(103 days)
The MyClearALIGN Dental Aligner System is indicated for the alignment of all permanent teeth/full permanent dentition during orthodontic treatment of malocclusions by way of continuous gentle forces.
MyClearALIGN Dental Aligner System device is a clear aligner system fabricated using the commercially available Zendura FLX, a clear, thin thermoformed Copolyester – Polyurethane Composite with 2 outer layers of cycloaliphatic polyester and an inner layer of polyether polyurethane. The pressure areas create forces that combine to form couples and thus a moment around or near the center of resistance or center of rotation of the tooth. A series of aligners is fabricated to sequentially move the teeth from their original positions of better alignment. Each set of aligners is designed to be replaced after approximately 2 weeks of 20-22 hours of wear per day (Approximately 300 hours per aligner). Each tooth is moved no greater than 0.25mm in one aligner stage (approximately 2 weeks of 20-22 hours of wear per day).
Here's an analysis of the provided text regarding acceptance criteria and device performance.
Important Note: The provided document is an FDA 510(k) clearance letter and its associated summary. This type of document focuses on establishing substantial equivalence to a predicate device, rather than proving a device meets specific, quantitative performance acceptance criteria through the kind of study structure typically seen for novel diagnostic AI/ML devices. Therefore, many of the requested points related to AI/ML study design (e.g., sample size for test set, expert adjudication, MRMC studies, ground truth for training) are not applicable to this type of medical device (physical dental aligners).
This document demonstrates the MyClearALIGN Dental Aligner System's equivalence to a predicate device primarily through bench testing and biocompatibility studies, not performance studies based on human interpretation or AI algorithm output. There is no mention of an AI/ML component to this device.
Acceptance Criteria and Study for MyClearALIGN Dental Aligner System
Based on the provided FDA 510(k) summary for the MyClearALIGN Dental Aligner System (K221475), the device is cleared based on its substantial equivalence to a predicate device (ULab Systems Dental Aligner Kit, K192596), not on meeting specific quantitative performance acceptance criteria like those used for AI/ML diagnostic tools.
The "acceptance criteria" here are implicitly satisfied by demonstrating that the device is "as safe, as effective, and performs as well as the predicate device" through non-clinical (biocompatibility and bench) testing.
1. Table of Acceptance Criteria and Reported Device Performance
Given the nature of the device (physical dental aligner) and the 510(k) pathway, direct quantitative "acceptance criteria" for diagnostic performance (e.g., sensitivity, specificity, AUC) are not applicable. Instead, the acceptance is based on demonstrating equivalence in materials, function, and safety to a legally marketed predicate.
| Acceptance Criterion (Implicit) | Reported Device Performance/Conclusion |
|---|---|
| Biocompatibility: Device materials are safe for human contact. | Biocompatibility testing conducted per ISO 10993 series (Parts 3, 5, 10, 11) demonstrated the device is non-toxic. |
| Physical Performance/Durability: Material properties, strength, and resistance to environmental factors are adequate for intended use. | Various bench tests were performed on the Zendura FLX material, including: - Trouser Tear strength - Light Transmission and Haze - Thermoforming verification - Compatibility with Commercial Denture Disinfectant - Exposure to Common Beverages - Extended Saline Exposure (37°C and 47°C) - Outdoor Exposure Effects - Stain resistance - Flexural Properties - Tensile & Elongation - Stress Relaxation Conclusion: Results of all testing demonstrate the device is comparable to other currently marketed devices and is substantially equivalent. |
| Indications for Use Equivalence: Device performs the same function for the same patient population as the predicate. | The MyClearALIGN Dental Aligner System is indicated for "alignment of all permanent teeth/full permanent dentition during orthodontic treatment of malocclusions by way of continuous gentle forces," which is stated to be the same as the predicate device. |
| Technological Equivalence: Device operates on the same scientific principles and has similar design features as the predicate. | Both are clear plastic sequential aligners, have the same intended/indications for use, same principle of operation, fabricated from digital scans, and made of thermoplastic (Zendura-FLX). Conclusion: Substantially equivalent in terms of technological characteristics, features, specifications, materials, and mode of operation. |
| Safety and Effectiveness: Overall, the device is as safe and effective as the predicate. | Conclusion: "MyClearALIGN Dental Aligner System is as safe, as effective, and performs as well as the predicate device." |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Test Set: Not applicable in the context of this 510(k) submission for a physical device. The "test set" here refers to samples of the material and finished product used in bench testing (e.g., samples for tensile strength, biocompatibility). The document does not specify the exact number of samples for each bench test, but implies that sufficient samples were tested to gain confidence in the material properties and device performance.
- Data Provenance: Not directly applicable as this isn't clinical performance data from patients. The testing was conducted "in accordance with and in conformance to applicable device regulations and guidance," implying standard laboratory testing practices.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
Not applicable. This device is not an AI/ML diagnostic tool requiring expert-established ground truth. The "ground truth" for the non-clinical tests would be the established scientific and engineering principles for material properties and biocompatibility as defined by the referenced ISO standards and design verification reports.
4. Adjudication Method for the Test Set
Not applicable. There's no human interpretation or adjudication in the context of the bench and biocompatibility tests conducted for this device.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
No. A MRMC study is relevant for evaluating the performance of diagnostic imaging devices or AI tools that assist human readers. This product is a physical dental aligner.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
No. This is not an algorithm or software device.
7. The Type of Ground Truth Used
- For Biocompatibility: Ground truth is established by adherence to recognized international standards (ISO 10993 series) and their defined toxicological endpoints.
- For Performance Bench Testing: Ground truth is established by engineering specifications, material science principles, and adherence to internal design verification reports and standard testing methodologies (e.g., for tensile strength, flexural properties, chemical resistance).
8. The Sample Size for the Training Set
Not applicable. This is not an AI/ML device that requires a training set.
9. How the Ground Truth for the Training Set Was Established
Not applicable.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 31, 2022
Key Dental Technologies, LLC % Thomas Padula Vice President of Regulatory Compliance Schiff & Company, Inc. 583 Mountain Avenue North Caldwell, New Jersey 07006
Re: K221475
Trade/Device Name: MyClearALIGN Dental Aligner System Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic Plastic Bracket Regulatory Class: Class II Product Code: NXC Dated: May 19, 2022 Received: July 19, 2022
Dear Thomas Padula:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
Submission Number (if known)
Device Name
MyClearALIGN Dental Aligner System
Indications for Use (Describe)
The MyClearALIGN Dental Aligner System is indicated for the alignment of all permanent teeth/full permanent dentition during orthodontic treatment of may of continuous gentle forces.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| 510(k) #: | K221475 | 510(k) Summary | Prepared on: 2022-07-18 |
|---|---|---|---|
| Contact Details | 21 CFR 807.92(a)(1) | ||
| Applicant Name | Key Dental Technologies, LLC | ||
| Applicant Address | 920 MLK Blvd Suite 920-B Chapel Hill NC 27517 United States | ||
| Applicant Contact Telephone | 919-259-2280 | ||
| Applicant Contact | Dr. Larry Moray | ||
| Applicant Contact Email | dr.larrymoray@myorthodontistus.com | ||
| Correspondent Name | Schiff & Company, Inc. | ||
| Correspondent Address | 583 Mountain Avenue North Caldwell NJ 07006 United States | ||
| Correspondent Contact Telephone | 201-317-8810 | ||
| Correspondent Contact | Mr. Thomas Padula | ||
| Correspondent Contact Email | thomaspadula@schiffandcompany.com | ||
| Device Name | 21 CFR 807.92(a)(2) | ||
| Device Trade Name | MyClearALIGN Dental Aligner System | ||
| Common Name | Orthodontic plastic bracket | ||
| Classification Name | Aligner, Sequential | ||
| Regulation Number | 872.5470 | ||
| Product Code | NXC | ||
| Legally Marketed Predicate Devices | |||
| 21 CFR 807.92(a)(3) | |||
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code | |
| K192596 | ULab Systems Dental Aligner Kit | NXC | |
| Device Description Summary | 21 CFR 807.92(a)(4) | ||
| MyClearALIGN Dental Aligner System device is a clear aligner system fabricated using the commercially available Zendura FLX, a clear,thin thermoformed Copolyester – Polyurethane Composite with 2 outer layers of cycloaliphatic polyester and an inner layer of polyetherpolyurethane. The pressure areas create forces that combine to form couples and thus a moment around or near the center of resistanceor center of rotation of the tooth. A series of aligners is fabricated to sequentially move the teeth from their original positionsof better alignment. Each set of aligners is designed to be replaced afterapproximately 2 weeks of 20-22 hours of wear per day (Approximately 300 hours per aligner). Each tooth is moved no greater than0.25mm in one aligner stage (approximately 2 weeks of 20-22 hours of wear per day). | |||
| Intended Use/Indications for Use | 21 CFR 807.92(a)(5) |
The MyClearALIGN Dental Aligner System is indicated for the alignment teeth/full permanent dentition during
{4}------------------------------------------------
orthodontic treatment of malocclusions by way of continuous gentle forces.
Indications for Use Comparison
The indications for use are the same as the predicate device.
Technological Comparison
The subject and predicate device are based on the following same technological elements:
· Both devices are clear plastic sequential aligners.
· Both devices have the same intended use for alignment of teeth during orthodontic treatment of malocclusions by way of continuous gentle forces.
· Both devices have the same indications for use.
· Both devices have the same principle of operation.
- · Both devices are fabricated based on an orthodontic treatment plan designer manufacturer's technicians from digital
- scans of the patient's teeth in an untreated state.
- · Both devices are made of thermoplastic (Zendura-FLX).
MyClearALIGN Dental Aligner System is substantially equivalent in terms of the technological characteristics, features, specifications, materials, mode of operation and indications for use, to ul ab Systems Dental Aligner K192596, cleared for marketing under 510(K).
Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
A series of studies were completed to demonstrate the substantial equivalence of MyClearALIGN Dental Aligner System to the predicate device. All testing was conducted in accordance with and in conformance to applicable device regulations and guidance. Results of all testing demonstrate the device is non-toxic, is comparable to other currently marketed devices and is substantially equivalent to legally marketed predicate and included:
Biocompatibility
• ISO 1093-3:2014, biological evaluation of medical devices – part 3: tests for genotoxicity and reproductive toxicity.
· ISO 10993-5:2009, biological evaluation of medical devices -- part 5: tests for in vitro cytotoxicity.
· ISO 10993-10 : 2010, biological evaluation of medical devices - part 10: tests for irritation and skin sensitization. (Biocompatibility).
• ISO 10993-11 Third edition 2017-09 biological evaluation of medical devices part 11: Tests for systemic toxicity
Performance Testing Bench
The following Bench testing was performed:
· FLX E2 Zendura XL Design Verification Report_TROUSER TEAR_#18-14 Rev 1
- FLX E3 Zendura XL Design Verification Report Light Transmission and Haze #18-10 Rev 1
- · FLX E4 Zendura XL Thermoforming Verification Report #18-15 Rev 1
- · FLX E17 Zendura XL Report Compatible with Commercial Denture Disinfectant #17-31
- · FLX E18 Zendura XL Report Exposure to Common Beverages #17-022
- FLX E19 Zendura XL Report Extended 37C Saline Exposure #17-33
- FLX E20 Zendura XL Report Extended 47C Saline Exposure 17-26
- · FLX E21 Zendura XL Report Outdoor Exposure Effects#17-024
- Zendura FLX Stain _25Jan2018
- · Zendura XL Design Verification Report Flexural Properties #17-44 Rev 1
- Zendura XL Design Verification Report Tensile&Elongation #17-46 Rev 1
- · Zendura XL Design Verification Report- Stain Resistance #18-05 Rev1
- Zendura XL Design Verification Report- Stress Relaxation #18-24 Rev 1
Performance Testing Clinical There were no clinical studies performed.
MyClearALIGN Dental Aligner System has the same indications for use and technology characteristics as the predicate device. MyClearALIGN Dental Aligner System is as safe, as effective, and performs as well as the predicate device.
21 CFR 807.92(a)(5)
21 CFR 807.92(a)(6)
§ 872.5470 Orthodontic plastic bracket.
(a)
Identification. An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.(b)
Classification. Class II.