K Number
K222046

Validate with FDA (Live)

Date Cleared
2022-09-30

(81 days)

Product Code
Regulation Number
884.5470
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The unscented menstrual tampon is intended for insertion into the vagina for the absorption of menstrual or other vaginal discharge.

Device Description

The proposed device is a traditional unscented menstrual tampon, and it has two types (i.e. digital tampon and applicator tampon). Each device consists of a tampon, including an absorbent pledget ("absorbent core") completely surrounded by overwap ("security veil") and a removal string ("withdrawal cord"), and an applicator (only for the applicator tampon). The tampon is of the traditional cylindrical, bullet-like shape. The applicator has a smooth, rounded tip to ease insertion. The proposed device is provided in 4 absorbencies: light (≤6g), regular (69g), super (912g) and super plus (12~15g). Each device is individually wrapped and packaged in sealed multi-unit containers for retail sale. And it is provided non-sterile and for single use only.

AI/ML Overview

The provided text describes the 510(k) premarket notification for an Unscented Menstrual Tampon (K222046). It outlines the device's characteristics, its comparison to a predicate device, and the non-clinical testing performed to establish substantial equivalence.

Here's a breakdown of the requested information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The document references established guidance documents for menstrual tampons and pads for its performance characteristics. While specific numerical acceptance criteria are not explicitly listed in a consolidated table within the provided text, it states that "All samples met the predefined acceptance criteria" for performance testing.

Acceptance Criteria CategorySpecific Tests ConductedReported Device Performance
Performance TestingDimensionsMet predefined acceptance criteria
AbsorbencyMet predefined acceptance criteria
Removal string strengthMet predefined acceptance criteria
Fiber sheddingMet predefined acceptance criteria
Tampon integrityMet predefined acceptance criteria
Chemical residuesMet predefined acceptance criteria
Biocompatibility TestingIn vitro cytotoxicity (ISO 10993-5:2009)Non-cytotoxic
Skin sensitization (ISO 10993-10:2010)Non-sensitizing
Vaginal irritation (ISO 10993-10:2010)Non-irritating
Acute systemic toxicity (ISO 10993-11:2017) (tampon only)Non-systemically toxic
Microbiology TestingEnhancement of Staphylococcus aureus growthDoes not enhance growth
Increase of Toxic Shock Syndrome Toxin-1 (TSST-1) productionDoes not increase production
Alteration of normal vaginal microflora growthDoes not alter growth

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the sample sizes used for each specific test in the performance, biocompatibility, or microbiology testing. It only mentions "All samples met the predefined acceptance criteria."

The data provenance is from non-clinical testing performed by Suzhou Borage Medical Technology Co., Ltd. in China, or by their designated testing facilities. This is retrospective in the sense that the testing was conducted on manufactured devices for the purpose of the 510(k) submission.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not applicable as the described study is a non-clinical device performance and safety evaluation for a menstrual tampon. It does not involve establishing ground truth from expert interpretation of data like images or patient outcomes. The "ground truth" here is the objective measurement of physical, chemical, and biological properties according to established standards.

4. Adjudication Method for the Test Set

This is not applicable as the described study is a non-clinical device performance and safety evaluation. Adjudication methods like 2+1 or 3+1 are typically used for expert review of clinical data, not for objective laboratory testing.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done, If So, What was the Effect Size of How Much Human Readers Improve with AI vs without AI Assistance

This is not applicable. The submission is for a physical medical device (unscented menstrual tampon) and does not involve AI or human readers for diagnostic interpretation.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was Done

This is not applicable. The submission is for a physical medical device (unscented menstrual tampon) and does not involve any algorithm or AI performance.

7. The Type of Ground Truth Used

The "ground truth" for this study is based on objective measurements and laboratory testing conforming to international standards (ISO 10993 series) and FDA guidance documents specific to menstrual tampons. This includes:

  • Physical measurements: Dimensions, absorbency, removal string strength, fiber shedding, tampon integrity.
  • Chemical analysis: Chemical residues.
  • Biological testing: In vitro cytotoxicity, skin sensitization, vaginal irritation, acute systemic toxicity, and specific microbiological tests for Staphylococcus aureus growth, TSST-1 production, and impact on vaginal microflora.

8. The Sample Size for the Training Set

This is not applicable. There is no "training set" as this is a physical device evaluation, not an algorithm or AI development.

9. How the Ground Truth for the Training Set Was Established

This is not applicable for the same reason as point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

September 30, 2022

Suzhou Borage Medical Technology Co., Ltd. % Grace Liu Consultant Shenzhen Joyantech Consulting Co. Ltd 1713A, 17th Floor, Block A, Zhongguan Times Square, Nanshan District Shenzhen, Guangdong 518000 China

Re: K222046

Trade/Device Name: Unscented Menstrual Tampon Regulation Number: 21 CFR§ 884.5470 Regulation Name: Unscented Menstrual Tampon Regulatory Class: II Product Code: HEB Dated: June 13, 2022 Received: July 11, 2022

Dear Grace Liu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jason R. Roberts, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K222046

Device Name Unscented Menstrual Tampon

Indications for Use (Describe)

The unscented menstrual tampon is intended for insertion into the absorption of menstrual or other vaginal discharge.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary - K222046

1. Contact Details

1.1 Applicant information

Applicant NameSuzhou Borage Medical Technology Co., Ltd.
AddressNo.9 Shiheshan Road, Dongshan Town, Wuzhong District,Suzhou City, Jiangsu Province, 215107, China
Contact personZhou Nana
Phone No.+86-512-66280008
E-mailznn@borage.com.cn
Date Prepared2022-09-27

1.2 Submission Correspondent

Image: LogoShenzhen Joyantech Consulting Co., Ltd
1713A, 17th Floor, Block A, Zhongguan Times Square, Nanshan
卓远天成District, Shenzhen, Guangdong Province, China
Phone No.+86-755-86069197
Contact personGrace Liu; Field Fu;
Contact person's e-mailgrace@cefda.com; field@cefda.com
Websitehttp://www.cefda.com

2. Device Information

Trade nameUnscented Menstrual Tampon
Common nameUnscented Menstrual Tampon
ClassificationII
Classification nameTampon, Menstrual, Unscented
Product codeHEB
Regulation No.21 CFR 884.5470

3. Legally Marketed Predicate Device

Trade NameTosama 100% Organic Cotton Menstrual Tampon
510(k) NumberK15117
Product CodeHEB
ManufacturerTOSAMA, d.o.o.

The predicate device has not been subject to a design-related recall.

4. Device Description

The proposed device is a traditional unscented menstrual tampon, and it has two types (i.e. digital tampon and applicator tampon). Each device consists of a tampon, including an absorbent pledget ("absorbent core") completely surrounded by overwap ("security veil") and a removal string ("withdrawal cord"), and an applicator (only for the applicator tampon). The tampon is of the traditional cylindrical, bullet-like shape. The applicator

{4}------------------------------------------------

has a smooth, rounded tip to ease insertion.

The proposed device is provided in 4 absorbencies: light (≤6g), regular (69g), super (912g) and super plus (12~15g). Each device is individually wrapped and packaged in sealed multi-unit containers for retail sale. And it is provided non-sterile and for single use only.

5. Intended Use/Indication for Use

The unscented menstrual tampon is intended for insertion into the vagina for the absorption of menstrual or other vaginal discharge.

6. Technological Characteristics Comparison

Comparison itemProposed Device(K222046)Predicate Device(K15117)Comment
ManufacturerSuzhou Borage MedicalTechnology Co., Ltd.TOSAMA, d.o.o.None
Product NameUnscented MenstrualTamponsTosama 100% OrganicCotton Menstrual TamponNone
Product CodeHEBHEBSame
Regulation Number21 CFR § 884.547021 CFR § 884.5470Same
ClassificationClass IIClass IISame
Intended Use/ Indications for UseThe unscented menstrualtampon is intended forinsertion into the vagina fortheabsorptionofmenstrual or other vaginaldischarge.The Tosama 100% OrganicCotton Menstrual Tampon isintended for insertion intothevaginafortheabsorption of menstrual orother vaginal discharge.Same
Single UseYesYesSame
SterilityNon-sterileNon-sterileSame
DesignTampon with cylindricalshape and bullet-like tip.Applicator withsmooth,rounded tip.Tampon with cylindricalshape and bullet-like tip.Applicator withsmooth,rounded tip.Same
Absorbency
Light≤6gN/A
Regular6~9g6~9gDifferent
Super9~12g9~12g
Super plus12~15g12~15g
Product dimensions
Digital tampon
LightPledget length(40~50) mmN/ADifferent
Pledget diameter(9.5~12.5) mm
Removal stringlength(125~165) mm
Pledget length(40~50) mm(43.0~46.0) mm
RegularPledget diameter(10.5~13.5) mm(11.0~12.0) mm
Removal stringlength(125~165) mm(130~160) mm
Pledget length(45~55) mm(48.0~51.0) mm
SuperPledget diameter(11.5~14.5) mm(12.0~13.0) mm
Removal stringlength(125~165) mm(130~160) mm
Pledget length(45~55) mm(48.0~51.0) mm
Super plusPledget diameter(13.3~16.3) mm(14.0~15.0) mm
Removal stringlength(125~165) mm(130~160) mm
Applicator tampon
Pledget length(40~50) mm
Pledget diameter(9.5~12.5) mm
LightRemoval stringlength(125~165) mmN/A
Applicator length125 mm
Applicator diameter15 mm
Pledget length(40~50) mm(43.0~46.0) mm
Pledget diameter(10.5~13.5) mm(11.0~12.0) mm
RegularRemoval stringlength(125~165) mm(130~160) mm
Applicator length125 mm120 mm
Applicator diameter15 mm13 mm
Pledget length(45~55) mm(43.0~46.0) mm
Pledget diameter(11.5~14.5) mm(12.0~13.0) mm
SuperRemoval stringlength(125~165) mm(130~160) mm
Applicator length126 mm120 mm
Applicator diameter16.5 mm16 mm
Super plusPledget length(45~55) mm(43.0~46.0) mm
Pledget diameter(13.3~16.3) mm(14.0~15.0) mm
Removal stringlength(125~165) mm(130~160) mm
Applicator length126 mm120 mm
Applicator diameter16.5 mm18 mm
ComponentPledget100% Viscose100% Organic CottonDifferent
MaterialsOverwapPolyethylene andPolyethyleneterephthalate100% Organic Cotton
Removal stringPolyester and Viscose100% Organic Cotton
ApplicatorPolyethylene andPolypropyleneTPO
AdditivesandFinishingAgentsAnti-wicking agentof removal stringParaffin-based resinParaffin-based resinSame
Finishing agent ofpledgetPolyoxyethylene stearateNot availableDifferent
Complies with ISO 10993-1YesYesSame
Compliance with microbiologyrequirements of FDA Guidance forTamponsYesYesSame
LabelingComplied with 21 CFR part801Complied with 21 CFR part801Same

Table 1 Technological Characteristics Comparison Table

{5}------------------------------------------------

{6}------------------------------------------------

The proposed device has the similar indication for use as the predicate device as well as comparable technical characteristics, and the differences don't raise any additional questions for safety and effectiveness. Therefore, the subject device has the same intended use as the predicate.

7. Summary of Non-clinical Testing

Non-clinical testing was conducted to verify that the proposed device can meet all design specifications and perform similarly to the predicate device. The following tests were conducted.

Performance Testing

The following performance characteristics were assessed in accordance with the FDA guidance document "Guidance for Industry and FDA Staff - Menstrual Tampons and Pads: Information for Premarket Notification Submissions (510(k)s)" issued on July 27, 2005.

  • Dimensions

  • Absorbency

  • Removal string strength

  • Fiber shedding

  • Tampon integrity

  • Chemical residues

All samples met the predefined acceptance criteria.

Biocompatibility Testing

{7}------------------------------------------------

Biocompatibility studies were performed in accordance with the FDA guidance document "Guidance for Industry and Food and Drug Administration Staff - Use of International Standard ISO 10993-1, 'Biological Evaluation of Medical Devices - Part 1: Evaluation and testing within a risk management process" issued on September 4, 2020 as follows:

  • In vitro cytotoxicity test per ISO 10993-5:2009

  • Skin sensitization test per ISO 10993-10:2010

  • Vaginal irritation test per ISO 10993-10:2010

  • Acute systemic toxicity test per ISO 10993-11:2017 (tampon only)

All tests were performed on the tampon and applicator separately. The results demonstrated that the proposed device is non-cytotoxic, non-irritating, non-sensitizing, and non-systemically toxic.

Microbiology Testing

Per the FDA guidance document "Guidance for Industry and FDA Staff - Menstrual Tampons and Pads: Information for Premarket Notification Submissions (510(k)s)" issued on July 27, 2005, the following microbiology testing was conducted on the final, finished form of the proposed device, and the test results showed that the proposed device does not:

  • enhance the growth of Staphylococcus aureus;

  • increase the production of Toxic Shock Syndrome Toxin-1 (TSST-1);

  • alter the growth of normal vaginal microflora.

8. Clinical Testing

No clinical study is included in this submission.

9. Conclusions

The nonclinical tests demonstrate that the proposed device is as safe and effective, as the legally marketed device (K15117). Therefore, the subject device is substantially equivalent to the predicate.

§ 884.5470 Unscented menstrual tampon.

(a)
Identification. An unscented menstrual tampon is a device that is a plug made of cellulosic or synthetic material that is inserted into the vagina and used to absorb menstrual or other vaginal discharge. This generic type of device does not include menstrual tampons treated with scent (i.e., fragrance materials) or those with added antimicrobial agents or other drugs.(b)
Classification. Class II (performance standards).