K Number
K221559

Validate with FDA (Live)

Manufacturer
Date Cleared
2023-02-23

(268 days)

Product Code
Regulation Number
878.4040
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The StringKing Disposable Surgical Gown is a sterile, single use surgical apparel intended to be worn by healthcare professionals to help protect both the patient and the healthcare worker from the transfer of microorganisms, body fluids and particulate matter. The StringKing Surgical Gowns meet Level 3 and Level 4 requirements of ANSI/AAMI PB 70:2012 Liquid barrier performance and classification of protective apparel and drapes intended for use in health care facilities.

Device Description

The Disposable Surgical Gown consists of a multi-layer nonwoven spunbond-meltblownspunbond (SMS) polyolefin & Film Lamination in the critical zones and a single layer of lighter, more breathable nonwoven SMS polvolefin in the non-critical zones. The StringKing Disposable Surgical Gown is designed for easy donning and doffing with hook and loop neck tabs, belt ties, removable transfer accessory, and polyester wrist cuffs. The StringKing Disposable Surgical Gown is one gown design that meets both Level 4 requirements of AMI PB70:2012 Liquid barrier performance and classification of protective apparel and drapes intended for use in health care facilities. The gown will be marketed as a L3 gown as well as a L4 gown. The StringKing Disposable Surgical Gowns are single use, disposable medical devices that will be provided sterile.

AI/ML Overview

Here's a breakdown of the requested information based on the provided text, focusing on the acceptance criteria and the study that proves the device meets them:

1. Table of Acceptance Criteria and Reported Device Performance

Test StandardAcceptance CriteriaReported Device PerformancePass/Fail
ANSI/AAMI PB70:2012 (Overall classification)Meets Level 3 and Level 4 Liquid Barrier RequirementsMeets ANSI/AAMI PB 70:2012 Level 3 and Level 4 Liquid Barrier requirementsPass
Water Resistance Hydrostatic Pressure Test - AATCC 127:2017 (cm)$\ge$ 50 cmAll samples passed at $\ge$ 50 cmPass
Flammability of Clothing Textiles - 16 CFR Part 1610 (a)Class 1: $\ge$ 3.5 secondsAll samples passed at $\ge$ 3.5 secondsPass
Durability ASTM D-5034$\ge$ 7 lbfAll samples passed at greater than 7 lbfPass
Tensile Strength ASTM D-5034 Machine Direction and Cross Direction$\ge$ 7 lbfAll samples passed at greater than 7 lbfPass
Trap Tear ASTM D5587-15 Machine Direction and Cross Direction$\ge$ 2.3 lbfAll samples passed at $\ge$ 2.3 lbfPass
Water Resistance Impact Penetration Test AATCC 42$\le$ 1.0 gAll samples passed at $\le$ 1.0 gPass
ASTM D1683 Seam Strength$\ge$ 7 lbfAll samples passed at $\ge$ 7 lbfPass
ASTM F1671 Viral PenetrationLevel 4: No penetration at 2 psiAll samples showed no penetration at 2 psiPass
Ethylene oxide residuals ISO 10993-7EO <4mg, ECH <9mgEO residuals were not detected ("ND")Pass
Cytotoxicity – ISO 10993-5$\le$ Grade 2 (mild)Grade 0Pass
Skin Sensitization- ISO 10993-10Primary Irritation Index 0-0.4 (Negligible)0.0Pass
Skin Irritation - ISO 10993-23$\le$ 1 on Magnusson and Kligman Scale0.0Pass

2. Sample size used for the test set and the data provenance

The document does not explicitly state the specific sample sizes for each individual test. It generally uses phrases like "All samples passed," indicating that a sufficient number of samples were tested to meet the requirements of the respective standards.

The data provenance is from non-clinical testing conducted according to recognized international and national standards (e.g., ANSI/AAMI PB70:2012, AATCC, ASTM, ISO, CFR). The testing was for the StringKing Disposable Surgical Gown. The document does not specify the country of origin of the testing facilities, but the standards are international. This is retrospective in the sense that the studies were performed and results collected before this submission.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This section is not applicable as the device is a physical product (surgical gown) and the "ground truth" for its performance is established through objective, standardized laboratory testing (e.g., measuring forces, penetration, chemical residues) against predefined thresholds, not through expert human interpretation or consensus of observational data.

4. Adjudication method for the test set

This is not applicable for the same reasons as above. Performance is determined by meeting objective criteria in standardized tests, not by human adjudication of qualitative or ambiguous data.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This is not applicable. The device is a surgical gown, not an AI-powered diagnostic or assistive tool. Therefore, MRMC studies involving human readers and AI are not relevant.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

This is not applicable. The device is a physical product and does not involve any algorithms or AI for its function or evaluation.

7. The type of ground truth used

The ground truth for the performance of the surgical gown is established by objective measurements and adherence to predefined thresholds specified in international and national standards (e.g., ANSI/AAMI PB70:2012, ASTM D5034, AATCC 127, ISO 10993, etc.). These standards define what constitutes a "pass" or "fail" for various performance characteristics like liquid barrier, flammability, tensile strength, tear strength, seam strength, viral penetration, and biocompatibility.

8. The sample size for the training set

This is not applicable. The device is a physical medical device and does not involve any machine learning algorithms that require a "training set."

9. How the ground truth for the training set was established

This is not applicable as there is no training set for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".

February 23, 2023

StringKing Thomas Frasca Senior Partner 19100 S Vermont Avenue Gardena. California 90248

Re: K221559

Trade/Device Name: Disposable Surgical Gown-Level 3 Disposable Surgical Gown-Level 4 Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FYA Dated: January 20, 2023 Received: January 20, 2023

Dear Thomas Frasca:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Bifeng Qian -S

Bifeng Oian, M.D., Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K221559

Device Name Disposable Surgical Gown - Level 3 Disposable Surgical Gown - Level 4

Indications for Use (Describe)

The StringKing Surgical Gown is a sterile, single use surgical apparel intended to be worn by healthcare professionals to help protect both the patient and the healthcare worker from the transfer of microorganisms, body fluids and particulate matter. The StringKing Surgical Gowns meet Level 3 and Level 4 requirements of ANSVAAMI PB 70:2012 Liquid barrier performance and classification of protective apparel and drapes intended for use in health care facilities.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K221559 510(k) SUMMARY

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR §807.92:

SUBMITTER 1.

StringKing 19100 S. Vermont Avenue Gardena. CA 90248 Tel: +1.508.654.1988 Fax: N/A

Contact Person:Tom Frasca
Date Prepared:February 23, 2023
Prepared by:Sharon Morrow, Regulatory Consultant

ll. DEVICE

510(k):K221559
Name of Device:Disposable Surgical Gown, Level 4Disposable Surgical Gown, Level 3
Classification Name:Disposable Surgical Gown
Regulation:21 CFR §878.4040
Regulatory Class:Class II
Product Classification Code:FYA

= PREDICATE DEVICE

Predicate Manufacturer Predicate Trade Name: Predicate 510(k):

B.J.ZH.F.PANTHER MEDICAL EQUIPMENT CO., LTD. Surgical Gown K202706

No reference devices were used in this submission.

IV. DEVICE DESCRIPTION

The Disposable Surgical Gown consists of a multi-layer nonwoven spunbond-meltblownspunbond (SMS) polyolefin & Film Lamination in the critical zones and a single layer of lighter, more breathable nonwoven SMS polvolefin in the non-critical zones. The StringKing Disposable Surgical Gown is designed for easy donning and doffing with hook and loop neck tabs, belt ties, removable transfer accessory, and polyester wrist cuffs. The StringKing Disposable Surgical Gown is one gown design that meets both Level 4 requirements of AMI PB70:2012 Liquid barrier performance and classification of protective apparel and drapes intended for use in health care facilities. The gown will be marketed as a L3 gown as well as a L4 gown. The StringKing Disposable Surgical Gowns are single use, disposable medical devices that will be provided sterile.

{4}------------------------------------------------

V. INDICATIONS FOR USE

The StringKing Disposable Surgical Gown is a sterile, single use surgical apparel intended to be worn by healthcare professionals to help protect both the patient and the healthcare worker from the transfer of microorganisms, body fluids and particulate matter. The StringKing Surgical Gowns meet Level 3 and Level 4 requirements of ANSI/AAMI PB 70:2012 Liquid barrier performance and classification of protective apparel and drapes intended for use in health care facilities.

COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE VI. PREDICATE DEVCE

DeviceSubject DeviceStringKingDisposableIsolation Gown(K221559)Predicate: B.J.ZH.F.Panther MedicalEquipment Co., Ltd.(K202706)Justification forDifferences
ManufacturerStringKingB.J.ZH.F. Panther MedicalEquipmentN/A
510K NumberK221559K202706N/A
ProductCommon NameDisposable SurgicalGownSurgical GownSame
Product CodeFYAFYASame
ClassificationClass II (21 CFR878.4040)Class II (21 CFR878.4040)Same
Intended Use /Indications forUseThe StringKingDisposable SurgicalGown is a sterile, singleuse surgical apparelintended to be worn byhealthcareprofessionals to helpprotect both the patientand the healthcareworker from the transferof microorganisms,body fluids andparticulate matter. TheStringKing SurgicalGowns meet Level 3and Level 4requirements ofANSI/AAMI PB 70:2012Liquid barrierperformance andclassification ofprotective apparel anddrapes intended for useSurgical gowns are intended tobe worn by operating roompersonnel during surgicalprocedure to protect both thesurgical patient and theoperating room personnel fromtransfer of microorganisms,body fluids, and particulatematerial.Per ANSI/AAMI PB70:2012Liquid barrier performance andclassification of protectiveapparel and drapes intended foruse in health care facilities, thesurgical gowns met therequirements for Level 3classification.Intended Use is the Same.Subject Device also meetsLevel 4 requirements
in health care facilities.
Gown SizesSmall, Medium,Large, X-Large, 2X-LargeSmall, Medium, Large, X-Large, XX-Large and XXX-Large.Similar; Predicateoffers XXX-Large
MaterialCompositionSMS ReinforcedSMS/PE - gownmaterial and tiesSMS polypropylenenonwoven + PolyesterSimilar;
Design FeaturesNeck Closure: Hook& Loop TabBelt: TiesCuffs: KnitTransfer TabNeck Closure: not included in510k Summary.Belt: not listed in 510kSummaryCuffs: KnitSimilar
ColorBlueBlueSame
OTC UseYesYesSame
SterilitySterile (EO)Sterile (EO) and Bulk, non-sterileSimilar; predicatesavailable bulk, non-sterile
UseSingle Use,DisposableSingle Use, DisposableSame
ANSI/AAMI PB70Level 3 and Level 4Level 3Same ; Subject Devicealso meets Level 4requirements
Liquid BarrierPerformanceClassificationPropertiesDevice was tested inaccordance withAAMI PB70:2012and meets Level 4and Level 3requirements fora surgical gown.Per ANSI/AAMI PB70:2012 Liquidbarrier performance andclassification of protective appareland drapes intended for use inhealth care facilities, the proposeddevices meet the requirements forLevel 3classification.Same
FlammabilityMeets requirementsof Class 1 per 16CFR 1610Class 1 per 16 CFR 1610Same
Non-barrierpropertiesNot specified in 510k summarySame
Tensile Strength;ASTM D5034 (MachineDirection)Meets requirementsof ASTM D5034:$\ge$ 7lbf
Tensile Strength;ASTM D5034 (Crossdirection)Meets requirementsof ASTM D5034:$\ge$ 7lbf
Trap Tear; ASTMD5587 (MachineDirection)Meets requirementsof ASTM D5587:$\ge$ 2.3lbf
Trap Tear; ASTMD5587 (CrossDirection)Meets requirementsof ASTM D5587:$\ge$ 2.3lbf
Seam Strength;ASTM D1683Meets requirementsof ASTM D1683:$\ge$ 7lbf
BiocompatibilityMeets requirementsof ISO 10993-5, ISO10993-10, ISO10993-23 (non-cytotoxic, non-irritant, non-sensitizer)Meets requirements of ISO10993-5, ISO 10993-10(non-cytotoxic, non-irritant, non-sensitizer)Same
Ethylene oxideresidualsMeets requirementsof ISO 10993-7Meets requirements ofISO 10993-7Same
Test StandardAcceptanceCriteriaResults -ProposedDevice -StringKingDisposableSurgical Gown(K221559)Pass/Fail
ANSI/AAMI PB70:2012Meets ANSI/AAMIPB 70:2012 Level 3and Level 4 LiquidBarrier requirements
Water ResistanceHydrostatic PressureTest - AATCC 127:2017(cm)$\ge$ 50cmAll samples passedat $\ge$ 50 cmPass
Flammability ofClothing Textiles - 16CFR Part 1610 (a)Class 1: $\ge$ 3.5 secondsAll samples passed at$\ge$ 3.5 secondsPass
DurabilityASTM D-5034$\ge$ 7lbfAll samples passed atgreater than 7lbfPass
Tensile StrengthASTM D-5034Machine Directionand CrossDirection$\ge$ 7lbfAll samples passed atgreater than 7lbfPass
Trap Tear ASTM D5587-15Machine Direction andCross Direction≥2.3lbfAll samples passedat≥2.3lbfPass
Water ResistanceImpact PenetrationTest AATCC 42≤1.0gAll samples passed at≤1.0gPass
ASTM D1683Seam Strength≥7lbfAll samples passed at≥7lbfPass
ASTM F1671Viral PenetrationLevel 4: Blood andviralpenetrationresistanceNopenetration at 2psiAll samples showedno penetration at 2psiPass
Ethylene oxide residualsISO 10993-7EO <4mg*ECH <9mgEO residuals were notdetected ("ND")Pass

{5}------------------------------------------------

{6}------------------------------------------------

{7}------------------------------------------------

VII. PERFORMANCE DATA

The following performance data were provided in support of the substantial equivalence determination.

{8}------------------------------------------------

*Daily dose

Sterilization & Shelf-life Testing

Sterilization validation according to ISO 11135; Shelf-life testing was conducted according to ASTM F1980-16 as well as ASTM F2096-11(2019) and ASTM F88/F88M-15.

Biocompatibility Testing

Biocompatibility TestingEndpointsAcceptance CriteriaResultsPass/Fail
Cytotoxicity – ISO 10993-5≤Grade 2 (mild)Grade 0Pass

{9}------------------------------------------------

Skin Sensitization- ISO 10993-10Primary Irritation Index0-0.4 — Negligible0.5-1.9-Slight2-4.9-Moderate 5-8-Severe0.0Pass
Skin Irritation - ISO 10993-23≤1 on Magnusson andKligman Scale0.0Pass

Clinical testing: Clinical Studies are not required for this device and therefore are not included in this submission.

VIII. CONCLUSIONS

The conclusions drawn from the non-clinical tests demonstrate that the subject device in 510(k) submission K221559, the StringKing Surgical Gown is as safe, as effective, and performs as well as or better than the legally marketed predicate device (K202706), manufactured by B.J.ZH.F. Panther Medical Equipment Co., Ltd.

§ 878.4040 Surgical apparel.

(a)
Identification. Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.(b)
Classification. (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.