K Number
K222126

Validate with FDA (Live)

Date Cleared
2024-08-16

(760 days)

Product Code
Regulation Number
862.1345
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

VivaChek™ Fad Blood Glucose Monitoring System is intended to quantitatively measure the glucose concentration in fresh capillary whole blood samples drawn from the fingertips. It is intended for use by persons with diabetes at home as an aid to monitor the effectiveness of diabetes control. It is not intended for neonatal use or for the diagnosis of or screening for diabetes. This system is intended for self-testing outside the body (in vitro diagnostic use), and should only be used by a single person and should not be shared.

VivaChek™ Fad Smart Blood Glucose Monitoring System is intended to quantitatively measure the glucose concentration in fresh capillary whole blood samples drawn from the fingertips. It is intended for use by persons with diabetes at home as an aid to monitor the effectiveness of diabetes control. It is not intended for neonatal use or for the diagnosis of or screening for diabetes. This system is intended for self-testing outside the body (in vitro diagnostic use), and should only be used by a single person and should not be shared.

VivaChek™ Fad Sync Blood Glucose Monitoring System is intended to quantitatively measure the glucose concentration in fresh capillary whole blood samples drawn from the fingertips. It is intended for use by persons with diabetes at home as an aid to monitor the effectiveness of diabetes control. It is not intended for neonatal use or for the diagnosis of or screening for diabetes. This system is intended for self-testing outside the body (in vitro diagnostic use), and should only be used by a single person and should not be shared.

Device Description

VivaChek™ Fad Blood Glucose Monitoring System consists of VivaChek™ Fad Blood Glucose Meter and the VivaChek™ Fad Blood Glucose Test Strips. The glucose meter and test strips are packaged separately.

VivaChek™ Fad Smart Blood Glucose Monitoring System consists of VivaChek™ Fad Smart Blood Glucose Meter and the VivaChek™ Fad Blood Glucose Test Strips. The glucose meter and test strips are packaged separately.

VivaChek™ Fad Sync Blood Glucose Monitoring System consists of VivaChek™ Fad Sync Blood Glucose Meter and the VivaChek™ Fad Blood Glucose Test Strips. The glucose meter and test strips are packaged separately.

VivaChek Fad Control Solution, VivaChek Lancing Device, VivaChek Lancets are required for use but not included in meter box or test strips box and should be purchased separately. The VivaChek Fad Control Solution is for use with the above meter and test strip as a quality control check to verify that the meter and test strip are working together properly, and that the test is performing correctly. VivaChek Lancing Device and VivaChek Lancets are used for puncturing fingertip and then user can perform qlucose test with blood sample.

VivaChek™ Fad Blood Glucose Monitoring System, VivaChek™ Fad Smart Blood Glucose Monitoring System and VivaChek™ Fad Sync Blood Glucose Monitoring System are designed to quantitatively measure the glucose concentration in fresh capillary whole blood from the fingertip. The ducose measurement is achieved by using the amperometric detection method. The test is based on measurement of electrical current caused by the reaction of the glucose with the reagents on the electrode of the test strip. The blood sample is pulled into the tip of the test strip through capillary action. Glucose in the sample reacts with glucose dehydrogenase and the mediator. Electrons are generated, producing a current that is positive correlation to the glucose concentration in the sample. After the reaction time, the glucose concentration in the sample is displayed.

AI/ML Overview

Here's a breakdown of the requested information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The document mentions meeting "FDA SMBG OTC Guidance and industry standards" for clinical performance. A specific detailed table of acceptance criteria and reported performance is not explicitly provided in the excerpt. However, based on the context of Blood Glucose Monitoring Systems, the primary acceptance criteria would relate to the accuracy of glucose readings compared to a reference method. The document states that the clinical studies data showed that the clinical performance met the FDA SMBG OTC Guidance, implying the device successfully passed these criteria. Without the specific guidance document referenced, a detailed table cannot be created from this text alone.

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the numerical sample size used for the test set in the clinical studies. It mentions "non-professional, inexperienced lay persons" were used for user evaluations.

  • Data Provenance: The studies were conducted by Vivachek Biotech (Hangzhou) Co., Ltd, located in Zhejiang, China. The document does not explicitly state if the data was retrospective or prospective, but clinical studies (user evaluations) generally imply prospective data collection in a controlled environment.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not provided in the document. For blood glucose monitoring systems, the ground truth is typically established using a laboratory reference method (e.g., YSI analyzer), rather than expert consensus on interpretation.

4. Adjudication Method for the Test Set

This information is not provided in the document. Given that the ground truth for blood glucose is typically a laboratory reference measurement, adjudication by experts wouldn't be directly applicable in the same way it would be for image-based diagnostics.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

This is not applicable to a Blood Glucose Monitoring System. MRMC studies are relevant for AI in diagnostic imaging where human readers interpret cases. This device is a direct measurement system.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

The device itself is a standalone system for glucose measurement. The "study" here refers to the overall performance of the device in the hands of the intended users. The clinical studies (user evaluations) assess the device's performance when used by individuals (humans in the loop). The "algorithm only" concept doesn't apply as it's a physical meter and test strip system.

7. The Type of Ground Truth Used

While not explicitly stated for all studies, for blood glucose monitoring systems, the ground truth for accuracy studies is typically established using a laboratory reference method (e.g., YSI Glucose Analyzer) on venous blood samples. The document implies this by stating that clinical performance met FDA SMBG OTC Guidance, which mandates comparison to such reference methods.

8. The Sample Size for the Training Set

This information is not provided. For a physical device like a blood glucose meter, there isn't a "training set" in the same sense as machine learning algorithms. The device's calibration and performance characteristics are established during its design and manufacturing process, and then validated through laboratory and clinical studies.

9. How the Ground Truth for the Training Set Was Established

Not applicable as there is no traditional "training set" for an AI algorithm here. The performance is validated against established reference methods and clinical guidelines.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. FOOD & DRUG ADMINISTRATION".

Vivachek Biotech (Hangzhou) Co., Ltd Mark Qian Quality Director Level 2, Block 2, 146 East Chaofeng Road Yuhang Econimic Development Zone Hangzhou, 311100, Zhejiang, China

Re: K222126

Trade/Device Name: VivaChek™ Fad Blood Glucose Monitoring System, VivaChek™ Fad Smart Blood Glucose Monitoring System, VivaChek™ Fad Sync Blood Glucose Monitoring System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose Test System Regulatory Class: Class II Product Code: NBW Dated: July 21, 2023 Received: July 21, 2023

Dear Mark Qian:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

August 16, 2024

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Joshua Balsam -S

Joshua M. Balsam, Ph.D. Branch Chief Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K222126

Device Name

VivaChek™ Fad Blood Glucose Monitoring System

Indications for Use (Describe)

VivaChek™ Fad Blood Glucose Monitoring System is comprised of the VivaChek™ Fad Blood Glucose Meter and the VivaChek™ Fad Blood Glucose Test Strips.

VivaChek™ Fad Blood Glucose Monitoring System is intended to quantitatively measure the glucose concentration in fresh capillary whole blood samples drawn from the fingertips. It is intended for use by persons with diabetes at home as an aid to monitor the effectiveness of diabetes control. It is not intended for neonatal use or for the diagnosis of or screening for diabetes. This system is intended for self-testing outside the body (in vitro diagnostic use), and should only be used by a single person and should not be shared.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K222126

Device Name

VivaChek™ Fad Smart Blood Glucose Monitoring System

Indications for Use (Describe)

VivaChek™ Fad Smart Blood Glucose Monitoring System is comprised of the VivaChek™ Fad Smart Blood Glucose Meter and the VivaChek™ Fad Blood Glucose Test Strips.

VivaChek™ Fad Smart Blood Glucose Monitoring System is intended to quantitatively measure the glucose concentration in fresh capillary whole blood samples drawn from the fingertips. It is intended for use by persons with diabetes at home as an aid to monitor the effectiveness of diabetes control. It is not intended for neonatal use or for the diagnosis of or screening for diabetes. This system is intended for self-testing outside the body (in vitro diagnostic use), and should only be used by a single person and should not be shared.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Indications for Use

510(k) Number (if known) K222126

Device Name

VivaChek™ Fad Sync Blood Glucose Monitoring System

Indications for Use (Describe)

VivaChek™ Fad Sync Blood Glucose Monitoring System is comprised of the VivaChek™ Fad Sync Blood Glucose Meter and the VivaChek™ Fad Blood Glucose Test Strips.

VivaChek™ Fad Sync Blood Glucose Monitoring System is intended to quantitatively measure the glucose concentration in fresh capillary whole blood samples drawn from the fingertips. It is intended for use by persons with diabetes at home as an aid to monitor the effectiveness of diabetes control. It is not intended for neonatal use or for the diagnosis of or screening for diabetes. This system is intended for self-testing outside the body (in vitro diagnostic use), and should only be used by a single person and should not be shared.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{5}------------------------------------------------

510(k) Summary

This summary of 510(k) safety and effectiveness information is submitted in accordance with 21 CFR 807.92.

The Assigned 510(k) number is K222126.

1. Submitter's Information

1.1 Submitter's Name, Address and Telephone Number

Name: VivaChek Biotech (Hangzhou) Co., Ltd

Address: Level 2, Block 2, 146 East Chaofeng Rd.,

Yuhang Economy Development Zone,

Hangzhou, 311100, Zhejiang, China.

Telephone: +86 571 8918 9521

1.2 Contact Person

Name: Mark Qian Position: Quality Director Email: mark.qian@vivachekbio.com

1.3 Date Updated for Summary

Aug 12, 2024

2. Devices Name

2.1 Proprietary Name:

VivaChek™ Fad Blood Glucose Monitoring System VivaChek™ Fad Smart Blood Glucose Monitoring System VivaChek™ Fad Sync Blood Glucose Monitoring System

2.2 Common Name

Glucose Test System

2.3 Classification Name

Classification Name: Glucose Test System Regulation Number: 21 CFR 862.1345 Requlatory Class: Class II

{6}------------------------------------------------

Product Code: NBW

2.4 Predicate (Legally Marketed) Device

Predicate Name: TD-4183 Blood Glucose Monitoring System.

510(k) Number: K190579.

Applicant: TaiDoc Technology Corporation.

No reference device was used in this submission.

3. Devices Description

VivaChek™ Fad Blood Glucose Monitoring System consists of VivaChek™ Fad Blood Glucose Meter and the VivaChek™ Fad Blood Glucose Test Strips. The glucose meter and test strips are packaged separately.

VivaChek™ Fad Smart Blood Glucose Monitoring System consists of VivaChek™ Fad Smart Blood Glucose Meter and the VivaChek™ Fad Blood Glucose Test Strips. The glucose meter and test strips are packaged separately.

VivaChek™ Fad Sync Blood Glucose Monitoring System consists of VivaChek™ Fad Sync Blood Glucose Meter and the VivaChek™ Fad Blood Glucose Test Strips. The glucose meter and test strips are packaged separately.

VivaChek Fad Control Solution, VivaChek Lancing Device, VivaChek Lancets are required for use but not included in meter box or test strips box and should be purchased separately. The VivaChek Fad Control Solution is for use with the above meter and test strip as a quality control check to verify that the meter and test strip are working together properly, and that the test is performing correctly. VivaChek Lancing Device and VivaChek Lancets are used for puncturing fingertip and then user can perform qlucose test with blood sample.

VivaChek™ Fad Blood Glucose Monitoring System, VivaChek™ Fad Smart Blood Glucose Monitoring System and VivaChek™ Fad Sync Blood Glucose Monitoring System are designed to quantitatively measure the glucose concentration in fresh capillary whole blood from the fingertip. The ducose measurement is achieved by using the amperometric detection method. The test is based on measurement of electrical current caused by the reaction of the glucose with the reagents on the electrode of the test strip. The blood sample is pulled into the tip of the test strip through capillary action. Glucose in the sample reacts with glucose dehydrogenase and the mediator. Electrons are generated, producing a current that is positive correlation to the glucose concentration in the sample. After the reaction time, the glucose concentration in the sample is displayed.

4. Intended Use:

VivaChek™ Fad Blood Glucose Monitoring System is comprised of the VivaChek™ Fad Blood Glucose Meter and the VivaChek™ Fad Blood Glucose Test Strips. VivaChek™ Fad Blood Glucose Monitoring System is intended to quantitatively measure the glucose

{7}------------------------------------------------

VivaChek Biotech (Hangzhou) Co., Ltd Address: Level 2. Block 2. 146 East Chaofeng Rd., Yuhang Economy Development Zone, Hangzhou, 311100, Zhejiang, China Contact: Mark Qian Email: mark.gian@vivachekbio.com Tel: +86 571 8918 9521

concentration in fresh capillary whole blood samples drawn from the fingertips. It is intended for use by persons with diabetes at home as an aid to monitor the effectiveness of diabetes control. It is not intended for neonatal use or for the diagnosis of or screening for diabetes. This system is intended for self-testing outside the body (in vitro diagnostic use), and should only be used by a single person and should not be shared.

VivaChek™ Fad Smart Blood Glucose Monitoring System is comprised of the VivaChek™ Fad Smart Blood Glucose Meter and the VivaChek™ Fad Blood Glucose Test Strips. VivaChek™ Fad Smart Blood Glucose Monitoring System is intended to quantitatively measure the glucose concentration in fresh capillary whole blood samples drawn from the fingertips. It is intended for use by persons with diabetes at home as an aid to monitor the effectiveness of diabetes control. It is not intended for neonatal use or for the diagnosis of or screening for diabetes. This system is intended for self-testing outside the body (in vitro diagnostic use), and should only be used by a single person and should not be shared.

VivaChek™ Fad Sync Blood Glucose Monitoring System is comprised of the VivaChek™ Fad Sync Blood Glucose Meter and the VivaChek™ Fad Blood Glucose Test Strips. VivaChek™ Fad Sync Blood Glucose Monitoring System is intended to quantitatively measure the glucose concentration in fresh capillary whole blood samples drawn from the fingertips. It is intended for use by persons with diabetes at home as an aid to monitor the effectiveness of diabetes control. It is not intended for neonatal use or for the diagnosis of or screening for diabetes. This system is intended for self-testing outside the body (in vitro diagnostic use), and should only be used by a single person and should not be shared.

{8}------------------------------------------------

5. Comparison of Technological Characteristics with Predicate Device:

5.1 Comparison of VivaChek™ Fad Blood Glucose Monitoring System with the predicate.

CharacteristicsPredicate: TD-4183 Blood Glucose Monitoring System (K190579)Candidate: VivaChek™ Fad Blood Glucose Monitoring SystemCharacteristicsPredicate: TD-4183 Blood GlucoseMonitoring System (K190579)Candidate:VivaChek™ FadSmart Blood GlucoseMonitoring SystemCandidate:VivaChek™ Fad SyncBlood GlucoseMonitoring System
SimilaritiesSimilarities
Intended UseIt is intended for use in the quantitative measurement of glucose in fresh capillary whole blood from the fingertip. This system is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes mellitus at home as an aid in monitoring the effectiveness of the diabetes control program. It is intended to be used by single person and should not be shared. It is not intended for the diagnosis of or screening for diabetes mellitus, and is not intended for use on neonates.SameIntended UseIt is intended for use in the quantitativemeasurement of glucose in freshcapillary whole blood from thefingertip. This system is intended forself-testing outside the body (in vitrodiagnostic use) by people withdiabetes mellitus at home as an aid inmonitoring the effectiveness of thediabetes control program. It isintended to be used by single personand should not be shared. It is notintended for the diagnosis of orscreening for diabetes mellitus, and isnot intended for use on neonates.SameSame
Work (Operation) PrincipleElectrochemical biosensor technologySameWork (Operation)PrincipleElectrochemical biosensor technologySameSame
Detection MethodAmperometrySameDetection MethodAmperometrySameSame
Strip Chemical CompositionGlucose Dehydrogenase -FADSameStrip ChemicalCompositionGlucose Dehydrogenase -FADSameSame
Test MeasuredGlucose in fingertip capillary whole bloodSameTest MeasuredGlucose in fingertip capillary bloodSameSame
Sample TypeFresh capillary whole bloodSameSample TypeFresh capillary whole bloodSameSame
Unit of Measurementmg/dLSameUnit ofMeasurementmg/dLSameSame
Operating Relative Humidity10-90%SameOperatingRelative Humidity10-90%SameSame
DifferencesDifferences
Measurement Range20 to 650 mg/dL20 to 600 mg/dLMeasurementRange20 to 650 mg/dL20 to 600 mg/dL20 to 600 mg/dL
Test Time6 seconds5 secondsTest Time6 seconds5 seconds5 seconds
Sample Size0.5µL0.8µLSample Size0.5µL0.8µL0.8µL
Hematocrit Range20-65%20-70%Hematocrit Range20-65%20-70%20-70%
Strips Storage35.6-86°F, 10-85% RH36-86°F, 10-90% RHStrips Storage35.6-86°F, 10-85% RH36-86°F, 10-90% RH36-86°F, 10-90% RH
Power Source1 x 1.5V AAA batteryOne (1) CR 2032 3.0V coin cell batteryPower Source1 x 1.5V AAA batteryRechargeable,250mAh, 3.7 volts,lithium polymerbatteryRechargeable,250mAh, 3.7 volts,lithium polymerbattery

{9}------------------------------------------------

5.2 Comparison of VivaChek™ Fad Smart Blood Glucose Monitoring System and

{10}------------------------------------------------

6. Laboratory (Nonclinical) and Clinical Studies

6.1 Studies/Tests with Performance Data

Studies and testing activities included cleaning and disinfection, robustness, precision, linearity, user evaluation, usability evaluation, interference, flex studies including electrical safety test and electromagnetic compatibility test, software validation. The following laboratory and clinical studies with performance data were provided in the submission to support of the substantial equivalence determination.

No.Study/Test Item
1Cleaning and Disinfection Study
2Meter Robustness Study
3Within-Run Precision Evaluation
4Intermediate Precision Evaluation
5Linearity Evaluation Study
6User Evaluations
7Usability Evaluations
8Interference Agents Study
9Hematocrit Effect Study
10Accelerated Closed Vial Stability Studies
11Accelerated Open Vial Stability Studies
12Real Time Closed Vial Stability Studies
13Real Time Open Vial Stability Studies
14Sample Perturbation Study
15Short Sample Detection Study
16Low Battery Study
17Altitude Effect Evaluation
18Operating Conditions Evaluation
19Intermittent Sampling Study
20Vibration Test, Shock Test
21Meter Environmental Temperature Test
22Error Message Test
23Testing with Used Test Strips
24Test Strip Early Removal Validation
25Incorrect Strip Insertion Validation
26Recharging Temperature Limit Validation
27FCC Tests
28Electrical Safety Test
29Electromagnetic Compatibility Test
30Meter Software Documentation etc.

{11}------------------------------------------------

VivaChek Biotech (Hangzhou) Co., Ltd

Address: Level 2. Block 2. 146 East Chaofeng Rd., Yuhang Economy Development Zone, Hangzhou, 311100, Zhejiang, China Contact: Mark Qian Email: mark.gian@vivachekbio.com Tel: +86 571 8918 9521

6.2 Brief Discussion of Laboratory (Nonclinical) Studies:

Laboratory (nonclinical) studies were performed on the VivaChek™ Fad Blood Glucose Monitoring System, VivaChek™ Fad Smart Blood Glucose Monitoring System and VivaChek™ Fad Sync Blood Glucose Monitoring System in accordance with the FDA SMBG OTC Guidance, CLSI quidelines and industry standards. Test data and results from laboratory studies showed that these devices performed as intended and met associated guidance documents and industry standards.

6.3 Brief Discussion of Clinical Studies including Usability Evaluations:

Since the VivaChek™ Fad Sync Blood Glucose Monitoring System has the same intended use, work principle and technological characteristics as VivaChek™ Fad Smart Blood Glucose Monitoring System, clinical studies (user evaluations) were conducted with intended users using the VivaChek™ Fad Blood Glucose Monitoring System and VivaChek™ Fad Smart Blood Glucose Monitoring System. Usability evaluation was performed on VivaChek™ Fad Blood Glucose Monitoring System, VivaChek™ Fad Smart Blood Glucose Monitoring System, and VivaChek™ Fad Svnc Blood Glucose Monitoring System respectively.

Studies result indicated that non-professional, inexperienced lay persons were able to obtain blood glucose readings when using the VivaChek™ Fad Blood Glucose Monitoring System and VivaChek™ Fad Smart Blood Glucose Monitoring System. The clinical studies data showed that the clinical performance met the FDA SMBG OTC Guidance. And all participants had the ability of understanding of the labeling, could use of systems and could interpret the testing results and error messages based on the usability evaluation reports. In addition, there was no adverse effect or complication occurred or identified during the clinical studies including usability evaluations. Therefore, these studies support the intended use as described in the proposed labeling.

{12}------------------------------------------------

7. Conclusion:

The nonclinical studies and clinical studies results demonstrate that the VivaChek™ Fad Blood Glucose Monitoring System and VivaChek™ Fad Smart Blood Glucose Monitoring System performed as intended and met FDA SMBG OTC Guidance and industry standards. The different characteristics listed in Section 5 of this summary do not raise safety and effectiveness questions according to the completed and accepted nonclinical and clinical studies. VivaChek™ Fad Blood Glucose Monitoring System and VivaChek™ Fad Smart Blood Glucose Monitoring System are as safe, as effective and perform better than the predicate device.

As described in the Section 6.3 above, VivaChek™ Fad Sync Blood Glucose Monitoring System has the same intended use, work principle and technological characteristics as VivaChek™ Fad Smart Blood Glucose Monitoring System, and based on the laboratory studies results, the user evaluation result from VivaChek™ Fad Smart Blood Glucose Monitoring System, and usability evaluation for VivaChek™ Fad Sync Blood Glucose Monitoring System. VivaChek™ Fad Sync Blood Glucose Monitoring System is also as safe, as effective and performs better than the predicate device.

Therefore, VivaChek™ Fad Blood Glucose Monitoring System, VivaChek™ Fad Smart Blood Glucose Monitoring System, and VivaChek™ Fad Sync Blood Glucose Monitoring System are substantially equivalent to the predicate device.

§ 862.1345 Glucose test system.

(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.