K Number
K222274

Validate with FDA (Live)

Device Name
FONA XDC
Manufacturer
Date Cleared
2022-08-23

(25 days)

Product Code
Regulation Number
872.1800
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

FONA XDC is an intraoral dental X-ray device intended for dental radiographic examination and diagnosis of diseases related to the anatomical structures of the teeth in both adult and pediatric patients.

Device Description

The device consists of an X-ray generator, a CPU that manages the exposure time and a mechanical suspension system, for wall mounting or mobile stand.
The image detectors (a necessary component for a fully-functional diagnostic system) are not part of the current submission.
The device, unlike the previous IntraOs 70 with AC power supply of the tube, is powered at constant potential, i.e. direct current, guaranteeing excellent quality of the final image. From an application point of view, FONA XDC radiographic system is similar to the IntraOs 70; it is possible to select nine anatomical zones (two more than in IntraOs 70), two patient body sizes and three types of receptors that can be set at different sensitivities.
The basic radiographic system allows to operate at 20 cm (8") source-skin distance (SSD) with circular radiation beam. The FONA XDC primary collimator, as for the IntraOs 70, consists of a brass cylinder and the limitation of the beam on a circular surface of 6 cm in diameter from the focus for a working distance of 20 cm. It is possible to bring the working distance to 30 cm from the focus by adding a cone extension. It is also possible to reduce the exposed circular area both by working at 20 cm and at 30 cm. with the interposition of rectanqular 3x4 cm or 2x3 cm BLD adaptor, for image receptors of size 2 (adult) or size 0 (child) respectively

AI/ML Overview

The provided text is a 510(k) summary for a dental X-ray device (FONA XDC) seeking substantial equivalence to a predicate device. It details the device's technical specifications and compares them to the predicate, but it does not include any information about studies proving the device meets acceptance criteria related to AI/algorithm performance.

The document explicitly states: "Based on the device nature (an x-ray generator similar to the predicate), clinical testing is not required to demonstrate substantial equivalence. Successful bench testing results should be enough proof that the FONA XDC works as intended."

Therefore, the following information cannot be extracted from the provided text:

  1. A table of acceptance criteria and the reported device performance (for an AI/algorithm-based device): Not applicable, as this is an X-ray generator, not an AI device.
  2. Sample size used for the test set and the data provenance: Not applicable.
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable.
  4. Adjudication method: Not applicable.
  5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done: Not applicable.
  6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not applicable.
  7. The type of ground truth used: Not applicable.
  8. The sample size for the training set: Not applicable.
  9. How the ground truth for the training set was established: Not applicable.

The document focuses on demonstrating substantial equivalence of the FONA XDC to a predicate device (IntraOs 70) and a reference device (Heliodent Plus) based on technical specifications, intended use, and adherence to performance standards for X-ray generators. The table provided in the document (pages 6-7) compares various technical characteristics like power supply, line voltage, focal spot, tube voltage, tube current, and adherence to electrical safety and radiation protection standards, which would serve as "acceptance criteria" for an X-ray generator.

Summary of Non-Clinical Testing and Conclusion:
The document mentions that "FDA consensus standards have been employed for electrical safety, electromagnetic compatibility, performance and usability. Each produced device is checked against the FDA performance standards for Ionizing radiation emitting products. The performance of the predicate device and the subject device have been validated using the same testing models." This implies that the device underwent bench testing to ensure it met these standards, and these standards serve as the acceptance criteria for the X-ray generator.

{0}------------------------------------------------

August 23, 2022

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

FONA S.r.l % Krupa Srivastava Regulatory Consultant Via G.Galilei 11 Assago, MILAN 20057 ITALY

Re: K222274

Trade/Device Name: FONA XDC Regulation Number: 21 CFR 872.1800 Regulation Name: Extraoral source x-ray system Regulatory Class: Class II Product Code: EHD Dated: July 29, 2022 Received: July 29, 2022

Dear Krupa Srivastava:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Laurel Burk, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K222274

Device Name FONA XDC

Indications for Use (Describe)

FONA XDC is an intraoral dental X-ray device intended for dental radiographic examination and diagnosis of diseases related to the anatomical structures of the teeth in both adult and pediatric patients.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for FONA, a company that is "Closer to You". The logo consists of the word "FONA" in large, gray, sans-serif letters. Above the word "FONA" is a curved, yellow line. Below the word "FONA" is the phrase "CLOSER TO YOU" in smaller, gray, sans-serif letters.

510(k) SUMMARY

K222274

Date Prepared:August 10, 2022
Submitters Information
NameFONA s.r.l.
AddressVia G.Galilei 1120057 Assago (MI), Italy
Contact PersonLuigi Germanò
Contact Telephone+39 0245712171
Device Information
Device/ Trade NameFONA XDC
Common NameUnit, X-Ray, Extraoral With Timer
Classification NameExtraoral source x-ray system
Classification Regulation21 CFR 872.1800
Product codeEHD
Classification2

Predicate Device:

The predicate device for the FONA XDC along with its 510(k) number is provided below:

Trade NameIntraOs 70
510(k) NumberK031118
Classification NameExtraoral source x-ray system
Classification Regulation21 CFR 872.1800
Product codeEHD

Special 510(k)- FONA XDC FONA S.r.I

FONA S.r.l. - Sede legale Via Victor Hugo 4, 20123 Milano (M) (Ml) Italia - Galileo Gallie, 11 - 20057 Assago
(MI) Italia - Capitale Sociale 10.000 Euro i.v. - Codi 12380490966 - REA: MI - 2657690 - Telefono: +39 02 45712171 - Fax: +39 02 45703385 - E-mail: info@fonaitaly.com - Web: www.fonadental.it

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for FONA, a company that is "Closer to You". The logo consists of the word "FONA" in large, gray, sans-serif letters. Above the word "FONA" is a yellow swoosh. Below the word "FONA" are the words "CLOSER TO YOU" in smaller, gray, sans-serif letters.

Classification2
-------------------

Reference Device:

The reference device for the FONA XDC along with its 510(k) number is provided below:

Trade NameHeliodent Plus
510(k) NumberK083344
Classification NameExtraoral source x-ray system
Classification Regulation21 CFR 872.1800
Product codeEHD
Classification2

Indications for use

FONA XDC is an intraoral dental X-ray device intended for dental radiographic examination and diagnosis of diseases related to the anatomical structures of the teeth in both adult and pediatric patients.

Device Description:

The device consists of an X-ray generator, a CPU that manages the exposure time and a mechanical suspension system, for wall mounting or mobile stand.

The image detectors (a necessary component for a fully-functional diagnostic system) are not part of the current submission.

The device, unlike the previous IntraOs 70 with AC power supply of the tube, is powered at constant potential, i.e. direct current, guaranteeing excellent quality of the final image. From an application point of view, FONA XDC radiographic system is similar to the IntraOs 70; it is possible to select nine anatomical zones (two more than in IntraOs 70), two patient body sizes and three types of receptors that can be set at different sensitivities.

The basic radiographic system allows to operate at 20 cm (8") source-skin distance (SSD) with circular radiation beam. The FONA XDC primary collimator, as for the IntraOs 70, consists of a brass cylinder and the limitation Special 510(k)- FONA XDC FONA S.r.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for FONA. The word "FONA" is written in large, gray, sans-serif letters. Above the word is a yellow swoosh. Below the word "FONA" is the phrase "CLOSER TO YOU" in smaller, gray, sans-serif letters.

of the beam on a circular surface of 6 cm in diameter from the focus for a working distance of 20 cm. It is possible to bring the working distance to 30 cm from the focus by adding a cone extension. It is also possible to reduce the exposed circular area both by working at 20 cm and at 30 cm. with the interposition of rectanqular 3x4 cm or 2x3 cm BLD adaptor, for image receptors of size 2 (adult) or size 0 (child) respectively

Substantial Equivalence:

We believe that the FONA XDC, is substantially equivalent to the legally marketed device (the predicate device) in terms of safety and effectiveness.

The primary predicate is the FONA XDC as it is a further development to the IntraOs 70 and performs similar operations and functions.

Both the current FONA XDC and the cleared IntraOs 70 are extra oral source X-ray systems intended for dental radiographic examination and diagnosis of diseases of the teeth.

Both the IntraOs 70 and the FONA XDC are Intraoral dental X-ray devices to be used together with a proper receptor. The medical device is intended to be used in hospitals or in medical dental centers by radiology doctors, dentists and qualified staff who have received proper training. They both can be configured in the wall mounted as well as the mobile versions. The mechanical mobile stand base, the scissor arms and support arms are identical.

Although worded differently, there are no differences between the subject device and the predicate device with respect to indications and intended use.

S.r.l. - Sede legale Via Victor Hugo 4, 20123 Milano (M) Italia - Sede operativa: Via Galileo Galilei, 11 - 20057 Assago Capitale Sociale 10.000 Euro i.v. - Codice Fiscale, Partita I.V.A. e Iscrizione Registro Imprese di Milano: 80490966 - REA: MI - 2657690 - Telefono: +39 02 45712171 - Fax: +39 02 45703385 - E-mail: info@fonaitaly.com - Web www.fonadental.it

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for FONA, a company that is "Closer to You". The logo consists of the word "FONA" in large, gray, sans-serif letters, with the tagline "CLOSER TO YOU" in smaller, gray letters underneath. Above the word "FONA" is a curved, yellow line that resembles a smile or an arc.

Comparison of the subject modified device to the cleared predicate device

DevicePredicate DeviceReference Device
Device NameFONA XDCIntraOs 70Heliodent Plus
Manufacturer NameFONA S.r.lFONA S.r.I (Formerly known as BlueX Imaging S.r.l)Sirona Dental Systems GmbH
Product IllustrationImage: FONA XDCImage: IntraOs 70Image: Heliodent Plus
Device Classification Regulation21 CFR 872.180021 CFR 872.180021 CFR 872.1800
Regulation/Classification NameExtraoral source x-ray systemExtraoral source x-ray systemExtraoral source x-ray system
Common NameUnit, X-Ray, Extraoral with TimerUnit, X-Ray, Extraoral with TimerUnit, X-Ray, Extraoral with Timer
Product Code(s)EHDEHDEHD
510(k)K222274K031118K083344
DevicePredicate DeviceReference Device
Device NameFONA XDCIntraOs 70Heliodent Plus
Intended Use / Indications for UseFONA XDC is an intraoral dental X-ray device intended for dentalradiographic examination anddiagnosis of diseases related to theanatomical structures of the teeth inboth adult and pediatric patients.The IntraOs 70 (with Autoset Timer) isintended for the dental radiographicexamination and diagnosis ofdiseases related to the anatomicalstructures of the teeth. Such a devicemakes use of an extra oral source x-ray system commonly referred to asintraoral x-ray equipment.The Heliodent Plus is an X-ray devicefor established dental surgeries andclinics intended to be used forintraoral radiography for examinationand diagnosis of diseases of theteeth, jaw, and oral structures.
Power SupplyDCACDC
Line Voltage110-127 V ± 10%220-240 V ± 10%110-120 V (from 99 V to 132 V)220-24 OV (from 198 V to 264 V)120 V ± 10%200-240 V ± 10%
Line FuseT 8AH at 110-127 VT 5AH at 220-240 V6.3 A at 115 V4 A at 230 V16 A slow blow
Line Frequency50-60 Hz50/60 Hz ± 1 Hz50/60 Hz
Line Resistance≤ 0.5 Ohm at 110-127 V≤ 1.0 Ohm at 220-240 V≤ 0.4 Ohm at 115 V≤ 0.8 Ohm at 230 V0.3 Ohm at 120 V0.8 Ohm at 200-240 V
Maximum Line Current8 A at 110-127 V5 A at 220-240 V6 A at 120 V4 A at 230 V10 A at 120 V6-5 A at 200-240 V
Focal Spot0.4 as per IEC 603360.8 as per IEC 603360.4 as per IEC 60336
Inherent Filtration≥ 2.3 mm Al at 70 kV> 2.5 mm Al at 70 kVp> 1.5 mm Al at 70
Tube Voltage60 or 70 kV ± 5% selectable70 kVp ± 8% at nominal line voltage66 kVp ± 8% at nominal line voltage -10%74 kVp ± 8% at nominal line voltage+10%60 kV / 70 kV switchable(max. tolerance ± 5 kV)
Tube Current7 mA ± 10%7.0 mA ± 15% at nominal line voltage5.3 mA ± 15% at nominal line voltage –10%8.3 mA ± 15% at nominal line voltage+ 10%7 mA (max. tolerance ± 1.4 mA)
DevicePredicate DeviceReference Device
Device NameFONA XDCIntraOs 70Heliodent Plus
Anatomical AreasUpper incisor•Maxillary incisor•Maxillary front tooth•
Upper cuspid / premolar•Maxillary canine or premolar•Maxillary canine / premolar•
Upper molar,•Maxillary molar•Maxillary molar•
Occlusal upper / lower arch•Mandibular incisor•Bite-wing exposure•
Occlusal premolar crowns•Mandibular canine or premolar•Mandibular front tooth•
Occlusal molar crowns•Mandibular molar•Mandibular canine / premolar•
Lower incisor•Bite-wing premolar•Mandibular molar•
Lower cuspid / premolar•
Lower molar•
Image receptorsFilm, Phosphor plate, digital sensorFilm and digital sensorFilm and digital sensor
Patient populationAdult and ChildAdult and ChildAdult and Child
Exposure time0.01-3.2 s ± 5%0.06-3.2 s ± 5%0.01 – 3.2 s
R20 scaleR10 scale(max. tolerance ± 10% +1 ms)
Principle of OperationThe device consists of an X-rayThe device consists of an X-rayThe device consists ofan X-ray
generator, a CPU that manages thegenerator, a CPU that manages thegenerator, a CPU that manages the
exposure time and a mechanicalexposure time and a mechanicalexposure time and a mechanical
suspension system, for wall mountingsuspension system, for wall mountingsuspension system, for wall mounting
or mobile stand.or mobile stand.or mobile stand.
The device is powered at constantIntraOs 70 is a traditional AC tube-The device is powered at constant
potential, i.e. direct current.head, 70 kVp, 7 mA.potential, i.e. direct current.
Xray exposure time controlMicroprocessor controlledMicroprocessor controlledMicroprocessor controlled
SterilizationNot Sterile, disinfect and useNot Sterile, disinfect and useNot Sterile, disinfect and use
Operating ConditionsTemperature 10°C to 40°CTemperature 10°C to 40°CAmbient temperature +10°C - +35°C
Humidity 30 to 75%Humidity 30 to 75%With room temperatures > 35°C (>
Pressure 700 to 1060 hPaPressure 700 to 1060 hPa95°F)
Dentsply Sirona recommends the use
of
an air conditioning system.
Relative humidity: 30% - 85% (no
condensation)
DevicePredicate DeviceReference Device
Device NameFONA XDCIntraOs 70Heliodent Plus
Performance StandardsIEC 60601-1 (Electrical Safety)IEC 60601-1 (Electrical Safety)IEC 60601-1 (Electrical Safety)
IEC 60601-1-2 (EMC)IEC 60601-1-2 (EMC)IEC 60601-1-2 (EMC)
IEC 60601-1-3 (Radiation Protection)IEC 60601-1-3 (Radiation Protection)IEC 60601-1-3 (Radiation Protection)
IEC 60601-2-65 (Performance)IEC 60601-2-65 (Performance)IEC 60336 (Focal Spots)
IEC 60336 (Focal Spots)IEC 60336 (Focal Spots)IEC 60601-2-65 (performance)
IEC 62304 (Software)IEC 62304 (Software)
IEC 60601-1-6 & IEC 62366-1(Usability)IEC 60601-1-6 & IEC 62366-1(Usability)
IEC 61223-3-4 (performance)IEC 61223-3-4 (performance)
ISO 10993-1 (Biocompatibility)ISO 10993-1 (Biocompatibility)
21CFR1020.30 & 21CFR 1020.3121CFR1020.30 & 21CFR 1020.31

Page 4 of 8

Ill FON Sci. - Sede legale Via Victor Hugo 4, 20123 Millio Galler Callier Callier Catile 10,00 Euro in - Codice Sociale 10,00 Euro i. - Codice Fiscale, Partile III. - Corizio Registro Imprese di Miano: 12380490966 - REA: NI - 2657690 - Telefon: +39 02 45703385 - E-mail: info@fonaitay.com - Web: www.fonadental.i.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for FONA, a company whose slogan is "Closer to You". The logo consists of the word "FONA" in large, gray, sans-serif letters. Above the word "FONA" is a yellow swoosh. Below the word "FONA" are the words "CLOSER TO YOU" in smaller, gray, sans-serif letters.

Page 5 of 8

ll FONA S.r.I. - Sece legale Via Victor Hugo 4, 2012 Milen Callie, 11 - 2005 Assap (M) Jalia - Capitele Sociale 10,000 Euro i. - Codize Fisale, Partid V.A. e sozizine Registro Imprese di Miano: 1230049096 - REA: M - 2657690 - Telefon: +39 02 45703385 - E-mail: info@fonaitaly.com - Web: www.fonadental.i.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for FONA. The logo consists of the word "FONA" in large, gray, sans-serif letters. Above the word "FONA" is a curved, yellow line. Below the word "FONA" are the words "CLOSER TO YOU" in smaller, gray, sans-serif letters.

Page 6 of 8

■ FONA S.r.l. - Sede legale Via Vidia - Sede cperatia: Va Galles Galle, 11 - 2007 Assago (M) Jalia - Copitele Sociale 10.00 Euro i.r. - Codie Fisale, Partia V.A. e sozizn Registro Imprese di Miano: 1230049096 - REA: M - 2657690 - Telefon: +39 02 45703385 - E-mail: info@fonaitaly.com - Web: www.fonadental.i.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for FONA, a company whose slogan is "Closer to You". The logo consists of the word "FONA" in large, bold, gray letters. Above the word "FONA" is a yellow arc. Below the word "FONA" are the words "CLOSER TO YOU" in smaller, gray letters.

Page 7 of 8

ll FONA S.r.I. - Sece legale Via Victor Hugo 4, 2012 Milen Callie, 11 - 2005 Assap (M) Jalia - Capitele Sociale 10,000 Euro i. - Codize Fisale, Partid V.A. e sozizine Registro Imprese di Miano: 1230049096 - REA: M - 2657690 - Telefon: +39 02 45703385 - E-mail: info@fonaitaly.com - Web: www.fonadental.i.

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the logo for FONA. The word "FONA" is written in large, gray, sans-serif letters. Above the word is a yellow swoosh. Below the word "FONA" is the phrase "CLOSER TO YOU" in smaller, gray, sans-serif letters.

Summary of Non Clinical Testing:

The device is an evolution of the predicate device based on experience of the predicate device from the field and in compliance with the state of the art in the dental imaging area.

FDA consensus standards have been employed for electrical safety, electromagnetic compatibility, performance and usability. Each produced device is checked against the FDA performance standards for lonizing radiation emitting products. The performance of the predicate device and the subject device have been validated using the same testing models.

Based on the device nature (an x-ray generator similar to the predicate), clinical testing is not required to demonstrate substantial equivalence. Successful bench testing results should be enough proof that the FONA XDC works as intended.

Conclusion:

The proposed device has the same intended use and principles of operation as the predicate device.

There are no significant changes to the materials, dimensions or to the assembly process of the device. No new biocompatibility testing is deemed to be required as compared to the predicate device.

In conclusion, the FONA XDC is as safety and effective as the predicate device.

Special 510(k)- FONA XDC

§ 872.1800 Extraoral source x-ray system.

(a)
Identification. An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.(b)
Classification. Class II.