K Number
K222068

Validate with FDA (Live)

Manufacturer
Date Cleared
2022-10-12

(90 days)

Product Code
Regulation Number
884.5300
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Durex Penck III Regular condom is used for contraception and for prophylactic purposes (to help reduce the risk of pregnancy and the transmission of sexually transmitted infections, STIs).

Device Description

Durex Penck III Regular are natural rubber latex-based condoms that completely cover the penis with a closely fitted membrane. Durex Penck III Regular are made of a non-colored natural rubber latex with silicone lubricant and are taper shaft, teat ended smooth shaped condoms. Durex Penck III Regular have a nominal length of 195 mm, width of 56 mm, and thickness of 48 um and are provided in Regular and Extra Lubricant varieties with 400 mg and 480 mg silicone lubricant respectively. The condoms are packaged in individually sealed flexible laminate foils made of polyethylene terephthalate, polyethylene, and aluminum. The foils are packaged in an outer consumer cardboard carton. The number of condoms in the carton may vary. Durex Penck III Regular condoms are intended for over-the-counter (OTC) use. These condoms conform with FDA-recognized standards ASTM D3492-16 and ISO 4074:2015.

AI/ML Overview

The provided document is a 510(k) Summary for a medical device called "Durex Penck III Regular," which is a male natural rubber latex condom. It describes the device, its intended use, comparison with a predicate device, and a summary of non-clinical performance testing.

Here's an analysis of the acceptance criteria and study information:

1. Table of Acceptance Criteria and Reported Device Performance

The document explicitly states that "The Durex Penck III Regular was tested and met all the requirements of ISO 4074:2015 - Natural rubber latex male condoms – Requirements and test methods and ASTM D3492-16 - Standard Specification for Rubber Contraceptives (Male Condoms)."

Therefore, the acceptance criteria are defined by these two international standards. The document also provides specific performance characteristics for the Durex Penck III Regular and compares them to the predicate device.

Acceptance CriterionDurex Penck III Regular Performance (Reported)Predicate Device Performance
MaterialNatural Rubber LatexNatural Rubber Latex
WidthNominal: $56 \pm 2$ mm, Mid Body: $54 \pm 2$ mmDurex Penck Standard: $54 \pm 2$ mm, Durex Penck XL: $57 \pm 2$ mm
Nominal Length$195 \pm 10$ mm$190 \pm 10$ mm
Nominal Thickness$0.048 \pm 0.008$ mmDurex Penck Standard: 0.048 mm, Durex Penck XL: $0.050 +0.004 / -0.002$ mm
LubricantSiliconeSilicone
Lubricant QuantityRegular: $400 \pm 50$ mg, Extra Lubricant: $480 \pm 50$ mgDurex Penck Standard: $400 \pm 50$ mg, Durex Penck XL: $480 \pm 50$ mg
Air Burst Pressure$> 1.0$ kPa$> 1.0$ kPa
Air Burst Volume$18.0$ LDurex Penck Standard: Minimum 18.0 L, Durex Penck XL: Minimum 22.0 L
SterilizationNon-SterileNon-sterile
TextureTaper shaft, teat ended smooth condomSmooth, straight walled, and teat ended
Shelf Life5 Years5 Years
Color AdditivesN/AN/A
Flavor AdditivesN/AN/A
BiocompatibilityNon-irritating, non-sensitizing, not acutely systemically toxic (met ISO 10993-5, -10, -11 requirements)(Not explicitly detailed for predicate, but similar expectations apply for condoms)
Physical PerformanceMet all requirements of ISO 4074:2015 and ASTM D3492-16(Predicate also conforms to these standards as implied by its clearance)
Shelf Life StabilityAll samples met predefined acceptance criteria based on accelerated stability evaluations per 21 CFR 801.435(Similar evaluations would have been done for predicate)

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify the exact sample sizes for each test (e.g., air burst, physical dimensions, biocompatibility). It states that the device was "tested and met all the requirements."

For biocompatibility, the methods listed (ISO 10993 series) often involve in-vivo (animal) and in-vitro studies. For physical performance (ISO 4074 and ASTM D3492), samples would be taken from production lots.

Data Provenance: The document does not explicitly state the country of origin of the data. Given it's a 510(k) submission to the US FDA, the tests would have been performed by a recognized laboratory, potentially in various locations. The studies are retrospective in the sense that they are performed on manufactured devices and evaluated against established standards.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not applicable to this type of device and study. The "ground truth" for condom performance is established by recognized international and national standards (ISO 4074:2015 and ASTM D3492-16). These standards define the test methodologies and acceptance criteria for physical properties and safety. Expert consensus in medical imaging or diagnosis is not relevant here. The experts involved would be those who author and maintain these standards, and laboratory technicians and engineers who perform the tests according to these standards.

4. Adjudication Method for the Test Set

This is not applicable. Adjudication methods like 2+1 or 3+1 are used in studies where human readers evaluate medical images or clinical outcomes that might have subjective interpretations. For objective physical and chemical testing of medical devices like condoms, there is no "adjudication" in this sense. The results are quantitative and compared directly to predefined numerical criteria in the standards.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size

No, an MRMC comparative effectiveness study was not done. This type of study is typically used for diagnostic imaging devices where the performance of human readers, sometimes aided by AI, is evaluated against a ground truth. The Durex Penck III Regular is a physical barrier device, not a diagnostic tool, so an MRMC study is not relevant.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

No, a standalone "algorithm only" study was not done. This concept applies to AI/ML-driven devices. The Durex Penck III Regular is a physical medical device whose performance is assessed through physical, chemical, and biological testing as per established standards, not through algorithmic evaluation.

7. The Type of Ground Truth Used

The "ground truth" for the Durex Penck III Regular device is based on established international and national standards and regulatory requirements.

  • Physical Performance: ISO 4074:2015 (Natural rubber latex male condoms – Requirements and test methods) and ASTM D3492-16 (Standard Specification for Rubber Contraceptives (Male Condoms)). These standards specify the test methods and the acceptable ranges/thresholds for properties like dimensions, air burst pressure/volume, and integrity.
  • Biocompatibility: ISO 10993-1:2009/R 2014, ISO 10993-5:2009/R 2014, ISO 10993-10:2010/R 2014, and ISO 10993-11:2017. These standards define the acceptable biological responses for device materials.
  • Shelf Life: 21 CFR 801.435 for accelerated stability evaluations.

8. The Sample Size for the Training Set

This information is not applicable. The concept of a "training set" is relevant for machine learning algorithms. The Durex Penck III Regular is a physical medical device, not an AI/ML algorithm. Its performance is validated through defined physical, chemical, and biological tests, not by training a model on a dataset.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable for the same reasons as #8. There is no training set for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 12, 2022

RB Health (US) LLC Kyle Prince Regulatory Associate 399 Interpace Parkway Parsippany, NJ 07054

Re: K222068

Trade/Device Name: Durex Penck III Regular Regulation Number: 21 CFR§ 884.5300 Regulation Name: Condom Regulatory Class: II Product Code: HIS Dated: July 13, 2022 Received: July 14, 2022

Dear Kyle Prince:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Monica D. Garcia, Ph.D. Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K222068

Device Name Durex Penck III Regular

Indications for Use (Describe)

The Durex Penck III Regular condom is used for contraception and for prophylactic purposes (to help reduce the risk of pregnancy and the transmission of sexually transmitted infections, STIs).

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summarv K222068 Durex Penck III Regular

1. Submitter Information

Applicant:RB Health (US) LLC
Address:399 Interpace Parkway ParsippanyNJ 07054-0024
Phone:(862) 325-0012

2. Correspondent Information

Company:RB Health (US) LLC
Contact:Kyle Prince
Phone:(862) 325-0012
Email:kyle.prince@reckitt.com

3. Date prepared: October 7, 2022

4. Device Information

Device Name:Durex Penck III Regular
Common Name:Male Natural Rubber Latex Condom
Regulation Number:21 CFR 884.5300
Regulation Name:Condom
Product Code:HIS (Condom)
Regulatory Class:Class II

5. Predicate Device Information

Device Name:Durex Penck Standard, Durex Penck X
510(k) Number:K200672
Sponsor:RB Health (US) LLC
Manufacturer:Reckitt Benckiser LLC

The predicate device has not been subject to a design-related recall.

6. Device Description

Durex Penck III Regular are natural rubber latex-based condoms that completely cover the penis with a closely fitted membrane. Durex Penck III Regular are made of a non-colored natural rubber latex with silicone lubricant and are taper shaft, teat ended smooth shaped condoms. Durex Penck III Regular have a nominal length of 195 mm, width of 56 mm, and thickness of 48 um and are provided in Regular and Extra Lubricant varieties with 400 mg and 480 mg silicone lubricant respectively. The condoms are packaged in individually sealed flexible laminate foils made of polyethylene terephthalate, polyethylene, and aluminum. The foils are packaged in an outer consumer cardboard carton. The number of condoms in the carton may vary. Durex Penck III Regular condoms are intended for over-the-counter (OTC) use. These condoms conform with FDA-recognized standards ASTM D3492-16 and ISO 4074:2015.

{4}------------------------------------------------

Device specifications are listed in Table 1 below.

7. Indications for Use Statement

The Durex Penck III Regular condom is used for contraception and for prophylactic purposes (to help reduce the risk of pregnancy and the transmission of sexually transmitted infections, STIs).

8. Comparison of Intended Use and Technological Characteristics with the Predicate Device

The table below includes a comparison of the intended use and technological characteristics of the subject and predicate devices.

Subject DeviceDurex Penck III RegularK222068Predicate DeviceDurex Penck Standard,Durex Penck XLK200672
Device & Predicate DeviceDurex Penck III RegularDurex Penck StandardDurex Penck XL
510(K) NumberK222068K200672
Product CodeHISHIS
Regulation Number21 CFR 884.530021 CFR 884.5300
Regulation NameCondomCondom
Indications for UseThe Durex Penck III Regularcondom is used forcontraception and forprophylactic purposes (tohelp reduce the risk ofpregnancy and thetransmission of sexuallytransmitted infections, STIs).The Durex Penck StandardCondom is used forcontraception and forprophylactic purposes (tohelp reduce the risk ofpregnancy and thetransmission of sexuallytransmitted infections, STIs).The Durex Penck XLCondom is used forcontraception and forprophylactic purposes (tohelp reduce the risk ofpregnancy and thetransmission of sexuallytransmitted infections, STIs).
Prescription or Over-The-Counter UseOver-The-CounterOver-The-Counter
Condom MaterialNatural Rubber LatexNatural Rubber Latex
WidthNominal: $56 \pm 2$ mmMid Body: $54 \pm 2$ mmDurex Penck Standard: $54 \pm 2$ mmDurex Penck XL: $57 \pm 2$ mm
Nominal Length$195 \pm 10$ mm$190 \pm 10$ mm
Nominal Thickness$0.048 \pm 0.008$ mmDurex Penck Standard: 0.048 mm

Table 1. Comparison of Intended Use and Technological Characteristics

{5}------------------------------------------------

- 0.002 mm
Durex Penck XL:0.050 +0.004 /- 0.002 mm
LubricantSiliconeSilicone
Lubricant QuantityDurex Penck III Regular:400 ± 50 mgDurex Penck III RegularExtra Lubricant:480 ± 50 mgDurex Penck Standard: 400 ±50 mgDurex Penck XL: 480 ± 50mg
Air Burst Pressure> 1.0 kPa> 1.0 kPa
Air Burst Volume18.0 LDurex Penck Standard:Minimum 18.0 LDurex Penck XL: Minimum22.0 L
SterilizationNon-SterileNon-sterile
TextureTaper shaft, teat endedsmooth condomSmooth, straight walled, andteat ended
Shelf Life5 Years5 Years
Color AdditivesN/AN/A
Flavor AdditivesN/AN/A

The subject and predicate device have similar indications for use and have the same intended use. The technological characteristics of the subject device and predicate device are similar in that they are natural rubber latex-based, are lubricated with silicone, and have the same shelf-life duration. The subject and predicate devices do have different technological characteristics, including different dimensions and specifications (e.g., burst volume). However, the different technological characteristics of the subject devices do not raise different questions of safety and effectiveness.

9. Summary of Non-Clinical Performance Testing

Biocompatibility:

Biocompatibility studies, including Acute Systemic Toxicity, Vaginal Irritation Testing, Cytotoxicity and Sensitization testing were performed in accordance with the 2020 FDA guidance document Use of International Standard ISO 10993-1, "Biological Evaluation of Medical Devices – Part 1: Evaluation and testing within a risk management process" and ISO 10993-1:2009 as follows:

  • · Cytotoxicity (ISO 10993-5:2009/R 2014)
  • · Sensitization (ISO 10993-10:2010/R 2014)
  • · Vaginal Irritation (ISO 10993-10:2010/R 2014)
  • · Acute Systemic Toxicity (ISO 10993-11:2017)

The results of testing demonstrate that the subject device is non-irritating, non-sensitizing, and not acutely, systemically toxic.

{6}------------------------------------------------

Physical Performance Testing:

The Durex Penck III Regular was tested and met all the requirements of ISO 4074:2015 - Natural rubber latex male condoms – Requirements and test methods and ASTM D3492-16 - Standard Specification for Rubber Contraceptives (Male Condoms).

Shelf Life:

The Durex Penck III Regular has a five-year shelf life based on the results of accelerated stability evaluations conducted as required in 21 CFR 801.435. All samples met predefined acceptance criteria.

10. Conclusion

The results of the performance testing described above demonstrate that the Durex Penck III Regular is as safe and effective as the predicate device and supports a determination of substantial equivalence.

§ 884.5300 Condom.

(a)
Identification. A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.(b)
Classification. (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.