K Number
K222265

Validate with FDA (Live)

Device Name
Diosculpt
Date Cleared
2022-09-23

(57 days)

Product Code
Regulation Number
878.5400
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The DioSculpt laser system is intended for non-invasive lipolysis of the abdomen, flanks, back, and thighs in individuals with a Body Mass Index (BMI) of 30 or less.

Device Description

The DioSculpt laser lipolysis system is a diode laser system, Electrically efficient semiconductor generate optical radiation (1060nm) which is used to deliver laser energy to subcutaneous tissue layers. DioSculpt's cooling and electical systems assist in maintaining safe and comfortable skin surface temperatures.

AI/ML Overview

The provided text describes the 510(k) premarket notification for the DioSculpt laser system, which is intended for non-invasive lipolysis. However, it does not contain information about acceptance criteria or a study proving the device meets those criteria in the context of diagnostic performance or clinical outcomes.

The document focuses on demonstrating substantial equivalence to predicate devices through:

  • Technological characteristics comparison: Showing similar technology, wavelength, pulse width, cooling, and patient attachment methods.
  • Non-clinical testing: Biocompatibility, electrical safety, EMC, and performance testing (specifically temperature testing to ensure safety despite slightly higher energy density/peak power).
  • Software verification and validation.

Therefore, I cannot provide the requested information about acceptance criteria, reported device performance, sample sizes for test/training sets, data provenance, number or qualifications of experts, adjudication methods, MRMC studies, standalone performance, or types of ground truth used for clinical efficacy or diagnostic accuracy. These types of studies are not typically part of a 510(k) submission that relies on demonstrating substantial equivalence to a legally marketed predicate device through technological similarity and non-clinical safety/performance.

The only "performance" discussed is related to safety (temperature testing) and compliance with electrical and laser safety standards.

What can be extracted from the text related to performance/safety criteria are:

1. A table of acceptance criteria and the reported device performance (focused on safety performance):

Acceptance Criterion (Safety)Reported Device Performance (DioSculpt)
Maximum allowable temperature of applicatorsAll testing results do not exceed maximum allowable temperature of 41°C (Worst-case and real operation scenarios)
BiocompatibilityComplies with ISO 10993-1 for skin sensitization, in vitro cytotoxicity, and intracutaneous reactivity.
Electrical SafetyComplies with AAMI/ANSI ES60601-1:2005(R) 2012 and A1:2012 standards.
Electromagnetic Compatibility (EMC)Complies with IEC 60601-1-2:2014 standard.
Laser PerformanceComplies with IEC60601-2-22:2019 and IEC 60825-1:2014.
Software Safety Level"Moderate" level of concern (failure could directly result in minor injury). Software verified and validated.

The remaining requested information is not available in the provided document:

  1. Sample sizes used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective): Not applicable for this type of submission which focuses on non-clinical testing and substantial equivalence.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable.
  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable, as this is not an AI/diagnostic device.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc): For the safety tests, the "ground truth" is compliance with established international standards and the measured physical properties (e.g., temperature).
  7. The sample size for the training set: Not applicable (no machine learning model for training).
  8. How the ground truth for the training set was established: Not applicable.

In summary, the provided document details the non-clinical testing performed to demonstrate the safety and performance (in a technical compliance sense, not clinical efficacy) of the DioSculpt device for its 510(k) clearance, asserting substantial equivalence to predicate devices. It does not contain information on clinical studies that establish efficacy acceptance criteria or clinical performance outcomes.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.

September 23, 2022

Shanghai Wonderful Opto-Electrics Tech. Co., Ltd. Lily Zhou Management Representative 2F, Building 11, Lane 1175, Tongpu Rd. Shanghai, Shanghai 200333 China

Re: K222265

Trade/Device Name: Diosculpt Regulation Number: 21 CFR 878.5400 Regulation Name: Low level laser system for aesthetic use Regulatory Class: Class II Product Code: PKT Dated: November 14, 2021 Received: July 28, 2022

Dear Lily Zhou:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jianting Wang Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K22265

Device Name DioSculpt

Indications for Use (Describe)

The DioSculpt laser system is intended for non-invasive lipolysis of the abdomen, flanks, back, and thighs in individuals with a Body Mass Index (BMI) of 30 or less.

Type of Use (Select one or both, as applicable)

☑ Prescriptive Use (Part 21 CFR 201 Subpart D)☐ Over-The-Counter Use (21 CFR 201 Subpart C)
-----------------------------------------------------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

__ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for SWOT. The logo is blue and features the word "SWOT" in a stylized font. There are two lines going through the letters "S" and "W". There is a registered trademark symbol in the upper right corner of the logo.

Date Prepared: Sep. 6th, 2022

510(k) Summary

(As required by 21 CFR 807.92)

l. SUBMITTER

Name of Sponsor:Shanghai Wonderful Opto-Electrics Tech. Co., Ltd.
Address:2F, Building 11, Lane 1175, TongPu Rd.,Shanghai 200333, China
Contact Name:Lily Zhou
Telephone No.:0086-021-62642623
Fax No.:0086-021-52827988
Email Address:laser@wonderful-sh.com

II. DEVICE

Trade Name:Diosculpt
Common Name:Low Level Laser for Lipolysis
Model Name:Diosculpt
Regulation ClassificationLow Level Laser System for Aesthetic Use
Product Code:PKT
Classification Name:Laser for disruption of adipocyte cells for aesthetic use
Classification Panel:General & Plastic Surgery
Device Class:II
Regulation Number:21 CFR 878.5400

III. PREDICATE DEVICE

The identified predicates within this submission are as follows:

Primary Predicate Device: The Cynosure Sculpsure Low Level Laser System for Aesthetic Use has been cleared by FDA through 510(k) No.K171992 (Decision Date-September 26, 2017).

Secondary Predicate Device: K191068, Powersculp laser lipolysis system, Wuhan Lotuxs Technology Co.,Ltd.(Decision Date-July 17, 2019)

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for SWOT. The logo is blue and features the word "SWOT" in a stylized font. There are two horizontal lines that run through the word "SWOT", and there is a circle with an R in it in the upper right corner of the logo. The logo is simple and modern, and it is likely used to represent a company or organization.

DEVICE DESCRIPTION IV.

The DioSculpt laser lipolysis system is a diode laser system, Electrically efficient semiconductor generate optical radiation (1060nm) which is used to deliver laser energy to subcutaneous tissue layers. DioSculpt's cooling and electical systems assist in maintaining safe and comfortable skin surface temperatures.

V. INDICATIONS FOR USE

The DioSculpt laser system is intended for non-invasive lipolysis of the abdomen, flanks, back, and thighs in individuals with a Body Mass Index (BMI) of 30 or less.

VI. COMPARISION OF TECHNOLOGICAL CHARACTERISTICS WITH PREDICATE DEVICE

The subject device in this 510(k) uses the same technology that is utilized in the predicate device. A comparison of technological characteristics is provided in the following table:

510(k) NumberK 222265K171992K191068
Lipolysis MethodHeat-assistedHeat-assistedHeat-assisted
Device TypeDiode LaserDiode LaserDiode Laser
Wavelength1060±20nm(infrared)1060±20nm(infrared)1060±20nm(infrared)
Pulse WidthCWCWCW
Energy DensityUp to 1.8W/cm²Up to 1.4W/cm²Up to 1.7W/cm²
Spot Size4*8 cm² on eachof Applicators4*6 cm² on eachof Applicators4*8 cm² on eachof Applicators
Peak power perapplicator57.6W33.6W50W
CoolingSelf-containedclosed loopSelf-containedclosed loopSelf-containedclosed loop
Attachment topatientBeltBeltBelt

Discussion

The energy density of DioSculpt is 0.1W/cm² higher than the predicate device K191068, and the peak power energy is accordingly slightly higher. The higher energy may generate higher temperature to skin. To quarantee the safety, two temperature testing were conducted to measure the four applicators temperature in worst-case scenario and real operation

scenario. All testing results do not exceed maximum allowable temperature of 41°C .

The temperature testing result demonstrates the minor difference in energy and peak power per applicator does not raise safety concerns.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image contains a logo with the text "SWOT" in blue. The letters are stylized and connected by a blue line. There is a registered trademark symbol in the upper right corner of the logo. The logo appears to be for a company or product named SWOT.

VII. NON-CLINICAL TESTING

Biocompatibility testing

The biocompatibility evaluation for DioSculpt was conducted in accordance with the FDA Blue Book Memorandum #G95-1 "Use of International Standard ISO-10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing,"" May 1, 1995, and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA.

The applicators are considered tissue contacting, which is categorized as "Surface device", " Intact skin contact", and "with a limited contact duration of less than 24 hours", the following tests were evaluated:

  • O Skin Sensitization
  • O In Vitro Cytotoxicity
  • O Intracutaneous Reactivity

Electrical Safety and electromagnetic compatibility (EMC)

Electrical safety and EMC testing were conducted on DioSculpt. The system complies with the AAMI/ANSI ES60601-1:2005(R) 2012 and A1:2012 standards for safety and IEC 60601-1-2:2014 standard for EMC.

Performance Testing

The performance testing were conducted on DioSculpt. The system complied with IEC60601-2-22:2019 and IEC 60825-1:2014.

To guarantee the safety, two temperature testing were conducted to measure the four applicators temperature in worst-case scenario and real operation scenario. All testing results do not exceed maximum allowable temperature of 41℃ .

Software Verification and Validation Testing

Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "moderate" level of concern, since a failure or latent flaw in the software could directly result in minor injury, either to patient or to a user of device.

VIII. Conclusions

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807, it is concluded, basing on performance testing, indication for use, and technology, the subject device is the same or similar to the predicate device K171992 and K191068.

§ 878.5400 Low level laser system for aesthetic use

(a)
Identification. A Low Level Laser System for Aesthetic Use is a device using low level laser energy for the disruption of adipocyte cells within the fat layer for the release of fat and lipids from these cells for noninvasive aesthetic use.(b)
Classification. Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Industry and Food and Drug Administration Staff; Class II Special Controls Guidance Document: Low Level Laser System for Aesthetic Use.” See § 878.1(e) for the availability of this guidance document.