K Number
K222199

Validate with FDA (Live)

Date Cleared
2022-10-21

(91 days)

Product Code
Regulation Number
878.4430
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Collagen P.I.N. (Percutaneous Induction Needling system is indicated for use as a treatment to improve the appearance of facial acne scars in adults with Fitzpatrick Skin Types I - III, aged 22 years and older.

Device Description

The Collagen P.I.N. consists of a handheld device that creates microinjuries into the skin, by virtue of a 1 A DC motor that rapidly reciprocates (7000 - 9000 rpm) an array of 35 gauge (SWG) stainless steel microneedles. The device handpiece motor body is comprised of aluminum alloy, with a dial mechanism that controls the depth of penetration of the microneedle array from 0.0 mm to a maximum of 3.0 mm. The Collagen P.I.N. disposable needle cartridge is designed in a single configuration - a 36-pin array in a radial arrangement. The needles, which are 3.0 mm long, are composed of medical grade stainless steel, with a 1.0 mm conical taper to needle array is housed in a cartridge housing that prevents liquids from entering the handpiece motor body via the inside lumen of the cartridge. The needle cartridge is composed of a polycarbonate outer shell and piston, which houses the microneedle array. The microneedle array is attached to the polycarbonate shell using a silicone boot, which acts as spring mechanism which facilitates the microneedling mode of action. Each lot of disposable microneedle cartridges are individually packaged and sterilized by ethylene oxide gas ("EtO"). The device contains a rechargeable lithium-ion battery. The device is to be operated only cordlessly, and not used while the device is plugged into the wall charging station.

AI/ML Overview

This document describes a 510(k) premarket notification for the "Collagen P.I.N. (Percutaneous Induction Needling) System". This device is a microneedling system indicated for improving the appearance of facial acne scars. The provided text outlines the performance data, including non-clinical and clinical studies, to demonstrate substantial equivalence to a predicate device.

Based on the provided text, the device itself is a mechanical microneedling device, not an AI/algorithm-based device. Therefore, acceptance criteria, study design elements, and performance metrics typically associated with AI/ML-driven devices (like those in sections 1, 2, 3, 4, 5, 6, 7, 8, and 9 of your prompt, pertaining to AI performance, expert ground truth, multi-reader studies, training sets, etc.) are not applicable here.

The document primarily focuses on demonstrating the safety and efficacy of a medical device (microneedling system), not an AI algorithm. Its acceptance criteria and study designs are therefore geared towards verifying the mechanical function, sterility, biocompatibility, and clinical outcome for physical treatment, rather than the performance of an algorithm performing a diagnostic or predictive task.

However, I can extract the relevant performance data and acceptance criteria for this physical medical device as described in the document.

Here's a summary of the acceptance criteria and the study results for the Collagen P.I.N. device, focusing on what is available in the provided text:

Preamble:
The Collagen P.I.N. device is a microneedling system. The studies described are to demonstrate its safety and effectiveness for improving the appearance of facial acne scars, and to establish its substantial equivalence to a predicate device (Crown Aesthetics' SkinPen Precision System). There is no mention of AI or algorithmic components in this device or its testing. Therefore, the requested information pertaining to AI/ML device testing (e.g., sample size for test/training sets for AI, ground truth establishment for AI, MRMC studies, standalone AI performance) is not applicable here.

The evaluation focuses on the device's physical properties, biocompatibility, sterilization, and clinical outcomes based on human assessment.


1. Table of Acceptance Criteria and Reported Device Performance

The document does not present a single, consolidated table of "acceptance criteria" alongside "reported device performance" for each criterion in the way an AI/ML device submission might. Instead, it describes various tests and their successful outcomes. I will synthesize this information into a table format based on the detailed performance data provided.

Criterion Category / TestSpecific Acceptance CriteriaReported Device Performance / Test ResultConclusion
Biocompatibility - In Vitro Cytotoxicity (ISO 10993-5)Qualitative: Grade < 2 (non-cytotoxic) Quantitative: > 70% viability (non-cytotoxic)Qualitative: No change in morphology (Grade 0). Quantitative: Viability ranged from 91.30% - 97.95%.PASS
Biocompatibility - Intracutaneous Reactivity (ISO 10993-10)Avg Graded Score should be minimal/low (implied, as score of 0 was target). No mortality/morbidity.Avg Graded Score = 0. No mortality or morbidity. Increase in body weight observed. Difference of mean skin reaction scores for test item extracts and vehicle control was zero.PASS
Biocompatibility - Skin Sensitization (ISO 10993-10)Positive control gives clear positive results. No response in negative control. No evidence of sensitization by test item.No mortality or morbidity observed. All animals showed increase in body weight. No evidence of sensitization.PASS
Biocompatibility - Acute Systemic Toxicity (ISO 10993-11)No significantly greater biological reactivity than solvent control. No significant body weight loss in control. No mortality/abnormal behavior in control.Control animals showed no biological reactions. All animals showed increase in body weight. No signs of ill health or overt toxicity. No animals dead/moribund.PASS
Biocompatibility - Material Mediated Pyrogenicity (ISO 10993-11)Control temperature variation < 1°C. Test animal temperature < 39.8°C. No individual body temperature increase of 0.5°C or more compared to control.Control temperature variation < 1°C. Control temperatures < 39.8°C. No individual temperature increases of 0.5°C or more.PASS
Cleaning/Disinfection Method Validation3-log10 (99.9%) reduction of specified microorganisms: S. aureus, P. aeruginosa, E. coli, K. pneumoniae, M. terrae.Achieved microbial load reduction performance criteria for each microorganism (3-log10 reduction).PASS
ETO Sterilization Validation (ISO 11135:2014)Sterility Assurance Level (SAL) of 10^-6 (6-log10 reduction). EtO residuals below 0.5 µg/lens. ECH residuals below 2 µg/lens.Achieved 6-log10 reduction. EtO residuals not detectable (<0.5 µg/lens). ECH residuals not detectable (<2 µg/lens).PASS
Shelf Life Testing ValidationConformance to specified packaging integrity standards (visual inspection, internal pressure/bubble test, seal strength, dye penetration) after accelerated aging.All aged samples passed visual inspection, internal pressure, seal strength, and integrity-dye penetration tests at both 1-year and 2-year simulation time points.PASS (2-year shelf life validated for microneedle cartridges)
Transport Simulation Testing (ISTA 2A, ISO 11607-1)No adverse effects on device integrity/packaging after transport simulation (visual inspection, internal pressure/bubble test, seal strength, dye penetration).No visible changes for conditioning/vibration. No irreversible changes for compression. Needle cartridges passed visual inspection, internal pressure, seal strength, and integrity-dye penetration tests.PASS
In Vitro Study - Needle Length, Penetration, Cross-ContaminationVisible needle length within 3.0 mm specification. Max skin penetration depth within 2.5 mm specification. Max puncture duration within 0.180 seconds specification. No leakage or backflow.Visible length: 1.959 mm to 2.833 mm (within 3.0 mm spec). Max penetration depth: 2.866 mm (±0.047) (within 2.5 mm spec). Max puncture duration: 0.180 seconds (within 0.180 sec spec). No evidence of leakage or backflow.Meets specifications
Clinical Study - Efficacy (Clinician Global Aesthetic Improvement Assessment - CGAIS)Proportion of subjects with "CGAIS success" (grade 1-3) at Visit 5 (Day 150).Investigator (live assessment): 100% success for facial acne scars (n=23). Non-investigator panel (photographs): 47.8% (non-investigator 1) and 60.9% (non-investigator 2) success for facial acne scars (mITT population).Met (Investigator), Varies (Panel)
Clinical Study - Subject-Reported Efficacy (SSIS, SGAIS, Satisfaction)Improvement in appearance as reported by subjects.SSIS: 91.3% reported at least 25% improvement at Day 150. SGAIS: 82.6% reported at least improvement at Day 150. Subject Satisfaction: 91.3% satisfied or very satisfied at Day 150.Met
Clinical Study - Safety (Adverse Events)Acceptable incidence and severity of Adverse Events (AEs), Treatment-Related AEs (TRAEs), Serious AEs (SAEs), Unexpected Adverse Device Effects (UADEs).11 (22%) subjects experienced TEAEs. 6 (24%) subjects with facial scars experienced TRAEs (all post-inflammatory pigment change, mild/mod severity, in Fitzpatrick skin type IV-VI). No SAEs, no UADEs. No study discontinuation due to AEs.Acceptable safety profile
Clinical Study - Safety (Tolerability Assessments)Acceptable levels of dryness, roughness, tightness, erythema, edema, itching, burning.Most assessments (dryness, roughness, itching, burning) were none/mild for ≥ 94% post-treatment. Tightness: none/mild (60-84%), moderate (up to 38%), severe (2%). Erythema: none/mild (34-64%), moderate (up to 58%), severe (up to 11%). Edema: none/mild (62%), moderate (up to 38%), severe (up to 4%).Acceptable tolerability profile with expected, mostly mild, transient skin responses.

2. Sample size used for the test set and the data provenance

  • Clinical Study Test Set (Efficacy & Safety):

    • Sample Size: 50 subjects initially entered the study. 47 (94%) completed the study.
    • Primary Efficacy Population (mITT): 48 subjects (all who received at least one treatment and had at least one post-treatment evaluation).
    • Facial Acne Scars Efficacy Sub-group: 25 subjects (from the mITT group, 23 with complete efficacy data (CGAIS)).
    • Non-facial Scars Safety Sub-group: 25 subjects.
    • Data Provenance: The document explicitly states "Induction Therapies sponsored a clinical study... single-arm, two-site study". The locations of the sites are not specified by country, but the submission is to the U.S. FDA, implying a U.S. or internationally accepted clinical trial. The study was prospective.
  • Non-Clinical Test Sets: Sample sizes vary by test (e.g., 1000 aged samples for shelf-life, with 900 being the 36-needle cartridge; 2 pig specimens for animal study, 8 skin samples; specific numbers for biocompatibility tests are standard for those protocols, but not explicitly stated as "sample size" in this summary).

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Clinical Study (CGAIS - for efficacy ground truth):

    • Investigator Assessment (Live): Performed by the study Investigator. Their qualifications are not explicitly stated, but they are implied to be medical professionals conducting the study.
    • Non-Investigator Panel Assessment (Photographs): A panel of 2 board-certified dermatologists (non-investigators) graded the CGAIS post-final visit using photographs. Their years of experience are not specified.
  • Animal Study (Histopathology):

    • Morphological analysis performed by unstated "experts." No specific number or qualifications are given, but usually implies qualified pathologists.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Clinical Study - CGAIS by Non-Investigator Panel:
    • The document states "graded by a panel of 2 board-certified dermatologists". It then reports separate success rates for "non-investigators 1 and 2" (47.8% and 60.9% respectively). This indicates that their assessments were reported individually and potentially averaged or reported as a range, rather than strict adjudication like 2+1 or 3+1 consensus. There is no mention of an explicit adjudication method if their assessments differed.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not Applicable. This is a medical device (microneedling system), not an AI-assisted diagnostic or imaging device. The "readers" in this context (dermatologists) are providing a clinical assessment of a physical outcome, not interpreting images with or without AI assistance. Therefore, an MRMC comparative effectiveness study with AI assistance was not performed.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not Applicable. This device is a physical microneedling system, not an algorithm. There is no "standalone algorithm" performance to report.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • Clinical Study:
    • Efficacy: Primarily expert assessment (investigator and independent dermatologists) based on visual appearance and photographic evidence (CGAIS), and patient-reported outcomes (SSIS, SGAIS, Subject Satisfaction).
    • Safety: Clinician assessment of adverse events and tolerability (Investigator Tolerability Assessments), and patient-reported outcomes (Common Treatment Responses in Diaries).
  • Animal Study: Pathology (histopathology from pig specimens) for morphological analysis.
  • Non-Clinical Material Tests: Ground truth is established by universally accepted standardized test methods (ISO, ASTM standards), which define test conditions and expected results for material properties, sterility, and biocompatibility.

8. The sample size for the training set

  • Not Applicable. This is a physical medical device, not an AI/ML algorithm. There is no "training set" in the context of machine learning.

9. How the ground truth for the training set was established

  • Not Applicable. As there is no AI/ML training set, there is no ground truth establishment for it.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 21, 2022

% Laurie Clarke Executive Vice President & Principal Technology Sciences Group, Inc. 1150 18th Street NW, Suite 475 Washington, District of Columbia 20036

Re: K222199

Trade/Device Name: Collagen P.I.N. (Percutaneous Induction Needling) Regulation Number: 21 CFR 878.4430 Regulation Name: Microneedling Device For Aesthetic Use Regulatory Class: Class II Product Code: OAI Dated: July 22, 2022 Received: July 22, 2022

Dear Laurie Clarke:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may. therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm | identifies | combination | product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-howreport-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) CDRH Leam (https://www.fda.gov/training-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatorythe DICE website assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K222199

Device Name

Collagen P.I.N. (Percutaneous Induction Microneedling) System

Indications for Use (Describe)

The Collagen P.I.N. (Percutaneous Induction Needling system is indicated for use as a treatment to improve the appearance of facial acne scars in adults with Fitzpatrick Skin Types I - III, aged 22 years and older.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows a logo for a company called "Induction Therapies". The word "Induction" is stacked on top of the word "Therapies". The logo is in a light blue box with white text. The text is in a sans-serif font.

Induction Therapies, LLC 1920 Stanley Gault Parkway, Ste. 100 Louisville, Kentucky 40223 Phone: (502) 909-2837 Fax: www.inductiontherapies.com

510(k) SUMMARY

Induction Therapies LLC Collagen P.I.N. (Percutaneous Induction Needling) System (K222199)

Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared

Induction Therapies LLC

1920 Stanley Gault Parkway, Suite 100 Louisville, Kentucky 40223

Contact Tel:(502) 594-3116
Contact Person:Amelia Aslam, J.D.
Date Prepared:October 19, 2022

Submission Correspondent

Laurie A. Clarke, J.D., M.P.P. Vice President and Principal, Medical Device Regulatory TSG Consulting, Inc. 1150 18th Street, NW Suite 475 Laurie.Clarke@TSGConsulting.com (202) 828-8975

Name of Device and Name/Address of Sponsor

Collagen P.I.N. (Percutaneous Induction Needling) System

Induction Therapies LLC 1920 Stanley Gault Parkway, Suite 100 Louisville, Kentucky 40223

Proprietary Name - Collagen P.I.N. (Percutaneous Induction Needling) System

Common Name - Powered Microneedle Device

Classification Regulation Number - 21 C.F.R. 878.4430

Product Code - QAI

{4}------------------------------------------------

Predicate Device

Crown Aesthetics' SkinPen Precision System (K220506)

Collagen P.I.N's Intended Use

The Collagen P.I.N. (Percutaneous Induction Needling) microneedling system is intended to be used to create microinjuries in the skin for aesthetic purposes.

Comparison of Collagen P.I.N. and its Predicate Device's Intended Use

Collagen P.I.N and SkinPen are intended to be used to create microinjuries in the skin for aesthetic purposes. Thus, Collagen P.I.N. has the same intended use as its predicate device.

Collagen P.I.N.'s Indications for Use

The Collagen P.I.N. (Percutaneous Induction Needling) microneedling system is indicated for use as a treatment to improve the appearance of facial acne scars in adults with Fitzpatrick Skin Types I - III, aged 22 years and older.

Comparison of Collagen P.I.N. and its Predicate Device's Indications for Use

Collagen P.I.N.'s indication for use is identical to SkinPen's second indication for use as both of these devices are indicated as a treatment to improve the appearance of facial acne scars in adults with Fitzpatrick Skin Types I - III, aged 22 years or older.

Collagen P.I.N.'s Technological Characteristics

The Collagen P.I.N. consists of a handheld device that creates microinjuries into the skin, by virtue of a 1 A DC motor that rapidly reciprocates (7000 - 9000 rpm) an array of 35 gauge (SWG) stainless steel microneedles. The device handpiece motor body is comprised of aluminum alloy, with a dial mechanism that controls the depth of penetration of the microneedle array from 0.0 mm to a maximum of 3.0 mm.

The Collagen P.I.N. disposable needle cartridge is designed in a single configuration - a 36-pin array in a radial arrangement. The needles, which are 3.0 mm long, are composed of medical grade stainless steel, with a 1.0 mm conical taper to needle array is housed in a cartridge housing that prevents liquids from entering the handpiece motor body via the inside lumen of the cartridge. The needle cartridge is composed of a polycarbonate outer shell and piston, which houses the microneedle array. The microneedle array is attached to the polycarbonate shell using a silicone boot, which acts as spring mechanism which facilitates the microneedling mode of action. Each lot of disposable microneedle cartridges are individually packaged and sterilized by ethylene oxide gas ("EtO").

The device contains a rechargeable lithium-ion battery. The device is to be operated only cordlessly, and not used while the device is plugged into the wall charging station.

Comparison to the Predicate Device's Technological Characteristics

Collagen P.I.N. and SkinPen also have very similar technological characteristics in that both of these devices include a single-use cartridge containing needles that is inserted into the distal end of a reusable handpiece to which a battery pack containing a rechargeable lithium-ion battery is

{5}------------------------------------------------

attached at the proximal end of the handpiece. The primary technological differences between Collagen P.I.N. and its predicate are: (1) Collagen P.I.N.'s maximum treatment depth of 3.0 mm is 0.5 mm deeper than its predicate's maximum treatment depth of 2.5 mm; (2) its maximum puncture rate of 1915 punctures per second is 6.2% more than its predicate's maximum puncture rate of 1797 punctures per second : and (3) its maximum motor speed of 9,000 RPM is 14% faster than its predicate's maximum motor speed of 7,700 RPM. However, Collagen P.I.N. does not raise any different questions of safety and effectiveness compared to its predicate as these devices present the same risks, including whether the needles will injure blood vessels and nerves below the skin surface and whether the number and pattern of needle pricks it creates cause sufficient injury to the skin tissue to have the intended therapeutic effect of improving the appearance of facial acne scars in adults with Fitzpatrick Skin Types I - III, aged 22 years or older.

Performance Data - Non-Clinical

Biocompatibility

Collagen P.I.N. is an external communicating, indirect blood path device - with "limited" contact of < 24 hours. As such. Collagen P.I.N. requires Cytotoxicity. Irritation & Intracutaneous Reactivity, Skin Sensitization, Acute Systemic Toxicity, and Material-Mediated Pyrogencity testing per Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1". The following table summarizes biocompatibility testing conducted on Collagen P.I.N. Those tests demonstrate the biocompatibility of Collagen P.I.N for use as "a treatment to improve the appearance of facial acne scars in adults with Fitzpatrick Skin Type I - III, aged 22 years and older.

. •

{6}------------------------------------------------

Biocompatibility Test / StandardSuccess CriteriaTest ResultConclusion
In Vitro CytotoxicityISO 10993-5 Third Edition, 2009-06-01Qualitative EvaluationNumerical Grading Scale: 0 to 2Grade < 2, non-cytotoxicQuantitative EvaluationExtract viability validation>70% viability, non-cytotoxicQualitative EvaluationNo change in the morphology (Grade 0) was observed.Quantitative EvaluationViability ranged from 91.30% - 97.95%.PASS
Intracutaneous ReactivityISO 10993-10 Third Edition, 2010-08-01Qualitative EvaluationNumerical Grading Scale: 0 to 4, - Both Erythema& Oedema FormationMaximum Possible Score = 8Avg Graded Score = 0No mortality or morbidity were observed.All animals showed an increase in body weightThe difference of the mean skin reaction scores for thetest item extracts and the vehicle control was zero.PASS
Skin SensitizationISO 10993-10 Third Edition, 2010-08-01Acceptance Criteria1. Positive control trial conducted within the testfacility should give clear positive results.2. No response should be observed in negativecontrol treated animalsNo mortality or morbidity were observed.All animals showed an increase in body weight.No gross and histopathological examination were foundnecessary.No evidence of sensitization was seen in any of the testitem extracts treated animalsPASS
Acute System ToxicityISO 10993-11 Third Edition, 2017-09Acceptance Criteria1. None of the animals treated with test item shouldshow a significantly greater biological reactivitythan animals treated with solvent control.2. None of the animals in the control group shouldshow significant loss of body weight.3. No mortality or abnormal behavior such asconvulsions or prostration should occur incontrol group animals.Control animals showed no biological reactionsAll animals showed an increase in body weightNo signs of ill health or overt toxicityNo animals were found dead or in moribund conditionPASS
Material Mediated PyrogenicityISO 10993-11 Third Edition, 2017-09Acceptance Criteria1. The control temperature should not vary by morethan 1°C.2. The test animal's temperature should be less than39.8 °C.3. None of the treated animals should show anindividual body temperature increase of 0.5℃ ormore compared to its control temperature.The variation in control temperatures between the 3animals was less than 1 °C.The control temperatures in 3 animals did not exceed39.8 ℃ (38.4, 38.7, and 38.2 ℃)The animals didn't show any individual temperatureincreases of 0.5 ℃ or more when compared to itscontrol temperature.PASS

{7}------------------------------------------------

Cleaning/Disinfection Method Validation

The only component of Collagen P.I.N. that is reusable is the device's handpiece. The Collagen P.I.N's User Guide contains the following instructions for cleaning/disinfecting its handpieces after each use: remove all visible debris from the surface of the handpiece prior to intermediate level disinfection. Validation testing was conducted on handpieces cleaned and disinfected in accordance with those instructions the handpiece met the sponsor's performance criteria for a 3logio (99.9%) reduction of each microorganism bioload reduction for the following microorganisms; Staphylococcus aureus (S. aureus), Pseudomonas aeruginosa (P. Aeruginosa), Escherichia coli (E. coli), Klebsiella pneumoniae (K. pneumonia), and Mycolicibacter terrae (M. terrae) - in a nine (9) minute contact time. Disinfection Method Validation of the Collagen P.I.N. handpiece achieved the microbial load reduction performance criteria for each microorganism.

Validation of ETO Sterilization

Collagen P.I.N.'s 36-pin microneedle cartridge is supplied sterile. They are sterilized using an ethylene oxide sterilization cycle of 90% EtO and 10% CO2 to achieve a sterility assurance level of 10°, which equates to a 99.9999% reduction of targeted microorganism populations. This EtO sterilization was designed and validated per ISO 11135:2014 "Sterilization of health care products – Ethylene oxide – Requirements for development, validation and routine control of a sterilization process for medical devices", which is an FDA-recognized consensus standard.

The validation testing showed that this method of sterilizing Collagen P.I.N.'s 36-needle cartridge achieved a 6-log10 reduction. In addition, ethylene oxide ("EtO") residuals were not detectable on the needle cartridges and thus, they were below the specified limit of 0.5u. per lens. Their ethylene chlorohydrin ("ECH") residuals were also not detectable and thus below the specified limit of 2μ per lens. Accordingly, the results of this test validated this method of sterilizing the 36-pin microneedle cartridges.

Shelf Life Testing Validation

Induction Therapies tested the sterile microneedle cartridges for conformance to the following standards to evaluate the impact of storage on their sterility and performance.

  • Accelerated Aging per ISO 11607-1 + AMD 1:2014 "Packaging for terminally sterilized . medical devices - Part 1: Requirements for materials, sterile barrier and packaging systems" ASTM F 1980:2016. which equated to 2 years, real-time aging;
  • . Visual Inspection per ASTM 1886:2016 "Standard Test Method for Determining Integrity of Seal for Flexible Packaging by Visual Inspection",
  • Internal Pressure (Bubble test) per ASTM F2096:2011 "Standard Test Method for Detecting . Gross Leaks in Packaging Pressurization (Bubble Test)";
  • Seal Strength per ASTM F 88A:2015 Technique A and EN 868-5:2009 Annex D: and ●
  • Integrity Dye Penetration ASTM F 1929:2025 Technique A "Standard Test Method for ● Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration.

{8}------------------------------------------------

Testing was conducted on 1,000 aged samples, 900 of which are the 36-needle cartridge subject to this clearance. Accelerated aging for 46 days was used to simulate storage for one year and for 91 days to simulate storage for two years. The aged samples passed the visual inspection internal pressure (bubble test), the seal strength test, and the integrity - dye penetration test at both time points. Thus, the testing validates the microneedle cartridges labeled two-year shelf life.

Transport Simulation Testing

Validation testing on 900 units of the 36-needle cartridge for which Induction Therapies is seeking clearance in this submission, as well as 100 units of the 12-needle version of that cartridge, which the Company no longer intends to market in the U.S. The accelerated aging was calculated for 91 days for the 2-year shelf-life validation test and 46 days for the one-year shelf-life validation test. The needle cartridge passed the visual inspection internal pressure (bubble test), the seal strength test, and the integrity - dye penetration test for both the 2-year validation and the 1-year validation testing. Thus, the testing demonstrated the needle cartridge's 2-year shelf life. The 2-year shelflife validation test report is provided in Attachment 14B-1. The one-year shelf-life validation test report, which constitutes interim results, is provided in Attachment 14B-2.

Transport Simulation testing per ISTA 2A on Collagen P.I.N.'s needle cartridges was also conducted. The testing was conducted on 900 units of the 36-needle cartridge for which Induction Therapies is seeking clearance in this submission, as well as 100 units of the 12-needle version of that cartridge, which the Company no longer intends to market in the U.S. More specifically, the following testing was conducted:

  • Transport simulation ISTA 2A;
  • Visual Inspection per ASTM 1886:2016 Standard Test Method for Determining Integrity ● of Seal for Flexible Packaging by Visual Inspection;
  • Internal Pressure (Bubble test) per ASTM F2096:2011 Standard Test Method for Detecting ● Gross Leaks in Packaging Pressurization (Bubble Test);
  • Seal Strength per ASTM F 88A:2015 Technique A and EN 868-5:2009 Annex D; and ●
  • Integrity Dye Penetration ASTM F 1929:2025 Technique A Standard Test Method for . Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration.

No visible changes were observed for the conditioning and vibration test. In addition, no irreversible changes were observed for the compression test (using the machine apply and hold method) as deformations of approximately 127 mm were documented. The needle cartridges passed the visual inspection, internal pressure (bubble test), seal strength, and integrity - dye penetration tests. The validation test report is provided in Attachment 14C-1.

{9}------------------------------------------------

Induction Therapies' transport simulation of the Collagen P.I.N.'s needle cartridges in accordance with ISO 11607-1 "Packaging for terminally sterilized medical devices - Part 1 Requirements for materials, sterile barrier systems, and packaging systems." Samples were stressed using a transport simulation, according to ISTA 2A procedures. The tests were performed sequentially on the same shipping box. The shipping box remained unopened until the sequences of tests were completed. A visual inspection was performed after each sequence was finished. The following observations were noted for each test:

  • . Room Conditioning, including Extreme Cold Winter Conditioning and Elevated Temperature Summer Conditioning - no visible changes;
  • . Compression Test Machine Apply and Hold Method - Dents on carton box were observed. Large deformations of approximately 29.7, were documented;
  • Vibration Test 1 (Random) no visible changes were observed; ●
  • Drop-Test The box showed deformed drop corners and partially deformed drop edges after ● the test; and
  • . Vibration Test 2 (Random) - no visible changes were observed.

The ISO 11607 transport simulation test report is provided in Attachment 14C-2. The Collagen P.I.N.'s handpiece is labeled with a six-month shelf life.

In Vitro Study - Assessment of Needle Length, Penetration, and Cross-Contamination Risk

Induction Therapies conducted in vitro animal testing to verify the length of the Collagen P.I.N.'s needle, to assess their depth and rate penetration into skin, and to determine whether there is leakage and backflow through the needle cartridge silicone membrane into the device nosecone. This testing was conducting using the Collagen Pin's 36-needle cartridge on pig skin.

The testing showed the visible length for the 36-needle cartridge ranged from 1.959 mm to 2.833 which was within the manufacturer's specification of 3.0 mm. In addition, maximum skin penetration depth for the 36-needle cartridge was 2.866 mm (±0.047), which was within the manufacturer's specification of 2.5 mm. The test also showed that the maximum puncture duration (rate) of the needles was 0.180 seconds, which was within the manufacturer's specification of 0.180 seconds. Furthermore, there was no evidence of leakage or backflow. Thus, this study demonstrated that the Collagen P.I.N.'s needle cartridge met its specifications.

{10}------------------------------------------------

Animal Study - Assessment of Accuracy and Safety

Collagen P.I.N. conducted an In Vitro histopathology study, using two (2) pig specimens, to evaluate the accuracy and safety of the device's 36-pin needle cartridge. The test utilized eight (8) skin samples, one from the face, back, abdomen, and leg of each animal. The device was tested with and without ink. Ink was included as a mechanism to increase the visible assessment of microneedle accuracy.

The morphological analysis of the skin samples revealed small dermo-epidermal erosions secondary to punctures in all samples, and minor mononuclear inflammatory infiltrate of the papillary dermis in samples of abdomen and leg. The results and conclusions of this pig-skin histopathology report did not raise any different concerns or questions regarding safety and effectiveness - as related to the accuracy of the Collagen P.I.N. 36-pin microneedle cartridge.

Performance Data - Clinical

Induction Therapies sponsored a clinical study to evaluate the safety and effectiveness of Collagen P.I.N. as a treatment to improve the appearance of facial acne scars and nonfacial surgical or traumatic scars in adults with all Fitzpatrick Skin Types, aged 22 years and older. The clinical study was a nonsignificant risk, open-label, prospective, single-arm, two-site study in fifty subjects (n=50). The efficacy data from the subjects with facial acne scars (n = 25), as well the safety data from those subjects with nonfacial surgical or traumatic scars (n= 25), were used to demonstrate Collagen P.I.N.'s substantial equivalence for use in treating facial acne scars in the intended patient population.

Study Subjects

Eligible subjects were males or females at least 22 years of any race who had facial acne scars rated as mild to moderate (Grade 15 to 180) per the Echelle d'évaluation clinique des cicatrices d'acné (ECCA) acne scar scale or one non-facial trauma or surgical scar that was mild to moderate (score of 1 to 10) based on the Vancouver Scar Scale (VSS). The scars had to be amenable to microneedling and not have been treated with any alternative treatment modality within 6 months of the first study visit. Forty-seven of the subjects (94%) completed the study and 3 (6%) discontinued early.

Study Treatment

After washing the subject's skin in treatment area with Induction Therapies' Cleanse-It and applying DMC Medical Ltd.'s SupraGel to it, the investigators set the initial treatment depth by rotating the dial on the handpiece which has a scale from 0.0 mm to 3.0 mm increments. The investigator glided the tip perpendicular to the skin using a cross hatch motion (using horizontal and vertical movements) in the treatment area while applying gentle traction to the surrounding area. They gradually increased the needle depth until pinpoint bleeding was observed. Subjects were instructed to use Induction Therapies' Block-IT, which is a sunscreen, and Induction Therapies' Skin Renewal Hydrator after the treatment.

Subjects were treated with Collagen P.I.N. at Visit 1 (Day 1), Visit 2 (Day 30), Visit 3 (Day 60), and Visit 4 (Day 90). They were assessed for final efficacy and safety outcomes at Visit 5 (Day 150).

{11}------------------------------------------------

The mean number of treatments for subjects with facial acne scars was 3.7 of the planned 4 treatments. The maximum treatment depth ranged from 1.0 to 3.0 mm overall, with the mean ranging from 2.06 to 2.36 mm.

Study Endpoints and Statistical Analyses

Collagen P.I.N.'s efficacy was evaluated based on the following:

  • Clinician Global Aesthetic Improvement Assessment (CGAIS): a 5-point scale from 1 (Very Much Improved) to 5 (Worse). The CGAIS was assessed in two ways - i.e., live by the Investigator pre-treatment at Day 30, Day 90 (pre-treatment), and Day 150 using pretreatment photographs as the baseline. In addition, CGAIS was graded by a panel of 2 board-certified dermatologists (noninvestigators) after the final visit using photographs taken pre-treatment at each visit. CGAIS success was the proportion of subjects for whom the appearance of their scars improved, i.e., grade 1-3.
  • Self-Assessed Scar Improvement Scale (SSIS): 6-point scale from -1 (Exacerbation of . Scars) to 4 (75% to 99% improvement in appearance of acne scars).
  • Subject Global Aesthetic Improvement Assessment (SGAIS): 5-point scale from 1 (Very ● Much Improved) to 5 (Worse).
  • Subject Satisfaction: 5-point scale from 1 (Very Satisfied) to 5 (Very dissatisfied). ●

The primary population for efficacy analyses was the modified intent-to-treat ("mITT") population, which consisted of all subjects who received at least one study treatment and provided at least one post-treatment evaluation (n=48). The primary efficacy endpoint was CGAIS success in the mITT population.

The safety of Collagen P.I.N. was evaluated based on:

  • Adverse Events (AEs), including treatment-emergent AEs ("TEAEs") and treatment-. related AEs (TRAEs), serious adverse events ("SAEs"), unexpected adverse device effects ("UADEs"), and for TRAE by severity and relationship to study treatment.
  • . Investigator Tolerability Assessments on a scale of zero (none) to three (severe) of dryness, roughness of skin, tightness, erythema, and edema prior to each study treatment and 15 minutes following each treatment. The Investigator also asked the subject about itching and burning prior to and 15 minutes after treatment.
  • . Common Treatment Responses (CTR Diary) - the subjects recorded the following responses that typically occur following microneedling treatment in their CTR Diaries: redness, tightness, rough skin, dryness, tenderness, itching, burning, discomfort, and peeling.

{12}------------------------------------------------

Study Results:

Collagen P.I.N. improved the appearance of mild to moderate facial acne scars as assessed by the CGAIS, SSIS, SGAIS, and subject satisfaction. The primary efficacy endpoint of the proportion of subjects with CGAIS success at Visit 5 (Day 150) in the mITT population based on live assessment by the Investigator was 100% for subjects with facial acne scars (n=23) with no subject experiencing a worsening of their scars. The proportion of subjects with CGAIS success at Visit 5 in the mITT population graded by non- investigators 1 and 2, which was the other primary efficacy endpoint, was 47.8% and 60.9% for subjects with facial acne scars.

The proportion of subjects in the mITT population who reported at least a 25% improvement in the appearance of treated scars on the SSIS was 91.3% at Visit 5 (Day 150) for subjects with facial acne scars. The proportion of subjects in the mITT population who reported at least improvement on the SGAIS increased was 82.6% at Visit 5 (Day 150) for subjects with facial acne scars. The proportion of subjects reporting that they were satisfied or very satisfied increased was 91.3% at Visit 5 for subjects with facial acne scars. The increase in the proportion of subjects reporting improvement on the SSIS and on SGAIS are consistent with the CGAIS success assessed by the Investigator at Visit 5. The increase in subject satisfaction over time supports the investigator's CGAIS results.

Of the 50 treated subjects, 11 (22%) experienced one or more TEAEs during the study. Eleven of the 25 subjects with facial acne scars (44%) had a total of 12 TEAEs and none of the 25 subjects with nonfacial scars had a TEAE. Six of the 25 subjects with facial scars (24%) experienced one or more TRAEs during the study. The only TRAE was post-inflammatory pigment change; all of those events were mild or moderate and all 6 subjects who experienced TRAEs had Fitzpatrick skin type IV to VI. No SAEs and no UADEs were reported during the study and no subjects had study treatment temporarily or permanently discontinued due to AEs.

All of the treatment tolerability assessments were reported as none or mild pre-treatment by at least 96% of subjects on Days 1, 30, 60, and 90 post-treatments of dryness, roughness of skin, itching, and burning were reported as none or mild for at least 94% of subjects. Tightness was assessed as none or mild post-treatment for 60% to 84% of subjects, moderate for up to 38% of subjects (Day 90), and severe for 2% on Day 90. Erythema was reported post-treatment as none or mild for 34% to 64% of subjects, moderate for up to 58% of subjects (Day 1), and severe for up to 11% (Day 30). Edema was reported post-treatment as none or mild for 62% of subjects, moderate for up to 38% of subjects (Day 60), and severe for up to 4% (Day 90).

A similar pattern was observed across most of the Common Treatment Responses ("CTRs"). From Day 1 to Day 7 after each visit, the proportion of subjects reporting none to mild CTRs generally increased, reaching 85% to 100% at Day 7. The severity of rough skin and peeling tended to increase for a few days before decreasing.

The results of this open-label study have shown that the Collagen P.I.N. improved the appearance of mild to moderate facial acne scars and was well-tolerated by all study subjects.

{13}------------------------------------------------

Conclusion

The Collagen P.I.N. has the same intended uses and similar indications and technological characteristics as SkinPen. Collagen P.I.N.'s slightly deeper maximum treatment depth, higher maximum puncture rate and slightly higher motor speed do not raise new questions of safety or effectiveness. Bench, animal, and clinical data demonstrate the safety and efficacy of Collagen P.I.N. for "use as a treatment to improve the appearance of facial acne scars in adults with Fitzpatrick Skin Types I - III, aged 22 years and older." Thus, Collagen P.I.N. is substantially equivalent to its predicate device for its proposed indication for use.

§ 878.4430 Microneedling device for aesthetic use.

(a)
Identification. A microneedling device for aesthetic use is a device using one or more needles to mechanically puncture and injure skin tissue for aesthetic use. This classification does not include devices intended for transdermal delivery of topical products such as cosmetics, drugs, or biologics.(b)
Classification. Class II (special controls). The special controls for this device are:(1) The technical specifications and needle characteristics must be identified, including needle length, geometry, maximum penetration depth, and puncture rate.
(2) Non-clinical performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(i) Accuracy of needle penetration depth and puncture rate;
(ii) Safety features built into the device to protect against cross-contamination, including fluid ingress protection; and
(iii) Identification of the maximum safe needle penetration depth for the device for the labeled indications for use.
(3) Performance data must demonstrate the sterility of the patient-contacting components of the device.
(4) Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the intended shelf life.
(5) Performance data must demonstrate the electrical safety and electromagnetic compatibility (EMC) of all electrical components of the device.
(6) Software verification, validation, and hazard analysis must be performed for all software components of the device.
(7) The patient-contacting components of the device must be demonstrated to be biocompatible.
(8) Performance data must validate the cleaning and disinfection instructions for reusable components of the device.
(9) Labeling must include the following:
(i) Information on how to operate the device and its components and the typical course of treatment;
(ii) A summary of the device technical parameters, including needle length, needle geometry, maximum penetration depth, and puncture rate;
(iii) Validated methods and instructions for reprocessing of any reusable components;
(iv) Disposal instructions; and
(v) A shelf life.
(10) Patient labeling must be provided and must include:
(i) Information on how the device operates and the typical course of treatment;
(ii) The probable risks and benefits associated with use of the device; and
(iii) Postoperative care instructions.