K Number
K221346

Validate with FDA (Live)

Manufacturer
Date Cleared
2022-07-08

(60 days)

Product Code
Regulation Number
888.3030
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Stryker VariAx 2 MIS Calcaneus is indicated for fractures of the calcaneus.

Device Description

This Traditional 510(k) submission is submitted to the U.S. FDA to provide authorization to market the VariAx 2 MIS Calcaneus plates. The goal of the Subject Device VariAx 2 MIS Calcaneus is to offer a comprehensive range of calcaneal implants and instruments in one system utilizing the VariAx 2 platform technology and SOMA database for the design of the plates.

The Subject Device VariAx 2 MIS Calcaneus has an equivalent Intended Use and Indications for Use as the Predicate Device and shares the same principle of operation.

Stryker VariAx 2 MIS Calcaneus is a non-active implant(s) intended to provide temporary stabilization for bones or bone fragments.

AI/ML Overview

The provided text describes a 510(k) submission for a medical device called "Stryker VariAx 2 MIS Calcaneus." This document does not include information about AI/ML device performance or clinical studies typically associated with evaluating AI/ML models. Instead, it focuses on the substantial equivalence of a bone fixation appliance to an existing predicate device.

Therefore, I cannot provide the requested information regarding acceptance criteria and studies that prove an AI/ML device meets them, as the provided text is for a traditional medical device (metallic bone fixation appliances) and explicitly states:

"Clinical testing was not required for this submission." (Page 4)

Here's why the specific points you asked for cannot be extracted from this document:

  1. A table of acceptance criteria and the reported device performance: This document specifies non-clinical testing for mechanical properties (ASTM F382-17) and MRI compatibility, but it does not present a table of performance metrics for an AI/ML device.
  2. Sample size used for the test set and the data provenance: Not applicable.
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable.
  4. Adjudication method for the test set: Not applicable.
  5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable.
  6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
  7. The type of ground truth used: Not applicable.
  8. The sample size for the training set: Not applicable.
  9. How the ground truth for the training set was established: Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

July 8, 2022

Stryker GmbH Jonathan Schell Sr. Staff Regulatory Affairs Specialist 325 Corporate Drive Mahwah, New Jersey 07430

Re: K221346

Trade/Device Name: Stryker VariAx 2 MIS Calcaneus Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS Dated: May 6, 2022 Received: May 9, 2022

Dear Jonathan Schell:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K221346

Device Name VariAx 2 MIS Calcaneus

Indications for Use (Describe)
Stryker VariAx 2 MIS Calcaneus is indicated for fractures of the calcaneus.
Type of Use (Select one or both, as applicable)
-------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary
Proprietary Name:VariAx 2 MIS Calcaneus
Common Name:Single/multiple component metallic bone fixation appliancesand accessories (21 CFR 888.3030)
Regulation Description:Single/multiple component metallic bone fixation appliancesand accessories (21 CFR 888.3030)
Regulation Number:21 CFR 888.3030
Product Code:HRS
Device Class:Class II
Sponsor:Stryker GmbHBohnackerweg 12545 Selzach / Switzerland
Contact Person:Jonathan Schell, RACSr. Staff Regulatory Affairs Specialist325 Corporate DriveMahwah, NJ 07430Phone: 484-889-5804Fax: 201-831-6500
Date Prepared:May 6, 2022
Predicate:K063875: Stryker Foot Plating System
Additional Predicate:K132898: Zimmer Biomet A.L.P.S. Calcaneal Plating System

Description

This Traditional 510(k) submission is submitted to the U.S. FDA to provide authorization to market the VariAx 2 MIS Calcaneus plates. The goal of the Subject Device VariAx 2 MIS Calcaneus is to offer a comprehensive range of calcaneal implants and instruments in one system utilizing the VariAx 2 platform technology and SOMA database for the design of the plates.

The Subject Device VariAx 2 MIS Calcaneus has an equivalent Intended Use and Indications for Use as the Predicate Device and shares the same principle of operation.

Intended Use

Stryker VariAx 2 MIS Calcaneus is a non-active implant(s) intended to provide temporary stabilization for bones or bone fragments.

K221346 - Page 1 of 2

{4}------------------------------------------------

Indications for Use

Stryker VariAx 2 MIS Calcaneus is indicated for fractures of the calcaneus.

Summary of Technologies

A comparison of the Subject Device VariAx 2 MIS Calcaneus shows that the Subject Device is substantially equivalent to the Predicate Device regarding Intended Use, material, design, and operational principle.

Non-Clinical Testing

  • . The Subject Device(s) biocompatibility profile is equivalent to the previously cleared Reference Device(s) regarding material formulation, processing, sterilization, and geometry.
  • The Subject Device sterilization methods and parameters remain the same as the Predicate Device.
  • Verification testing (ASTM F382-17) was performed on the Subject Device and the acceptance criterion was fulfilled.
  • MR compatibility testing was conducted on the Subject Device.

Clinical Testing

Clinical testing was not required for this submission.

Conclusion

The Subject Device VariAx 2 MIS Calcaneus has an equivalent Intended Use and Indications for Use as the Predicate Device, and equivalent technological characteristics to the Predicate Device. Therefore, the information provided in this submission demonstrates substantial equivalence of the Subject Device to the Predicate Device.

§ 888.3030 Single/multiple component metallic bone fixation appliances and accessories.

(a)
Identification. Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.