K Number
K221871

Validate with FDA (Live)

Date Cleared
2022-08-30

(63 days)

Product Code
Regulation Number
862.3100
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Healstone Accurate ™ Drug of Abuse Urine Test Cup is a rapid qualitative immunoassay. The device provides preliminary results for the detection of potential abuse of one or more drugs. This is not a screening device to monitor prescription medication. This drug test cup may contain any combination of the drug tests listed in the table above. This drug tests cup provides only a preliminary result. An alternative laboratory test must be used to confirm the results provided by this drug test. Gas chromatography/mass spectrometry (CCMS) is the preferred method confirmation test. For in vitro diagnostic use only.

The Healstone Accurate ™ Drug of Abuse Urine Test Cup Rx is a rapid qualitative immunoassay. The device provides preliminary results for the detection of potential abuse of one or more drugs. This is not a screening device to monitor prescription medication. This drug test cup may contain any combination of the drug tests listed in the table above. This drug tests cup provides only a preliminary result. An alternative laboratory test must be used to confirm the results provided by this drug test. Gas chromatography/mass spectrometry (GC/MS) is the preferred method confirmation test. It is intended for prescription use. For in vitro diagnostic use only.

Device Description

The Healstone Accurate™ Drug of Abuse Urine Test Cup and Healstone Accurate™ Drug of Abuse Urine Test Cup Rx are rapid, single-use in vitro diagnostic devices. Each test kit contains a test device in one pouch. One pouch contains a test Accurate™ Cup and two desiccants, and a package insert. The Healstone Accurate™ Drug of Abuse Urine Test Cup is intended for over-the-counter use and the Healstone Accurate™ Drug of Abuse Urine Test Cup Rx is intended for prescription use.

AI/ML Overview

The provided text describes the performance characteristics of the "Healstone Accurate™ Drug of Abuse Urine Test Cup" and "Healstone Accurate™ Drug of Abuse Urine Test Cup Rx", but does not specify explicit acceptance criteria in the form of numerical thresholds for sensitivity, specificity, accuracy, or other performance metrics. The tables presented show the reported device performance across various drug tests at different concentrations relative to the cutoff level.

Here's an attempt to extract the information requested, based on what is available in the document:

1. Table of Acceptance Criteria and Reported Device Performance

As noted above, no explicit acceptance criteria are stated in the document. However, the reported performance in both the precision studies and lay-user studies consistently shows a high level of agreement with expected outcomes at concentrations well below and well above the cutoff. At the cutoff concentration, the device exhibits approximately 50% positive and 50% negative results, which is a typical characteristic for qualitative immunoassays designed to detect substances at a specific threshold.

For the precision studies, for concentrations at -100%, -75%, -50% cut off and +50%, +75%, +100% cut off, the results were almost always 100% negative and 100% positive, respectively. For concentrations at -25% cut off, a small number of false positives (1-3 out of 50) were observed across different lots and drugs. For concentrations at +25% cut off, a small number of false negatives (1-3 out of 50) were observed. This indicates that the device has a narrow window around the cutoff where results can be variable, which is expected for such tests.

For the lay-user studies, the agreement percentages for concentrations clearly below (-100%, -75%, -50% cutoff) and clearly above (+50%, +75% cutoff) the cutoff are all 100%. At -25% cutoff, agreement is generally high (85-95% negative). At +25% cutoff, agreement is also high (85-95% positive).

Without explicit acceptance criteria, it's impossible to state whether these results "meet" them. However, they demonstrate consistent performance around the designed cutoff levels.

2. Sample Size and Data Provenance

  • Precision Test Set Sample Size: For each drug and each concentration level, 50 tests were performed (2 runs per day for 25 days). This was done for three different lots of test cups. So, for each drug and each concentration, the total number of tests across three lots is 50 * 3 = 150.

  • Comparison Studies (Professional Operator) Test Set Sample Size: For each drug, 80 unaltered urine samples were used (40 negative and 40 positive). This was tested by three operators.

  • Lay-user Study Test Set Sample Size:

    • Configuration 1 (AMP 1000, MET 1000, MOP 300, COC 300): 140 participants (81 males, 59 females). Each participant tested one blind-labeled test solution. Given that there were 7 concentration levels tested per drug in the table, and 14 drugs are listed for configuration 1, this would imply (20 samples per concentration level, 7 concentration levels, for 14 drugs means) a total of 14 * 7 * 20 = 1960 samples for Configuration 1.
    • Configuration 2 (AMP 500, MET 500, OPI 2000, COC 150): 140 participants (90 males, 50 females). Similar breakdown of sample numbers as Configuration 1.
    • Total participants across both configurations = 280. Total samples tested would be approximately 1960 samples per configuration, totaling around 3920 samples.
  • Data Provenance: The document does not specify the country of origin for the data. The studies appear to be prospective as samples were prepared at specific concentrations and then blinded for testing.

3. Number of Experts and Qualifications for Ground Truth

  • Comparison Studies (Professional Operator): The ground truth for the comparison studies was established using LC/MS or GC/MS results. These are laboratory-based, highly accurate analytical methods, widely considered the "gold standard" for drug concentration determination. The document does not specify the number of experts (e.g., lab technicians, chemists) that performed or interpreted the LC/MS/GC/MS results, nor their specific qualifications.
  • Lay-user Study: The ground truth for the lay-user study was also based on the known concentrations established by LC/MS.

4. Adjudication Method for the Test Set

  • Precision Studies: Not applicable. The results are reported directly (number of negative/positive results for the given concentration).
  • Comparison Studies (Professional Operator): The results from the test device (Healstone Accurate™) were compared directly against the LC/MS or GC/MS results. Discordant results are noted for each drug and operator. This is a direct comparison rather than an adjudication among human readers.
  • Lay-user Study: The lay users read the device results, and these were compared to the known concentrations (ground truth from LC/MS). The document does not describe any adjudication process for lay-user interpretations if they differed from a pre-determined "correct" visual interpretation.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No MRMC comparative effectiveness study was done regarding human readers improving with AI vs without AI assistance. This device is a rapid qualitative immunoassay (a test strip/cup that shows lines), not an AI-powered diagnostic tool requiring human interpretation of AI-generated insights. The "lay-user study" involved human readers (lay individuals) interpreting the device directly, without AI assistance.

6. Standalone (Algorithm Only) Performance Study

  • This device is a physical immunoassay test cup, not an algorithm. Therefore, a standalone algorithm-only performance study is not applicable. The performance described is the standalone performance of the physical device.

7. Type of Ground Truth Used

  • Precision, Comparison, and Lay-user Studies: The ground truth was established by known concentrations of spiked drugs in urine samples, confirmed by LC/MS or GC/MS (laboratory-based analytical methods). LC/MS and GC/MS are considered definitive methods for identifying and quantifying substances.

8. Sample Size for the Training Set

  • This information is not provided in the document. The document describes performance evaluation studies but does not detail the development or training of the immunoassay. Immunoassays are not "trained" in the same way machine learning algorithms are; their performance is inherent to their chemical and biological design.

9. How the Ground Truth for the Training Set Was Established

  • As mentioned above, information on a "training set" and its ground truth is not provided in this document as it pertains to the development of an immunoassay, not an AI/ML model.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 30, 2022

Healstone Biotech Inc % Joe Shia Director LSI International Inc. 504 East Diamond Ave. Suite I Gaithersburg, MD 20877

Re: K221871

Trade/Device Name: Healstone Accurate™ Drug of Abuse Urine Test Cup, Healstone Accurate™ Drug of Abuse Urine Test Cup Rx Regulation Number: 21 CFR 862.3100 Regulation Name: Amphetamine Test System Regulatory Class: Class II Product Code: NFT, NGL, PTH, NFV, NFY, PTG, QBF, NGG, QAW, NFW, NGM Dated: June 26, 2022 Received: June 28, 2022

Dear Joe Shia:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Paula Caposino, Ph.D. Acting Deputy Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K221871

Device Name

Healstone Accurate™ Drug of Abuse Urine Test Cup

Indications for Use (Describe)

The Healstone Accurate ™ Drug of Abuse Urine Test Cup is a rapid qualitative immunoassay. The device provides preliminary results for the detection of potential abuse of one or more drugs. This is not a screening device to monitor prescription medication.

CODESUBSTANCECUT-OFF(ng/mL)
AMPAmphetamine1000 or 500
BUPBuprenorphine10
BARSecobarbital300
BZOOxazepam300
COCCocaine300 or 150
EDDP2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine300
MET/mAMPMethamphetamine1000 or 500
MDMA3,4-methylenedioxy-methamphetamine500
OPI/MOPMorphine2000 or 300
MTDMethadone300
OXYOxycodone100
PCPPhencyclidine25
PPXPropoxyphene300
TCANortriptyline1000
THCMarijuana50

This drug test cup may contain any combination of the drug tests listed in the table above.

This drug tests cup provides only a preliminary result. An alternative laboratory test must be used to confirm the results provided by this drug test. Gas chromatography/mass spectrometry (CCMS) is the preferred method confirmation test. For in vitro diagnostic use only.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Indications for Use

510(k) Number (if known)

K221871

Device Name

Healstone Accurate™Drug of Abuse Urine Test Cup Rx

Indications for Use (Describe)

The Healstone Accurate ™ Drug of Abuse Urine Test Cup Rx is a rapid qualitative immunoassay. The device provides preliminary results for the detection of potential abuse of one or more drugs. This is not a screening device to monitor prescription medication.

CODESUBSTANCECUT-OFF(ng/mL)
AMPAmphetamine1000 or 500
BUPBuprenorphine10
BARSecobarbital300
BZOOxazepam300
COCCocaine300 or 150
EDDP2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine300
MET/mAMPMethamphetamine1000 or 500
MDMA3,4-methylenedioxy-methamphetamine500
OPI/MOPMorphine2000 or 300
MTDMethadone300
OXYOxycodone100
PCPPhencyclidine25
PPXPropoxyphene300
TCANortriptyline1000
THCMarijuana50

This drug test cup may contain any combination of the drug tests listed in the table above.

This drug tests cup provides only a preliminary result. An alternative laboratory test must be results provided by this drug test. Gas chromatography/mass spectrometry (GC/MS) is the preferred method confirmation test. It is intended for prescription use. For in vitro diagnostic use only.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{5}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{6}------------------------------------------------

  • Date August 30, 2022 1 Healstone Biotech Inc. 2 Submitter 655 W Kent Ave N #650 Vancouver, BC V6P 6T7, Canada
  • Contact Person Joe Shia 3 LSI International Inc. 504 East Diamond Ave., Suite I Gaithersburg, MD 20877 Telephone: 240-505-7880 Fax: 301-916-6213 Email: shiajl@yahoo.com
  • Healstone Accurate™ Drug of Abuse Urine Test Cup 4 Device Name Healstone Accurate™ Drug of Abuse Urine Test Cup Rx
  • 5 Classification Class II
Product CodeRegulation SectionCup
Target Drug
NFTAmphetamine (AMP)862.3100, Amphetamine TestSystemToxicology
NGLBuprenorphine (BUP)862.3650, Opiate Test SystemToxicology
PTHSecobarbital (BAR)862.3150, Barbiturate TestSystemToxicology
NFVOxazepam (BZO)862.3170,Benzodiazepine Test SystemToxicology
NFYCocaine (COC)862.3250, Cocaine Test SystemToxicology
PTG2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine(EDDP)862.3620, Methadone Test SystemToxicology
NGGMethamphetamine(MET)862.3610,Methamphetamine Test SystemToxicology
NGGMethylenedioxymethamphetamine (MDMA)862.3610,Methamphetamine Test SystemToxicology

{7}------------------------------------------------

NGL862.3650, Opiate Test SystemToxicology
Morphine (MOP/OPI)
PTG862.3620, Methadone Test SystemToxicology
Methadone (MTD)
NGL862.3650, Opiate Test SystemToxicology
Oxycodone (OXY)
NGMUnclassified, EnzymeToxicology
Phencyclidine (PCP)Immunoassay Phencyclidine
QBF862.3700 Propoxyphene testToxicology
Propoxyphene (PPX)system.
QAW862.3910 Tricyclic antidepressantToxicology
Nortriptyline (TCA)drugs test system
NFW862.3870, Cannabinoids TestToxicology
Cannabinoids (THC 50)System

Predicate Device K191841 6.

Healstone Accurate™ Multi Panel Drug Urine Test Cup

7. Intended Use

The Healstone Accurate TM Drug of Abuse Urine Test Cup is a rapid qualitative immunoassay. The device provides preliminary results for the detection of potential abuse of one or more drugs. This is not a screening device to monitor prescription medication.

CODESUBSTANCECUT-OFF (ng/mL)
AMPAmphetamine1000 or 500
BUPBuprenorphine10
BARSecobarbital300
BZOOxazepam300
COCCocaine300 or 150
EDDP2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine300
MET/mAMPMethamphetamine1000 or 500
MDMA3,4-methylenedioxy-methamphetamine500
OPI / MOPMorphine2000 or 300
MTDMethadone300
OXYOxycodone100
PCPPhencyclidine25
PPXPropoxyphene300
TCANortriptyline1000
THCMarijuana50

This drug test cup may contain any combination of the drug tests listed in the table above.

This drug tests cup provides only a preliminary result. An alternative laboratory test must be used to confirm the results provided by this drug test. Gas chromatography/mass spectrometry (GC/MS) is the preferred method confirmation test.

For in vitro diagnostic use only.

The Healstone Accurate ™ Drug of Abuse Urine Test Cup Rx is a rapid qualitative immunoassay.

{8}------------------------------------------------

The device provides preliminary results for the detection of potential abuse of one or more drugs. This is not a screening device to monitor prescription medication.

CODESUBSTANCECUT-OFF (ng/mL)
AMPAmphetamine1000 or 500
BUPBuprenorphine10
BARSecobarbital300
BZOOxazepam300
COCCocaine300 or 150
EDDP2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine300
MET/mAMPMethamphetamine1000 or 500
MDMA3,4-methylenedioxy-methamphetamine500
OPI/MOPMorphine2000 or 300
MTDMethadone300
OXYOxycodone100
PCPPhencyclidine25
PPXPropoxyphene300
TCANortriptyline1000
THCMarijuana50

This drug test cup may contain any combination of the drug tests listed in the table above. This drug tests cup provides only a preliminary result. An alternative laboratory test must be used to confirm the results provided by this drug test. Gas chromatography/mass spectrometry (GC/MS) is the preferred method confirmation test. It is intended for prescription use. For in vitro diagnostic use only.

Device Description 8.

The Healstone Accurate™ Drug of Abuse Urine Test Cup and Healstone Accurate™ Drug of Abuse Urine Test Cup Rx are rapid, single-use in vitro diagnostic devices. Each test kit contains a test device in one pouch. One pouch contains a test Accurate™ Cup and two desiccants, and a package insert. The Healstone Accurate™ Drug of Abuse Urine Test Cup is intended for over-the-counter use and the Healstone Accurate™ Drug of Abuse Urine Test Cup Rx is intended for prescription use.

9. Substantial Equivalence Information

ItemProposed DevicePredicate(K191841)
Indication(s) foruseFor the qualitative determination of Amphetamine,Buprenorphine, Secobarbital, Oxazepam, Cocaine, 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine, Methamphetamine,Methylenedioxymethamphetamine, Morphine, Methadone,Oxycodone, Phencyclidine, Propoxyphene, Nortriptyline andCannabinoids in human urine.Same
MethodologyCompetitive binding, lateral flow immunochromatographic assaybased on antigen-antibody reactionSame

{9}------------------------------------------------

Type of TestQualitativeSame
Specimen TypeHuman urineSame
Target Drug andCut Off ValuesTarget DrugCutoff (ng/mL)Same except that
Amphetamine (AMP)1000 or 500AMP and MET have
Buprenorphine (BUP)10no 500 cutoff, and
Secobarbital (BAR)300COC has no 150
Oxazepam (BZO)300cutoff.
Cocaine (COC)300 or 150
2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP)300
Methamphetamine (MET)1000 or 500
Methylenedioxymethamphetamine (MDMA)500
Morphine (MOP 300/OPI 2000)2000 or 300
Methadone (MTD)300
Oxycodone (OXY)100
Phencyclidine (PCP)25
Propoxyphene (PPX)300
Nortriptyline (TCA)1000
Cannabinoids (THC 50)50
ConfigurationsTest CupCup
Intended UsePrescription Use and over-the-counter useFor over-the-counteruse

10. Test Principle

Healstone Accurate™ Drug of Abuse Urine Test Cup and Healstone Accurate™ Drug of Abuse Urine Test Cup Rx are rapid tests for the qualitative detection of Amphetamine, Buprenorphine, Secobarbital, Oxazepam, Cocaine, 2-ethylidene-1,5-dimethyl-3,3-dipheny|pyrrolidine,Methamphetamine, Methylenedioxymethamphetamine, Morphine, Methadone, Oxycodone, Phencyclidine, Propoxyphene, Nortriptyline and Cannabinoids in urine samples. They are lateral flow chromatographic immunoassay. When urine sample is added to the cup device, urine is absorbed into the test strip and migrates upwards by capillary action. If the concentration of target drug presented in the urine sample is below the cutoff level, the target drug will not saturate the binding sites of its specific monoclonal antibody-coated particles. The antibody-coated particles will then be captured by immobilized drug-conjugate and a visible colored band will be formed on the test line region. If the concentration of target is beyond the cutoff level, the target drug will saturate the binding sites of its specific monoclonal antibody-particles, thus the antibody-coated particles will not be captured by immobilized drug-conjugate hence no colored band will be formed on the test line region.

A band should be formed on the control line region regardless of the presence of target drug or metabolite in the sample to indicate that the tests have been performed properly.

{10}------------------------------------------------

11. Performance Characteristics

1. Analytical Performance

Precision a.

Precision studies were carried out for samples with concentrations of -100% cut off, -75% cut off, -50% cut off, -25% cut off, cutoff, +25% cut off, +50% cut off, +75% cut off and +100% cut off. Samples with concentration of -100% cutoff were drug-free urines samples. Other samples were prepared by spiking target drug in drug-free urine samples. Each drug concentration was confirmed by LC/MS or GC/MS. For each concentration, tests were performed two runs per day for 25 days using three lots of test Cups. The results obtained are summarized in the following tables:

Concentration(ng/mL)Lot Number-100%cut off-75%cut off-50%cut off-25%cutoffcut off+25%cut off+50%cut off+75%cut off+100%cut off
02.45.17.510.212.314.817.619.6
Lot 150-/0+50-/0+50-/0+49-/1+25-/25+48+/2-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+48-/2+26-/24+48+/2-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+47-/3+27-/23+49+/1-50+/0-50+/0-50+/0-

Accurate™ Drug of Abuse Urine Test Cup BUP 10

Accurate™ Drug of Abuse Urine Test Cup PCP 25

Concentration(ng/mL)Lot Number-100%cut off-75%cut off-50%cut off-25%cutoffcut off+25%cut off+50%cut off+75%cut off+100%cut off
05.811.918.725.232.038.743.150.8
Lot 150-/0+50-/0+50-/0+48-/2+22-/28+49+/1-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+49-/1+22-/28+48+/2-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+48-/2+23-/27+48+/2-50+/0-50+/0-50+/0-

Accurate™ Drug of Abuse Urine Test Cup THC 50

Concentration-100%cut off-75%cut off-50%cut off-25%cutoffcut off+25%cut off+50%cut off+75%cut off+100%cut off
(ng/mL)012.426.939.353.564.577.789.6102.4
Lot Number
Lot 150-/0+50-/0+50-/0+49-/1+23-/27+48+/2-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+48-/2+22-/28+48+/2-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+49-/1+23-/27+48+/2-50+/0-50+/0-50+/0-

Accurate™ Drug of Abuse Urine Test Cup OXY 100

Concentration-100%-75%-50%-25%cut off+25%+50%+75%+100%
--------------------------------------------------------------------------

{11}------------------------------------------------

(ng/mL)cut offcut offcut offcutoffcut offcut offcut offcut offcut off
Lot Number026.849.281.1101.5131.6153.4174.1192.0
Lot 150-/0+50-/0+50-/0+49-/1+25-/25+48+/2-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+49-/1+24-/26+49+/1-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+48-/2+27-/23+48+/2-50+/0-50+/0-50+/0-

Accurate™ Drug of Abuse Urine Test Cup BAR 300

Concentration(ng/mL)Lot Number-100%cut off-75%cut off-50%cut off-25%cutoffcut off+25%cut off+50%cut off+75%cut off+100%cut off
078.2145.0230.9316.6382.5455.2533.0587.7
Lot 150-/0+50-/0+50-/0+49-/1+24-/26+49+/1-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+48-/2+25-/25+48+/2-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+49-/1+21-/29+49+/1-50+/0-50+/0-50+/0-

Accurate™ Drug of Abuse Urine Test Cup BZO 300

Concentration(ng/mL)-100%cut off-75%cut off-50%cut off-25%cutoffcut off+25%cut off+50%cut off+75%cut off+100%cut off
Lot Number076.6148.6222.9300.0374.3455.7516.0598.6
Lot 150-/0+50-/0+50-/0+48-/2+24-/26+49+/1-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+48-/2+23-/27+49+/1-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+49-/1+22-/28+48+/2-50+/0-50+/0-50+/0-

Accurate™ Drug of Abuse Urine Test Cup EDDP 300

Concentration(ng/mL)-100%cut off-75%cut off-50%cut off-25%cutoffcut off+25%cut off+50%cut off+75%cut off+100%cut off
Lot Number072.4149.2233.2312.4385.2451.2522.0599.0
Lot 150-/0+50-/0+50-/0+49-/1+23-/27+49+/1-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+21-/29+48+/2-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+49-/1+22-/28+49+/1-50+/0-50+/0-50+/0-

Accurate™ Drug of Abuse Urine Test Cup MTD 300

Concentration(ng/mL)Lot Number-100%cut off-75%cut off-50%cut off-25%cutoffcut off+25%cut off+50%cut off+75%cut off+100%cut off
073.7156.5229.9314.4365.1456.9521.2590.2
Lot 150-/0+50-/0+50-/0+49-/1+23-/27+49+/1-50+/0-50+/0-50+/0-

{12}------------------------------------------------

Lot 250-/0+50-/0+50-/0+48-/2+23-/27+48+/2-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+48-/2+24-/26+48+/2-50+/0-50+/0-50+/0-

Accurate™ Drug of Abuse Urine Test Cup MOP 300

Concentration(ng/mL)Lot Number-100%cut off-75%cut off-50%cut off-25%cutoffcut off+25%cut off+50%cut off+75%cut off+100%cut off
069.1148.2228.7302.0382.1445.0509.1583.2
Lot 150-/0+50-/0+50-/0+49-/1+23-/27+48+/2-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+48-/2+24-/26+48+/2-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+49-/1+23-/27+49+/1-50+/0-50+/0-50+/0-

Accurate™ Drug of Abuse Urine Test Cup PPX 300

Concentration(ng/mL)Lot Number-100%cut off-75%cut off-50%cut off-25%cutoffcut off+25%cut off+50%cut off+75%cut off+100%cut off
073.0160.0236.0320.7384.7458.7538.6579.9
Lot 150-/0+50-/0+50-/0+48-/2+26-/24+49+/1-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+49-/1+26-/24+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+48-/2+25-/25+48+/2-50+/0-50+/0-50+/0-

Accurate™ Drug of Abuse Urine Test Cup COC 150

Concentration(ng/mL)Lot Number-100%cut off-75%cut off-50%cut off-25%cutoffcut off+25%cut off+50%cut off+75%cut off+100%cut off
Lot 150-/0+50-/0+50-/0+49-/1+22-/28+48+/2-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+48-/2+21-/29+49+/1-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+49-/1+23-/27+49+/1-50+/0-50+/0-50+/0-

Accurate™ Drug of Abuse Urine Test Cup MDMA 500

Concentration(ng/mL)Lot Number-100%cut off-75%cut off-50%cut off-25%cutoff+25%cut off+50%cut off+75%cut off+100%cut off
0122.9272.8370.6499.5628.4752.1870.4935.7
Lot 150-/0+50-/0+50-/0+49-/1+22-/28+48+/2-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+48-/2+21-/29+48+/2-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+48-/2+23-/27+49+/1-50+/0-50+/0-50+/0-

{13}------------------------------------------------

Accurate™ Drug of Abuse Urine Test Cup TCA 1000

Concentration(ng/mL)-100%cut off-75%cut off-50%cut off-25%cutoffcut off+25%cut off+50%cut off+75%cut off+100%cut off
Lot Number0248.7489.5758.2988.31288.61536.51701.31958.2
Lot 150-/0+50-/0+50-/0+50-/0+23-/27+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+22-/28+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+22-/28+50+/0-50+/0-50+/0-50+/0-

Accurate™ Drug of Abuse Urine Test Cup AMP 500

Concentration(ng/mL)-100%cut off-75%cut off-50%cut off-25%cutoffcut off+25%cut off+50%cut off+75%cut off+100%cut off
Lot Number0123.4267.8354.5498.5630.4716.0850.7
Lot 150-/0+50-/0+50-/0+48-/2+24-/26+48+/2-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+48-/2+26-/24+49+/1-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+49-/1+25-/25+48+/2-50+/0-50+/0-50+/0-

Accurate™ Drug of Abuse Urine Test Cup MET 500

Concentration(ng/mL)Lot Number-100%cut off-75%cut off-50%cut off-25%cutoffcut off+25%cut off+50%cut off+75%cut off+100%cut off
0128.8262.7365.1512.9680.8790.4909.21021.3
Lot 150-/0+50-/0+50-/0+49-/1+26-/24+49+/1-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+49-/1+26-/24+48+/2-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+48-/2+25-/25+49+/1-50+/0-50+/0-50+/0-

Accurate™ Drug of Abuse Urine Test Cup OPI 2000

Concentration(ng/mL)Lot Number-100%cut off-75%cut off-50%cut off-25%cutoffcut off+25%cut off+50%cut off+75%cut off+100%cut off
0535.81070.41636.62122.32632.63014.83337.53762.8
Lot 150-/0+50-/0+50-/0+50-/0+24-/26+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+23-/27+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+24-/26+50+/0-50+/0-50+/0-50+/0-

Accurate™ Drug of Abuse Urine Test Cup COC 300

Concentration(ng/mL)Lot Number-100%cut off-75%cut off-50%cut off-25%cutoffcut off+25%cut off+50%cut off+75%cut off+100%cut off
079.9159.9229.9301.1380.2463.4529.1593.2
Lot 150-/0+50-/0+50-/0+49-/1+22-/28+49+/1-50+/0-50+/0-50+/0-

{14}------------------------------------------------

Lot 250-/0+50-/0+50-/0+48-/2+22-/28+49+/1-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+49-/1+24-/26+49+/1-50+/0-50+/0-50+/0-

Accurate™ Drug of Abuse Urine Test Cup AMP 1000

Concentration(ng/mL)Lot Number-100%cut off0-75%cut off235.8-50%cut off525.0-25%cutoff785.9cut off1086.6+25%cut off1315.3+50%cut off1599.2+75%cut off1816.7+100%cut off2040.5
Lot 150-/0+50-/0+50-/0+50-/0+26-/24+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+26-/24+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+27-/23+50+/0-50+/0-50+/0-50+/0-

Accurate™ Drug of Abuse Urine Test Cup MET 1000

Concentration(ng/mL)Lot Number-100%cut off-75%cut off-50%cut off-25%cutoffcut off+25%cut off+50%cut off+75%cut off+100%cut off
0252.5512.2767.21021.91251.61496.41781.42005.1
Lot 150-/0+50-/0+50-/0+50-/0+24-/26+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+23-/27+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+24-/26+50+/0-50+/0-50+/0-50+/0-

The following cutoff values are verified:

Target DrugCut-off level
Amphetamine (AMP)1000 ng/mL or 500 ng/mL
Buprenorphine (BUP)10 ng/mL
Secobarbital (BAR)300 ng/mL
Oxazepam (BZO)300 ng/mL
Cocaine (COC)300 ng/mL or 150 ng/mL
2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP)300 ng/mL
Methamphetamine (MET)1000 ng/mL or 500 ng/mL
Methylenedioxymethamphetamine (MDMA)500 ng/mL
Morphine (OPI 2000/MOP 300)2000 ng/mL or 300 ng/mL
Methadone (MTD)300 ng/mL
Oxycodone (OXY)100 ng/mL
Phencyclidine (PCP)25 ng/mL
Propoxyphene (PPX)300 ng/mL
Nortriptyline (TCA)1000 ng/mL
Cannabinoids (THC 50)50 ng/mL

{15}------------------------------------------------

b. Linearity

Not applicable

  • c. Stability
    The devices are stable at 2-30℃ for 24 months based on real-time stability studies.

  • d. Interference
    Potential interfering substances were added to drug-free urine samples with target drugs of -50% cutoff and +50% cutoff level.

Compounds that show no interference at a concentration of 100μg/mL are summarized in the following table.

(-) CotinineDiphenhydramine HClNoscapine
(±)-4-Methylephedrine-D3D,L-PropranololO-Hydroxyhippuric acid
3-HydroxytyramineD,L-TyrosineOlanzapine
AcetaminophenDoxylamineOxalic acid
AcetophenetidinD-PseudoephedrineOxolinic acid
Acetylsalicylic acidEcgonine methyl esterOxymetazoline
Albumin(100mg/dL)EMDPPapaverine
Alpha MethadolErythromycinPenicillin-G
AminopyrineEthanol (1%)Perphenazine
AmoxicillinFenoprofenPhenacetin
AmpicillinFurosemidePhenelzine
ApomorphineGabapentinPhenethylamine
Ascorbic acidGatifloxacinPhenylpropanolamine
AspartameGentisic acidPrednisone
AspirinGlucoseProcaine
AtropineHemoglobinPromethazine
AzithromycinHydralazineQuetiapine
Benzilic acidHydrochlorothiazideQuinine
Benzoic acidHydrocortisoneRanitidine
BenzphetamineIbuprofenSalicylic acid
BilirubinIsoxsuprineSerotonin

{16}------------------------------------------------

CannabidiolKetamineSerotonin (5- Hydroxytyramine)
CarfentanilKetoprofenSertraline
ChloralhydrateLAAM HClSulfamethazine
ChloramphenicolLabetalolSulindac
ChloroquineL-EphedrineTelmisartan
ChlorothiazideL-EpinephrineTetrahydrocortisone 3-(β-Dglucuronide)
ChlorpromazineLoperamideTetrahydrocortisone, 3-acetate
CholesterolLoratadineTetrahydrozoline
ClonidineL-phenylephrineThiamine
CortisoneMagnesiumThioridazine
CreatinineMaprotilineTriamterene
D,L-TryptophanMeperidineTrifluoperazine
D,L-IsoproterenolMeprobamateTrimethobenzamide
D,L-OctopamineMethoxyphenamineTrimethoprim
D,L-EpinephrineN-AcetylprocainamideUric acid
DelorazepamNalidixic acidVerapamil
DeoxycorticosteroneNaproxenVitamin B2
DesloratadineNiacinamideVitamin C
DextromethorphanNicotineZaleplon
DiclofenacNifedipineZomepirac
Diclofenac sodiumNordoxepinβ-Estradiol
DiflunisalNorethindrone
DigoxinNorfentanyl
  • e. Specificity
    To test the specificity, drug metabolites and other components that are likely to cross-react in urine samples were spiked into drug-free urine. These urine samples were tested using three lots of each device.

Percent cross-reactivity, provided in the below table, was calculated by dividing the cutoff concentration by the minimum concentration required to obtain a positive result, multiplied by 100; compounds that did not yield a positive result at the highest concentration tested have relative cross reactivity results represented by a dash in the table below:

{17}------------------------------------------------

BUP 10 (Buprenorphine,Cutoff=10 ng/mL)Minimum concentrationrequired to obtain apositive result (ng/mL)% Cross-Reactivity
Buprenorphine -3-D-Glucuronide1606.25%
Norbuprenorphine10100%
Norbuprenorphine-3-D-Glucuronide2005%
Morphine>100,000-
Oxymorphone>100,000--
Hydromorphone>100,000--
PCP (Phencyclidine)(Phencyclidine,Cutoff=25 ng/mL)Minimum concentrationrequired to obtain apositive result (ng/mL)% Cross-Reactivity
4-Hydroxyphencyclidine125000.2%
THC 50(11-nor-Δ9-THC-9-COOH,Cutoff=50 ng/mL)Minimum concentrationrequired to obtain apositive result (ng/mL)% Cross-Reactivity
11-nor-Δ8-THC 50-9-COOH30167%
(-)-11-nor-9-carboxy-Δ 9-THC 5050100%
11-nor-Δ9-THC 50-carboxy glucuronide10050%
11-hydroxy-Δ9-Tetrahydrocannabinol50001%
Δ8- Tetrahydrocannabinol13003.8%
Δ9- Tetrahydrocannabinol50001%
Cannabinol200000.25%
Cannabidiol>100000--
OXY 100(Oxycodone, Cutoff=100 ng/mL)Minimum concentrationrequired to obtain apositive result (ng/mL)% Cross-Reactivity
Ethyl Oxycodone75,0000.13%
Hydrocodone5,0002%
Hydromorphone25,0000.4%
Levorphanol tartrate25,0000.4%
Naloxone hydrochloride10,0001%
Naltrexone hydrochloride50,0000.2%
Oxymorphone20050%
Oxymorphone-D320050%
Dihydrocodeine>100,000--
Codeine>100,000--
Morphine>100,000--
Acetylmorphine>100,000--

{18}------------------------------------------------

Buprenorphine>100,000--
Ethylmorphine>100,000--
Thebaine>100,000--
COC 150(Benzoylecgonine, Cutoff=150 ng/mL)Minimum concentrationrequired to obtain apositive result (ng/mL)% Cross-Reactivity
Cocaethylene150100%
Cocaine HCl150100%
Ecgonine25,0000.6%
Norcocaine50,0000.3%
Ecgonine methyl Ester>100000--
BAR 300(Secobarbital, Cutoff=300ng/mL)Minimum concentrationrequired to obtain apositive result (ng/mL)% Cross-Reactivity
Alphenal150200%
Amobarbital300100%
Aprobarbital250120%
Butabarbital2,50012%
Butethal100300%
Cyclopentobarbital60050%
Pentobarbital250120%
Phenobarbital250120%
Butalbital2,50012%
BZO 300(Oxazepam, Cutoff=300ng/mL)Minimum concentrationrequired to obtain apositive result (ng/mL)% Cross-Reactivity
a-Hydroxyalprazolam1,26023.8%
Alprazolam200150%
Bromazepam1,56019.2%
Chlordiazepoxide1,56019.2%
Clobazam100300%
Clonazepam2,50012%
Clorazepate Dipotassium200150%
Desalkylflurazepam40075%
Diazepam200150%
Estazolam2,50012%
Flunitrazepam40075%
D,L-Lorazepam1,56019.2%
Midazolam12,5002.4%

{19}------------------------------------------------

Nitrazepam100300%
Norchlordiazepoxide200150%
Nordiazepam40075%
R,S-Lorazepam glucuronide160187.5%
Temazepam100300%
Triazolam2,50012%
Demoxepam2,00015%
Flurazepam50060%
Delorazepam>100,000--
EDDP 300(2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine, Cutoff = 300 ng/mL)Minimum concentrationrequired to obtain apositive result (ng/mL)% Cross-Reactivity
Methadone>100,000--
EMDP>100,000--
Doxylamine>100,000--
Disopyramide>100,000--
LAAM (Levo-alpha-acetylmethadol) HCl>100,000--
Alpha Methadol>100,000--
MTD 300(Methadone, Cutoff=300ng/mL)Minimum concentrationrequired to obtain apositive result (ng/mL)% Cross-Reactivity
Doxylamine>100,000--
EDDP>100,000--
EMDP>100,000--
LAAM>100,000--
Alpha Methadol>100,000--
MOP 300(Morphine, Cutoff=300ng/mL)Minimum concentrationrequired to obtain apositive result (ng/mL)% Cross-Reactivity
6-acetylmorphine40075%
Codeine300100%
Dihydrocodeine1,00030%
EthylMorphine100300%
Heroin60050%
6-Monoacetylmorphine150200%
Hydrocodone10,0003%
Hydromorphone50060%
Levorphanol tartrate10,0003%
Nalorphine HCl50,0000.6%

{20}------------------------------------------------

Thebaine6,2404.8%
s-Monoacetylmorphine300100%
Morphine-3-β-d-glucuronide1,00030%
Normorphine>100,000--
Oxycodone>100000--
Oxymorphone>100000--
Norcodeine>100000--
Procaine>100000--
PPX 300(Propoxyphene, Cutoff=300ng/mL)Minimum concentrationrequired to obtain apositive result (ng/mL)% Cross-Reactivity
d-Norpropoxyphene300100%
MDMA 500(3,4-Methylenedioxymethamphetamine HCl,Cutoff=500ng/mL)Minimum concentrationrequired to obtain apositive result (ng/mL)% Cross-Reactivity
(+/-)3,4-Methylenedioxy-n-ethylamphetamine(MDEA)300166.7%
(+/-)-Methylenedioxyamphetamine(MDA)3,00016.7%
L-Methamphetamine50,0001%
D-Methamphetamine>100,000--
D-Amphetamine>100,000--
L-Amphetamine>100,000--
AMP 500(Amphetamine, Cutoff=500ng/mL)Minimum concentrationrequired to obtain apositive result (ng/mL)% Cross-Reactivity
Hydroxyamphetamine4,00012.5%
(+/-)-Methylenedioxyamphetamine(MDA)200250%
D,L-Amphetamine500100%
Diethylstilbestrol2,50020%
L-Amphetamine25,0002%
p-Hydroxyamphetamine50,0001%
Phentermine4,00012.5%
β-Phenylethylamine50,0001%
Tyramine50,0001%
p-Hydroxynorephedrine50,0001%
D,L-Norephedrine50,0001%
(+/-)3,4-Methylenedioxy-n-ethylamphetamine(MDEA)100,0000.5%

{21}------------------------------------------------

D-Methamphetamine>100,000--
L-Methamphetamine>100,000--
(+/-)3,4-Methylenedioxymethamphetamine(MDMA)>100,000--
Ephedrine HCl>100,000--
Phenylpropanolamine>100,000--
Benzphetamine>100,000--
L-Ephedrine>100,000--
L-Epinephrine>100,000--
D,L-Epinephrine>100,000--
MET 500(D(+)-Methamphetamine, Cutoff=500ng/mL)Minimum concentrationrequired to obtain apositive result (ng/mL)% Cross-Reactivity
(+/-)3,4-Methylenedioxy-n-ethylamphetamine(MDEA)10,0005%
(±)-MDMA1,25040%
L-Methamphetamine12,5004%
Fenfluramine25,0002%
p-Hydroxymethamphetamine5,00010%
D,L-Methamphetamine500100%
β-Phenylethylamine25,0002%
Mephetermine25,0002%
L-Amphetamine40,0001.25%
Ephedrine HCl100,0000.5%
(1R,2S)-(-)-Ephedrine100,0000.5%
D-Amphetamine>100,000--
Chloroquine>100,000--
(+/-)3,4- Methylenedioxyamphetamine(MDA)>100,000--
L-Phenylephrine>100,000--
TCA 1000(Nortriptyline, Cutoff=1000ng/mL)Minimum concentrationrequired to obtain apositive result (ng/mL)% Cross-Reactivity
Amitriptyline1,50066.7%
Chlorpheniramine50,0002%
Clomipramine10,00010%
Cyclobenzaprine Hydrochloride5,00020%
Desipramine1,000100%
Doxepine2,00050%
Duloxetine10,00010%

{22}------------------------------------------------

Imipramine1,000100%
Norclomipramine12,5008%
Nordoxepine1,000100%
Promazine50,0002%
Trimipramine10,00010%
Maprotiline>100,000--
Promethazine hydrochloride>100,000--
COC 300(Benzoylecgonine, Cutoff=300ng/mL)Minimum concentrationrequired to obtain apositive result (ng/mL)% Cross-Reactivity
Cocaethylene300100%
Cocaine HCl300100%
Ecgonine50,0000.6%
Norcocaine100,0000.3%
Ecgonine methyl ester>100,000--
AMP 1000(d-Amphetamine, Cutoff=1000ng/mL)Minimum concentrationrequired to obtain apositive result (ng/mL)% Cross-Reactivity
Hydroxyamphetamine800012.5%
(+/-)-Methylenedioxyamphetamine(MDA)400250%
D,L-Amphetamine1,000100%
Diethylstilbestrol5,00020%
L-Amphetamine50,0002%
p-Hydroxyamphetamine100,0001%
Phentermine8,00012.5%
β-Phenylethylamine100,0001%
Tyramine100,0001%
p-Hydroxynorephedrine100,0001%
D,L-Norephedrine100,0001%
D-Methamphetamine>100000--
L-Methamphetamine>100000--
Ephedrine HCl>100000--
(+/-)3,4- Methylenedioxymethamphetamine(MDMA)>100000--
Phenylpropanolamine>100000--
Benzphetamine>100000--
L-Ephedrine>100,000--
L-Epinephrine>100,000--
D,L-Epinephrine>100,000--

{23}------------------------------------------------

MET 1000(D(+)-Methamphetamine, Cutoff=1000ng/mL)Minimum concentrationrequired to obtain apositive result (ng/mL)% Cross-Reactivity
(+/-)3,4-Methylenedioxy-n-ethylamphetamine(MDEA)20,0005%
(+/-)3,4-Methylenedioxymethamphetamine(MDMA)2,50040%
L-Methamphetamine25,0004%
Fenfluramine50,0002%
p-Hydroxymethamphetamine10,00010%
D,L-Methamphetamine1,000100%
β-Phenylethylamine50,0002%
Mephetermine50,0002%
L-Amphetamine75,0001.33%
D-Amphetamine>100000--
D,L-Amphetamine>100000--
Chloroquine>100000--
Ephedrine HCl>100000--
(+/-)3,4-Methylenedioxyamphetamine (MDA)>100000--
Trimethobenzamide>100000--
L-Phenylephrine>100000--
(1R,2S)-(-)-Ephedrine>100000--
OPI 2000(Morphine, Cutoff=2000ng/mL)Minimum concentrationrequired to obtain apositive result (ng/mL)% Cross-Reactivity
6-acetylmorphine2,50080%
Codeine1,000200%
Dihydrocodeine1,500133.3%
EthylMorphine2,50080%
6-Monoacetylmorphine1500133%
Heroin5,00040%
Hydrocodone5,00040%
Hydromorphone25,0008%
Levorphanol tartrate10,00020%
Nalorphine HCl5,00040%
Norcodeine4,00050%
Normorphine5,00040%
Oxymorphone75,0002.7%
s-Monoacetylmorphine2,000100%

{24}------------------------------------------------

Thebaine13,00015.4%
Morphine 3-β-D-glucuronide2,000100%
Oxycodone>100,000--
Procaine>100,000--
  • f. Effect of Urine Specific Gravity and Urine pH
    To investigate the effect of urine specific gravity, urine samples with specific gravity from 1.000 to 1.035 were spiked with target drugs at +50% cutoff and -50% cutoff levels. Three operators tested each sample using test devices from three different lots. The results were all positive for samples at +50% cutoff and all negative for samples at -50% cutoff. No interference was observed at any specific gravity condition tested.

To investigate the effect of urine pH, urine samples with pH value from 4 to 9 were spiked with target drugs at +50% cutoff and -50% cutoff levels. Three operators tested each sample using test devices from three different lots. The results were all positive for samples at +50% cutoff and all negative for samples at -50% cutoff, indicating that urine pH value between 4.0 and 9.0 has no interference with the test device.

    1. Comparison Studies
      The method comparison studies for Healstone Accurate™ Drug of Abuse Urine Test Cup were performed in-house with three operators.

Operators ran 80 (40 negative and 40 positive) unaltered urine samples were blind labeled and compared to LC/MS or GC/MS results. The results are presented in the table below:

For Accurate™ Drug of Abuse Urine Test Cup:

HealstoneAccurate™Drug-FreeLowNegative byLC/MS (lessthan -50%)Near CutoffNegative byLC/MS(Between -50%and the Cutoff)Near CutoffPositive byLC/MS(Between thecutoff and+50%)High Positiveby LC/MS(greater than+50%)
Operator APositive0022018
Negative1019920
Operator BPositive0011918
Negative10191030
Operator CPositive0012018
Negative10191020
AMP 500

Discordant Results for AMP 500:

OperatorSample NumberLC/MS Result (ng/mL)Healstone AccurateTM Result
----------------------------------------------------------------------------

{25}------------------------------------------------

Operator ATN-SAMP06072450.4+
Operator ATN-SAMP06044467.2+
Operator BTN-SAMP06072450.4+
Operator CTN-SAMP06044467.2+
Operator ATN-SAMP06006518.5-
Operator ATN-SAMP06022526.4-
Operator BTN-SAMP06041532.0-
Operator BTN-SAMP06057532.5-
Operator BTN-SAMP06073539.1-
Operator CTN-SAMP06006518.5-
Operator CTN-SAMP06057532.5-

BUP 10

HealstoneAccurateTMDrug-FreeLowNegative byLC/MS (lessthan -50%)Near CutoffNegative byLC/MS(Between -50%and the Cutoff)Near CutoffPositive byLC/MS(Between thecutoff and+50%)High Positiveby LC/MS(greater than+50%)
Operator APositive0001324
Negative10151530
Operator BPositive0011324
Negative10151430
Operator CPositive0031624
Negative10151200

Discordant Results for BUP 10:

OperatorSample NumberLC/MS Result (ng/mL)Healstone AccurateTMResult
Operator BTN-BUP060429.6+
Operator CTN-BUP060409.5+
Operator CTN-BUP060429.6+
Operator CTN-BUP060139.8+
Operator ATN-BUP0604810.0-
Operator ATN-BUP0605810.1-
Operator ATN-BUP0600210.8-
Operator BTN-BUP0603710.2-
Operator BTN-BUP0600210.8-
Operator BTN-BUP0602311.0-

{26}------------------------------------------------

HealstoneAccurate™Drug-FreeLowNegative byLC/MS (lessthan -50%)Near CutoffNegative byLC/MS(Between -50%and the Cutoff)Near CutoffPositive byLC/MS(Between thecutoff and+50%)High Positiveby LC/MS(greater than+50%)
Operator APositive0021722
Negative10151310
Operator BPositive0021422
Negative10151340
Operator CPositive0001522
Negative10151530

Discordant Results for BAR 300:

OperatorSample NumberLC/MS Result (ng/mL)Healstone Accurate™Result
Operator ATN-BAR06061266.1+
Operator ATN-BAR06013288.3+
Operator BTN-BAR06046286.3+
Operator BTN-BAR06076299.7+
Operator ATN-BAR06054317.2-
Operator BTN-BAR06011314.7-
Operator BTN-BAR06074315.8-
Operator BTN-BAR06039320.1-
Operator BTN-BAR06004324.9-
Operator CTN-BAR06074315.8-
Operator CTN-BAR06054317.2-
Operator CTN-BAR06039320.1-

BZO 300

HealstoneAccurate™Drug-FreeLowNegative byLC/MS (lessthan -50%)Near CutoffNegative byLC/MS(Between -50%and the Cutoff)Near CutoffPositive byLC/MS(Between thecutoff and+50%)High Positiveby LC/MS(greater than+50%)
Operator APositive0021624
Negative10161200
Operator BPositive0021624
Negative10161200
Operator CPositive0021624

{27}------------------------------------------------

Negative10161200
------------------------------

Discordant Results for BZO 300:

OperatorSample NumberLC/MS Result (ng/mL)Healstone AccurateTM Result
Operator ATN-BZO06041294.7+
Operator ATN-BZO06006296.2+
Operator BTN-BZO06022239.6+
Operator BTN-BZO06041294.7+
Operator CTN-BZO06076233.4+
Operator CTN-BZO06006296.2+

COC 150

HealstoneAccurate™Drug-FreeLowNegative byLC/MS(less than -50%)Near CutoffNegative byLC/MS(Between -50% and theCutoff)Near CutoffPositive byLC/MS(Between thecutoff and+50%)HighPositive byLC/MS(greaterthan +50%)
Operator APositive0031425
Negative10131410
Operator BPositive0021425
Negative10131510
Operator CPositive0011525
Negative10131600

Discordant Results for COC 150:

OperatorSample NumberLC/MS Result (ng/mL)HealstoneAccurateTM Result
Operator ATN-SCOC06066126.0+
Operator ATN-SCOC06067134.4+
Operator ATN-SCOC06005141.4+
Operator BTN-SCOC06005141.4+
Operator BTN-SCOC06040146.6+
Operator CTN-SCOC06066126.0+
Operator ATN-SCOC06025178.4-
Operator BTN-SCOC06049153.2-

{28}------------------------------------------------

HealstoneAccurateTMDrug-FreeLowNegative byLC/MS (lessthan -50%)Near CutoffNegative byLC/MS(Between -50% and theCutoff)Near CutoffPositive byLC/MS(Between thecutoff and+50%)High Positiveby LC/MS(greater than+50%)
Operator APositive003
Negative10151220
Operator BPositive0011623
Negative10151410
Operator CPositive0011623
Negative10151410

Discordant Results for EDDP 300:

OperatorSample NumberLC/MS Result (ng/mL)HealstoneAccurate™ Result
Operator ATN-EDDP06031236.0+
Operator ATN-EDDP06054292.9+
Operator ATN-EDDP06078294.9+
Operator BTN-EDDP06065251.0+
Operator CTN-EDDP06074274.7+
Operator ATN-EDDP06038341.8-
Operator ATN-EDDP06003342.4-
Operator BTN-EDDP06047331.2-
Operator CTN-EDDP06047331.2-

MET 500

HealstoneAccurate™Drug-FreeLowNegative byLC/MS (lessthan -50%)Near CutoffNegative byLC/MS(Between -50%and the Cutoff)Near CutoffPositive byLC/MS(Between thecutoff and+50%)High Positiveby LC/MS(greater than+50%)
Operator APositive0021820
Negative1020820
Operator BPositive0011720
Negative1020930
Operator CPositive0001620
Negative10201040

Discordant Results for MET 500:

{29}------------------------------------------------

OperatorSample NumberLC/MS Result (ng/mL)HealstoneAccurateTM Result
Operator ATN-SMET06057433.5+
Operator ATN-SMET06033489.7+
Operator BTN-SMET06068497.6+
Operator ATN-SMET06014569.7-
Operator ATN-SMET06078594.0-
Operator BTN-SMET06012540.8-
Operator BTN-SMET06014569.7-
Operator BTN-SMET06077571.4-
Operator CTN-SMET06012540.8-
Operator CTN-SMET06014569.7-
Operator CTN-SMET06007587.6-
Operator CTN-SMET06042610.4-

MDMA 500

HealstoneAccurateTMDrug-FreeLowNegative byLC/MS (lessthan -50%)Near CutoffNegative byLC/MS(Between -50%and the Cutoff)Near CutoffPositive byLC/MS(Between thecutoff and+50%)High Positiveby LC/MS(greater than+50%)
Operator APositive0031920
Negative10131410
Operator BPositive0031720
Negative10131430
Operator CPositive0011720
Negative10131630

Discordant Results for MDMA 500:

OperatorSample NumberLC/MS Result (ng/mL)HealstoneAccurateTM Result
Operator ATN-MDMA06032478.7+
Operator ATN-MDMA06004479.3+
Operator ATN-MDMA06005496.8+
Operator BTN-MDMA06038459.1+
Operator BTN-MDMA06076473.3+
Operator BTN-MDMA06005496.8+
Operator CTN-MDMA06005496.8+
Operator ATN-MDMA06075518.6-
Operator BTN-MDMA06065501.8-

{30}------------------------------------------------

Operator BTN-MDMA06057529.3-
Operator BTN-MDMA06022587-
Operator CTN-MDMA06057529.3-
Operator CTN-MDMA06067529.3-
Operator CTN-MDMA06047595.7-

MOP 300

HealstoneAccurate™Drug-FreeLowNegative byLC/MS (lessthan -50%)Near CutoffNegative byLC/MS(Between -50%and the Cutoff)Near CutoffPositive byLC/MS(Between thecutoff and+50%)High Positiveby LC/MS(greater than+50%)
Operator APositive0011028
Negative10171220
Operator BPositive0001128
Negative10171310
Operator CPositive0001128
Negative10171310

Discordant Results for MOP 300:

OperatorSample NumberLC/MS Result (ng/mL)HealstoneAccurate™ Result
Operator ATN-MOP06014295.8+
Operator ATN-MOP06061308.6-
Operator ATN-MOP06074314.9-
Operator BTN-MOP06061308.6-
Operator CTN-MOP06074314.9-

MTD 300

HealstoneAccurateTMDrug-FreeLowNegative byLC/MS (lessthan -50%)Near CutoffNegative byLC/MS(Between -50%and the Cutoff)Near CutoffPositive byLC/MS(Between thecutoff and+50%)High Positiveby LC/MS(greater than+50%)
Operator APositive0011722
Negative10181110
Operator BPositive0001522
Negative10181230

{31}------------------------------------------------

Operator C
Operator CPositive0011722
Negative10181110
OperatorSample NumberLC/MS Result (ng/mL)Healstone AccurateTM Result
Operator ATN-MTD06078273.6+
Operator CTN-MTD06078273.6+
Operator ATN-MTD06072305.1-
Operator BTN-MTD06042301.7-
Operator BTN-MTD06072305.1-
Operator BTN-MTD06077333.9-
Operator CTN-MTD06024339.6-

Discordant Results for MTD 300:

OXY 100

HealstoneAccurate™Drug-FreeLowNegative byLC/MS (lessthan -50%)Near CutoffNegative byLC/MS(Between -50%and the Cutoff)Near CutoffPositive byLC/MS(Between thecutoff and+50%)High Positiveby LC/MS(greater than+50%)
Operator APositive0001621
Negative10151530
Operator BPositive0011821
Negative10151410
Operator CPositive0011621
Negative10151430

Discordant Results for OXY 100:

OperatorSample NumberLC/MS Result (ng/mL)HealstoneAccurateTM Result
Operator BTN-OXY0602193.3+
Operator CTN-OXY0602193.3+
Operator ATN-OXY06040107.1-
Operator ATN-OXY06010109.6-
Operator ATN-OXY06002128.7-
Operator BTN-OXY06010109.6-
Operator CTN-OXY06005106.8-
Operator CTN-OXY06040107.1-
Operator CTN-OXY06010109.6-

{32}------------------------------------------------

PCP 25

HealstoneAccurate™Drug-FreeLowNegative byLC/MS (lessthan -50%)Near CutoffNegative byLC/MS(Between -50%and the Cutoff)Near CutoffPositive byLC/MS(Between thecutoff and+50%)High Positiveby LC/MS(greater than+50%)
Operator APositive00112
Negative10181110
Operator BPositive0021227
Negative10181010
Operator CPositive0001227
Negative10181210

Discordant Results for PCP 25:

OperatorSample NumberLC/MS Result (ng/mL)Healstone AccurateTMResult
Operator ATN-PCP0605222.0+
Operator BTN-PCP0605222.0+
Operator BTN-PCP0603120.6+
Operator ATN-PCP0606927.1-
Operator BTN-PCP0606927.1-
Operator CTN-PCP0607627.5-

PPX 300

HealstoneAccurateTMDrug-FreeLowNegative byLC/MS (lessthan -50%)Near CutoffNegative byLC/MS(Between -50%and the Cutoff)Near CutoffPositive byLC/MS(Between thecutoff and+50%)High Positiveby LC/MS(greater than+50%)
Operator APositive0021524
Negative10171110
Operator BPositive0001324
Negative10171330
Operator CPositive0021524
Negative10171110

Discordant Results for PPX 300:

OperatorSample NumberLC/MS Result (ng/mL)Healstone AccurateTM Result

{33}------------------------------------------------

Operator ATN-PPX06035289.0+
Operator ATN-PPX06070292.4+
Operator CTN-PPX06035289.0+
Operator CTN-PPX06070292.4+
Operator ATN-PPX06020335.1-
Operator BTN-PPX06038331.9-
Operator BTN-PPX06040331.9-
Operator BTN-PPX06003350.5-
Operator CTN-PPX06040331.9-

TCA 1000

HealstoneAccurate™Drug-FreeLowNegative byLC/MS (lessthan -50%)Near CutoffNegative byLC/MS(Between -50%and the Cutoff)Near CutoffPositive byLC/MS(Between thecutoff and+50%)High Positiveby LC/MS(greater than+50%)
Operator APositive0031422
Negative10141340
Operator BPositive0031622
Negative10141320
Operator CPositive0011622
Negative10141520

Discordant Results for TCA 1000:

OperatorSample NumberLC/MS Result (ng/mL)HealstoneAccurate™ Result
Operator ATN-TCA06066888.4+
Operator ATN-TCA06050908.9+
Operator ATN-TCA06015919.8+
Operator BTN-TCA06018890.8+
Operator BTN-TCA06031917.1+
Operator BTN-TCA06015919.8+
Operator CTN-TCA06050908.9+
Operator ATN-TCA060541071.4-
Operator ATN-TCA060441085.0-
Operator ATN-TCA060551097.2-
Operator ATN-TCA060191108.7-
Operator BTN-TCA060441085.0-
Operator BTN-TCA060191108.7-
Operator CTN-TCA060441085.0-

{34}------------------------------------------------

Operator CTN-TCA060731100.4-
------------------------------------

THC 50

HealstoneAccurate™Drug-FreeLowNegative byLC/MS (lessthan -50%)Near CutoffNegative byLC/MS(Between -50%and the Cutoff)Near CutoffPositive byLC/MS(Between thecutoff and+50%)High Positiveby LC/MS(greater than+50%)
Operator APositive0011522
Negative10161330
Operator BPositive0001422
Negative10161440
Operator CPositive0011522
Negative10161330

Discordant Results for THC 50:

OperatorSample NumberLC/MS Result (ng/mL)HealstoneAccurateTM Result
Operator ATN-THC0604147.3+
Operator CTN-THC0607246.6+
Operator ATN-THC0607051.8-
Operator ATN-THC0606056.8-
Operator ATN-THC0600157.4-
Operator BTN-THC0607051.8-
Operator BTN-THC0602152.4-
Operator BTN-THC0603653.0-
Operator BTN-THC0600157.4-
Operator CTN-THC0603653.0-
Operator CTN-THC0606957.7-
Operator CTN-THC0604057.9-

AMP 1000

HealstoneAccurate™Drug-FreeLow Negative byLC/MS (lessthan -50%)Near Cutoff Negative byLC/MS(Between -50%and the Cutoff)Near Cutoff Positive byLC/MS(Between thecutoff and+50%)High Positive byLC/MS(greater than+50%)
Operator APositive0011620
Negative10141540

{35}------------------------------------------------

Operator BPositive0041720
Negative10141230
Operator CPositive0021820
Negative10141420

Discordant Results for AMP 1000:

OperatorSample NumberLC/MS Result (ng/mL)HealstoneAccurate™ Result
Operator ATN-AMP06043955.5+
Operator BTN-AMP06043955.5+
Operator BTN-AMP06077960.5+
Operator BTN-AMP06007986.2+
Operator BTN-AMP06008996.1+
Operator CTN-AMP06059948.8+
Operator CTN-AMP06008996.1+
Operator ATN-AMP060481087.4-
Operator ATN-AMP060381097.7-
Operator ATN-AMP060031147.4-
Operator ATN-AMP060781147.4-
Operator BTN-AMP060481087.4-
Operator BTN-AMP060381097.7-
Operator BTN-AMP060781147.4-
Operator CTN-AMP060031147.4-
Operator CTN-AMP060581178.0-

COC 300

HealstoneAccurate™Drug-FreeLowNegative byLC/MS (lessthan -50%)Near CutoffNegative byLC/MS(Between -50%and the Cutoff)Near CutoffPositive byLC/MS(Between thecutoff and+50%)High Positiveby LC/MS(greater than+50%)
Operator APositive0021722
Negative10131510
Operator BPositive0031722
Negative10131410
Operator CPositive0021622
Negative10131520

Discordant Results for COC 300:

OperatorSample NumberLC/MS ResultHealstoneAccurate™ Result
----------------------------------------------------------------------

{36}------------------------------------------------

Operator ATN-COC06012278.7+
Operator ATN-COC06059287.7+
Operator BTN-COC06035269.9+
Operator BTN-COC06038287.7+
Operator BTN-COC06061296.1+
Operator CTN-COC06038287.7+
Operator CTN-COC06059287.7+
Operator ATN-COC06011313.9-
Operator BTN-COC06052304.4-
Operator CTN-COC06040336.6-
Operator CTN-COC06026337.1-

MET 1000

HealstoneAccurate™Drug-FreeLowNegative byLC/MS (lessthan -50%)Near CutoffNegative byLC/MS(Between -50%and the Cutoff)Near CutoffPositive byLC/MS(Between thecutoff and+50%)High Positiveby LC/MS(greater than+50%)
Operator A Positive0031326
Operator A Negative10141310
Operator B Positive0021226
Operator B Negative10141420
Operator C Positive0021226
Operator C Negative10141420

Discordant Results for MET 1000:

OperatorSample NumberLC/MS ResultHealstoneAccurateTM Result
Operator ATN-MET06004932.7+
Operator ATN-MET06038935.0+
Operator ATN-MET06061950.8+
Operator BTN-MET06060924.1+
Operator BTN-MET06061950.8+
Operator CTN-MET06038935.0+
Operator CTN-MET06061950.8+
Operator ATN-MET060541001.0-
Operator BTN-MET060541001.0-
Operator BTN-MET060261045.7-
Operator CTN-MET060541001.0-
Operator CTN-MET060261045.7-

{37}------------------------------------------------

OPI 2000

HealstoneAccurate™Drug-FreeLowNegative byLC/MS (lessthan -50%)Near CutoffNegative byLC/MS(Between -50%and the Cutoff)Near CutoffPositive byLC/MS(Between thecutoff and +50%)High Positiveby LC/MS(greater than+50%)
Operator APositive0001917
Negative10161440
Operator BPositive0002017
Negative10161430
Operator CPositive0011917
Negative10161340

Discordant Results for OPI 2000:

OperatorSample NumberLC/MS ResultHealstoneAccurateTM Result
Operator CTN-SMOP060181727.0+
Operator ATN-SMOP060132218.9-
Operator ATN-SMOP060782244.7-
Operator ATN-SMOP060542309.0-
Operator ATN-SMOP060192340.1-
Operator BTN-SMOP060432064.9-
Operator BTN-SMOP060082094.7-
Operator BTN-SMOP060402226.2-
Operator CTN-SMOP060432064.9-
Operator CTN-SMOP060082094.7-
Operator CTN-SMOP060132218.9-
Operator CTN-SMOP060712276.4-

Lay-user study:

A lay user study was performed using urine samples prepared at the following concentrations; -100%, +/-75%, +/-50%, +/-25% of the cutoff by spiking drug(s) into drug free-pooled urine specimens. The concentrations of the samples were confirmed by LC/MS. Each sample was aliquoted into individual containers and blind-labeled. A total of 280 participants with diverse educational and professional backgrounds aged 20 years and older were recruited from three sites. Eighty-one males and 59 females tested AccurateTM Drug of Abuse Urine Test Cup Configuration 1 (including AMP 1000, MET 1000, MOP 300, COC 300); 90 male and 50 females tested AccurateTM Drug of Abuse Urine Test Cup Configuration 2 (including AMP 500, MET 500, OPI 2000, COC 150). Each participant was provided one package insert, one blind labeled test solution, and one test device. The results are summarized below:

{38}------------------------------------------------

Lay-User Study Results for Accurate™ Drug of Abuse Urine Test Cup Configuration 1 (including AMP 1000, MET 1000, MOP 300, COC 300):

DrugCutoff(ng/mL)ResultsConcentration
-100%cutoff-75%cutoff-50%cutoff-25%cutoff+25%cutoff+50%cutoff+75%cutoff
Negative20202017200
Positive0003182020
BZO300Total20202020202020
Agreement (%)100%100%100%85%90%100%100%
Negative20202018200
1000Positive0002182020
TCATotal20202020202020
Agreement (%)100%100%100%90%90%100%100%
Negative20202018200
300Positive0002182020
BARTotal20202020202020
Agreement (%)100%100%100%90%90%100%100%
Negative20202018200
Positive0002182020
MET1000Total20202020202020
Agreement (%)100%100%100%90%90%100%100%
Negative20202018200
Positive0002182020
COC300Total20202020202020
Agreement (%)100%100%100%90%90%100%100%
Negative20202017200
50Positive0003182020
THCTotal20202020202020
Agreement (%)100%100%100%85%90%100%100%
Negative20202018200
1000Positive0002182020
AMPTotal20202020202020
Agreement (%)100%100%100%90%90%100%100%
Negative20202018100
300Positive0002192020
PPXTotal20202020202020
Agreement (%)100%100%100%90%95%100%100%
Negative20202017200
BUP10Positive0003182020
Total20202020202020
Agreement (%)100%100%100%85%90%100%100%
OXY100Negative20202017200
Positive0003182020
Total20202020202020
Agreement (%)100%100%100%85%90%100%100%
MOP300Negative20202019200
Positive0001182020
Total20202020202020
Agreement (%)100%100%100%95%90%100%100%
PCP25Negative20202018300
Positive0002172020
Total20202020202020
Agreement (%)100%100%100%95%90%100%100%
MTD300Negative20202018200
Positive0002182020
Total20202020202020
Agreement (%)100%100%100%90%85%100%100%
MDMA500Negative20202018300
Positive0002172020
Total20202020202020
Agreement (%)100%100%100%90%90%100%100%
EDDP300Negative20202018000
Positive0002202020
Total20202020202020
Agreement (%)100%100%100%90%85%100%100%

{39}------------------------------------------------

Lay-User Study Results for Accurate™ Drug of Abuse Urine Test Cup Configuration 2 (AMP 500, MET 500, OPI 2000, COC 1500):

DrugCutoff(ng/mL)ResultsConcentration
- 100%cutoff-75%cutoff-50%cutoff-25%cutoff+25%cutoff+50%cutoff+75%cutoff
BZO300Negative20202018300
Positive0002172020
Total20202020202020
Agreement (%)100%100%100%90%85%100%100%
TCA1000Negative20202017300
Positive0003172020
Total20202020202020
Agreement (%)100%100%100%85%85%100%100%
BAR300Negative20202019200
Positive0001182020
Total20202020202020
Agreement (%)100%100%100%95%90%100%100%
MET500Negative20202018200
Positive0002182020
Total20202020202020
Agreement (%)100%100%100%90%90%100%100%
COC150Negative20202018200
Positive0002182020
Total20202020202020
Agreement (%)100%100%100%90%90%100%100%
THC50Negative20202018200
Positive0002182020
Total20202020202020
Agreement (%)100%100%100%90%90%100%100%
AMP500Negative20202018200
Positive0002182020
Total20202020202020
Agreement (%)100%100%100%90%90%100%100%
PPX300Negative20202018200
Positive0002182020
Total20202020202020
Agreement (%)100%100%100%90%90%100%100%
BUP10Negative20202018200
Positive0002182020
Total20202020202020
Agreement (%)100%100%100%90%90%100%100%
OXY100Negative20202017200
Positive0003182020
Total20202020202020
Agreement (%)100%100%100%85%90%100%100%
OPI2000Negative20202017200
Positive0003182020
Total20202020202020
Agreement (%)100%100%100%85%90%100%100%
PCP25Negative20202017200
Positive0003182020
Total20202020202020
Agreement (%)100%100%100%85%90%100%100%
MTD300Negative20202018200
Positive0002182020
Total20202020202020
Agreement (%)100%100%100%90%90%100%100%
Total20202020202020
Agreement (%)100%100%100%90%90%100%100%
MDMA500Negative20202017200
Positive0003182020
Total20202020202020
Agreement (%)100%100%100%85%90%100%100%
EDDP300Negative20202017200
Positive0003182020
Total20202020202020
Agreement (%)100%100%100%85%90%100%100%

{40}------------------------------------------------

{41}------------------------------------------------

Participants were given surveys on the ease of understanding the instruction for use. All participants indicated that the device instruction is easy to understand and follow. A Flesch-Kincaid reading analysis was performed on each package insert and the scores revealed a reading Grade Level of 7.

12. Conclusion

Based on the test principle and performance characteristics of the device including precision, cut-off, interference, specificity, method comparison and lay-user studies of the devices, it's concluded that Healstone Accurate™ Drug of Abuse Urine Test Cup and Healstone Accurate™ Drug of Abuse Urine Test Cup Rx are substantially equivalent to the predicate devices.

§ 862.3100 Amphetamine test system.

(a)
Identification. An amphetamine test system is a device intended to measure amphetamine, a central nervous system stimulating drug, in plasma and urine. Measurements obtained by this device are used in the diagnosis and treatment of amphetamine use or overdose and in monitoring levels of amphetamine to ensure appropriate therapy.(b)
Classification. Class II (special controls). An amphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).