Intended Use

Polyisoprene Powder Free Surgical Glove Tested for Use with Chemotherapy Drugs is intended to be worn by operating room personnel to protect surgical wound from contamination. It is also tested for use against Chemotherapy Drugs.

The gloves were tested for use with chemotherapy drugs as per ASTM D6978-05 (Reapproved 2019) Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs.

Chemotherapy Drug and ConcentrationMinimum Breakthrough Detection Time in Minutes
Carmustine (3.3 mg/ml)12.3
Cisplatin (1.0 mg/ml)>240
Cyclophosphamide (20.0 mg/ml)>240
Dacarbazine (10.0 mg/ml)>240
Doxorubicin Hydrochloride (2.0 mg/ml)>240
Etoposide (20.0 mg/ml)>240
Fluorouracil (50.0 mg/ml)>240
Methotrexate (25.0 mg/ml)>240
Mitomycin C (0.5 mg/ml)>240
Paclitaxel (6.0 mg/ml)>240
Thiotepa (10.0 mg/ml)17.4
Vincristine Sulfate (1.0 mg/ml)>240

Please note that Carmustine and Thiotepa have extremely low permeation times of 12.3 minutes and 17.4 minutes respectively.

Warning: Do not use with Carmustine and Thiotepa

Device Description

Polyisoprene Powder Free Surgical Glove Tested for Use with Chemotherapy Drugs is a disposable single-use, sterile, natural-colored and powder-free surgical glove made from synthetic polyisoprene latex.

AI/ML Overview

The provided documents describe the acceptance criteria and performance of the Polyisoprene Powder Free Surgical Glove Tested for Use with Chemotherapy Drugs. This is a medical device, and the information is presented as part of an FDA 510(k) premarket notification.

Here's a breakdown of the requested information based on the provided text:


1. Table of Acceptance Criteria and Reported Device Performance

The core of the acceptance criteria and device performance is detailed in the "TECHNOLOGICAL CHARACTERISTICS COMPARISON TABLE" and the "SUMMARY OF NON-CLINICAL TESTING" sections.

Characteristic / Test MethodologyAcceptance Criteria (Standard)Reported Device Performance
Chemotherapy Drugs PermeationASTM D6978-05 (Reapproved 2019): Under the conditions of the study, no permeation.Carmustine (3.3 mg/ml): 12.3 minutesCisplatin (1.0 mg/ml): > 240 minutesCyclophosphamide (20.0 mg/ml): > 240 minutesDacarbazine (10.0 mg/ml): > 240 minutesDoxorubicin Hydrochloride (2.0 mg/ml): > 240 minutesEtoposide (20.0 mg/ml): > 240 minutesFluorouracil (50.0 mg/ml): > 240 minutesMethotrexate (25.0 mg/ml): > 240 minutesMitomycin C (0.5 mg/ml): > 240 minutesPaclitaxel (6.0 mg/ml): > 240 minutesThiotepa (10.0 mg/ml): 17.4 minutesVincristine Sulfate (1.0 mg/ml): > 240 minutesNote: Carmustine and Thiotepa had extremely low permeation times, leading to a "Do not use" warning.
Freedom from HolesASTM D3577-19 requirements of AQL 1.5Meets ASTM D3577-19 requirements of AQL 1.5
Length (ASTM D3577-19)Varies by size (e.g., 5.5: Min 245mm, 6.0-9.0: Min 265mm)Meets ASTM D3577-19 requirements for length (e.g., Size 5.5: 285mm, Size 9.0: 291mm)
Width (ASTM D3577-19)Varies by size (e.g., 5.5: 70 ± 6mm, 9.0: 114 ± 6mm)Meets ASTM D3577-19 requirements for width (e.g., Size 5.5: 74mm, Size 9.0: 116mm)
Thickness (ASTM D3577-19)Palm, Finger, Cuff Minimum 0.10 mmMeets ASTM D3577-19 requirements for thickness (e.g., Palm: 0.19-0.21mm, Finger: 0.23-0.24mm, Cuff: 0.15-0.16mm)
Tensile Strength (ASTM D3577-19)Before Aging: ≥ 17 MPaAfter Aging: ≥ 12 MPaBefore Aging: Average 17.9 MPaAfter Aging: Average 15.2 MPa
Ultimate Elongation (ASTM D3577-19)Before Aging: ≥ 650 %After Aging: ≥ 490 %Before Aging: Average 952 %After Aging: Average 940 %
Stress at 500% Elongation (ASTM D3577-19)Max 7.0 MPa (Before Aging)Average 2.2 MPa (Before Aging)
Powder Residual (ASTM D6124-06 (2017))≤ 2 mg per gloveAverage 0.34 mg/glove
In Vitro Cytotoxicity (ISO 10993-5)Not cytotoxicFound to be cytotoxic, but further evaluation (ISO 10993-11) showed no systemic toxicity.
Primary Skin Irritation (ISO 10993-10)Not an irritantNot an irritant
Dermal Sensitization (ISO 10993-10)Not a sensitizerNot a sensitizer
Acute Systemic Toxicity (ISO 10993-11)Does not pose a toxicity concernNo signs of toxicity
Pyrogenicity Test (USP <151>)Non-pyrogenicNon-pyrogenic

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the specific sample sizes for each test in the "test set" (i.e., the data used to prove the device meets acceptance criteria). However, it lists recognized industry standards (ASTM, ISO, USP) for each test. These standards typically define minimum sample sizes for testing to ensure statistical validity.

The data provenance can be inferred based on the applicant information:

  • Country of Origin of the Data: The applicant is Hartalega NGC Sdn. Bhd., located in Sepang, Selangor, Malaysia. Therefore, the testing was likely conducted in Malaysia or by laboratories commissioned by the Malaysian company.
  • Retrospective or Prospective: These are non-clinical (laboratory) tests performed on the device to demonstrate its properties. This would be considered prospective in the sense that the tests were specifically conducted to evaluate this new device for regulatory submission, not retrospectively analyzed from existing data.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

This information is not applicable as the evaluation relies on objective, standardized laboratory tests (e.g., chemical permeation, physical property measurements, biocompatibility assays) rather than expert interpretation of a "test set" in the context of medical imaging or diagnostic algorithms. The "ground truth" is established by the defined parameters and methodologies of the ASTM, ISO, and USP standards.


4. Adjudication Method for the Test Set

This is not applicable as the evaluation relies on objective, standardized laboratory tests. Adjudication methods like 2+1 or 3+1 are typically used in studies involving human readers and subjective interpretations, such as radiology image reviews.


5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

No, an MRMC comparative effectiveness study was not done. The document explicitly states: "CLINICAL PERFORMANCE DATA: Not applicable. There was no clinical data required to support the subject device as the indication for use is equivalent to the predicate device." MRMC studies are used to evaluate the diagnostic performance of devices, often medical imaging AI, with human-in-the-loop scenarios. This document describes the performance of a physical medical device (surgical glove) through non-clinical testing.


6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

This is not applicable as the device is a physical surgical glove, not an algorithm or AI system. The tests evaluate the physical and chemical properties of the glove itself.


7. The Type of Ground Truth Used

The "ground truth" for this device's evaluation is based on established standard test methodologies and defined acceptance criteria set by organizations like ASTM, ISO, and USP. These are objective measurements:

  • Physical measurements (length, width, thickness)
  • Mechanical properties (tensile strength, elongation)
  • Chemical resistance (breakthrough time for chemotherapy drugs)
  • Biological safety (cytotoxicity, irritation, sensitization, systemic toxicity, pyrogenicity)

It does not involve expert consensus, pathology, or outcomes data in the typical sense of AI/diagnostic device evaluation.


8. The Sample Size for the Training Set

This is not applicable. This is a physical medical device, not an AI or machine learning model that requires a training set. The term "training set" refers to data used to train an algorithm.


9. How the Ground Truth for the Training Set was Established

This is not applicable for the same reason as Section 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

September 24, 2022

Hartalega NGC SDN BHD Nurul Kong Deputy General Manager - Quality Assurance NO. 1 Persiaran Tanjung Kawasan Perindustrian Tanjung Sepang, Selangor Darul Ehsan 43900 Malaysia

Re: K221718

Trade/Device Name: Polyisoprene Powder Free Surgical Glove Tested for Use with Chemotherapy Drugs (Natural) Regulation Number: 21 CFR 878.4460 Regulation Name: Non-Powdered Surgeon's Glove Regulatory Class: Class I, reserved Product Code: KGO, LZC Dated: August 15, 2022 Received: August 23, 2022

Dear Nurul Kong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Bifeng Oian.M.D.,Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K221718

Device Name

Polyisoprene Powder Free Surgical Glove Tested for Use with Chemotherapy Drugs

Indications for Use (Describe)

Polyisoprene Powder Free Surgical Glove Tested for Use with Chemotherapy Drugs is intended to be worn by operating room personnel to protect surgical wound from contamination. It is also tested for use against Chemotherapy Drugs.

The gloves were tested for use with chemotherapy drugs as per ASTM D6978-05 (Reapproved 2019) Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs.

Chemotherapy Drug and ConcentrationMinimum Breakthrough Detection Time in Minutes
Carmustine (3.3 mg/ml)12.3
Cisplatin (1.0 mg/ml)>240
Cyclophosphamide (20.0 mg/ml)>240
Dacarbazine (10.0 mg/ml)>240
Doxorubicin Hydrochloride (2.0 mg/ml)>240
Etoposide (20.0 mg/ml)>240
Fluorouracil (50.0 mg/ml)>240
Methotrexate (25.0 mg/ml)>240
Mitomycin C (0.5 mg/ml)>240
Paclitaxel (6.0 mg/ml)>240
Thiotepa (10.0 mg/ml)17.4
Vincristine Sulfate (1.0 mg/ml)>240

Please note that Carmustine and Thiotepa have extremely low permeation times of 12.3 minutes and 17.4 minutes respectively.

Warning: Do not use with Carmustine and Thiotepa

Type of Use (Select one or both, as applicable)

] Prescription Use (Part 21 CFR 801 Subpart D)

× Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY (K221718)

FOR

POLYISOPRENE POWDER FREE SURGICAL GLOVE TESTED FOR USE WITH CHEMOTHERAPY DRUGS

(The information contained herein is being provided in accordance with the requirements of 21 CFR 807.92)

APPLICANT INFORMATION

Date Prepared: April 27, 2022
Name: Hartalega NGC Sdn. Bhd.
Address: No. 1, Persiaran Tanjung,Kawasan Perindustrian Tanjung,43900 Sepang, Selangor,Malaysia
Establishment Registration Number: 3011200663

CORRESPONDENT AND/OR PREPARER INFORMATION

Contact Name:Nurul Aisyah Kong
Contact Title:Deputy General Manager – Quality Assurance
Phone Number:(603) 3280 3888
Fax Number:(603) 3271 0135
Contact Email:wkkong@hartalega.com.my

DEVICE IDENTIFICATION

Common Name of the Device:Surgeon's Glove
Trade Name (Proprietary Name):Polyisoprene Powder Free Surgical Glove Tested for Use withChemotherapy Drugs
Device Class:1
Product Code:KGO, LZC
Regulation Number:21 CFR 878.4460
Reason for 510(k) Submission:New device

{4}------------------------------------------------

PREDICATE DEVICE INFORMATION

510(k) NumberTradenameProduct Code
K190018Gammex Non-Latex Polyisoprene White Surgical Gloves Tested forUse with Chemotherapy DrugsKGO

DESCRIPTION OF THE DEVICE:

Polyisoprene Powder Free Surgical Glove Tested for Use with Chemotherapy Drugs is a disposable single-use, sterile, natural-colored and powder-free surgical glove made from synthetic polyisoprene latex.

INDICATIONS FOR USE:

Polyisoprene Powder Free Surgical Glove Tested for Use with Chemotherapy Drugs is intended to be worn by operating room personnel to protect surgical wound from contamination. It is also tested for use against Chemotherapy Drugs.

The gloves were tested for use with chemotherapy drugs as per ASTM D6978-05 (Reapproved 2019) Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs.

Chemotherapy Drug and ConcentrationMinimum Breakthrough Detection Time in Minutes
Carmustine (3.3 mg/ml)12.3
Cisplatin (1.0 mg/ml)> 240
Cyclophosphamide (20.0 mg/ml)> 240
Dacarbazine (10.0 mg/ml)> 240
Doxorubicin Hydrochloride (2.0 mg/ml)> 240
Etoposide (20.0 mg/ml)> 240
Fluorouracil (50.0 mg/ml)> 240
Methotrexate (25.0 mg/ml)> 240
Mitomycin C (0.5 mg/ml)> 240
Paclitaxel (6.0 mg/ml)> 240
Thiotepa (10.0 mg/ml)17.4
Vincristine Sulfate (1.0 mg/ml)> 240

Please note that Carmustine and Thiotepa have extremely low permeation times of 12.3 minutes and 17.4 minutes respectively.

Warning: Do not use with Carmustine and Thiotepa

{5}------------------------------------------------

TECHNOLOGICAL CHARACTERISTICS COMPARISON TABLE:

CharacteristicsandParametersSubject DevicePredicate Device(K190018)Discussion
Trade NamePolyisoprene Powder Free Surgical Glove Tested for Use with Chemotherapy DrugsGammex Non-Latex Polyisoprene White Surgical Gloves Tested for Use with Chemotherapy DrugsDifferent
ApplicantHartalega NGC Sdn. Bhd.Ansell Healthcare Products LLCDifferent
Product CodeKGO, LZCKGOSimilar
Classification11Similar
RegulationNumber21 CFR 878.446021 CFR 878.4460Similar
RegulationNameSurgeon's GloveSurgeon's GloveSimilar
Indications forUsePolyisoprene Powder Free Surgical Glove Tested for Use with Chemotherapy Drugs is intended to be worn by operating room personnel to protect surgical wound from contamination. It is also tested for use against Chemotherapy Drugs.The gloves were tested for use with chemotherapy drugs as per ASTM D6978-05 (Reapproved 2019) Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs.Gammex Non Latex PI White Surgical Gloves Tested for Use with Chemotherapy Drugs are intended to be worn by operating room personnel to protect a surgical wound from contamination.These gloves were tested for use with Chemotherapy Drugs as per ASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.Similar
ChemotherapyDrug andConcentrationMinimumBreakthroughDetection Time inMinutesChemotherapyDrug andConcentrationMinimumBreakthroughDetection Time inMinutes
Carmustine(3.3 mg/ml)12.3Carmustine(3.3 mg/ml)10.2Performancelevel forCarmustine andThiotepabetween subjectdevice andpredicate deviceis comparable.AdditionalChemotherapydrugs weretested on thepredicatedevice.
Cisplatin(1.0 mg/ml)> 240Cisplatin(1.0 mg/ml)> 240
Cyclophosphamide(20.0 mg/ml)> 240Cyclophosphamide(20.0 mg/ml)> 240
TestChemotherapyDrugsDacarbazine(10.0 mg/ml)> 240Dacarbazine(10.0 mg/ml)> 240
DoxorubicinHydrochloride(2.0 mg/ml)> 240DoxorubicinHydrochloride(2.0 mg/ml)> 240
Etoposide(20.0 mg/ml)> 240Etoposide(20.0 mg/ml)> 240
Fluorouracil(50.0 mg/ml)> 240Fluorouracil(50.0 mg/ml)> 240
Methotrexate(25.0 mg/ml)> 240Methotrexate(25.0 mg/ml)> 240
Mitomycin C(0.5 mg/ml)> 240Mitomycin C(0.5 mg/ml)> 240
Paclitaxel(6.0 mg/ml)> 240Paclitaxel(6.0 mg/ml)> 240
Thiotepa(10.0 mg/ml)17.4Thiotepa(10.0 mg/ml)11.5

{6}------------------------------------------------

CharacteristicsandParametersSubject DevicePredicate Device (K190018)Discussion
Vincristine Sulfate(1.0 mg/ml)> 240Please note that Carmustine and Thiotepa have extremely low permeation times of 12.3 minutes and 17.4 minutes respectively.Warning: Do not use with Carmustine and ThiotepaVincristine Sulfate (1.0 mg/ml) > 240Blenoxane (15 mg/ml) > 240Busulfan (6 mg/ml) > 240Cytarabine (100 mg/ml) > 240Daunorubicin (5 mg/ml) > 240Docetaxel (10.0 mg/ml) > 240Fludarabine (25 mg/ml) > 240Gemcitabine (38 mg/ml) > 240Idarubicin (1 mg/ml) > 240Ifosfamide (50.0 mg/ml) > 240Irinotecan (20.0 mg/ml) > 240Mechlorethamine HCI (1.0 mg/ml) > 240Melphalan (5 mg/ml) > 240Mitoxantrone (2.0 mg/ml) > 240Oxaliplatin (2.0 mg/ml) > 240Paraplatin (10 mg/ml) > 240Ellence (2 mg/ml) > 240Rituximab (10 mg/ml) > 240Please note that the following drugs have extremely low permeation times:Carmustine (BCNU): 10.2 minutes and Thiotepa:11.5 minutes.Warning: Do not use with Carmustine and Thiotepa
Type of useOver the counter useOver the counter useSimilar
MaterialsPolyisoprenePolyisopreneSimilar
ColorNaturalWhiteDifferent
Design• Single Use• Sterile• Powder-Free• Hand Specific• Single Use• Sterile• Powder-Free• Hand SpecificSimilar
CharacteristicsandParametersSubject DevicePredicate Device(K190018)Discussion
• Beaded Cuff• Beaded Cuff
SterilitySterileSterileSimilar
SterilizationRadiation10-6 SALRadiation10-6 SALSimilar
Freedom fromholesMeets ASTM D3577-19 requirements ofAQL 1.5Meets ASTM D3577-19 requirements ofAQL 1.5Similar
LengthLength (mm): ≥ 265 mmLength (mm): ≥ 265 mmSimilar.Meets ASTMD3577-19
Dimensions5.5: 70 ± 6 (mm)6.0: 76 ± 6 (mm)6.5: 83 ± 6 (mm)7.0: 89 ± 6 (mm)7.5: 95 ± 6 (mm)8.0: 102 ± 6 (mm)8.5: 108 ± 6 (mm)9.0: 114 ± 6 (mm)5.5: 70 ± 6 (mm)6.0: 76 ± 6 (mm)6.5: 83 ± 6 (mm)7.0: 89 ± 6 (mm)7.5: 95 ± 6 (mm)8.0: 102 ± 6 (mm)8.5: 108 ± 6 (mm)9.0: 114 ± 6 (mm)Similar.Meets ASTMD3577-19
ThicknessCuff Thickness: ≥ 0.10 mmPalm Thickness: ≥ 0.10 mmFinger Thickness: ≥ 0.10 mmCuff Thickness: ≥ 0.10 mmPalm Thickness: ≥ 0.10 mmFinger Thickness: ≥ 0.10 mmSimilar.Meets ASTMD3577-19
PhysicalPropertiesTensile Strength Before Aging: ≥ 17 MPaTensile Strength After Aging: ≥ 12 MPaUltimate Elongation Before Aging: ≥ 650 %Ultimate Elongation After Aging: ≥ 490 %Tensile Strength Before Aging: ≥ 17 MPaTensile Strength After Aging: ≥ 12 MPaUltimate Elongation Before Aging: ≥ 650 %Ultimate Elongation After Aging: ≥ 490 %Similar.Meets ASTMD3577-19
Powder residualResidual Powder: ≤ 2 mg per gloveResidual Powder: ≤ 2 mg per gloveSimilar.Meets ASTMD3577-19
In VitroCytotoxicityISO 10993-5Under the conditions of the study, the devicewas found to be cytotoxicUnder the conditions of the study, the devicewas found to be cytotoxicSimilar
Primary SkinIrritationISO 10993-10Under the conditions of the study, the deviceis not an irritantUnder the conditions of the study, the devicenot an irritantSimilar
DermalSensitizationISO 10993-10Under the conditions of the study, the deviceis not a sensitizerUnder the conditions of the study, the devicenot a sensitizerSimilar
Acute SystemicToxicityISO 10993-11Under the conditions of the study, therewas no evidence of systemic toxicityUnder the conditions of the study, therewas no evidence of systemic toxicitySimilar
PyrogenicityTestUSP <151>Under the conditions of the study, the testarticle was non-pyrogenicN/AAdditionalPyrogenicityTest was testedon the subjectdevice
Test Methodology/StandardsAcceptance Criteria of theStandardsResult Summary
SizeLength(mm)Width(mm)
DimensionsASTM D3577-19Standard Specification forRubber Surgical Gloves5.5Min 24570 ± 6Meets ASTM D3577-19 requirements for lengthand width. Similar to predicate device.
6.0Min 26576 ± 6
6.5Min 26583 ± 6
7.0Min 26589 ± 6
7.5Min 26595 ± 6
8.0Min 265102 ± 6
8.5Min 265108 ± 6
9.0Min 265114 ± 6
SizeAverage Length (mm)
5.528574
6.028380
6.529084
7.029091
7.529298
8.0300103
8.5297109
9.0291116
Meets ASTM D3577-19 requirements for thickness.Similar to predicate device
DimensionsThickness (mm)SizeAverage Palm Thickness (mm) MinAverage Finger Thickness (mm)Average Cuff Thickness (mm)
ASTM D3577-19Standard Specification forRubber Surgical GlovesPalm Minimum 0.10Finger Minimum 0.10Cuff Minimum 0.105.50.210.230.15
6.00.200.230.16
6.50.200.230.16
7.00.190.230.15
7.50.200.240.15
8.00.190.230.15
8.50.200.240.16
9.00.200.240.15
Physical PropertiesASTM D3577-19Standard Specification forRubber Surgical GlovesParameterBefore AgingAfter AgingMeets ASTM D3577-19 requirements for tensile strength and elongation at break before and after accelerated aging. Similar to predicate device.
Tensile StrengthMin 17 MPaMin 12 MPaBefore AgeTensile Strength (MPa)Average 17.9Elongation at Break (%)Average 952Stress at 500% Elongation (MPa)Average 2.2
Ultimate ElongationMin 650%Min 490%After AgeTensile Strength (MPa)Average 15.2Elongation at Break (%)Average 940
Stress at 500% ElongationMax 7.0 MPaN/A
Freedom from holesASTM D5151-19Standard Test Method forDetection of Holes inMedical GlovesASTM D3577-19Standard Specification forRubber Surgical GlovesAQL 1.5Meets ASTM D3577-19 and ASTM D5151-19 requirements of AQL 1.5. Similar to predicate device.
Powder residualASTM D6124-06(2017)Standard Test Method forResidual Powder onMedical GlovesASTM D3577-19Standard Specification forRubber Surgical GlovesPowder Free; ≤ 2 mg per gloveMeets ASTM D3577-19 and ASTM D6124-06 (2017) requirements for Powder Free; ≤ 2 mg per glove. Similar to predicate device.Average 0.34 mg/glove
Chemotherapy DrugsPermeationASTM D6978-05(2019)Standard Practice forAssessment of Resistance ofMedical Gloves toPermeation byChemotherapy DrugsUnder the conditions of the study, no permeation.See results in the Table above.

{7}------------------------------------------------

{8}------------------------------------------------

SUMMARY OF NON-CLINICAL TESTING:

Non-clinical testing was performed to verify that the subject device meets the acceptance criteria of the performance test and all design specifications. The test results demonstrated that the subject device with the following standards as shown below.

  • . ASTM D3577-19 Standard Specification for Rubber Surgical Gloves
  • . ASTM D5151-19 Standard Test Method for Detection of Holes in Medical Gloves
  • ASTM D6124-06 (Reapproved 2017) Standard Test Method for Residual Powder on Medical Gloves .
  • . ASTM D6978-05 (Reapproved 2019) Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs
  • . ASTM F1929 -15 Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration
  • . ISO 11137-1:2006 Sterilization of health care products - Radiation - Part 1: Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices
  • . ISO 11137-2:2013 Sterilization of health care products - Radiation - Part 2: Establishing the sterilization dose
  • ISO 10993-5 In Vitro Cytotoxicity .
  • . ISO 10993-10 Primary Skin Irritation
  • ISO 10993-10 Dermal Sensitization
  • . ISO 10993-11 Acute Systemic Toxicity
  • . USP <151> Pyrogen Test (USP Rabbit Test)
  • . USP <85> Bacterial Endotoxins Test

Physical Characteristics

{9}------------------------------------------------

{10}------------------------------------------------

Pyrogenicity TestUSP <151>Under the conditions of the study, thetest article was non-pyrogenicUnder the conditions of the study, the test article wasnon-pyrogenic
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

Biocompatibility

Test Methodology/StandardsAcceptance Criteria of theStandardsResult Summary
In Vitro CytotoxicityISO 10993-5:2009Biological evaluation ofmedical devices — Part 5:Tests for in vitrocytotoxicityUnder the conditions of the study, thedevice is not cytotoxicUnder the conditions of the study, the device wasfound to be cytotoxic and therefore the devicewere evaluated under ISO 10993-11 – Test for acutesystemic toxicity.From Acute Systemic Toxicity test, none of the testarticles were observed with signs of toxicity. Similarto predicate device
Primary Skin IrritationISO 10993-10:2010Biological evaluation ofmedical devices — Part 10:Tests for irritation and skinsensitizationUnder the conditions of the study, thedevice is not an irritantUnder the conditions of the study, the device is notan irritant. Similar to predicate device
DermalSensitizationISO 10993-10:2021Biological evaluation ofmedical devices — Part 10:Tests for skin sensitizationUnder the conditions of the study, thedevice is not a sensitizerUnder the conditions of the study, the device is nota sensitizer. Similar to predicate device
Acute SystemicToxicityISO 10993-11:2017Biological evaluation ofmedical devices —Part 11: Tests for systemictoxicityUnder the conditions of the study, thedevice does not pose a toxicityconcernUnder the conditions of the study, there was no signsof toxicity. Similar to predicate device

CLINICAL PERFORMANCE DATA:

Not applicable. There was no clinical data required to support the subject device as the indication for use is equivalent to the predicate device.

CONCLUSION:

The conclusions drawn from the non-clinical tests demonstrate that the subject device is as safe, as effective, and performs as well as or better than the legally marketed predicate device.

§ 878.4460 Non-powdered surgeon's glove.

(a)
Identification. A non-powdered surgeon's glove is a device intended to be worn on the hands of operating room personnel to protect a surgical wound from contamination. A non-powdered surgeon's glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls).