K Number
K221134

Validate with FDA (Live)

Date Cleared
2022-05-19

(30 days)

Product Code
Regulation Number
878.4040
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SafeSource Direct Procedure Mask with Ear Loops is intended to be worn to protect both the patient and healthcare personnel from the transfer of microorganisms, bodily fluids, and particulate material. These procedure masks are intended for use in infection control practices to reduce to blood and bodily fluid. The procedure masks are single use, disposable devices, provided non-sterile.

The SafeSource Direct Surgical Mask with Ties is intended to be worn to protect both the patient and healthcare personnel from the transfer of microorganisms, bodly fluids, and particulate material. These surgical masks are intended for use in infection control practices to reduce the potential exposure to blood and bodily fluid. The surgical masks are single use, disposable devices, provided non-sterile.

Device Description

The Procedure Mask with Ear Loops (PM3001, PM3003) is a non-sterile, single use, three-layer, flat-folded mask with ear loops, and a nose wire. The Procedure Mask with Ear Loops is constructed of inner and outer layers consisting of spunbond polypropylene and a middle layer of meltblown polypropylene. The welded ear loops made of spandex/elastane fiber and polyester filament keep the user's mouth and nose. The ethylene propylene copolymer coated aluminum nose wire allows the user to fit the mask around their nose.

The Surgical Mask with Ties (SM3001, SM3003) is a non-sterile, single use, three-layer, flat-folded mask with ties, and a nose wire. The Surgical Mask with Ties is constructed of inner and outer layers consisting of spunbond polypropylene and a middle layer of meltblown polypropylene. The welded polypropylene tie straps keep the mask close to the user's mouth and nose. The ethylene propylene coated aluminum nose wire allows the user to fit the mask around their nose.

AI/ML Overview

This document describes the performance of a medical device, specifically Procedure Masks with Ear Loops and Surgical Masks with Ties, submitted for FDA 510(k) clearance (K221134). The device is compared to a predicate device (K202899) to establish substantial equivalence.

Here's a breakdown of the requested information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are based on ASTM F2100-19 for medical face mask performance, and other relevant standards like ASTM F1862/F1862M-17 (Fluid Resistance), ASTM F2299/F2299M-03 (Particulate Filtration Efficiency), ASTM F2101-19 (Bacterial Filtration Efficiency), EN 14683:2019 + AC:2019 Annex C (Differential Pressure), 16 CFR 1610 (Flammability), ISO 10993-5: 2009 (Cytotoxicity), and ISO 10993-10: 2010 (Irritation and Sensitization).

There are two tables provided in the document:

  • Table 1: Comparison of Performance Testing for Proposed ASTM Level 1 Procedure Mask with Ear Loops (PM3001) and Surgical Mask with Ties (SM3001) with Predicate Device
  • Table 2: Comparison of Performance Testing for Proposed ASTM Level 3 Procedure Mask with Ear Loops (PM3003) and Surgical Mask with Ties (SM3003) with Predicate Device

Table 1: ASTM Level 1 Masks (PM3001 & SM3001)

Non-Clinical TestAcceptance Criteria (ASTM Level 1)Proposed Device Performance (PM3001)Proposed Device Performance (SM3001)
Fluid Resistance≥ 29 of 32 pass at 80 mmHgPass (96/96) at 80 mmHg, 32 samples from 3 non-consecutive lotsPass (96/96) at 80 mmHg, 32 samples from 3 non-consecutive lots
Particulate Filtration Efficiency≥ 95% at 0.1 micronPass (96/96) at > 95%, 32 samples from 3 non-consecutive lotsPass (96/96) at > 95%, 32 samples from 3 non-consecutive lots
Bacterial Filtration Efficiency≥ 95%Pass (96/96) at > 95%, 32 samples from 3 non-consecutive lotsPass (96/96) at > 95%, 32 samples from 3 non-consecutive lots
Differential Pressure< 5.0 mmH2O/cm²Pass (96/96) at < 5.0 mmH2O/cm², 32 samples from 3 non-consecutive lotsPass (95/96) at < 5.0 mmH2O/cm², 32 samples from 3 non-consecutive lots
FlammabilityClass 1Pass (96/96) at Class 1, 32 samples from 3 non-consecutive lotsPass (96/96) at Class 1, 32 samples from 3 non-consecutive lots
Cytotoxicity(Not specified, assumed "Non-cytotoxic")Non-cytotoxicNon-cytotoxic
Irritation(Not specified, assumed "Non-irritating")Negligibly irritating (Primary Irritation Index 0.0)Negligibly irritating (Primary Irritation Index 0.0)
Sensitization(Not specified, assumed "Non-sensitizing")Not a contact sensitizerNot a contact sensitizer

Table 2: ASTM Level 3 Masks (PM3003 & SM3003)

Non-Clinical TestAcceptance Criteria (ASTM Level 3)Proposed Device Performance (PM3003)Proposed Device Performance (SM3003)
Fluid Resistance≥ 29 of 32 pass at 160 mmHgPass (96/96) at 160 mmHg, 32 samples from 3 non-consecutive lotsPass (96/96) at 80 mmHg, 32 samples from 3 non-consecutive lots
Particulate Filtration Efficiency≥ 98% at 0.1 micronPass (95/96) at > 98%, 32 samples from 3 non-consecutive lotsPass (95/96) at > 98%, 32 samples from 3 non-consecutive lots
Bacterial Filtration Efficiency≥ 98%Pass (96/96) at > 98%, 32 samples from 3 non-consecutive lotsPass (96/96) at > 98%, 32 samples from 3 non-consecutive lots
Differential Pressure< 6.0 mmH2O/cm²Pass (96/96) at < 6.0 mmH2O/cm², 32 samples from 3 non-consecutive lotsPass (96/96) at < 6.0 mmH2O/cm², 32 samples from 3 non-consecutive lots
FlammabilityClass 1Pass (96/96) at Class 1, 32 samples from 3 non-consecutive lotsPass (96/96) at Class 1, 32 samples from 3 non-consecutive lots
Cytotoxicity(Not specified, assumed "Non-cytotoxic")Non-cytotoxicNon-cytotoxic
Irritation(Not specified, assumed "Non-irritating")Negligibly irritating (Primary Irritation Index 0.0)Negligibly irritating (Primary Irritation Index 0.0)
Sensitization(Not specified, assumed "Non-sensitizing")Not a contact sensitizerNot a contact sensitizer

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: For each non-clinical performance test (Fluid Resistance, Particulate Filtration Efficiency, Bacterial Filtration Efficiency, Differential Pressure, Flammability), the tests were conducted on 32 samples from 3 non-consecutive lots, totaling 96 samples for each test. For cytotoxicity, irritation, and sensitization, specific sample sizes are not explicitly stated, but the results indicate evaluation was performed.
  • Data Provenance: The document does not specify the country of origin of the data. It indicates these are "non-clinical testing" results, which implies laboratory testing. It does not state whether the data is retrospective or prospective, but as it concerns performance of the newly manufactured device, it would inherently be prospective testing.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of those Experts

No information is provided regarding human experts or ground truth establishment in the context of the performance tests for the masks. The "ground truth" for these tests is the quantitative measurement against established ASTM/EN/CFR standards, which are objective and do not require expert human interpretation in the way medical image analysis might.

4. Adjudication Method for the Test Set

Not applicable. The tests are laboratory-based and yield quantitative results against defined criteria, not subjective interpretations requiring adjudication.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No. The document explicitly states: "No clinical testing is included in this submission." This type of study would typically be done for AI/CAD devices, not for basic personal protective equipment like masks where performance is evaluated through standardized material tests.

6. Standalone (i.e., algorithm only without human-in-the-loop performance) Performance

Not applicable. This is a physical medical device (mask), not an algorithm or software.

7. Type of Ground Truth Used

The ground truth for the performance tests is based on pre-defined, internationally recognized standards and test methods (e.g., ASTM, EN, CFR, ISO). Each test has a specific methodology and acceptance criterion for properties like filtration efficiency, fluid resistance, breathability, and biological safety.

8. Sample Size for the Training Set

Not applicable. This is a physical medical device, not an AI/ML algorithm that requires a training set.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set for a physical device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.

May 19, 2022

SafeSource Direct, LLC % Prithul Bom Most Repsonsible Person Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k Saint Paul, Minnesota 55114

Re: K221134

Trade/Device Name: Procedure Mask with Ear Loops and Surgical Mask with Ties Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: April 18, 2022 Received: April 19, 2022

Dear Prithul Bom:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K221134

Device Name

Procedure Mask with Ear Loops and Surgical Mask with Ties

Indications for Use (Describe)

The SafeSource Direct Procedure Mask with Ear Loops is intended to be worn to protect both the patient and healthcare personnel from the transfer of microorganisms, bodily fluids, and particulate material. These procedure masks are intended for use in infection control practices to reduce to blood and bodily fluid. The procedure masks are single use, disposable devices, provided non-sterile.

The SafeSource Direct Surgical Mask with Ties is intended to be worn to protect both the patient and healthcare personnel from the transfer of microorganisms, bodly fluids, and particulate material. These surgical masks are intended for use in infection control practices to reduce the potential exposure to blood and bodily fluid. The surgical masks are single use, disposable devices, provided non-sterile.

Type of Use (Select one or both, as applicable)
☐ Prescription Use (Part 21 CFR 801 Subpart D)☑ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

Version: 2

K221134

The following information is provided in accordance with 21 CFR 807.92 for the Premarket 510(k) Summary:

Submitter Information

Company:Robbie Wicks
Vice President / COO
SafeSource Direct, LLC
200 St Nazaire Rd.
Broussard, LA 70518 United States
Telephone: 337-210-7943
Fax: 337-202-8099
rwicks@safesourcedirect.com)
Contact:Robbie Wicks
Vice President / COO
SafeSource Direct, LLC
200 St Nazaire Rd.
Broussard, LA 70518 United States
Telephone: 337-210-7943
Fax: 337-202-8099
rwicks@safesourcedirect.com)
Date Summary Prepared:April 12, 2022
Trade Name:Procedure Mask with Ear Loops and Surgical Maskwith Ties
Common Name:Mask, surgical
Classification Name:Surgical apparel
Review Panel:General & Plastic Surgery (SU)
Regulation:878.4040
Class:Class II
Product Code:FXX

Equivalence Claimed to Predicate Device

The Procedure Mask with Ear Loops and Surgical Mask with Ties is equivalent to the Procedure Mask, Surgical Mask (K202899), manufactured by Kenpax International Limited.

Trade Name:Procedure Mask / Surgical Mask
Common Name:Surgical Mask
Classification Name:Surgical Face Mask
Product Code:FXX

{4}------------------------------------------------

Device Description

The Procedure Mask with Ear Loops (PM3001, PM3003) is a non-sterile, single use, three-layer, flat-folded mask with ear loops, and a nose wire. The Procedure Mask with Ear Loops is constructed of inner and outer layers consisting of spunbond polypropylene and a middle layer of meltblown polypropylene. The welded ear loops made of spandex/elastane fiber and polyester filament keep the user's mouth and nose. The ethylene propylene copolymer coated aluminum nose wire allows the user to fit the mask around their nose.

The Surgical Mask with Ties (SM3001, SM3003) is a non-sterile, single use, three-layer, flat-folded mask with ties, and a nose wire. The Surgical Mask with Ties is constructed of inner and outer layers consisting of spunbond polypropylene and a middle layer of meltblown polypropylene. The welded polypropylene tie straps keep the mask close to the user's mouth and nose. The ethylene propylene coated aluminum nose wire allows the user to fit the mask around their nose.

Substantial Equivalence Table
CharacteristicProposed Device: ProcedureMask with Ear Loops andSurgical Mask with TiesPrimary Predicate Device:Procedure Mask, Surgical Mask(K202899)
Intended UseTo protect both patients andhealthcare personnel from thetransfer of microorganisms, bodyfluids, and particulate material.To protect both patients andhealthcare personnel from thetransfer of microorganisms, bodyfluids, and particulate material.
Indication for UseThe SafeSource Direct ProcedureMask with Ear Loops is intended tobe worn to protect both the patientand healthcare personnel from thetransfer of microorganisms, bodilyfluids, and particulate material.These procedure masks areintended for use in infectioncontrol practices to reduce thepotential exposure to blood andbodily fluid. The procedure masksare single use, disposable devices,provided non-sterile.The SafeSource Direct SurgicalMask with Ties is intended to beworn to protect both the patient andhealthcare personnel from thetransfer of microorganisms, bodilyfluids, and particulate material.These surgical masks are intendedfor use in infection controlpractices to reduce the potentialexposure to blood and bodily fluid.The surgical masks are single use,disposable devices, provided non-The Procedure Masks/ SurgicalMasks are intended to be worn toprotect both the patient andhealthcare personnel from transferof microorganisms, body fluids,and particulate material. These facemasks are intended for use ininfection control practices toreduce the potential exposure toblood and body fluids. This is asingle use, disposable device(s),provided non-sterile.
sterile.
510(k) No.PendingK202899
ManufacturerSafeSource Direct, LLCKenpax International Limited
ClassificationClass II, FXX (21 CFR 878.4040)Class II, FXX (21 CFR 878.4040)
Model DescriptionPM3001: Procedure Mask with EarLoops, ASTM Level 1PM3003: Procedure Mask with EarLoops, ASTM Level 3SM3001: Surgical Mask with Ties,ASTM Level 1SM3003: Surgical Mask with Ties,ASTM Level 368-8506-G: Procedure Mask,Earloop, ASTM Level 165-8508-G: Procedure Mask,Earloop, ASTM Level 368-8536-B: Surgical Mask, Tie-on,ASTM Level 168-8538-B: Surgical Mask, Tie-on,ASTM Level 3
Design FeaturesEar Loops, Tie-onEar Loops, Tie-on
Mask StyleFlat PleatedFlat Pleated
Outer Facing LayerBlue spunbond (polypropylene)Level 1: 25 gsmLevel 3: 25 gsmSpunbond polypropyleneLevel 1: 22 gsmLevel 3: 25 gsm
Middle LayerWhite meltblown (polypropylene)Level 1: 25 gsmLevel 3: 25 gsmMelt-blown polypropyleneLevel 1: 22 gsmLevel 3: 25 gsm
Inner Facing LayerBlue spunbond (polypropylene)Level 1: 25 gsmLevel 3: 25 gsmSpunbond polypropyleneLevel 1: 22 gsmLevel 3: 25 gsm
Nose PieceDual aluminum wire, whiteethylene propylene copolymerPolyethylene coated steel wire
Ear LoopsWhite knitted spandex/elastanefiber, polyester filamentNylon, spandex
Tie-OnWhite spunbond (polypropylene)Spunbond polypropylene
ColorBlue, WhiteBlue, Green
Dimension (Width)3.74" ± 0.40"95 ± 5 mm
Dimension (Length)6.89" ± 0.40"178 ± 5 mm
OTC UseYesYes
SterilityNon-sterileNon-sterile
UseSingle Use, DisposableSingle Use, Disposable
ASTM F2100 LevelLevel 1, Level 3Level 1, Level 3
Fluid ResistanceMeet ASTM F1862/F1862M-17Meet ASTM F1862-17
Particulate Filtration Efficiency(PFE)Meet ASTM F2299/F2299M-03(2017)Meet ASTM F2299-17
Bacterial Filtration(BFE)Efficiency
Meet ASTM F2101-19Meet ASTM F2101-19
Differential PressureMeet EN 14683:2019+AC:2019Annex CMeet EN 14683:2019, Annex C
FlammabilityMeet 16 CFR 1610 (Oct 2008)Meet 16 CFR 1610
CytotoxicityNon-cytotoxicNon-cytotoxic
IrritationNegligibly irritating (PrimaryIrritation Index 0.0)Non-irritating
SensitizationNot a contact sensitizerNon-sensitizing

Substantial Equivalence Comparison

{5}------------------------------------------------

K221134

Page 3 of 6

{6}------------------------------------------------

Additional Information

Summary of Non-Clinical Testing Conducted

Non-clinical testing was conducted to verify the proposed device meets the same design specifications as the predicate device. Test results demonstrated the proposed device complies with the requirements of US FDA Guidance for Industry and FDA Staff / Surgical Masks - Premarket Notification [510(k)] Submissions posted on July 14, 2004 and the following standards:

  • · ASTM F2100-19, Standard Specification for Performance of Materials Used In Medical Face Masks
  • · ASTM F2101-19, Standard Test Method for Evaluating the Bacterial Filtration Efficiency (BFE) of Medical Face Mask Materials, Using A Biological Aerosol of Staphylococcus aureus
  • · EN 14683:2019 + AC:2019 Annex C, Medical Face Masks-Requirements and Test Methods
  • · ASTM F2299/F2299M-03(2017), Standard Test Method for Determining the Initial Efficiency of Materials Used in Medical Face Masks to Penetration by Particulates Using Latex Spheres
  • · ASTM F1862/F1862M-17, Standard Test Method for Resistance of Medical Face Masks To Penetration by Synthetic Blood (Horizontal Projection of Fixed Volume at a Known Velocity)
  • · 16 CFR 1610, Standard for the Flammability of Clothing Textiles
  • · ISO 10993-5: 2009, Biological Evaluation of Medical Devices -- Part 5: Tests For In Vitro Cytotoxicity, and
  • · ISO 10993-10: 2010, Biological Evaluation of Medical Devices Part 10: Tests For Irritation And Skin Sensitization.

Tables 1 and 2 summarize the performance test data.

Table 1 Comparison of Performance Testing for Proposed ASTM Level 1 ProcedureMask with Ear Loops (PM3001) and Surgical Mask with Ties (SM3001) with Predicate
Device
Non-ClinicalTestAcceptanceCriteria(ASTMLevel 1)ProposedDevicePredicateDevice(K202899)ProposedDevicePredicateDevice(K202899)
PM300168-8506-GSM300168-8536-B

{7}------------------------------------------------

K221134 Page 5 of 6

FluidResistance≥ 29 of 32pass at 80mmHgPass (96/96)at 80 mmHg,32 samplesfrom 3 non-consecutivelots32 of 32passed at 80mmHg, 3 lotsPass (96/96)at 80 mmHg,32 samplesfrom 3 non-consecutivelots32 of 32passed at 80mmHg, 3 lots
ParticulateFiltrationEfficiency≥ 95% at 0.1micronPass (96/96)at > 95%,32 samplesfrom 3 non-consecutivelots97.4%,97.5%,97.5%Pass (96/96)at > 95%,32 samplesfrom 3 non-consecutivelots97.2%,97.1%,97.1%
BacterialFiltrationEfficiency≥ 95%Pass (96/96)at > 95%, 32samples from3 non-consecutivelots99.9%, 3 lotsPass (96/96)at > 95%, 32samples from3 non-consecutivelots99.9%, 3 lots
DifferentialPressure< 5.0mmH2O/cm2Pass (96/96)at < 5.0mmH2O/cm²,32 samplesfrom 3 non-consecutivelots2.9, 2.8, 2.7mmH20/cm²Pass (95/96)at < 5.0mmH2O/cm²,32 samplesfrom 3 non-consecutivelots3.7, 3.4, 3.7mmH20/cm²
FlammabilityClass 1Pass (96/96)at Class 1, 32samples from3 non-consecutivelotsClass 1Pass (96/96)at Class 1, 32samples from3 non-consecutivelotsClass 1
CytotoxicityNon-cytotoxicNon-cytotoxicNon-cytotoxicNon-cytotoxic
IrritationNegligiblyirritating(PrimaryIrritationIndex 0.0)Non-irritatingNegligiblyirritating(PrimaryIrritationIndex 0.0)Non-irritating
SensitizationNot a contactsensitizerNon-sensitizingNot a contactsensitizerNon-sensitizing

Table 2 Comparison of Performance Testing for Proposed ASTM Level 3 Procedure Mask with Ear Loops (PM3003) and Surgical Mask with Ties (SM3003) with Predicate Device

Non-ClinicalTestAcceptanceCriteria(ASTMLevel 3)ProposedDevicePredicateDevice(K202899)ProposedDevicePredicateDevice(K202899)
PM300368-8508-GSM300368-85368-B
Fluid Resistance≥ 29 of 32 pass at 160 mmHgPass (96/96) at 160 mmHg, 32 samples from 3 non-consecutive lots32 of 32 passed at 160 mmHg, 3 lotsPass (96/96) at 80 mmHg, 32 samples from 3 non-consecutive lots32 of 32 passed at 160 mmHg, 3 lots
Particulate Filtration Efficiency≥ 98% at 0.1 micronPass (95/96) at > 98%, 32 samples from 3 non-consecutive lots98.2%, 98.4%, 98.4%Pass (95/96) at > 98%, 32 samples from 3 non-consecutive lots98.4%, 98.4%, 98.3%
Bacterial Filtration Efficiency≥ 98%Pass (96/96) at > 98%, 32 samples from 3 non-consecutive lots99.9%, 3 lotsPass (96/96) at > 98%, 32 samples from 3 non-consecutive lots99.9%, 3 lots
Differential Pressure< 6.0 mmH2O/cm2Pass (96/96) at < 6.0 mmH2O/cm2, 32 samples from 3 non-consecutive lots3.4, 3.0, 3.0 mmH2O/cm²Pass (96/96) at < 6.0 mmH2O/cm2, 32 samples from 3 non-consecutive lots4.1, 3.4, 3.4 mmH2O/cm2
FlammabilityClass 1Pass (96/96) at Class 1, 32 samples from 3 non-consecutive lotsClass 1Pass (96/96) at Class 1, 32 samples from 3 non-consecutive lotsClass 1
CytotoxicityNon-cytotoxicNon-cytotoxicNon-cytotoxicNon-cytotoxic
IrritationNegligibly irritating (Primary Irritation Index 0.0)Non-irritatingNegligibly irritating (Primary Irritation Index 0.0)Non-irritating
SensitizationNot a contact sensitizerNon-sensitizingNot a contact sensitizerNon-sensitizing

{8}------------------------------------------------

Summary of Clinical Testing Conducted

No clinical testing is included in this submission.

Substantial Equivalence Conclusion

Based on comparison of the subject device to the predicate device and the non-clinical tests performed, it can be concluded that the proposed Procedure Mask with Ear Loops and Surgical Mask with Ties are as safe and effective, and perform as well as the legally marketed predicate device Kenpax International Limited Procedure Mask/Surgical Mask cleared under premarket notification K202899.

K221134 Page 6 of 6

§ 878.4040 Surgical apparel.

(a)
Identification. Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.(b)
Classification. (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.