K Number
K222219

Validate with FDA (Live)

Date Cleared
2022-12-06

(134 days)

Product Code
Regulation Number
892.1750
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

R2 STUDIO Q/RCT820 is CBCT and panoramic x-ray imaging system with cephalometric. Which is intended to radiographic examination of the dento-maxillofacial, sinus, TMJ, Airway for diagnostic support for adult and pediatric patients. And a model scan is included as an option. Cephalometric image is also includes wrist to obtain carpus images for growth and maturity assessment for orthodontic treatment. The device is to be operated and used by dentists or other legally qualified health care professionals. This device is not intended for use on patients less than approximately 21 kg (46 lb) in weight and 113 cm (44.5 in) in height; these height and weight measurements approximately correspond to that of an average 5 year old.

Device Description

R2 STUDIO Q/RCT820 is 3D computed tomography for scanning hard tissues like bone and teeth. By rotating the c-arm which is embedded with high voltage generator all-in-one x-ray tube and a detector on each end, CBCT images of dental maxillofacial is attained by recombining data from the same level that are scanned from different angle. Panoramic image scanning function for attaining image of whole teeth, cephalometric scanning option for attaining cephalic image are included. and Model Scan option for attaining dental model CBCT image are included.

AI/ML Overview

The provided document is a 510(k) premarket notification for the R2 STUDIO Q/RCT820 dental X-ray system. Here's a breakdown of the requested information:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state acceptance criteria in a quantitative table format alongside reported performance for specific imaging metrics. Instead, it states that "All test results were satisfactory" following performance testing. The comparison table (page 5) lists various technical specifications for the proposed device and two predicate devices but does not provide specific performance values against pre-defined acceptance limits.

2. Sample size used for the test set and the data provenance

The document mentions "clinical evaluation report" but does not specify the sample size for any clinical test sets. The data provenance (country of origin, retrospective/prospective) is not explicitly stated.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This information is not provided in the document.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not provided in the document.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

The document does not mention any multi-reader multi-case (MRMC) comparative effectiveness study or any studies involving AI assistance. This device is an imaging system, not an AI-powered diagnostic tool.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

This information is not applicable as the device is an imaging system and not designed for standalone algorithmic performance without human-in-the-loop.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The document mentions "clinical evaluation report" and that "Sample images with a clinical evaluation report is provided to support the intended use of the device." However, it does not specify the type of ground truth used for this evaluation.

8. The sample size for the training set

The document does not mention a "training set" in the context of device performance, as it is describing an imaging system rather than an AI/ML model that requires training data.

9. How the ground truth for the training set was established

This information is not applicable as the document does not refer to a "training set" for an AI/ML model.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 6, 2022

MegaGen Implant CO., Ltd. % Yong Jae Kim Research Engineer 45, Secheon-ro, 7-gil Daegu, Dasa-eup, Dalesong-gun 42921 KOREA

Re: K222219

Trade/Device Name: R2 Studio Q/RCT820 Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: Class II Product Code: OAS Dated: July 18, 2022 Received: November 10, 2022

Dear Yong Jae Kim:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

2022.12.06
07:13:02
-05'00'
Lu Jiang

Lu Jiang, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K222219

Device Name R2 STUDIO Q/RCT820

Indications for Use (Describe)

R2 STUDIO Q/RCT820 is CBCT and panoramic x-ray imaging system with cephalometric.

Which is intended to radiographic examination of the dento-maxillofacial, sinus, TMJ, Airway for diagnostic support for adult and pediatric patients. And a model scan is included as an option.

Cephalometric image is also includes wrist to obtain carpus images for growth and maturity assessment for orthodontic treatment.

The device is to be operated and used by dentists or other legally qualified health care professionals.

This device is not intended for use on patients less than approximately 21 kg (46 lb) in weight and 113 cm (44.5 in) in height; these height and weight measurements approximately correspond to that of an average 5 year old.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

1. 510(k) Summary

The summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR Part 807.92.

  1. Date: July 18, 2022

3. Administrative Information

APPLICANTMegaGen Implant Co., Ltd.
ADDRESS45, Secheon-ro, 7gil, Dasa-eup, Dalseong-gun, Daegu, Republic of Korea
ManufacturerMegaGen Implant Co., Ltd.45, Secheon-ro, 7gil, Dasa-eup, Dalseong-gun, Daegu, Republic of Korea
TEL: +82-53-247-2261FAX: +82-53-247-2254

Contact Person e-mail: ra3@imegagen.com

4. Device Information

Device Name

Trade/Proprietary Name: R2 STUDIO Q/RCT820 Common Name: Dental panoramic/tomography and cephalometric x-ray system

Classification

Classification Name: Computed tomography x-ray system Regulation Number: 21 CFR 892.1750 Class: Il Product code: OAS Panel: Radiology

5. Predicate device & Reference Device

ParameterPredicate DeviceReference Device
Device NameRCT800RCT700
ManufacturerRAY Co., LtdRAY Co., Ltd
510(k) NumberK192737K213226
Classification nameComputed tomography x-ray systemComputed tomography x-ray system
Regulation number892.1750892.1750
Primary product codeOASOAS

{4}------------------------------------------------

7. Device Description

System purpose R2 STUDIO Q/RCT820 is 3D computed tomography for scanning hard tissues like bone and teeth. By rotating the c-arm which is embedded with high voltage generator all-in-one x-ray tube and a detector on each end, CBCT images of dental maxillofacial is attained by recombining data from the same level that are scanned from different angle.

Panoramic image scanning function for attaining image of whole teeth, cephalometric scanning option for attaining cephalic image are included. and Model Scan option for attaining dental model CBCT image are included.

8. Indication for use

R2 STUDIO Q/RCT820 is CBCT and panoramic x-ray imaging system with cephalometric.

Which is intended to radiographic examination of the dento-maxillofacial, sinus, TMJ, Airway for diagnostic support for adult and pediatric patients. And a model scan is included as an option.

Cephalometric image is also includes wrist to obtain carpus images for growth and maturity assessment for orthodontic treatment.

The device is to be operated and used by dentists or other legally qualified heath care professionals.

This device is not intended for use on patients less than approximately 21 kg (46 lb) in weight and 113 cm (44.5 in) in height; these height measurements approximately correspond to that of an average 5 vear old.

9. Patient population

The patient population can be the possible person who can be taken X-ray diagnostic radiation exposure. There is no restriction for ethnic group, Gender, weight, health, or condition. We recommend patients for x-ray diagnostic radiation exposure to be over 5 years old.

10. Operating principle

The purpose of this unit is diagnose clinical structures of a tooth and head areas mainly by using the characteristics of permeability from X-ray. The principle of functions are as following. The machine is made of X-ray generator and arms in which transfers X-ray signals to a sensor in 2D. Also, an object that has a magnification is required in a distance. Moreover, the unit has to be adjustable depending on height of a patient and PC system to reconstruct an image. The arm parts are controlled for rotating and linear moving to synchronize between the sensor and X-ray generator to get the image of interests. The purpose of this mechanism is to provide the images in 2D or/and 3D as preferred to diagnose in a monitor. CBCT provides in 3D images as reconstructed and Panorama is to diagnose the structures in a panoramic view. Cephalometric allows for orthodontic treatment. These 3 functions could be in 1 system, Panorama with Cephalometric, or Panorama only system depending on the needs. To provide the features as mentioned above, digital transferring from permeated X-ray to absorbing to the sensor is essential and all the process are proceed in Detector. Detector transfers X-ray to light depending on the structure materials. Detector is separated into indirect method that the light is changed to digital signals on photodiode and direct method in which the light is directly transferred to digital signal. This unit is using both direct method depending on the interior structure materials.

{5}------------------------------------------------

11. Comparison with predicate devices

The following table provides the summary of the technological characteristics of R2 STUDIO Q/RCT820 compared to the ence.

ParameterProposed DevicePredicate DeviceReference Device
ManufacturerMegaGen Implant Co., Ltd.RAY Co., Ltd.RAY Co., Ltd.
Device nameR2 STUDIO Q/RCT820RCT800RCT700
510(k) Number(Traditional 510(k))K192737K213226
(Traditional 510(k))(Traditional 510(k))
Common NameDental panoramic/tomography andcephalometric x-ray systemDental panoramic/tomography andcephalometric x-ray systemDental panoramic/tomography andcephalometric x-ray system
Indications for useR2 STUDIO Q/RCT820 is CBCT and panoramicx-ray imaging system with cephalometric.Which is intended to radiographicexamination of the dento-maxillofacial, sinus,TMJ, Airway for diagnostic support for adultand pediatric patients. And a model scan isincluded as an option.Cephalometric image is also includes wrist toobtain carpus images for growth and maturityassessment for orthodontic treatment.The device is to be operated and used bydentists or other legally qualified heath careprofessionals.This device is not intended for use on patientsless than approximately 21 kg (46 lb) in weightand 113 cm (44.5 in) in height; these heightand weight measurements approximatelycorrespond to that of an average 5 year old.RCT800 is CBCT and panoramic x-ray imagingsystem with cephalometric.Which is intended to radiographicexamination of the dento-maxillofacial, sinus,TMJ, Airway for diagnostic support for adultand pediatric patients. And a model scan isincluded as an option.Cephalometric image is also includes wrist toobtain carpus images for growth and maturityassessment for orthodontic treatment.The device is to be operated and used bydentists or other legally qualified heath careprofessionals.RCT700 is CBCT and panoramic x-ray imagingsystem with cephalometric. Which is intendedto radiographic examination of the dento-maxillofacial, sinus, TMJ, Airway and ENTstructure for diagnostic support for adult andpediatric patients. And a model scan isincluded as an option. Cephalometric image isalso includes wrist to obtain carpus images forgrowth and maturity assessment fororthodontic treatment. The device is to beoperated and used by dentists or other legallyqualified heath care professionals.
Mode of OperationContinuous operation with intermittent,stated permissible loadingContinuous operation with intermittent,stated permissible loadingContinuous operation with intermittent,stated permissible loading
3D technologyCBCT Cone beamCBCT Cone beamCBCT Cone beam
Computed TomographyComputed TomographyComputed Tomography
Performance Specification1) CBCT Computed tomography1) CBCT Computed tomography1) CBCT Computed tomography
- Patient- Patient- Patient
- Dental Model Scan(Optional)- Dental Model Scan(Optional)- Dental Model Scan(Optional)
2) Panoramic2) Panoramic2) Panoramic
3) Cephalometric(optional)3) Cephalometric(optional)3) Cephalometric(optional)
- Scan type- One shot type- One shot type
- Scan type
- Scan type
BaseBaseBase
CT + PANOCT + PANOCT + PANO
Option(CEPH)Option(CEPH)Option(CEPH)
CT + PANO + SCAN CEPHCT + PANO + SCAN CEPHCT + PANO + SCAN CEPH
Functional Option
CT + PANO + One shot(One shot, StandardCT + PANO + One shot(One shot, Standard
Type)Type)
CT + PANO + One shot(One shot, Large Type)CT + PANO + One shot(One shot, Large Type)
CTFXDD-1724RFXDD-0606CAFXDD-0606CA
FXDD-1012CHAJupi0606X
PANOFXDD-1724RFXDD-0606CAFXDD-0606CA
Detector TypeFXDD-1012CHAJupi0606X
Ceph (Scan)XID-C24DCXID-C24DCXID-C24DC
Ceph (One shot)N/A1717SCCFXRD-1717VA
N/APaxScan2530CFXDD1012CA
Exposure switch Type"Deadman" Button type"Deadman" Button type"Deadman" Button type
Ceph ApparatusCeph ApparatusCeph Apparatus
Vertical CarriageVertical CarriageVertical Carriage
RotatorRotatorRotator
X-RAY GeneratorX-RAY GeneratorX-RAY Generator
X-ray tubeX-ray tubeX-ray tube
High Frequency GeneratorHigh Frequency GeneratorHigh Frequency Generator
ColumnColumnColumn
Touch monitor (panel)Touch monitor (panel)Touch monitor (panel)
DetectorDetectorDetector
- CT- CT
Main ComponentsFXDD-1724RFXDD-0606CAFXDD-0606CA
FXDD-1012CHAJupi0606X
- PANO- PANO- PANO
FXDD-1724RFXDD-0606CAFXDD-0606CA
FXDD-1012CHAJupi0606X
- Ceph- Ceph-Ceph
XID-C24DC(Scan)XID-C24DC(Scan)XID-C24DC(Scan)
PaxScan 2530C(One shot, Standard Size)FXDD-1012CA(One shot, Standard Size)
1717SCC(One shot, Large Size)FXRD-1717VA(Oneshot, Large Size)
ChinrestChinrestChinrest
Head restHead restHead rest
Automatic CollimatorAutomatic CollimatorAutomatic Collimator
Exposure switchExposure switchExposure switch
Emergency stop switchEmergency stop switchEmergency stop switch
Console PC setConsole PC setConsole PC set
CT examsCT examsCT exams
Automatic CollimatorPanoramic examsPanoramic examsPanoramic exams
Cephalometric examsCephalometric examsCephalometric exams
Display TypeTFT LCD type(Normally black)*1280x800 pixelTFT LCD type(Normally black)*1280x800 pixelTFT LCD type(Normally black)*1280x800 pixel
ClassClass I with type B applied parts according toIEC 60601-1Class I with type B applied parts according toIEC 60601-1Class I with type B applied parts according toIEC 60601-1
Focal size0.5(Patient/Model scan)0.5(Patient)0.04 (Model scan)0.5
Field of View(CT)Max. 180x160 mmFXDD-0606CA : Max. 160x100 mmFXDD-1012CHA : Max. 200x200 mmMax. 160x100 mm
X-ray Voltage(Patient)60~100kVp60~100kVp60~100kVp
X-ray Current(Patient)1~17mA4~17mA1~17mA
X-ray Voltage(Model Scan, Optional)N/A50~80kVpN/A
X-ray Current(Model Scan, Optional)N/A0.4~0.7mAN/A
Total FiltrationMin. 2.8 mm Al equivalentMin. 2.8 mm Al equivalentMin. 2.8 mm Al equivalent
Detector PixelsizeCTFXDD-1724R : 95 μmFXDD-0606CA : 119 μmFXDD-1012CHA : 124 μmFXDD-0606CA : 119μmJupi0606X : 100 μm
PANON/AFXDD-1724R : 95 μmFXDD-0606CA : 119 μmFXDD-1012CHA : 124 μmFXDD-0606CA : 119 μmJupi0606X : 100 μm
Ceph (Scan)XID-C24DC : 100 μmXID-C24DC : 100 μmXID-C24DC : 100 μm
Ceph (One shot)N/AN/A1717SCC : 127 μmPaxScan 2530C : 139 μmFXRD-1717VA : 140 μmFXDD-1012CA : 124 μm
MagnificationCT1.44(Patient/Model scan)1.44(Patient)1.91(Model Scan)1.44
PANO1.351.31.31
Ceph (Scan)1.111.111.11
Ceph (One shot)N/AStandard Size : 1.12Large Size : 1.13Standard Size : 1.12Large Size : 1.13
N/A
Scan timeCT: below 20sec (Patient/Model scan)CT : below 20sec(Patient)CT : below 180sec(Model Scan)CT : below 14sec
Pano : below 14secPano : below 14secPano : below 14sec
Ceph(Scan) : below 20secCeph[Scan type] : below 20secCeph[Scan type] : below 20sec
N/ACeph[One shot type] : below 2secCeph[One shot type] : below 2sec
Format compatibleDICOM 3.0 Format compatibleDICOM 3.0 Format compatibleDICOM 3.0 Format compatible
Image Viewing SoftwareR2 STUDIO QRayScan(Cleared under K192737)RayScan(Cleared under K182614)
Image acquisitionGiga-Ethernet NetworkGiga-Ethernet NetworkGiga-Ethernet Network
Total HeightMax 2,296mmMax 2,296mmMax 2,296mm
Weight1) Computed Tomography(CT) +Panoramic(PANΟ) = 240kg(529.1lb) ± 10%1) Computed Tomography(CT) +Panoramic(PANΟ) = 189kg(416.6lb) ± 10%1) Computed Tomography(CT) +Panoramic(PANΟ) = 185kg(407.9lb) ± 10%
2) Computed Tomography(CT) +Panoramic(PANΟ) + Ceph(Scan type) =270kg(595.2lb) ± 10%2) Computed Tomography(CT) +Panoramic(PANO) + Ceph(Scan type) =219kg(482.8lb) ± 10%2) Computed Tomography(CT) +Panoramic(PANO) + Ceph(Scan type) =212.5kg(468.5lb) ± 10%
3) Computed Tomography(CT) +Panoramic(PANO) + Ceph(One shot type,installed in Standard size) = 217kg(478.4lb) ± 10%3) Computed Tomography(CT) +Panoramic(PANO) + Ceph(One shot type,installed in Standard size) = 211kg(465.2lb) ± 10%
4) Computed Tomography(CT) +Panoramic(PANO) + Ceph(One shot type,installed in Large size) = 212kg(467.3lb) ± 10%4) Computed Tomography(CT) +Panoramic(PANO) + Ceph(One shot type,installed in Large size) = 211kg(465.2lb) ± 10%
Type of installationWall or floor mountWall or floor mountWall or floor mount
Patient positionStanding / WheelchairStanding / WheelchairStanding / Wheelchair
Face scan modeOptionOptionNA
Applicable StandardsIEC 60601-1IEC 60601-1IEC 60601-1
IEC 60601-1-2IEC 60601-1-3IEC 60601-1-3
IEC 60601-1-3IEC 60601-2-63IEC 60601-2-63
IEC 60601-2-63IEC 60601-1-2IEC 60601-1-2

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

Similarities and differences

1. Similarities

The subject device has the same characteristic for the following compared to the predicate device

  • Indication for use, mode of operation, 3D technology, performal option, exposure switch type, main components(except for detector), autonativ collimator, display type, class, fora voltage(patient), }-ray current(patient), total filtration, magnification(CT, Ceph), scan time, format compatible, image aquisition, total height, y installation, patient position, applicable standards

2. Differences

The subject device has the different characteristic for the following compared to the predicate device and reference device - Detector (using for CT, PANO), Field of view, magnification(PANO),Weight, Image Viewing Software

{9}------------------------------------------------

12. Safety and Effectiveness Information

RCT820 system described in this 510(k) is similar to the predicate device in terms of indications for use, materials, safety characteristics, and X-ray source.

The fundamental technological characteristics of the subject and predicate device are similar. The imaging modes are similar; PANO, CEPH (Optional), CBCT, Model Scan.

The differences are as follows.

  • Detector (using for CT, PANO), Field of view, magnification (PANO)
  • Weight
  • Image Viewing Software

These differences are explained not affecting on the substantial equivalence, also the subject device was verified with non-clinical performance test and clinical test. The test result supports that the subject device is substantially equivalent to the predicate device and the diffecting the substantial equivalence.

Biocompatibility Testing

The biocompatibility evaluation for RCT820 components was conducted in accordance with the FDA Guidance Document and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices – Part1: Evaluation and Testing within a Risk Management Process ", as recognized by FDA.

The components of the RCT820 are considered surface medical device or external communicating medical device for a duration of less than 24 hours.

The biocompatibility testing was performed for Cytotoxicity, Skin sensitization and Irritation according to ISO 10993-5 and ISO 10993-10.

Electrical Safety and Electromagnetic compatibility (EMC)

The Electrical Safety and Electromagnetic compatibility were performed in accordance with the following standards.

  • · IEC 60601-1:2005/AMD1:2012
  • ・ IEC 60601-1-3:2008/AMD1:2013
  • · IEC 60601-1-6:2010/AMD1:2013
  • ・ IEC 60601-2-63:2017/AMD1:2017
  • · IEC 60601-1-2: 2014

Software Validation

The software(named R2 STUDIO Q) used in the subject device has an imaging viewing option similar to the predicate devices software. Likewise, the software of RCT820 saves the patient and image data and offers an inquiry function. In addition, supports the image generate function intended to obtain images using the RCT820 equipment and various sensors for diagnosis. And that has been validated according to FDA "Guidance for the Contents of Premarket Submissions for Software Contained in Medical Devices" and "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices" to assure substantial equivalence. The software for this device was considered as a "moderate" level of concern, since a failure or latent flaw in the

{10}------------------------------------------------

software would not directly result in serious injury or death to the patient or operator. As a result, we identified the level of concern associated with new device and provided documentation consistent with that level. Based on our risk analysis of software, the difference does not affect its safety and effectiveness.

Performance Test – Bench Test

Bench testing was conducted according to FDA Guidance "Format for Traditional and Abbreviated 510(k)s, section 18, Performance Testing - Bench testing is used to assess whether or not the parameter measured required for describing functionalities related to imaging properties of the dental X-ray device and patient dosage satisfies the designated tolerance. Performance (Imaging performance) testing was conducted according to standard of IEC 61223-3-4 and IEC 61223-3-7.

All test results were satisfactory.

Non-clinical considerations were conducted in accordance with FDA Guidance for the submissions of 510(k)'s for Solid State X-ray Imaging Devices".

RCT820 is equipped with FXDD-1724R. FXDD-1724R is a new SSXI detector, which is used to capture an image in panoramic, CBCT. Sample images with a clinical evaluation report is provided to support the intended use of the device.

Based on Non-Clinical and Clinical Test results of FXDD-1724R for the subject device, is similar to that of the Jupi0606X for the reference device.

Pediatric Information

Pediatric information related to the use of this device is provided to users in compliance with FDA guidance "Pediatric Information for X-ray Imaging Device Premarket Notifications".

13. Conclusions

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification. MegaGen Implant Co., Ltd. concludes that the newly RCT820 is safe and effective and substantially equivalent to the predicate devices as described herein.

§ 892.1750 Computed tomography x-ray system.

(a)
Identification. A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.