ANNE Pediatric
K221530 · Sibel, Inc. · DRG · Nov 22, 2022 · Cardiovascular
Device Facts
| Record ID | K221530 |
| Device Name | ANNE Pediatric |
| Applicant | Sibel, Inc. |
| Product Code | DRG · Cardiovascular |
| Decision Date | Nov 22, 2022 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 870.2910 |
| Device Class | Class 2 |
| Attributes | Pediatric |
Intended Use
ANNE Pediatric is a wireless vital signs and physiological data monitoring platform indicated for the measurement of heart rate, skin temperature, and body temperature by qualified healthcare professionals in healthcare settings. The device is intended for use on neonatal patients of any gestational age up to infants 2 years of age as a general patient monitor to provide continuous physiological information as an aid to diagnosis and treatment. The device is not intended to monitor critically ill patients. The device is not intended to monitor or measure heart rate during motion. The device is not an apnea alarm.
Device Story
ANNE Pediatric is a wireless vital signs monitoring platform for neonates and infants up to 2 years old. System consists of two skin-mounted, bio-integrated sensors (chest and foot) and the ANNE View mobile application. Sensors collect ECG and temperature biosignals; data transmitted via Bluetooth to mobile device for real-time display. Used by healthcare professionals in clinical settings to aid diagnosis and treatment. Device provides continuous physiological information; does not provide alarms or monitor heart rate during motion. Benefits include non-invasive, continuous monitoring of vital signs in non-critically ill pediatric patients.
Clinical Evidence
Clinical study conducted at a tertiary facility with n=137 neonates (median gestational age 38 weeks). HR performance compared against Masimo Rad-97 in one open-label and three closed-label rounds (n=70). Final closed-label round (n=20) demonstrated a normalized root mean square deviation (RMSD) of 1.2%, meeting the ±10% or ±5bpm accuracy requirement per IEC 60601-2-27. No adverse events or skin injuries reported.
Technological Characteristics
Wireless vital signs monitor using Bluetooth connectivity. Sensors are skin-mounted and bio-integrated. Biocompatibility per ISO 10993-1, -5, -10. Electrical safety per ANSI/AAMI ES60601-1 and IEC 60601-1-2. Wireless coexistence per ANSI IEEE C63.27. Temperature measurement per ISO 80601-2-56. Heart rate measurement per IEC 60601-2-27. Software developed per IEC 62304.
Indications for Use
Indicated for neonatal patients of any gestational age up to infants 2 years of age. Used by healthcare professionals in healthcare settings for continuous monitoring of heart rate, skin temperature, and body temperature. Not for critically ill patients, not for heart rate monitoring during motion, and not an apnea alarm.
Regulatory Classification
Identification
A radiofrequency physiological signal transmitter and receiver is a device used to condition a physiological signal so that it can be transmitted via radiofrequency from one location to another, e.g., a central monitoring station. The received signal is reconditioned by the device into its original format so that it can be displayed.
Predicate Devices
Reference Devices
Related Devices
- K211305 — ANNE One · Sibel, Inc. · Sep 14, 2021
- K240251 — ANNE Chest · Sibel Health, Inc. · Jun 3, 2024
- K223721 — Masimo Stork · Masimo Corporation · Dec 15, 2023
- K223711 — ANNE One · Sibel Health, Inc. · Aug 10, 2023
- K130709 — VISI MOBILE MONITORING SYSTEM · Sotera Wireless, Inc. · Oct 7, 2013
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 22, 2022
Sibel Inc. Sarah Coughlin OARA Engineer 6650 W. Touhy Ave. Niles, Illinois 60714
Re: K221530
Trade/Device Name: ANNE Pediatric Regulation Number: 21 CFR 870.29870.2910 Regulation Name: Radiofrequency physiological signal transmitter and receiver Regulatory Class: Class II Product Code: DRG, MWI, FLL Dated: Mav 23, 2022 Received: May 26, 2022
Dear Sarah Coughlin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Jennifer W. Shih -S
Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known)
K221530
Device Name ANNE Pediatric
Indications for Use (Describe)
ANNE Pediatric is a wireless vital signs and physiological data monitoring platform indicated for the measurement of heart rate, skin temperature by qualified healthcare professionals in healthcare settings. The device is intended for use on neonatal patients of any gestational age up to infants 2 years of age as a general patient monitor to provide continuous physiological information as an aid to diagnosis and treatment. The device is not intended to monitor critically ill patients. The device is not intended to monitor or measure heart rate during motion. The device is not an apnea alarm.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (6/20)
{3}------------------------------------------------
# 510(K) Summary
#### -Submitter:
Sibel Inc. 6650 W. Touhy Avenue, Niles, IL 60714 Tel: (224) 251-8859
Official Correspondent: Sarah Coughlin, Senior Regulatory Affairs and Quality Assurance Engineer 6650 W. Touhy Avenue, Niles, IL 60714 Tel: (224) 251-8859
Date Prepared: 10/7/2022
#### II. Device Information
Name of Device: ANNE Pediatric
510(k) Number: K221530
Common or Usual Name: Wireless Remote Monitoring System
Classification Name: Radiofrequency Physiological Signal Transmitter and Receiver
Review Panel: Cardiovascular
Regulation: 21 CFR §870.2910
Regulatory Class: Class II
Product Classification Code: DRG, MWI, FLL
#### . Predicate Device
Trade Name: ANNE One Predicate 510(k): K211305 Predicate device manufacturer: Sibel Inc.
#### IV. Device Description
ANNE Pediatric is a wireless vital signs and physiological data monitoring platform that collects electrocardiography (ECG) and temperature biosignals to measure vital signs such as heart rate (HR), skin temperature, and body temperature. The system features two skin-mounted, bio-integrated sensors that pair with the ANNE View application for the display of real-time vital measurements. Data is transmitted wirelessly from the sensors via Bluetooth to the mobile device.
{4}------------------------------------------------
#### V. Indications for Use
ANNE Pediatric is a wireless vital signs and physiological data monitoring platform indicated for the measurement of heart rate, skin temperature, and body temperature by qualified healthcare professionals in healthcare settings. The device is intended for use on neonatal patients of any gestational age up to infants 2 years of age as a general patient monitor to provide continuous physiological information as an aid to diagnosis and treatment. The device is not intended to monitor critically ill patients. The device is not intended to monitor or measure heart rate during motion. The device is not an apnea alarm.
| | Subject device<br>ANNE Pediatric | Predicate device<br>Sibel Inc. | Variances/Equivalence |
|-------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------|
| Trade Name | ANNE Pediatric | ANNE One | |
| Class | II | II | Equivalent |
| Product Code | DRG, MWI, FLL | DRG, MWI, FLL | Equivalent |
| Regulation<br>Number and Name | 870.2910<br>Transmitters and<br>Receivers, Physiological<br>Signal, Radiofrequency | 870.2910<br>Transmitters and<br>Receivers, Physiological<br>Signal, Radiofrequency | Equivalent |
| Indications for Use | ANNE Pediatric is a<br>wireless vital signs and<br>physiological data<br>monitoring platform<br>indicated for the<br>measurement of heart<br>rate, skin temperature,<br>and body temperature by<br>qualified healthcare<br>professionals in<br>healthcare settings. The<br>device is intended for<br>use on neonatal patients<br>of any gestational age up<br>to infants 2 years of age<br>as a general patient<br>monitor to provide<br>continuous physiological<br>information as an aid to<br>diagnosis and treatment.<br>The device is not<br>intended to monitor<br>critically ill patients. The<br>device is not intended to<br>monitor or measure<br>heart rate during motion.<br>The device is not an<br>apnea alarm. | ANNE One is a wireless<br>vital signs and<br>physiological data<br>monitoring platform<br>indicated for the<br>measurement of heart<br>rate, respiratory rate,<br>step count, fall count,<br>skin temperature, and<br>body temperature by<br>qualified healthcare<br>professionals in<br>healthcare settings. The<br>device is intended for<br>use on general care<br>patients who are 18<br>years of age or older as<br>a general patient monitor<br>to provide continuous<br>physiological information<br>as an aid to diagnosis<br>and treatment. The<br>device is not intended for<br>use on critical care<br>patients. The device is<br>not intended to monitor<br>or measure respiratory<br>rate or heart rate on<br>ambulatory patients. | Substantially<br>Equivalent |
| Target Population | Neonates and infants up<br>to 2 years of age | Adults, 18 years of age<br>and older | Different |
| Use Environment | Healthcare Settings | Healthcare Settings | Equivalent |
| Sensor Placement | Chest and Foot | Chest and Finger | Substantially<br>Equivalent |
| Data | Data transmitted<br>wirelessly via Bluetooth<br>from sensors to mobile<br>device | Data transmitted<br>wirelessly via Bluetooth<br>from sensors to mobile<br>device | Equivalent |
| Notification | No notification ability | No notification ability | Equivalent |
| Heart Rate | Yes | Yes | Equivalent |
| Skin Temperature | Yes | Yes | Equivalent |
| Body Temperature | Yes | Yes | Equivalent |
| Respiratory Rate | No | Yes | Different |
| Step Count | No | Yes | Different |
| Fall Count | No | Yes | Different |
{5}------------------------------------------------
#### VI. Clinical Data
The functionality and performance of the ANNE Pediatric system was assessed in n=137 neonates at Aga Khan University, a tertiary healthcare facility in Nairobi, Kenya. The study compared the performance of HR measurements with ANNE Pediatric against the Masimo Rad-97 in one open label and three closed-label rounds with n=70 neonates. Median gestational age of neonates included in HR analysis was 38 (range 25 to 42) weeks. The HR analysis cohort was 44.7% female, 52.9% male, and 1.4% intersex. The normalized root mean square deviation (RMSD) in the final closed-label round of testing with n=20 neonates was 1.2%, meeting the accuracy requirement of ±10% or ±5bpm, whichever is greater, defined in IEC 60601-2-27. No adverse events or skin injuries were reported during the study.
#### VII. Performance Data
The following consensus standards and bench testing were used to evaluate the safety and effectiveness of ANNE Pediatric:
- Electrical safety and electromagnetic compatibility testing according to ANSI/AAMI ES60601-1:2005/(R)2012 and IEC 60601-1-2:2014 standards.
- . Biocompatibility testing according to ISO 10993-1:2018, ISO 10993-5:2009, and ISO 10993-10:2010 for patient contacting materials.
- Wireless coexistence testing according to ANSI IEEE C63.27-2017. ●
- Defibrillation testing according to Section 8.5.5 of ANSI/AAMI ES60601-1:2005/(R)2012 ●
{6}------------------------------------------------
- . Software verification and validation testing according to IEC 62304:2015 and the FDA guidance document, Content of Premarket Submissions for Software Contained in Medical Devices.
- Shelf life testing of the adhesive to demonstrate safe and effective performance over the . intended device life cycle.
- . Bench testing to demonstrate the mechanical durability of the sensors.
- o Usability testing in accordance with the FDA guidance document, Applying Human Factors and Usability Engineering to Medical Devices
- Performance testing of body temperature measurements according to ISO 80601-2-56:2017.
- 0 Performance testing of heart rate measurements according to Sections 201.12.1.101.15 and 201.12.1.101.17 of IEC 60601-2-27:2011.
- Cybersecurity evaluation according to the requirements of the FDA draft guidance . document, Content of Premarket Submissions for Management of Cybersecurity in Medical Devices
- Assessment of Software of Unknown Provenance per the FDA guidance document, Off-The-Shelf Software Use in Medical Devices
#### VIII. Conclusion
The results of the substantial equivalence assessment, taken together with the clinical and performance testing data, demonstrate that the ANNE Pediatric system's performance characteristics are substantially equivalent to the predicate device in both technology and intended use.