K Number
K222406

Validate with FDA (Live)

Device Name
Clarius AI
Date Cleared
2023-01-23

(167 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Clarius AI is intended to semi-automatically place calipers for non-invasive measurements of musculoskeletal structures (e.g., Achilles' tendon, plantar fascia, patellar tendon) on ultrasound data acquired by the Clarius Ultrasound Scanner (i.e., L 7 and L15). The user shall be a healthcare professional trained and qualified in MSK (musculoskeletal) ultrasound. The user shall retain the ultimate responsibility of ascertaining the measurements based on standard practices and clinical iudgment.

Device Description

Clarius AI is a radiological (ultrasound) image processing software application which implements artificial intelligence (Al), including non-adaptive machine learning algorithms, and is incorporated into the Clarius App software for use as part of the complete Clarius Ultrasound Scanner system product offering in musculoskeletal (MSK) ultrasound imaging applications. Clarius Al (MSK model) is intended for use by trained healthcare practitioners for non-invasive measurements of ultrasound data from musculoskeletal (MSK) ultrasound imaging acquired by the Clarius Ultrasound Scanner system using an artificial intelligence (AI) image segmentation algorithm. Clarius AI (MSK model) is intended to semi-automatically place adjustable calipers and provide supplementary information to the user regarding tendon thickness measurements (i.e., foot/plantar fascia, ankle/Achilles' tendon, knee/patellar tendon). Clarius Al is intended to inform clinical management and is not intended to replace clinical decision-making. The clinician retains the ultimate responsibility of ascertaining the measurements based on standard practices and clinical judgment. Clarius Al is indicated for use in adult patients only.

During the ultrasound imaging procedure, the anatomical site is selected through a preset software selection (e.g., foot, ankle, knee), in which the Clarius Al will engage to segment the correlating tendon. Clarius Al analyzes ultrasound images in real-time and outputs probabilities for each pixel within the image for determination of the particular tendon thickness.

The combination of all the pixels meeting a programmed threshold will render an overlay being displayed on top of the ultrasound image with a pre-programmed transparency so that the ultrasound greyscale is still visible. Once the user has obtained the best view, imaging can be manually paused, in which the Clarius Al will further analyze the tendon segmentation to determine the greatest thickness, in number of pixels, and subsequently place two measurement calipers that correspond to the top and bottom of the tendon at its thickest region, outputting a value in millimeters. The user can then manually alter the measurement calipers to make any necessary adjustments if desired. Clarius Al does not perform any functions that could not be accomplished manually by a trained and qualified user.

Clarius AI (MSK model) is incorporated into the Clarius App software and is intended for use with the following Clarius Ultrasound Scanner system transducers (previously 510(k)-cleared in K180799, K192107, and K213436):

Clarius Ultrasound Transducers: L7 and L15
Clarius App Software: Clarius Ultrasound App (Clarius App) for iOS; Clarius Ultrasound App (Clarius App) for Android

AI/ML Overview

The Clarius AI device is intended for semi-automatic placement of calipers for non-invasive measurements of musculoskeletal structures (e.g., Achilles' tendon, plantar fascia, patellar tendon) on ultrasound data. The device's performance was evaluated through non-clinical and clinical testing to demonstrate its safety and effectiveness.

Here’s a breakdown of the acceptance criteria and the study that proves the device meets them:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria for the Clarius AI device, particularly regarding measurement accuracy, are based on achieving non-inferiority to manual measurements made by expert clinicians. The specific criterion for clinical significance was defined as a difference of greater than 20% from normal thickness measurements.

Acceptance Criteria (Measurement Accuracy)Reported Device Performance
Automatic thickness measurement to be non-inferior to mean manual measurements.Non-inferior (p-value of 9.0 x 10^-5). The mean difference between automated and manual measurements was 0.03% (95% CI: -0.05% to -0.01%).
Clinically significant difference defined as >20% absolute percent difference between automatic measurement and mean reviewer measurement, corresponding to: - 0.6 mm for plantar fascia - 1 mm for patellar tendon - 1.2 mm for Achilles' tendon.The automatic measurement was found to be non-inferior to manual measurements within this clinically significant margin.
Automatic segmentation to have a high degree of overlap with ground truth.Average Dice score of 96% and mean IoU of 94% for tendon segmentation.

2. Sample Sizes and Data Provenance

  • Test Set (Validation Phase):

    • Sample Size: 73 subjects, resulting in a total of 2,503 ultrasound images/frames.
    • Data Provenance: Retrospective analysis of anonymized ultrasound images. Images were captured in-house from volunteer subjects and by clinical partners, mainly located in the USA. The data was anonymized and queried from Clarius Cloud storage.
  • Training Set:

    • Sample Size: A total of 20,287 images.
    • Data Provenance: Images were captured in-house from volunteer subjects and by clinical partners, mainly located in the USA. The data was anonymized and queried from Clarius Cloud storage.

3. Number of Experts and Qualifications for Ground Truth - Test Set

The document mentions "expert clinicians" and "licensed clinicians with relevant (i.e., musculoskeletal) ultrasound experience" were involved in establishing the ground truth for the test set (manual measurements). However, the exact number of experts and their specific qualifications (e.g., years of experience, board certification) are not explicitly stated in the provided text. The study states "reviewer pairs" for calculating manual measurement differences, implying at least two reviewers per case for comparison.

4. Adjudication Method for the Test Set

The provided text does not explicitly detail an adjudication method (e.g., 2+1, 3+1). It states that the "difference between auto-measurements and mean manual measurements" was compared to the "mean difference between manual measurements within the clinically significant margin." This suggests that multiple manual measurements were obtained, and their mean was used for comparison, rather than a specific consensus or adjudication process described.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

The provided text describes a verification test where "Clarius AI auto-measurements are non-inferior to manual measurements performed by licensed clinicians." While this compares AI performance against human performance, it doesn't describe a typical MRMC study designed to assess how human readers improve with AI assistance versus without AI assistance (i.e., an effect size of AI assistance on human readers). The focus was on the AI's standalone accuracy relative to human measurements.

6. Standalone Performance (Algorithm Only without Human-in-the-Loop)

Yes, a standalone performance evaluation was largely conducted for the AI algorithm. The core of the "Measurement Accuracy" testing compared the automatic thickness measurement directly against the mean manual measurements, indicating the algorithm's performance without specific human-in-the-loop assistance influencing the AI's initial output for that specific evaluation. The segmentation accuracy (Dice score and IoU) also represents standalone algorithm performance. While the AI semi-automatically places calipers and allows for manual adjustment, the non-inferiority study specifically evaluated the AI's auto-measurement against expert manual measurements.

7. Type of Ground Truth Used

  • Expert Consensus / Manual Measurements: For the measurement accuracy study, the ground truth was established by "mean manual measurements performed by licensed clinicians with relevant (i.e., musculoskeletal) ultrasound experience."
  • Expert Annotation (Segmentation): For the segmentation evaluation, the ground truth was described as "segmentation ground truth" where "the tendon regions in the images were annotated by a clinical scientist as the ground truth."

8. Sample Size for the Training Set

A total of 20,287 images were used for the training phase.

9. How the Ground Truth for the Training Set was Established

For the training phase, "the tendon regions in the images were annotated by a clinical scientist as the ground truth." This indicates that human experts (clinical scientists) manually outlined or labeled the tendon regions in the ultrasound images to create the reference data used to train the machine learning model.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION".

Clarius Mobile Health Corp. Agatha Szeliga Director, Regulatory Affairs 130-2985 Virtual Way Vancouver, British Columbia V5M 4X7 Canada

Re: K222406

January 23, 2023

Trade/Device Name: Clarius AI Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: QIH Dated: December 16, 2022 Received: December 21, 2022

Dear Agatha Szeliga:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jessica Lamb

Jessica Lamb, Ph.D. Assistant Director Imaging Software Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K222406

Device Name Clarius AI

Indications for Use (Describe)

Clarius AI is intended to semi-automatically place calipers for non-invasive measurements of musculoskeletal structures (e.g., Achilles' tendon, plantar fascia, patellar tendon) on ultrasound data acquired by the Clarius Ultrasound Scanner (i.e., L 7 and L15). The user shall be a healthcare professional trained and qualified in MSK (musculoskeletal) ultrasound. The user shall retain the ultimate responsibility of ascertaining the measurements based on standard practices and clinical iudgment.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

This 510(k) summary of safety and effectiveness information is submitted in accordance with the requirements of 21 CFR § 807.92.

Subject Device Trade Name:Clarius Al
Subject Device Model:MSK
Device Classification Name:Automated Radiological Image Processing Software

Regulation Number, Name and Product Code:

Regulation NumberRegulation NameProduct Code
21 CFR § 892.2050Medical Image Management and Processing SystemQIH
FDA 510(k) Review Panel:Radiology
Classification:Class II
Manufacturer:Clarius Mobile Health Corp.130-2985 Virtual WayVancouver, BC V5M 4X7 Canada
Contact Name:Agatha SzeligaDirector, Regulatory Affairsagatha.szeliga@clarius.com
Date 510(k) Summary Prepared:December 15, 2022

Predicate Device Information:

Device Trade Name:LVivo Software Application
510(k) Reference:K200232
Manufacturer Name:DiA Imaging Analysis Ltd
Regulation Name:Medical Image Management and Processing System
Device Classification Name:Automated Radiological Image Processing Software
Product Code(s):QiH
Regulation Number:21 CFR § 892.2050
Regulatory Class:Class II

Note: The predicate device has not been subject to a design-related recall.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the Clarius Ultrasound logo. The logo consists of an orange icon that resembles three concentric letter C's, followed by the word "clarius" in dark teal. To the right of "clarius" are the words "ultrasound anywhere" stacked on top of each other, also in dark teal.

Reference Device #1 Information:

Device Trade Name:CoLumbo
510(k) Reference:K220497
Manufacturer Name:Smart Soft Healthcare AD
Regulation Name:Medical Image Management and Processing System
Device Classification Name:Automated Radiological Image Processing Software
Product Code(s):QIH
Regulation Number:21 CFR § 892.2050
Regulatory Class:Class II

Reference Device #2 Information:

Device Trade Name:Al-Rad Companion (Musculoskeletal)
510(k) Reference:K193267
Manufacturer Name:Siemens Medical Solutions USA, Inc.
Regulation Name:Computed tomography x-ray system
Device Classification Name:Computed tomography x-ray system
Product Code(s):JAK
Regulation Number:21 CFR § 892.1750
Regulatory Class:Class II

Device Description

Clarius AI is a radiological (ultrasound) image processing software application which implements artificial intelligence (Al), including non-adaptive machine learning algorithms, and is incorporated into the Clarius App software for use as part of the complete Clarius Ultrasound Scanner system product offering in musculoskeletal (MSK) ultrasound imaging applications. Clarius Al (MSK model) is intended for use by trained healthcare practitioners for non-invasive measurements of ultrasound data from musculoskeletal (MSK) ultrasound imaging acquired by the Clarius Ultrasound Scanner system using an artificial intelligence (AI) image segmentation algorithm. Clarius AI (MSK model) is intended to semi-automatically place adjustable calipers and provide supplementary information to the user regarding tendon thickness measurements (i.e., foot/plantar fascia, ankle/Achilles' tendon, knee/patellar tendon). Clarius Al is intended to inform clinical management and is not intended to replace clinical decision-making. The clinician retains the ultimate responsibility of ascertaining the measurements based on standard practices and clinical judgment. Clarius Al is indicated for use in adult patients only.

During the ultrasound imaging procedure, the anatomical site is selected through a preset software selection (e.g., foot, ankle, knee), in which the Clarius Al will engage to segment the correlating tendon. Clarius Al analyzes ultrasound images in real-time and outputs probabilities for each pixel within the image for determination of the particular tendon thickness.

The combination of all the pixels meeting a programmed threshold will render an overlay being displayed on top of the ultrasound image with a pre-programmed transparency so that the ultrasound greyscale is still visible. Once the user has obtained the best view, imaging can be manually paused, in which the Clarius Al will further analyze the tendon segmentation to determine the greatest thickness, in number of

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the Clarius logo. The logo consists of an orange symbol that looks like three concentric letter C's, followed by the word "clarius" in dark teal. To the right of "clarius" are the words "ultrasound anywhere" stacked on top of each other, also in dark teal.

pixels, and subsequently place two measurement calipers that correspond to the top and bottom of the tendon at its thickest region, outputting a value in millimeters. The user can then manually alter the measurement calipers to make any necessary adjustments if desired. Clarius Al does not perform any functions that could not be accomplished manually by a trained and qualified user.

Clarius AI (MSK model) is incorporated into the Clarius App software and is intended for use with the following Clarius Ultrasound Scanner system transducers (previously 510(k)-cleared in K180799, K192107, and K213436):

Clarius Ultrasound TransducersL7 and L15
Clarius App SoftwareClarius Ultrasound App (Clarius App) for iOS;Clarius Ultrasound App (Clarius App) for Android

Indications for Use for Clarius AI (MSK)

Clarius AI is intended to semi-automatically place calipers for non-invasive measurements of musculoskeletal structures (e.g., Achilles' tendon, plantar fascia, patellar tendon) on ultrasound data acquired by the Clarius Ultrasound Scanner (i.e., L7 and L15). The user shall be a healthcare professional trained and qualified in MSK (musculoskeletal) ultrasound. The user shall retain the ultimate responsibility of ascertaining the measurements based on standard practices and clinical iudgment.

Comparison of the Subject Device and Legally Marketed Devices for Demonstration of Substantial Equivalence

The following table provides a comparison of the subject device, Clarius Al, to the predicate device and reference devices. A comparison of the subject device to the predicate and reference devices shows that the subject device has similar indications for use, is based on a similar Al algorithm, and provides automated radiological image processing with segmentation and measurement of anatomical structures, comparable to the legally marketed devices referenced herein.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the Clarius logo. The logo consists of an orange symbol that looks like three concentric letter C's, followed by the word "clarius" in dark teal. To the right of "clarius" are the words "ultrasound anywhere" stacked on top of each other, also in dark teal.

Table 1 - Comparison of the Subject Device to the Legally Marketed Devices

CriteriaSUBJECT DEVICEPREDICATE DEVICEREFERENCE DEVICE #1REFERENCE DEVICE #2RATIONALE(if subject devicediffers from predicatedevice)
Clarius AlLVivo SoftwareApplicationColumboAl-Rad Companion(Musculoskeletal)
510(k) Holder/ManufacturerClarius Mobile HealthCorp.DiA Imaging AnalysisLtdSmart Soft HealthcareADSiemens MedicalSolutions USA, Inc.Not applicable
Submission ReferenceCurrent SubmissionK200232K220497K193267Not applicable
Product Code(s)QIHQIHQIHJAKSame as predicatedevice and referencedevice #1
Device ClassificationNameAutomatedRadiological ImageProcessing SoftwareAutomatedRadiological ImageProcessing SoftwareAutomatedRadiological ImageProcessing SoftwareComputedtomography x-raysystemSame as predicatedevice and referencedevice #1
Regulation Number21 CFR 892.205021 CFR 892.205021 CFR 892.205021 CFR 892.1750Same as predicatedevice and referencedevice #1
Regulation NameMedical ImageManagement andProcessing SystemMedical ImageManagement andProcessing SystemMedical ImageManagement andProcessing SystemComputedtomography x-raysystemSame as predicatedevice and referencedevice #1
Intended UseNon-invasiveprocessing ofultrasound imagesusing automatic imagesegmentation andmeasurement ofanatomical structuresutilizing artificialintelligencealgorithms.Non-invasiveprocessing ofultrasound imagesusing automatic imagesegmentation andmeasurement ofanatomical structuresutilizing artificialintelligencealgorithms.Non-invasiveprocessing of MRimages usingautomatic imagesegmentation andmeasurement ofanatomical structuresutilizing artificialintelligencealgorithms.Non-invasiveprocessing of CTimages usingautomatic imagesegmentation andmeasurement ofanatomical structuresutilizing artificialintelligencealgorithms.Same as predicatedevice and similar toreference devices
Indications for UseClarius Al is intendedfor use by trainedhealthcareprofessionals to semi-automatically placeLVivo platform isintended for non-invasive processing ofultrasound images todetect, measure, andColumbo is an imagepost-processing andmeasurementsoftware tool thatprovides quantitativeAl-Rad Companion(Musculoskeletal) is animage processingsoftware that providesquantitative andClarius Al indicationsfor use are similar tothe predicate device'sindications for use asboth devices are
CriteriaSUBJECT DEVICEPREDICATE DEVICEREFERENCE DEVICE #1REFERENCE DEVICE #2RATIONALE (if subject device differs from predicate device)
Clarius AlLVivo Software ApplicationCoLumboAl-Rad Companion (Musculoskeletal)
calipers for non-invasive anatomical measurements on ultrasound data acquired by the Clarius Ultrasound Scanner.calculate relevant medical parameters of structures and function of patients with suspected disease.spine measurements from previously-acquired DICOM lumbar spine Magnetic Resonance (MR) images for users' review, analysis, and interpretation. It provides the following functionality to assist users in visualizing, measuring and documenting out-of-range measurements:-Feature segmentation;-Feature measurement;Threshold based labeling of out of range measurement; and-Export of measurement results to a written report for user's review, revise and approval. CoLumbo does not produce or recommend any type of medical diagnosis or treatment. Instead, it simply helps users toqualitative analysis from previously acquired Computed Tomography DICOM images to support radiologists and physicians from emergency medicine, specialty care, urgent care, and general practice in the evaluation and assessment of musculoskeletal disease. It provides the following functionality:-Segmentation of vertebrae-Labelling of vertebrae-Measurements of heights in each vertebra and indication if they are critically different-Measurement of mean Hounsfield value in volume of interest within vertebra. Only DICOM images of adult patients are considered to be valid input.indicated for non-invasive processing of ultrasound data/ images for measurements of anatomical structures.
CriteriaSUBJECT DEVICEPREDICATE DEVICEREFERENCE DEVICE #1REFERENCE DEVICE #2RATIONALE(if subject devicediffers from predicatedevice)
Clarius AlLVivo SoftwareApplicationCoLumboAl-Rad Companion(Musculoskeletal)
more easily identifyand classify features inlumbar MR images andcompile a report. Theuser is responsible forconfirming/ modifyingsettings, reviewing andverifying the software-generatedmeasurements,inspecting out-of-range measurements,and approving draftreport content usingtheir medicaljudgment anddiscretion.The device is intendedto be used only byhospitals and othermedical institutions.Only DICOM images ofMRI acquired fromlumbar spine exams ofpatients aged 18 andabove are consideredto be valid input.Columbo does notsupport DICOM imagesof patients that arepregnant, undergoMRI scan with contrastmedia, or have post-
CriteriaSUBJECT DEVICEPREDICATE DEVICEREFERENCE DEVICE #1REFERENCE DEVICE #2RATIONALE(if subject device
Clarius AlLVivo SoftwareApplicationCoLumboAl-Rad Companion(Musculoskeletal)differs from predicatedevice)
Radiologicalapplication/ SupportedmodalityUltrasoundUltrasoundoperationalcomplications,scoliosis, tumors,infections, fractures.CTSame as predicatedevice
Principle of Operation/TechnologyUltrasound imageprocessing softwareimplementing artificialintelligence includingnon-adaptive machinelearning algorithmstrained with clinicaland/or artificial dataintended for non-invasive segmentationand measurements ofultrasound data.Ultrasound imageprocessing softwareimplementing artificialintelligence includingnon-adaptive machinelearning algorithmstrained with clinicaland/or artificial dataintended for non-invasive segmentationand measurements ofultrasound data.Radiological imageprocessing softwareimplementing artificialintelligence includingnon-adaptive machinelearning algorithmstrained with clinicaland/or artificial dataintended for non-invasive segmentationand measurements ofMR dataRadiological imageprocessing softwareimplementing artificialintelligence includingnon-adaptive machinelearning algorithmstrained with clinicaland/or artificial dataintended for non-invasive segmentationand measurements ofCT dataSame as predicatedevice
SegmentationYes - Segmentation ofanatomical structures(tendons)Yes - Segmentation ofanatomical structures(LV, RV, bladder)Yes - Segmentation ofanatomical structures(spinal vertebrae)Yes - Segmentation ofanatomical structures(spinal vertebrae)Similar to predicatedevice and referencedevices
MeasurementYes - Measurement ofanatomical structures(tendons)Yes - Measurement ofanatomical structures(LV, RV, bladder)Yes - Measurement ofanatomical structures(spinal vertebrae)Yes - Measurement ofanatomical structures(spinal vertebrae)Similar to predicatedevice and referencedevices
Al AlgorithmImage segmentation,anatomicalidentification, andmeasurement utilizingthe machine learning-based algorithmException U-NetImage segmentationfor border detection.Algorithm combinesimage processing andDeep Learning NeuralNetwork for analysis.Deep ConvolutionalImage-to-Image NeuralNetwork3D Deep Image-to-Image NetworkSimilar to predicatedevice and referencedevices
CriteriaSUBJECT DEVICEPREDICATE DEVICEREFERENCE DEVICE #1REFERENCE DEVICE #2RATIONALE(if subject devicediffers from predicatedevice)
Clarius AlLVivo SoftwareApplicationCoLumboAl-Rad Companion(Musculoskeletal)
Automation(Yes or No)YesYesYesYesSame as predicatedevice and referencedevices
Display CalipersYesYesYesYesSame as predicatedevice and referencedevices
Manual adjustment/Manual editing by usercapability(Yes or No)YesYesYesYesSame as predicatedevice and referencedevices
Anatomical SiteFoot, Ankle, KneeBladder, HeartLumbar SpineThoracic SpineAlthough theanatomical sites/structures for use ofthe predicate device(bladder, heart) aredifferent from thesubject device (foot,ankle, knee), thepredicate device andsubject device share avery similar intendeduse in terms ofidentifying/ viewing,measuring/quantifying andreporting resultsacquired by ultrasounddevices for non-invasivemeasurements ofanatomical structuresutilizing artificial
CriteriaSUBJECT DEVICEPREDICATE DEVICEREFERENCE DEVICE #1REFERENCE DEVICE #2RATIONALE(if subject device
Clarius AlLVivo SoftwareApplicationCoLumboAl-Rad Companion(Musculoskeletal)differs from predicatedevice)
intelligencealgorithms.
Environment of UseHealthcare setting(e.g., hospital, clinic)Healthcare setting(e.g., hospital, clinic)Healthcare setting(e.g., hospital, clinic)Healthcare setting(e.g., hospital, clinic)Same as predicatedevice and referencedevices
Intended UsersLicensed healthcareprofessionalsLicensed healthcareprofessionalsLicensed healthcareprofessionalsLicensed healthcareprofessionalsSame as predicatedevice and referencedevices
Patient PopulationAdultsAdultsAdultsAdultsSame as predicatedevice and referencedevices

{7}------------------------------------------------

ound C (

{8}------------------------------------------------

asound y anywhere

{9}------------------------------------------------

sound y vhere

{10}------------------------------------------------

sound y where

{11}------------------------------------------------

C C rasound anywhere

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image shows the Clarius logo. The logo consists of an orange icon on the left and the word "clarius" in dark teal on the right. To the right of the word "clarius" is the phrase "ultrasound anywhere" in a smaller font, also in dark teal. The icon on the left is a stylized "C" made up of three curved lines.

Non-Clinical Performance Testing Summary

Clarius Al (MSK) was designed and developed by Clarius Mobile Health Corp. in accordance with the applicable requirements, design controls, and standards to establish safety and effectiveness of the device.

Non-clinical performance testing has demonstrated that Clarius Al (MSK) complies with the following FDArecognized consensus standards:

StandardRecognitionNumberTitle of Standard
13-79IEC 62304:2006 + A1:2015 - Medical device software — Software life cycle processes
5-40ISO 14971:2019 Medical devices — Application of risk management to medical devices
12-300NEMA PS 3.1 - 3.20 (2016) Digital Imaging and Communications in Medicine (DICOM) Set
5-114IEC 62366-1:2015 + A1:2020 Medical devices — Part 1: Application of usability engineering tomedical devices
5-117ISO 15223-1:2016 Medical devices — Symbols to be used with medical device labels, labellingand information to be supplied

Safety and performance of Clarius Al (MSK) has been evaluated through verification and validation testing in accordance with software specifications and applicable performance standards. The traceability analysis provides traceability between requirements, design specifications, risks, and verification testing of the subject device. All software requirements and risk analysis have been successfully verified and traced. Software verification and validation activities were conducted per IEC 62304:2006 + AMD1:2015 – Medical device software – Software lifecycle processes and ISO 14971:2019 Medical devices – Application of risk management to medical devices, and in accordance with FDA guidance documents, General Principles of Software Validation, Final Guidance for Industry and FDA Staff (issued January 11, 2002), Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (issued May 11, 2005), and Content of Premarket Submissions for Management of Cybersecurity in Medical Devices (issued October 2, 2014).

Applicable software documentation for a Moderate Level of Concern software, per FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (May 11, 2005) has been provided.

A comprehensive risk analysis was performed for the subject device and appropriate risk controls have been implemented to mitigate hazards.

Clarius conforms to the cybersecurity requirements by implementing a process of preventing unauthorized access, modifications, misuse or denial of use, or the unauthorized use of information that is stored, accessed or transferred from a medical device to an external recipient.

Clarius Al (MSK) was tested and found to be safe and effective for the intended users, uses, and use environments, as demonstrated through verification and validation testing was performed to ensure that the final product is capable of meeting the requirements for the specified clinical application and performs as intended to meet users' needs, while demonstrating substantial equivalence to the predicate device.

{13}------------------------------------------------

Image /page/13/Picture/0 description: The image shows the Clarius logo. On the left is an orange symbol that looks like three curved lines that form a C shape. To the right of the symbol is the word "clarius" in dark teal. To the right of "clarius" are the words "ultrasound anywhere" stacked on top of each other, also in dark teal.

Clinical Performance Testing Summary

Clarius performed Al model training to create a documented baseline of the Al model. Training consisted of a basic hyperparameter exploration over a static dataset optimizing towards a static validation set sampled from the training data. U-net with MobileNetV2 backbone was used as the model architecture for Clarius AI.

Three datasets of ultrasound images of Achilles' Tendons, Plantar Fascia and Patellar Tendons, were used in the study. All the images had segmentation ground truth. The images in the dataset were classified into tendon and background. Images were captured in-house from volunteer subjects and by clinical partners in outpatient clinics mainly located in the USA. The data was anonymized and queried from Clarius Cloud storage. Ultrasound recordings were acquired with the subjects in various positions depending on preference. Tendons were imaged in longitudinal views. The plantar fascia was imaged in dorsiflexion, plantar flexion, and neutral positions. The Achilles' tendon was imaged with the ankle at 90 degrees. The knee was imaged in 30 degrees flexion. Images with and without the structures of interest were collected to create a dataset that was large and heterogeneous. Images were acquired in different depths to mitigate different use cases for users. The data were acquired using two Clarius ultrasound linear probes with a frequency of 5 - 15 MHz and 4 - 13 MHz. A total of 20,287 images were acquired. For the training phase, the tendon regions in the images were annotated by a clinical scientist as the ground truth.

A total of 73 subjects were used to evaluate the performance of the Al model in the validation phase. Each subject had multiple studies of the above-mentioned tendons, resulting in a total of 2,503 ultrasound images/frames. The reported average Dice score was 96% and the mean loU was 94% for tendon segmentation.

Verification testing was performed to verify if Clarius Al auto-measurements are non-inferior to manual measurements performed by licensed clinicians with relevant (i.e., musculoskeletal) ultrasound experience. Verification testing of Clarius Al was performed through a retrospective analysis of anonymized ultrasound images. The objective was to determine whether the Clarius Al (MSK model) software measurement output adequately aligned with expert clinicians' manual caliper placement. The features tested during the verification study were tendon thickness measurement and segmentation mask. The difference between auto-measurements and mean manual measurements was found to be no greater than the mean difference between manual measurements within the clinically significant margin and with a statistical significance level of 0.05.

Measurement Accuracy:

The absolute percent (%) difference between reviewer pairs was calculated and compared to the absolute percent (%) difference between the automatic thickness measurement and mean reviewer measurement using a one-sided t-test and an equivalence margin of 20%. The automatic thickness measurement was found to be non-inferior (p value of 9.0 x 10'). The mean difference between differences was 0.03% (95% CI (-0.05)-(-0.01)).

A difference of >20% constitutes a clinically significant difference for all structures, which corresponds to the following differences based on normal thickness measurements: 0.6 mm for plantar fascia; 1 mm for patellar tendon; 1.2 mm for Achilles' tendon.

{14}------------------------------------------------

Image /page/14/Picture/0 description: The image shows the Clarius Ultrasound logo. The logo consists of an orange symbol resembling three curved lines forming a C shape, followed by the word "clarius" in dark teal. To the right of "clarius" is the text "ultrasound anywhere" also in dark teal, with "ultrasound" stacked above "anywhere".

Verification testing of Clarius Al was successfully completed with the results demonstrating that Clarius Al (MSK model) software measurement output adequately aligned with expert clinicians' manual caliper placement. As such, Clarius Al has been verified for tendon thickness measurement and segmentation mask for use in musculoskeletal ultrasound applications.

A design validation study was performed to validate the essential functionality of Clarius Al as incorporated into the Clarius App software for the specified application and intended use in automatic segmentation highlight, semi-automatic caliper placement with manual caliper adjustment, and persistent storage, to determine if Clarius Al is clinically usable and meets user needs in musculoskeletal ultrasound applications. The results of the validation study demonstrated that Clarius as intended and meets user needs in musculoskeletal ultrasound applications for use in automatic segmentation highlight, semi-automatic caliper placement with manual caliper adjustment, and persistent storage.

Conclusion & Summary of Substantial Equivalence

Based on the information presented in this Traditional 510(k) premarket notification and based on the fundamental scientific technology utilizing artificial intelligence algorithms, technological characteristics, principle of operation, intended use, environment of use, and indications for use, Clarius Al has been determined to be substantially equivalent in terms of safety and effectiveness to the legally marketed predicate/reference devices.

Performance testing of Clarius Al, including results from verification studies, has demonstrated that Clarius Al software measurement output adequately aligns with expert clinicians' manual measurements, and thereby performs as intended for automatic segmentation with semiautomatic caliper placement with manual caliper adjustment in musculoskeletal ultrasound applications, meeting user needs.

The subject device and the predicate device are both radiological (ultrasound) image processing software applications which implement artificial intelligence (Al) including non-adaptive learning algorithms trained with clinical and/or artificial data intended for non-invasive measurements of ultrasound data, utilizing similar machine-learning based algorithms to detect, measure, and calculate relevant medical parameters of structures with manual adjustment capability by the user.

Any minor differences in the indications for use/anatomical structures between the subject device and the legally marketed devices do not raise any issues related to safety or effectiveness, thereby demonstrating that Clarius Al is as safe and effective as the legally marketed devices, and therefore substantially equivalent.

N/A