K Number
K221425

Validate with FDA (Live)

Device Name
Primary Relief
Manufacturer
Date Cleared
2022-09-13

(120 days)

Product Code
Regulation Number
882.5890
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Primary Relief is a percutaneous electrical nerve stimulator (PENS) system indicated for up to 3 days for symptomatic relief of post-operative pain following cesarean section delivery or up to 3 days adjunctive symptomatic relief of post-operative pain following cardiac surgery.

Device Description

The Primary Relief is designed to aid in the reduction of post-operative pain by the method of cranial electrical stimulation at the auricular stimulation points. The Primary Relief is a wearable, battery-operated device that is designed to administer periodical low level electrical pulses to the ear over three days / 72 hours from the time of activation of the device. The electrical pulse from the device will be delivered to the branches of cranial nerves on the ear through a set of wire assembly and stimulation needles. Three zinc air batteries with 1.4 V each provide the required stimulation energy for 72 hours. There stimulation electrode and one ground electrode which constitute of a needle and lead/ wire with the snap-fit ring. The stimulation needles are inserted at three specific points, which have the ability to stimulate the cranial nerves. The ground electrode is inserted at one specific point (constant in all treatments) which forms the functional earthing to the device. This constant current source guarantees equivalent stimulation energy regardless of the individual impedance of the stimulation pattern consists of rectangular pulses with differing inter-pulse intervals. A 3-pin connector is provided, which is used to check the output voltage of the device once it is activated and before applying to the patient with any one of the voltage measuring devices available in the market with the appropriate regulatory compliance.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the Primary Relief device, based on the provided FDA 510(k) summary:

Description of Acceptance Criteria and Proving Study

The acceptance criteria for the Primary Relief device appear to be primarily focused on demonstrating its effectiveness in reducing post-operative pain, specifically after cardiac surgery, to support its expanded indications for use. The study aimed to show a statistically significant reduction in pain scores compared to a sham device.


1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Implied)Reported Device Performance (Primary Endpoints)
Statistically significant reduction in pain intensity at rest.Group A (Treatment) score was 1.19 units (SE=0.16; P-value < 0.001) less than the control group (Group B), for the group-wise change in pain score at rest. This indicates a statistically significant improvement in pain relief for the treatment group at rest.
Statistically significant reduction in pain intensity during cough.Group A (Treatment) score was 1.25 units (SE=0.23; P-value < 0.001) less than the control group (Group B), for the group-wise change in the pain score during cough. This also indicates a statistically significant improvement in pain relief for the treatment group during cough.
Safety: No serious adverse events.No complications or adverse events were observed in any of the subjects during the study period across both arms. Device met safety expectations.
Effect maintained throughout the study period.The effect was maintained throughout the study period and during rest and cough situations. This suggests sustained efficacy over the 72-hour treatment period.

2. Sample size used for the test set and the data provenance

  • Sample Size: 60 subjects (29 in the intervention group and 31 in the control/sham group).
  • Data Provenance: Single center, prospective study. The country of origin is not explicitly stated, but the submitter's address is in San Mateo, California, USA, and the FDA is a US regulatory body, suggesting the study likely took place in the US or was conducted under US regulatory standards.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

The ground truth for pain was established through Numeric Rating Scale (NRS) scores reported by the patients themselves. This is a patient-reported outcome measure and does not involve expert adjudication in the traditional sense for image interpretation or diagnosis. The study notes that the primary and secondary endpoints were measured using "validated and widely used tools."


4. Adjudication method for the test set

There was no external adjudication method (like 2+1 or 3+1) described for the pain scores, as these are direct patient-reported outcomes via NRS. The study itself was a double-blinded prospective study, meaning neither the subjects nor the study staff administering the treatment (and presumably collecting pain scores) knew whether the patient received the active or sham device. This blinding helps maintain the integrity of the data collection for patient-reported outcomes.


5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No, an MRMC study was not done. This study is for a medical device (Transcutaneous Electrical Nerve Stimulator for Pain Relief), not an AI-powered diagnostic tool requiring human reader interpretation of images. Therefore, the concept of "human readers improve with AI vs without AI assistance" is not applicable here.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

  • Yes, in spirit, a standalone performance was assessed. The device (Primary Relief) delivers electrical stimulation for pain relief. Its performance is directly measured by its effect on patient pain scores. While a human applies the device, the device itself functions autonomously for 72 hours, delivering stimulation. The clinical study evaluated its direct effect (standalone effect) on pain scores compared to a sham device.

7. The type of ground truth used

The ground truth for the primary efficacy endpoint (pain intensity) was patient-reported outcomes using the Numeric Rating Scale (NRS). For secondary efficacy endpoints, the summary mentions "five measures intended to assess the pulmonary function," which would also be objective physiological measurements or patient-reported outcomes, but the specific ground truth establishment method for these is not detailed in the summary.


8. The sample size for the training set

The document does not explicitly mention a separate "training set" in the context of the clinical study for the expanded indication. The 60 subjects were part of the test set or the clinical validation study for this specific 510(k) submission. For medical devices like this, there isn't typically a "training set" in the machine learning sense for the device's functionality. The "training" for the device's design and parameters would have occurred during its development and prior predicate clearance (K213188).


9. How the ground truth for the training set was established

As there's no explicitly defined "training set" for an AI model in this context, the question regarding its ground truth establishment is not directly applicable. If we consider the development of the device itself (including its electrical parameters), the ground truth would have been established through preclinical studies, engineering specifications, and previous clinical trials that led to the predicate device's clearance (K213188), demonstrating its general safety and effectiveness for a similar indication (post-operative pain following cesarean section delivery).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

September 13, 2022

DyAnsys, Inc. Srini Nageshwar CEO 300, North Bayshore Boulevard San Mateo. California 94401

Re: K221425

Trade/Device Name: Primary Relief Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief Regulatory Class: Class II Product Code: NHI Dated: August 9, 2022 Received: August 11, 2022

Dear Srini Nageshwar:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Amber Ballard, PhD Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K221425

Device Name Primary Relief

Indications for Use (Describe)

The Primary Relief is a percutaneous electrical nerve stimulator (PENS) system indicated for up to 3 days for symptomatic relief of post-operative pain following cesarean section delivery or up to 3 days adjunctive symptomatic relief of post-operative pain following cardiac surgery.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for DyAnsys. The logo consists of a stylized red symbol to the left of the word "DyAnsys" in black font. A circled "R" symbol is located to the upper right of the word "DyAnsys", indicating that the name is a registered trademark.

510(k) SUMMARY

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR 807.87 and 21 CFR 807.92.

510(k) Number: K221425

I. SUBMITTER

Date Prepared:September 13, 2022
Name:DyAnsys, Inc
Address:300, North Bayshore Boulevard,San Mateo, CA 94401, USA
Contact Person:Srini Nageshwar
Phone Number:408.480.4700
Facsimile Number:(650)-556-1621

II. DEVICE INFORMATION

Trade Name:Primary Relief
Classification Name:Percutaneous Electrical Nerve Stimulation (PENS) device
(21 CFR 882.5890)
Device Class:Class II
Product Code:NHI

III. PREDICATE DEVICE INFORMATION

Model Name: Primary Relief (K213188) Manufacturer: DyAnsys, Inc. Classification Name: Percutaneous Electrical Nerve Stimulation (PENS) devices (21 CFR 882.5890) Device Class: Class II Product Code: NHI

The predicate device has not been subject to a design-related recall.

IV. DEVICE DESCRIPTION

The Primary Relief is designed to aid in the reduction of post-operative pain by the method of cranial electrical stimulation at the auricular stimulation points. The Primary Relief is a wearable, battery-operated device that is designed to administer periodical low level electrical pulses to the ear over three days / 72 hours from the time of activation of the device.

The electrical pulse from the device will be delivered to the branches of cranial nerves on the ear through a set of wire assembly and stimulation needles. Three zinc air batteries with 1.4 V each provide the required stimulation energy for 72 hours. There stimulation electrode and one ground electrode which constitute of a needle and lead/ wire with the snap-fit ring. The stimulation needles are inserted at three specific points, which have the ability to stimulate the cranial nerves.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for DyAnsys. The logo consists of a stylized red symbol resembling an arrow or abstract shape, followed by the company name "DyAnsys" in bold, black font. A registered trademark symbol is placed to the upper right of the company name.

The ground electrode is inserted at one specific point (constant in all treatments) which forms the functional earthing to the device. This constant current source guarantees equivalent stimulation energy regardless of the individual impedance of the stimulation pattern consists of rectangular pulses with differing inter-pulse intervals.

A 3-pin connector is provided, which is used to check the output voltage of the device once it is activated and before applying to the patient with any one of the voltage measuring devices available in the market with the appropriate regulatory compliance

V. INDICATIONS FOR USE

The Primary Relief is a percutaneous electrical nerve stimulator (PENS) system indicated for up to 3 days for symptomatic relief of post-operative pain following cesarean section delivery or up to 3 days adjunctive symptomatic relief of post-operative pain following cardiac surgery..

The Indications for Use statement of Primary Relief is not identical to that of the predicate device; The only difference is the expansion in indications for use which is adequately supported through the clinical study results. The difference does not alter the intended therapeutic use of the device nor do they affect the safety and effectiveness of the device relative to the predicate. Both the subject and predicate devices have the same intended use of symptomatic relief and management of pain through the delivery of short-term electrical stimulation.

VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

The DyAnsys, Inc Primary Relief system is substantially equivalent to the legally marketed predicate device Primary Relief(K213188). It was evaluated through performance and non-clinical testing.

MODEL NAMEPrimary Relief(Subject)Primary Relief(K213188 - Predicate)
MANUFACTURERDyAnsys Inc.DyAnsys Inc.
Intended Use / Indications for Use Statement
INDICATIONS FOR USEThe Primary Relief is apercutaneous electrical nervestimulator (PENS) systemindicated for up to 3 days forsymptomatic relief of post-operative pain following cesareansection delivery or up to 3 daysadjunctive symptomatic relief ofpost-operative pain followingcardiac surgery.The Primary Relief is apercutaneous electrical nervestimulator (PENS) systemindicated for up to 3 days forsymptomatic relief of post-operative pain followingcesarean section delivery.
PATIENT POPULATIONAdultsAdults

Table 6.1: Comparison of Technological Characteristics

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for DyAnsys. The logo consists of a stylized red symbol resembling an arrow pointing to the right, followed by the text "DyAnsys" in black. A registered trademark symbol is present to the upper right of the text.

SINGLE USEELECTRODESYesYes
PORTABLE COMPONENTSYesYes

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image contains the logo for DyAnsys. The logo consists of a stylized red symbol resembling an arrow or a waveform, followed by the text "DyAnsys" in a clean, sans-serif font. A registered trademark symbol is placed to the upper right of the word "DyAnsys".

TECHNOLOGICAL CHARACTERISTICS
PULSE GENERATOR
PRODUCT DIMENSION(mm)50 * 23 * 750 * 23 * 7
MASS (g)8 g (including battery)8 g (including battery)
PRODUCT SHAPERectangleRectangle
POWER
FREQUENCY (Hz)1 - 100 (Sweep Pattern)1 - 100 (Sweep Pattern)
WAVEFORMBiphasic withRectangular PulseBiphasic withRectangular Pulse
BATTERY TYPEP13 Zinc Air batteries that are non-rechargeableP13 Zinc Air batteries
BATTERY CAPACITY310 mAh310 mAh
NO. x VOLTAGE (V)3 x 1.4 V3 x 1.4 V
PULSE WIDTH (ms)~ 0.980~ 0.980
DUTY CYCLEContinuous StimulationContinuous Stimulation
TYPICAL BATTERYOPERATING TIME (hours)7272
ENVIRONMENTAL
OPERATINGTEMPERATURE5 °C to 45 °CNot publicly available
OPERATING HUMIDITY40% to 80%Not publicly available
ENVIRONMENT OF USEClinics, Hospital and HomeenvironmentsClinics, Hospital and Homeenvironments
STERILIZATION OFELECTRODESEtO SterilizationEtO Sterilization
RE-USESingle use DeviceSingle use Device
SHELF LIFE6 months6 months
ACCESSORIES- Primary Relief with non activated batteries- Top Cover of Primary Relief- Sterile pack of needles for use with the Primary Relief device- Adhesive to fasten the needles- Adhesive for the Primary Relief deviceNot publicly available

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for DyAnsys. The logo consists of a stylized red symbol resembling an arrow pointing upwards, followed by the text "DyAnsys" in black font. A circled "R" symbol, indicating a registered trademark, is positioned to the upper right of the text.

Table 6.2: Comparison of System Characteristics

System CharacteristicPrimary Relief(Subject)Primary Relief(K213188 - Predicate)
PULSE GENERATOR
Software ControlledYesYes
Mass (g)8 g8 g
Dimensions(mm)50 * 23 * 750 * 23 * 7
Housing MaterialABS Plastic MaterialABS Plastic Material
ELECTRODE NEEDLE ARRAYS
Introduction MethodPercutaneousPercutaneous
Electrode ConfigurationSingle needle in all the 4 leadsSingle needle in all the 4 leads
Tissue ContactSkin/TissueSkin/Tissue
Electrode TypePercutaneous fine needlePercutaneous fine needle
StimulatingElectrode MaterialTitaniumTitanium
StimulatingElectrode ShapeStraightStraight
Number of leads4 (3 signal and 1 ground)4 (3 signal and 1 ground)
Needle Dimensions -Diameter (mm) xLength (mm)0.4 x 20.4 x 2
Surface area ofneedle (cm2)0.0201 cm20.0201 cm2
Supplied SterileYesYes
SYSTEM CHARACTERISTICS (OUTPUT SPECS)
Net Charge(microcoulomb (µC) perpulse;0 (Due to biphasic nature of thewaveform)Not publicly available
Where and how it isusedLow levels of electrical current aredelivered by the externalstimulator through electrodesplaced percutaneously on the ear totarget the associated cranial andoccipital nerves.Low levels of electrical current aredelivered by the external stimulatorthrough electrodes placedpercutaneously on the ear to targetthe associated cranial and occipitalnerves.
Anodes / Cathodesplease identify thegroundThe ground electrode of PrimaryRelief is a single isolated wireseparated from the assembly ofsignal wires, which creates aclosed circuit for safe and properfunctioning of stimulation device.The ground electrode of PrimaryRelief is a single isolated wireseparated from the assembly ofsignal wires, which creates a closedcircuit for safe and properfunctioning of stimulation device.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for DyAnsys. The logo consists of a red and black abstract symbol on the left, followed by the text "DyAnsys" in black. A registered trademark symbol is located to the upper right of the text.

The subject device. Primary Relief, has identical intended use, form factor, technological characteristics, mechanism of action and operating principles compared to that of the predicate device. Both devices are intended to be a prescription (Rx) devices for use by or on the order of a licensed healthcare practitioner. Both devices are body-worn, have similar indications for use and deliver electrical stimulation therapy for symptomatic relief and management of post-operative pain. Both devices deliver biphasic electrical stimulation waveforms hence are charge balanced due to the a positive and negative phase between active electrode(s) and the ground electrode.

The purpose of this submission is to expand the indications of use to include adjunctive use in patients with post- operative pain following cardiac surgery along with the already 510(k) cleared indication for post- operative pain following cesarean section delivery (K213188). The electrical stimulation parameters and the other characteristics remain identical to the predicate device.

The results from the clinical study (summarized below) have demonstrated the safety of the therapy and there were no serious adverse events observed during the study period.

VII. Performance Data

The nonclinical testing of Primary Relief device included biocompatibility testing, electrical safety (electromagnetic compatibility and safety), performance bench testing and software verification and validation.

The Primary Relief device and its components are subjected to performance bench testing to validate the effectiveness of each unit. The final product testing is performed to verify and compare the effectual output along with that of the predicate device. The functional test is performed for 72 hours to monitor the continuous performance. The pulse duration, amplitude and current values are captured for the Primary Relief device. The Primary Relief has equivalent performance specifications when compared to the predicate device.

Clinical testing results presented with this submission have demonstrated the safety of the therapy. No serious adverse events were observed during the study.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for DyAnsys, a medical device company. The logo consists of a stylized red symbol resembling an abstract letter 'A' or an arrow pointing upwards, followed by the company name "DyAnsys" in black, sans-serif font. A registered trademark symbol is placed to the upper right of the company name.

VIII. Summary of Clinical Information

This was a single center, double arm, randomized, double-blinded prospective study involving 60 subjects who have undergone elective cardiac surgery and satisfied all the inclusion criteria. The subjects were split into the intervention group (29 patients: mean age = 53.86 ± 11.3 vears) who received the Primary Relief device (Group A - treatment arm) and the control group (mean age = 54.90 ± 9.2 years) who received the sham device (Group B - identical to the subject device but delivers no stimulation and acts as placebo). The subject and sham devices were placed on the patients post-operatively prior to being awoken from anesthesia. Analgesics were administered upon patients' request to both treatment and control groups.

The primary efficacy endpoint is pain intensity at rest and on cough (measured through NRS -Numeric Rating Scale of 11 points with 0 signifying no pain and 10 being the worst pain imaginable) and the secondary efficacy endpoints included five measures intended to assess the pulmonary function. Post-operative baseline pain scores were obtained at 6-8 hours following placement of the subject and sham devices upon waking of the subjects from anesthesia. The primary and secondary endpoints were measured according to validated and widely used tools at 9 intervals spanning 72 hours.

The analysis showed that minimally invasive percutaneous electrical nerve stimulation treatment using Primary Relief device reduces the pain score compared to that of the control group :

  • · On average, the Group A score was 1.19 units (SE=0.16; P-value < 0.001) less than the control group (Group B), for the group-wise change in pain score at rest.
  • · On average, the Group A score was 1.25 units (SE=0.23; P-value < 0.001) less than the control group (Group B), for the group-wise change in the pain score during cough.

The effect was maintained throughout the study period and during rest and cough situations.

No complications or adverse events were observed in any of the subjects during the study period across both arms.

IX. Conclusion

Primary Relief system has been shown to be substantially equivalent to the identified predicate device based on comparison of device classification, intended use, indications for use statement and basic operating principle. Clinical and non-clinical testing shows the suitability of Primary Relief for its intended use of relieving post-operative pain following C-section delivery and for adjunctive use following cardiac surgery.

§ 882.5890 Transcutaneous electrical nerve stimulator for pain relief.

(a)
Identification. A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.(b)
Classification. Class II (performance standards).