Intended Use

The Norlase ECHO Photocoagulator is intended to be used in ophthalmic laser procedures including retinal and macular photocoagulation, iridotomy, and trabeculoplasty.

Device Description

The Norlase ECHO device is a modification to the Norlase Leaf Photocoagulator (K190083) and also consists of a laser unit, a wireless control unit and a foot switch. The laser unit must be attached to the Doctors own compatible slit lamp to work as a full system. The Norlase ECHO Device is controlled by the wireless control unit that allows the Doctor to set the desired parameters for treatment. A shrouded foot switch is connected to the laser unit to control the emission of laser light. The Doctor will use the user supplied slit lamp to identify the area to be treated and after setting the appropriate parameters, will target the ophthalmic tissue utilizing a visible red aiming laser. The foot switch will be pressed by the Doctor to deliver therapeutic green laser light to the target tissue. A mechanically fixed eve safety filter is built into the Laser Unit to protect the Doctor from any stray or reflected green laser light travelling along the visual path of the slit lamp. The Norlase Echo device consists of the following modules: Laser unit + power supply + power cable, Foot switch + cable, Control unit (Tablet computer with installed Software) + power supply + USB cable. The wavelength of the treatment beam is centered around ~520 nm, and the maximum optical output power is 1.5W (to tissue). The laser is operated in continuous wave (CW) mode and electronically pulsing the laser output power to achieve durations from 50 usec to 1 second. Laser parameters are controlled by a Control Unit (tablet) User Interface that utilizes a touchscreen control to select or change a setting. Voice Control of selected parameters is an optional feature that can assist in the increment or selection of a parameter. The ECHO modification consists of capability for pattern scanning.

AI/ML Overview

The provided text is a 510(k) summary for the Norlase ECHO Photocoagulator. This document focuses on demonstrating substantial equivalence to a predicate device, rather than proving a device meets specific acceptance criteria based on clinical study performance. Medical devices cleared via the 510(k) pathway often rely on bench testing, engineering verification, and comparison to legally marketed predicate devices, especially when the changes are related to technological characteristics rather than new indications for use or significant safety/efficacy concerns.

The document explicitly states: "Clinical data was not required for this type of device." This means there was no clinical study conducted to evaluate device performance against specific acceptance criteria in a human population. Instead:

  • Engineering verification and validation were performed.
  • Bench testing was conducted to demonstrate compliance with various safety and performance standards.
  • Software testing was performed.
  • Human factors evaluation was carried out in a simulated use environment.
  • The key finding for substantial equivalence is that the differences (primarily the addition of pattern scanning) do not introduce new issues related to safety and efficacy.

Therefore, I cannot provide an answer that aligns with the structure of your requested output (acceptance criteria and proof from a clinical study) because the provided document does not contain that information. The performance data discussed is limited to engineering and bench testing, not clinical performance against ground truth established by experts.

To directly address your points based on the provided text, while acknowledging the lack of clinical study data:

  1. A table of acceptance criteria and the reported device performance:
    The document does not present "acceptance criteria" in the sense of a clinical study endpoint (e.g., sensitivity, specificity, accuracy). Instead, it lists standards the device complies with and types of testing performed. The "performance" is stated as successful verification and validation according to documented procedures.

    Acceptance Criteria (as implied by testing)Reported Device Performance (as stated in submission)
    Compliance with electrical and mechanical safety standards (e.g., IEC 60601-1)Successful bench testing for electrical and mechanical safety in compliance with cited standards.
    Compliance with laser performance standards (e.g., IEC 60825-1)Successful bench testing for laser performance in compliance with cited standards.
    Compliance with ophthalmic device standards (e.g., ISO 15004-2)Successful bench testing specific to ophthalmic devices in compliance with the cited standard.
    Specified performance of pattern scanning featureSuccessful bench testing for verification and validation of the specified performance of the pattern scanning feature.
    Software compliance (e.g., IEC 62304), including cybersecuritySuccessful software testing (verification and validation), including test cases for off-the-shelf software and cybersecurity features.
    Usability in simulated environmentSuccessful human factors evaluation to demonstrate usability in a simulated use environment when used by healthcare professionals.
    Risk mitigation effectiveness (ISO 14971)All risk mitigations satisfactorily verified and validated.
    Device meets design requirements and user needsResults of performance testing demonstrated that the device met the design requirements and user needs.
  2. Sample size used for the test set and the data provenance: Not applicable. No clinical test set. The data provenance is from internal company verification and validation testing, not from a patient population.

  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. No clinical test set requiring expert ground truth in this context.

  4. Adjudication method for the test set: Not applicable. No clinical test set.

  5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is a laser photocoagulator, not an AI diagnostic or assistance tool.

  6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is a medical device, not an AI algorithm. Its "performance" is related to its physical and software functionality, which was tested during engineering verification.

  7. The type of ground truth used: For the engineering and bench testing, "ground truth" would be established by validated measurement equipment, engineering specifications, and compliance with industry standards. For instance, a power meter would provide the "ground truth" for laser power output, and test protocols would define the "ground truth" for software functionality. There is no clinical "ground truth" (e.g., pathology, outcomes data) relevant to this 510(k) submission.

  8. The sample size for the training set: Not applicable. This is not an AI/ML device that requires a training set.

  9. How the ground truth for the training set was established: Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 11, 2022

Norlase Jan Forstberg Vice President of Regulatory Affairs and Quality Assurance Brydehusvej 13 Ballerup, Hovedstaden DK-2750 Denmark

Re: K222157

Trade/Device Name: ECHO Photocoagulator Regulation Number: 21 CFR 886.4390 Regulation Name: Ophthalmic Laser Regulatory Class: Class II Product Code: HOF Dated: July 15, 2022 Received: July 20, 2022

Dear Jan Forstberg:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Anjana Jain, PhD Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K222157

Device Name ECHO Photocoagulator

Indications for Use (Describe)

The Norlase ECHO Photocoagulator is intended to be used in ophthalmic laser procedures including retinal and macular photocoagulation, iridotomy, and trabeculoplasty.

Type of Use (Select one or both, as applicable):

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

Prescription Use (Part 21 CFR 801 Subpart D)

__ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary - K222157 Prepared September 16, 2022

Sponsor:Norlase ApSBrydehusvej 132750 BallerupDenmark
Contact Person:Jan Forstberg
Telephone:+ 45-40-109302
Submission Date:September 16, 2022
Device Name:ECHO Photocoagulator
Common Name:Photocoagulator
Classification:
Regulatory Class:II
Review Category:Ophthalmology Photocoagulator21CFR 886.4390 (HQF)
Classification Panel:Ophthalmology

A. Legally Marketed Predicate Devices

The predicate device is the Norlase Leaf Photocoagulator (K190083) also classified according to 21CFR 886.4390 (HQF).

B. Device Description:

The Norlase ECHO device is a modification to the Norlase Leaf Photocoagulator (K190083) and also consists of a laser unit, a wireless control unit and a foot switch. The laser unit must be attached to the Doctors own compatible slit lamp to work as a full system. The Norlase ECHO Device is controlled by the wireless control unit that allows the Doctor to set the desired parameters for treatment. A shrouded foot switch is connected to the laser unit to control the emission of laser light.

The Doctor will use the user supplied slit lamp to identify the area to be treated and after setting the appropriate parameters, will target the ophthalmic tissue utilizing a visible red aiming laser. The foot switch will be pressed by the Doctor to deliver therapeutic green laser light to the target tissue. A mechanically fixed eve safety filter is built into the Laser Unit to protect the Doctor from any stray or reflected green laser light travelling along the visual path of the slit lamp.

The Norlase Echo device consists of the following modules:

  • Laser unit + power supply + power cable ।
  • । Foot switch + cable
  • Control unit (Tablet computer with installed Software) + power supply + USB cable ।

{4}------------------------------------------------

The wavelength of the treatment beam is centered around ~520 nm, and the maximum optical output power is 1.5W (to tissue). The laser is operated in continuous wave (CW) mode and electronically pulsing the laser output power to achieve durations from 50 usec to 1 second. Laser parameters are controlled by a Control Unit (tablet) User Interface that utilizes a touchscreen control to select or change a setting. Voice Control of selected parameters is an optional feature that can assist in the increment or selection of a parameter. The ECHO modification consists of capability for pattern scanning.

C. Intended Use

The ECHO Photocoagulator is intended to be used in ophthalmic laser procedures including retinal and macular photocoagulation, iridotomy and trabeculoplasty.

Device NamePrimary Predicate DeviceNorlase LEAF greenlaser photocoagulatorK190083Subject DeviceNorlase ECHO greenlaser photocoagulatorSubstantialEquivalence
ManufacturerNorlaseNorlaseIdentical
510(k) numberK190083K222157N/A
510(k) decision date05/03/2019N/AN/A
Regulation medicalspecialtyOphthalmicOphthalmicIdentical
Review panelOphthalmicOphthalmicIdentical
Product codeHQFHQFIdentical
Regulation number886.4390886.4390Identical
Regulation descriptionOphthalmicphotocoagulatorOphthalmicphotocoagulatorIdentical
Classification22Identical
Indication for use andIntended useThe Norlase LEAFPhotocoagulator is intendedto be used in ophthalmicsurgical proceduresincluding retinal andmacular photocoagulation,iridotomy andtrabeculoplastyThe Norlase ECHOPhotocoagulator is intendedto be used in ophthalmicsurgical proceduresincluding retinal andmacular photocoagulation,iridotomy andtrabeculoplasty.Identical
User interfaceManual and voice controlManual and voice controlIdentical
ConnectivityWirelessWirelessIdentical
Treatment Beam
Laser typeLaser diodeLaser diodeIdentical
Wavelength520 nm520 nmIdentical
Laser modeContinuous wave (CW)Continuous wave (CW)Identical
Power output50 to 1500 mW50 - 1500 mWIdentical
Exposure time0.05 ms to 1 s.0.05 ms to 1 s.Identical
Interval time50msec-3sec50msec-3secIdentical
Cooling methodAir cooledAir cooledIdentical
Single spotYesYesIdentical
Pattern scanningNoYesSimilar

D. Substantial Equivalence

{5}------------------------------------------------

Pulse counter0 - 9,9990 - 9,999Identical
Laser beam diameter at focus50, 100, 200, 300, and 500 μm50, 100, 200, 300, and 500 μmIdentical
Other specifications:
Attachable slit lamp delivery unitFor HAAG 900BM/900BQ and similarZeiss 30SL, SL130 and similarFor HAAG 900BM/900BQ and similarZeiss 30SL, SL130 and similarIdentical
Spot size50 to 500 μm50 to 500 μmIdentical
Voltage100 to 240 Vac 50/60 Hz100 to 240 Vac 50/60 HzIdentical
Power consumption120 VA120 VAIdentical
Dimension253 (H) x 153 (W) x 43 (D) mm253 (H) x 153 (W) x 43 (D) mmIdentical
Weight4 kg4 kgIdentical

In summary, based on the comparison of indications for use and technological characteristics of the subject device is substantially equivalent to the predicate device. Based on the performance data provided in the submission these differences do not introduce new issues related to safety and efficacy.

E. Performance Data

Every specification of the ECHO Photocoagulator device has been verified and validated according to the company's documented development and test procedures. The verification and validation testing included testing to the following applicable standards:

ISO 14971Application of risk management to medical devices
ISO 15004-2Ophthalmic instruments - Fundamental requirements andtest methods - Part 1: Light hazard protection
IEC 60601-1+Medical electrical equipment- General requirements forbasic safety and essential performance
IEC 60601-1-2Electromagnetic disturbances
IEC 60601-2-22Medical electrical equipment - Part 2: Particularrequirements for the safety of diagnostic and therapeuticlaser equipment
IEC 60825-1Safety of laser products - Part 1: Equipment classification,and requirements [Including: Technical Corrigendum 1(2008), Interpretation Sheet 1 (2007), Interpretation Sheet 2(2007)]11
IEC 62304+Medical device software - Software life-cycle processes
IEC 62366-1Application of usability engineering to medical devices
ANSI Z136-1American National Standard for Safe Use of Lasers
ANSI Z136-3American National Standard for Safe Use of Lasers inHealth Care

{6}------------------------------------------------

Verification and validation testing were completed in accordance with the company's Design Control process in compliance with 21 CFR Part 820.30. Successful results for the following tests were included in the submission as performance data supporting substantial equivalence:

    1. Bench testing for electrical and mechanical safety in compliance with the electrical safety and laser standards cited above
    1. Bench testing for laser performance in compliance with the laser standards cited above
    1. Bench testing specific to ophthalmic devices in compliance with the standard cited above
    1. Bench testing for verification and validation of the specified performance of the pattern scanning feature for the subject device.
    1. Software testing, consisted of verification and validation testing in compliance with ISO 62304, including test cases related to off the shelf software as well as cybersecurity features and the software used to implement the patten scanning feature for the subject device.
    1. Human factors evaluation to demonstrate usability in a simulated use environment when used by health care professionals

To address the added capability for pattern scanning, an additional comparison was provided to a reference device. the Topcon PASCAL Synthesis TwinStar Ophthalmic Scanning Laser System (K170409).

Clinical data was not required for this type of device.

F. Conclusion

Potential risks were identified according to the ISO 14971 Standard. The risks were analyzed with regard to risk/benefit category and mitigations were implemented and tested as part of the performance testing described above. All risk mitigations were satisfactorily verified and validated. Where there were technological differences from the predicate, these were shown not to result in any new issues of safety or efficacy according to the performance data submitted.

Therefore, the ECHO Photocoagulator is substantially equivalent to the predicate device with regards to intended use and technological characteristics. Results of performance testing demonstrated that the device met the design requirements and as well as the user needs

§ 886.4390 Ophthalmic laser.

(a)
Identification. An ophthalmic laser is an AC-powered device intended to coagulate or cut tissue of the eye, orbit, or surrounding skin by a laser beam.(b)
Classification. Class II.