K Number
K221996

Validate with FDA (Live)

Manufacturer
Date Cleared
2023-03-20

(257 days)

Product Code
Regulation Number
892.5050
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The P-Cure Proton Therapy System is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other condition susceptible to treatment by radiation in the head, neck and thorax.

Device Description

The P-CURE system is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other condition susceptible to treatment by radiation in the head, neck and thorax.

The P-CURE proton beam therapy system comprises four main subsystems that function in tandem to generate the desired dose level and distribution at the target site:
Beam production system (Synchrotron based accelerator)

  • Injector produces and delivers protons to the synchrotron
  • Synchrotron ring accelerates the proton beam in circular orbit (within the ring) to the desired energy level
  • Extraction system extracts the beam from the ring to the beam delivery subsystem
    Beam delivery system for a single fixed beam treatment room. Steers and monitors the extracted proton pencil beam from the synchrotron to the desired treatment location (Nozzle).
    Patient Positioning System (P-ARTIS). Mechanically orients the seated patient; provides independent means of patient registration using CT (3D) and X-ray (2D)
  • CT system (P-ARTIS CT)
  • Robotic arm and chair (6 Degree of freedom Couch) (P-ARTIS PPS)
  • X-ray system (P-ARTIS XR)
  • Positioning Software (P-ARTIS SW)
    Control and Safety Systems
  • Control Subsystem (TSM). Synchronizes the various subsystem actions and connects with hospital oncology information systems and PACS.
  • Safety Subsystem. Includes hardware and software means to ensure safe system operation for patient and personnel. It includes subsystem interlocks, treatment beam parameters monitoring, and others.
AI/ML Overview

The provided text describes a 510(k) premarket notification for the P-Cure Proton Beam Therapy System, demonstrating its substantial equivalence to a predicate device (ProTom Radiance 330). However, it does not contain the specific information required to answer your request regarding acceptance criteria and a detailed study proving the device meets those criteria for an AI/CADe device.

The document primarily focuses on:

  • Legal/Regulatory Status: FDA clearance (K221996), device classification, and general controls.
  • Device Description: Components of the P-Cure system (synchrotron, beam delivery, patient positioning, control/safety).
  • Comparison to Predicate: Highlighting similarities in core technology (proton beam, synchrotron, energy range, pencil beam scanning) and differences (fixed beam vs. gantry, seated vs. supine patient position).
  • Performance Data (General): A high-level list of tests performed (mechanical, beam performance, safety interface, simulation/validation, repeatability/reproducibility) and a general statement that the device "functioned as intended and met its specifications."

Therefore, I cannot provide the requested information for the following reasons:

  • No Acceptance Criteria Table or Reported Performance: The document states "In all instances, the P-Cure System functioned as intended and met its specifications" but does not provide a table with specific acceptance criteria (e.g., precision, accuracy, sensitivity, specificity values) or the numerical results of performance for each criterion.
  • No Information on AI/CADe Study: The P-Cure device is a Proton Beam Therapy System, not an AI/CADe (Computer-Assisted Detection/Diagnosis) device for image analysis. The performance data mentioned (mechanical, beam, safety, positioning) are related to the physical operation and output of the radiation therapy system itself, not to the performance of an AI algorithm in detecting or diagnosing conditions from medical images.
  • No Test Set Details: Since it's not an AI/CADe study, there's no mention of sample size for a test set, data provenance, ground truth establishment methods (expert consensus, pathology), adjudication, or MRMC studies.
  • No Training Set Details: Similarly, for a physical device, there isn't a training set in the context of an AI model.

If the request was based on a misunderstanding of the document's content and aimed at describing the performance validation of a complex medical device (like the proton therapy system) in a general sense, the document provides the following limited details:

  • General Performance Testing Categories: Mechanical, beam performance, safety interface, simulation/validation, and repeatability/reproducibility testing.
  • Applicable Standards: IEC 60601-1, IEC 600601-1-2, EN 606601-2-44, IEC 60601-1-3, IEC 60601-1-8, IEC 60601-2-54, IEC 60601-1-64, IEC 60601-2-68, IEC 62667, and AAPM TG-224. These standards are typically related to general medical electrical equipment safety, electromagnetic compatibility, radiation therapy equipment, and specific performance measurements for such devices.
  • Overall Conclusion: "In all instances, the P-Cure System functioned as intended and met its specifications. Testing demonstrated substantial equivalence in terms of performance and safety to the predicate."

In summary, the provided document is a 510(k) clearance letter for a proton beam therapy system, not a study report for an AI/CADe device. Therefore, it does not contain the specific information required for your detailed questions about AI model acceptance criteria and validation.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA) along with the Department of Health & Human Services. The FDA logo features a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG" in blue, and "ADMINISTRATION" in a smaller font size below. To the left of the FDA logo is the symbol of the Department of Health & Human Services, which consists of a stylized human figure.

March 20, 2023

P-Cure Ltd. % Jonathan Kahan Partner Hogan Lovells US LLP 555 Thirteenth Street. N.W. WASHINGTON DC 20004

Re: K221996

Trade/Device Name: P-Cure Proton Beam Therapy System Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: Class II Product Code: LHN Dated: February 16, 2023 Received: February 17, 2023

Dear Jonathan Kahan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Digitally signed by Lora D. Weidner -S Lora D. Weidner -S
Date: 2023.03.20 09:44:39 -04'00'

Lora D. Weidner, Ph.D. Assistant Director Radiation Therapy Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K221996

Device Name

P-Cure Proton Beam Therapy System

Indications for Use (Describe)

The P-Cure Proton Therapy System is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other condition susceptible to treatment by radiation in the head, neck and thorax.

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY P-Cure, Ltd.'s Proton Beam Therapy System

Submitter

P-Cure, Ltd. 14 Carmel St. Shilat Industrial Zone, 7318800 lsrael

Phone: +972 54 488 1399

Contact Person: Ori Lubin Date Prepared: March 15, 2023

Name of Device:

P-Cure Proton Beam Therapy System

Common or Usual Name:

Proton beam therapy systems

Classification Name:

Medical Charged-Particle Radiation Therapy System, 21 CFR 892.5050

Regulatory Class:

Class II

Product Code:

LHN

Predicate Device

ProTom International Holding Corp. Radiance 330 Proton Beam Therapy System (K191521)

Device Description:

The P-CURE system is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other condition susceptible to treatment by radiation in the head, neck and thorax.

The P-CURE proton beam therapy system comprises four main subsystems that function in tandem to generate the desired dose level and distribution at the target site:

Beam production system (Synchrotron based accelerator) .

  • o Injector produces and delivers protons to the synchrotron
  • Synchrotron ring accelerates the proton beam in circular orbit (within the o ring) to the desired energy level

{4}------------------------------------------------

  • Extraction system extracts the beam from the ring to the beam delivery o subsvstem
  • Beam delivery system for a single fixed beam treatment room. Steers and . monitors the extracted proton pencil beam from the synchrotron to the desired treatment location (Nozzle).
  • . Patient Positioning System (P-ARTIS). Mechanically orients the seated patient; provides independent means of patient registration using CT (3D) and X-ray (2D)
    • CT system (P-ARTIS CT) o
    • Robotic arm and chair (6 Degree of freedom Couch) (P-ARTIS PPS) o
    • o X-ray system (P-ARTIS XR)
    • Positioning Software (P-ARTIS SW) O

· Control and Safety Systems

  • Control Subsystem (TSM). Synchronizes the various subsystem actions and o connects with hospital oncology information systems and PACS.
  • Safety Subsystem. Includes hardware and software means to ensure safe o system operation for patient and personnel. It includes subsystem interlocks, treatment beam parameters monitoring, and others.

Intended Use / Indications for Use

The P-CURE system is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other condition susceptible to treatment by radiation in the head, neck and thorax.

Summary of Technological Characteristics and Comparison to the Predicate

The P-Cure proton beam therapy system and its predicate Protom Radiance 330 (K191521) are based on the same technological elements where the system is designed to achieve radiation therapy by accelerating protons for delivery through a nozzle to the patient for the treatment of conditions which are susceptible to radiation. The P-cure system uses a similar synchrotron as utilized by the predicate. Both systems use CT to obtain 3D images for planning and positioning. The P-Cure system delivers a treatment by a fixed beam assembly, and uses a robotic patient positioning system to move the patient to ensure delivery of the beam to the target area whereas the predicate system incorporates a gantry to move the beam in conjunction with a robotic patient positioning system. The P-Cure system positions the patient in a seated position compared to the supine position used with the predicate system. These differences do not raise different questions of safety or efficacy and testing demonstrates substantially equivalent performance compared to the predicate. A table comparing the key features of the subject and predicate devices is provided below.

DeviceP-CURE SystemRadiance 330 Proton BeamTherapy System
510(k) NumberK221996K191521
DeviceP-CURE SystemRadiance 330 Proton BeamTherapy System
Intended Use/Indications forUseThe P-CURE system is a medicaldevice designed to produce anddeliver a proton beam for thetreatment of patients with localizedtumors and other conditionsusceptible to treatment byradiation in the head, neck andthorax.The ProTom Radiance 330 is amedical device designed toproduce and deliver a protonbeam for the treatment ofpatients with localized tumorsand other condition susceptibleto treatment by radiation
AcceleratorSynchrotronSynchrotron
ParticleProtonsProtons
VariableEnergyYes; 70-250 MeVYes; 70-250 MeV
Cycle TimeVariableVariable
BeamTransportBending and focusing magnetsBending and focusing magnets
NozzlesPencil Beam ScanningPencil Beam Scanning
Beam Rangein Patient(TissueDepth)3 cm to 38 cm clinical use3 cm to 38 cm clinical use
Range VerifierOptionalOptional
Control andSafety SystemSoftware controls accelerator,beam transport and delivery,sets operational parameters,monitors systems andprovides alerts regarding errorSoftware controls accelerator,beam transport and delivery,sets operational parameters,monitors systems and providesalerts regarding error conditions.
MechanicalBeamYesYes
BeamContinuously variable over rangeContinuously variable over
Intensitiesup to 1E10 protons per cycle time.range up to 1E10 protons per cycle time.
ShieldingSteel and ConcreteSteel and Concrete
GantryrotatingN/A. Fixed beamRotating gantry
PatientPositionerYes - floor mounted chairYes - floor mounted couch
ImagingYes - Diagnostic quality CT andx-ray systemYes - couch mounted CBCTand x-ray system
TreatmentRoomFixed beam orientationand rotating chairFixed couch position androtating gantry

{5}------------------------------------------------

Performance Data

The Company performed testing as follows:

  • Mechanical testing to verify performance of the positioning system .
  • Beam performance testing to evaluate beam dose shape, beam dose, dose rate, . dose monitoring, and spot positioning
  • · Safety interface testing to verify collision sensors, mechanical interlocks

{6}------------------------------------------------

  • Simulation and validation testing to verify integration with oncology information . systems
  • · Validation testing to verify integration with positioning system and treatment planning svstem
  • · Testing to support repeatability and reproducibility of patient positioning and immobilization

Product safety and essential performance electrical testing was conducted based upon the following consensus standards: IEC 60601-1, IEC 600601-1-2, EN 606601-2-44, IEC 60601-1-3, IEC 60601-1-8, IEC 60601-2-54, IEC 60601-1-64, IEC 60601-2-68, IEC 62667, and AAPM TG-224.

In all instances, the P-Cure System functioned as intended and met its specifications. Testing demonstrated substantial equivalence in terms of performance and safety to the predicate.

Substantial Equivalence

The P-Cure System is as safe and effective as the ProTom International Holding Corp. Radiance 330 Proton Beam Therapy System (K191521). The P-Cure System has the same intended use and substantially similar indications for use technological characteristics, and principles of operation as its predicate device. The minor differences in the specific indications for use statement and technological differences between the P-Cure System and its predicate devices do not raise different questions of safety or effectiveness. Performance data demonstrate that the P-Cure System is as safe and effective as the Radiance 330 Proton Beam Therapy System. Thus, the P-Cure System is substantially equivalent.

§ 892.5050 Medical charged-particle radiation therapy system.

(a)
Identification. A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.(b)
Classification. Class II. When intended for use as a quality control system, the film dosimetry system (film scanning system) included as an accessory to the device described in paragraph (a) of this section, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.