(307 days)
CarnaLife® Holo is a software device for display of medical images and other healthcare data. It includes functions for image review, image manipulation, basic measurements, annotations, and 3D visualization.
Lossy compressed mammography images and digitized film screen images must not be reviewed for primary image interpretations. Mammographic images may only be interpreted using an FDA-cleared display that meets technical specifications reviewed and accepted by FDA or displays accepted by the appropriate regulatory agency for the country in which it is used.
Display monitors used for reading medical images for diagnostic purposes must comply with the applicable regulatory approvals and quality control requirements for their use and maintenance.
CarnaLife® Holo software is indicated for use by qualified healthcare professionals including, but not restricted to radiologists, non-radiology specialists, physicians, and technicians.
When accessing CarnaLife® Holo software from a wireless stereoscopic head-mounted display (HMD), images viewed are for informational purposes only and not intended for diagnostic use.
CarnaLife® Holo is a Software as a Medical Device (SaMD) to be used by qualified healthcare professionals including, but not restricted to radiologists, non-radiology specialists, physicians, and technicians for the visualization of medical images in 3D to allow for surgical planning activities. The device makes it possible to visualize 3D Digital Imaging and Communications in Medicine (DICOM) files using volume and surface rendering in 3D to be viewed on a classic display or on a stereoscopic, holographic display. CarnaLife® Holo is intended to be used in operating rooms; office environments within hospitals or at any other location with a computer; and for informational purposes only at any location using the head-mounted display (HMD). The images can be viewed on desktop PCs with a monitor allowing the interaction with CarnaLife® Holo to be performed with a mouse and/or keyboard; and on an HMD, where the interaction with CarnaLife® Holo is performed using hand gestures, voice commands and virtual menus.
CarnaLife® Holo is used to:
- Load patient DICOM data; ●
- . Image review, image manipulation, basic measurements, annotations, and 3D visualization; and
- . View DICOM data using a traditional computer monitor, display monitors used for reading medical images for diagnostic purposes, or in mixed reality (MR) using a head-mounted display (HMD).
The provided document is a 510(k) summary for the CarnaLife® Holo device. It outlines the device description, indications for use, comparison to a predicate device (SurgicalAR K190764), and performance data. However, it does not provide detailed acceptance criteria or extensive study information typically found in a clinical study report for proving performance.
The document focuses on demonstrating substantial equivalence to a predicate device, which is a different regulatory pathway than proving de novo clinical effectiveness. The "Performance Data" section describes design verification and validation activities, but these are primarily focused on software functionality and accuracy rather than clinical performance based on a substantial patient cohort.
Here's an attempt to extract and infer the requested information based on the provided text, heavily noting what is not present:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not present acceptance criteria in a formal, quantifiable table with corresponding performance metrics like sensitivity, specificity, or reader agreement for diagnostic tasks. Instead, it describes functional accuracy tests.
| Acceptance Criteria (Inferred from testing description) | Reported Device Performance |
|---|---|
| Accuracy of Measuring Function: | "Measurement accuracy met specified requirements" for: |
| * X - distance of P1-P2 and P3-P4 | * X - distance of P1-P2 and P3-P4 |
| * Y - distance of P1-P3 and P2-P4 | * Y - distance of P1-P3 and P2-P4 |
| * Angles defined by points P1-P2-P3 | * Angles defined by points P1-P2-P3 |
| * Diagonals defined by P1-P4 and P3-P2 | * Diagonals defined by P1-P4 and P3-P2 |
| Accuracy of Volume Rendering: | "The accuracy of Volume Rendering meeting specified requirements was demonstrated by comparing the size and dimensions of resulting displayed holograms to the known size and dimensions of the two geometric objects." (Implicitly, it met the requirements) |
| * Comparison of displayed hologram size/dimensions to known object size | "Design validation was successfully completed, and testing met all predetermined acceptance criteria based on user needs and intended use." |
| General Design Validation: | "The design verification was successful in that the design output specifications satisfactorily met the design input requirements." |
| * Design output specifications met design input requirements |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Testing: The document mentions "two data sets characterized by different voxel sizes (isotropic and anisotropic data)" for the measuring function test, and "CT scans of two different geometric objects of known size and dimensions" for the volume rendering test. This implies a very small, controlled, and synthetic/phantom-based test set, not a clinical patient image dataset.
- Data Provenance: The document does not specify the country of origin of the data or whether it was retrospective or prospective. Given the nature of the tests (geometric objects, voxel sizes), it's highly likely to be controlled lab data, possibly simulated or phantom studies. It is not indicated as a clinical patient dataset.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
- The document describes functional accuracy testing against known geometric properties or defined points, implying a ground truth established by engineering specifications or measurement tools, not human expert consensus from medical professionals.
- Therefore, the number of experts and their qualifications are not applicable in the context of the described performance tests.
4. Adjudication Method for the Test Set
- Given that the ground truth establishment involves direct measurement or comparison to known geometric properties, and not human expert interpretation, an adjudication method like 2+1 or 3+1 is not applicable and is not mentioned.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- No MRMC study was done or reported. The document focuses on the device's functional accuracy (measurements, rendering) and its substantial equivalence to a predicate, not on a direct comparison of human reader performance with and without AI assistance using this specific device.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
- The described "Performance Data" section details tests of the software's accuracy in performing measurements and rendering. This can be considered a form of standalone performance assessment of these specific functionalities. However, it's not a standalone diagnostic accuracy study in the traditional sense (e.g., classifying disease vs. no disease). The device itself is described as an image display and manipulation tool, not an AI diagnostic algorithm.
7. Type of Ground Truth Used
- The ground truth used was based on known geometric properties/dimensions of objects or precise definitions for measurements (e.g., P1, P2, P3, P4 points for distance and angle calculations). It was not expert consensus, pathology, or outcomes data.
8. Sample Size for the Training Set
- The document does not mention a training set or any machine learning/AI model training. The described tests are for the accuracy of features like measurement and 3D rendering, which are typically based on deterministic algorithms, not learned models. Therefore, the concept of a training set is not applicable to the performance data presented.
9. How the Ground Truth for the Training Set Was Established
- Since no training set is mentioned or implied for the performance data provided, this information is not applicable.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
MedApp S.A. % Albert Pacheco Alliance Network Independent Consultant PAREXEL International 8 Federal Street BILLERICA, MASSACHUSETTS 01821
May 1, 2023
Re: K221870
Trade/Device Name: CarnaLife® Holo Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: LLZ Dated: March 21, 2023 Received: March 22, 2023
Dear Albert Pacheco:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica Lamb
Jessica Lamb, Assistant Director Imaging Software Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K221870
Device Name CarnaLife® Holo
Indications for Use (Describe)
CarnaLife® Holo is a software device for display of medical images and other healthcare data. It includes functions for image review, image manipulation, basic measurements, annotations, and 3D visualization.
Lossy compressed mammography images and digitized film screen images must not be reviewed for primary image interpretations. Mammographic images may only be interpreted using an FDA-cleared display that meets technical specifications reviewed and accepted by FDA or displays accepted by the appropriate regulatory agency for the country in which it is used.
Display monitors used for reading medical images for diagnostic purposes must comply with the applicable regulatory approvals and quality control requirements for their use and maintenance.
CarnaLife® Holo software is indicated for use by qualified healthcare professionals including, but not restricted to radiologists, non-radiology specialists, physicians, and technicians.
When accessing CarnaLife® Holo software from a wireless stereoscopic head-mounted display (HMD), images viewed are for informational purposes only and not intended for diagnostic use.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| Prescription Use (Part 21 CFR 801 Subpart D) | ☑ |
| Over-The-Counter Use (21 CFR 801 Subpart C) | ☐ |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary as Required by 21 CFR 807.92(c)
For CarnaLife® Holo
| Date Prepared: | April 26, 2023 |
|---|---|
| Submitter: | MedApp,S.A.Armii Krajowej 2530-150 Kraków, Poland |
| Official Contact: | Al PachecoIndependent ConsultantParexel International2520 Meridian Parkway, Suite 200Durham, NC 27713, USAPhone: +1 760 421 2919Email: albert.pacheco@parexel.com |
| Proprietary Name: | CarnaLife® Holo |
| Classification: | Class IIMedical Image Management and Processing System21 CFR 892.2050Product Code: LLZ |
| Predicate Device: | SurgicalAR (K190764) |
| Reason for Submission: | New Device |
Device Description
CarnaLife® Holo is a Software as a Medical Device (SaMD) to be used by qualified healthcare professionals including, but not restricted to radiologists, non-radiology specialists, physicians, and technicians for the visualization of medical images in 3D to allow for surgical planning activities. The device makes it possible to visualize 3D Digital Imaging and Communications in Medicine (DICOM) files using volume and surface rendering in 3D to be viewed on a classic display or on a stereoscopic, holographic display. CarnaLife® Holo is intended to be used in operating rooms; office environments within hospitals or at any other location with a computer; and for informational purposes only at any location using the head-mounted display (HMD). The images can be viewed on desktop PCs with a monitor allowing the interaction with CarnaLife® Holo to be performed with a mouse and/or keyboard; and on an HMD, where the interaction with CarnaLife® Holo is performed using hand gestures, voice commands and virtual menus.
CarnaLife® Holo is used to:
- Load patient DICOM data; ●
- . Image review, image manipulation, basic measurements, annotations, and 3D visualization; and
- . View DICOM data using a traditional computer monitor, display monitors used for reading medical images for diagnostic purposes, or in mixed reality (MR) using a head-mounted display (HMD).
{4}------------------------------------------------
Indications for Use
CarnaLife® Holo is a software device for display of medical images and other healthcare data. It includes functions for image review, image manipulation, basic measurements, annotations, and 3D visualization.
Lossy compressed mammography images and digitized film screen images must not be reviewed for primary image interpretations. Mammographic images may only be interpreted using an FDA-cleared display that meets technical specifications reviewed and accepted by FDA or displays accepted by the appropriate regulatory agency for the country in which it is used.
Display monitors used for reading medical images for diagnostic purposes must comply with the applicable regulatory approvals and quality control requirements for their use and maintenance.
CarnaLife® Holo software is indicated for use by qualified healthcare professionals including, but not restricted to radiologists, non-radiology specialists, physicians, and technicians.
When accessing CarnaLife® Holo software from a wireless stereoscopic head-mounted display (HMD), images viewed are for informational purposes only and not intended for diagnostic use.
Substantial Equivalence Discussion
CarnaLife® Holo is substantially equivalent to the predicate device, SurgicalAR (K190764), as both devices have the same intended use and similar technological characteristics. While both devices do not have the same software features, the differences do not raise different questions of safety and effectiveness.
| DeviceCharacteristic | Subject Device:CarnaLife® Holo (K221870) | Predicate Device:SurgicalAR (K190764) | Comparison |
|---|---|---|---|
| Classification | Class II | Class II | Same |
| Product Code | LLZ | LLZ | Same |
| Indications forUse | CarnaLife® Holo is a softwaredevice for display of medicalimages and other healthcaredata. It includes functions forimage review imagemanipulation, basicmeasurements, and 3Dvisualization.Lossy compressedmammography images anddigitized film screen imagesmust not be reviewed forprimary image interpretations.Mammographic images mayonly be interpreted using anFDA- cleared display that meetstechnical specificationsreviewed and accepted by FDAor displays accepted by theappropriate regulatory agency | SurgicalAR is a software device fordisplay of medical images andother healthcare data. It includesfunctions for image review imagemanipulation, basicmeasurements, and 3Dvisualization (MPR reconstructionsand 3D volume rendering).Lossy compressed mammographyimages and digitized film screenimages must not be reviewed forprimary image interpretations.Mammographic images may onlybe interpreted using an FDA-cleared display that meetstechnical specifications reviewedand accepted by FDA or displaysaccepted by the appropriateregulatory agency for the countryin which it is used. | Same |
| DeviceCharacteristic | Subject Device:CarnaLife® Holo (K221870) | Predicate Device:SurgicalAR (K190764) | Comparison |
| for the country in which it is used.Display monitors used for reading medical images for diagnostic purposes must comply with the applicable regulatory approvals and quality control requirements for their use and maintenance.CarnaLife® Holo software is indicated for use by qualified healthcare professionals including, but not restricted to radiologists, non-radiology specialists, physicians, and technicians.When accessing CarnaLife® Holo software from a wireless stereoscopic head-mounted display (HMD), images viewed are for informational purposes only and not intended for diagnostic use. | Display monitors used for reading medical images for diagnostic purposes must comply with the applicable regulatory approvals and quality control requirements for their use and maintenance.SurgicalAR software is indicated for use by qualified healthcare professionals including, but not restricted to radiologists, non-radiology specialists, physicians, and technologists.When accessing SurgicalAR software from a wireless stereoscopic head-mounted display (HMD) or mobile device, images viewed are for informational purposes only and not intended for diagnostic use. | ||
| PrescriptionUse or OTCUse | Prescription use | Prescription use | Same |
| Annotation andMeasurementTools | Line Angle Ruler Arrow | Line Angle Ruler Arrow | Same |
| Data TypeSupported | DICOM Non-DICOM | DICOM Non-DICOM | Same |
| Image InputTypesSupported | CT/angio CT MR/4D MR 3DRA ECHO PET USG | CT MR | Different:Thedifferencedoes not alterintended useand issupported bysoftwareverificationandvalidation. |
| Image View/Manipulation | Image Zoom/Zoom Pan Window Level AutoWindow/Presets | Image Zoom Pan Window Level AutoWindow | Different:Thepredicatedevice does |
| DeviceCharacteristic | Subject Device:CarnaLife® Holo (K221870) | Predicate Device:SurgicalAR (K190764) | Comparison |
| Level Reset Scout Lines/Intersection Lines Image Rotate Image Flipor Intersection Image Invert Magnify | Level Reset Scout Lines Image Rotate Image Flip Magnify | not have the"ImageInvert"functionality. | |
| Data Encryption | HTTPS SSL | HTTPS SSL | Same |
| PatientDemographicDisplay | Capable of displaying patientdemographic information | Capable of displaying patientdemographic information | Same |
| Data Security | Stored locally on the workstation(PC) | Stored on server | Same |
| Audit Trail | Audit trail logged | Audit trail logged | Same |
| File Type Used | JPEG for lossy data PNG for lossless data | JPEG for lossy data PNG for lossless data | Same |
| MPR Viewing | This viewing feature enables thedisplay of reformatted CT andMR images into axial, coronal,and sagittal orientations. | This viewing feature enables thedisplay of reformatted CT and MRimages into axial, coronal, andsagittal orientations. | Same |
| 3D VolumeRenderedViewing | This viewing feature enables thedisplay of 3D perspective viewsof CT and MR image sets thathave been transformed intovolumes. It also provides presetsto enable users to alter thevisualization parameters of the3D views to highlight features. | This viewing feature enables thedisplay of 3D perspective views ofCT and MR image sets that havebeen transformed into volumes. Italso provides presets to enableusers to alter the visualizationparameters of the 3D views tohighlight features. | Same |
| CrosshairNavigation andSynchronization | This viewing feature provides afacility to synchronize and scrollthrough multiple views at thesame time. | This viewing feature provides afacility to synchronize and scrollthrough multiple views at the sametime. | Same |
| Ability to closean image byclicking an "X"in the upper-leftportion of theview port | Ability to close an image byclicking an "X" in the examinationtab. | Ability to close an image by clickingan "X" in the upper-left portion ofthe viewport. | Same |
| Support for TIFFiles | Yes | Yes | Same |
| Support forBMP Files | Yes | No | Different:Thedifferencedoes not alterintended useand issupported bysoftwareverification |
| DeviceCharacteristic | Subject Device:CarnaLife® Holo (K221870) | Predicate Device:SurgicalAR (K190764) | Comparison |
| andvalidation. | |||
| Support ofStored ECHOData | Yes | No | Different:Thedifferencedoes not alterintended useand issupported bysoftwareverificationandvalidation. |
| HMD supportforinformationalpurposes only(not fordiagnostic use) | This viewing feature providesaccess of CarnaLife® Holosoftware on consumer, off-the-shelf wireless, Wi-Fi enabled,stereoscopic head-mounteddisplay with minimum of 2GBRAM. | This viewing feature providesaccess of SurgicalAR software onconsumer, off-the-shelf wireless,Wi-Fi enabled, stereoscopic head-mounted display with minimum of2GB RAM. | Same |
| Diagnosticquality medicalimage review | Ability to provide diagnosticquality medical image review formulti-dimensional digital imagesacquired from a variety ofimaging devices. | Ability to provide diagnostic qualitymedical image review for multi-dimensional digital images acquiredfrom a variety of imaging devices. | Same |
{5}------------------------------------------------
{6}------------------------------------------------
{7}------------------------------------------------
Performance Data
Design validation and verification activities were performed for CarnaLife® Holo as a result of the risk analysis assessment and product requirements. The purpose of software design verification activities was to ensure that the design output specifications met the design input requirements. It is concluded that the design verification was successful in that the design output specifications satisfactorily met the design input requirements.
Design validation was successfully completed, and testing met all predetermined acceptance criteria based on user needs and intended use. Human factors activities were completed per ANSI/AAMI/IEC 62366.
Performance testing included:
- . Accuracy of the Measuring function: This testing was conducted using two data sets characterized by different voxel sizes (isotropic and anisotropic data). For each set, points P1, P2, P3 and P4 were defined for both Axial and Coronal projections. Measurement accuracy met specified requirements for the below measurements of distance:
- o X - distance of P1-P2 and P3-P4
- o Y - distance of P1-P3 and P2-P4
- Angles defined by points P1-P2-P3 o
- Diagonals defined by P1-P4 and P3-P2. o
{8}------------------------------------------------
- Accuracy of Volume Rendering: This testing was conducted by performing Volume Rendering on . CT scans of two different geometric objects of known size and dimensions. The accuracy of Volume Rendering meeting specified requirements was demonstrated by comparing the size and dimensions of resulting displayed holograms to the known size and dimensions of the two geometric objects.
Conclusion
The performance data demonstrate the CarnaLife® Holo is as safe, as effective, and performs as well as or better than the predicate device, SurgicalAR (K190764). In conclusion, the subject device, CarnaLife® Holo, is substantially equivalent to the predicate device, SurgicalAR (K190764).
N/A