K Number
K221979

Validate with FDA (Live)

Date Cleared
2022-10-16

(103 days)

Product Code
Regulation Number
870.2700
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Fingertip Pulse Oximeter (Model:C101A2,C101B1,C101A3) is a non-invasive device intended for spot checking of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR). This portable device is indicated for use in adult patients in hospitals.

Device Description

SpO2 is based on the absorption of pulse blood oxygen to red and infrared light by means of finger sensor and SpO2 measuring unit. The light-electronic transducer in finger sensor converts the pulse red and infrared light modulated by pulse blood oxygen into electrical signal, the signal is processed by hardware and software of the unit. The PLETH curve and numeral value of SpO2 will be obtained. The pulse oximeter, model C101A2, C101B1, C101A3 is designed for spot checking of the pulse oxygen saturation and pulse rate for adults in a clinic environment. This medical device can be reused. Not for continuously monitoring.

The device is not for life-supporting or life-sustaining, not for implant.

The device is not provided sterile and is NOT a reprocessed single-use device.

The device is a spot-check pulse oximeter and does not include alarms.

The device does not support the measurement in the condition of low perfusion.

The device is not intended for life-supporting or life-sustaining.

The device is reusable and does not need sterilization.

AI/ML Overview

Acceptance Criteria and Device Performance Study for Pulse Oximeter (Model C101A2, C101B1, C101A3)

The provided document, a 510(k) Summary for the Pulse Oximeter (Model C101A2, C101B1, C101A3), indicates that the device is substantially equivalent to a previously cleared predicate device (Pulse Oximeter, Model C101H1). This means that the acceptance criteria and performance data for the subject device are considered equivalent to those established for the predicate device.

Crucially, the document states: "As there were no hardware, software or performance changes made to the subject device when compared to the primary predicate device, no additional non-clinical testing was considered necessary to support the substantial equivalence." and "As the subject device utilizes the same monitoring technology as the predicate device, additional testing was not considered necessary to support the substantial equivalence."

Therefore, the information below refers to the performance that the predicate device (K173123) demonstrated, which the subject device (K221979) is claimed to meet by virtue of substantial equivalence. The summary does not include new testing data specifically for the subject device's performance against these criteria, except for biocompatibility due to a color change in patient-contacting materials.

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are implied by the specifications of the predicate device, which the subject device is deemed to meet.

ItemAcceptance Criteria (from Predicate Device K173123)Reported Device Performance (for Subject Device K221979)
SpO2 Measuring Range0%-100%0%-100%
SpO2 Resolution1%1%
SpO2 Accuracy70-100%, ±3%70-100%, ±3%
PR Measuring Range30-240 BPM30-240 BPM
PR Resolution1 bpm1 bpm
PR Accuracy±1 bpm or ±1%, whichever is greater±1 bpm or ±1%, whichever is greater
Electrical SafetyComply with IEC 60601-1, IEC 60601-1-11Comply with IEC 60601-1, IEC 60601-1-11
EMCComply with IEC 60601-1-2Comply with IEC 60601-1-2
Performance StandardISO 80601-2-61ISO 80601-2-61
BiocompatibilityComply with ISO 10993-1, FDA Guidance (for previous device)Cytotoxicity per ISO 10993-5, Irritation per ISO 10993-10, Sensitization per ISO 10993-10 (for new device due to color change)

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document explicitly states that "As there were no hardware, software or performance changes made to the subject device when compared to the primary predicate device, no additional non-clinical testing was considered necessary to support the substantial equivalence." and "As the subject device utilizes the same monitoring technology as the predicate device, additional testing was not considered necessary to support the substantial equivalence."

Therefore, no new performance test data for SpO2 or PR accuracy is included for the subject device (K221979) in this submission. The assumption is that the previously cleared predicate device (K173123) met these criteria, and since the technology is the same, no new testing was deemed necessary for these aspects. Details regarding the sample size and data provenance for the predicate device's original performance testing are not provided in this summary.

For Biocompatibility testing (which was performed for the subject device due to a color change in patient-contacting materials):

  • Sample Size: Not specified for each test (Cytotoxicity, Irritation, Sensitization), but standard procedures would involve multiple samples of the actual patient-contacting materials.
  • Data Provenance: Not specified; assumed to be conducted in a laboratory setting as per international standards (ISO 10993).

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not provided in the 510(k) summary, as no new clinical or performance studies for SpO2/PR accuracy were conducted for the subject device. For the original predicate device, pulse oximeter accuracy studies typically involve comparing the device's readings to arterial blood gas analysis (co-oximetry) measurements, which serve as the ground truth. This process usually involves clinical personnel (e.g., nurses, physicians, respiratory therapists) performing the measurements and often statistical experts involved in data analysis, but not necessarily "experts establishing ground truth" in the sense of image interpretation.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not applicable/not provided as no new clinical or performance studies involving adjudication were conducted for the subject device.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This is not applicable as the device is a Pulse Oximeter, not an AI-assisted diagnostic tool involving human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

The device is a standalone pulse oximeter that provides direct measurements of SpO2 and PR. Its performance evaluation would inherently be "standalone" in the sense that it directly measures physiological parameters without human interpretation in the loop (beyond reading the display). The performance criteria listed (SpO2/PR accuracy) are indicative of standalone performance against a clinical reference standard. The document states "As there were no hardware, software or performance changes made to the subject device when compared to the primary predicate device, no additional non-clinical testing was considered necessary to support the substantial equivalence." This implies that the standalone performance requirements previously established for the predicate device were considered met.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

For pulse oximetry, the ground truth for SpO2 accuracy is typically arterial blood gas analysis (co-oximetry), which provides a highly accurate measure of arterial oxygen saturation. For Pulse Rate (PR), the ground truth is often established by electrocardiography (ECG) or a manually measured pulse using a stopwatch. This information is not explicitly stated in this 510(k) summary for the predicate device's original testing, but these are the standard methods for validating pulse oximeter performance.

8. The sample size for the training set

This is not applicable/not provided. The device is a traditional medical device (pulse oximeter) that relies on physical principles (spectrophotometry and photoplethysmography) rather than machine learning or AI models that require training data.

9. How the ground truth for the training set was established

This is not applicable/not provided for the same reason as above; there is no "training set" in the context of this traditional device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 16, 2022

Shenzhen IMDK Medical Technology CO., Ltd. % Boyle Wang Shanghai Truthful Information Technology Co., Ltd. RM. 1801, No. 161, East Lujiazui Rd., Pudong Shanghai, Shanghai 200120 China

Re: K221979

Trade/Device Name: Pulse Oximeter (Model C101A2, C101B1, C101A3) Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DQA Dated: September 16, 2022 Received: September 16, 2022

Dear Boyle Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

James J. Lee, Ph.D. Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K221979

Device Name Pulse Oximeter(Model:C101A2,C101B1,C101A3)

Indications for Use (Describe)

Fingertip Pulse Oximeter (Model:C101A2,C101B1,C101A3) is a non-invasive device intended for spot checking of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR). This portable device is indicated for use in adult patients in hospitals.

Type of Use (Select one or both, as applicable)

Research Use (21 CFR 201.3 Subpart D)
Over-The-Counter Use (21 CFR 201.3 Subpart G)

Prescription Use (Part 21 CFR 801 Subpart D)

__ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

K221979

This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR 807.92.

1.0 Submitter's Information

Name:Shenzhen IMDK Medical Technology Co., Ltd.
Address:C Zone, 10F, Building 16, Yuanshan Industrial B Area,Gongming Street, Guangming District, Shenzhen,Guangdong, P.R. China
Tel:86-755-27155684
Contact:Yuan Xia
Registration Number:3015007456

Designated Submission Correspondent

Contact:Mr. Boyle Wang
Name:Shanghai Truthful Information Technology Co., Ltd.
Address:Room 1801, No. 161 East Lujiazui Rd., Pudong Shanghai200120 China
Tel:+86-21-50313932
Email:Info@truthful.com.cn

Date of Preparation: Sep.27,2022

2.0 Device Information

Trade name:Pulse Oximeter
Common name:Pulse Oximeter
Classification name:Oximeter
Model(s):C101A2, C101B1, C101A3
Production code:DQA
Regulation number:21 CFR 870.2700
Classification:Class II
Panel:Cardiovascular

3.0 Predicate Device Information

Manufacturer:Shenzhen IMDK Medical Technology Co., Ltd
Trade name:Pulse Oximeter, Model: C101H1

{4}------------------------------------------------

510(k) number: K173123

4.0 Indication for Use Statement

Fingertip Pulse Oximeter (Model: C101A2, C101B1, C101A3) is a non-invasive device intended for spot checking of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR). This portable device is indicated for use in adult patients in hospitals.

5.0 Device Description

SpO2 is based on the absorption of pulse blood oxygen to red and infrared light by means of finger sensor and SpO2 measuring unit. The light-electronic transducer in finger sensor converts the pulse red and infrared light modulated by pulse blood oxygen into electrical signal, the signal is processed by hardware and software of the unit. The PLETH curve and numeral value of SpO2 will be obtained. The pulse oximeter, model C101A2, C101B1, C101A3 is designed for spot checking of the pulse oxygen saturation and pulse rate for adults in a clinic environment. This medical device can be reused. Not for continuously monitoring.

The device is not for life-supporting or life-sustaining, not for implant.

The device is not provided sterile and is NOT a reprocessed single-use device.

The device is a spot-check pulse oximeter and does not include alarms.

The device does not support the measurement in the condition of low perfusion.

The device is not intended for life-supporting or life-sustaining.

The device is reusable and does not need sterilization.

6.0 Technological Characteristics

Principle of Operation

There is no change in the principle of operation as part of this submission from the previous clearance under K173123. The module still utilizes the same principles of operation for pulse oximetry governed by the following principles:

  1. Oxyhemoglobin (oxygenated blood) and deoxyhemoglobin (non-oxygenated blood) differ in their absorption of red and infrared light (spectrophotometry).

  2. The amount of arterial blood in tissue changes with your pulse (photoplethysmography). Therefore, the amount of light absorbed by the varying quantities of arterial blood changes as well.

Mechanism of Action for Achieving the Intended Effect

There is no change to the Mechanism of Action of the Pulse Oximeter (Model: C101A2, C101B1, C101A3) from the previous clearance K173123- Pulse Oximeter (Model: C101H1):

A mathematic formula is established making use of Lambert Beer Law according to Spectrum Absorption Characteristics of Reductive hemoglobin (RHb) and

{5}------------------------------------------------

Oxyhemoglobin (Hbo2) in red and near-infrared zones. Operation principle of the instrument: Photoeletric Oxyhemoglobin Inspection Technology adopted in accordance with Capacity Pulse Scanning and Recording Technology, so that two beams of different wavelength of lights (660nm red and 905nm near infrared light) can be focused on a human nail tip through a clamping finger-type sensor. A measured signal obtained by a photosensitive element, will be shown on the oximeters display through process in electronic circuits and microprocessor.

7.0 Summary of Technological Characteristics of the Subject Device Compared to the Predicate Device

The subject device, Pulse Oximeter (Model: C101A2, C101B1, C101A3) and the predicate device Pulse Oximeter (Model: C101H1), have the following key similarities:

  • · Both devices have the same intended use
  • · Both devices are indicated for the same patient population
  • Both devices have the same principle of operation and mechanism of action

The subject device and the primary predicate device have the following differences: The appearance of the Subject device is different with that of the Predicate device. But this difference does not affect the basic safety and essential performance, the subject and the predicate are the same and substantially equivalent.

8.0 Non-clinical Testing

In this current submission, just add four model C101A2, C101B1,C101A3 to the Legally marketed predicate device K173123.

As there were no hardware, software or performance changes made to the subject device when compared to the primary predicate device, no additional non-clinical testing was considered necessary to support the substantial equivalence.

Performance Bench Testing

As there are no hardware or software changes as part of this submission from the previous clearance, no performance bench testing was included as part of this submission.

Biocompatibility Testing

As the color of enclosure changes to the patient contacting materials as part of this submission from the previous clearance, biocompatibility testing was included as part of this submission.

The biocompatibility evaluation for the subject device was conducted in accordance with the FDA for Industry and Food and Drug Administration Staff "Use of International Standard ISO 10993-1, 'Biological Evaluation of Medical Devices -Part 1: Evaluation and Testing within a risk management process". The biocompatibility

{6}------------------------------------------------

testing included the following tests:

  • Cytotoxicity;
  • Irritation;
  • · Sensitization.

Electromagnetic Compatibility, Electrical Safety, Environmental, Mechanical and Cleaning

As there were no changes to the hardware or software as part of this submission from the previous clearance, no Electromagnetic Compatibility, Electrical Safety, Environmental, Mechanical and Cleaning testing was included with this submission.

Software Verification and Validation Testing

As there are no software changes as part of this submission from the previous clearance, no software testing was included as part of this submission.

9.0 Clinical Test Conclusion

As the subject device utilizes the same monitoring technology as the predicate device, additional testing was not considered necessary to support the substantial equivalence.

10.0 Technological Characteristic Comparison Table

{7}------------------------------------------------

Table1-General Comparison

ItemSubject DeviceK221979Predicated DeviceK173123Remark
Product NamePulse OximeterPulse Oximeter--
ManufacturerShenzhen IMDK Medical TechnologyCo., Ltd.Shenzhen IMDK Medical TechnologyCo., Ltd.--
Product CodeDQADQASame
Regulation No.21CFR 870.270021CFR 870.2700Same
ClassClass IIClass IISame
ModelC101A2, C101B1, C101A3C101H1--
IntendedUse/Indicationfor UseFingertip Pulse Oximeter (Model:C101A2, C101B1, C101A3) is anon-invasive device intended forspot checking of functional oxygensaturation of arterial hemoglobin(SpO2) and pulse rate (PR). Thisportable device is indicated for usein adult patients in hospitals.Fingertip Pulse Oximeter C101H1 is anon-invasive device intended for spotchecking of functional oxygensaturation of arterial hemoglobin(SpO2) and pulse rate (PR). Thisportable device is indicated for use inadult patients in hospitals.Same
PrincipleThe device displays numericalvalues for functional oxygensaturation of arterial hemoglobin(SpO2) and pulse rate bymeasuring the absorption of red andinfrared (IR) light passing throughperfused tissue. Changes in theabsorption caused by the pulsationof blood in the vascular bed areused to determine oxygen saturationand pulse rate.The device displays numerical valuesfor functional oxygen saturation ofarterial hemoglobin (SpO2) and pulserate by measuring the absorption of redand infrared (IR) light passing throughperfuse tissue. Changes in theabsorption caused by the pulsationof blood in the vascular bed are used todetermine oxygen saturation and pulserate.Same
AppliedPopulationAdults in a clinic environmentAdults in a clinic environmentSame
Application sitesFingerFingerSame
Display TypeOLEDOLEDSame
Display ContentDisplay the SPO2, PR, pulse bargraph, pulse wave and batterypower statusDisplay the SPO2, PR, pulse bargraph, pulse wave and battery powerstatusSame
ContactingmaterialShell: ABSClip pad: SiliconButton: PCShell: ABSClip pad: SiliconButton: PCSame
OverallDimensionC101A2: 603633mmC101B1: 663633mmC101A3: 583633mm603635mm--

Table 2 Performance Comparison

ItemSubject DeviceK221979Predicate DeviceK173123Remark
LED wavelengthRed= 660 nm; Infrared=904nmRed= 660 nm; Infrared=904nmSame
Power source2 AAA alkaline batteries2 AAA alkaline batteriesSame
Display dataSpO2%; PRSpO2%; PRSame
SpO2 MeasuringRange0%-100%0%-100%Same
SpO2 Resolution1%1%Same
SpO2 Accuracy70~100%, ±3%;70~100%, ±3%;Same

{8}------------------------------------------------

510(k) Summary

PR MeasuringRange30-240BPM30-240BPMSame
PR Resolution1 bpm1 bpmSame
PR Accuracy±1 bpm or ±1%, whichever isgreater±1 bpm or ±1%, whichever isgreaterSame

Table 3 Safety Comparison

ItemSubject DeviceK221979Predicate DeviceK173123Remark
Electrical SafetyComply with IEC 60601-1IEC 60601-1-11Comply with IEC 60601-1IEC 60601-1-11Same
EMCComply with IEC 60601-1-2Comply with IEC 60601-1-2Same
PerformanceISO 80601-2-61ISO 80601-2-61Same
BiocompatibilityCytotoxicity per ISO 10993-5;Irritation per ISO 10993-10;Sensitization per ISO 10993-10Comply with ISO 10993-1, FDA GuidanceSame

11.0 Conclusion

The conclusions drawn from the comparison and analysis above demonstrate that the subject device is substantially equivalent to the legally marketed predicated device.

§ 870.2700 Oximeter.

(a)
Identification. An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.(b)
Classification. Class II (performance standards).