K Number
K222160

Validate with FDA (Live)

Date Cleared
2022-11-22

(125 days)

Product Code
Regulation Number
870.1130
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Globalcare GUS610 Automatic Blood Pressure Monitor is indicated for the non-invasive measurement of diastolic and systolic blood pressures and pulse rate of adults by means of an inflatable cuff which is wrapped around the upper arm. The cuff circumference is limited to 22 to 42 cm.

Device Description

The Globalcare GUS610 Blood Pressure Monitor is an automatic upper arm blood pressure monitor intended for non-invasive measurement or monitoring of adults' arterial blood pressure and pulse rate. The device is powered by batteries or by a mains-connected power supply. The Globalcare GUS610 Blood Pressure Monitor comprises the following parts: Blood pressure monitor main device x1, Upper arm cuff x1, 1.5 V LR03 AAA batteries x4, Storage bag x1, Instructions for use x1, 100-240Vac 50/60Hz power supply (optional) with USB cable x1. None of the parts included with the device are supplied in a sterile condition. The Globalcare GUS610 Blood Pressure Monit is an electronic unit incorporating a screen that displays results and other information relevant to device operation. The device is powered by batteries or by a mains-connected power adapter and is designed for home ('OTC') use. The clinical outputs from the GUS610 Blood Pressure Monitor are: Systolic pressure (mm Hq), Diastolic pressure (mm Hq), Pulse rate (1/min), Cardiac arrhythmia (Irregular heartbeat (IHB)) symbol, Risk indicator (WHO color scale), Hemodynamic Rest Condition (HSD) indication. The device operates on the oscillometric method: pressure sensors on the cuff are used to capture pulse waves during constricted blood flow and the device then computes the results for display by means of an algorithm which analyses the pressures transmitted during arterial oscillations that occur during cuff inflation. Irreqular heartbeat (IHB) is detected by measuring the interval time between each heartbeat and comparing these values with the average heartbeat interval (heartbeat per minutes shown on the display). If the difference between any heartbeat interval and the average is higher or lower than 25 %, the device shows the irregular heartbeat (IHB) symbol on the display. The pulse rate is determined by calculating the frequency of the cuff, which are cardiac synchronous. The GUS610 is controlled by software which calculates the blood pressure (diastolic and pulse rate by the oscillometric method. To do this, it also controls the pneumatic components to inflate the cuff (with a pump), holds the cuff pressure, and then deflates the cuff (by means of a solenoid valve). The software collects the relevant data from the pressure sensor during inflation. The software also drives the LCD display, from which the user can read the result of the blood pressure and pulse rate measurements and undertakes the Hemodynamic Rest Condition (HSD) detection calculations, and reports whether any irregular heartbeat is occurring.

AI/ML Overview

The provided text describes the regulatory clearance of a blood pressure monitor, the Globalcare GUS610, and its substantial equivalence to a predicate device. The primary study presented focuses on the clinical accuracy of the device rather than the performance of an AI algorithm or human reader improvement.

Here's an analysis of the acceptance criteria and the study, addressing your specific points:

Acceptance Criteria and Reported Device Performance

The device, Globalcare GUS610 Blood Pressure Monitor, is subject to the ISO 81060-2:2018/AMD1:2020 standard for clinical accuracy of non-invasive automated sphygmomanometers. The study aims to demonstrate compliance with this standard.

Acceptance Criteria (from ISO 81060-2:2018+A1:2020)Reported Device Performance (from Clinical Investigation)
Clinical Accuracy (subclause 5.2.4.1.2)Device capable of accurately measuring blood pressure values in all specified subgroups.
SafetyNo adverse events reported during the investigation.
Distribution of SubjectsMet or exceeded minimum proportions for gender, limb size, and blood pressure ranges (SBP, DBP).

Note: The document explicitly states: "The clinical results indicated that the subject device meets the requirements of ISO 81060-2:2018+A1: 2020." The specific numerical accuracy metrics (e.g., mean difference and standard deviation of differences between the device and reference method) which are central to ISO 81060-2 are not directly presented in this summary, but the conclusion states that the requirements were met.

Study Details

  1. Sample size used for the test set and the data provenance:

    • Sample Size: 255 blood pressure datasets from 85 adult subjects.
    • Data Provenance: The document states "A clinical investigation was performed to validate the clinical accuracy of the subject device... The investigation was performed on Globalcare model GCE610 blood pressure monitor, a device that is technologically identical to the subject device, but which is cleared for marketing in a non-US regulatory jurisdiction." This implies the data originates from a clinical study conducted for a similar device, likely outside the US (given "non-US regulatory jurisdiction"). The method described is a prospective clinical accuracy study.
    • The demographic breakdown of the 85 subjects is provided in Table 1 (e.g., gender, limb size, SBP/DBP ranges), demonstrating adherence to the ISO 81060-2 standard's requirements for subject distribution.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • The ground truth was established using an auscultatory reference sphygmomanometer. Per ISO 81060-2, this typically involves two trained observers simultaneously and independently taking readings. The document does not specify the number or qualifications of these observers, but good clinical practice for such studies requires highly trained individuals to minimize observer bias.
  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • The method used is the arm sequential method described in subclause 5.2.4.1 of ISO 81060-2:2018+A1:2020. This method involves sequential measurements by the test device and the reference standard(s) on the same arm, with a specific protocol for timing and number of repetitions. ISO 81060-2 generally requires two independent observers for the reference measurement, with a pre-defined threshold for disagreement (e.g., within 4 mmHg); if they disagree by more than this, a third observer might be involved or the measurement discarded. The document does not explicitly detail the observer adjudication process but implies adherence to the ISO standard.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, an MRMC comparative effectiveness study was not done. This study is for a standalone medical device (blood pressure monitor) that accurately measures blood pressure, not an AI-powered diagnostic tool that assists human readers.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Yes, a standalone performance evaluation was done. The GUS610 Blood Pressure Monitor itself, an automated device, operates without continuous human input during the measurement process. The "performance data" and "clinical performance" sections directly describe the accuracy of the device's algorithm in measuring blood pressure and pulse rate. The device's software is described as controlling the pneumatic components, collecting data from the pressure sensor, and calculating results (systolic, diastolic, pulse rate, IHB, HSD). This is a direct assessment of the device's algorithmic performance against a clinical reference.
  6. The type of ground truth used (expert concensus, pathology, outcomes data, etc):

    • The ground truth used was auscultatory reference measurements obtained using a legally available sphygmomanometer, aligning with the "gold standard" for blood pressure measurement in clinical validation studies as specified by ISO 81060-2.
  7. The sample size for the training set:

    • Not applicable / Not specified. This document describes the validation study for a medical device (blood pressure monitor) against an established international standard. It's not a submission for an AI/ML device where a separate "training set" would be typically documented as part of the regulatory submission process. The device's algorithm for calculating blood pressure is pre-programmed, refined during development processes, but not typically "trained" in the machine learning sense with a discrete training set that requires specific regulatory submission.
  8. How the ground truth for the training set was established:

    • Not applicable. As above, the concept of a "training set" for the type of device described (non-invasive blood pressure monitor) is not relevant in the context of this regulatory summary. The device's underlying algorithm is based on established oscillometric principles and calibration/design processes, not on a machine learning training dataset established with ground truth in the way AI/ML systems are developed.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. FOOD & DRUG ADMINISTRATION".

November 22, 2022

Globalcare Medical Technology Co., Ltd. % Roger Gray VP Quality and Regulatory Donawa Lifescience Consluting Srl Piazza Albania 10 Rome, 00153 Italy

Re: K222160

Trade/Device Name: GUS610 Blood pressure monitor Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: October 20, 2022 Received: October 27, 2022

Dear Roger Gray:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Stephen G. Browning -S

LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K222160

Device Name

GUS610 Automatic Blood Pressure Monitor

Indications for Use (Describe)

The Globalcare GUS610 Automatic Blood Pressure Monitor is indicated for the non-invasive measurement of diastolic and systolic blood pressures and pulse rate of adults by means of an inflatable cuff which is wrapped around the upper arm. The cuff circumference is limited to 22 to 42 cm.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.qov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

Device Name:GUS610 Blood Pressure Monitor
Type of 510(k) submission:Traditional
Date of submission:21 November 2022
Manufacturer:Globalcare Medical Technology Co., LtdNo.7 Factory, European Industrial ParkNo. 39 Mid Industrial Main RoadXiaolan Town, Zhongshan CityGuangdong, CHINA 528415
Phone:+86-760-22589901
FDA Establishment Reg. Number:3010880718
510(k) Owner and Submitter:Globalcare Medical Technology Co., LtdNo.7 Factory, European Industrial ParkNo. 39 Mid Industrial Main RoadXiaolan Town, Zhongshan CityGuangdong, CHINA 528415
Owner/Operator Reg. Number:10046812
510(k) Contact:Mr Roger GrayVP Quality and RegulatoryDonawa Lifescience ConsultingPiazza Albania 1000153 RomeItaly
Phone:Fax:Email:+39 06 578 2665+39 06 574 3786rgray@donawa.com
FDA Product Code:DXN
FDA Regulation Number:21 CFR 870.1130
FDA Classification Name:Noninvasive blood pressure measurement system
Classification Panel:Cardiovascular
Common Name:Upper arm blood pressure monitor
FDA Classification:Class II
Submission Type:510(k)
Indications for Use:The Globalcare GUS610 automatic Blood Pressure Monitor isindicated for home use for the non-invasive measurement ofdiastolic and systolic blood pressures and pulse rate of adults bymeans of an inflatable cuff which is wrapped around the upperarm.

{4}------------------------------------------------

Device Description:

The Globalcare GUS610 Blood Pressure Monitor is an automatic upper arm blood pressure monitor intended for non-invasive measurement or monitoring of adults' arterial blood pressure and pulse rate.

The device is powered by batteries or by a mains-connected power supply.

The Globalcare GUS610 Blood Pressure Monitor comprises the following parts:

• Blood pressure monitor main devicex1
• Upper arm cuffx1
• 1.5 V LR03 AAA batteriesx4
• Storage bagx1
• Instructions for usex1
• 100-240Vac 50/60Hz power supply (optional) with USB cablex1

None of the parts included with the device are supplied in a sterile condition.

The Globalcare GUS610 Blood Pressure Monit is an electronic unit incorporating a screen that displays results and other information relevant to device operation. The device is powered by batteries or by a mains-connected power adapter and is designed for home ('OTC') use.

The clinical outputs from the GUS610 Blood Pressure Monitor are:

  • . Systolic pressure (mm Hq)
  • Diastolic pressure (mm Hq) ●
  • . Pulse rate (1/min)
  • Cardiac arrhythmia (Irregular heartbeat (IHB)) symbol ●
  • Risk indicator (WHO color scale)
  • Hemodynamic Rest Condition (HSD) indication

The device operates on the oscillometric method: pressure sensors on the cuff are used to capture pulse waves during constricted blood flow and the device then computes the results for display by means of an algorithm which analyses the pressures transmitted during arterial oscillations that occur during cuff inflation.

Irreqular heartbeat (IHB) is detected by measuring the interval time between each heartbeat and comparing these values with the average heartbeat interval (heartbeat per minutes shown on the display). If the difference between any heartbeat interval and the average is higher or lower than 25 %, the device shows the irregular heartbeat (IHB) symbol on the display.

The pulse rate is determined by calculating the frequency of the cuff, which are cardiac synchronous.

The GUS610 is controlled by software which calculates the blood pressure (diastolic and pulse rate by the oscillometric method. To do this, it also controls the pneumatic components to inflate the cuff (with a pump), holds the cuff pressure, and then deflates the cuff (by means of a solenoid valve). The software collects the relevant data from the pressure sensor during inflation.

The software also drives the LCD display, from which the user can read the result of the blood pressure and pulse rate measurements and undertakes the Hemodynamic Rest Condition (HSD) detection calculations, and reports whether any irregular heartbeat is occurring.

Performance data:

The GUS610 has been tested and found to be in compliance with the following standards:

{5}------------------------------------------------

Safety and EMC:

  • IEC 60601-1:2005/AMD1:2012 + AMD2:2020 ●
  • IEC 60601-1-11:2015/AMD1:2020 .
  • IEC 60601-1-2:2014/A1:2020 ●
  • IEC 80601-2-30:2018 ●

Clinical performance:

  • ISO 81060-2:2018/AMD1:2020 ●
    Cuff biocompatibility:

  • ISO 10993-5:2009 ●

  • . ISO 10993-10:2010

Bench tests:

  • Internal tests to verify performance ●
    • Mains and Battery comparison O
    • After 500 hour Life Test has been completed O
    • After 12 months storage O
    • After drop test O

The results of the above testing assist in the demonstration of substantial equivalence of the subject device with the predicate device.

Clinical performance

A clinical investigation was performed to validate the clinical accuracy of the subject device, using the arm sequential method described in subclause 5.2.4.1 of ISO 81060-2:2018+A1:2020 to collect 255 blood pressure datasets from 85 adult subjects.

A clinical investigation was carried out to validate device clinical accuracy. The investigation was performed on Globalcare model GCE610 blood pressure monitor, a device that is technologically identical to the subject device, but which is cleared for marketing in a non-US regulatory jurisdiction. The clinical validation was performed using the same arm sequential method described in subclause 5.2.4.1 of ISO 81060-2:2018+A1:2020 to collect 255 blood pressure datasets from 85 adult subjects.

According to the requirements of ISO 81060-2:2018+A1:2020, a legally available in the US auscultatory reference sphygmomanometer was used for comparison purposes.

The clinical results indicated that the subject device meets the requirements of ISO 81060-2:2018+A1: 2020, and that there were no adverse events during the investigation. Baseline demographic information regarding participants in the study is provided in Table 1.

DistributionCategoryRequirementActual distributionComment
MinimumproportionMinimumnumberNumberProportion
GenderMale30%265665.9%P
Female30%202934.1%P
Limb sizerange(cm)22-2720%172630.6%P
27-3220%172023.5%P
32-3720%171821.2%P
37-4210%172124.7%P
22-24.510%91011.8%P
39.5-4210%91112.9%P
SBP ≤1005%41214.1%P
Bloodpressure(mmHg)SBP ≥14020%171922.4%P
SBP ≥1605%41517.6%P
DBP ≤605%41011.8%P
DBP ≥8520%172832.9%P
DBP >1005%41214.1%P

Table 1: Baseline demographic information

{6}------------------------------------------------

Analysis of data from the investigation is based on subclause 5.2.4.1.2 of ISO 81060-2:2018+A1:2020 with the following conclusions:

Effectiveness: The results of the investigation show that the subject device is capable of accurately measuring the blood pressure values in all the subgroups foreseen by standard ISO 81060-2:2018+A1: 2020.

Safety: The study reported no adverse events for the subject device.

Summary: Based on the clinical performance as documented in the clinical study report, the subject device was found to have a safety and effectiveness profile that is substantially equivalent to the predicate device.

Substantial equivalence

The predicate device selected for comparison with the GUS610 Blood Pressure Monitor is:

Predicate Device:Blood Pressure Monitor Model GUS622
Sponsor:Globalcare, China
510(k) Number:K192609
Clearance Date:21 April 2020
FDA Product Code:DXN
Classification Name:Noninvasive blood pressure measurement system
Regulation No:21 CFR 870.1130
Class:II

Predicate device comparison table:

The following Table 2 provides evidence of substantial equivalence of the subject device with the selected predicate device.

Table 2: Predicate device comparison table
FeatureSubject devicePredicate deviceSimilarity
Device nameGUS610GUS622N/A
DeviceManufacturerGlobalcare, ChinaGlobalcare, ChinaN/A
510(k) ReferenceThis submissionK192609N/A
FDA ProductCodeDXNDXNSame
FDAClassificationNameNoninvasive blood pressuremeasurement systemNoninvasive blood pressuremeasurement systemSame
FDA RegulationNumber21 CFR 870.113021 CFR 870.1130Same
DevicedescriptionAutomatic upper arm bloodpressure monitor intended fornon-invasive measurement ormonitoring of adults' arterialblood pressure and pulse ratevia an upper arm cuff.Automatic upper arm bloodpressure monitor intended fornon-invasive measurement ormonitoring of adults' arterialblood pressure and pulse ratevia an upper arm cuff.Same
Indications foruseThe Globalcare GUS610automatic Blood PressureMonitor is indicated for home usefor the non-invasivemeasurement of diastolic andsystolic blood pressures andpulse rate of adults by means ofan inflatable cuff which iswrapped around the upper arm.The cuff circumference is limitedto 22 to 42 cm.The Globalcare GUS622automatic Blood PressureMonitor is indicated for home usefor the non-invasivemeasurement of diastolic andsystolic blood pressures andpulse rate of adults by means ofan inflatable cuff which iswrapped around the upper arm.The cuff circumference is limitedto 22 to 42 cm.Same
Table 2: Predicate device comparison table
FeatureSubject devicePredicate deviceSimilarity
Use environmentIndoor use, home environmentIndoor use, home environmentSame
MeasuringmethodOscillometric method, automaticinflation and measurementduring inflationOscillometric method, automaticinflation and measurementduring deflationDifferent
Devicemeasurements(outputparameters)Diastolic and systolic bloodpressurePulse rateDiastolic and systolic bloodpressurePulse rateSame
Additional outputindicationsIrregular heart beatHemodynamic instabilityIrregular heart beatHemodynamic instabilitySame
Output resultcalculationSoftware algorithmSoftware algorithmSame
MeasurementrangeCuff pressure: 0-300 mmHgSystolic: 50-280 mmHgDiastolic: 30-200 mmHgPulse: 40-200 beats/minCuff pressure: 0-300 mmHgSystolic: 50-280 mmHgDiastolic: 30-200 mmHgPulse: 40-199 beats/minSame
AccuracyPressure: ±3 mmHgPulse rate: ±5 %Pressure: ±3 mmHgPulse rate: ±5 %Same
External PowerSupplyInput 100-240 VAC 50/60 Hz 0.5A max; Output 6 VDC(+/-5%) / 600mA (optional)Input 100-240 VAC 50/60 Hz 0.5A max; Output 6 VDC(+/-5%) / 600mA (optional)Same
Unit voltage6V dc6V dcSame
Batteries4 x AAA alkaline 1.5 VDC4 x AAA alkaline 1.5 VDCSame
Battery lifeApprox 500 cyclesApprox 500 cyclesSame
StandardscomplianceIEC 60601-1:2005/AMD1:2012 +AMD2:2020IEC 60601-1-11:2015/AMD1:2020IEC 60601-1-11:2015/AMD1:2020IEC 60601-1-11:2015/AMD1:2020ANSI/AAMI ES60601-1:2005 /A2:2010IEC 60601-1-11:2015IEC 60601-1-2:2014IEC 80601-2-30:2009/AMD1:2013ISO 81060-2:2013EN 1060-3:1997+A2:2009Subject device complieswith more recent standards
Device ProtectionIEC 60601-1: Class 2IEC 60601-1: Class 2Same
Applied PartIEC 60601-1: Type BFIEC 60601-1: Type BFSame
Memory90 x 1 user60 x 2 userDifference has no effect onsafety or effectiveness
DimensionsL 85 mm x W 170 mm x H 48mmL 98 mm x W 140 mm x H 53mmDifferences have no effecton safety or effectiveness
Weight275 g device without cuff andbatteries207 g device without cuff andbatteriesDifference has no effect onsafety or effectiveness

{7}------------------------------------------------

The subject device and the predicate device have many identical or similar properties or features. The differences that exist and are identified in the above table include:

  • Blood pressure and pulse rate calculation take place during inflation in the subject device, but ● during deflation in the predicate device. This method produces a faster measurement result display and a lower cuff compression on the patient's arm.
  • . The subject device complies with more recent standards than the predicate device.
  • The memory function holds different numbers of readings. .
  • . Weight and size are different.

None of the identified differences introduce new aspects of safety or effectiveness.

{8}------------------------------------------------

Conclusion

The subject and predicate devices have very similar intended uses and fundamental technological characteristics. Any differences in technological characteristics between subject and predicate devices do not raise different questions of safety and effectiveness. The performance data demonstrate that the subject device performs in a similar manner to the predicate device. The clinical data demonstrate that the subject device performs comparably to the predicate device which is legally available in the US for the same intended use. The performance and clinical data together demonstrate that the subject device is substantially equivalent to the predicate device

§ 870.1130 Noninvasive blood pressure measurement system.

(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).