Load for Tray Models: VP0045, VP0048, VP0051, VP0052, VP0053

STERRAD 100NX and 100NX with ALLClear Technology Standard Cycle:

Metal and nonmetal medical devices including instruments with have diffusion-restricted spaces, such as the hinged portion of forceps and scissors

Medical devices with a single stainless steel lumen with:

≥ 0.7 mm ID and ≤ 500 mm in length

STERRAD 100NX and 100NX with ALLClear Technology Flex Scope Cycle:

Metal and nonmetal medical devices including instruments which have diffusion-restricted spaces, such as the hinged portion of forceps and scissors.

Medical devices, including most flexible endoscopes, with:

· Single channel polyethylene and Teflon (polytetrafluoroethylene)

≥ 1mm ID and < 850 mm in length

STERRAD 100NX and 100NX with ALLClear Technology Express Cycle:

Metal and nonmetal devices surfaces and instruments which have diffusion-restricted spaces, such as the hinged portion of forceps and scissors.

STERRAD 100NX and 100NX with ALLClear Technology Duo Cycle:

Medical devices including:

· most flexible endoscopes with a single channel of polyethylene and Teflon (polytetraffluoroethylene) with ≥ 1 mm ID and ≤ 875 mm in length

· accessory devices that are normally connected to a flexible endoscope during use

{9}------------------------------------------------

· flexible endoscopes without lumens

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

区 Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{10}------------------------------------------------

Image /page/10/Picture/1 description: The image shows the logo for STERIS. The logo consists of the word "STERIS" in a bold, sans-serif font, with a registered trademark symbol to the right of the word. Below the word is a graphic of several horizontal, wavy lines in blue, resembling water.

510(k) Summary for Vis-U-All Low Temperature Sterilization Pouches/Tubing

Sponsor Facility

STERIS Corporation 5960 Heisley Road Mentor, OH 44060 Phone: (440) 354-2600 Fax No: (440) 357-9198

Contact: Anthony Piotrkowski Director, Regulatory Affairs

Telephone: (440) 392-7437 Fax No: (440) 357-9198 e-mail: tony piotrkowski(@steris.com

Submission Date: September 7, 2022

Premarket Notification Number: K222440

STERIS Corporation ■ 5960 Heisley Road ■ Mentor, OH 44060-1834 USA ■ 440-354-2600

{11}------------------------------------------------

1. Device Name

Trade Name:Vis-U-All Low Temperature SterilizationPouches/Tubing
Device Classification:Class II
Common/Usual Name:Sterilization pouch
Classification Name:Sterilization wrap
Classification Number:21 CFR 880.6850
Product Code:FRG

2. Predicate Device

Vis-U-All Low Temperature Sterilization Pouches/Tubing, K183401

3. Description of Device

The proposed Vis-U-All Low Temperature Sterilization Pouches/Tubing is identical to the predicate and is a Tyvek/plastic film sterilization containment pouch designed for devices to be sterilized by the health care provider in V-PRO or STERRAD Low Temperature Sterilization Systems. As is the predicate device, the proposed device is available as a self-seal pouch, a heat-seal pouch, or heat-seal tubing. Available sizes and configurations are shown in Table 1.

Table 1. Sizes and Configurations of Vis-U-All Low Temperature Sterilization Pouches/Tubing
---------------------------------------------------------------------------------------------------------
TypeSize *TypeSize*TypeSize*
HeatSealPouch3 x 73 x 7Tubing3" x 100'
4 x 94 x 94" x 100'
4 x 12Self SealPouch4 x 126" x 100'
4 x 224 x 229" x 100'
6 x 106 x 1014" x 100'
8 x 128 x 12
10 x 1510 x 15
12 x 1812 x 18
8 x 21
8 x 27
9 x 27
11 x 22
12 x 27

*Sizes are in inches unless specified otherwise

The purpose of this submission is to qualify use of the Vis-U-All Low Temperature Sterilization Pouches and Tubing for extended claims in the V-PRO 60, s2, maX and maX 2 Sterilizer Cycles.

{12}------------------------------------------------

Intended Use/ Indications for Use 4.

The Vis-U-All Low Temperature Sterilization Pouches/Tubing Intended use remains the same. The indications for use are being modified to expand lumen claims. A detailed comparison of the cleared and proposed indications for use are provided in Table 2.

5. Comparison of the Modified Device to the Predicate

The proposed and predicate devices are single use sterilization pouches for use in V-PRO Sterilizers. Table 2 summarizes the difference between the proposed device and predicate device cleared under K183401.

Table 2. Technical Comparison to the Predicate.
-------------------------------------------------
FeatureModified Vis-U-All Low TemperatureSterilization Pouch (proposed) K222440Vis-U-All Low TemperatureSterilization Pouch (K183401)Comparison
IntendedUse /Indicationsfor UseThe Vis-U-All Low TemperatureSterilization Pouches/Tubing aresterilization containment pouches for useby health care providers to enclose:• medical devices in a single or doublepouch configuration• trays* containing medical devices in asingle or double pouch configuration• small items requiring surfacesterilization in a single pouchconfiguration within a tray*NOTE: Trays must be legally marketedfor use in the V-PRO Low Temperatureor STERRAD Sterilization Systems andcontain a vent surface area to trayvolume ratio ≥ $0.135 in^{-1}$ with themaximum number of instrumentorganizers installed.to be sterilized in the:• Lumen, Non Lumen, Flexible, FastNon Lumen and Fast Cycles of the V-PRO® Low Temperature SterilizationSystems• Default Cycle of the STERRAD 100SSterilizer• Standard and Advanced Cycles of theSTERRAD NX and NX withALLClear Technology Sterilizers• Express, Standard, Flex Scope andDUO Cycles of the STERRAD 100NXand 100NX with ALLClearTechnology SterilizersThe Vis-U-All Low TemperatureSterilization Pouches/Tubing aresterilization containment pouches for useby health care providers to enclose:• medical devices in a single or doublepouch configuration• trays containing medical devices in asingle or double pouch configuration• small items requiring surfacesterilization in a single pouchconfiguration within a trayNOTE: Trays must be legally marketedfor use in the V-PRO Low Temperatureor STERRAD Sterilization Systems andcontain a vent surface area to trayvolume ratio ≥ $0.135 in^{-1}$ with themaximum number of instrumentorganizers installed.to be sterilized in the:• Lumen, Non Lumen, Flexible, FastNon Lumen and Fast Cycles of the V-PRO® Low Temperature SterilizationSystems• Default Cycle of the STERRAD 100SSterilizer• Standard and Advanced Cycles of theSTERRAD NX and NX with ALLClearTechnology Sterilizers• Express, Standard, Flex Scope andDUO Cycles of the STERRAD 100NXand 100NX with ALLClearTechnology SterilizersIdentical exceptfor:• Clarificationof lumenclaims for V-PRO lumencycles• Clarificationof endoscopeclaims for V-PRO flexiblecycles• Removal of"stainless steelor titanium"from V-PRONon Lumencycles.
FeatureModified Vis-U-All Low TemperatureSterilization Pouch (proposed) K222440Vis-U-All Low TemperatureSterilization Pouch (K183401)Comparison
*STERRAD and ALLClear aretrademarks of Advanced SterilizationProducts*STERRAD and ALLClear aretrademarks of Advanced SterilizationProducts
The pouches maintain the sterility of theenclosed devices until used.The pouches maintain the sterility of theenclosed devices until used.
When used to enclose medical devices,the pouches are intended to contain thedevices in such a manner as to leave aminimum of one inch between thedevices and seal on all sides. When usedto enclose a tray, the tray must fit looselywithin the pouch.When used to enclose medical devices,the pouches are intended to contain thedevices in such a manner as to leave aminimum of one inch between thedevices and seal on all sides. When usedto enclose a tray, the tray must fit looselywithin the pouch.
Intended Sterilization Cycles andIntended Pouch Loads when MedicalDevices are:• Directly pouched• Placed inside of a tray and the traypouchedIntended Sterilization Cycles andIntended Pouch Loads when MedicalDevices are:• Directly pouched• Placed inside of a tray and the traypouched
V-PRO 60 & s2 Lumen Cycle• Non-lumened devices with diffusion-restricted spaces such as the hingedportion of forceps and scissors.• Non-lumened devices including non-lumened rigid and semi-rigidendoscopes• Medical devices, including single, dualand triple channeled rigid and semi-rigid endoscopes, with the followingconfigurations:o single or dual lumen devices• $\ge$ 0.77 mm internal diameter(ID) and $\le$ 410 mm in length• $\ge$ 1.8 mm ID x $\le$ 542 mm inlengtho triple lumen devices• $\ge$ 1.2 mm ID and $\le$ 275 mm inlength• $\ge$ 1.8 mm ID and $\le$ 310 mm inlengthOr• $\ge$ 2.8 mm ID and $\le$ 317 mm inlengthV-PRO 60 & s2 Lumen Cycle• Instruments with diffusion-restrictedspaces such as the hinged portion offorceps and scissors• Non-lumened devices including non-lumened rigid and semi-rigidendoscopes• Medical devices, including single, dualand triple channeled rigid and semi-rigid endoscopes, with the followingconfigurations:o single or dual lumen devices• $\ge$ 0.77 mm internal diameter(ID) and $\le$ 410 mm in lengtho triple lumen devices• $\ge$ 1.2 mm ID and $\le$ 275 mm inlength• $\ge$ 1.8 mm ID and $\le$ 310 mm inlengthor• $\ge$ 2.8 mm ID and $\le$ 317 mm inlength
FeatureModified Vis-U-All Low TemperatureSterilization Pouch (proposed) K222440Vis-U-All Low TemperatureSterilization Pouch (K183401)Comparison
V-PRO 60 & s2 Non Lumen CycleNon-lumened devices including non-lumened rigid, semi-rigid and flexibleendoscopes and non-lumened deviceswith diffusion-restricted spaces such asthe hinged portion of forceps andscissors.V-PRO 60 & s2 Non Lumen CycleNon-lumened devices including non-lumened rigid, semi-rigid and flexibleendoscopes and non-lumened deviceswith diffusion-restricted spaces such asthe hinged portion of forceps andscissors.
V-PRO 60 & s2 Flexible CycleLoad 1: One flexible surgical endoscopeor bronchoscope with a light cord (if notintegral to endoscope) and mat withoutany additional load. The flexibleendoscope may be a:• single or dual lumen device withlumens that are $\ge$ 1 mm ID and $\le$990 mm in lengthLoad 2: Non-lumened devices includingnon-lumened rigid semi-rigid, andflexible endoscopes and non-lumeneddevices with diffusion-restricted areassuch as the hinged portion of forceps orscissors. Medical devices, including rigidand semi-rigid endoscopes, with thefollowing configurations:• $\ge$ 0.76 mm ID and $\le$ 233 mm in length• $\ge$ 1.0 mm ID and $\le$ 254 mm in length• $\ge$ 1.8 mm ID and $\le$ 542 mm in lengthV-PRO 60 & s2 Flexible CycleLoad 1: One flexible surgical endoscopeor bronchoscope with a light cord (if notintegral to endoscope) and mat withoutany additional load. The flexibleendoscope may be a:• single or dual lumen device withlumens that are $\ge$ 1 mm ID and $\le$ 990mm in lengthLoad 2: Non-lumened devices includingnon-lumened rigid, semi-rigid, andflexible endoscopes and non-lumeneddevices with diffusion-restricted areassuch as the hinged portion of forceps orscissors. Medical devices, including rigidand semi-rigid endoscopes, with thefollowing dimensions:• $\ge$ 2 mm ID and $\le$ 400 mm in length• $\ge$ 0.76 mm ID and $\le$ 233 mm in length• $\ge$ 1.0 mm ID and $\le$ 254 mm in length
V-PRO s2 Fast Cycle• Non-lumened devices including non-lumened rigid and semi-rigid andflexible endoscopes and non-lumeneddevices with diffusion-restricted spacessuch as the hinged portion of forcepsand scissors.• Medical devices (including single, dualand triple channeled rigid and semi-rigid endoscopes) with the followingconfigurations:• single or dual lumen devices$\ge$ 0.77 mm ID and $\le$ 410 mm inlength$\ge$ 1.8 mm ID x $\le$ 542 mm inlength• Triple channeled devices with stainlessV-PRO s2 Fast Cycle• Non-lumened devices including non-lumened rigid and semi-rigid andflexible endoscopes and non-lumeneddevices with diffusion-restricted spacessuch as the hinged portion of forcepsand scissors.• Medical devices (including single, dualand triple channeled rigid and semi-rigid endoscopes) with the followingconfigurations:single or dual channeled deviceswith stainless steel lumens thatare $\ge$ 0.77 mm ID and $\le$ 410 mmin length• Triple channeled devices with stainless
steel lumens that are either:steel lumens that are either:
FeatureModified Vis-U-All Low TemperatureSterilization Pouch (proposed) K222440Vis-U-All Low TemperatureSterilization Pouch (K183401)Comparison
$\geq$ 1.2 mm ID and $\leq$ 275 mm in length $\geq$ 1.8 mm ID and $\leq$ 310 mm in length Or$\geq$ 2.8 mm ID and $\leq$ 317 mm in length V-PRO 1, 1 Plus, maX & maX 2 Lumen Cycle Non-lumened devices with diffusion-restricted spaces such as the hinged portion of forceps and scissors. Non-lumened devices including non-lumened rigid and semi-rigid endoscopes Medical devices, including single, dual or triple channeled stainless steel lumens that are: $\geq$ 0.77 mm ID and $\leq$ 527 mm in length $\geq$ 0.8 mm ID and $\leq$ 542 mm in length $\geq$ 0.48 mm ID and $\leq$ 100 mm in length Medical devices with Dead end lumens that are $\geq$ 1.3 mm ID and $\leq$ 73 mm in length Devices with rigid non-metallic lumens (such as those used in endoscope sheaths, take-apart forceps and trocars) that are: $\geq$ 3 mm ID and $\leq$ 298 mm in length $\geq$ 4 mm ID and $\leq$ 424 mm in length$\geq$ 1.2 mm ID and $\leq$ 275 mm in length $\geq$ 1.8 mm ID and $\leq$ 310 mm in length Or$\geq$ 2.8 mm ID and $\leq$ 317 mm in length V-PRO 1, 1 Plus, maX & maX 2 Lumen Cycle Instruments with diffusion-restricted spaces such as the hinged portion of forceps and scissors Non-lumened devices including non-lumened rigid and semi-rigid endoscopes Medical devices, including single, dual and triple channeled rigid and semi-rigid endoscopes, with the following configurations: single or dual lumen devices $\geq$ 0.77 mm ID and $\leq$ 410 mm in length triple lumen devices $\geq$ 1.2 mm ID and $\leq$ 275 mm in length $\geq$ 1.8 mm ID and $\leq$ 310 mm in length or$\geq$ 2.8 mm ID and $\leq$ 317 mm in length
V-PRO 1, 1 Plus, maX & maX2 Non Lumen CycleNon-lumened devices including non-lumened rigid, semi-rigid and flexible endoscopes and non-lumened devices with diffusion-restricted spaces such as the hinged portion of forceps and scissors.V-PRO 1, 1 Plus, maX & maX2 Non Lumen CycleNon-lumened devices including non-lumened rigid, semi-rigid and flexible endoscopes and non-lumened devices with stainless steel or titanium diffusion-restricted spaces such as the hinged portion of forceps and scissors.
FeatureModified Vis-U-All Low TemperatureSterilization Pouch (proposed) K222440Vis-U-All Low TemperatureSterilization Pouch (K183401)Comparison
Load 1: Single or dual lumen surgicalflexible endoscopes (such as those usedin ENT, Urology and Surgical Care) andbronchoscopes with a light cord (if notintegral to endoscope) and mat with noadditional load.The flexible endoscopes may containeither a single or dual channel lumen thatis $\geq$ 1 mm ID and $\leq$ 1050 mm in lengthLoad 2:• Non-lumened devices including non-lumened rigid, semi-rigid and flexibleendoscopes and non-lumened deviceswith diffusion-restricted spaces suchas the hinged portion of forceps andscissors• Single, dual or triple channel stainlesssteel lumen that is $\geq$ 0.48 mm ID and$\leq$ 100 mm in length.Load 1: Single or dual lumen surgicalflexible endoscopes (such as those usedin ENT, Urology and Surgical Care) andbronchoscopes with a light cord (if notintegral to endoscope) and mat with noadditional load.The flexible endoscopes may containeither:• a single lumen that is $\geq$ 1 mm IDand $\leq$ 1050 mm in length• or two lumens with:• one lumen that is $\geq$ 1 mm ID and$\leq$ 990 mm in length• and the other lumen that is $\geq$ 1mm ID and $\leq$ 850 mm in lengthLoad 2: Non-lumened instrumentsincluding instruments with diffusion-restricted areas such as the hingedportion of forceps or scissors.
V-PRO maX 2 Fast Non Lumen CycleNon-lumened devices including non-lumened rigid, semi-rigid and flexibleendoscopes and non-lumened deviceswith diffusion-restricted spaces such asthe hinged portion of forceps andscissors.V-PRO maX 2 Fast Non Lumen CycleNon-lumened devices including non-lumened rigid, semi-rigid and flexibleendoscopes and non-lumened deviceswith stainless steel or titanium diffusion-restricted spaces such as the hingedportion of forceps and scissors.
STERRAD 100S Default CycleMetal and nonmetal medical devicesincluding instruments which havediffusion-restricted spaces, such as thehinged portion of forceps and scissors.Metal and nonmetal lumened instrumentswith:• $\geq$ 6 mm ID and $\leq$ 310 mm in lengthMedical devices with a single stainlesssteel lumen with:• $\geq$ 1 mm ID and $\leq$ 125 mm in length• $\geq$ 2 mm ID and $\leq$ 250 mm in length• $\geq$ 3 mm ID and $\leq$ 400 mm in lengthSTERRAD 100S Default CycleMetal and nonmetal medical devicesincluding instruments which havediffusion-restricted spaces, such as thehinged portion of forceps and scissors.Metal and nonmetal lumened instrumentswith:• $\geq$ 6 mm ID and $\leq$ 310 mm in lengthMedical devices with a single stainlesssteel lumen with:• $\geq$ 1 mm ID and $\leq$ 125 mm in length• $\geq$ 2 mm ID and $\leq$ 250 mm in length• $\geq$ 3 mm ID and $\leq$ 400 mm in length
STERRAD NX and NX with ALLClearTechnology Standard CycleMetal and nonmetal medical devicesincluding instruments which havediffusion-restricted spaces, such as thehinged portion of forceps and scissors.STERRAD NX and NX with ALLClearTechnology Standard CycleMetal and nonmetal medical devicesincluding instruments which havediffusion-restricted spaces, such as thehinged portion of forceps and scissors
FeaturePRO-LITE Sterilization TraysModified Vis-U-All Low TemperatureSterilization Pouch (proposed) K222440Vis-U-All Low TemperatureSterilization Pouch (K183401)Comparison
Medical devices with a single stainlesssteel lumen with:• ≥ 1 mm ID and ≤ 150 mm in length• ≥ 2 mm ID and ≤ 400 mm in lengthMedical devices with a single stainlesssteel lumen with:• ≥ 1 mm ID and ≤ 150 mm in length• ≥ 2 mm ID and ≤ 400 mm in length
STERRAD NX and NX with ALLClearTechnology Advanced CycleMetal and nonmetal medical devicesincluding instruments which havediffusion-restricted spaces, such as thehinged portion of forceps and scissors.STERRAD NX and NX with ALLClearTechnology Advanced CycleMetal and nonmetal medical devicesincluding instruments which havediffusion-restricted spaces, such as thehinged portion of forceps and scissors.
Medical Devices, including most flexibleendoscopes, with:• a single stainless steel lumen with:○ ≥ 1 mm ID and ≤ 500 mm inlength• Single channel polyethylene andTeflon (polytetrafluoroethylene)○ ≥1 mm ID and ≤ 850 mm in lengthMedical Devices, including most flexibleendoscopes, with:• a single stainless steel lumen with:• ≥ 1 mm ID and ≤ 500 mm inlength• Single channel polyethylene andTeflon (polytetrafluoroethylene)• ≥1 mm ID and ≤ 850 mm in length
STERRAD 100NX and 100NX withALLClear Technology Standard CycleMetal and nonmetal medical devicesincluding instruments which havediffusion-restricted spaces, such as thehinged portion of forceps and scissors.Medical devices with a single stainlesssteel lumen with:• ≥ 0.7 mm ID and ≤ 500 mm in lengthSTERRAD 100NX and 100NX withALLClear Technology Standard CycleMetal and nonmetal medical devicesincluding instruments which havediffusion-restricted spaces, such as thehinged portion of forceps and scissors.Medical devices with a single stainlesssteel lumen with:• ≥ 0.7 mm ID and ≤ 500 mm in length
STERRAD 100NX and 100NX withALLClear Technology Flex Scope CycleMetal and nonmetal medical devicesincluding instruments which havediffusion-restricted spaces, such as thehinged portion of forceps and scissors.Medical Devices, including most flexibleendoscopes, with:• Single channel polyethylene andTeflon (polytetrafluoroethylene)• ≥ 1 mm ID and ≤ 850 mm inlengthSTERRAD 100NX and 100NX withALLClear Technology Flex Scope CycleMetal and nonmetal medical devicesincluding instruments which havediffusion-restricted spaces, such as thehinged portion of forceps and scissors.Medical Devices, including most flexibleendoscopes, with:Single channel polyethylene andTeflon (polytetrafluoroethylene)• ≥ 1 mm ID and ≤ 850 mm inlength
STERRAD 100NX and 100NX withALLClear Technology Express CycleMetal and nonmetal medical devices(surfaces sterilization only) andinstruments which have diffusion-STERRAD 100NX and 100NX withALLClear Technology Express CycleMetal and nonmetal medical devices(surfaces sterilization only) and
PRO-LITE Sterilization Trays
FeatureModified Vis-U-All Low TemperatureSterilization Pouch (proposed) K222440Vis-U-All Low TemperatureSterilization Pouch (K183401)Comparison
restricted spaces, such as the hingedportion of forceps and scissors.STERRAD 100NX and 100NX withALLClear Technology Duo CycleMedical devices including:most flexible endoscopes with asingle channel of polyethylene andTeflon (polytetrafluoroethylene) with$\geq$ 1 mm ID and $\leq$ 875 mm in length accessory devices that are normallyconnected to a flexible endoscopeduring use flexible endoscopes without lumensinstruments which have diffusion-restricted spaces, such as the hingedportion of forceps and scissors.STERRAD 100NX and 100NX withALLClear Technology Duo CycleMedical devices including:most flexible endoscopes with asingle channel of polyethylene andTeflon (polytetrafluoroethylene) with$\geq$ 1 mm ID and $\leq$ 875 mm in length accessory devices that are normallyconnected to a flexible endoscopeduring use flexible endoscopes without lumens
DeviceFeaturesChevron end of pouches for ease ofopeningChemical process indicator for EOChevron end of pouches for ease ofopeningChemical process indicator for EOIdentical
Maintenanceof Sterility1 year1 yearIdentical
Materials ofConstructionTyvek and plasticTyvek and plasticIdentical
TypesSelf Seal, Heat Seal, TubingSelf Seal, Heat Seal, TubingIdentical

{13}------------------------------------------------

{14}------------------------------------------------

{15}------------------------------------------------

{16}------------------------------------------------

{17}------------------------------------------------

{18}------------------------------------------------

6. Summary of Performance Testing

Table 3 summarizes the testing of the Vis-U-All Low Temperature Sterilization Pouches/Tubing to demonstrate that the proposed pouch is qualified for use in V-PRO Low temperature Sterilization Systems.

Table 3. Performance Test Summary

TestResultConclusion
½ CycleVerification ofMated SurfacesSterile efficacy was demonstrated for mated surfaces under worst case conditions in the V-PRO Sterilizer cycles.PASS
½ Cycle EfficacyThe standard injection weight resulted in all sterile results within the validation load used to qualify each sterilizer cycle.PASS
Simulated UseTestSimulated use testing verified the ability of the sterilizer cycles to sterilize medical devices under worst-case processing conditions.PASS
In Use TestThe in use investigation demonstrated the ability of the V-PRO Sterilizer cycles to sterilize patient-soiled, clinically-cleaned, medical instruments.PASS

{19}------------------------------------------------

7. Conclusion

Based on the intended use, technological characteristics and non-clinical performance data, the subject device is as safe, as effective and performs as well as or better than the legally marketed predicate device (K183401), Class II (21 CFR 880.6850), product code FRG.

{20}------------------------------------------------

510(k) Summary For PRO-LITE™ Sterilization Tray

Sponsor Facility

STERIS Corporation 5960 Heisley Road Mentor, OH 44060 Phone: (440) 354-2600 Fax No: (440) 357-9198

Anthony Piotrkowski Contact: Director, Regulatory Affairs

Telephone: (440) 392-7437 Fax No: (440) 357-9198 e-mail: tony piotrkowski(@steris.com

Submission Date: September 7, 2022

Premarket Notification Number: K222440

STERIS Corporation ■ 5960 Heisley Road ■ Mentor, OH 44060-1834 USA ■ 440-354-2600

{21}------------------------------------------------

Device Name 7.

Trade Name:PRO-LITE Sterilization Tray
Common/usual Name:Sterilization Trays, cassettes and other accessories
Classification Name:Sterilization Wrap
Classification21 CFR 880.6850
ClassII
Product CodeKCT, FRG

8. Predicate Device

PRO-LITE Sterilization Tray, K183402

9. Description of Device

The PRO-LITE Sterilization Trays contain medical devices for sterilization in the V-PRO Low Temperature Sterilization Systems and STERRAD Sterilizers. The trays are available in various sizes, outlined in Table 4, to accommodate the loads to be processed. The proposed trays are identical in design to the predicate Sterilization Tray (K183402) and are composed of a base and a lid. The lid includes clamping mechanisms designed to secure the lid onto the base. There are numerous holes in the base and lid for sterilant penetration. The tray is categorized as a cassette and requires complete enclosure in a legally-marketed sterilization wrap or pouch to maintain sterility of the enclosed devices. Both the base and the lid for the proposed tray are made of a mineral-filled polypropylene material.

ModelDescription (in)ModelDescription (in)ModelDescription (in)
VP004513 x 4.5 x 2.25VP004813 x 7.75 x 2.25VP005112 x 11.75 x 4
VP004619 x 4.5 x 2.25VP004919 x 7.75 x 2.25VP005225 x 11.75 x 4
VP004725 x 4.5 x 2.25VP005027 x 7.75 x 4VP005325 x 14 x 5

Table 4. External Dimensions of Trav Line

Optional instrument organizers are provided as accessories to the trays and allow stabilization of various cylindrical medical devices during processing. Table 5 lists the organizer sizes. The organizers are either "blank" and are used to partition the tray or have a device holding portion into which the devices are inserted. At the organizer base is a flapped groove that is used to position the organizer within a PRO-LITE Sterilization Tray.

Table 5. Instrument Organizer Model Numbers

ModelDescriptionModelDescription
VP0054Blank, TallVP0055Blank, Short
VP00633 mm - 7mm, TallVP00673 mm - 7 mm, Short
VP00647 mm - 11mm, TallVP00687 mm - 11 mm, Short
VP006511 mm - 15 mm, TallVP006911 mm - 15 mm, Short
VP006615 mm - 19mm, TallVP007015 mm - 19 mm, Short

STERIS Corporation ■ 5960 Heisley Road ■ Mentor, OH 44060-1834 USA ■ 440-354-2600

{22}------------------------------------------------

Optional sterilization mats are provided as accessories to the trays. The mats, which are used to cushion and stabilize devices placed into the trays, are available in sizes as shown in Table 6 to fit the nine PRO-LITE Sterilization Trays. The mats are a diamond grid design with "fingers" that extend from each corner of the diamond and at the midpoint of each diamond side. The fingers cushion and stabilize instruments, helping to prevent the instruments from freely moving in the tray during packaging, sterilization and storage. The cushioning and stabilization qualities help protect delicate instruments placed into the trays.

ModelDescription (in)ModelDescription (in)ModelDescription (in)
VP007113 x 4.5VP007413 x 7.75VP007712 x 11.75
VP007219 x 4.5VP007519 x 7.75VP007825 x 11.75
VP007325 x 4.5VP007627 x 7.75VP007925 x 14

Tahle 6 Silicone Mat Model Numbers

The purpose of this submission is to expand claims for the use of these tray models in the following sterilizer cycles:

  • V-PRO 1, 1 Plus, maX and maX 2 Lumen Cycle ●
  • V-PRO 1 Plus, maX and maX 2 Non Lumen Cycle ●
  • V-PRO maX and maX 2 Flexible Cycle ●
  • V-PRO maX 2 Fast Non Lumen Cycle ●
  • V-PRO 60 and s2 Lumen, Non Lumen and Flexible Cycles ●
  • V-PRO s2 Fast Cycle

10. Intended Use/ Indications for Use

The tray indications for use are included in Table 7 which compares the proposed device to the predicate.

Instrument organizers are optional accessories intended to stabilize cylindrical medical instruments within the PRO-LITE Sterilization Trays.

ModelDescriptionModelDescription
VP0054Blank, TallVP0055Blank, Short
VP00633 mm - 7 mm, TallVP00673 mm - 7 mm, Short
VP00647 mm - 11 mm, TallVP00687 mm - 11 mm, Short
VP006511 mm - 15 mm, TallVP006911 mm - 15 mm, Short
VP006615 mm - 19 mm, TallVP007015 mm - 19 mm, Short

Sterilization mats are optional accessories intended to cushion and stabilize instruments within the PRO-LITE Sterilization Trays.

ModelDescription (in)ModelDescription (in)ModelDescription (in)
VP007113 x 4.5VP007413 x 7.75VP007712 x 11.75
VP007219 x 4.5VP007519 x 7.75VP007825 x 11.75
VP007325 x 4.5VP007627 x 7.75VP007925 x 14

{23}------------------------------------------------

11. Summary of Technical Characteristics

The proposed PRO-LITE sterilization trays, sterilization mats and instrument organizers are identical in composition to the claimed predicate device. The technical characteristics are summarized below in Table 7.

Table 7. Summary of Tray Physical Description and Technological Properties
----------------------------------------------------------------------------
FeaturePRO-LITE Sterilization Tray (proposed) K222440PRO-LITE Sterilization Tray (K183402)Comparison
Intended Use / Indications for UseThe PRO-LITE Sterilization Trays are used to contain medical devices for sterilization in the following Cycles:• Lumen, Non Lumen, Flexible, Fast Non Lumen and Fast Cycles of the V-PRO Low Temperature Sterilization Systems• Default Cycle of the STERRAD®* 100S Sterilizer• Standard and Advanced Cycles of the STERRAD NX and NX with ALLClear Technology Sterilizers• Standard, Flex Scope, Express and DUO Cycles of the STERRAD 100NX and 100NX with ALLClear Technology Sterilizers*STERRAD and ALLClear are trademarks of Advanced Sterilization ProductsPrior to placing in the Sterilizer, the trays must either be:• wrapped with a legally marketed sterilization wrap for use in the Sterilizers listed aboveor• placed inside a legally marketed pouch for enclosing trays in the Sterilizers listed above.The PRO-LITE Sterilization Trays are not intended to maintain sterility; they are intended to be used in conjunction with a validated, FDA-cleared sterilization wrap or pouch in order to maintain sterility of the enclosedThe PRO-LITE Sterilization Trays are used to contain medical devices for sterilization in the following Cycles:• Lumen, Non Lumen, Flexible, Fast Non Lumen and Fast Cycles of the V-PRO Low Temperature Sterilization Systems• Default Cycle of the STERRAD®* 100S Sterilizer• Standard and Advanced Cycles of the STERRAD NX and NX with ALLClear Technology Sterilizers• Standard, Flex Scope, Express and DUO Cycles of the STERRAD 100NX and 100NX with ALLClear Technology Sterilizers*STERRAD and ALLClear are trademarks of Advanced Sterilization ProductsPrior to placing in the Sterilizer, the trays must either be:• wrapped with a legally marketed sterilization wrap for use in the Sterilizers listed aboveor• placed inside a legally marketed pouch for enclosing trays in the sterilizers listed aboveThe PRO-LITE Sterilization Trays are not intended to maintain sterility; they are intended to be used in conjunction with a validated, FDA-cleared sterilization wrap or pouch in order to maintain sterility of the enclosedIdentical except for:• Clarification of lumen claims for V-PRO lumen cycles• Clarification of endoscope claims for V-PRO flexible cyclesRemoval of “stainless steel or titanium” from V-PRO Non Lumen cycles.
FeaturePRO-LITE Sterilization Tray(proposed) K222440PRO-LITE Sterilization Tray(K183402)Comparison
Intended Sterilization Cycles andIntended Tray Load for Tray Models:VP0045, VP0046, VP0047, VP0048,VP0049, VP0050, VP0051, VP0052Intended Sterilization Cycles andIntended Tray Load for Tray Models:VP0045, VP0046, VP0047, VP0048,VP0049, VP0050, VP0051, VP0052
V-PRO 60 and s2 Lumen Cycle:• Non-lumened devices with diffusion-restricted spaces such as the hingedportion of forceps and scissors• Non-lumened devices including non-lumened rigid and semi-rigidendoscopes• Medical devices, including single,dual and triple channeled rigid andsemi-rigid endoscopes with thefollowing configurations:o Single or dual channeled deviceswith stainless steel lumens that are:■ $\ge$ 0.77 mm ID and $\le$ 410 mm inlength■ $\ge$ 1.8 mm ID x $\le$ 542 mm inlengtho Triple channeled devices withstainless steel lumens that are:■ $\ge$ 1.2 mm ID and $\le$ 275 mm inlength■ $\ge$ 1.8 mm ID and $\le$ 310 mm inlengthor■ $\ge$ 2.8 mm ID and $\le$ 317 mm inlengthV-PRO 60 and s2 Lumen Cycle:• Instruments with diffusion-restrictedspaces such as the hinged portion offorceps and scissors• Non-lumened devices including non-lumened rigid and semi-rigidendoscopes• Medical devices, including single,dual and triple channeled rigid andsemi-rigid endoscopes with thefollowing configurations:o Single or dual lumen devices■ $\ge$ 0.77 mm ID and $\le$ 410 mm inlength• Triple lumen devices■ $\ge$ 1.2 mm ID and $\le$ 275 mm inlength■ $\ge$ 1.8 mm ID and $\le$ 310 mm inlengthor■ $\ge$ 2.8 mm ID and $\le$ 317 mm inlength
V-PRO 60 and s2 Non Lumen Cycle:Non-lumened devices including non-lumened rigid, semi-rigid and flexibleendoscopes and non-lumened deviceswith diffusion-restricted spaces such asthe hinged portion of forceps andscissors.V-PRO 60 and s2 Non Lumen Cycle:Non-lumened devices including non-lumened rigid, semi-rigid and flexibleendoscopes and non-lumened deviceswith diffusion-restricted spaces such asthe hinged portion of forceps andscissors.
V-PRO 60 and s2 Flexible Cycle:Load 1: One flexible surgicalendoscope or bronchoscope with a lightcord (if not integral to endoscope) andmat without any additional load. Theflexible endoscope may be a:V-PRO 60 and s2 Flexible Cycle:Load 1: One flexible surgicalendoscope or bronchoscope with a lightcord (if not integral to endoscope) andmat without any additional load. Theflexible endoscope may be a:
FeaturePRO-LITE Sterilization Tray(proposed) K222440PRO-LITE Sterilization Tray(K183402)Comparison
Single or dual lumen device withlumens that are $\geq$ 1 mm ID and $\leq$ 990 mm in length Load 2: Non-lumened devices including non-lumened rigid, semi-rigid, and flexible endoscopes and non-lumened devices with diffusion-restricted areas such as the hinged portion of forceps or scissors. Medical devices, including rigid and semi-rigid endoscopes with the following configurations: $\geq$ 0.76 mm ID and $\leq$ 233 mm in length $\geq$ 1.0 mm ID and $\leq$ 254 mm in length $\geq$ 1.8 mm ID and $\leq$ 542 mm in length Intended Sterilization Cycle and Intended Tray Load for Tray Models:VP0045, VP0046, VP0047, VP0048, VP0049Single or dual lumen device withlumens that are $\geq$ 1 mm ID and $\leq$ 990 mm in length Load 2: Non-lumened devices including non-lumened rigid, semi-rigid, and flexible endoscopes and non-lumened devices with diffusion-restricted areas such as the hinged portion of forceps or scissors. Medical devices, including rigid and semi-rigid endoscopes with the following configurations: $\geq$ 2.0 mm ID and $\leq$ 400 mm in length $\geq$ 0.76 mm ID and $\leq$ 233 mm in length $\geq$ 1.0 mm ID and $\leq$ 254 mm in length
V-PRO s2 Fast Cycle: Non-lumened devices including non-lumened rigid, semi-rigid and flexible endoscopes, and non-lumened devices with diffusion-restricted areas such as the hinged portion of forceps and scissors. Medical devices (including single, dual and triple channeled rigid and semi-rigid endoscopes) with the following configurations: Single or dual channeled devices with stainless steel lumens $\geq$ 0.77 mm ID and $\leq$ 410 mm in length $\geq$ 1.8 mm ID and $\leq$ 542 mm in length Triple channeled devices with stainless steel lumens that are either $\geq$ 1.2 mm ID and $\leq$ 275 mm in length $\geq$ 1.8 mm ID and $\leq$ 310 mm in length $\geq$ 2.8 mm ID and $\leq$ 317 mm in lengthV-PRO s2 Fast Cycle: Non-lumened devices including non-lumened rigid, semi-rigid and flexible endoscopes and non-lumened devices with diffusion-restricted spaces such as the hinged portion of forceps and scissors Medical devices (including single, dual and triple channeled rigid and semi-rigid endoscopes) with the following configurations: Single or dual channeled devices with stainless steel lumens that are $\geq$ 0.77 mm ID and $\leq$ 410 mm in length Triple channeled devices with stainless steel lumens that are either $\geq$ 1.2 mm ID and $\leq$ 275 mm in length $\geq$ 1.8 mm ID and $\leq$ 310 mm in length $\geq$ 2.8 mm ID and $\leq$ 317 mm in length
FeaturePRO-LITE Sterilization Tray(proposed) K222440PRO-LITE Sterilization Tray(K183402)Comparison
Intended Sterilization Cycles andIntended Tray Load for Tray Models:VP0045, VP0046, VP0047, VP0048,VP0049, VP0050, VP0051, VP0052,VP0053Intended Sterilization Cycles andIntended Tray Load for Tray Models:VP0045, VP0046, VP0047, VP0048,VP0049, VP0050, VP0050, VP0051,VP0052, VP0053
V-PRO 1, 1 Plus, maX, and maX 2Lumen Cycle:• Non-lumened devices with diffusion-restricted spaces such as the hingedportion of forceps and scissors• Medical devices, including single,dual or triple channeled stainless steellumens that are:○ ≥ 0.77 mm ID and ≤ 527 mm inlength○ ≥ 0.8 mm ID and ≤ 542 mm inlength○ ≥ 0.48 mm ID and ≤ 100 mm inlength• Medical devices with dead endstainless steel lumens that are ≥ 1.3mm ID and ≤ 73 mm in length• Devices with rigid non-metalliclumens (such as those used inendoscope sheaths, take-apart forcepsand trocars) that are:○ ≥ 3 mm ID and ≤ 298 mm in length○ ≥ 4 mm ID and ≤ 424 mm in lengthV-PRO 1, 1 Plus, maX, and maX 2Lumen Cycle:• Instruments with diffusion-restrictedspaces such as the hinged portions offorceps and scissors• Non-lumened devices including non-lumened rigid and semi-rigidendoscopes• Medical devices, including single,dual and triple channeled rigid andsemi-rigid endoscopes, with thefollowing configurations:o single or dual lumen devices■ ≥ 0.77 mm ID and ≤ 410 mm inlengtho triple lumen devices■ ≥ 1.2 mm ID and ≤ 275 mm inlength■ ≥ 1.8 mm ID and ≤ 310 mm inlength■ ≥ 2.8 mm ID and ≤ 317 mm inlength
V-PRO 1 Plus, maX, and maX 2 NonLumen Cycle:Non-lumened devices including non-lumened rigid, semi-rigid and flexibleendoscopes and non-lumened deviceswith diffusion-restricted spaces such asthe hinged portion of forceps andscissorsV-PRO 1 Plus, maX, and maX 2 NonLumen Cycle:Non-lumened devices including non-lumened rigid, semi-rigid and flexibleendoscopes and non-lumened deviceswith stainless steel or titaniumdiffusion-restricted spaces such as thehinged portion of forceps and scissors.
V-PRO maX, and maX 2 FlexibleCycle:Load 1: Single or dual lumen surgicalflexible endoscopes (such as those usedin ENT, Urology and Surgical Care)and bronchoscopes with a light cord (ifnot integral to endoscope) and mat withno additional load.V-PRO maX, and maX 2 FlexibleCycle:Load 1: Single or dual lumen surgicalflexible endoscopes (such as those usedin ENT, Urology and Surgical Care)and bronchoscopes with a light cord (ifnot integral to endoscope) and mat withno additional load.
FeaturePRO-LITE Sterilization Tray(proposed) K222440PRO-LITE Sterilization Tray(K183402)Comparison
The flexible endoscopes may containeither a single or dual lumen that is ≥ 1mm ID and ≤ 1050 mm in lengthLoad 2:• Non-lumened devices including non-lumened rigid, semi-rigid and flexibleendoscopes and non-lumened deviceswith diffusion-restricted spaces suchas the hinged portion of forceps andscissors• Single, dual or triple channel stainlesssteel lumens that are ≥ 0.48 mm IDand ≤ 100 mm in lengthThe flexible endoscopes may containeither:• a single lumen that is ≥ 1 mm ID and≤ 1050 mm in length• or two lumens with:○ one lumen that is ≥ 1 mm ID and< 990 mm in length○ and the other lumen that is ≥ 1mm ID and ≤ 850 mm in lengthLoad 2: Non-lumened instrumentsincluding instruments with diffusion-restricted areas such as the hingedportion of forceps or scissors.
V-PRO maX 2 Fast Non Lumen Cycle:Non-lumened devices including non-lumened rigid, semi-rigid and flexibleendoscopes and non-lumened deviceswith diffusion-restricted spaces such asthe hinged portion of forceps andscissors.V-PRO maX 2 Fast Non Lumen Cycle:Non-Lumened devices including non-lumened rigid, semi-rigid and flexibleendoscopes and non-lumened deviceswith stainless steel or titaniumdiffusion-restricted spaces such as thehinged portion of forceps and scissors
Intended Sterilization Cycles andIntended Tray Load for Tray Models:VP0045, VP0046, VP0047, VP0048,VP0049, VP0050, VP0051, VP0052Intended Sterilization Cycles andIntended Tray Load for Tray Models:VP0045, VP0046, VP0047, VP0048,VP0049, VP0050, VP0051, VP0052
STERRAD 100S Default Cycle:Metal and nonmetal medical devicesincluding instruments which havediffusion-restricted spaces, such as thehinged portion of forceps and scissors.Metal and nonmetal lumenedinstruments with• ≥ 6 mm ID and ≤ 310 mm in lengthMedical devices with a single stainlesssteel lumen with:• ≥ 1 mm ID and ≤ 125 mm in length• ≥ 2 mm ID and ≤ 250 mm in length• ≥ 3 mm ID and ≤ 400 mm in lengthSTERRAD 100S Default Cycle:Metal and nonmetal medical devicesincluding instruments which havediffusion-restricted spaces, such as thehinged portion of forceps and scissors.Metal and nonmetal lumenedinstruments with• ≥ 6 mm ID and ≤ 310 mm in lengthMedical devices with a single stainlesssteel lumen with:• ≥ 1 mm ID and ≤ 125 mm in length• ≥ 2 mm ID and ≤ 250 mm in length• ≥ 3 mm ID and ≤ 400 mm in length
Intended Sterilization Cycles andIntended Tray Load for Tray Models:VP0045, VP0046, VP0048, VP0049Intended Sterilization Cycles andIntended Tray Load for Tray Models:VP0045, VP0046, VP0048, VP0049
STERRAD NX and NX with ALLClearTechnology Standard Cycle:STERRAD NX and NX with ALLClearTechnology Standard Cycle:
FeaturePRO-LITE Sterilization Tray (proposed) K222440PRO-LITE Sterilization Tray (K183402)Comparison
Metal and non-metal medical devices including instruments which have diffusion-restricted spaces, such as the hinged portion of forceps and scissors. Medical devices with a single stainless steel lumen with:• $\geq$ 1 mm ID and $\leq$ 150 mm in length• $\geq$ 2 mm ID and $\leq$ 400 mm in lengthSTERRAD NX and NX with ALLClear Technology Standard Cycle:Metal and non-metal medical devices including instruments which have diffusion-restricted spaces, such as the hinged portion of forceps and scissors. Medical devices with a single stainless steel lumen with:• $\geq$ 1 mm ID and $\leq$ 150 mm in length• $\geq$ 2 mm ID and $\leq$ 400 mm in length
STERRAND NX and NX with ALLClear Technology Advanced Cycle:Metal and non-metal medical devices including instruments which have diffusion-restricted spaces, such as the hinged portion of forceps and scissors Medical devices, including most flexible endoscopes, with:o a single stainless steel lumen with:○ $\geq$ 1 mm ID and $\leq$ 500 mm in lengtho single channel polyethylene and Teflon (polytetrafluoroethylene)○ $\geq$ 1mm ID and $\leq$ 850 mm in lengthSTERRAND NX and NX with ALLClear Technology Advanced Cycle:Metal and non-metal medical devices including instruments which have diffusion-restricted spaces, such as the hinged portion of forceps and scissors Medical devices, including most flexible endoscopes, with:o a single stainless steel lumen with:○ $\geq$ 1 mm ID and $\leq$ 500 mm in lengtho single channel polyethylene and Teflon (polytetrafluoroethylene)• $\geq$ 1mm ID and $\leq$ 850 mm in length
Intended Sterilization Cycles and Intended Tray Load for Tray Models:VP0045, VP0046, VP0048, VP0049, VP0051, VP0052, VP0053Intended Sterilization Cycles and Intended Tray Load for Tray Models:VP0045, VP0046, VP0048, VP0049, VP0051, VP0052, VP0053
STERRAD 100NX and 100NX with ALLClear Technology Standard Cycle:Metal and nonmetal medical devices including instruments with have diffusion-restricted spaces, such as the hinged portion of forceps and scissors Medical devices with a single stainless steel lumen with:• $\geq$ 0.7 mm ID and $\leq$ 500 mm in lengthSTERRAD 100NX and 100NX with ALLClear Technology Standard Cycle:Metal and nonmetal medical devices including instruments with have diffusion-restricted spaces, such as the hinged portion of forceps and scissors Medical devices with a single stainless steel lumen with:$\geq$ 0.7 mm ID and $\leq$ 500 mm in length
STERRAD 100NX and 100NX with ALLClear Technology Flex Scope Cycle:Metal and nonmetal medical devices including instruments which have diffusion-restricted spaces, such as the hinged portion of forceps and scissors. Medical devices, including most flexible endoscopes withSTERRAD 100NX and 100NX with ALLClear Technology Flex Scope Cycle:Metal and nonmetal medical devices including instruments which have diffusion-restricted spaces, such as the hinged portion of forceps and scissors
FeaturePRO-LITE Sterilization Tray(proposed) K222440PRO-LITE Sterilization Tray(K183402)Comparison
• Single channel polyethylene andTeflon (polytetrafluoroethylene)○ ≥ 1mm ID and ≤ 850 mm in lengthMedical devices, including mostflexible endoscopes, with:• Single channel polyethylene andTeflon (polytetrafluoroethylene)○ ≥ 1mm ID and ≤ 850 mm in length
STERRAD 100NX and 100NX withALLClear Technology Express Cycle:Metal and nonmetal devices surfacesand instruments which have diffusion-restricted spaces, such as the hingedportion of forceps and scissors.STERRAD 100NX and 100NX withALLClear Technology Express Cycle:Metal and nonmetal devices surfacesand instruments which have diffusion-restricted spaces, such as the hingedportion of forceps and scissors.
STERRAD 100NX and 100NX withALLClear Technology Duo Cycle:Medical devices including:• most flexible endoscopes with asingle channel of polyethylene andTeflon (polytetrafluoroethylene) with$\geq$ 1 mm ID and ≤ 875 mm in length• accessory devices that are normallyconnected to a flexible endoscopeduring use• flexible endoscopes without lumensSTERRAD 100NX and 100NX withALLClear Technology Duo Cycle:Medical devices including:• most flexible endoscopes with asingle channel of polyethylene andTeflon (polytetrafluoroethylene) with$\geq$ 1 mm ID and ≤ 875 mm in length• accessory devices that are normallyconnected to a flexible endoscopeduring use• flexible endoscopes without lumens
Vent toVolumeRatioAll trays are the same:0.135 in-1All trays are the same:0.135 in-1Identical
TrayCompositionMineral-filled polypropylene, stainlesssteelMineral-filled polypropylene, stainlesssteelIdentical
InstrumentOrganizerCompositionMedical Grade Silicone, UPS grade VIMedical Grade Silicone, UPS grade VIIdentical
MatCompositionMedical Grade Silicone, UPS grade VIMedical Grade Silicone, UPS grade VIIdentical

{24}------------------------------------------------

{25}------------------------------------------------

{26}------------------------------------------------

{27}------------------------------------------------

{28}------------------------------------------------

{29}------------------------------------------------

Summary of Non-clinical Tests 12.

Performance testing to demonstrate substantial equivalence to the predicate has been completed and is summarized in Table 8 below.

TestResultConclusion
½ Cycle Verification ofMated SurfacesSterile efficacy was demonstrated for mated surfaces underworst case conditions in the V-PRO Sterilizer cycles.PASS
½ Cycle EfficacyThe standard injection weight resulted in all sterile resultswithin the validation load used to qualify each sterilizer cycle.PASS

Table 8. Performance Test Summary

{30}------------------------------------------------

PRO-ELITE Sterilization Trays
TestResultConclusion
Simulated Use TestSimulated use testing verified the ability of the sterilizer cyclesto sterilize medical devices under worst-case processingconditions.PASS
In Use TestThe in use investigation demonstrated the ability of the V-PROSterilizer cycles to sterilize patient-soiled, clinically-cleaned,medical instruments.PASS

7. Conclusion

Based on the intended us, technological characteristics and non-clinical performance data, the subject device is as safe, as effective and performs as well as or better than the legally marketed predicate device (K183402), Class II (21 CFR 880.6850), product code KCT.

§ 880.6850 Sterilization wrap.

(a)
Identification. A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.(b)
Classification. Class II (performance standards).