Intended Use

The NaviCam Small Bowel Capsule Endoscopy System is intended for visualization of the small bowel mucosa in adults. It may be used as a tool in the detection of abnormalities of the small bowel.

Device Description

The NaviCam Small Bowel Capsule Endoscopy System is an endoscopic capsule imaging system intended to obtain images of the small bowel. It is comprised of the following components:

  1. Capsule (AKES-11SW, AKES-11SI): The disposable, ingestible NaviCam Small Bowel Capsule is designed to acquire video images during the natural propulsion through the GI tract. The capsule transmits the acquired images via an RF communication channel to the NaviCam Data Recorder located outside the body.
  2. Data recorder (AKR-1, AKRI-1): The Data Recorder is an external receiving and recording unit that receives and stores the acquired images from the capsule.
  3. ESView Software: The ESView is a software application for processing, analyzing, storing, and viewing the acquired images collected from the NaviCam Data Recorder to create a video of the images. The software also includes a reporting function to create detailed clinical reports and a capsule endoscopy atlas.
  4. Locator: The Locator is a handheld device that is used to turn the NaviCam Capsule on. It is also used for determining if the capsule is still in the body when the patient is not sure whether he/she expelled it.
AI/ML Overview

The NaviCam Small Bowel Capsule Endoscopy System was assessed for its performance primarily through a comparative clinical study against a predicate device, the PillCam SB3 Capsule Endoscopy System, and also through various bench/in-vitro tests.

Here's a breakdown of the acceptance criteria and study details:

1. Table of Acceptance Criteria and Reported Device Performance

Test TypeAcceptance Criteria (Stated or Implied)Reported Device Performance
Bench/In-Vitro TestsSuccessfully passed all listed tests.
Biting TestAbility to withstand applied forces similar to accidental biting.Pass
Angular Resolution TestMeasurement of MTF using ISO 12233 slanted edge methodology and new angular resolution method using LEDs. (Specific thresholds for "pass" are not provided in the document but implied successful meeting of design specifications.)Pass
Temperature Safety TestTemperature change during operation within safe limits. (Specific thresholds for "pass" are not provided in the document but implied successful meeting of design specifications.)Pass
pH TestIntegrity of the capsule during exposure to simulated extreme pH levels. (Specific thresholds for "pass" are not provided in the document but implied successful meeting of design specifications.)Pass
Image Intensity UniformityUniformity of image intensity. (Specific thresholds for "pass" are not provided in the document but implied successful meeting of design specifications.)Pass
Image Frame Rate TestHigher frame rate provides good transmission property. (Specific thresholds for "pass" are not provided in the document but implied successful meeting of design specifications.)Pass
Geometric Distortion TestDetermination of geometric distortion and local magnification. (Specific thresholds for "pass" are not provided in the document but implied successful meeting of design specifications.)Pass
Field of View (FOV) TestDetermination of FOV value. (Specific thresholds for "pass" are not provided in the document but implied successful meeting of design specifications.)Pass
Battery Life TestBattery life of at least 8 hours and capturing over 57,500 images.Pass (demonstrated to last at least 8 hours and capture over 57,500 images).
Image Resolution TestTesting of image resolution. (Specific thresholds for "pass" are not provided in the document but implied successful meeting of design specifications.)Pass
Magnetic Field TestMeasurement of magnetic flux density on capsule surface and non-optical bottom, and determination of safety distance. (Specific thresholds for "pass" are not provided in the document but implied successful meeting of design specifications.)Pass
DOV TestMeasurement of MTF in air and underwater at different distances within claimed DOV range using ISO 12233 slanted edge methodology and angular resolution method. (Specific thresholds for "pass" are not provided but implied successful meeting of design specifications.)Pass
Color and Gray Scale TestEvaluation of optical performance. (Specific thresholds for "pass" are not provided but implied successful meeting of design specifications.)Pass
Data Integrity TestData transmission between capsule, data recorder, and ESView software. (Specific thresholds for "pass" are not provided but implied successful meeting of design specifications.)Pass
Clinical Study
Diagnostic Overall Percent Agreement with predicate deviceNot explicitly stated as a strict threshold, but achieving substantial equivalence to the predicate device (PillCam SB3) in diagnostic performance. The document implies that the observed agreement rate demonstrates similarity.89.66% (81.50%, 94.46%) overall percent agreement with the PillCam SB3. Overall percent agreement Kappa of 0.6652 (0.4653, 0.8652). This was deemed to demonstrate similar performance and effectiveness.

2. Sample Size Used for the Test Set and Data Provenance

The document mentions that the clinical study was a prospective study (NCT05086471).
However, the specific sample size (number of patients or cases) used for the test set is not provided in the given text.
The data provenance (e.g., country of origin) for the clinical study is also not stated.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

This information is not provided in the given text.

4. Adjudication Method for the Test Set

This information is not provided in the given text. The text only states that the NaviCam system was compared to the PillCam SB3 in terms of diagnostic performance, but it doesn't detail how discrepancies or consensus was handled for establishing ground truth or comparing diagnostic findings.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size of Human Readers Improvement with AI vs Without AI Assistance

A comparative clinical study was performed, but it was to compare the device's performance (NaviCam Small Bowel Capsule Endoscopy System) to a predicate device (PillCam SB3 Capsule Endoscopy System), not specifically an MRMC study assessing human reader improvement with or without AI assistance. The study evaluates the diagnostic agreement between the two capsule endoscopy systems. Therefore, the effect size of human readers improving with AI vs. without AI assistance is not applicable/not reported in this context.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

The NaviCam Small Bowel Capsule Endoscopy System includes "ESView Software...for processing, analyzing, storing, and viewing the acquired images... The software also includes a reporting function". While the software performs analysis, the clinical study appears to evaluate its overall diagnostic performance in detecting abnormalities, which would typically involve human review of the generated images/reports. The text does not explicitly state if a standalone algorithm-only performance study was conducted separate from human interpretation. The reported "diagnostic Overall Percent Agreement" is for the system, which implies the combined interpretation of the images.

7. The Type of Ground Truth Used

The ground truth for the clinical study is based on the detection of abnormalities of the small bowel by both the NaviCam system and the predicate PillCam SB3 system. The phrase "diagnostic Overall Percent Agreement" implies that the agreement was measured against the findings of another diagnostic tool (the predicate device), which often serves as a form of "ground truth" in equivalence studies when a gold standard (like pathology) is not universally available for every finding. The document does not explicitly state that pathology or outcomes data were used as the definitive ground truth for every finding. It strongly suggests the predicate device's findings were used as the reference point for comparison.

8. The Sample Size for the Training Set

This information is not provided in the given text. The document focuses on the performance study data, not the training dataset for any underlying AI/software components.

9. How the Ground Truth for the Training Set Was Established

This information is not provided in the given text, as details about a training set are absent.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 2, 2022

ANKON Technologies Co., Ltd. % Shoshana Friedman Senior Regulatory Affairs Consultant ProMedoss Inc 3521 Hatwynn Road Charlotte, NC 28269

Re: K221590

Trade/Device Name: NaviCam Small Bowel Capsule Endoscopy System Regulation Number: 21 CFR 876.1300 Regulation Name: Ingestible telemetric gastrointestinal capsule imaging system Regulatory Class: Class II Product Code: NEZ Dated: November 2, 2022 Received: November 2, 2022

Dear Shoshana Friedman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Shanil P. Haugen -S

Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K221590

Device Name

NaviCam Small Bowel Capsule Endoscopy System

Indications for Use (Describe)

The NaviCam Small Bowel Capsule Endoscopy System is intended for visualization of the small bowel mucosa in adults. It may be used as a tool in the detection of abnormalities of the small bowel.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Ankon. The logo features a stylized, light blue, rounded shape resembling a 'U' with a small vertical line inside, suggesting a power button symbol. Below the symbol is the word "ankon" in a bold, sans-serif font, with the tagline "INNOVATION FOR BETTER HEALTH" in smaller letters underneath.

510(K) SUMMARY NaviCam Small Bowel Capsule Endoscopy System 510(k) NumberK221590

SUBMITTER 1.

Applicant's Name:

ANKON Technologies Co., Ltd. B3-2, B3-3, D3-4 Biolake, No.666, Hi-Tech Road, East Lake New Technology Development Zone, Wuhan, 430075 Hubei, China Phone: +86- 27-87056201 Fax: +86- 27-87705217

Primary Contact:

Shoshana (Shosh) Friedman Senior Regulatory Affairs Consultant Phone: (704) 430-8695 s.friedman@promedoss.com

DATE PREPARED 2.

11/20/2022

DEVICE 3.

Trade Name: NaviCam Small Bowel Capsule Endoscopy System

Classification: Name: System, Imaging, Gastrointestinal, Wireless, Capsule Product Code: NEZ Regulation No: 876.1300 Class: II Classification Panel: Gastroenterology/Urology

PREDICATE DEVICES 4.

  • · Primary Predicate Device: Given PillCam (SB3) Endoscopy System cleared under K123864 (hereafter referred to as "PillCam SB3").
  • · Reference Device: NaviCam Capsule Endoscope System with NaviCam Stomach Capsule (also referred to as "NaviCam Stomach System") granted De-Novo under DEN190037

5. DEVICE DESCRIPTION

The NaviCam Small Bowel Capsule Endoscopy System is an endoscopic capsule imaging system intended to obtain images of the small bowel. It is comprised of the following components:

  • [1.] Capsule (AKES-11SW, AKES-11SI): The disposable, ingestible NaviCam Small Bowel Capsule is designed to acquire video images during the natural propulsion

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Ankon. The logo features a stylized, rounded shape in a light blue color, resembling a water droplet or a stylized letter 'U'. Below the symbol, the word "ankon" is written in a modern, sans-serif font. Underneath the company name, the tagline "INNOVATION FOR BETTER HEALTH" is displayed in smaller, uppercase letters.

through the GI tract. The capsule transmits the acquired images via an RF communication channel to the NaviCam Data Recorder located outside the body.

  • [2.] Data recorder (AKR-1, AKRI-1): The Data Recorder is an external receiving and recording unit that receives and stores the acquired images from the capsule.
  • [3.] ESView Software: The ESView is a software application for processing, analyzing, storing, and viewing the acquired images collected from the NaviCam Data Recorder to create a video of the images. The software also includes a reporting function to create detailed clinical reports and a capsule endoscopy atlas.
  • [4.] Locator: The Locator is a handheld device that is used to turn the NaviCam Capsule on. It is also used for determining if the capsule is still in the body when the patient is not sure whether he/she expelled it.

INDICATIONS FOR USE 6.

The NaviCam Small Bowel Capsule Endoscopy System is intended for visualization of the small bowel mucosa in adults. It may be used as a tool in the detection of abnormalities of the small bowel.

SUBSTANTIAL EQUIVALENCE 7.

The NaviCam Small Bowel Capsule Endoscopy System is substantially equivalent to the predicate devices based on the following:

Indications and Contraindications

The indications for use of the NaviCam Small Bowel Capsule Endoscopy System are included within the indications for use of the PillCam SB3 Capsule Endoscopy System primary predicate device. The contraindications for use of the NaviCam Small Bowel Capsule Endoscopy System are the same as these of the PillCam SB3 Capsule Endoscopy System

Technology

The capsule, data recorder, and software of the NaviCam Small Bowel Capsule Endoscopy System are substantially equivalent to these of the PillCam SB3 Capsule Endoscopy System primary predicate system.

The capsule, data recorder, and locator of the NaviCam Small Bowel Capsule Endoscopy System are also substantially equivalent to these of the reference device, the NaviCam Stomach System.

Performance

The performance of the NaviCam Small Bowel Capsule Endoscopy System was compared to this of the PillCam SB3 Capsule Endoscopy System primary predicate device and found to be substantially equivalent.

Conclusion

The NaviCam Small Bowel Capsule Endoscopy System has substantially similar indications and contraindications, technological, and performance characteristics as these of the predicate and reference devices. Any minor differences that may exist do not affect its safety and effectiveness when used as intended.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image contains the logo for Ankon. The logo features a stylized, rounded shape in a gradient of blue, resembling a water droplet or a stylized letter 'U'. Inside the shape is a vertical line. Below the shape is the word "ankon" in a bold, sans-serif font. Underneath the word "ankon" is the phrase "INNOVATION FOR BETTER HEALTH" in a smaller, sans-serif font.

8. PERFORMANCE DATA

The NaviCam Small Bowel Capsule Endoscopy System was tested in accordance with the FDA guidance titled "Ingestible Telemetric Gastrointestinal Capsule Imaging System - Final Class II Special Controls Guidance Document for Industry and FDA".

Bench/In-Vitro Testing

The NaviCam Small Bowel Capsule Endoscopy System successfully passed the following test:

TestPurpose
Biting TestTo demonstrate that the NaviCam SB capsule can withstand applied forces as may occur in case of accidental biting.
Angular Resolution TestTo measure the MTF using ISO 12233 slanted edge methodology, and to provide a new angular resolution method using LEDs in the capsule.
Temperature Safety TestTo test the temperature change during NaviCam SB capsule operation.
pH TestTo evaluate the integrity of the NaviCam SB capsule during exposure to simulated condition of extreme pH levels.
Image Intensity UniformityTo evaluate the intensity uniformity of the NaviCam SB capsule image.
Image Frame Rate TestTo demonstrate that the higher frame rate of NaviCam SB capsule image provides good transmission property.
Geometric Distortion TestTo determine geometric distortion of the NaviCam SB capsule and provide the local magnification of the image.
Field of View (FOV) TestTo determine the FOV value of the NaviCam SB capsule.
Battery Life TestTo demonstrate that the battery life of the NaviCam SB capsule lasts at least 8 hours and captures over 57500 images.
Image Resolution TestTo test the image resolution of the NaviCam SB capsule.
Magnetic Field TestTo measure the maximum value of magnetic flux density on the surface of the NaviCam SB capsule, the magnetic flux density on the non-optical bottom of the capsule, and to determine the magnetic field safety distance of the capsule.
DOV TestTo measure the MTF in air and underwater at different distances within the claimed DOV range using ISO 12233 slanted edge methodology and angular resolution method.
Color and Gray Scale TestTo evaluate the optical performance of the NaviCam SB capsule.
Data Integrity TestTo test the data transmission between the capsule through the data recorder and the ESView software.

The NaviCam Small Bowel Capsule System was tested for compliance with the following standards and regulations:

  • IEC 60601-1:2005+A1:2012 / EN 60601-1:2006 + A1:2013 + A12:2014 ●
  • ANSI/AAMI ES60601-1 (2005/(R)2012 + A1:2012, C1:2009/(R)2012 + A2:2010/(R)2012) - Amendment 1 - Revision Date 2012/08/21.
  • CAN/CSA-C22.2 No. 60601-1:14 Edition 3 Revision Date 2014/03. ●

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Ankon. The logo features a stylized blue "O" shape with a smaller white "i" inside, resembling a power button. Below the symbol is the word "ANKON" in a modern, sans-serif font, and beneath that is the tagline "INNOVATION FOR BETTER HEALTH" in smaller letters.

  • IEC 60601-1-11:2015. ANSI/AAMI HA60601-1-11: 2015 ●
  • CAN/CSA-C22.2 No. 60601-1-11: 15
  • IEC 60601-1-2:2014 ●
  • IEC 61000-3-2:2014
  • IEC 61000-3-3:2013 ●
  • IEC 62471 (1st edition 2006-07) ●
  • IEC 62133:2012; IEC 62133-2:2017
  • ANSI C63.27-2017
  • UL 1642:2020 ●
  • . 47 CFR Part 15

The system was found to comply with all applicable requirements of these standards.

Clinical Data

The NaviCam Small Bowel Capsule Endoscopy System was evaluated in a prospective study, comparing its performance to that of the FDA cleared PillCam SB3 Capsule Endoscopy System (NCT05086471).

In this prospective clinical study, the NaviCam Small Bowel Capsule Endoscopy System and PillCam SB3 Capsule Endoscopy System had a diagnostic Overall Percent Agreement rate of 89.66% (81.50%, 94.46%), and an overall percent agreement Kappa of 0.6652 (0.4653, 0.8652). The results of this comparative study demonstrate that the NaviCam Small Bowel Capsule Endoscopy System performs similarly to the PillCam SB3 Capsule Endoscopy System and provides an effective non-invasive method for the detection of small bowel diseases.

9. CONCLUSION

ANKON Technologies Co., Ltd. believes that the NaviCam Small Bowel Capsule Endoscopy System is substantially equivalent to its primary predicate device, the Given PillCam (SB3) Endoscopy System and its reference device, the NaviCam Stomach System. It has substantially equivalent indications and contraindication, technological characteristics, and performance characteristics to these of the primary and reference systems, and therefore does not introduce any new safety or effectiveness concerns.

§ 876.1300 Ingestible telemetric gastrointestinal capsule imaging system.

(a)
Identification. An ingestible telemetric gastrointestinal capsule imaging system is used for visualization of the small bowel mucosa as an adjunctive tool in the detection of abnormalities of the small bowel. The device captures images of the small bowel with a wireless camera contained in a capsule. This device includes an ingestible capsule (containing a light source, camera, transmitter, and battery), an antenna array, a receiving/recording unit, a data storage device, computer software to process the images, and accessories.(b)
Classification. Class II (special controls). The special control is FDA's guidance, “Class II Special Controls Guidance Document: Ingestible Telemetric Gastrointestinal Capsule Imaging Systems; Final Guidance for Industry and FDA.”