K Number
K221510

Validate with FDA (Live)

Date Cleared
2022-10-12

(141 days)

Product Code
Regulation Number
872.3690
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

TrusFIL-Flow Flowable Composite Restorative is a visible light curing dental restorative material indicated for:

  • Class III and V restorations
  • Restoration of minimally invasive cavity preparations (including small, non stress-bearing occlusal restorations)
  • Base/liner under direct restorations
  • Repair of small defects in indirect restorations
  • Pit and fissure sealant
  • Undercut blockout
  • Repair of resin and acrylic temporary materials
Device Description

TrusFIL-Flow Flowable Composite Restorative is a visible light curing dental restorative material intended to restore carious lesions or structural defects in teeth in combination of a bonding agent and with or without other materials such as luting, etching agents, cavity liners, universal composites and others commonly used in a tooth restoration. The principal organic components are mixtures of dental methacrylate resins (Bis-GMA, TEGDMA, EBPADMA). The inorganic filler loading is about 46% by volume having particle size range of about 0.01 to 3 microns. It is packaged in syringes and single-dose capsules. Single-dose delivery is intended for single patient (single use) only to prevent cross-contamination between patients.
Available shades include opaque dentin shades and regular body shades. All shades are radiopaque.

AI/ML Overview

The provided document is a 510(k) premarket notification for a dental restorative material, TrusFIL-Flow Flowable Composite Restorative. It aims to demonstrate substantial equivalence to a predicate device, not to prove the device meets acceptance criteria for an AI/algorithmic medical device, as implied by the user's request.

Therefore, the document does not contain the information needed to answer the user's questions about acceptance criteria and study details for an AI/algorithmic device. The document primarily focuses on physical and chemical properties, biocompatibility, and intended use for a conventional dental material.

Specifically, the following information is not available in the provided text:

  • A table of acceptance criteria and reported device performance for an AI/algorithmic device. The "acceptance criteria" presented are pass/fail criteria for physical properties of a dental material according to ISO standards and FDA guidance, not for an AI.
  • Sample size used for a test set or data provenance for an AI.
  • Number of experts or their qualifications for establishing ground truth for an AI test set.
  • Adjudication method for an AI test set.
  • Information on Multi-Reader Multi-Case (MRMC) comparative effectiveness study or human reader improvement with AI assistance.
  • Standalone performance for an algorithm only.
  • Type of ground truth used (expert consensus, pathology, outcomes data, etc.) for an AI.
  • Sample size for a training set for an AI.
  • How ground truth for a training set was established for an AI.

The document explicitly states under "10. Clinical Performance Data": "Not applicable. Clinical performance testing has not been performed on the subject device." This further confirms that no clinical studies or AI performance evaluations, as requested, were conducted or reported here.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 12, 2022

Rizhao HuGe Biomaterials Company, Ltd. Maggie Zheng Regulatory Affairs Manager No.2 North Zhaoyang Road, District of Donggang Rizhao City, Shandong Province 276800 CHINA

Re: K221510

Trade/Device Name: TrusFIL-Flow Flowable Composite Restorative Regulation Number: 21 CFR 872.3690 Regulation Name: Tooth Shade Resin Material Regulatory Class: Class II Product Code: EBF Dated: July 6, 2022 Received: July 14, 2022

Dear Maggie Zheng:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K221510

Device Name

TrusFIL-Flow Flowable Composite Restorative

Indications for Use (Describe)

TrusFIL-Flow Flowable Composite Restorative is a visible light curing dental restorative material indicated for:

  • Class III and V restorations
  • Restoration of minimally invasive cavity preparations (including small, non stress-bearing occlusal restorations)
  • Base/liner under direct restorations
  • Repair of small defects in indirect restorations
  • Pit and fissure sealant
  • Undercut blockout
  • Repair of resin and acrylic temporary materials
Type of Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

005_510 (k) Summary K221510

This summary of 510(k) for the subjective device equivalence information is being submitted in accordance with the requirements of 21 C.F.R. 807.92.

1. Date Summary Prepared: Oct. 11, 2022

2. Submitter Information:

Owner's NameRizhao HuGe Biomaterials Company, Ltd.
AddressNo.2 North Zhaoyang Road, District of Donggang, Rizhao City,Shandong Province, China 276800
Telephone0086 633 2277268
Contact PersonMs. Maggie Zheng
Contact TitleRegulatory Affairs Manager
E-mailzhengxy@hugedent.com

3. Device Name

Trade name: TrusFIL-Flow Flowable Composite Restorative Common name: Flowable Composite Restorative Classification name: Material, Tooth Shade, Resin Regulatory Class: II Product Code: EBF

4. Predicate Device Information

Owner/OperatorDevice Trade Name510 (k) No.Product CodePredicate
3M ESPE Dental ProductsFILTEK SUPREME ULTRAFLOWABLE RESTORATIVEK100235EBFPrimary

This predicate device has not been subject to a design-related recall.

No reference devices were used in this submission.

{4}------------------------------------------------

5. Description of Device

TrusFIL-Flow Flowable Composite Restorative is a visible light curing dental restorative material intended to restore carious lesions or structural defects in teeth in combination of a bonding agent and with or without other materials such as luting, etching agents, cavity liners, universal composites and others commonly used in a tooth restoration. The principal organic components are mixtures of dental methacrylate resins (Bis-GMA, TEGDMA, EBPADMA). The inorganic filler loading is about 46% by volume having particle size range of about 0.01 to 3 microns. It is packaged in syringes and single-dose capsules. Single-dose delivery is intended for single patient (single use) only to prevent cross-contamination between patients.

Available shades include opaque dentin shades and regular body shades. All shades are radiopaque.

6. Indications for Use

TrusFIL-Flow Flowable Composite Restorative is a visible light curing dental restorative material indicated for:

  • Class III and V restorations

  • Restoration of minimally invasive cavity preparations (including small, non stress-bearing occlusal restorations)

  • Base/liner under direct restorations

  • Repair of small defects in indirect restorations

  • Pit and fissure sealant

  • Undercut blockout

  • Repair of resin and acrylic temporary materials

7. Summary of Physical Properties Tests

TrusFIL-Flow Flowable Composite Restorative is classified as Type 1 Class 2 Group 1 in accordance with ISO 4049: 2019. The physical properties were determined and tested according to ISO 4049: 2019 and FDA guidance "Guidance for Industry and FDA Staff Dental Composite Resin Devices -Premarket Notification [510(k)|Submissions Document issued on: October 26, 2005".

{5}------------------------------------------------

HUGE

In-vitro bench tests were performed on the TrusFIL-Flow Flowable Composite Restorative including the items listed in the table below. The test results indicated that the TrusFIL-Flow Flowable Composite Restorative meets the pass/fail criteria and supports substantial equivalence when compared to the predicate device on physical properties.

ItemsPass/fail criteriaConclusion
ISO 4049: 20195.2.7 Sensitivity to ambient light,Class 2 materialsThe material shall remain physicallyhomogeneous.Meet the criteria per ISO 4049
ISO 4049: 20195.2.8 Depth of cure, Class 2 materialsOpaque restorative materials: ≥1 mmOther restorative materials: ≥1.5 mmMeet the criteria per ISO 4049
ISO 4049: 20195.2.9 Flexural strengthThe flexural strength shall be equal to orgreater than 80 MPa.Meet the criteria per ISO 4049
ISO 4049: 20195.2.10 Water sorption and solubilitya) The water sorption shall be ≤ 40ug/mm³.b) The solubility shall be shall be ≤ 7.5ug/mm3.Meet the criteria per ISO 4049
ISO 4049: 20195.3 Shade, restorative materialsThe shade of the set material shall matchclosely that of the manufacturer's shadeguide. The set material shall be evenlypigmented when viewed withoutmagnification.Meet the criteria per ISO 4049
ISO 4049: 20195.4 Colour stability after irradiationand water sorptionNo more than a slight change in colourshall be observed.Meet the criteria per ISO 4049
ISO 4049: 20195.5 Radio-opacityThe radio-opacity shall be equal to orgreater than that of the same thickness ofaluminium (1 mm of material).Meet the criteria per ISO 4049
FDA guidanceElastic modulusThe elastic modulus shall be equal to orgreater than 2 GPa.Meet the criteria per internalstandard
FDA guidanceSurface hardnessThe surface hardness shall be equal to orgreater than 30HV.Meet the criteria per internalstandard
ItemsPass/fail criteriaConclusion
FDA guidanceCompressive strengthThe compressive strength shall be equalto or greater than 200 MPa.Meet the criteria per internalstandard

{6}------------------------------------------------

미래문

8. Technological Characteristics

The following table shows the significant technological characteristics and indicates the

similarities and differences with the predicate device:

Subject devicePrimary predicate device
TechnologicalCharacteristicsTrusFIL-Flow Flowable CompositeRestorative(K221510)Filtek Supreme Ultra FlowableRestorative(K100235)
Compositionof MaterialsMethacrylate-based resins, photo initiator,fillers and pigmentsMethacrylate-based resins, photo initiator,fillers and pigments
Physical FormFlowable PasteFlowable Paste
Indications of UseTrusFIL-Flow Flowable CompositeRestorative is a visible light curing dentalrestorative material indicated for:- Class III and V restorations- Restoration of minimally invasive cavitypreparations (including small, nonstress-bearingocclusal restorations)- Base/liner under direct restorations- Repair of small defects in indirectrestorations- Pit and fissure sealant- Undercut blockout- Repair of resin and acrylic temporarymaterialsFiltek Supreme Flow restorative is indicatedfor use in:- Class III and V restorations-Restoration of minimally invasive cavitypreparations (including small, nonstress-bearingocclusal restorations)-Base/liner under direct restorations- Repair of small defects in indirectrestorations- Pit and fissure sealant- Undercut blockout-Repair of resin and acrylic temporarymaterials
Prescription/over-the-counter usePrescriptionPrescription
Curing methodLight cureLight cure
Delivery formSyringe and single-dose capsuleSyringe and single-dose capsule
Radio-opacityMeet the criteria per ISO 4049Meet the criteria per ISO 4049
Physical PropertiesThe subject device and the predicate device have substantially equivalent physicalproperties as they all meet the criteria per ISO 4049 and conform to FDA guidance"Guidance for Industry and FDA Staff Dental Composite Resin Devices -PremarketNotification [510(k)]Submissions Document issued on: October 26, 2005".
TechnologicalCharacteristicsSubject devicePrimary predicate device
TrusFIL-Flow Flowable CompositeRestorative(K221510)Filtek Supreme Ultra FlowableRestorative(K100235)
FDA-RecognizedStandardsISO 4049; ISO 7405; ISO 10993-1ISO 4049; ISO 7405; ISO 10993-1

{7}------------------------------------------------

All compositions of the subject device are based upon industry well-known chemistry. The technological characteristics of the subject device(K221510) are very similar to those of the predicate device, Filtek Supreme Ultra Flowable Restorative(K100235). The subject device is a similar product, manufactured with similar materials and used in the same way by the same types of users and patient populations. The subject device and primary predicate device have minor different Indications for Use in language. However, the difference does not change the intended use or substantial equivalence, both products are intended for direct anterior and posterior restorations. Besides, other comparison items such as description of material, physical form, curing method, delivery form and physical properties, etc. are the same or very similar. And both products are supplied for prescription use.

9. Summary of Biocompatibility

The subject device, TrusFIL-Flow Flowable Composite Restorative, is substantially equivalent to the predicate device that have been legally marketed for decades and with no clinical adverse events. The formulation of the subject device does not contain any non-conventional chemicals compared to the legally marketed predicate device.

Biocompatibility tests were performed fully following the ISO 10993 standards. The test items include Cytotoxicity, Sensitization, Irritation, Systemic Toxicity, Subchronic Toxicity, Genotoxicity and Pulp and Dentine Usage.

10. Clinical Performance Data

Not applicable. Clinical performance testing has not been performed on the subject device.

11. Risk Analysis

The risk analysis of TrusFIL-Flow Flowable Composite Restorative was conducted according to ISO 14971. As analyzed, the residual risk of the TrusFIL-Flow Flowable Composite Restorative is considered acceptable. The benefits of the product are considered to outweigh the

{8}------------------------------------------------

HUGE

risks outlined in the risk analysis.

12. Conclusions

Based on the indications for use, technological characteristics, performance testing and comparison to predicate device, the subject device has been shown to be as safe and as effective as the predicate device for its intended use and the minor differences in indications for use fall within the intended use of the predicate device affecting neither the general intended use nor substantial equivalence. Rizhao HuGe Biomaterials Company, Ltd. concludes that the subject device is substantially equivalent to the predicate device described herein.

§ 872.3690 Tooth shade resin material.

(a)
Identification. Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.(b)
Classification. Class II.