K Number
K221861

Validate with FDA (Live)

Date Cleared
2023-02-18

(236 days)

Product Code
Regulation Number
876.1500
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Upon intravenous administration and use of an ICG consistent with its approved label, the FloNavi Endoscopic Fluorescence Imaging System is intended to provide real-time endoscopic visible and near-infrared fluorescence imaging. During minimally invasive surgery, the FloNavi Endoscopic Fluorescence Imaging System enables surgeons to perform minimally invasive surgery using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile ducts, using near-infrared imaging.

Upon interstitial administration and use of an ICG consistent with its approved label, the FloNavi Endoscopic Fluorescence Imaging System is used to perform intraoperative fluorescence imaging and visualization of the lymphatic system, including lymphatic vessels and lymph nodes.

Device Description

The proposed system, FloNavi Endoscopic Fluorescence Imaging System (FloNavi System) is comprised of an image processing unit, a camera head, and a light source (including a flexible light guide cable).

There are two models of the proposed system. The primary components of each model are provided in Table 1.

Model 1: Image Processing Unit: OPTO-CAM214K, Camera Head: OPTO-CHD214KE/OPTO-CHD214KH, Light Source: OPTO-LED214K
Model 2: Image Processing Unit: OPTO-CAM2100, Camera Head: OPTO-CHD2100, Light Source: OPTO-LED2100

FloNavi Endoscopic Fluorescence Imaging System is capable of providing real-time endoscopic visible and near-infrared fluorescence imaging.

During surgical procedures, FloNavi may be operated to provide visualization similar to that provided by conventional imaging systems used in surgical endoscopy. The area of interest is illuminated with visible light from the light source and the resulting reflecting light is imaged by the camera and displayed on the video monitor. When used with the VIS-only laparoscopes, the System is only capable of the conventional mode of visualization described herein.

To provide NIR fluorescence imaging, FloNavi is used with the imaging agent, indocyanine green (ICG). The patient is injected with ICG imaging agent. The ICG fluoresces when illuminated through the laparoscope with NIR excitation light from the light source, and the fluorescence response is then imaged with the camera, processed and displayed on a video monitor.

The proposed system is designed to be used with rigid endoscopes, monitors and other ancillary equipment.

AI/ML Overview

The requested information about the acceptance criteria and study proving device performance is not available in the provided text. The document is an FDA 510(k) clearance letter and summary, which confirms substantial equivalence to a predicate device based on non-clinical testing, but does not detail acceptance criteria or specific performance study results of the type requested.

The document explicitly states:
"No clinical tests were performed." and "Performance testing was conducted on imaging performance and light source performance to support the marketing claims and to confirm that safety and effectiveness of the FloNavi Endoscopic Fluorescence Imaging System is at least equivalent to the predicate device."

Therefore, I cannot provide:

  1. A table of acceptance criteria and the reported device performance.
  2. Sample size used for the test set and data provenance.
  3. Number of experts used to establish ground truth and their qualifications.
  4. Adjudication method for the test set.
  5. Effect size of human readers with vs. without AI assistance (as no clinical AI assistance study is mentioned).
  6. Standalone performance (algorithm only) was done. (While non-clinical performance was done, it's not described in terms of specific standalone algorithm performance, but rather overall system performance.)
  7. Type of ground truth used.
  8. Sample size for the training set.
  9. How the ground truth for the training set was established.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

February 18, 2023

Guangdong OptoMedic Technologies, Inc. % Joyce Yang Consultant Shenzhen Joyantech Consulting Co., Ltd. 1713A, 17th Floor, Block A, Zhongguan Times Square Nanshan District Shenzhen, Guangdong 518100 China

Re: K221861

Trade/Device Name: FloNavi Endoscopic Fluorescence Imaging System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: GCJ, IZI Dated: January 13, 2023 Received: January 17, 2023

Dear Joyce Yang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Jianting Wang -S

Jianting Wang Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K221861

Device Name

FloNavi Endoscopic Fluorescence Imaging System

Indications for Use (Describe)

Upon intravenous administration and use of an ICG consistent with its approved label, the FloNavi Endoscopic Fluorescence Imaging System is intended to provide real-time endoscopic visible and near-infrared fluorescence imaging. During minimally invasive surgery, the FloNavi Endoscopic Fluorescence Imaging System enables surgeons to perform minimally invasive surgery using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile ducts, using near-infrared imaging.

Upon interstitial administration and use of an ICG consistent with its approved label, the FloNavi Endoscopic Fluorescence Imaging System is used to perform intraoperative fluorescence imaging and visualization of the lymphatic system, including lymphatic vessels and lymph nodes.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary - K221861

Date of Summary prepare: February 7, 2023

1. Submission Sponsor

Applicant Name:Guangdong OptoMedic Technologies, Inc.
Address:Suite 503, Building A, Golden Valley IntellicreationCommunity, No.2 Yonganbei Street, Daxu,Guicheng, Nanhai, 528200, Foshan, Guangdong,P.R.China
Contact person:Jane Guo
Phone:+86 (757) 8670 2920
Email:guoweijuan@optomedic.com

2. Submission correspondent

Name:Shenzhen Joyantech Consulting Co., Ltd
Address:1713A, 17th Floor, Block A, Zhongguan TimesSquare, Nanshan District, Shenzhen518000
Contact person:Joyce Yang
Phone:+86-755-86069197
Email:joyce@cefda.com

3. Device Identification

Trade Name:FloNavi Endoscopic Fluorescence Imaging System
Common or Usual Name:Endoscopic Imaging System
Model:OPTO-CAM214K, OPTO-CHD214KE, OPTO-CHD214KH, OPTO-LED214K;OPTO-CAM2100, OPTO-CHD2100, OPTO-LED2100;
Classification name:Endoscope and accessories
Review Panel:Gastroenterology/Urology
Product Code:GCJ
Device Class:Class II
Regulation Number:21 CFR § 876.1500

Legally Marketed Predicate Device 4.

Predicate Device:
Trade Name:PINPOINT Endoscopic Fluorescence ImagingSystem
Regulation number:21 CFR § 876.1500
Regulation class:Class II

{4}------------------------------------------------

Regulation name:Endoscope and accessories
510(k) Number:K182606
Product Code:GCJ; IZI
Manufacturer:Novadaq Technologies ULC.
Reference Device:
Trade Name:Image 1 SPIES System
Regulation number:21 CFR § 876.1500
Regulation class:Class II
Regulation name:Endoscope and accessories
510(k) Number:K160044
Product Code:FET
Manufacturer:Karl Storz Endoscopy America, Inc.
Trade Name:Power LED 175
Regulation number:21 CFR § 876.1500
Regulation class:Class II
Regulation name:Endoscope and accessories
510(k) Number:K123956
Product Code:FCW, NTN
Manufacturer:Karl Storz Endoscopy America, Inc.

Device Description 5.

The proposed system, FloNavi Endoscopic Fluorescence Imaging System (FloNavi System) is comprised of an image processing unit, a camera head, and a light source (including a flexible light guide cable).

There are two models of the proposed system. The primary components of each model are provided in Table 1.

FloNavi EndoscopicFluorescenceImaging SystemPrimary component
Model 1Image Processing Unit: OPTO-CAM214KCamera Head: OPTO-CHD214KE/OPTO-CHD214KHLight Source: OPTO-LED214K
Model 2Image Processing Unit: OPTO-CAM2100Camera Head: OPTO-CHD2100Light Source: OPTO-LED2100
Table 1 System Model and primary component
------------------------------------------------------

{5}------------------------------------------------

Except slight differences in appearance, other differences between components of either model are shown in Table 2.

Image Processing Unit
ItemOPTO-CAM214KOPTO-CAM2100
Video outputsDVI, HDMI, 4×3G-SDI, 12G-SDISDI, DVI, CVBS, S-VIDEO
Video output resolution$4096 \times 2160p, 3840 \times 2160p, 1920 \times 1080p, 50/60Hz$$1920 \times 1080p, 720 \times 576i, 50/60Hz$
Size (mm)403370150403370150
Weight (kg)9±15%9±15%
Camera Head
ItemOPTO-CHD214KHOPTO-CHD214KEOPTO-CAM2100
Image sensor4K CMOS sensor assembly4K CMOS sensor assemblyHD CMOS sensor assembly
Focusing mechanismManualMotor-drivenManual
Weight (kg)0.6±15%0.5±15%
Light Source
ItemOPTO-LED214KOPTO-LED2100
Size (mm)403370150403370150
Weight (kg)11±15%11±15%
Light guide cableDiameter: 4.5mmDiameter: 4.8 mm

FloNavi Endoscopic Fluorescence Imaging System is capable of providing real-time endoscopic visible and near-infrared fluorescence imaging.

During surgical procedures, FloNavi may be operated to provide visualization similar to that provided by conventional imaging systems used in surgical endoscopy. The area of interest is illuminated with visible light from the light source and the resulting reflecting light is imaged by the camera and displayed on the video monitor. When used with the VIS-only laparoscopes, the System is only capable of the conventional mode of visualization described herein.

{6}------------------------------------------------

To provide NIR fluorescence imaging, FloNavi is used with the imaging agent, indocyanine green (ICG). The patient is injected with ICG imaging agent. The ICG fluoresces when illuminated through the laparoscope with NIR excitation light from the light source, and the fluorescence response is then imaged with the camera, processed and displayed on a video monitor.

The proposed system is designed to be used with rigid endoscopes, monitors and other ancillary equipment. The compatible rigid endoscope is O-Mec laparoscopes 690 Series (Model: 690-331030H), which was cleared under K201151.

Intended Use/ Indications for Use

Upon intravenous administration and use of an ICG consistent with its approved label, the FloNavi Endoscopic Fluorescence Imaging System is intended to provide realtime endoscopic visible and near-infrared fluorescence imaging. During minimally invasive surgery, the FloNavi Endoscopic Fluorescence Imaging System enables surgeons to perform minimally invasive surgery using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile ducts, using near-infrared imaging.

Upon interstitial administration and use of an ICG consistent with its approved label, the FloNavi Endoscopic Fluorescence Imaging System is used to perform intraoperative fluorescence imaging and visualization of the lymphatic system, including lymphatic vessels and lymph nodes.

Comparison itemSubject Device (K221861)Predicate Device(K182606)
Product NameFloNaviFluorescenceImaging SystemPINPOINTFluorescence Imaging System
Product CodeGCJGCJ, IZI
Regulation Number21 CFR § 876.150021 CFR § 876.1500
ClassificationClass IIClass II
Type of usePrescription UsePrescription Use
Comparison itemSubject Device (K221861)Predicate Device(K182606)
Intended use &Indication s for UseUpon intravenousadministration and use ofan ICG consistent with itsapproved label, the FloNaviEndoscopic FluorescenceImaging System is intendedto provide real-timeendoscopic visible andnear-infrared fluorescenceimaging. During minimallyinvasive surgery, theFloNavi EndoscopicFluorescence ImagingSystem enables surgeonsto perform minimallyinvasive surgery usingstandard endoscopic visiblelight as well as visualassessment of vessels,blood flow and relatedtissue perfusion, and atleast one of the major extra-hepatic bile ducts, usingnear-infrared imaging.Upon interstitialadministration and use ofan ICG consistent with itsapproved label, the FloNaviEndoscopic FluorescenceImaging System is used toperform intraoperativefluorescence imaging andvisualization of thelymphatic system, includinglymphatic vessels andlymph nodes.Upon intravenous administration ofTRADENAME (ICG drug product),the PINPOINT EndoscopicFluorescence Imaging System isused with TRADENAME to performintraoperative fluorescenceangiography, and it is also indicatedfor use in fluorescence imaging ofbiliary ducts, and when indicated,during intraoperativecholangiography.The PINPOINT EndoscopicFluorescence Imaging System isindicated for use to provide real timeendoscopic visible and near-infraredfluorescence imaging. ThePINPOINT System enablessurgeons to perform minimallyinvasive surgery using standardendoscope visible light as well asvisual assessment of vessels, bloodflow and related tissue perfusion,and at least one of the major extra-hepatic bile ducts (cystic duct,common bile duct or commonhepatic duct), using near infraredimaging.Fluorescence imaging of biliaryducts with the PINPOINT System isintended for use with standardof care white light, and whenindicated, intraoperativecholangiography. The device is notintended for standalone use forbiliary duct visualization.Upon interstitial administration ofTRADENAME (ICG drug product),the PINPOINT System isused to perform intraoperativefluorescence imaging andvisualization of the lymphaticsystem,including lymphatic vessels andlymph nodes.
Applicable userPhysiciansPhysicians
Environment of useHealthcare facility/hospitalHealthcare facility /hospital
Single use / ReusableReusableReusable
Sterile /non-sterileMarketed as non-sterileMarketed as non-sterile
Device Systemcomponents- Image Processing Unit- Camera Head- Light Source- Endoscopic video processor /illuminator (VPI)- Laparoscope
Comparison itemSubject Device (K221861)Predicate Device(K182606)
- Light Guide Cable- Camera head
- Light guide cable
Video output signalsOPTO-CAM214K:DVI, HDMI, 4×3G-SDI,12G-SDIHD-SDI, DVI
Video output signalsOPTO-CAM2100:SDI, DVI, CVBS, S-VIDEO
Video output resolutionOPTO-CAM214K:4096×2160p; 3840×2160p;1920×1080p1920×1080
Video output resolutionOPTO-CAM2100:1920×1080p; 720×576i
Voltage110-240V~100-240V~
Frequency50/60 Hz50/60 Hz
Power consumptionImage Processing Unit:150VALight Source: 200VA300VA
Image sensorsOPTO-CHD214KE/OPTO-CHD214KH :4K CMOS sensor assemblyCMOS HD sensor assembly
Image sensorsOPTO-CHD2100:HD CMOS sensor assembly
Aspect ratio16:916:9
Light sources- Visible (VIS): Light-emitting diode array- Near infrared (NIR): NIRlaser diode- Visible (VIS): Light-emitting diodearray- Near infrared (NIR): NIR laserdiode
Light guidecableTransmissionspectrumVisible + NIRVisible + NIR
FiberdiameterOPTO-LED214K: 4.5mm;OPTO-LED2100: 4.8 mm4.9 mm
Length3 m/3.5m3 m
SterilizationAutoclaveAutoclave
Type ofprotectionagainst electric shock(as per IEC 60601-1)Class IClass I
Comparison itemSubject Device (K221861)Predicate Device(K182606)
Degree of protectionagainst electric shocks(as per IEC 60601-1)CF-typeCF-type
Laser classification(as per IEC 60825-1)Class 3RClass 3R
Radiofrequencyemissions(as per CISPR 11)Group 1, Class AGroup 1, Class A

Table 3 Technological characteristics comparison

6. Technological characteristics comparison

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

Summary of Technological Characteristics:

As shown in the comparison table, the proposed system and its predicate has the same intended use and indications for use and similar technological characteristics.

7. Summary of non-clinical testing

The non-clinical test results demonstrated that the proposed device complies with the following standards:

  • AAMI/ANSI ES 60601-1:2005/(R)2012 Medical Electrical Equipment- Part 1: . General requirements for basic safety and essential performance.
  • 60601-1-2:2014 Medical electrical equipment- Part 1-2: General IEC . requirements for basic safety and essential performance- Collateral standard: Electromagnetic compatibility- Requirements and tests.
  • IEC 60601-2-18:2009 Medical electrical equipment Part 2-18: Particular . requirements for the basic safety and essential performance of endoscopic equipment
  • IEC 60825:2014 Safety of laser products Part 1: Equipment classification and . requirements was assessed and showed that subject device is a Class 3R laser device.
  • IEC 62471:2006 Photobiological safety of lamps and lamp systems .

Performance testing was conducted on imaging performance and light source performance to support the marketing claims and to confirm that safety and effectiveness of the FloNavi Endoscopic Fluorescence Imaging System is at least equivalent to the predicate device.

8. Brief discussion of clinical tests

No clinical tests were performed.

9. Conclusions

The subject device and the predicate device have the same intended use, similar technological characteristics. The technological differences will not cause safety and effectiveness problems for the subject device as compared to its predicate device. Performance tests demonstrate that the FloNavi Endoscopic Fluorescence Imaging

{10}------------------------------------------------

System performs according to specifications and functions as intended. Therefore,
the proposed system is substantially equivalent to its predicate device.

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.