(47 days)
GS Medical's Pegasus-X Expandable PLIF System is indicated for use with autogenous bone graft in skeletally mature patients with degenerative disc disease (DDD) of the lumbar spine at one or two contiguous spinal levels from L2-S1. DDD is defined as discogenic pain with degeneration of the disc confirmed by patient history and radios. These patients should have had six months of non-operative treatment. These DDD patients may have up to Grade 1 spondylolisthesis at the involved level(s). The Pegasus-X Expandable PLIF Cage is to be combined with internal supplemental fixation cleared for use in the lumbar spine.
The Pegasus-X Expandable PLIF Cage devices are designed for restoring the height of the intervertebral space after resection of the disc. The Pegasus-X Expandable PLIF Cage devices consist of implants in various heights, footprints, and lordotic configurations with architecture designed to accept pre and post packing of bone graft material. The intervertebral body devices are made of titanium alloy (Ti-6AL-4V ELI). The Pegasus-X Expandable PLIF Cage devices are not radio lucent but have large windows through the sides that will allow visualization of the placement and subsequent bone fusion. The Expandable PLIF Cage (Posterior Lumbar Interbody Fusion Cage) are designed for segments L2 to S1.
The provided text is a 510(k) summary for the Pegasus-X Expandable PLIF System, a medical device for spinal fusion. It focuses on demonstrating substantial equivalence to already cleared predicate devices.
Crucially, this document does NOT contain information about acceptance criteria or a study proving the device meets those criteria in the context of an AI/ML-based medical device performance study.
The "Performance Data" section solely refers to mechanical and biomechanical testing of the physical implant itself (e.g., static and dynamic compression, expulsion, subsidence testing) against ASTM standards, not clinical performance or AI/ML algorithm performance.
Therefore, I cannot provide the requested information based on the given text.
To address your query regarding acceptance criteria and a study proving an AI/ML device meets them, the document would need to describe:
- Clinical performance metrics: Such as sensitivity, specificity, accuracy, F1-score, AUC for a diagnostic AI, or an effect size for an MRMC study if it's an AI-assisted tool.
- A clinical study design: Whether prospective or retrospective, the patient population, image acquisition protocols, and how ground truth was established for AI model evaluation.
- Details about the dataset: Sample size, data provenance (e.g., types of medical images, where they came from), and how it was split into training and test sets.
- Expert ground truth establishment: Number and qualifications of experts, and their adjudication methods.
Without such information, any attempt to answer your questions would be speculative and incorrect based only on the provided text.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath. The logo is simple and professional, and it is easily recognizable.
July 27, 2022
GS Medical Co. Ltd. Barry Sands President and Founder RQMIS Inc. 110 Haverhill Road, Suite 524 Amesbury, Massachusetts 01913
Re: K221687
Trade/Device Name: Pegasus-X Expandable PLIF System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: MAX Dated: June 8. 2022 Received: June 10, 2022
Dear Barry Sands:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K221687
Device Name Pegasus-X Expandable PLIF System
Indications for Use (Describe)
GS Medical's Pegasus-X Expandable PLIF System is indicated for use with autogenous bone graft in skeletally mature patients with degenerative disc disease (DDD) of the lumbar spine at one or two contiguous spinal levels from L2-S1. DDD is defined as discogenic pain with degeneration of the disc confirmed by patient history and radios. These patients should have had six months of non-operative treatment. These DDD patients may have up to Grade 1 spondylolisthesis at the involved level(s). The Pegasus-X Expandable PLIF Cage is to be combined with internal supplemental fixation cleared for use in the lumbar spine.
| Type of Use (Select one or both, as applicable) |
|---|
| ------------------------------------------------- |
| Prescription Use (Part 21 CFR 801 Subpart D) |
|---|
| ---------------------------------------------- |
| Over-The-Counter Use (21 CFR 801 Subpart C) |
|---|
| --------------------------------------------- |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) SUMMARY
GS Medical's Pegasus-X Expandable PLIF System
l. SUBMITTER'S ADDRESS, TELEPHONE NUMBER, CONTACT PERSON
Sam Camp Senior Engineer GS Medical Co. Ltd. 90 Osongsaengmyong 4-ro Osong-eup, Heungdeok-gu, Gheongju-si, Chungcheongbuk-do 28161 Korea Tel.: 405-535-2719 Email: scamp@gsmedicalusa.com
Contact Person:
Barry E. Sands RQMIS Inc. 110 Haverhill Road, Suite 524 Amesbury, MA 01913 Phone: (978) 358-7307 Email: barrysands@rqmis.com Facsimile: (978) 477-0206
Date Prepared: June 3, 2022
ll. SUBJECT DEVICE
| Trade/proprietary name of device: | Pegasus-X Expandable PLIF Cage |
|---|---|
| Common or Usual Name: | Intervertebral Body Fusion Device |
| Classification Name: | Intervertebral Body Fusion Device with Bone Graft, Lumbar |
| Regulation Number: | 21 CFR 888.3080 |
| Classification: | Class II |
| Product Code(s) | MAX |
{4}------------------------------------------------
III. PREDICATE DEVICES
Primary Predicate
| Name | Globus Rise Spacer |
|---|---|
| 510(k) number | K113447 |
| Product Code | MAX |
| CFR | 21 C.F.R. 888.3080 |
| Classification | Class II |
Additional Predicate
| Name | AnyPlus PEEK PLIF |
|---|---|
| 510(k) number | K100516 |
| Product Code | MAX |
| CFR | 21 C.F.R. 888.3080 |
| Classification | Class II |
IV. DEVICE DESCRIPTION
The Pegasus-X Expandable PLIF Cage devices are designed for restoring the height of the intervertebral space after resection of the disc. The Pegasus-X Expandable PLIF Cage devices consist of implants in various heights, footprints, and lordotic configurations with architecture designed to accept pre and post packing of bone graft material. The intervertebral body devices are made of titanium alloy (Ti-6AL-4V ELI). The Pegasus-X Expandable PLIF Cage devices are not radio lucent but have large windows through the sides that will allow visualization of the placement and subsequent bone fusion. The Expandable PLIF Cage (Posterior Lumbar Interbody Fusion Cage) are designed for segments L2 to S1.
V. INTENDED USE/INDICATIONS FOR USE
GS Medical's Pegasus-X Expandable PLIF Cage is indicated for use with autogenous bone graft in skeletally mature patients with degenerative disc disease (DDD) of the lumbar spine at one or two contiguous spinal levels from L2-S1. DDD is defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies. These patients should have had six months of non-operative treatment. These DDD patients may have up to Grade 1 spondylolisthesis at the involved level(s). The Pegasus-X Expandable PLIF Cage is to be combined with internal supplemental fixation cleared for use in the lumbar spine.
The intented use of the subject device and the predicates are identical.
{5}------------------------------------------------
VI. TECHNOLOGICAL CHARACTERISTICS
The subject device cages are provided in rectangular shape and have space to hold bone graft material. Upon implantation, the device can be expanded to any height within the allowed range as specified for each device description (cage combination). The expansion of the cage is achieved via the Torque Limiting Handle. The components are manufactured from Titanium Alloy (Ti-6Al-4V ELI) meeting the specifications of ASTM F136 Standard.
The Subject cage implants are available in various heights, footprints, and lordotic angles to suit the individual patient's pathology and anatomical conditions. The primary and additional predicates have rectagular shape, with various footprints, height and lordotic angles. Both the subject device, primary and additional predicate are rectangular in shape, have lordotic angles, and are available in various heights and widths. The shape, width, lordosis are withing the maximum primary predicate and additional predicate range. The highet expansion of the subject cage is within the range of the primary predicate highet expansion range. Since there is no significant difference in technological characteristcs between the subject and predicate devices, we conclude that the subject device is substantially equivalent to Primary and additional predicate.
VII. PERFORMANCE DATA
The worst-case cage of the Pegasus-X Expandable PLIF Cage system underwent testing according to ASTM 2077, specifically static and dynamic axial compression testing, shear static and dynamic compression, expulsion testing, and subsidence testing according to ASTM F2267. The results met all acceptance criteria, and the subject device cage is equivalent to additional predicate biomechanical performance.
CONCLUSION VIII.
The Pegasus-X Expandable PLIF Cage has been found to be substantially equivalent to the primary predicate with respect to technical characteristics, design, materials, and intended use. And to the additional predicate with respect to performance. The technological differences between the subject device and the predicate do not raise new questions of safety and effectiveness.
N/A