K Number
K221240

Validate with FDA (Live)

Device Name
BriefCase
Date Cleared
2022-05-17

(18 days)

Product Code
Regulation Number
892.2080
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

BriefCase is a radiological computer aided triage and notification software indicated for use in the analysis of nonenhanced head CT images in adults or transitional adolescents aged 18 and older. The device is intended to assist hospital networks and appropriately trained medical specialists in workflow triage by flagging and communication of suspected positive findings of Intracranial hemorrhage (ICH) pathologies.

BriefCase uses an artificial intelligence algorithm to analyze images and highlight cases with detected findings on a standalone desktop application in parallel to the ongoing standard of care image interpretation. The user is presented with notification for cases with suspected findings. Notifications include compressed preview images that are meant for informational purposes only and not intended for diagnostic use beyond notification. The device does not alter the original medical image and is not intended to be used as a diagnostic device.

The results of BriefCase are intended to be used in conjunction with other patient information and based on their professional judgment, to assist with triage/prioritization of medical images. Notified clinicians are responsible for viewing full images per the standard of care.

Device Description

BriefCase is a radiological computer-assisted triage and notification software device. The software system is based on an algorithm programmed component and consists of a standard offthe-shelf operating system, the Microsoft Windows server 2012 64bit, and additional applications, which include PostgreSQL, DICOM module and the BriefCase Image Processing Application. The device consists of the following three modules: (1) Aidoc Hospital Server (AHS/Orchestrator) for image acquisition; (2) Aidoc Cloud Server (ACS) for image processing; and (3) Aidoc Desktop Application for workflow integration.

DICOM images are received, saved, filtered and de-identified before processing. Filtration matches metadata fields with keywords. Series are processed chronologically by running the algorithms on each series to detect suspected cases. The software then flags suspect cases by sending notifications to the desktop application, thereby facilitating triage and prioritization by the user. As the BriefCase software platform harbors several triage algorithms, the user may opt to filter out notifications by pathology, e.q., a chest radiologist may choose to filter out alerts on ICH cases, and a neuro-radiologist would opt to divert PE alerts. Where several medical centers are linked to a shared PACS, a user may read cases for a certain center but not for another, and thus may opt to filter out alerts by center. Activating the filter does not impact the order in which notifications are presented in the Aidoc desktop application.

The desktop application feed displays all incoming suspect cases, each notified case in a line. Hovering over a line in the feed pops up a compressed, low-quality, grayscale, unannotated image that is captioned "not for diagnostic use" and is displayed as a preview function. This compressed preview is meant for informational purposes only, does not contain any marking of the findings, and is not intended for primary diagnosis beyond notification.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study details for the Aidoc BriefCase device, based on the provided text:

Acceptance Criteria and Device Performance

1. A table of acceptance criteria and the reported device performance

Acceptance CriteriaReported Device Performance
Sensitivity $\ge$ 80%96.15% (95% CI: 90.44%, 98.94%)
Specificity $\ge$ 80%94.83% (95% CI: 89.08%, 98.08%)
Time-to-notificationComparable to predicate device: 33.5 seconds (mean estimate)

Study Details

2. Sample size used for the test set and the data provenance

  • Test Set Sample Size: 220 cases
  • Data Provenance: Retrospective, multicenter study from 5 US-based clinical sites.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

The information provided does not specify the number of experts used to establish ground truth or their qualifications. It only states that cases were "identified as positive both by the reviewers as well as the BriefCase device" (implying human review for ground truth).

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

The document does not explicitly state the adjudication method used for the test set.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • MRMC Study: No, a multi-reader multi-case (MRMC) comparative effectiveness study involving human readers with and without AI assistance was not conducted or reported. The study evaluated the standalone performance of the AI and its time-to-notification compared to a predicate AI device.
  • Effect Size: Not applicable, as no MRMC study with human readers was performed. The device is a "workflow triage" tool and not intended for primary diagnosis by itself.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Yes, a standalone performance evaluation of the algorithm was conducted and reported, demonstrating its sensitivity and specificity in identifying ICH. The "Primary Endpoint" results (Sensitivity and Specificity) are indicative of the algorithm's standalone performance.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The ground truth for the test set was established by "reviewers." While not explicitly stated as "expert consensus," this implies a form of human expert review. It is not mentioned that pathology or outcomes data were used as ground truth for ICH detection in this context.

8. The sample size for the training set

The document states that the subject device was trained on a "larger data set" compared to the predicate device, but the specific sample size for the training set is not provided.

9. How the ground truth for the training set was established

The document does not specify how the ground truth for the training set was established. It only mentions that the device uses an "artificial intelligence, deep-learning algorithm trained on medical images."

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG ADMINISTRATION" in blue as well. The FDA is a federal agency responsible for regulating and supervising the safety of food, drugs, and other products.

Aidoc Medical, Ltd. % John J. Smith Partner Hogan Lovells US LLP 555 Thirteenth Street WASHINGTON DC 20004

May 17, 2022

Re: K221240

Trade/Device Name: BriefCase Regulation Number: 21 CFR 892.2080 Regulation Name: Radiological computer aided triage and notification software Regulatory Class: Class II Product Code: QAS Dated: April 18, 2022 Received: April 19, 2022

Dear John J. Smith:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Jessica Lamb, Ph.D. Assistant Director Imaging Software Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known)

K221240

Device Name

BriefCase

Indications for Use (Describe)

BriefCase is a radiological computer aided triage and notification software indicated for use in the analysis of nonenhanced head CT images in adults or transitional adolescents aged 18 and older. The device is intended to assist hospital networks and appropriately trained medical specialists in workflow triage by flagging and communication of suspected positive findings of Intracranial hemorrhage (ICH) pathologies.

BriefCase uses an artificial intelligence algorithm to analyze images and highlight cases with detected findings on a standalone desktop application in parallel to the ongoing standard of care image interpretation. The user is presented with notification for cases with suspected findings. Notifications include compressed preview images that are meant for informational purposes only and not intended for diagnostic use beyond notification. The device does not alter the original medical image and is not intended to be used as a diagnostic device.

The results of BriefCase are intended to be used in conjunction with other patient information and based on their professional judgment, to assist with triage/prioritization of medical images. Notified clinicians are responsible for viewing full images per the standard of care.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.

{3}------------------------------------------------

510(k) Summarv Aidoc Medical, Ltd.'s BriefCase

Submitter:

Aidoc Medical, Ltd.92 Yigal Alon St.Tel-Aviv, IsraelPhone:+972-73-7946870
Contact Person:N. Epstein, Ph.D.
Date Prepared:April 18, 2022
Name of Device:BriefCase
Classification Name:Radiological computer-assisted triage and notification softwaredevice
Regulatory Class:Class II
Product Code:QAS (21 C.F.R. 892.2080)
Primary Predicate Device:BriefCase (ICH triage, K203508)

Device Description

BriefCase is a radiological computer-assisted triage and notification software device. The software system is based on an algorithm programmed component and consists of a standard offthe-shelf operating system, the Microsoft Windows server 2012 64bit, and additional applications, which include PostgreSQL, DICOM module and the BriefCase Image Processing Application. The device consists of the following three modules: (1) Aidoc Hospital Server (AHS/Orchestrator) for image acquisition; (2) Aidoc Cloud Server (ACS) for image processing; and (3) Aidoc Desktop Application for workflow integration.

DICOM images are received, saved, filtered and de-identified before processing. Filtration matches metadata fields with keywords. Series are processed chronologically by running the algorithms on each series to detect suspected cases. The software then flags suspect cases by sending notifications to the desktop application, thereby facilitating triage and prioritization by the user. As the BriefCase software platform harbors several triage algorithms, the user may opt to filter out notifications by pathology, e.q., a chest radiologist may choose to filter out alerts on ICH cases, and a neuro-radiologist would opt to divert PE alerts. Where several medical centers are linked to a shared PACS, a user may read cases for a certain center but not for another, and thus may opt to filter out alerts by center. Activating the filter does not impact the order in which notifications are presented in the Aidoc desktop application.

The desktop application feed displays all incoming suspect cases, each notified case in a line. Hovering over a line in the feed pops up a compressed, low-quality, grayscale, unannotated image that is captioned "not for diagnostic use" and is displayed as a preview function. This compressed preview is meant for informational purposes only, does not contain any marking of the findings, and is not intended for primary diagnosis beyond notification.

{4}------------------------------------------------

Presenting the users with worklist prioritization facilitates earlier triage by prompting the user to assess the relevant original images in the PACS. Thus, the suspect case receives attention earlier than would have been the case in the standard of care practice alone.

Intended Use / Indications for Use

BriefCase is a radiological computer aided triage and notification software indicated for use in the analysis of non-enhanced head CT images in adults or transitional adolescents aged 18 and older. The device is intended to assist hospital networks and appropriately trained medical specialists in workflow triage by flagging and communication of suspected positive findings of Intracranial hemorrhage (ICH) pathologies.

BriefCase uses an artificial intelligence algorithm to analyze images and highlight cases with detected findings on a standalone desktop application in parallel to the ongoing standard of care image interpretation. The user is presented with notification for cases with suspected findings. Notifications include compressed preview images that are meant for informational purposes only and not intended for diagnostic use beyond notification. The device does not alter the original medical image and is not intended to be used as a diagnostic device.

The results of BriefCase are intended to be used in conjunction with other patient information and based on their professional judgment, to assist with triage/prioritization of medical images. Notified clinicians are responsible for viewing full images per the standard of care.

Comparison of Technological Characteristics

The subject BriefCase for ICH triage and primary predicate BriefCase for ICH triage (K203508) are identical in most aspects and differ with respect to their algorithm performance, due to training the subject device on a larger data set. Additional minor changes have been made to reflect software and the UI.

Both the primary predicate and subject devices are radiological computer-aided triage and notification software programs. Both devices are artificial intelligence, deep-learning algorithms incorporated software packages for use with DICOM 3.0 compliant CT scanners, PACS, and radiology workstations.

Both devices are intended to aid in triage and prioritization of radiological images and utilize the same design of deep learning algorithm trained on medical images. Both devices are intended to provide the specialists with notifications and unannotated low-quality preview images of suspect studies for the purpose of preemptive triage.

The subject and predicate BriefCase devices raise the same types of safety and effectiveness questions, namely, accurate detection of findings within the processed study. It is important to note that, like the predicate, the subject device neither removes cases from the standard of care reading queue nor deprioritized cases. Both devices operate in parallel with the standard of care, which remains the default option for all cases. A table comparing the key features of the subject and the primary predicate devices is provided below.

Table 1. Key feature comparison
Predicate DeviceSubject Device
Aidoc Briefcase for ICH triage(K203508)Aidoc Briefcase for ICH triage
Intended Use /Indications forUseBriefCase is a radiological computeraided triage and notification softwareindicated for use in the analysis of non-enhanced head CT images. The deviceis intended to assist hospital networksand appropriately trained medicalspecialists in workflow triage byflagging and communication ofsuspected positive findings ofpathologies in head CT images, namelyIntracranial Hemorrhage (ICH).BriefCase uses an artificial intelligencealgorithm to analyze images andhighlight cases with detected ICH on astandalone desktop application inparallel to the ongoing standard of careimage interpretation. The user ispresented with notifications for caseswith suspected ICH findings.Notifications include compressedpreview images that are meant forinformational purposes only and notintended for diagnostic use beyondnotification. The device does not alterthe original medical image and is notintended to be used as a diagnosticdevice.The results of BriefCase are intended tobe used in conjunction with otherpatient information and based onprofessional judgment, to assist withtriage/prioritization of medical images.Notified clinicians are responsible forviewing full images per the standard ofcare.BriefCase is a radiological computeraided triage and notification softwareindicated for use in the analysis of non-enhanced head CT images in adults ortransitional adolescents aged 18 andolder. The device is intended toassist hospital networks andappropriately trained medicalspecialists in workflow triageby flagging and communication ofsuspected positive findings ofIntracranial Hemorrhage (ICH)pathologies.BriefCase uses an artificial intelligencealgorithm to analyze images andhighlight cases with detected findingson a standalone desktop application inparallel to the ongoing standard of careimage interpretation. The user ispresented with for cases withsuspected ICH findings. Notificationsinclude compressed preview imagesthat are meant for informationalpurposes only and not intended fordiagnostic use beyond notification. The device doesnot alter the original medical imageand is not intended to be used as adiagnostic device.The results of BriefCase are intendedto be used in conjunction with otherpatient information and based on theirprofessional judgment, to assist withtriage/prioritization of medical images.Notified clinicians are responsible forviewing full images per the standard ofcare.
User populationHospital networks and appropriatelytrained medical specialistsHospital networks and appropriatelytrained medical specialists
Anatomicalregion ofinterestHeadHead
DataacquisitionprotocolNon-enhanced CT scan of the head orneckNon-enhanced CT scan of the head orneck
Notification-only/notificationYesYes
alerts), parallelworkflow tool
ImagesformatDICOMDICOM
Interferencewith standardworkflowNo. No cases are removed fromWorklist or deprioritized.No. No cases are removed fromdesktop app or deprioritized
Inclusion/Exclusioncriteria forclinicalperformancetestingInclusion criteria• Non-enhanced head or neck CTprotocol with a 64-slice scanner orhigher.• Scans performed on adults/transitional adults ≥ 18 years of age.• Slice thickness 0.625mm – 5.0mm.Exclusion Criteria• All scans that are technicallyinadequate, including motionartifacts, severe metal artifacts,suboptimal bolus timing or aninadequate field of view.Inclusion criteria• Non-enhanced head or neck CTprotocol with a 64-slice scanner orhigher.• Scans performed on adults/transitional adults ≥ 18 years of age.• Slice thickness 0.625 mm - 5.0 mm.Exclusion Criteria• All scans that are technicallyinadequate, including motionartifacts, severe metal artifacts,suboptimal bolus timing or aninadequate field of view.
AlgorithmArtificial intelligence algorithm withdatabase of images.Artificial intelligence algorithm withdatabase of images.
Structure• AHS module (image acquisition);• ACS module (image processing));• Aidoc Worklist application forworkflow integration (worklist andnon-diagnostic Image Viewer).• AHS module ( image acquisition);• ACS module (image processing));• Aidoc Desktop application forworkflow integration (non-diagnostic Image Viewer).

{5}------------------------------------------------

{6}------------------------------------------------

Performance Data

Pivotal Study Summary

Aidoc conducted a retrospective, blinded, multicenter, study with the BriefCase software to evaluate the software's performance in identifying non-contrast head CT images containing intracranial hemorrhage (ICH) in 220 cases from 5 US-based clinical sites.

Primary endpoints were sensitivity and specificity with an 80% performance goal. Secondary endpoints were BriefCase time-to-notification compared to the predicate device, Positive Predictive Value (PPV), Negative Predictive Value (NPV), Positive Likelihood Ratio (PLR), and Negative Likelihood Ratio (NLR).

The mean age of patients whose scans were reviewed in the study was 65.1 years, with standard deviation of 18.8 years. Gender distribution was 49.8% male, and 50.2% female. Race distribution within the study data patient population was unavailable. None of the potential covariates demonstrated statistical significance, thus, device performance did not meaningfully interact with location, gender or age.

{7}------------------------------------------------

Primary Endpoint

Sensitivity and specificity exceeded the 80% performance goal. Sensitivity was 96.15% (95% Cl: 90.44%, 98.94%) and specificity was 94.83% (95% Cl: 89.08%, 98.08%).

Secondary Endpoint

In addition, the time-to-notification metric observed for the BriefCase software in the three medical centers was compared to the equivalent metric of the predicate devices.

The BriefCase time-to-notification includes the time to get the DICOM exam, de-identify it, upload it to the cloud, analyze and send a notification on a positive suspect case back to the desktop application.

The BriefCase time-to-notification was measured for all True Positive cases (i.e., identified as positive both by the reviewers as well as the BriefCase device) and is given in Table 2 below. The Table also displays the same metric reported for the predicate BriefCase ICH.

The time-to-notification results obtained for the subject BriefCase device show comparability with the primary predicate with regard to time savings to the standard of care review.

Time -to-notification(in seconds)MeanEstimate95% LowerCL95% UpperCLMedianIQR
Predicate K203508Processing Time267.6246.0289.8N/AN/A
BriefCase Time-to-notification33.530.936.130.420.2

Table 2. Time-to- notification comparison for BriefCase devices

NPV was 99.29% (95% Cl: 98.16% - 99.73%) and PPV was 76.64% (95% Cl: 60.06% - 87.74%).

PLR was 18.59 (95% Cl: 8.52 - 40.56) and NLR was 0.04 (95% Cl: 0.02 - 0.11)

Thus, the reported similar time-to-notification data demonstrates that when using the subject BriefCase for ICH triage the clinician may have the same benefit in time saving as with the predicate BriefCase for ICH triage.

In summary, performance goals were achieved. The time-to-notification results for the proposed device are comparable to the results for the predicate, demonstrating preemptive triage in the range of several minutes.

Conclusions

The subject BriefCase for ICH triage and the predicate BriefCase for ICH triage are intended to aid in prioritization and triage of radiological images for the indications for suspected positive findings

{8}------------------------------------------------

of Intracranial hemorrhage (ICH) pathologies. Both devices are software packages with the same technological characteristics and principles of operation, incorporating deep learning Al algorithms that process images, and software to send notifications and display unannotated compressed lowquality preview images. In both devices, the labeling clearly states that the devices are not for diagnostic use and instructs the user to further evaluate and diagnose based only on the original images in the local PACS.

Both devices operate in parallel to the standard of care workflow in the sense that they do not change the original image, do not provide any marking on the output preview, do not remove images from the standard of care FIFO queue and do not de-prioritize cases, thus not disturbing standard interpretation of the images. Both devices notify the radiologist of time-sensitive critical cases within the range of several minutes, and thus contribute similarly to the standard of care workflow turnaround time reduction through preemptive triage.

The subject BriefCase device for ICH triage is thus substantially equivalent to the primary predicate BriefCase for ICH.

§ 892.2080 Radiological computer aided triage and notification software.

(a)
Identification. Radiological computer aided triage and notification software is an image processing prescription device intended to aid in prioritization and triage of radiological medical images. The device notifies a designated list of clinicians of the availability of time sensitive radiological medical images for review based on computer aided image analysis of those images performed by the device. The device does not mark, highlight, or direct users' attention to a specific location in the original image. The device does not remove cases from a reading queue. The device operates in parallel with the standard of care, which remains the default option for all cases.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) A detailed description of the notification and triage algorithms and all underlying image analysis algorithms including, but not limited to, a detailed description of the algorithm inputs and outputs, each major component or block, how the algorithm affects or relates to clinical practice or patient care, and any algorithm limitations.
(ii) A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide effective triage (
e.g., improved time to review of prioritized images for pre-specified clinicians).(iii) Results from performance testing that demonstrate that the device will provide effective triage. The performance assessment must be based on an appropriate measure to estimate the clinical effectiveness. The test dataset must contain sufficient numbers of cases from important cohorts (
e.g., subsets defined by clinically relevant confounders, effect modifiers, associated diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals for these individual subsets can be characterized with the device for the intended use population and imaging equipment.(iv) Stand-alone performance testing protocols and results of the device.
(v) Appropriate software documentation (
e.g., device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results).(2) Labeling must include the following:
(i) A detailed description of the patient population for which the device is indicated for use;
(ii) A detailed description of the intended user and user training that addresses appropriate use protocols for the device;
(iii) Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (
e.g., poor image quality for certain subpopulations), as applicable;(iv) A detailed description of compatible imaging hardware, imaging protocols, and requirements for input images;
(v) Device operating instructions; and
(vi) A detailed summary of the performance testing, including: test methods, dataset characteristics, triage effectiveness (
e.g., improved time to review of prioritized images for pre-specified clinicians), diagnostic accuracy of algorithms informing triage decision, and results with associated statistical uncertainty (e.g., confidence intervals), including a summary of subanalyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.