K Number
K222075

Validate with FDA (Live)

Device Name
CellFX System
Date Cleared
2022-09-02

(50 days)

Product Code
Regulation Number
878.4400
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CellFX® System is intended for dermatological procedures requiring ablation and resurfacing of the skin.

Device Description

The Pulse Biosciences® CellFX® System is a proprietary platform technology. The CellFX System consists of the CellFX Console, CellFX Handpiece, CellFX Treatment Tips, and CellFX Software. The CellFX System delivers nanosecond duration electrical pulses that disrupt the function of cells leading to cell death, while sparing non-cellular tissue. The CellFX System delivers a series of timed, nanosecond electrical pulses (referred to as a "Cycle") to ablate and resurface tissue areas in dermatologic conditions.

The CellFX Console is capable of delivering short electric pulses at amplitudes up to 15 kV and pulse widths up to 700 ns. The electrical energy pulses are applied directly to targeted tissue using sterile Treatment Tips with stainless steel microneedles. The treatment parameters are selected by the user through a user interface on the Touchscreen Display of the CellFX Console.

AI/ML Overview

The provided document is a 510(k) summary for the CellFX System, which is an electrosurgical device for dermatological procedures. It outlines the substantial equivalence determination to a previously cleared predicate device (K211444). The document does not describe an AI/ML-driven device or an acceptance criteria table with reported device performance, sample sizes for test/training sets, expert adjudication methods, MRMC studies, or specific ground truth establishment for AI models.

Therefore, I cannot fulfill the request as the provided text lacks the necessary information about an AI/ML device and its performance evaluation. The document focuses on the CellFX System's electrical safety, biocompatibility, bench testing, software validation, and an animal study to demonstrate its safety and performance for ablation and resurfacing.

The term "acceptance criteria" is used in a general sense for biocompatibility tests and performance of the device in the animal study, but not in the context of an AI/ML model's diagnostic or predictive performance against a ground truth.

Here's what I can extract given the limitations of the provided text, related to the general performance data presented:

General Performance Data from the Document (Not AI/ML Specific):

  1. A table of acceptance criteria and the reported device performance:

    • Biocompatibility Testing:
      • Acceptance Criteria: All biocompatibility tests (Pyrogenicity, Systemic Toxicity, Sensitization, Irritation, Cytotoxicity, Hemolysis) met their respective acceptance criteria per FDA Guidance.
      • Reported Performance: "All biocompatibility tests met their respective acceptance criteria."
    • Electrical Safety and Electromagnetic Compatibility (EMC):
      • Acceptance Criteria: Compliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-2, IEC 60601-1-6, IEC 62366, and IEC 62304.
      • Reported Performance: "All electrical safety and EMC tests passed."
    • Bench Testing:
      • Acceptance Criteria: Device performs as intended and meets product specifications.
      • Reported Performance: "The performance testing for the CellFX System (Console, Handpiece and Tip) verified that the CellFX System performed as intended and met product specifications."
    • Software Verification and Validation Testing:
      • Acceptance Criteria: Supports safety and effectiveness; cybersecurity controls mitigate risk.
      • Reported Performance: "Completed software testing supports the safety and effectiveness of the device." "Cybersecurity controls have been implemented..."
    • Animal Study (Safety and Performance):
      • Acceptance Criteria (Safety): No acute procedural-related adverse events or complications. Full and acceptable healing of treated sites, no effect on draining lymph nodes or non-target organs.
      • Reported Performance (Safety): "All CellFX treatments were successfully completed with no acute procedural-related adverse events or procedural complications..." "Pathology results support the safety of the CellFX System as there was full and acceptable healing of treated sites, no effect on draining lymph nodes or nontarget organs."
      • Acceptance Criteria (Performance): All test articles meet the acceptance criteria; treatment zones successfully achieve degeneration and necrosis of active treatment sites while epidermis remains intact in majority of sites.
      • Reported Performance (Performance): "All test articles met the acceptance criteria for device performance. The treatment zones met the acceptance criteria and successfully achieved degeneration and necrosis of active treatment sites while the epidermis remained intact in the majority of the treatment sites. The device performed as intended."
  2. Sample size used for the test set and the data provenance:

    • Animal Study: 10 Yucatan mini-pigs.
    • Provenance: Animal study (swine skin). The document does not specify a country of origin, nor whether it's retrospective or prospective, but animal studies for regulatory submission are typically prospective.
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable/Not mentioned for any of the tests described. The "ground truth" for the animal study appears to be a pathological assessment of tissue changes and healing, implicitly performed by qualified veterinary pathologists, though details are not provided.
  4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable/Not mentioned.
  5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No MRMC study was mentioned. This device is not an AI-assisted diagnostic tool.
  6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is not an AI/ML device. The "performance" assessment focuses on the physical and biological effects of the electrosurgical device.
  7. The type of ground truth used:

    • For the animal study: Pathological assessment of tissue degeneration, necrosis, healing, and lack of adverse effects on lymph nodes/organs.
  8. The sample size for the training set:

    • Not applicable. This is not an AI/ML device.
  9. How the ground truth for the training set was established:

    • Not applicable. This is not an AI/ML device.

{0}------------------------------------------------

September 2, 2022

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Pulse Biosciences, Inc. Punam Gollamudi Director US Regulatory Affairs 3957 Point Eden Way Hayward. California 94545

Re: K222075

Trade/Device Name: CellFX System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: GEI Dated: July 13, 2022 Received: July 14, 2022

Dear Punam Gollamudi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

Device Name CellFX® System

Indications for Use (Describe)

The CellFX® System is intended for dermatological procedures requiring ablation and resurfacing of the skin.

Type of Use (Select one or both, as applicable)

× | Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(K) SUMMARY

510(k) Summary

I. Submitter

Applicant:Pulse Biosciences, Inc.3957 Point Eden WayHayward, CA 94545Phone: (510) 906-4649
Contact Person:Punam GollamudiDirector of US Regulatory AffairsPulse Biosciences, Inc.Phone: (415) 305-4404
Date Prepared:August 30, 2022

II. DEVICE INFORMATION

Trade Name:CellFX® System
Regulation Number21 CFR § 878.4400
Regulation Name:Electrosurgical Cutting and Coagulation Device andAccessories
Regulation Class:Class II
Product Code:GEI
Classification Panel:General and Plastic Surgery

III. PREDICATE DEVICE

K211444 - CellFX® System, Pulse Biosciences, Inc. No reference devices were used in this submission.

IV. DEVICE DESCRIPTION

The Pulse Biosciences® CellFX® System is a proprietary platform technology. The CellFX System consists of the CellFX Console, CellFX Handpiece, CellFX Treatment Tips, and CellFX Software. The CellFX System delivers nanosecond duration electrical pulses that disrupt the function of cells leading to cell death, while sparing non-cellular tissue. The CellFX System delivers a series of timed, nanosecond electrical pulses (referred to as a "Cycle") to ablate and resurface tissue areas in dermatologic conditions.

{4}------------------------------------------------

The CellFX Console is capable of delivering short electric pulses at amplitudes up to 15 kV and pulse widths up to 700 ns. The electrical energy pulses are applied directly to targeted tissue using sterile Treatment Tips with stainless steel microneedles. The treatment parameters are selected by the user through a user interface on the Touchscreen Display of the CellFX Console.

V. INDICATIONS FOR USE STATEMENT

The CellFX® System is intended for dermatological procedures requiring ablation and resurfacing of the skin.

VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

The subject device has the same technological characteristics as the predicate device. The fundamental scientific technology, energy source, materials, and processes for packaging and sterilization have not been changed from the predicate device. The device design is similar to the predicate. To support a determination of substantial equivalence, Pulse Biosciences performed verification and validation testing demonstrating the subject device performs as intended. Based on the data provided, the additional Treatment Tips do not raise any new or different questions of safety and effectiveness.

Subject DevicePredicate Device (K211444)
Trade NameCellFX® SystemCellFX® System
Prescription OnlyYesYes
Regulation Number878.4400878.4400
Product CodeGEIGEI
Indications for Use
Indications for UseStatementThe CellFX System is intended fordermatological procedures requiringablation and resurfacing of the skin.The CellFX System is intended fordermatological procedures requiringablation and resurfacing of the skin.
Technological Characteristics
Mechanism of ActionNano-Pulse Stimulation: ultrafastnanosecond electrical pulses to thetargeted tissue via treatment tips withan array of microneedlesNano-Pulse Stimulation: ultrafastnanosecond electrical pulses to thetargeted tissue via treatment tips withan array of microneedles
Pulse Frequency1-10 Pulses per second1-10 Pulses per second
PulseAmplitude/Voltage300V to 15kV300V to 15kV
Pulse Width100 ns - 700 ns100 ns - 700 ns
Power Input100-240 VAC, 50/60 Hz, 2A100-240 VAC, 50/60 Hz, 2A
Power Output30 Watts30 Watts

{5}------------------------------------------------

Energy DensityRange7.5 x 7.5 mm – 40-70 mJ/mm³10.0 x 10.0 mm– 40-70 mJ/mm³1.5 x 1.5 mm – 110-190 mJ/mm³2.5 x 2.5 mm – 60-110 mJ/mm³5.0 x 5.0 mm – 45-85 mJ/mm³
Use of LocalAnesthesiaYesYes
Sterile TipEO SterilizationEO Sterilization
HandpieceUniversal HandpieceUniversal Handpiece
System Physical Characteristics
Height132 cm132 cm
Length46 cm46 cm
Width53 cm53 cm
Weight54 kg54 kg
Power Input100-240 VAC, 50/60 Hz, 2A100-240 VAC, 50/60 Hz, 2A
Tip Characteristics
Treatment Tip3-Row2-Row
MicroneedleConfiguration
1.5 x 1.5 mmN/A2 rows of 3 needles per row
2.5 x 2.5 mmN/A2 rows of 3 or 4 needles per row
5.0 x 5.0 mmN/A2 rows of 4 or 6 needles per row
7.5 x 7.5mm3 rows of 9 needles per rowN/A
10 x 10mm3 rows of 11 needles per rowN/A
Tip Insertion Depth1 mm and 2 mm1 mm and 2 mm

VII. PERFORMANCE DATA

The following performance data was referenced or provided in support of the substantial equivalence determination.

Biocompatibility Testing

The testing was conducted in accordance with FDA Guidance, issued September 4, 2020: Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process." All biocompatibility tests met their respective acceptance criteria.

The battery of testing included the following:

  • · Pyrogenicity, Material Mediated
  • · Systemic Toxicity
  • · Sensitization
  • Irritation
  • · Cytotoxicity
  • Hemolysis

All biocompatibility tests met their respective acceptance criteria.

{6}------------------------------------------------

Electrical safety and electromagnetic compatibility (EMC)

The CellFX System complies with IEC 60601-1. IEC 60601-1-2. 60601-2-2. IEC 60601-1-6, IEC 62366, and IEC 62304. All test conditions were performed as outlined within the IEC standards. All electrical safety and EMC tests passed.

Bench Testing

The performance testing for the CellFX System (Console, Handpiece and Tip) verified that the CellFX System performed as intended and met product specifications.

Software Verification and Validation Testing

Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "moderate" level of concern since a failure of the device could directly result in minor injury to the patient or operator. Completed software testing supports the safety and effectiveness of the device.

Cybersecurity controls have been implemented to mitigate the risk of malware being introduced into the CellFX System as recommended by FDA's Draft Guidance for Industry and FDA Staff, "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices."

Animal Study

Safety and Performance Evaluation of the CellFX® System Using 3 Row Tips in Swine Skin

An animal safety study was conducted to evaluate the safety and performance of the CellFX System utilizing the 7.5 mm and 10.0 mm Treatment Tips at acute and chronic timepoints. The treatment zones were characterized across a range of energy settings (minimum, default, and maximum). The study was conducted on 10 Yucatan mini-bigs that underwent skin treatments with the subject device including 3-Row CellFX Treatment Tips (7.5 mm and 10.0 mm). Animals were divided into five cohorts of two animals each for the following timepoints: 0, 2, 6, 14, and 30 days.

Safety

All CellFX treatments were successfully completed with no acute procedural-related adverse events or procedural complications during the skin treatment or in-life phase of the study. Pathology results support the safety of the CellFX System as there was full and acceptable healing of treated sites, no effect on draining lymph nodes or nontarget organs.

Performance

All test articles met the acceptance criteria for device performance. The treatment zones met the acceptance criteria and successfully achieved degeneration and necrosis of active treatment sites while the epidermis remained intact in the majority of the treatment sites. The device performed as intended.

{7}------------------------------------------------

Conclusion

CellFX treatments in the swine skin using the 3-Row Treatment Tips were shown to be safe and performed as intended. Therefore, this animal study data confirmed that the device performs as intended and does not raise safety concerns.

Summary

Based on the animal safety study, the CellFX System was found to have a safety and performance profile that is equivalent to the predicate device and does not raise any different questions of safety or effectiveness.

CONCLUSION

The CellFX System has the same indication for use, technological characteristics and principles of operation as its predicate device. The non-clinical performance testing and in vivo animal data provided in this submission demonstrate and support that the CellFX System is as safe and as effective as the predicate device. Therefore, the CellFX System is substantially equivalent to the predicate device.

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.