Intended Use

AVATAR MEDICAL Software V1 is intended as a medical imaging system that allows the processing, review, analysis, communication and media interchange of multi-dimensional digital images acquired from CT or MR imaging devices. It is also intended as software for preoperative surgical planning, and as software for the intraoperative display of the aforementioned multi-dimensional digital images. AVATAR MEDICAL Software V1 is designed for use by health care professionals and is intended to assist the clinician who is responsible for making all final patient management decisions.

Device Description

The Avatar Medical Software V1 (AMS V1) is a software-only device that allows trained medical professionals to review CT and MR image data in three-dimensional (3D) format and/or in virtual reality (VR) interface. The 3D and VR images are accessible through the software desktop application and, if desired, through compatible VR headsets which are used by users for preoperative surgical planning and for display during intervention/surgery.

The AMS V1 product is to be used to assist in medical image review. Intended users are trained medical professionals, including imaging technicians, clinicians and surgeons.

AMS V1 includes two main software-based user interface components, the Desktop Interface and the VR Interface. The Desktop Interface runs on a compatible off-the-shelf (OTS) workstation provided by the hospital and only accessed by authorized personnel. The Desktop Interface contains a graphical user interface where a user can retrieve DICOM-compatible medical images locally or on a Picture Archiving Communication System (PACS). Retrieved CT and MR images can be viewed in 2D and 3D formats. Users are able to make measurements, annotations, and apply fixed and manual image filters.

The VR Interface is accessible via a compatible OTS headset to allow users to review the medical images in a VR format. VR formats can be viewed only when the user connects a compatible VR headset directly to the workstation being used to view the Desktop Interface. Additionally. AMS V1 enables the intended users to remotely stream the Desktop Interface to another workstation on the same local area network (LAN).

The 3D images generated using AMS V1 are intended to be used in relation to surgical procedures in which CT or MR images are used for preoperative planning and/or during intervention/surgery.

The intraoperative use of the AMS V1 solely corresponds to the two following cases:

  • Display of the AMS V1 Desktop Interface on existing monitors/screens in the operating room
  • Use in a non-sterile image review room accessible from the operating room during the procedure (AMS V1 operates on VR headsets which are not approved to be used in the sterile environment of the operating room)
AI/ML Overview

Here's a detailed breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Feature/FunctionTool in Subject Device: AMS V1Tool in Reference Device: Osirix MD (K101342)Acceptance CriteriaReported Device Performance
Linear Measurements (polylines)CurveClose PolygonNo statistical difference between distributions of measurements obtained for AMS V1 or the reference device Osirix MD, as evaluated per t-test statistics. Tests performed for a series of objects in reference MR and CT phantom images.No statistical difference was reported, as implied by the statement "No statistical difference between distributions of measurements obtained for AMS V1 or the reference device Osirix MD, as evaluated per t-test statistics. Tests performed for a series of objects in reference MR and CT phantom images." The device met this criterion.
Diameter MeasurementsRulerLengthNo statistical difference between distributions of measurements obtained for AMS V1 or the reference device Osirix MD, as evaluated per t-test statistics. Tests performed for a series of objects in reference MR and CT phantom images.No statistical difference was reported, as implied by the statement "No statistical difference between distributions of measurements obtained for AMS V1 or the reference device Osirix MD, as evaluated per t-test statistics. Tests performed for a series of objects in reference MR and CT phantom images." The device met this criterion.
Display Quality (Luminance, Contrast)N/AN/A (evaluated against guidance)Successfully evaluated against the AAPM guidance recommendation for visual evaluation of luminance and contrast."The quality of the display was successfully evaluated against the AAPM guidance recommendation for visual evaluation of luminance and contrast." The device met this criterion.
Optical Testing (VR Platforms)N/AN/A (evaluated against standard)Homogeneity of luminance, resolution, and contrast evaluated as acceptable per IEC63145-20-20 in the center of the displays for the specified VR platforms."Additional optical testing was conducted on compatible VR platforms as per IEC63145-20-20 and passed as expected. The homogeneity of luminance, resolution, and contrast was evaluated as acceptable per these standards in the center of the displays for the specified VR platforms." The device met this criterion.
Filter Technology FunctionalityImage filters(Similar to cleared device Osirix MD (K101342))Opacity and color of specific voxels in the image demonstrated to be controllable as intended by the filtering principle."The functioning of the filter technology was assessed by visual inspection. Using a reference DICOM, the opacity and color of specific voxels in the image was demonstrated to be controllable as intended by the filtering principle, which is similar to the cleared device Osirix MD (K101342)." The device met this criterion.
VR Experience FluidityN/AN/AAveraged Frame Per Second (FPS) superior to a specific threshold for the minimal hardware configuration."The fluidity of the VR experience was assessed by evaluating the average Frame Per Second rate. The averaged FPS was superior to the specific threshold for the minimal hardware configuration." The device met this criterion.

The provided text for points 2-9 below is sparse. The information below is limited to what can be directly inferred or is explicitly stated in the document.

2. Sample Size Used for the Test Set and Data Provenance

The document states that tests for linear and diameter measurements were "performed for a series of objects in reference MR and CT phantom images." This indicates that phantom data was used for these specific tests. The exact sample size (number of phantom images or objects within them) for the test set is not specified.

The data provenance is described as "reference MR and CT phantom images," which implies controlled, synthetic data rather than patient data from a specific country or collected retrospectively/prospectively.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The document does not specify the number of experts used or their qualifications for establishing ground truth on the test set. For the measurement tests, the ground truth was based on "reference MR and CT phantom images," meaning the inherent dimensions of the phantom objects served as the ground truth. For visual evaluations (display quality, filter functionality), it can be inferred that qualified personnel performed these, but no details are provided.

4. Adjudication Method for the Test Set

The document does not specify an adjudication method. For the measurement tests, the ground truth was objective (phantom dimensions), and for visual assessments, it seems a direct assessment against standards or intended functionality was performed, without mention of a multiple-reader adjudication process.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No MRMC comparative effectiveness study was done or reported in this document. The study focused on standalone performance evaluation as described below.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

Yes, a standalone performance evaluation was done. The performance data section describes "Measurement performance testing was conducted by leveraging reference digital phantoms and the comparison with a cleared device (Osirix MD K101342)." This compares the algorithm's measurement capabilities against a reference, which is a standalone assessment. Similarly, the display quality, optical testing, filter technology, and VR fluidity assessments are focused on the device's inherent performance.

7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

The ground truth used for specific tests was:

  • phantom images (for linear and diameter measurements) where the actual dimensions of the objects in the phantoms served as the objective truth.
  • AAPM guidance recommendation and IEC63145-20-20 standard (for display quality and optical testing).
  • Reference DICOM with intended filtering principles (for filter technology functionality).

8. The Sample Size for the Training Set

The document does not provide any information regarding the sample size for a training set. This submission focuses on verification and validation testing, suggesting that the core algorithms might have been developed prior, and details about their training data are not included in this 510(k) summary.

9. How the Ground Truth for the Training Set Was Established

The document does not provide any information on how the ground truth for a training set was established, as it does not discuss a training set.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Avatar Medical % Rory Carrillo Quality and Regulatory Consultant Cosm 45 Bartlett St. SAN FRANCISCO CA 94110

May 24, 2023

Re: K222035

Trade/Device Name: Avatar Medical Software V1 Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: LLZ Dated: April 21, 2023 Received: April 24, 2023

Dear Rory Carrillo:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jessica Lamb

Jessica Lamb, Ph.D. Assistant Director Imaging Software Team DHT8B: Division of Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K222035

Device Name AVATAR MEDICAL Software V1

Indications for Use (Describe)

AVATAR MEDICAL Software V1 is intended as a medical imaging system that allows the processing, review, analysis, communication and media interchange of multi-dimensional digital images acquired from CT or MR imaging devices. It is also intended as software for preoperative surgical planning, and as software for the intraoperative display of the aforementioned multi-dimensional digital images. AVATAR MEDICAL Software V1 is designed for use by health care professionals and is intended to assist the clinician who is responsible for making all final patient management decisions.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Section 5. 510(k) Summary

1. General Information

K222035

510(k) SponsorAvatar Medical
Address11 rue de Lourmel75015 Paris France
Correspondence PersonRory A. CarrilloQuality and Regulatory ConsultantCosm
Contact InformationEmail: rory@cosmhq.comPhone: 562-533-7010
Date PreparedNovember 22, 2022

2. Subject Device

Proprietary NameAvatar Medical Software V1
Common NameAMS V1
Classification NameSystem, Image Processing, Radiological
Regulation Number21 CFR 892.2050
Regulation NameMedical Image Management and Processing System
Product CodeLLZ
Regulatory ClassII

Predicate Device 3.

Proprietary NameCeevra Reveal 2.0
Premarket NotificationK173274
Classification NameSystem, Image Processing, Radiological
Regulation Number21 CFR 892.2050
Regulation NameMedical Image Management and Processing System
Product CodeLLZ
Regulatory ClassII

4. Device Description

The Avatar Medical Software V1 (AMS V1) is a software-only device that allows trained medical professionals to review CT and MR image data in three-dimensional (3D) format and/or in virtual reality (VR) interface. The 3D and VR images are accessible through the software desktop application and, if desired, through compatible VR headsets which are used by users for preoperative surgical planning and for display during intervention/surgery.

{4}------------------------------------------------

The AMS V1 product is to be used to assist in medical image review. Intended users are trained medical professionals, including imaging technicians, clinicians and surgeons.

AMS V1 includes two main software-based user interface components, the Desktop Interface and the VR Interface. The Desktop Interface runs on a compatible off-the-shelf (OTS) workstation provided by the hospital and only accessed by authorized personnel. The Desktop Interface contains a graphical user interface where a user can retrieve DICOM-compatible medical images locally or on a Picture Archiving Communication System (PACS). Retrieved CT and MR images can be viewed in 2D and 3D formats. Users are able to make measurements, annotations, and apply fixed and manual image filters.

The VR Interface is accessible via a compatible OTS headset to allow users to review the medical images in a VR format. VR formats can be viewed only when the user connects a compatible VR headset directly to the workstation being used to view the Desktop Interface. Additionally. AMS V1 enables the intended users to remotely stream the Desktop Interface to another workstation on the same local area network (LAN).

The 3D images generated using AMS V1 are intended to be used in relation to surgical procedures in which CT or MR images are used for preoperative planning and/or during intervention/surgery.

The intraoperative use of the AMS V1 solely corresponds to the two following cases:

  • Display of the AMS V1 Desktop Interface on existing monitors/screens in the operating ● room
  • . Use in a non-sterile imaqe review room accessible from the operating room during the procedure (AMS V1 operates on VR headsets which are not approved to be used in the sterile environment of the operating room)

Usage During Interventional and Surgical Procedures

The AVATAR MEDICAL Software V1 is not meant to be used in a sterile environment.

The AVATAR MEDICAL Software V1 is not meant to register 3D images to patients.

Two usage scenarios of the AVATAR MEDICAL Software V1 software are possible during an interventional or surqical procedure:

1. Inside the operating room

The Desktop Interface (no VR) can be displayed on a computer monitor (previously installed respecting the management of sterility) inside the operating room. It is not expected that users will analyze the image, optimize filters, and realize measurements

{5}------------------------------------------------

in this intraoperative context. This usage is for visualizing results of a previous surgery planning session. The user will follow their scrub-in and scrub-out facility protocols when exiting and entering the sterile field of the operating room.

2. Outside the operating room

The full software experience, including VR, can be accessed in an adjacent non-sterile image review room. In this usage, the user will need to scrub-out to access the review room and, after review, scrub-in following their facility protocols as required. The user will not be in contact nor in the presence of the patient while wearing the VR headset

The software is not meant to be used with the user's hands on the patient during an interventional or surqical procedure.

5. Intended Use and Indications for Use

AVATAR MEDICAL Software V1 is intended as a medical imaging system that allows the processing, review, analysis, communication and media interchange of multi-dimensional digital images acquired from CT or MR imaging devices. It is also intended as software for preoperative surgical planning, and as software for the intraoperative display of the aforementioned multi-dimensional digital images. AVATAR MEDICAL Software V1 is designed for use by health care professionals and is intended to assist the clinician who is responsible for making all final patient management decisions.

6. Comparison of Technological Characteristics

Feature/FunctionSubject Device:AMS V1Predicate Device:Ceevra Reveal 2.0(K173274)
Intended UsersHealth care ProfessionalsHealth care Professionals
IntendedEnvironmentHealthcare facilities such ashospitals and clinicsHealthcare facilities such ashospitals and clinics
Device ClassClass IIClass II

{6}------------------------------------------------

Feature/FunctionSubject Device:AMS V1Predicate Device:Ceevra Reveal 2.0(K173274)
PreoperativeUseYesYes
IntraoperativeDisplayYesYes
Image AnalysisFeaturesInteractivemanipulation,filtering, create annotations andmeasurementsInteractivemanipulation andsegment
MeasurementsYes - linear and diametermeasurements onlyNo
RemoteStreamingYesNo
Pan imagePan image in any directionPan image in any direction
PIN CodeYes, for remote streaming onlyOptional 4-digit security code

7. Predicate Device Comparison

The subject device and predicate device have similar indications for use in that they are both intended for viewing CT and MR data in multi-dimensional views as well as for preoperative and intraoperative surgical planning using similar technologies. Additionally, both devices have similar technological characteristics with only minor differences that do not raise questions of safety or effectiveness.

8. Performance Data

Safety and performance of the AMS V1 has been evaluated and verified in accordance with software specifications and applicable performance standards through software verification and validation testing. Additionally, the software validation activities were performed in accordance with ANSI AAMI IEC 62304:2006/41:2016 - Medical device software - Software life cycle processes, in addition to the FDA Guidance documents, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" and "Content of Premarket Submission for Management of Cybersecurity in Medical Devices." All functional and

AMS V1 Traditional 510(k)

{7}------------------------------------------------

performance testing occurred on the minimum hardware configuration and the specified operating systems and VR platforms. Measurement performance testing was conducted by leveraging reference digital phantoms and the comparison with a cleared device (Osirix MD K101342). The table below summarizes the acceptance criteria for the measurement tools. The quality of the display was successfully evaluated against the AAPM guidance recommendation for visual evaluation of luminance and contrast . Additional optical testing was conducted on compatible VR platforms as per IEC63145-20-20 and passed as expected. The homogeneity of luminance, resolution, and contrast was evaluated as acceptable per these standards in the center of the displays for the specified VR platforms. The functioning of the filter technology was assessed by visual inspection. Using a reference DICOM, the opacity and color of specific voxels in the image was demonstrated to be controllable as intended by the filtering principle, which is similar to the cleared device Osirix MD (K101342) . The fluidity of the VR experience was assessed by evaluating the average Frame Per Second rate. The averaged FPS was superior to the specific threshold for the minimal hardware configuration.

Feature/FunctionTool in SubjectDevice:AMS V1Tool in ReferenceDevice:Osirix MD(K101342)Acceptance criteria
LinearMeasurements(polylines)CurveClose PolygonNo statistical differencebetween distributions ofmeasurements obtained forAMS V1 or the referencedevice Osirix MD, as evaluatedper t-test statistics.Tests performed for a series ofobjects in reference MR andCT phantom images.
DiameterMeasurementsRulerLength

9. Conclusion

Based on the information submitted in this premarket notification, and based on the indications for use, technological characteristics and performance testing, the AMS V1 raises no new questions of safety and effectiveness and demonstrates substantial equivalence to the predicate device in terms of safety, effectiveness, and performance.