K Number
K221677

Validate with FDA (Live)

Date Cleared
2022-11-10

(154 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Synapse 3D Base Tools is medical imaging software that is intended to provide trained medical professionals with tools to aid them in reading, interpreting, and treatment planning. Synapse 3D Base Tools accepts DICOM compliant medical images acquired from a variety of imaging devices including, CT, MR, CR, US, NM, PT, and XA, etc. This product is not intended for use with or for the primary diagnostic interpretation of Mammography images.

Synapse 3D Base Tools provides several levels of tools to the user:

Basic imaging tools for general images, including 2D viewing, volume rendering and 3D volume viewing, orthogonal / oblique / curved Multi-Planar Reconstructions (MPR), Average (RaySum) and Minimum (MinIP) Intensity Projection, 4D volume viewing, image subtraction, surface rendering, sector and rectangular shape MPR image viewing, MPR for dental images, creating and displaying multiple MPR images along an object, time-density distribution, basic image processing, noise reduction, CINE, measurements, annotations, reporting, printing, storing, distribution, and general image management and administration tools, etc.

  • Tools for regional segmentation of anatomical structures within the image data, path definition through vascular and other tubular structures, and boundary detection.

-Image viewing tools for modality specific images, including CT PET fusion and ADC image viewing for MR studies. -Imaging tools for CT images including virtual endoscopic viewing and dual energy image viewing.

-Imaging tools for MR images including delayed enhancement image viewing, diffusion-weighted MRI image viewing.

Device Description

The 3D image analysis software Synapse 3D Base Tools (V6.6) is medical application software running on Windows server/client configuration installed on commercial general-purpose Windows-compatible computers. It offers software tools which can be used by trained professionals to interpret medical images obtained from various medical devices, to create reports, or to develop treatment plans.

Synapse 3D Base Tools is connected through DICOM standard to medical devices such as CT, MR, CR, US, NM, PT, XA, etc. and to a PACS system storing data generated by these medical devices, and it retrieves image data via network communications based on the DICOM standard. The retrieved image data are stored on the local disk managed by Synapse 3D Base Tools (V6.6), and the associated image-related information of the image data is registered in its database and is used for display, image processing, analysis, etc. Images newly created by Synapse 3D Base Tools (V6.6) not only can be display, but also can be printed on a hardcopy using a DICOM printer or a Windows printer.

Synapse 3D Base Tools (V6.6) is a basic software module that works with other cleared clinical applications, including Synapse 3D Cardiac Tools (K200973), Synapse 3D Perfusion Analysis (K162287), Synapse 3D Lung and Abdomen Analysis (K130542), Synapse 3D Liver and Kidney Analysis (K142521), Synapse 3D Nodule Analysis (K120679), Synapse 3D Colon Analysis (K123566), Synapse 3D Tensor Analysis (K141514) and Synapse 3D Blood Flow Analysis (K191544). All these software modules consist of the Synapse 3D product family.

Synapse 3D Base Tools can be integrated with Fujifilm's Synapse PACS, and can be used as a part of a Synapse system. Synapse 3D Base Tools also can be integrated with Fujifilm's Synapse Cardiovascular for cardiology purposes.

AI/ML Overview

The provided text does not contain detailed acceptance criteria and a study proving that the device meets these criteria in the typical format expected for a medical device with an AI/ML component affecting diagnostic accuracy. The device, "Synapse 3D Base Tools v6.6", is described as medical imaging software for viewing, interpreting, and planning, and its predicate device (Synapse 3D Base Tools v6.1) and other reference devices are also imaging software tools.

Instead, the submission focuses on demonstrating substantial equivalence to a predicate device, Synapse 3D Base Tools (V6.1) (K203103), by confirming similar indications for use and technical characteristics. The key performance information provided is about software development processes, verification and validation, and cybersecurity, rather than specific diagnostic performance metrics (e.g., sensitivity, specificity, AUC) for novel AI features impacting clinical decisions.

However, based on the information provided, particularly about the Dual Energy Analysis module and PixelShine, we can infer some aspects related to acceptance and performance.

Here's an analysis of the provided text in relation to your questions:

1. A table of acceptance criteria and the reported device performance

The document does not explicitly state a table of quantitative acceptance criteria for diagnostic performance metrics (e.g., sensitivity, specificity, accuracy) that the device must meet to be considered "accepted." This is likely because the device is primarily a "Medical image management and processing system" (21 CFR 892.2050) and not a primary diagnostic AI intended to make a diagnosis itself, but rather to provide "tools to aid them [trained medical professionals] in reading, interpreting, and treatment planning."

However, we can infer the acceptance for new features based on comparative testing for substantial equivalence:

Acceptance Criterion (Inferred from Substantial Equivalence Review)Reported Device Performance / Method of Proof
Overall Functionality and Design Specifications"Test results showed that all tests passed successfully according to the design specifications." "All of the different components... have been stress tested to ensure that the system as a whole provides all the capabilities necessary to operate according to its intended use and in a manner substantially equivalent to the predicate devices."
Dual Energy Image Viewing FeatureThe added "Dual Energy image viewing" feature in v6.6 is stated to be "the same as the feature available on the syngo.CT Dual Energy ('Reference Device'), which was cleared by the FDA under K133648." The module was "evaluated via comparative testing on patient data with the reference device syngo.CT Dual Energy (K133648)."
PixelShine FeatureThe "PixelShine" feature is "the embedded functionality of previously-cleared PixelShine (cleared by CDRH via K161625)." This implies that its integration into Synapse 3D Base Tools v6.6 maintains the already cleared performance of the standalone PixelShine.
Software Development Process AdherenceSoftware development plan, hazard analysis, risk management, requirements analysis, architectural design, detailed design, unit implementation/verification, integration testing, system testing, release, and maintenance processes were followed and described.
CybersecurityConfidentiality, integrity, and availability are maintained in accordance with FDA guidance (Section 6, October 2, 2014 guidance). DICOM standard communication assures adequate protection.
Adherence to Performance StandardsCompliance with DICOM Set (PS 3.1-3.20) (2016), IEC 62304 Ed 1.1 2015-06, and ISO 14971:2019.

2. Sample size used for the test set and the data provenance

The document mentions "actual clinical images" for benchmark performance testing and "patient data" for comparative testing of the Dual Energy Analysis module. However, it does not specify the sample size for these test sets, nor does it explicitly state the country of origin or whether the data was retrospective or prospective.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

The document does not provide any information regarding the number or qualifications of experts used to establish ground truth for any test set. Given the nature of the device as a processing and viewing tool, the "ground truth" for its functions would typically be the accuracy of the image processing (e.g., correct segmentation boundaries, accurate measurements compared to manual gold standard) rather than a clinical diagnosis.

4. Adjudication method for the test set

The document does not specify any adjudication method (e.g., 2+1, 3+1) for establishing ground truth or evaluating performance.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

The document does not mention an MRMC comparative effectiveness study directly assessing human reader improvement with or without AI assistance. The comparative testing described is for the Dual Energy Analysis module against a reference device's feature, not a human reader study. The device provides "tools to aid" professionals, implying assistance, but no study is presented to quantify this aid's effect on reader performance.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

The document states: "benchmark performance testing was conducted using actual clinical images to help demonstrate that the semi-automatic or automatic segmentation, detection, and registration functions implemented in Synapse 3D Base Tools achieved the expected accuracy performance." This indicates that standalone performance testing was conducted for these specific algorithmic functions. However, it does not provide specific metrics or results for these standalone algorithms.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The document does not explicitly state the type of ground truth. For "segmentation, detection, and registration functions," the ground truth would typically be manual delineation or measurement by experts (expert consensus or "gold standard" annotations), confirmed by visual inspection or perhaps clinical follow-up for specific detection tasks, but this is not detailed in the provided text.

8. The sample size for the training set

The document does not specify any training set sample size. This device is presented more as an update to an existing image processing software and an integration of already cleared functionalities (like PixelShine and the Dual Energy feature from K133648), rather than a new AI/ML algorithm that requires a detailed description of its training data.

9. How the ground truth for the training set was established

Since no training set information is provided, there is no information on how ground truth for any training set was established.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

FUJIFILM Corporation % Jeffrey Wan Manager, Regulatory Affairs FUJIFILM Healthcare Americas Corporation 81 Hartwell Avenue, Suite 300 LEXINGTON MA 02421

Re: K221677

November 10, 2022

Trade/Device Name: Synapse 3D Base Tools v6.6 Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: LLZ Dated: October 21, 2022 Received: October 21, 2022

Dear Jeffrey Wan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Jessica Lamb

Jessica Lamb, Ph.D. Assistant Director Imaging Software Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K221677

Device Name SYNAPSE 3D Base Tools v6.6

Indications for Use (Describe)

Synapse 3D Base Tools is medical imaging software that is intended to provide trained medical professionals with tools to aid them in reading, interpreting, and treatment planning. Synapse 3D Base Tools accepts DICOM compliant medical images acquired from a variety of imaging devices including, CT, MR, CR, US, NM, PT, and XA, etc. This product is not intended for use with or for the primary diagnostic interpretation of Mammography images.

Synapse 3D Base Tools provides several levels of tools to the user:

Basic imaging tools for general images, including 2D viewing, volume rendering and 3D volume viewing, orthogonal / oblique / curved Multi-Planar Reconstructions (MPR), Average (RaySum) and Minimum (MinIP) Intensity Projection, 4D volume viewing, image subtraction, surface rendering, sector and rectangular shape MPR image viewing, MPR for dental images, creating and displaying multiple MPR images along an object, time-density distribution, basic image processing, noise reduction, CINE, measurements, annotations, reporting, printing, storing, distribution, and general image management and administration tools, etc.

  • Tools for regional segmentation of anatomical structures within the image data, path definition through vascular and other tubular structures, and boundary detection.

-Image viewing tools for modality specific images, including CT PET fusion and ADC image viewing for MR studies. -Imaging tools for CT images including virtual endoscopic viewing and dual energy image viewing.

-Imaging tools for MR images including delayed enhancement image viewing, diffusion-weighted MRI image viewing.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW! *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

Date Prepared:November 7, 2022
Submitter's Information:FUJIFILM Corporation26-30 NISHIAZABU, 2-CHOMEMINATO-KU, TOKYO 106-8620
Contact Person:Jeffrey WanManager, Regulatory AffairsTelephone: (201) 675-8947Email: jeffrey.wan@fujifilm.com
Device Trade Name:Synapse 3D Base Tools
Device Common Names:Medical image management and processing system
Device Classification Name:System, Image Processing, Radiological
Product Code:LLZ
Regulation Number:21 CFR 892.2050
Device Class:Class II
Panel:Radiology
Predicate Devices:Synapse 3D Base Tools (K203103)FUJIFILM Corporation
Reference Devicessyngo.CT Dual Energy (K133648)Siemens Medical Solutions USA, Inc.PixelShine (K161625)AlgoMedica

{4}------------------------------------------------

1. Description of the Device

Synapse 3D Base Tools (V6.6) (this submission) is updated software of previously-cleared Synapse 3D Base Tools (V6.1) (cleared by CDRH via K203103 on February 9, 2021.

The 3D image analysis software Synapse 3D Base Tools (V6.6) is medical application software running on Windows server/client configuration installed on commercial general-purpose Windows-compatible computers. It offers software tools which can be used by trained professionals to interpret medical images obtained from various medical devices, to create reports, or to develop treatment plans.

Synapse 3D Base Tools is connected through DICOM standard to medical devices such as CT, MR, CR, US, NM, PT, XA, etc. and to a PACS system storing data generated by these medical devices, and it retrieves image data via network communications based on the DICOM standard. The retrieved image data are stored on the local disk managed by Synapse 3D Base Tools (V6.6), and the associated image-related information of the image data is registered in its database and is used for display, image processing, analysis, etc. Images newly created by Synapse 3D Base Tools (V6.6) not only can be display, but also can be printed on a hardcopy using a DICOM printer or a Windows printer.

Synapse 3D Base Tools (V6.6) is a basic software module that works with other cleared clinical applications, including Synapse 3D Cardiac Tools (K200973), Synapse 3D Perfusion Analysis (K162287), Synapse 3D Lung and Abdomen Analysis (K130542), Synapse 3D Liver and Kidney Analysis (K142521), Synapse 3D Nodule Analysis (K120679), Synapse 3D Colon Analysis (K123566), Synapse 3D Tensor Analysis (K141514) and Synapse 3D Blood Flow Analysis (K191544). All these software modules consist of the Synapse 3D product family.

Synapse 3D Base Tools can be integrated with Fujifilm's Synapse PACS, and can be used as a part of a Synapse system. Synapse 3D Base Tools also can be integrated with Fujifilm's Synapse Cardiovascular for cardiology purposes.

{5}------------------------------------------------

2. Indications for Use

Synapse 3D Base Tools is medical imaging software that is intended to provide trained medical professionals with tools to aid them in reading, interpreting, reporting, and treatment planning. Synapse 3D Base Tools accepts DICOM compliant medical images acquired from a variety of imaging devices including, CT, MR, CR, US, NM, PT, and XA, etc. This product is not intended for use with or for the primary diagnostic interpretation of Mammography images. Synapse 3D Base Tools provides several levels of tools to the user: Basic imaging tools for general images, including 2D viewing, volume rendering and 3D volume viewing, orthogonal / oblique / curved Multi-Planar Reconstructions (MPR), Maximum (MIP), Average (RaySum) and Minimum (MinIP) Intensity Projection, 4D volume viewing, image fusion, image subtraction, surface rendering, sector and rectangular shape MPR image viewing, MPR for dental images, creating and displaying multiple MPR images along an object, time-density distribution, basic image processing, noise reduction, CINE, measurements, annotations, reporting, printing, storing, distribution, and general image management and administration tools, etc.

  • Tools for regional segmentation of anatomical structures within the image data, path definition through vascular and other tubular structures, and boundary detection.
  • Image viewing tools for modality specific images, including CT PET fusion, ADC image viewing for MR studies.
  • Imaging tools for CT images including virtual endoscopic viewing and dual energy image viewing.
  • Imaging tools for MR images including delayed enhancement image viewing, diffusion-weighted MRI image viewing.

3. Substantial Equivalence Comparison

Synapse 3D Base Tools has the same intended use, similar labeling, and clinical application tools as those of the cleared predicate device Synapse 3D Base Tools (K203103). The device features and technical characteristics comparison with predicates is shown as Table 1 Device Features and Technical Characteristics Comparison Matrix.

{6}------------------------------------------------

Device ParametersSynapse 3DBase Tools(V6.6)(Thissubmission)Synapse 3D BaseTools(V6.1)(K203103)(Primarypredicate device)Comparison
Classification NameSystem, ImageProcessing,RadiologicalSystem, ImageProcessing,RadiologicalSame
Regulatory Number892.2050892.2050Same
Product CodeLLZLLZSame
ClassificationClass IIClass IISame
Review PanelRadiologyRadiologySame
2D ViewingYesYesSame
ImageStoring(DICOM SCP)YesYesSame
Image Communication(DICOM SCU)YesYesSame
DICOMInterface(SCP/SCU)YesYesSame
Printing(DICOMSCU)YesYesSame
Measurements (2D and3D)YesYesSame
Annotations -Standardized and FreeTextYesYesSame
ReportingYesYesSame
CineYesYesSame
Volume Rendering and3D ViewingYesYesSame
MPR- orthogonal /oblique / curvedMulti-PlanarReconstructions(MPR),- Sector andrectangular shapeMPR imageYesYesSame
Device ParametersSynapse 3DBase Tools(V6.6)(Thissubmission)Synapse 3D BaseTools(V6.1)(K203103)(Primarypredicate device)Comparison
MPR for dentalimagesMultiple MPRimages along anobject (Slicer)YesYes
Maximum, Average,Minimum IntensityProjectionYesYesSame
4D viewingYesYesSame
Image fusionYesYesSame
Surface renderingYesYesSame
Image subtraction (3D)YesYesSame
Time-densitydistributionYesYesSame
General image datamanagement andadministration toolsYesYesSame
SegmentationYesYesSame
Path definitionYesYesSame
Boundary detectionYesYesSame
CT PET fusionYesYesSame
ADC image viewing(MRI)YesYesSame
Virtual EndoscopicSimulatorYesYesSame
Diffusion-weightedMRI Data AnalysisYesYesSame
Delayed EnhancementImage ViewingYesYesSame
Dual Energy imageYesNoAdded application of dual
Device ParametersSynapse 3DBase Tools(V6.6)(Thissubmission)Synapse 3D BaseTools(V6.1)(K203103)(Primarypredicate device)Comparison
viewingenergy image viewing.Note: The dual energyimage viewing feature isthe same as the featureavailable on the syngo.CTDual Energy ("ReferenceDevice"), which wascleared by the FDA underK133648. Therefore, thisadded feature does notraise different questions ofsafety and effectiveness.
PixelShineYesNoThis feature is theembedded functionality ofpreviously-clearedPixelShine (cleared byCDRH via K161625).
Product AvailabilitySoftwareProductSoftware ProductSame
Hardware PlatformWindows PCWindows PCSame

Table 1 Device Features and Technical Characteristics Comparison Matrix

{7}------------------------------------------------

{8}------------------------------------------------

4. Safety Information

Synapse 3D Base Tools introduces no new safety or efficacy issues other than those already identified with the predicate devices. As part of the Risk Management process, appropriate preventive measures in response to the results of the Hazard Analysis have been taken in accordance with the November 4, 2021 issue of the "DRAFT Guidance for the Content of Premarket Submissions for Device Software Functions." The Synapse 3D Base Tools labeling contains instructions for use and necessary cautions, warnings and notes to provide the safe and effective use of the device.

5. Testing and Performance Information

Nonclinical testing result:

The purpose of Software Development Process for Synapse 3D Base Tools is to carry out the activities relating to the establishment of the software development plan (or plans) for definitely conducting software hazard analysis, risk management, requirement analysis, architectural

{9}------------------------------------------------

design, the design specification, unit implementation and verification, software integration and integration testing, software system test, software release, software maintenance. The main activities in software development process are described as follows.

  • . Software development plan
  • Software hazard analysis and risk management .
  • Software requirements analysis/specification ●
  • Software architectural design ●
  • Software detailed design specification ●
  • Software unit module implementation and verification
  • Software integration and system testing ●

Clinical tests:

The subject of this 510(k) notification, Synapse 3D Base Tools does not require clinical studies to support safety and effectiveness of the software.

Verification and Validation:

Testing for verification and validation involved system level functionality test, component testing, verification testing, integration testing, usability testing, installation/upgrade testing, labeling testing, as well as the testing for risk mitigations associated with the risk management process. In addition, benchmark performance testing was conducted using actual clinical images to help demonstrate that the semi-automatic or automatic segmentation, detection, and registration functions implemented in Synapse 3D Base Tools achieved the expected accuracy performance. Pass/Fail criteria were based on the requirements and intended use of the product. Test results showed that all tests passed successfully according to the design specifications. All of the different components of the Synapse 3D Base Tools software have been stress tested to ensure that the system as a whole provides all the capabilities necessary to operate according to its intended use and in a manner substantially equivalent to the predicate devices. The Dual Energy Analysis module was evaluated via comparative testing on patient data with the reference device syngo.CT Dual Energy (K133648).

Cybersecurity:

The confidentiality, integrity and availability are maintained by Synapse 3D Base Tools in accordance with Section 6 of the Content of Premarket Submissions for Management of Cybersecurity in Medical Devices; Guidance for Industry and Food and Drug Administration Staff (October 2, 2014).

Synapse 3D Base Tools is connected through DICOM standard to medical devices and to a PACS system storing data generated by these medical devices, and it retrieves image data via network communication based on the DICOM standard. Therefore Synapse 3D Base Tools assures an adequate degree of protection for cybersecurity.

{10}------------------------------------------------

Performance standards:

  • Digital Imaging and Communications in Medicine (DICOM) Set (PS 3.1 3.20) (2016). .
  • IEC 62304 Edition 1.1 2015-06, Medical Device Software Software Life Cycle Processes.
  • ISO 14971:2019 2019-12-10, Medical Devices Application of Risk Management to Medical Devices.

6. Conclusion

Performance tests were conducted to test the functionality of the subject device, Synapse 3D Base Tools. Results of all conducted testing were acceptable in supporting the claim of substantial equivalence.

N/A