K Number
K221765

Validate with FDA (Live)

Manufacturer
Date Cleared
2022-12-23

(189 days)

Product Code
Regulation Number
862.3170
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Benzodiazepines II (BNZ2) is an in vitro diagnostic test for the qualitative and semiquantitative detection of benzodiazepines in human urine on cobas c systems at cutoff concentrations of 100 ng/mL, and 300 ng/mL. Semiquantitative test results may be obtained that permit laboratories to assess assay performance as part of a quality control program.

Semiquantitative assays are intended to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as gas chromatography/mass spectrometry (GC-MS), or Liquid Chromatography coupled with Tandem Mass Spectrometry (LC-MS/MS).

Benzodiazepines II provides only a preliminary analytical test result. A more specific alternical method must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC-MS) or Liquid Chromatography coupled with Tandem Mass Spectrometry (LC-MS/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.

Device Description

The Benzodiazepines II assay is an in vitro diagnostic test for the qualitative and semi-quantitative detection of benzodiazepines in human urine on automated clinical chemistry analyzers at cutoff concentrations of 100 ng/mL and 300 ng/mL. The semi quantitative test results may be obtained that permit laboratories to assess assay performance as part of a quality control program.

The device consists of two wet reagents which contain the key components of the immunoassay: monoclonal/ polyclonal antibody against the drug, substrate, and enzyme-labeled drug (conjugate).

AI/ML Overview

Here's an analysis of the acceptance criteria and the study proving the device meets them, based on the provided text:

Acceptance Criteria and Device Performance for ONLINE DAT Benzodiazepines II

1. Table of Acceptance Criteria and Reported Device Performance

The direct "acceptance criteria" for performance are not explicitly listed in a single table, but rather demonstrated through various performance studies. The table below summarizes the key performance aspects tested and the results obtained for the ONLINE DAT Benzodiazepines II.

Device Name: ONLINE DAT Benzodiazepines II
Intended Use: Qualitative and semiquantitative detection of benzodiazepines in human urine on cobas c systems at cutoff concentrations of 100 ng/mL, 200 ng/mL, and 300 ng/mL.

Performance CharacteristicAcceptance Criteria (Implied/Demonstrated)Reported Device Performance (ONLINE DAT Benzodiazepines II)
Precision (Repeatability & Intermediate)Low CV% for semiquantitative results; >95% agreement for qualitative results around specified cutoffs.100 ng/mL Cutoff (SQ): Repeatability CV: 1.8-5.7%; Intermediate Precision CV: 2.0-6.2%. 100 ng/mL Cutoff (Qualitative): >95% negative reading for -50%, -25% samples; >95% positive reading for +25%, +50% samples. 200 ng/mL Cutoff (SQ): Repeatability CV: 0.8-2.5%; Intermediate Precision CV: 1.5-3.1%. 200 ng/mL Cutoff (Qualitative): >95% negative reading for -50%, -25% samples; >95% positive reading for +25%, +50% samples. 300 ng/mL Cutoff (SQ): Repeatability CV: 1.2-2.9%; Intermediate Precision CV: 1.6-3.5%. 300 ng/mL Cutoff (Qualitative): >95% negative reading for -50%, -25% samples; >95% positive reading for +25%, +50% samples.
Analytical Recovery and LinearityAcceptable percent recovery across the assay range, demonstrating linearity.100 ng/mL Cutoff: Recovery generally within 94-116% across various concentrations (0-1000 ng/mL). 200 ng/mL Cutoff: Recovery generally within 95-119% across various concentrations (0-1000 ng/mL). 300 ng/mL Cutoff: Recovery generally within 92-117% across various concentrations (0-3000 ng/mL).
Analytical Specificity/Cross-ReactivityAppropriate cross-reactivity with common benzodiazepines and metabolites, and minimal cross-reactivity with structurally unrelated compounds.Cross-reactivity with Benzodiazepines and Metabolites: Varies (e.g., Nordiazepam: 98-101%, Alprazolam: 109-115%, Oxazepam: 104-113%, Lorazepam Glucuronide: 56-58%, 7-Acetamidonitrazepam: 0.3-0.5%). No Interference (Structurally Unrelated Compounds): None of the tested compounds (e.g., Acetone, Ascorbic Acid, Creatinine, Ethanol, Glucose, Hemoglobin, Ibuprofen, Morphine, Nicotine, etc.) at high concentrations caused positive or negative interference.
Endogenous InterferencesNo interference from common endogenous compounds at specified concentrations.None of the listed endogenous compounds (e.g., Acetone, Ascorbic Acid, Calcium, Creatinine, Ethanol, Glucose, Hemoglobin, Human Albumin, Urea, Uric Acid) caused interference at the tested concentrations.
Interference DrugsNo interference from common drugs at specified concentrations, with noted exceptions.Most tested drugs (e.g., Acetaminophen, Aspirin, Amitryptyline, Caffeine, Codeine, Ibuprofen, Morphine, Nicotine) caused no interference at very high concentrations. Exception: Oxaprozin may yield falsely elevated results when present with benzodiazepine at negative control levels due to cross-reactivity (13% at 100 ng/mL cutoff, 6% at 200 ng/mL cutoff, 10% at 300 ng/mL cutoff).
Specific Gravity and pH InterferenceProper recovery of results across specified pH and specific gravity ranges.Urine samples with pH 4.0-9.0 and specific gravities 1.001-1.034 resulted in proper recovery in both semi-quantitative and qualitative modes.
Method Comparison to Predicate (Clinical Correlation)High agreement with LC-MS/MS, especially for negative and confirmed positive samples; acceptable performance near cutoff.100 ng/mL Cutoff: 100% negative agreement with normal urines. 100% positive agreement with confirmed positive samples. Overall good agreement with LC-MS/MS with few discrepancies near cutoff. 200 ng/mL Cutoff: 100% negative agreement with normal urines. 100% positive agreement with confirmed positive samples. Overall good agreement with LC-MS/MS with few discrepancies near cutoff. 300 ng/mL Cutoff: 100% negative agreement with normal urines. 100% positive agreement with confirmed positive samples. Overall good agreement with LC-MS/MS with few discrepancies near cutoff.
Stability (On-board Stability)Stable performance (agreement, mean, SD, CV) over 12 weeks of on-board use after transport stress.100/200/300 ng/mL Cutoffs (SQ & Qualitative): 100% agreement for negative and positive controls and clinical samples for 91 days (13 weeks) onboard after transport stress. Mean, SD, and CV values remained consistent and within acceptable ranges over the 91-day period.

Study Details:

  1. Sample Size used for the test set and the data provenance:

    • Precision Study: 84 determinants (n=84) for each cutoff concentration (100, 200, 300 ng/mL) in both qualitative and semi-quantitative modes. These samples were drug-free negative urine spiked with Oxazepam. Data provenance is internal to Roche Diagnostics, as samples were prepared for the study.
    • Analytical Recovery and Linearity Study: 17 levels per cutoff, run in triplicate, for three reagent lots. Samples were "prepared" (implied internal spiking/dilution).
    • Analytical Specificity/Cross-Reactivity Study: Concentration series prepared for each cross-reactant by spiking drug-free urine. Data provenance is internal.
    • Endogenous Interferences: Two sets of samples per interferent, containing benzodiazepine and/or glucuronidated benzodiazepine. Data provenance is internal.
    • Interference Drugs: Samples spiked with benzodiazepine in the presence of potentially interfering drugs. Data provenance is internal.
    • Specific Gravity and pH Interference: Urine samples prepared with varying pH and specific gravities, spiked with benzodiazepine. Data provenance is internal.
    • Method Comparison (Clinical Correlation) Study:
      • 100 ng/mL Cutoff: 137 native human unaltered samples in total (48 screened negative, 54 screened preliminary positive/confirmed, 8 near cutoff negative by LC-MS/MS, 7 near cutoff positive by LC-MS/MS).
      • 200 ng/mL Cutoff: 134 native human unaltered samples in total (56 screened negative, 57 screened preliminary positive/confirmed, 12 near cutoff negative by LC-MS/MS, 9 near cutoff positive by LC-MS/MS).
      • 300 ng/mL Cutoff: 114 native human unaltered samples in total (40 screened negative, 45 screened preliminary positive/confirmed, 17 near cutoff negative by LC-MS/MS, 12 near cutoff positive by LC-MS/MS).
      • Data Provenance: Samples were "purchased from clinical laboratories" and "obtained from a clinical laboratory," implying a retrospective collection of clinical urine samples from the United States or a jurisdiction with similar clinical laboratory practices.
    • Stability Study: Two control samples per cutoff, the cutoff calibrator, and two clinical samples (pooled urine from individual donors). Data provenance is internal.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • No human experts were explicitly used to establish a ground truth in the sense of clinical interpretation for this in vitro diagnostic device study.
    • The "ground truth" for the method comparison study was established by a confirmatory analytical method: Liquid Chromatography coupled with Tandem Mass Spectrometry (LC-MS/MS), which is considered the gold standard for drug confirmation. This is an objective analytical measurement, not an expert opinion.
    • For the precision, linearity, and interference studies, the ground truth was based on precisely prepared spiked samples with known concentrations.
  3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • There was no adjudication method described for the test set in the conventional sense of expert review for image-based diagnostics.
    • The method comparison data used LC-MS/MS as the reference method. Discrepant samples in the method comparison were cross-referenced against the LC-MS/MS values, which served as the definitive reference.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No MRMC comparative effectiveness study was performed or described. This device is an in vitro diagnostic (IVD) immunoassay for detecting substances in urine. It does not involve human readers interpreting images with or without AI assistance. Therefore, this question is not applicable.
  5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Yes, the primary performance evaluation is a standalone (algorithm only) performance. The device itself is an automated chemical analyzer (cobas c systems) that performs the immunoassay and generates a result (qualitative or semi-quantitative concentration). There is no "human-in-the-loop" performance component for the direct operation or result generation of the device. The results are then used by clinical laboratories.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For the method comparison study, the primary ground truth was Liquid Chromatography coupled with Tandem Mass Spectrometry (LC-MS/MS), which is an objective chemical confirmatory method.
    • For other analytical performance studies (precision, linearity, cross-reactivity, interferences, stability), the ground truth was based on known concentrations in spiked samples prepared in the laboratory.
  7. The sample size for the training set:

    • The document describes performance evaluation studies (validation studies) for a cleared device. It does not explicitly mention a "training set" or its size, as such devices are typically developed and optimized by the manufacturer using internal data and then validated through these types of pre-market performance studies. The presented data represents the validation of the final product.
  8. How the ground truth for the training set was established:

    • As no "training set" is explicitly mentioned in the context of device development or machine learning in this submission, the method for establishing its ground truth is not provided. For an IVD like this, ground truth during development would typically involve similarly prepared spiked samples with known concentrations and/or analysis by definitive reference methods like LC-MS/MS.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.

December 23, 2022

Roche Diagnostics Khoa Tran Regulatory Affairs Program Manager 9115 Hague Road Indianapolis, IN 46250

Re: K221765

Trade/Device Name: ONLINE DAT Benzodiazepines II Regulation Number: 21 CFR 862.3170 Regulation Name: Benzodiazepine test system Regulatory Class: Class II Product Code: JXM Dated: June 15, 2022 Received: June 17, 2022

Dear Khoa Tran:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Digitally signed by Paula_________________________________________________________________________________________________________________________________________________________________________ Paula Caposino -S Caposino -S Date: 2022.12.23

Paula Caposino, Ph.D. Acting Deputy Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K221765

Device Name ONLINE DAT Benzodiazepines II

Indications for Use (Describe)

Benzodiazepines II (BNZ2) is an in vitro diagnostic test for the qualitative and semiquantitative detection of benzodiazepines in human urine on cobas c systems at cutoff concentrations of 100 ng/mL, and 300 ng/mL. Semiquantitative test results may be obtained that permit laboratories to assess assay performance as part of a quality control program.

Semiquantitative assays are intended to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as gas chromatography/mass spectrometry (GC-MS), or Liquid Chromatography coupled with Tandem Mass Spectrometry (LC-MS/MS).

Benzodiazepines II provides only a preliminary analytical test result. A more specific alternical method must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC-MS) or Liquid Chromatography coupled with Tandem Mass Spectrometry (LC-MS/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

ONLINE DAT Benzodiazepines II K221765 - 510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.

In accordance with 21 CFR 807.87. Roche Diagnostics hereby submits official notification as required by Section 510(k) of the Federal Food, Drug and Cosmetics Act of our intention to market the device described in this Premarket Notification 510(k).

The purpose of this Traditional 510(k) Premarket Notification is to obtain FDA review and clearance for the ONLINE DAT Benzodiazepines II

{4}------------------------------------------------

Submitter NameRoche Diagnostics
Address9115 Hague RoadP.O. Box 50416Indianapolis, IN 46250-0457
ContactKhoa TranPhone: (317) 946-7843Email: khoa.tran@roche.comSecondary Contact Name: Leslie PattersonPhone: (317) 225-8563Email: leslie.patterson@roche.com
Date PreparedJune 6, 2022
Proprietary NameONLINE DAT Benzodiazepines II
Common NameBenzodiazepines Enzyme Immunoassay
Classification Name andPanelBenzodiazepine Test System, 91 – Toxicology
Product Codes,Regulation NumbersJXM, Class II, 21 CFR 862. 3170
Predicate DevicesONLINE DAT Benzodiazepines Plus
Establishment RegistrationRoche Diagnostics GmbH Mannheim, Germany: 9610126Roche Diagnostics GmBH Penzberg, Germany: 9610529Roche Diagnostics Indianapolis, IN United States: 1823260.

{5}------------------------------------------------

1. DEVICE DESCRIPTION

The Benzodiazepines II assay is an in vitro diagnostic test for the qualitative and semi-quantitative detection of benzodiazepines in human urine on automated clinical chemistry analyzers at cutoff concentrations of 100 ng/mL and 300 ng/mL. The semi quantitative test results may be obtained that permit laboratories to assess assay performance as part of a quality control program.

1.1. ONLINE DAT Benzodiazepines II

The device consists of two wet reagents which contain the key components of the immunoassay: monoclonal/ polyclonal antibody against the drug, substrate, and enzyme-labeled drug (conjugate).

2. INDICATIONS FOR USE

Benzodiazepines II (BNZ2) is an in vitro diagnostic test for the qualitative and semiquantitative detection of benzodiazepines in human urine on cobas c systems at cutoff concentrations of 100 ng/mL, 200 ng/mL, and 300 ng/mL.

Semiquantitative test results may be obtained that permit laboratories to assess assay performance as part of a quality control program. Semiquantitative assays are intended to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as gas chromatography/mass spectrometry (GC-MS), or Liquid Chromatography coupled with Tandem Mass Spectrometry (LC-MS/MS).

Benzodiazepines II provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC-MS) or Liquid Chromatography coupled with Tandem Mass Spectrometry (LC-MS/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.

{6}------------------------------------------------

TECHNOLOGICAL CHARACTERISTICS 3.

The assay is based on the kinetic interaction of microparticles in a solution (KIMS) as measured by changes in light transmission. In the absence of sample drug, free antibody binds to drugmicroparticle conjugates causing the formation of particle aggregates that are photometrically detected by turbidity measurements. As the aggregation reaction proceeds in the absence of sample drug, the absorbance increases.

When a urine sample contains the drug in question, this drug competes with the particle-bound drug derivative for free antibody. Antibody bound to sample drug is no longer available to promote particle aggregation, and subsequent particle lattice formation is inhibited. The presence of sample drug diminishes the increasing absorbance in proportion to the concentration of drug in the sample. Sample drug content is determined relative to the value obtained for a known cutoff concentration of drug.

The presence of p-glucuronidase enzyme enhances the Benzodiazepines II assay cross-reactivity to some of the glucuronidated metabolites. Enzymatic cleavage makes the benzodiazepine part of the glucuronides more accessible for the antibody.

The assay consists of two liquid reagents. Benzodiazepines antibody buffer and conjugated benzodiazepine derivative microparticles.

The following table compare the ONLINE DAT Benzodiazepines II with its predicate device, ONLINE DAT Benzodiazepines Plus (K043327).

{7}------------------------------------------------

FeatureCandidate DeviceONLINE DAT BenzodiazepinesIIPredicate DeviceONLINE DAT BenzodiazepinesPlus (K043327)
Intended UseBenzodiazepines II (BNZ2) is anin vitro diagnostic test for thequalitative and semiquantitativedetection of benzodiazepines inhuman urine on Roche/Hitachicobas c systems at cutoffconcentrations of 100 ng/mL, 200ng/mL, and 300 ng/mL.Semiquantitative test results maybe obtained that permitlaboratories to assess assayperformance as part of a qualitycontrol program.Benzodiazepines II provides only apreliminary analytical test result. Amore specific alternate chemicalmethod must be used in order toobtain a confirmed analyticalresult. Gas chromatography/massspectrometry (GC-MS) or LiquidChromatography coupled withTandem Mass Spectrometry(LC-MS/MS) is the preferredconfirmatory method. Clinicalconsideration and professionaljudgment should be applied to anydrug of abuse test result,particularly when preliminarypositive results are used.The ONLINE DAT BenzodiazepinesPlus is an in vitro diagnostic test forthe qualitative and semi-quantitativedetection of benzodiazepines inhuman urine on automated clinicalchemistry analyzers at cutoffconcentrations of 100 ng/mL, 200ng/mL, and 300 ng/mL. Semi-quantitative test results may beobtained that permit laboratories toassess assay performance as part of aquality control program.
Detection MethodKIMS, Kinetic interaction ofmicroparticles in a solutionSame
Instrument Platformcobas c 501Mod P
Test MatrixUrineSame
Measured AnalyteBenzodiazepine and itsmetabolitesBenzodiazepine
Cutoff Levels100 ng/mL, 200 ng/mL and 300ng/mLSame
Reagent Stability2-8 °C until expiration dateSame

Technical Characteristics Comparison Table between ONLINE DAT Table 1: Benzodiazepines II and ONLINE DAT Benzodiazepines Plus

{8}------------------------------------------------

NON-CLINICAL PERFORMANCE EVALUATION 4.

The following performance data are provided in support of the substantial equivalence determination:

  • Precision according to CLSI EP5-A3 .
  • Linearity according to CLSI EP6-A .
  • Analytical Specificity/Cross-Reactivity .
  • Endogenous Interferences .
  • Interference Drugs .
  • Interference Testing of Specific Gravity and pH .
  • Method Comparison to Predicate .
  • Stability .

All performance specifications were met.

4.1. Precision

Repeatability and Intermediate Precision 4.1.1.

The precision study was performed using CLSI Guideline EP05-A3 as a guideline. Precision experiments were conducted using one reagent lot on one cobas c 501 instrument. Testing was carried out for 21 days with two runs per day, and two replicates per run in both Qualitative and Semi-quantitative modes, giving a total of 84 determinants (n = 84). Drug-free negative urine was spiked with Oxazepam to final concentrations of -50%, below cutoff and +25%, +50% above cutoff.

100 ng/mL SQRepeatabilityMean (ng/mL)SD (ng/mL)CV (%)
Urine -50%45.02.565.7
Urine -25%68.82.443.6

{9}------------------------------------------------

DAT2N(control 75 ng/mL)78.81.992.5
Cutoff Urine99.72.382.4
Urine +25%1232.432.0
DAT2P(control 125 ng/mL)1272.241.8
Urine +50%1462.591.8
IntermediatePrecisionMean (ng/mL)SD (ng/mL)CV (%)
Urine -50%45.02.796.2
Urine -25%68.82.653.9
DAT2N(control 75 ng/mL)78.82.793.5
Cutoff Urine99.72.632.6
Urine +25%1232.922.4
DAT2P(control 125 ng/mL)1273.212.5
Urine +50%1462.942.0
100 ng/mLQualitativeNumber TestedNegative/PositiveConfidence Level
Urine -50%8484/0> 95 % negative reading
Urine -25%8484/0> 95 % negative reading
DAT2N(control 75 ng/mL)8484/0> 95 % negative reading

{10}------------------------------------------------

Cutoff Urine8473/11N/A
Urine +25%840/84> 95 % positive reading
DAT2P(control 125 ng/mL)840/84> 95 % positive reading
Urine +50%840/84> 95 % positive reading
Repeatability200 ng/mL SQMean (ng/mL)SD (ng/mL)CV (%)
Urine -50%98.22.502.5
Urine -25%1462.511.7
DAT3N(control 150 ng/mL)1501.911.3
Cutoff Urine1992.481.2
Urine +25%2423.151.3
DAT3P(control 250 ng/mL)2482.671.1
Urine +50%2792.340.8
IntermediatePrecisionMean (ng/mL)SD (ng/mL)CV (%)
Urine -50%98.23.093.1
Urine -25%1462.872.0
DAT3N(control 150 ng/mL)1503.492.3
Cutoff Urine1993.331.7
Urine +25%2423.721.5

{11}------------------------------------------------

DAT3P(control 250 ng/mL)2485.682.3
Urine +50%2794.881.7
200 ng/mLQualitativeNumber TestedNegative/PositiveConfidence Level
Urine -50%8484/0> 95 % negative reading
Urine -25%8484/0> 95 % negative reading
DAT3N(control 150 ng/mL)8484/0> 95 % negative reading
Cutoff Urine8460/24N/A
Urine +25%840/84> 95 % positive reading
DAT3P(control 250 ng/mL)840/84> 95 % positive reading
Urine +50%840/84> 95 % positive reading
300 ng/mL SQRepeatabilityMean (ng/mL)SD (ng/mL)CV (%)
Urine -50%1514.412.9
Urine -25%2113.991.9
DAT1N(control 225 ng/mL)2235.502.5
Cutoff Urine2764.171.5
Urine +25%3545.211.5

{12}------------------------------------------------

DAT1P(control 375 ng/mL)3634.351.2
Urine +50%4325.141.2
IntermediatePrecisionMean (ng/mL)SD (ng/mL)CV (%)
Urine -50%1515.313.5
Urine -25%2115.402.6
DAT1N(control 225 ng/mL)2236.232.8
Cutoff Urine2766.072.2
Urine +25%3546.171.7
DAT1P(control 375 ng/mL)3637.352.0
Urine +50%4326.701.6
300 ng/mLQualitativeNumber TestedNegative/PositiveConfidence Level
Urine -50%8484/0> 95 % negative reading
Urine -25%8484/0> 95 % negative reading
DAT1N(control 225 ng/mL)8484/0> 95 % negative reading
Cutoff Urine8483/1N/A
Urine +25%840/84> 95 % positive reading
DAT1P(control 375 ng/mL)840/84> 95 % positive reading

{13}------------------------------------------------

Urine +50%840/84> 95 % positive reading
-----------------------------------------------

Analytical Recovery and Linearity 4.2.

The recovery study was evaluated on a single cobas c 501 in according to CLSI guideline EP06-A. The study protocol consisted of three reagent lots, the total number of samples was 17 levels per cutoff and ran in triplicate. The recovery study experiment was conducted using three reagent lots on one cobas c 501 instrument. Two series of samples were prepared for each cutoff to support that the recovery performance of the device is acceptable over the whole range between the lowest and the highest calibrator.

For each sample, the percentage recovery was calculated as the percent of the three results with regard to the target value. The average percent recovery is summarized in the table below.

100 CutoffLot 1Lot 2Lot 3
Target(ng/mL)Mean(ng/mL)Recovery(%)Mean(ng/mL)Recovery(%)Mean(ng/mL)Recovery(%)
01N/A0N/A4N/A
25271082911629115
50541075310552105
75761027610175100
10099999999100100
125122981229812197
150146971459714899
175169971689617298
200188941899519397

Results for 0-200% of 100 ng/mL cutoff

{14}------------------------------------------------

100 CutoffLot 1Lot 2Lot 3
Target(ng/mL)Mean(ng/mL)Recovery(%)Mean(ng/mL)Recovery(%)Mean(ng/mL)Recovery(%)
01N/A0N/A4N/A
1009898100100101101
200191961919619597
300288962919729598
400399100399100404101
500517103524105522104
600638106644107638106
700753108763109752107
800842105856107856107
900921102928103921102
1000982989789898899

Results for 0-1000% of 100 ng/mL cutoff

Results for 0-200% of 200 ng/mL cutoff

200 CutoffLot 1Lot 2Lot 3
Target(ng/mL)Mean(ng/mL)Recovery(%)Mean(ng/mL)Recovery(%)Mean(ng/mL)Recovery(%)
01N/A0N/A1N/A
50521045410752103
10097971001009999
150142951449614396
200191951909519296
250243972389524297

{15}------------------------------------------------

300292972899629598
35035110034799349100
400408102403101406101

Results for 0-1000% of 200 ng/mL cutoff

200 CutoffTarget(ng/mL)Mean(ng/mL)Lot 1Recovery(%)Mean(ng/mL)Lot 2Recovery(%)Mean(ng/mL)Lot 3Recovery(%)
01N/A0N/A1N/A
10099991001009999
200193961919519397
300293982899629699
400404101399100399100
500526105518104517103
600646108643107634106
700760109761109751107
800857107852107850106
900927103927103922102
1000986999829898699

Results for 0-200% of 300 ng/mL cutoff

300 CutoffTarget(ng/mL)Lot 1Lot 2Lot 3
Mean(ng/mL)Recovery(%)Mean(ng/mL)Recovery(%)Mean(ng/mL)Recovery(%)
00N/A0N/A9N/A
75891198411287116
150155104154103160106

{16}------------------------------------------------

225216962209822499
3002819428394300100
375349તે તે જેવી તેમ જ દૂધની ડેરી જેવી સવલતો પ્રાપ્ય થયેલી છે. આ ગામમાં પ્રાથમિક શાળા, પંચાયતઘર, આંગણવાડી તેમ જ દૂધની ડેરી જેવી સવલતો પ્રાપ્ય થયેલી છે. આ ગામમાં પ્રાથમિક શાળા, પંચ348ਰੇਤੇ35194
45041692418તે તે જેવી તેમ જ દૂધની ડેરી જેવી સવલતો પ્રાપ્ય થયેલી છે. આ ગામમાં પ્રાથમિક શાળા, પંચાયતઘર, આંગણવાડી તેમ જ દૂધની ડેરી જેવી સવલતો પ્રાપ્ય થયેલી છે. આ ગામમાં પ્રાથમિક શાળા, પંચ432તે તે રેણ
52548292487ਰੇਤੋ49194
60055092556તે તે જેવી તેમ જ દૂધની ડેરી જેવી સવલતો પ્રાપ્ય થયેલી છે. આ ગામમાં પ્રાથમિક શાળા, પંચાયતઘર, આંગણવાડી તેમ જ દૂધની ડેરી જેવી સવલતો પ્રાપ્ય થયેલી છે. આ ગામમાં પ્રાથમિક શાળા, પંચર્સાતે તે ઉ

Results for 0-1000% of 300 ng/mL cutoff

300 CutoffTarget(ng/mL)Mean(ng/mL)Lot 1Recovery(%)Mean(ng/mL)Lot 2Recovery(%)Mean(ng/mL)Lot 3Recovery(%)
00N/A0N/A9N/A
300284952819428896
600554925579356193
900869978889986997
1200125310412751061228102
1500168711217561171639109
1800209611621431192069115
2100242211524891192424115
2400267911227111132674111
2700283710528331052831105
3000290397286295290797

{17}------------------------------------------------

Analytical Specificity/Cross-Reactivity 4.3.

The determination of cross reactivity by common benzodiazepines was conducted on one cobas c 501, using one reagent lot. Concentration series were prepared for each cross reactant by spiking drug free urine. The percent cross-reactivity was calculated from the ratio of the cutoff concentration and the calculated equivalent concentration of the cross reactant. The study was conducted for the semi-quantitative application. Results are summarized below:

Benzodiazepine andmetabolitesConcentration TestedCross-reactivity (%)
Nordiazepam10199%
Alprazolam92109%
Diazepam90111%
Lorazepam10595%
Lorazepam glucuronide17856%
Oxazepam89113%
Oxazepam glucuronide13574%
Temazepam94106%
Temazepam glucuronide16063%
3-Hydroxybromazepam15366%
3-Hydroxyflubromazepam13077%
3-Hydroxyflunitrazepam11885%
4-Hydroxyalprazolam63160%
4-Hydroxytriazolam98102%
7-Acetamidonitrazepam214210.5%
Benzodiazepine andmetabolitesConcentration TestedCross-reactivity (%)
7-Aminoclonazepam12480%
7-Aminoflunitrazepam10991%
7-Aminonimetazepam87115%
7-Aminonitrazepam73137%
Bentazepam12878%
Bromazepam76132%
Brotiazolam13873%
Chlordiazepoxide10992%
Clobazam95106%
Clonazepam10397%
Clonazolam11091%
Clorazepate18953%
Delorazepam10992%
Demoxepam76131%
Desalkylflurazepam97103%
Deschloretizolam81124%
Desmethylchlordiazepoxide10893%
Desmethylflunitrazepam100100%
Desmethylmedazepam16859%
Diclazepam99101%
Benzodiazepine andmetabolitesConcentration TestedCross-reactivity (%)
Didesethylflurazepam11686%
Estazolam88114%
Etizolam11885%
Flubromazepam13276%
Flubromazolam10595%
Flunitrazepam11388%
Flurazepam16162%
Halazepam13276%
Hydroxyethylflurazepam10397%
Lormetazepam10794%
Meclonazepam12382%
Medazepam13872%
Midazolam10695%
Nifoxipam12978%
Nimetazepam99101%
Nitrazepam96105%
Phenazepam12481%
Pinazepam11091%
Prazepam12480%
Pyrazolam10397%
Benzodiazepine andmetabolitesConcentration TestedCross-reactivity (%)
Tetrazepam11686%
Triazolam10397%
α-Hydroxyalprazolam85118%
α-Hydroxyalprazolamglucuronide19053%
α-Hydroxymidazolam10397%
α-Hydroxymidazolamglucuronide17956%
α-Hydroxytriazolam96105%

Cross Reactivity of Benzodiazepines and Metabolites for 100 ng/mL Cutoff

{18}------------------------------------------------

{19}------------------------------------------------

{20}------------------------------------------------

Cross Reactivity of Benzodiazepines and Benzodiazepines Metabolites for 200 ng/mL

Benzodiazepine andmetabolitesConcentration TestedCross-reactivity (%)
Nordiazepam198101%
Alprazolam174115%
Diazepam175115%
Lorazepam20896%
Lorazepam glucuronide36056%
Oxazepam193104%
Oxazepam glucuronide28271%
Temazepam193103%
Temazepam glucuronide31863%
3-Hydroxybromazepam37154%
Benzodiazepine andmetabolitesConcentration TestedCross-reactivity (%)
3-Hydroxyflubromazepam25678%
3-Hydroxyflunitrazepam25977%
4-Hydroxyalprazolam119169%
4-Hydroxytriazolam20399%
7-Acetamidonitrazepam420050.5%
7-Aminoclonazepam28371%
7-Aminoflunitrazepam21991%
7-Aminonimetazepam21394%
7-Aminonitrazepam177113%
Bentazepam28869%
Bromazepam154130%
Brotiazolam26576%
Chlordiazepoxide28271%
Clobazam185108%
Clonazepam20896%
Clonazolam22689%
Clorazepate39650%
Delorazepam21792%
Demoxepam182110%
Desalkylflurazepam194103%
Benzodiazepine andmetabolitesConcentration TestedCross-reactivity (%)
Deschloretizolam159126%
Desmethylchlordiazepoxide30466%
Desmethylflunitrazepam196102%
Desmethylmedazepam39950%
Diclazepam20299%
Didesethylflurazepam23187%
Estazolam168119%
Etizolam23685%
Flubromazepam26675%
Flubromazolam20398%
Flunitrazepam21294%
Flurazepam31264%
Halazepam26974%
Hydroxyethylflurazepam20697%
Lormetazepam21494%
Meclonazepam30665%
Medazepam27972%
Midazolam20697%
Nifoxipam31064%
Nimetazepam20199%
Benzodiazepine andmetabolitesConcentration TestedCross-reactivity (%)
Nitrazepam188106%
Phenazepam25279%
Pinazepam21394%
Prazepam24083%
Pyrazolam20399%
Tetrazepam23386%
Triazolam20796%
α-Hydroxyalprazolam168119%
α-Hydroxyalprazolamglucuronide39950%
α-Hydroxymidazolam201100%
α-Hydroxymidazolamglucuronide37254%
α-Hydroxytriazolam198101%

{21}------------------------------------------------

{22}------------------------------------------------

{23}------------------------------------------------

Cross Reactivity of Benzodiazepines and Benzodiazepines Metabolites for 300 ng/mL Cutoffs

Benzodiazepine andmetabolitesConcentration TestedCross-reactivity (%)
Nordiazepam30598%
Alprazolam275109%
Diazepam273110%
Lorazepam32393%
Lorazepam glucuronide51958%
Benzodiazepine andmetabolitesConcentration TestedCross-reactivity (%)
Oxazepam275109%
Oxazepam glucuronide44268%
Temazepam272110%
Temazepam glucuronide49661%
3-Hydroxybromazepam50459%
3-Hydroxyflubromazepam40175%
3-Hydroxyflunitrazepam39975%
4-Hydroxyalprazolam192156%
4-Hydroxytriazolam31994%
7-Acetamidonitrazepam931480.3%
7-Aminoclonazepam39077%
7-Aminoflunitrazepam34387%
7-Aminonimetazepam283106%
7-Aminonitrazepam230131%
Bentazepam39376%
Bromazepam229131%
Brotiazolam40774%
Chlordiazepoxide37480%
Clobazam277108%
Clonazepam31795%
Benzodiazepine andmetabolitesConcentration TestedCross-reactivity (%)
Clonazolam33889%
Clorazepate60150%
Delorazepam33490%
Demoxepam256117%
Desalkylflurazepam30798%
Deschloretizolam257117%
Desmethylchlordiazepoxide37081%
Desmethylflunitrazepam301100%
Desmethylmedazepam54955%
Diclazepam31795%
Didesethylflurazepam35285%
Estazolam270111%
Etizolam36283%
Flubromazepam43569%
Flubromazolam32792%
Flunitrazepam32991%
Flurazepam48762%
Halazepam40674%
Hydroxyethylflurazepam32492%
Lormetazepam32892%
Benzodiazepine andmetabolitesConcentration TestedCross-reactivity (%)
Meclonazepam40973%
Medazepam39476%
Midazolam33290%
Nifoxipam41273%
Nimetazepam30997%
Nitrazepam294102%
Phenazepam38278%
Pinazepam32792%
Prazepam36383%
Pyrazolam31196%
Tetrazepam36083%
Triazolam31595%
α-Hydroxyalprazolam275109%
α-Hydroxyalprazolamglucuronide56753%
α-Hydroxymidazolam32692%
α-Hydroxymidazolamglucuronide56953%
α-Hydroxytriazolam31296%

{24}------------------------------------------------

{25}------------------------------------------------

{26}------------------------------------------------

Endogenous Interference Testing of Structurally Unrelated Compounds 4.4.

Interference from structurally unrelated compounds was evaluated with two sets of samples were prepared and measured for the analysis of each individual interferent. One set of the samples

{27}------------------------------------------------

contained the interferent in the presence of a benzodiazepine at both target concentrations (75% and 125% of the cutoff of the assay), and one set of the samples contained the interferent in the presence of a glucuronidated benzodiazepine (oxazepam-glucuronide) at both target concentrations (75% or 125% of the cutoff of the assay). Testing was performed in both qualitative and semiquantitative modes. The compounds listed in the table below did not cause any positive or negative interference at the concentrations shown:

Compound NameConcentration in presence of benzodiazepine (mg/dL)
Acetone1000
Ascorbic Acid1500
Calcium(as CaCl2)133
Citrate(K3-Citrate x H2O)357
Creatinine1000
Ethanol1000
Glucose7000
Hemoglobin750
Human Albumin250
Human IgG110
Magnesium(MgCl2)238
Oxalate(Na2-Oxalate)20
Phosphate(NaH2PO4 x H2O)2028
Sodium chloride5844
Urea18000
Compound NameConcentration inpresence ofbenzodiazepine(mg/dL)
Uric Acid100
Urobilinogen15

{28}------------------------------------------------

Interference Testing of Specific Gravity and pH 4.5.

Urine samples within a pH range from 4.0 to 9.0 and samples with specific gravities ranging from 1.001 to 1.034 containing benzodiazepine at the level of the negative control (75 ng/mL, 150 ng/mL or 225 ng/mL) and at the level of the positive control (125 ng/mL, 250 ng/mL or 375 ng/mL) corresponding respectively to the given cutoff (100 ng/mL, 200 ng/mL or 300 ng/mL) recovered properly in both semi-quantitative and qualitative modes.

Drug Interferences 4.6.

The drug interference study was conducted using one reagent lot on the cobas c 501. Samples spiked with benzodiazepine were measured in the presence of potentially interfering drugs. The target concentration for the benzodiazepine was 75% or the cutoff of the assay. For all samples three replicates were recorded with the semi-quantitative applications and the qualitative applications. The test results were checked for exceeding the cutoff value. The maximum drug concentration which doesn't interfere is reported and established as interferent claim. For the tested drugs and drug concentrations compare table below.

For Oxaprozin, the interference was additionally evaluated following an alternative protocol. In a drug-free matrix, the approximate quantity of Oxaprozin that is equivalent in assay reactivity to the 100 ng/mL, 200 ng/mL, and 300 ng/mL cutoff was determined to be 790 ng/mL, 3091 ng/mL and 3049 ng/mL respectively. This equals to a cross reactivity of 13% (at cutoff 100 ng/mL), 6% (at cutoff 200 ng/mL), and 10% (at cutoff 300 ng/mL).

When oxaprozin was added pooled human urine containing benzodiazepine at the level of the negative control (75 ng/mL, 150 ng/mL or 225 ng/mL), positive results were observed at >100 ng/mL, >200 ng/mL, and >300 ng/mL for the 100 ng/mL cutoff, 200 ng/mL cutoff, and 300 ng/mL

{29}------------------------------------------------

cutoff, respectively. Patient samples containing benzodiazepines in the presence of oxaprozin may yield falsely elevated results. Results should always be assessed in conjunction with the patient's medical history, clinical examinations, and other clinicopathological findings.

CompoundsConcentration (ng/mL)
Cutoff 100ng/mLCutoff 200ng/mLCutoff 300ng/mL
Acetaminophen300000030000003000000
Acetylsalicylic acid100000100000100000
Amitryptyline100000100000100000
Amobarbital100000100000100000
d-Amphetamine100000100000100000
1-Amphetamine100000100000100000
Ampicillin100000100000100000
Aspartame100000100000100000
Atropine100000100000100000
Benzocaine100000100000100000
Benzoylecgonine100000100000100000
Benzphetamine100000100000100000
Buspirone100000100000100000
Butabarbital100000100000100000
Ca-dobesilate100000010000001000000
Caffeine100000100000100000
Calcium hypochlorite100000100000100000
CompoundsConcentration (ng/mL)
Cutoff 100ng/mLCutoff 200ng/mLCutoff 300ng/mL
Cannabidiol100000100000100000
Captopril100000100000100000
Cefoxitin200000040000006000000
Chloroquine100000100000100000
Chlorpheniramine40000100000100000
Chlorpromazine100000100000100000
Cocaine100000100000100000
Codeine100000100000100000
Desipramine100000100000100000
Dextromethorphan100000100000100000
Dextropropoxyphene (d-Propoxyphene)100000100000100000
Digoxin100000100000100000
Diphenhydramine40000100000100000
Doxepine100000100000100000
Ecgonine100000100000100000
Ecgonine methyl ester100000100000100000
EDDP (2-Ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine)250005000075000
EMDP (2-Ethyl-5-methyl-3,3-diphenylpyrroline)250004000080000
Enalapril100000100000100000
CompoundsConcentration (ng/mL)
Cutoff 100ng/mLCutoff 200ng/mLCutoff 300ng/mL
d-Ephedrine100000100000100000
1-Ephedrine100000100000100000
Epinephrine100000100000100000
Erythromycin100000100000100000
Estriol100000100000100000
Fenoprofen40000100000100000
Flumazenil100000100000100000
Furosemide100000100000100000
Gentamicine sulfate400000400000400000
Gentisic acid100000100000100000
Glutethimide100000100000100000
Guaiacol glycerol ether100000100000100000
Hydrochlorothiazide100000100000100000
Hydroxyindole acetic acid100000100000100000
Hydroxyindole carboxylic acid100000100000100000
Ibuprofen400000040000004000000
Imipramine100000100000100000
Isoproterenol100000100000100000
Ketamine100000100000100000
Concentration (ng/mL)
CompoundsCutoff 100ng/mLCutoff 200ng/mLCutoff 300ng/mL
Levodopa100000010000001000000
Lidocaine100000100000100000
LSD100000100000100000
Melanin100000100000100000
Meperidine (Pethidin)100000100000100000
Methadone400004000040000
d-Methamphetamine100000100000100000
1-Methamphetamine100000100000100000
Methaqualone100000100000100000
Methyldopa200000020000002000000
Methylenedioxyamphetamine(MDA)100000100000100000
Methylenedioxymethamphetamine(MDMA)100000100000100000
Methylphenidate100000100000100000
Morphine100000100000100000
N-acetyl cysteine100001000010000
Naloxone100000100000100000
Naltrexone100000100000100000
Naproxen100000100000100000
Niacinamide100000100000100000
CompoundsConcentration (ng/mL)
Cutoff 100ng/mLCutoff 200ng/mLCutoff 300ng/mL
Nicotine100000100000100000
Norethindrone100000100000100000
1-Norpseudoephedrine100000100000100000
Ofloxacin900000900000900000
Omeprazol100000100000100000
Oxaprozin100200300
Penicillin G100000100000100000
Pentazocine100000100000100000
Pentobarbital100000100000100000
Phenazopyridine300000300000300000
Phencyclidine100000100000100000
Phenobarbital100000100000100000
Phenothiazine100000100000100000
Phenylbutazone100000100000100000
Phenylpropanolamine100000100000100000
Phenytoin100000100000100000
Procaine100000100000100000
Promethazine100000100000100000
d-Pseudoephedrine100000100000100000
CompoundsConcentration (ng/mL)
Cutoff 100ng/mLCutoff 200ng/mLCutoff 300ng/mL
Quetiapine500050005000
Quinidine100000100000100000
Quinine100000100000100000
Salicyluric acid600000060000006000000
Secobarbital100000100000100000
Sulindac100000100000100000
Tetracycline300000300000300000
Tetrahydrozoline100000100000100000
Δ9 THC-9-carboxylic acid100000100000100000
Thioridazine100000100000100000
Tolmetin100000100000100000
Trifluoperazine100000100000100000
Trimipramine100000100000100000
Tyramine100000100000100000
Verapmil100000100000100000
Zaleplone100000100000100000
Zolpidem500005000050000
Zopiclone100000100000100000

{30}------------------------------------------------

{31}------------------------------------------------

{32}------------------------------------------------

{33}------------------------------------------------

{34}------------------------------------------------

{35}------------------------------------------------

4.7. Method Comparison to Predicate

The method comparison study was performed in accordance with CLSI Guideline EP09-A3. A collection of 137 native human unaltered samples were purchased from clinical laboratories where they screened negative and preliminary positive flanking the cutoffs 100 ng/mL, 200 ng/mL and 300 ng/mL. The total of 117 of the 137 samples were measured with the assays with the cutoff 100 ng/mL, 134 samples were measured with the assays with cutoff 200 ng/mL and 114 samples were measured with the assays with cutoff 300 ng/mL. The results were compared to LC-MS/MS where samples were also treated with ß-glucuronidase.

A total of 48 urine samples, obtained from a clinical laboratory, where they screened negative in a drug test panel, were evaluated with Benzodiazepines II. 100 % of these normal urines were negative relative to the 100 ng/mL cutoff.

A total of 54 urine samples obtained from a clinical laboratory, where they screened preliminary positive with a commercially available immunoassay and were subsequently confirmed by Liquid Chromatography coupled with Tandem Mass Spectrometry (LC-MS/MS), were evaluated with the Benzodiazepines II assay. 100 % of these samples were positive to the 100 ng/mL cutoff.

In addition, 8 urine samples obtained from a clinical laboratory were found by LC-MS/MS in a concentration of 50-100 % of the cutoff concentration and 7 samples obtained from a clinical laboratory were found by LC-MS/MS in a concentration of 100-150 % of the cutoff concentration. The following results were obtained with the Benzodiazepines II assay on the cobas c 501 analyzer relative to the LC-MS/MS values. The results are summarized in the table below. Values in brackets represent the results for the two qualitative applications (Qualitative/Qualitative Clinical).

Benzodiazepines II Clinical Correlation (Cutoff = 100 ng/mL)
Negative SamplesLC-MS/MS values (ng/mL)
Near Cutoff
50-99104-147152-1558
cobas c 501analyzer+02 (1/1)754
-486 (7/7)00

{36}------------------------------------------------

The correlation between the semi-quantitative and the qualitative applications is summarized in the tables below.

Correlation (100 ng/mL Cutoff)
Semi-Quantitative
-+
Qualitative
+062
-541
Correlation (100 ng/mL Cutoff)
Semi-Quantitative -Semi-Quantitative +
Qualitative Clinical +062
Qualitative Clinical -541

The samples found to be discrepant at the 100 ng/mL cutoff with one of the predicate device applications are listed below.

SampleID#Semi-QuantitativeValues[ng/mL]Semi-QuantitativePos/NegQualitativePos/NegQualitativeClinicalPos/NegLC-MSConfirmationValues[ng/mL]LC-MSConfirmationPos/NegComment
UR_065104POSNEGNEG68NEGnear cutoff
UR_131150POSPOSPOS99NEGnear cutoff

A total of 56 urine samples obtained from a clinical laboratory, where they screened negative in a drug test panel, were evaluated with the Benzodiazepines II assay. 100 % of these normal urines were negative relative to the 200 ng/mL cutoff.

{37}------------------------------------------------

A total of 57 urine samples obtained from a clinical laboratory, where they screened preliminary positive with a commercially available immunoassay and were subsequently confirmed by Liquid Chromatography coupled with Tandem Mass Spectrometry (LC-MS/MS), were evaluated with the Benzodiazepines II assay. 100 % of these samples were positive to the 200 ng/mL cutoff. In addition, 12 urine samples obtained from a clinical laboratory were found by LC-MS/MS in a concentration of 50-100 % of the cutoff concentration and 9 urine samples obtained from a clinical laboratory were found by LC-MS/MS in a concentration of 100-150 % of the cutoff concentration. The following results were obtained with the Benzodiazepines II assay on the cobas c 501 analyzer relative to the LC-MS/MS values. The results are summarized in the table below. Values in brackets represent the results for the qualitative application.

Benzodiazepines II Clinical Correlation (Cutoff = 200 ng/mL)
Negative SamplesLC-MS/MS values (ng/mL)
Near Cutoff
137-196211-295304-2971
cobas c 501 analyzer+02 (3)757
-5610 (9)20

The correlation between the semi-quantitative and the qualitative applications is summarized in table below.

Correlation (200 ng/mL Cutoff)
Semi-Quantitative
-+
Qualitative+166
-670

The samples found to be discrepant at the 200 ng/mL cutoff with one of the predicate device applications are listed below.

{38}------------------------------------------------

Sample ID#Semi-QuantitativeValues [ng/mL]Semi-QuantitativePos/NegQualitativePos/NegLC-MSValues [ng/mL]LC-MSPos/Neg ConfirmationComment
UR_051132NEGNEG282POSnear cutoff
UR_060235POSPOS147NEGnear cutoff
UR_073197NEGPOS196NEGnear cutoff
UR_134249POSPOS142NEGnear cutoff

A total of 40 urine samples obtained from a clinical laboratory, where they screened negative in a drug test panel, were evaluated with the Benzodiazepines II assay. 100 % of these normal urines were negative relative to the 300 ng/mL cutoff.

A total of 45 urine samples obtained from a clinical laboratory, where they screened preliminary positive with a commercially available immunoassay and were subsequently confirmed by Liquid Chromatography coupled with Tandem Mass Spectrometry (LC-MS/MS), were evaluated with the Benzodiazepines II assay. 100 % of these samples were positive to the 300 ng/mL cutoff. In addition, 17 urine samples obtained from a clinical laboratory were found by LC-MS/MS in a concentration of 50-100% of the cutoff concentration and 12 urine samples obtained from a clinical laboratory were found by LC-MS/MS in a concentration of 100-150 % of the cutoff concentration. The following results were obtained with the Benzodiazepines II assay on the cobas c 501 analyzer relative to the LC-MS/MS values. The results are summarized in the table below. Values in brackets represent the results for the qualitative application.

Benzodiazepines II Clinical Correlation (Cutoff = 300 ng/mL)
NegativeSamplesLC-MS/MS values (ng/mL)
Near Cutoff456-2971
152-295304-441
cobas c 501analyzer+01 (4)1145
-4016 (13)10

The samples found to be discrepant at the 300 ng/mL cutoff with one of the predicate device applications are listed below.

SampleSemi-QuantitativeQualitativeLC-MSComment
--------------------------------------------------------

{39}------------------------------------------------

ID#Values[ng/mL]Pos/NegPos/NegValues[ng/mL]Pos/NegConfirmation
UR_062276NEGNEG304POSnear cutoff
UR_076297NEGPOS277NEGnear cutoff
UR_077296NEGPOS216NEGnear cutoff
UR_078295NEGPOS272NEGnear cutoff
UR_083348POSPOS295NEGnear cutoff

Stability 4.8.

The measurements to prove the on board stability of 12 weeks were carried out using one reagent lot on the cobas c 501 instrument. The reagent transport stress was simulated using the following incubation condition: ≥ 120 h at 35°C (± 2°C).

Two control samples, the cutoff calibrator and two clinical samples, were tested per each cutoff. The two clinical samples were prepared by pooling urine of individual donor samples in a way to achieve concentrations around 50% and 150% of the cutoff respectively.

SpecimenCutoff100 ng/mLCutoff200 ng/mLCutoff300 ng/mL
negative control(target value)DAT2N, DATCN(75 ng/mL)DAT3N(150 ng/mL)DAT1N(225 ng/mL)
positive control(target value)DAT2P, DATCP(125 ng/mL)DAT3P(250 ng/mL)DAT1P(375 ng/mL)
calibrator(target value)Preciset DAT Plus II / S3(100 ng/mL)Preciset DAT Plus II / S4(200 ng/mL)Preciset DAT Plus I / S3(300 ng/mL)
clinical negapprox. 50 ng/mLapprox. 100 ng/mLapprox. 150 ng/mL
clinical posapprox. 150 ng/mLapprox. 300 ng/mLapprox. 450 ng/mL

Overview: samples for the on board stability experiment

100 ng/mL for Semi-Quantitative

time point1234
ReagentStatus:starttransport stressed(≥120 h 35°C)transport stressed+84 days onboardtransport stressed+91 days onboard
negative control (n=21)
agreement [%]100100100100
mean [ng/mL]76.577.580.777.0
SD [ng/mL]1.692.161.821.76

{40}------------------------------------------------

time point1234
ReagentStatus:starttransport stressed(≥120 h 35°C)transport stressed+84 days onboardtransport stressed+91 days onboard
CV [%]2322
positive control (n=21)
agreement [%]100100100100
mean [ng/mL]125126129124
SD [ng/mL]2.321.642.022.16
CV [%]2122
negative clinical sample (n=3)
agreement [%]100100100100
positive clinical sample (n=3)
agreement [%]100100100100
calibrator (n=3)
mean [ng/mL]100101106101

200 ng/mL for Semi-Quantitative

time point1234
Reagentstarttransport stressed(≥120 h 35°C)transport stressed+ 84 days onboardtransport stressed+ 91 days onboard
Status:
negative control (n=21)
agreement [%]100100100100
mean [ng/mL]151152151147
SD [ng/mL]1.512.552.801.63
CV [%]1221
positive control (n=21)
agreement [%]100100100100
mean [ng/mL]255257254251
SD [ng/mL]2.823.912.592.46
CV [%]1211
negative clinical sample (n=3)
agreement [%]100100100100
positive clinical sample (n=3)
agreement [%]100100100100
calibrator (n=3)
mean [ng/mL]199201199197

300 ng/mL for Semi-Quantitative

time point1234
Reagentstarttransport stressed(≥120 h 35°C)transport stressed+84 days onboardtransport stressed+91 days onboard
Status:

{41}------------------------------------------------

negative control (n=21)
agreement [%]100100100100
mean [ng/mL]226.0231.3230.7227.4
SD [ng/mL]3.512.855.706.45
CV [%]2223
positive control (n=21)
agreement [%]100100100100
mean [ng/mL]363.0371.4370.0364.0
SD [ng/mL]4.855.425.616.20
CV [%]1122
negative clinical sample (n=3)
agreement [%]100100100100
positive clinical sample (n=3)
agreement [%]100100100100
calibrator (n=3)
mean [ng/mL]291301296287

100 ng/mL for Qualitative

time point1234
Reagent Status:starttransport stressed(≥120 h 35°C)transport stressed+ 84 days onboardtransport stressed+ 91 days onboard
negative control (n=21)
agreement [%]100100100100
positive control (n=21)
agreement [%]100100100100
negative clinical sample (n=3)
agreement [%]100100100100
positive clinical sample (n=3)
agreement [%]100100100100

200 ng/mL for Qualitative

time point1234
Reagent Status:starttransport stressed(≥120 h 35°C)transport stressed +84 days onboardtransport stressed +91 days onboard
negative control (n=21)
agreement [%]100100100100
positive control (n=21)
agreement [%]100100100100
negative clinical sample (n=3)
agreement [%]100100100100
positive clinical sample (n=3)
agreement [%]100100100100

{42}------------------------------------------------

300 ng/mL for Qualitative

time point1234
Reagent Status:starttransport stressed(≥120 h 35°C)transport stressed+84 days onboardtransport stressed+91 days onboard
negative control (n=21)
agreement [%]100100100100
positive control (n=21)
agreement [%]100100100100
negative clinical sample (n=3)
agreement [%]100100100100
positive clinical sample (n=3)
agreement [%]100100100100

100 ng/mL for Qualitative Clinical

time point1234
Reagent Status:starttransport stressed(≥120 h 35°C)transport stressed +84 days onboardtransport stressed +91 days onboard
negative control (n=21)
agreement [%]100100100100
positive control (n=21)
agreement [%]100100100100
negative clinical sample (n=3)
agreement [%]100100100100
positive clinical sample (n=3)
agreement [%]100100100100

CONCLUSIONS 5.

The information provided in this 510(k) Premarket Notification will support a determination of substantial equivalence for ONLINE DAT Benzodiazepines II. The data supports a safe and effective device, which performs as well or better than the predicate devices.

§ 862.3170 Benzodiazepine test system.

(a)
Identification. A benzodiazepine test system is a device intended to measure any of the benzodiazepine compounds, sedative and hypnotic drugs, in blood, plasma, and urine. The benzodiazepine compounds include chlordiazepoxide, diazepam, oxazepam, chlorzepate, flurazepam, and nitrazepam. Measurements obtained by this device are used in the diagnosis and treatment of benzodiazepine use or overdose and in monitoring levels of benzodiazepines to ensure appropriate therapy.(b)
Classification. Class II (special controls). A benzodiazepine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).