Intended Use

The Intense Pulsed Light (IPL) System is an over-the-counter device intended for the removal of unwanted body hair.

Device Description

Intense Pulsed Light (IPL) System is a small over-the-counter device for the permanent reduction of hair growth based on Intense Pulsed Light (IPL). It works below the skin's surface and does not involve any cutting or pulling, reducing hair growth with minimal pain. The Intense Pulsed Light (IPL) System includes main unit, an adaptor and goggles. The device is only powered by the external power adapter and its IPL emission activation is by a switch or auto light emission. Based on the ice-cool technology, the hair removal device with ice-cold compress functions.

AI/ML Overview

I apologize, but the provided text does not contain information about the acceptance criteria or a study that proves the device meets those criteria in the context of typical AI algorithm performance evaluation.

The document discusses the substantial equivalence of an "Intense Pulsed Light (IPL) System" (models T013C, T015C, T015K) to a predicate device (model T012C) for the purpose of removing unwanted body hair.

Here's a breakdown of why it doesn't fit the request for AI algorithm performance:

  • Device Type: This is a physical medical device (an IPL system), not an AI algorithm.
  • "Acceptance Criteria" in this context: The "acceptance criteria" here refer to meeting regulatory standards for medical device safety and effectiveness to demonstrate substantial equivalence to a predicate device, rather than performance metrics for an AI algorithm (e.g., sensitivity, specificity, AUC).
  • "Study that proves the device meets the acceptance criteria": The document refers to non-clinical testing (electrical safety, EMC, photobiological safety, biocompatibility, software life cycle processes) to demonstrate substantial equivalence and mitigate risks associated with minor design changes. There is explicitly no clinical testing performed (as stated in Section 8).

Therefore, I cannot populate the table or answer the specific questions about sample sizes, expert qualifications, adjudication methods, MRMC studies, standalone performance, or ground truth types, as these are related to AI algorithm validation, which is not what this document describes.

The document focuses on non-clinical testing and regulatory compliance for a hardware device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a smaller font.

June 30, 2022

Shenzhen Fansizhe Science And Technology Co., Ltd % You Yijie Manager Qimmiq Medical Consulting Service Co., Ltd. RM.406, Building C, Run Science Park, No.18 Shenzhou Road, Huangpu Guangzhou, Guangdong 510663 China

Re: K221569

Trade/Device Name: Intense Pulsed Light (IPL) System, model: T013C, T015K Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OHT, ONF Dated: May 15, 2022 Received: May 31, 2022

Dear You Yijie:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Jianting Wang Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K221569

Device Name

Intense Pulsed Light (IPL) System, model:T013C, T015C, T015K

Indications for Use (Describe)

The Intense Pulsed Light (IPL) System is an over-the-counter device intended for the removal of unwanted body hair.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

|X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Section 5: 510(K) Summary

1. Submitter's Information

Establishment Registration Information:
Name:Shenzhen Fansizhe Science And Technology Co.,Ltd
Address:2nd and 3rd floors, No.17, Xinfa 2nd Road, XinqiaoCommunity, Xinqiao Street, Baoan District,Shenzhen, Guangdong, China
Contact Person of applicant:
Name:Endang Cheng
Address:2nd and 3rd floors, No.17, Xinfa 2nd Road, XinqiaoCommunity, Xinqiao Street, Baoan District,Shenzhen, Guangdong, China
TEL:+86 0755-85275266
Email:ChengEndang1001@163.com
Contact Person of the Submission:
Name:You Yijie
Address:RM.406, Building C, Run Science Park, No.18Shenzhou Road, Huangpu, Guangzhou,Guangdong 510663 P.R.China
TEL:(+86)020-82245821
FAX:(+86)020-82245821
Email:jet.you@qimmiq-med.com

Date prepared: May 13, 2022

2. Device Information

Trade Name:Intense Pulsed Light (IPL) System
Model:T013C, T015C, T015K
Classification name:Light Based Over-The-Counter Hair Removal
Common or Usual Name:Powered Light Based Non-Laser Surgical Instrument WithThermal Effect
Review panel:General & Plastic Surgery
Product code:OHT
Subsequent Product code:ONF
Regulation Class:II
Regulation Number:878.4810

{4}------------------------------------------------

3. Predicate Device Information

510(k) submitter/holder:Shenzhen Fansizhe Science And TechnologyCo., Ltd
510(K) Number:K212881
Trade Name:Intense Pulsed Light (IPL) System
Model:T012C
Classification name:Powered Light Based Non-Laser SurgicalInstrument With Thermal Effect
Review panel:General & Plastic Surgery
Product code:OHT, ONF
Regulation Class:II
Regulation Number:878.4810

4. Device description

Intense Pulsed Light (IPL) System is a small over-the-counter device for the permanent reduction of hair growth based on Intense Pulsed Light (IPL). It works below the skin's surface and does not involve any cutting or pulling, reducing hair growth with minimal pain. The Intense Pulsed Light (IPL) System includes main unit, an adaptor and goggles. The device is only powered by the external power adapter and its IPL emission activation is by a switch or auto light emission.

Based on the ice-cool technology, the hair removal device with ice-cold compress functions.

5. Indications for Use

The Intense Pulsed Light (IPL) System is an over-the-counter device intended for the removal of unwanted body hair.

6. Summary of technological characteristics of device

SEComparisonsSubject device(Intense Pulsed Light (IPL) System,model: T013C, T015C, T015K)Predicate device(Intense Pulsed Light (IPL)System,model: T012C)Discussion ofdifference
510K NumberK221569K212881/
Classification21CFR 878.481021CFR 878.4810Same
Product CodeOHT, ONFOHT, ONFSame
FDA ClassIIIISame
ModelT013C, T015C, T015KT012C/
Indications for UseThe Intense Pulsed Light (IPL) System isan over-the-counter device intended forthe removal of unwanted body hair.The Intense Pulsed Light (IPL)System is an over-the-counterdevice intended for the removal ofunwanted body hair.Same
Type of useOver-The-Counter UseOver-The-Counter UseSame
Environment of UseHome useHome useSame
DesignHand-holdHand-holdSame
Patient PopulationAdultAdultSame
Material of PatientcontactcomponentsThe patients' skin or operator contact:T013C:PC for shell(Lamp shell, button, lightguide column, right shell, face shell,front upper shell), aluminum for coldcompress panelT015C&T015K: ABS for ABS shell andbuttons, aluminum for cold compresspanelThe patients' skin or operatorcontact:ABS for ABS shell and buttons,aluminum for cold compresspanelThe material ofT015C andT015K isidentical withpredicate device,T013C isdifferent withpredicate device,(Discussion isindicated in D1)
Biocompatibilitytesting1.Type of contact: direct contact forusers and patients.2.Nature of body contact category:Surface Contact class: A (<24 h)3.Meets ISO 10993- 5, ISO 10993-101.Type of contact: direct contactfor users and patients.2.Nature of body contactcategory: Surface Contact class:A (<24 h)3.Meets ISO 10993- 5, ISO 10993-10Same
Single Patient,multi-useYesYesSame
Patient InterfaceButtonsButtonsSame
TechnologyIntense Pulse Light (IPL)Intense Pulse Light (IPL)Same
Dimensions171×76×50mm for T013C206×113×91.5mm for T015C206×113×91.5mm for T015K148×190×56mmDifferent (Discussion is indicated inD2)
Power sourcean external power supplyInput: 100-240V ~ 50/60Hz, 1.5A Max.Output: 24Vdc 2.2Aan external power supplyInput: 100-240V ~ 50/60Hz, 1.5AMax. Output: 24Vdc 2.2ASame
Light sourceXenon Arc FlashlampXenon Arc FlashlampSame
Wavelength510nm~1200nm510nm~1200nmSame
Spot Size4cm² for T013C3.5cm² for T015C3.7cm² for T015KLight outlet: 4cm²Lamp cover: 1.85cm²Minordifferent(Discussion is indicated inD3)
Max. Fluence(J/cm²)4.03J/cm² for T013C4.69J/cm² for T015C3.6J/cm² for T015K5.18J/cm²(Attach Probe Cover:3.41J/cm²)Different (Discussion is indicated inD4)
Pulse duration412ms for T013C and T015C310ms for T015K4~12msThe parameter ofT013C andT015C isidentical withpredicate device,T013K is minordifferent withpredicate device(Discussion isindicated in D5)
Output energy3.8J16.1J for T013C4.1J16.4J for T015C5.2J~13.4J for T015K5.5J~20.7JDifferent(Discussion is indicated in D6)
PulsingControlFinger switchFinger switchSame
OutputChannelOne channelOne channelSame
DeliveryDirect Illumination to TissueDirect Illumination to TissueSame
SoftwareControlYesYesSame
OperatingenvironmentTemperature: 5-30 °CRelative humidity: 20-90%,without condensationAtmospheric Pressure: 70 kPa to 106 kPaTemperature: 5-30 °CRelative humidity: 20-90%,without condensationAtmospheric Pressure: 70 kPa to 106 kPaSame
Storage andtransportationenvironmentTemperature: -20 -55 °CRelative humidity: 5-95%,without condensationAtmospheric Pressure: 70 kPa to 106 kPaTemperature: -20 -55 °CRelative humidity: 5-95%,without condensationAtmospheric Pressure:70 kPa to 106 kPaSame
Electricalsafety, EMC,BiologicalEvaluationANSI AAMI ES60601-1IEC 60601-1-2IEC 60601-1-11IEC 60601-2-83IEC 62471ISO 10993-5ISO 10993-10ANSI AAMI ES60601-1IEC 60601-1-2IEC 60601-1-11IEC 60601-2-83IEC 62471ISO 10993-5ISO 10993-10Same

compared to the predicate devices (K212881)

{5}------------------------------------------------

{6}------------------------------------------------

Based on the comparison chart above, there have been no changes to the intended use or product specifications of subject device from those of the predicate device, the fundamental operating principle of the device is identical to that of the predicate device. The change associated with the updated industrial design which include a different PCB layout and some different hardware components, patient contact material, and minor different specifications, have been verified and validated via laboratory testing. Through the verification and validation process, it has been shown that the differences do not raise new questions of safety and effectiveness.

The discussion of differences exist between the subject and predicate devices is listed in following:

  • D1: The material of T015C and T015K is identical with predicate device,T013C is different with predicate device. The subject devices have been validated for cytotoxicity per ISO 10993- 5 and Irritation as well as Sensitization per ISO 10993-10 with positive results, therefore, the material difference of subject device with Predicate device TB-1755 (K183217) do not raise new questions of safety and effectiveness.
  • D2: The difference of dimensions will not affect the safety and effectiveness.
  • D3: Spot Size of subject device is same or similar with predicate device. The safety and effectiveness of the subject device is verified via tests according to ANSI AAMI ES60601-1, IEC60601-1-2, IEC60601-1-11, IEC 62471 and IEC 60601-2-83, so the differences do not affect the safety and effectiveness.

{7}------------------------------------------------

  • D4: Max. Fluence of subject device is within the range of predicate device. The safety and effectiveness of the subject device is verified via tests according to ANSI AAMI ES60601-1, IEC60601-1-2, IEC60601-1-11, IEC 62471 and IEC 60601-2-83, so the differences do not affect the safety and effectiveness.
  • D5: Pulse duration of T013C and T015C is identical with predicate device,T013K is minor different with predicate device, but similar and within the range of predicate device. The safety and effectiveness of the subject device is verified via tests according to ANSI AAMI ES60601-1, IEC60601-1-2, IEC60601-1-11, IEC 62471 and IEC 60601-2-83, so the differences do not affect the safety and effectiveness.
  • D6: Output energy of subject device is within the range of predicate device. The safety and effectiveness of the subject device is verified via tests according to ANSI AAMI ES60601-1, IEC60601-1-2, IEC60601-1-11, IEC 62471 and IEC 60601-2-83, so the differences do not affect the safety and effectiveness.

7. Discussion of Non-Clinical Tests Performed for Safety and

effectiveness are as follows

The modifications to the device have been designed and assessed under design control processes compliant with FDA 21 CFR 820. Based on modification clearance, a risk analysis was conducted to assess the impact of the changes on the subject device using internal design control procedures and a fault tree analysis described in the FDA-recognized version of ISO 14971.

These risks were mitigated using planned measures that included testing to recognized FDA consensus standards. Changes in software were verified and validated using the software development process.

Non-clinical testing as below table has been performed to demonstrate that the safety and efficacy of the device remains substantially equivalent to its predicate.

StandardsStandards NameResults
ANSI AAMI ES60601-1:2005/(R)2012 andA1:2012Medical Electrical Equipment -Part 1: General RequirementsFor Basic Safety And EssentialPerformancePass
IEC 60601-1-2: 2014Medical Electrical Equipment --Part 1-2: General RequirementsFor Basic Safety And EssentialPerformance -- CollateralStandard: ElectromagneticDisturbances -- RequirementsAnd TestsPass
IEC 60601-1-11: 2015Medical Electrical Equipment -Part 1-2: General RequirementsFor Basic Safety And EssentialPerformance - CollateralStandard: ElectromagneticDisturbances - RequirementsAnd TestsPass

{8}------------------------------------------------

IEC 60601-2-83:2019Medical electrical equipmentPart 2-83:Particularrequirements for the basicsafety and essentialperformance of home lighttherapy equipmentPass
IEC 62471: 2006Photobiological safety of lampsand lamp systemsPass
ISO 10993-5:2009Biological evaluation of medicaldevices - Part 5: Tests for invitro cytotoxicityPass
ISO 10993-10:2010Biological evaluation of medicaldevices - Part 10: Tests forirritation and skin sensitizationPass
IEC62304:2006+A1:2015Medical device software -Software life cycle processesPass

8. Discussion of Clinical Accuracy Testing Performed

There was no clinical testing performed.

9. Conclusions

A risk analysis was performed to identify risks associated with the device modifications. Verification and validation tests have been performed to demonstrate that the identified risks have been mitigated. Thus, based on the information provided in this premarket notification, it is concluded that the modified device Intense Pulsed Light (IPL) Systems, model: T013C, T015C, T015K are substantially equivalent to the legally marketed predicate device.

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.