K Number
K222234

Validate with FDA (Live)

Date Cleared
2022-12-21

(148 days)

Product Code
Regulation Number
862.1345
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The GlucoSure ADVANCE Link Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertips, forearm, or palm. Alternative site testing should be performed only during steady-state (when glucose is not changing rapidly). Testing is done outside the body (In Vitro diagnostic use). It is intended for self-testing by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. It should only be used by a single patient and should not be shared. It is not indicated for the diagnosis of or screening for diabetes or for neonatal use.

Device Description

The GlucoSure ADVANCE Link Blood Glucose Monitoring System consists of the GlucoSure ADVANCE Link Blood Glucose Meter, GlucoSure ADVANCE Link Blood Glucose Test Strips and Contrex Plus 5 glucose control solution. It is used for testing of blood glucose by self-testers at home. The GlucoSure ADVANCE Link Blood Glucose Test Strips and Contrex Plus 5 glucose control solution are purchased separately. The modified device of GlucoSure ADVANCE Link Blood Glucose Meter is derived from the existing device of BGM014 Blood Glucose Meter and the modified device contain the Bluetooth function to transfer glucose results to the mobile device. The blood glucose test strips and glucose control solution utilized in the GlucoSure ADVANCE Link Blood Glucose Monitoring System are the same as the BGM014 Blood Glucose Test Strips and Contrex Plus 5 Glucose Control Solution, previously cleared in K161299. The meter materials of GlucoSure ADVANCE Link Blood Glucose Meter are the same as the BGM014 Blood Glucose Meter. Therefore, the disinfection performance (robustness of meter to multiple cleanings and disinfections) was previously cleared in K161299.

AI/ML Overview

Here's an analysis of the provided text, focusing on the acceptance criteria and study information:

Acceptance Criteria and Device Performance

The provided 510(k) summary does not explicitly list a table of quantifiable acceptance criteria with corresponding reported device performance, which is typical for performance claims related to new algorithms or AI-driven devices. Instead, it describes various tests performed to demonstrate substantial equivalence to a predicate device.

However, based on the nature of a Blood Glucose Monitoring System, the primary performance characteristic is system accuracy. While not explicitly a table, the document mentions that a "Usability study confirmed the system accuracy." The common standards for blood glucose meter accuracy are from ISO 15197. Without the full test reports, the specific numerical acceptance criteria and reported performance values related to accuracy are not present in this document.

The document also implies acceptance criteria for:

  • Battery Life: The new device must meet an acceptable battery life, which is noted to be decreased from 1000 tests to 750 tests compared to the predicate. The implicit acceptance criterion here would be that 750 tests is still considered acceptable for the intended use.
  • Memory Capacity: The new device's memory capacity of 700 test results is a decrease from 1000. Similar to battery life, the implicit acceptance criterion is that 700 results are sufficient.
  • EMC and Electrical Safety: The device must comply with relevant Electromagnetic Compatibility and Electrical Safety standards.
  • Software Verification and Validation (including cybersecurity): The software must be verified and validated to ensure proper function and data security, especially with the added Bluetooth functionality.
  • Disinfection Performance: The meter materials are the same as the predicate, so the disinfection performance must remain acceptable, as it was previously cleared.

Table of Implicit Acceptance Criteria and Reported Performance (based on available information):

Acceptance Criteria CategoryImplicit Acceptance Criteria / StandardReported Device Performance / Outcome
System AccuracyMeets accepted standards for blood glucose monitoring systems (e.g., ISO 15197, though not explicitly stated, clinical evidence is required for such devices). The system should show "system accuracy.""A Usability study confirmed the system accuracy" and "Testing showed that the GlucoSure ADVANCE Link Blood Glucose Monitoring System perform in a substantially equivalent manner to that of the predicate." (Specific numerical accuracy values are not provided in this summary.)
Battery LifeAn acceptable number of tests per battery charge for a blood glucose meter (the predicate was 1000 tests).Decreased to 750 tests. (Implied that 750 tests is acceptable for the self-testing patient population).
Memory CapacityAn acceptable number of glucose test results that can be stored on the device (the predicate was 1000 results).Decreased to 700 test results. (Implied that 700 results is acceptable).
EMC and Electrical SafetyCompliance with relevant Electromagnetical Compatibility and Electrical Safety standards."EMC and Electrical Safety" testing was conducted. (Implied that the device passed these tests and meets the necessary standards).
Software Verification and ValidationSoftware functions correctly, securely transfers data via Bluetooth, and handles associated error messages (Er5, Er6, Er7, Er8). Includes cybersecurity management."Software verification and validation including cybersecurity management" was conducted. (Implied that the software performs as intended and securely).
Disinfection PerformanceMeter materials must withstand multiple cleanings and disinfections without degradation of performance or integrity, as previously cleared for the predicate."The meter materials of GlucoSure ADVANCE Link Blood Glucose Meter are the same as the BGM014 Blood Glucose Meter. Therefore, the disinfection performance (robustness of meter to multiple cleanings and disinfections) was previously cleared in K161299." (Implied that the disinfection performance remains acceptable due to identical materials and prior clearance).
Usability / Ease of UseSystem operation should be according to design and easy for self-testers to use."A Usability study confirmed the system accuracy, operation according to design, and ease of use to support the intended use as described in the proposed labeling."
Substantial Equivalence to Predicate DeviceThe overall performance and safety profile of the new device must be comparable to the predicate device, especially regarding the quantitative measurement of glucose in fresh capillary whole blood samples from specified sites for self-testing by people with diabetes."Testing showed that the GlucoSure ADVANCE Link Blood Glucose Monitoring System perform in a substantially equivalent manner to that of the predicate."

Study Details from the Provided Text:

  1. Sample size used for the test set and the data provenance:

    • The document mentions "A Usability study" and "Testing," but does not provide specific sample sizes (e.g., number of patients, number of blood samples) for any test set.
    • The provenance of the data (e.g., country of origin, retrospective or prospective) is not explicitly stated. It's likely prospective for a usability/clinical accuracy study, but this is an inference.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • The document states "A Usability study confirmed the system accuracy." For blood glucose monitoring systems, ground truth is typically established by comparing capillary whole blood glucose readings to laboratory reference methods (e.g., YSI blood glucose analyzer) using venous plasma or serum.
    • The number of experts and their qualifications used to establish ground truth are not specified in this summary.
  3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • No adjudication method is mentioned in the provided text. For blood glucose accuracy studies, adjudication methods (like expert consensus) are typically not directly applicable as the ground truth is established by a quantitative laboratory reference method.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No MRMC study was mentioned or performed. This device is a Blood Glucose Monitoring System, which does not involve "human readers" or "AI assistance" in the typical diagnostic imaging sense. It's a direct measurement device.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Yes, implicitly. The "system accuracy" would be a standalone performance measure of the device itself (meter + test strips + algorithm) without requiring a human interpretation component, beyond the user following instructions. The device provides a quantitative number.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The specific type of ground truth is not explicitly stated in this summary but, in the context of blood glucose monitoring systems, ground truth for accuracy studies is typically established using a laboratory reference method (e.g., YSI 2300 STAT Plus Glucose & Lactate Analyzer measuring plasma glucose).
  7. The sample size for the training set:

    • This device is a measurement system, not a machine learning/AI diagnostic system in the sense that would require a distinct "training set" for an algorithm to learn from data in a traditional ML paradigm. The "algorithm" here refers to the embedded logic for glucose measurement. Therefore, no training set in the typical AI sense is mentioned or implied.
  8. How the ground truth for the training set was established:

    • As there's no mention of a traditional "training set" for an AI algorithm, this point is not applicable based on the provided text. The device's measurement algorithm is likely developed through engineering principles and calibration rather than data-driven machine learning training.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 21, 2022

Apex BioTechnology Corp. Lisa Liu Manager of Quality Assurance Division No. 7, Li-Hsin Road V, Hsinchu Science Park Hsinchu. 30078 China

Re: K22234

Trade/Device Name: GlucoSure ADVANCE Link Blood Glucose Monitoring System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose Test System Regulatory Class: Class II Product Code: NBW Dated: July 25, 2022 Received: July 26, 2022

Dear Lisa Liu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Digitally signed by Paula____ Paula Caposino -S Caposino -S Date: 2022.12.21 15:50:17 -05'00'

Paula Caposino, Ph.D. Acting Deputy Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.

510(k) Number (if known) K22234

Device Name

GlucoSure ADVANCE Link Blood Glucose Monitoring System

Indications for Use (Describe)

The GlucoSure ADVANCE Link Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertips, forearm, or palm. Alternative site testing should be performed only during steady-state (when glucose is not changing rapidly). Testing is done outside the body (In Vitro diagnostic use). It is intended for self-testing by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. It should only be used by a single patient and should not be shared. It is not indicated for the diagnosis of or screening for diabetes or for neonatal use.

The GlucoSure ADVANCE Link Blood Glucose Monitoring System is comprised of the GlucoSure ADVANCE Link Blood Glucose Meter and GlucoSure ADVANCE Link Blood Glucose Test Strips.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (6/20)

{3}------------------------------------------------

510(k) Summary

510(k) numberK222234
SubmitterApex Biotechnology Corp.No. 7, Li-Hsin Road V, Hsinchu Science ParkHsinchu, 30078CHINA (TAIWAN)
Contact PersonLisa LiuManager of Quality Assurance DivisionApex Biotechnology Corp.No. 7, Li-Hsin Road V, Hsinchu Science ParkHsinchu, 30078CHINA (TAIWAN)email: lisaliu@apexbio.comPhone: 011-886-3-5641952FAX: 011-886-3-5678021
Date PreparedJuly 25, 2022
Trade NamesGlucoSure ADVANCE Link Blood Glucose Monitoring System
ClassificationGlucose test system, 21 CFR 862.1345, Class II
Product CodesNBW
PredicateDevicesBGM014 Blood Glucose Monitoring System (K161299)
DeviceDescriptionThe GlucoSure ADVANCE Link Blood Glucose Monitoring System consists ofthe GlucoSure ADVANCE Link Blood Glucose Meter, GlucoSure ADVANCELink Blood Glucose Test Strips and Contrex Plus 5 glucose control solution. Itis used for testing of blood glucose by self-testers at home. The GlucoSureADVANCE Link Blood Glucose Test Strips and Contrex Plus 5 glucose controlsolution are purchased separately.The modified device of GlucoSure ADVANCE Link Blood Glucose Meter isderived from the existing device of BGM014 Blood Glucose Meter and themodified device contain the Bluetooth function to transfer glucose results to themobile device.The blood glucose test strips and glucose control solution utilized in theGlucoSure ADVANCE Link Blood Glucose Monitoring System are the same as
the BGM014 Blood Glucose Test Strips and Contrex Plus 5 Glucose ControlSolution, previously cleared in K161299.
The meter materials of GlucoSure ADVANCE Link Blood Glucose Meter arethe same as the BGM014 Blood Glucose Meter. Therefore, the disinfectionperformance (robustness of meter to multiple cleanings and disinfections) waspreviously cleared in K161299.
Intended UseThe GlucoSure ADVANCE Link Blood Glucose Monitoring System isintended for the quantitative measurement of glucose in fresh capillary wholeblood samples drawn from the fingertips, forearm, or palm. Alternative sitetesting should be performed only during steady-state (when glucose is notchanging rapidly). Testing is done outside the body (In Vitro diagnostic use).It is intended for self-testing by people with diabetes at home as an aid tomonitor the effectiveness of diabetes control. It should only be used by a singlepatient and should not be shared. It is not indicated for the diagnosis of orscreening for diabetes or for neonatal use.The GlucoSure ADVANCE Link Blood Glucose Monitoring System is
comprised of the GlucoSure ADVANCE Link Blood Glucose Meter andGlucoSure ADVANCE Link Blood Glucose Test Strips.
Comparison ofTechnologicalCharacteristicsThe GlucoSure ADVANCE Link Blood Glucose Meter uses the same test stripand test algorithm as the predicate. The changes was for The color of strip ejection button changed from gray to white. The printing color on the backside of the LCD cover changed from gray toblue. LCD pattern modification Bluetooth connectivity was added to the meter for the wireless transferof data to mobile devices. Added Bluetooth related error messageEr5: Bluetooth status failedEr6: The Bluetooth device pairing failedEr7: Failed data transmission via BluetoothEr8: Bluetooth disconnected during data upload Minor modification to control solution test process. Decrease memory capacity from 1000 test results to 700 test results. Battery life decreased from 1000 tests to 750 tests.
Change in name from BGM014 Blood Glucose Monitoring System toGlucoSure ADVANCE Link Blood Glucose Monitoring System. Change in name from BGM014 Blood Glucose Meter to GlucoSureADVANCE Link Blood Glucose Meter. The GlucoSure ADVANCE Link Blood Glucose Monitoring Systemshares identical blood glucose test strips with BGM014 Blood GlucoseMonitoring System, but only change in brand name from BGM014Blood Glucose Test Strips to GlucoSure ADVANCE Link BloodGlucose Test Strips.
Non-ClinicalTestingTesting was conducted as follows: Battery life test, EMC and Electrical Safety,Software verification and validation including cybersecurity managementdemonstrate substantial equivalence to the predicate system.
Clinical TestingA Usability study confirmed the system accuracy, operation according to design,and ease of use to support the intended use as described in the proposed labeling.
ConclusionTesting showed that the GlucoSure ADVANCE Link Blood Glucose MonitoringSystem perform in a substantially equivalent manner to that of the predicate. Weconclude that the candidate devices are substantially equivalent to the predicatedevice.

{4}------------------------------------------------

{5}------------------------------------------------

§ 862.1345 Glucose test system.

(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.