Freestyle Lancing Device II, Autolet, Autolet Lite, Unilet Lancets
K221613 · Owen Mumford, Ltd. · QRL · Oct 3, 2022 · General, Plastic Surgery
Device Facts
Record ID
K221613
Device Name
Freestyle Lancing Device II, Autolet, Autolet Lite, Unilet Lancets
Applicant
Owen Mumford, Ltd.
Product Code
QRL · General, Plastic Surgery
Decision Date
Oct 3, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4850
Device Class
Class 2
Indications for Use
Freestyle Lancing Device II: Use with compatible lancets for capillary blood sampling. Autolet: Use with compatible lancets for capillary blood sampling. Autolet Lite: Use with compatible lancets for capillary blood sampling. Unilet Lancets: Use with compatible lancing device for capillary blood sampling.
Device Story
Hand-held, reusable, spring-driven lancing devices (Freestyle Lancing Device II, Autolet, Autolet Lite) and associated single-use sterile lancets (Unilet Lancets). Devices used in non-clinical environments by patients for capillary blood sampling. User inserts lancet, adjusts penetration depth via dial or cap, and triggers spring-loaded mechanism to puncture skin (finger or alternate sites). Lancets are disposed of after single use; devices are cleaned with isopropyl alcohol or soap. Output is a capillary blood sample for diagnostic testing. Benefits include frequent, hygienic blood glucose monitoring for diabetes management.
Clinical Evidence
Bench testing only. Design verification testing confirmed all devices met pre-determined specifications. Biocompatibility assessed per ISO 10993-1:2009 and ISO 10993-5:2009; all materials scored 0 (non-cytotoxic). Sterilization of lancets validated per ISO 11137.
Technological Characteristics
Spring-driven, hand-held, reusable lancing devices. Penetration depth adjustable (0.5mm-2.5mm). Lancets provided sterile via gamma irradiation (25-50 kGy, SAL 10^-6). Materials assessed for biocompatibility per ISO 10993. No electronic components or software.
Indications for Use
Indicated for capillary blood sampling in patients requiring frequent blood glucose monitoring, such as diabetics. For single-patient, home-use only.
Regulatory Classification
Identification
The regulation covers four types of blood lancets: (a) Single use only blood lancet with an integral sharps injury prevention feature; (b) Single use only blood lancet without an integral sharps injury prevention feature; (c) Multiple use blood lancet for single patient use only; and (d) Multiple use blood lancet for multiple patient use. All are defined as disposable or reusable devices comprised of a blade attached to a base used to puncture the skin to obtain a drop of blood for diagnostic purposes.
Special Controls
*Classification.* Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that the structure and material composition are consistent with the intended use and must include a sharps injury prevention feature.
(ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use and that the integral sharps injury prevention feature will irreversibly disable the device after one use.
(iii) The device must be demonstrated to be biocompatible.
(iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin (
*e.g.,* blade).(v) Labeling must include:
(A) Detailed descriptions, with illustrations, of the proper use of the device and its sharps injury prevention feature.
(B) Handwashing instructions for the user before and after use of the device.
(C) Instructions on preparation (
*e.g.,* cleaning, disinfection) of the skin to be pierced.(D) Instructions for the safe disposal of the device.
(E) Labeling must be appropriate for the intended use environment.
(
*1* ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.(
*2* ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vi) Labeling must also include the following statements, prominently placed:
(A) “For use only on a single patient. Discard the entire device after use.”
(B) “Warning: Not intended for more than one use. Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested.”
(b)
*Single use only blood lancet without an integral sharps injury prevention feature* —(1)*Identification.* A disposable blood lancet intended for a single use that is comprised of a single use blade attached to a solid, non-reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2)
*Classification.* Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that the structure and material composition are consistent with the intended use and address the risk of sharp object injuries and bloodborne pathogen transmissions.
(ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use.
(iii) The device must be demonstrated to be biocompatible.
(iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin (
*e.g.,* blade).(v) Labeling must include:
(A) Detailed descriptions, with illustrations, of the proper use of the device.
(B) Handwashing instructions for the user before and after use of the device.
(C) Instructions on preparation (
*e.g.,* cleaning, disinfection) of the skin to be pierced.(D) Instructions for the safe disposal of the device.
(E) Labeling must be appropriate for the intended use environment.
(
*1* ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.(
*2* ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vi) Labeling must also include the following statements, prominently placed:
(A) “For use only on a single patient. Discard the entire device after use.”
(B) “Warning: Not intended for more than one use. Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested.”
(c)
*Multiple use blood lancet for single patient use only* —(1)*Identification.* A multiple use capable blood lancet intended for use on a single patient that is comprised of a single use blade attached to a solid, reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2)
*Classification.* Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that:
(A) The lancet blade can be changed with every use, either manually or by triggering a blade storage unit to discard the used blade and reload an unused blade into the reusable base; and
(B) The structure and material composition are consistent with the intended use and address the risk of sharp object injuries and bloodborne pathogen transmissions and allow for validated cleaning and disinfection.
(ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use.
(iii) The device must be demonstrated to be biocompatible.
(iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin (
*e.g.,* blade).(v) Validation testing must demonstrate that the cleaning and disinfection instructions are adequate to ensure that the reusable lancet base can be cleaned and low level disinfected.
(vi) Labeling must include:
(A) Detailed descriptions, with illustrations, of the proper use of the device.
(B) The Environmental Protection Agency (EPA) registered disinfectant's contact time for disinfectant use.
(C) Handwashing instructions for the user before and after use of the device.
(D) Instructions on preparation (
*e.g.,* cleaning, disinfection) of the skin to be pierced.(E) Instructions on the cleaning and disinfection of the device.
(F) Instructions for the safe disposal of the device.
(G) Instructions for use must address the safe storage of the reusable blood lancet base between uses to minimize contamination or damage and the safe storage and disposal of the refill lancet blades.
(H) Labeling must be appropriate for the intended use environment.
(
*1* ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.(
*2* ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vii) Labeling must also include the following statements, prominently placed:
(A) “For use only on a single patient. Disinfect reusable components according to manufacturer's instructions between each use.”
(B) “Used lancet blades must be safely discarded after a single use.”
(C) “Warning: Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested. The cleaning and disinfection instructions for this device are intended only to reduce the risk of local use site infection; they cannot render this device safe for use for more than one patient.”
(d)
*Multiple use blood lancet for multiple patient use* —(1)*Identification.* A multiple use capable blood lancet intended for use on multiple patients that is comprised of a single use blade attached to a solid, reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2)
*Classification.* Class III (premarket approval).(3)
*Date PMA or notice of completion of a PDP is required:* A PMA or a notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before May 22, 2024, for any multiple use blood lancet for multiple patient use described in paragraph (d)(1) of this section that was in commercial distribution before May 28, 1976, or that has, on or before May 22, 2024, been found to be substantially equivalent to a multiple use blood lancet for multiple patient use described in paragraph (d)(1) of this section that was in commercial distribution before May 28, 1976. Any other multiple use blood lancet for multiple patient use shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
K222617 — EasyTouch Lancing Device · Stat Medical Devices · Mar 21, 2023
K250813 — MICROLET®NEXT 2 Lancing Device · Ascensia Diabetes Care U.S., Inc. · May 14, 2025
K221175 — Multi-Lancet Device 2, ReliOn Premier Lancing Device · Arkray, Inc. · Oct 17, 2022
K222034 — RightLance Blood Lancing System · Beijing Ruicheng Medical Supplies Co., Ltd. · Sep 7, 2022
K223099 — Facet Manatee Reusable Lancing Base · Facet Technologies, LLC · Nov 28, 2022
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
10/3/2022
Owen Mumford Ltd % Patty Cronan Quality Manager Owen Mumford USA Inc. 1755 West Oak Commons Ct. Marietta, Georgia 30062
Re: K221613
Trade/Device Name: Freestyle Lancing Device II, Autolet, Autolet Lite, Unilet Lancets Regulation Number: 21 CFR 878.4850 Regulation Name: Blood Lancets Regulatory Class: Class II Product Code: QRL, QRK Dated: Mav 30, 2022 Received: June 3, 2022
Dear Patty Cronan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
#### 510(k) Number (if known) K221613
Device Name
Freestyle Lancing Device II, Autolet, Autolet Lite, Unilet Lancets
Indications for Use (Describe)
Freestyle Lancing Device II: Use with compatible lancets for capillary blood sampling.
Autolet: Use with compatible lancets for capillary blood sampling.
Autolet Lite: Use with compatible lancets for capillary blood sampling.
Unilet Lancets: Use with compatible lancing device for capillary blood sampling.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# SECTION 5.0
# 510(k) SUMMARY
# 1.Submitter
| Prepared by: | Matthew Willoughby<br>Regulatory Affairs Officer<br>Owen Mumford Ltd.<br>Tel: +44(0)1993 812021<br>Fax: +44(0)1993 813466 |
|-----------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Prepared for: | Owner/ Operator<br>Owen Mumford Ltd<br>Brook Hill<br>Woodstock<br>Oxfordshire<br>OX20 1TU<br>United Kingdom<br>Establishment Registration Number: 3003348846 |
| Contact Person: | Darren Mansell<br>Regulatory Affairs Manager<br>Owen Mumford Ltd,<br>Tel: +44(0)1993 812021<br>Fax: +44(0)1993 813466<br>Email: darren.mansell@owenmumford.com<br>Paul Smith<br>Head of Quality<br>Owen Mumford Ltd,<br>Tel: +44(0)1993 812021<br>Fax: +44(0)1993 813466<br>Email:paul.smith@owenmumford.com |
| Date Prepared: | 16 AUG 2022 |
Section 5-1
{4}------------------------------------------------
# 2. Device
| Name of Device: | Freestyle Lancing Device II |
|--------------------------|-------------------------------------------------------|
| | Autolet |
| | Autolet Lite |
| | Unilet Lancets |
| Common Name: | Home-use Lancing Device |
| Classification Name: | Multiple Use Blood Lancet for Single Patient Use Only |
| Product Code: | QRL |
| Subsequent Product Code: | QRK |
| FDA Classification: | Class II |
#### 3.Predicate Devices
Predicate Device Name: Accu-Chek Softclix Blood Lancing System, 510(k) number K214022.
#### 4. Description of The Device
The purpose of this 510(k) Premarket Notification is to obtain FDA clearance for the following devices:
- . Freestyle Lancing Device II
- . Autolet
- . Autolet Lite
- Unilet Lancets
The intended use for these devices remains the same as the predicate device.
The Autolet design is an Owen Mumford branded variant of the "Freestyle Lancing Device II". The sole difference is the moulded branding on the outer of the rear body component, the component is functionally identical and made of the same material. Therefore, much of the information for the Freestyle Lancing Device II and Autolet devices is the same and they can be considered variants of the same device.
For brevity, in the sections within this submission pertaining to both the "Freestyle Lancing Device II" and "Autolet" devices, they will be referred to as the "FS & A Device". When information provided is specific to one variant, its full name will be used. Additionally, any reference within the documentation referring to 'Perceval' can be considered as pertaining to the FS & A Device, as the original design work was performed under the project name Perceval.
The FS & A Device is a hand-held non-sterile, reusable lancing device intended for single patient capillary blood sampling in non-clinical environments. The ADC Freestyle Lancing Device II variant is sold for use with ADC 'Freestyle' single-use lancets; these are not included in this submission. The OM Autolet variant is sold for use with Unilet Lancets and as 'universally compatible with most lancets'. The lancets are disposed of after each use and the device is maintained by using isopropyl alcohol or soap to wipe the outside of the device. The FS & A Device is for use only on a single patient.
The FS & A Device is intended to be used by patients with medical conditions who require frequent capillary blood sampling, such as diabetics. Frequency of use will depend greatly on the patients need for using the device, although typically diabetics may monitor their blood qlucose levels several times per day. The primary site for lancing is the finger, although an 'alternate site adapter' is provided, so lancing can be performed on sites other than the fingers as necessary.
{5}------------------------------------------------
The Autolet Lite is a hand-held non-sterile, reusable lancing device intended for single patient capillary blood sampling in non-clinical environments. The device was designed for use as a system with Unilet Lancets (and as 'universally compatible with most lancets') which are included in this 510(k) The Unilet Lancets feature 3 lancet gauges, 3 with labelling "Unilet Lancets" and the same 3 lancet gauges labelled "Unilet ComforTouch Lancets". These lancets have different gauge sizes and styling but are functionally identical and for the purpose of this 510(k), all references to "Unilet Lancets" can be considered to apply to Unilet Lancets and Unilet ComforTouch Lancets.
### 5.Indications for Use
Freestyle Lancing Device II: Use with compatible lancets for capillary blood sampling.
Autolet: Use with compatible lancets for capillary blood sampling.
Autolet Lite: Use with compatible lancets for capillary blood sampling.
Unilet Lancets: Use with compatible lancing device for capillary blood sampling.
# 6. Technological Characteristics
The Autolet, Freestyle Lancing Device II device and the Autolet Lite and Unilet Lancets are substantially equivalent to the predicate device, the Accu-Chek Softclix Blood Lancing System, 510(k) number K214022.
A comparison of the intended uses and technological characteristics is summarized in the table 5.1 below.
{6}------------------------------------------------
0(k) Number: K221
Autolet Lite is a hand-held inon-held inoningle parting along managing memallian inglang
ing dimination manusias manus and sanar lingin and analin
garad by usaint minuman m anged from the predicate utolet Lite and Unilet Lance changed from the predicate dev Submission Device – FS
Device he Aoou-Chek Softclix Lancing Development
Sansatible Accu-Check Scou-Check Shorts Comments
Times to optrain and constitution from the Archives
Tingers to option the Alexand ( Accu-Ohek Softolix Blood Land
on is intended for the differ the children of the more the more the more of the more of the man in the comment of the supper and the upper and
e Accu-Chek Softclix Blood Lanci
System Cap. evice Characteristi evice Descripti ntended Us
le 5.1 Comparison of the intended uses and technological characteristics between the predicate and submission dev
{7}------------------------------------------------
| 1613<br>て<br>K2<br>Number: |
|----------------------------|
| |
| 0(k) |
| ਦੇ ਹ |
| Device Characteristic | Accu-Chek Softclix Blood Lancing System | Submission Device - FS & A Device | Autolet Lite and Unilet Lancets |
|-----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | The Accu-Chek Softclix Blood Lancing System is intended for the hygienic collection of capillary blood for testing purposes from the side of a fingertip and from alternative sites, such as the palm, the upper arm, and the forearm. | Freestyle Lancing Device II: Use with compatible lancets for capillary blood sampling.<br><br>Autolet: Use with compatible lancets for capillary blood sampling. | Autolet Lite: Use with compatible lancets for capillary blood sampling.<br><br>Unilet Lancets: Use with compatible lancing device for capillary blood sampling. |
| | The sterile, single-use lancets are to be used with the resuseable lancing device that is to be cleaned and disinfected between each use, and then the lancets are to be disposed of. | | |
| | This system if for use only on a single patient in a home setting. | | |
| | This system is not suitable for use by healthcare professionals with multiple patients in a healthcare setting. | | |
| Number of Uses | Base (lancing device): multiple useLancet: single use | Base (lancing device): multiple useLancet: single use | Base (lancing device): multiple useLancet: single use |
| Depth Adjustment | 11 levels by twisting cap | Levels 1-5, with half settings for a total of 9 different settings, set by twisting dial. | 10 different settings, set by twisting cap |
| Depth of Penetration | Minimum depth of penetration: 0.8mm<br>Maximum depth of penetration: 2.3mm | Minimum depth of penetration: 0.5mm<br>Maximum depth of penetration: 2.5mm | Minimum depth of penetration: 0.60mm<br>Maximum depth of penetration: 1.95mm |
{8}------------------------------------------------
| 13<br>ਹਵ<br>ਟ<br>K2 |
|---------------------|
| er:<br>Numb |
| 10(k)<br>ம் |
| Device Characteristic | Accu-Chek Softclix Blood Lancing<br>System | Submission Device – FS & A<br>Device | Autolet Lite and Unilet Lancets |
|-----------------------|---------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|
| Load and firing | Load by pressing priming button when lancet is inserted Fire by pressing the release button | Load by pulling back grey slider until it clicks (after lancet has been inserted) Fire by pressing the lancing button | Load by inserting the lancet into the lancet holder and pushing down until fully seated Fire by pressing the release button |
| Mechanical Loading | Spring-driven | Spring-driven | Spring-driven |
| Lancet Sterility | Yes, gamma irradiation | Yes, gamma irradiation | Yes, Gamma irradiation |
| Lancet Dimensions | 28G, 0.4mm wide, beveled cut with 3 facets. Needle length not known. | N/A – this column covers the devices only. | Unilet Lancets: 23G 0.64mm wide, 28G 0.36mm wide, 30G 0.31mm wide<br>All lancets have the same needle length 3.2mm (+-0.3mm) |
| | | | Needle drawings can be found in section 11 – Device Description, of this submission. |
{9}------------------------------------------------
# 7.Performance Data
#### Non-clinical performance data:
Design verification testing of the FS & A has been carried out to evaluate the performance of the devices against defined acceptance criteria.
The following table provides a summary of the relevant design verification testing.
#### Bench Testing:
The objective of the design verification testing conducted was to verify that the submission devices met the pre-determined specifications, to support the conclusion that they are fit for purpose and are considered safe and effective for the intended use.
| Submission Device | Design Verification Testing | Results |
|-------------------|--------------------------------------------------|----------------------------------------|
| FS & A Device | Yes – 20 Design Input IDs<br>Tested and Verified | Passed all testing and<br>verification |
| Autolet Lite | Yes – 12 Design Inputs<br>Tested and Verified | Passed all testing and<br>verification |
| Unilet Lancets | Yes – 12 Design Inputs<br>Tested and Verified | Passed all testing and<br>verification |
#### Biocompatibility:
All patient contact materials of the FS & A Device were assessed against the requirements of ISO 10993-1:2009. All skin contact materials used in the device were tested for cytotoxicity in accordance with ISO 10993-5:2009. These results along with base material and master-batch data sheets, the proposed material application, the device assembly process and any examples of current usage were used by the Toxicologist (Medwise International Consultancy Limited, York UK) to assess the suitability of the materials for their intended use and provide a biocompatibility evaluation. All contact materials were found to be non-cytotoxic according to the test method stipulated in ISO 10993-5:2009: Each material scored 0; the lowest level achievable. Expert toxicological review for the materials concluded that "it is inconceivable that transient daily skin contact with the device would result in any incidence of irritation or sensitization" to users and therefore the requirements of ISO 10993 are met without the need for further animal testing.
Results of the testing and toxicological assessment demonstrate the materials of the FS & A Device are biocompatible and do not pose a biological risk to the user, therefore, safe for use.
#### Sterilization:
Unilet Lancets are the only device in this submission that is provided sterile.
Sterilization Method: Gamma radiation
Radiation dose range: 25-50 kGy
Sterility assurance level (SAL): 10-6
{10}------------------------------------------------
#### Description of the method used to validate the sterilization cycle:
Sterilization validation (Dose mapping) was conducted in accordance with Section 9 of ISO 11137: "Sterilization of Healthcare Products - Radiation - Part 1: Requirements for Development, Validation and Routine Control of a Sterilization Process for Medical Devices."
The verification dose for the Unilet Lancets manufactured by Tianjin Huahong Technology Co., Ltd is 6.5kGy and the minimum exposure dose has been determined as 25kGy.
#### Shelf life
The FS & A Device is designed to last for 2 years as per the design verification for the device.
Unilet Lite is designed to last for no less than 3000 complete uses.
Unilet Lancets have a sterile shelf life of 5 years.
### 8.Conclusion
In summary, the differences between the FS & A Device, Autolet Lite and Unilet Lancets and the predicate have no impact on safety and effectiveness and the products are therefore substantially equivalent to the predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.