K Number
K221100

Validate with FDA (Live)

Device Name
Viz RV/LV
Manufacturer
Date Cleared
2022-08-29

(137 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Viz RV/LV Software device is designed to measure the maximal diameters of the right and left ventricles of the heart from a volumetric CTPA acquisition and report the ratio of those measurements. Viz RV/LV analyzes cases using an artificial intelligence algorithm to identify the location and measurements of the ventricles. The Viz RV/LV software provides the user with annotated images showing ventricular measurements. Its results are not intended to be used on a stand-alone basis for clinical decision-making or otherwise preclude clinical assessment of CTPA cases.

Device Description

The Viz RV/LV is a software-only device that uses a locked artificial intelligence machine learning (AI/ML) algorithm to measure the maximal diameters of the right and left ventricles of the heart from a computed tomography pulmonary angiogram (CTPA) and report the ratio of those measurements. Viz RV/LV produces an Annotated Image Series (Figure 1) and an RV/LV Summary Report (Figure 2) in DICOM format. The Annotated Image Series shows an RGB overlay on each slice of the input scan: The red and blue solid lines indicate the maximum ventricular diameter for each ventricle. The dashed line indicates a diameter measured on a slice that is within 10 slices of the global maximum ventricular diameter. The interventricular septum is marked in solid green on all images where diameters are marked. The maximal diameter is presented along with solid lines on slices with global maximum diameter. The RV/LV Summary Report summarizes the results of the ventricle analysis and shows the slices with the maximum right and left ventricular diameters. The lines measuring the maximum RV and LV diameters are displayed over the original CTPA slice image, along with the lengths of the largest RV and LV diameters, and the RV/LV ratio. Viz RV/LV is hosted on Viz.ai's Backend Server and analyzes applicable CTPA scans that are acquired on CT scanners and are forwarded to Viz.ai's Backend Server. The results of the analysis are exported in DICOM format are sent to a PACS destination for review by thoracic radiologists, general radiologists, pulmonologists, cardiologists, or other similar physicians to assist in the assessment of right ventricle enlargement.

AI/ML Overview

Here's an analysis of the acceptance criteria and the study proving the device meets them, based on the provided text.

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are not explicitly stated as distinct numerical targets in the same way that a "performance goal" for MAE is mentioned. However, based on the wording, we can infer the primary acceptance criterion for the algorithm's accuracy from the clinical performance section.

Metric / CriterionAcceptance Criteria (Inferred)Reported Device Performance
Mean Absolute Error (MAE)MAE between the algorithm's measurement and established ground truth less than 7.2 mm (performance goal).The study demonstrated that the MAE was less than 7.2 mm between the established ground truth.
Agreement (General)High degree of agreement between algorithm measurements and manually obtained measurements."The algorithm's ventricle diameter measurements were aligned when compared against the measurements that were obtained manually." and "There was a high degree of agreement between the different trained radiologist as demonstrated by statical analysis."
Clinical PerformanceDemonstrate safety and effectiveness comparable to the predicate device."Clinical performance data demonstrated that the device is as safe and effective as the previously cleared Imbio RV/LV software (K203256)."
StrandingNo overlapping data between training and pivotal study."There was no overlapping data between the training sets and the pivotal study in terms of time and patient images."

2. Sample Size Used for the Test Set and Data Provenance

  • Test Set Sample Size: Not explicitly stated as a number of cases, but the study implies a sufficient sample for statistical analysis. It mentions "The 4 clinical sites used in the pivotal study" are a subset of 13 larger sites.
  • Data Provenance:
    • Country of Origin: Not explicitly stated, but the submission is to the U.S. FDA, implying compliance with U.S. regulatory standards. Clinical sites are mentioned, suggesting real-world data collection.
    • Retrospective or Prospective: Not explicitly stated, but the mention of "4 clinical sites used in the pivotal study were a subset of a larger 13 sites used as part of the training data set" and "no overlapping data between the training sets and the pivotal study in terms of time and patient images" suggests that the test set data was collected independently from the training data, likely retrospectively for the purpose of this analysis, but drawn from existing clinical acquisitions.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • Number of Experts: Not explicitly stated as a specific number, but referred to as "trained radiologists" (plural), indicating more than one.
  • Qualifications of Experts: Described as "trained radiologists." Specific experience (e.g., "10 years of experience") is not provided.

4. Adjudication Method for the Test Set

  • Adjudication Method: Not explicitly detailed. The text states, "There was a high degree of agreement between the different trained radiologist as demonstrated by statical analysis," suggesting that the ground truth was established through some form of consensus among these multiple radiologists. This could imply a majority vote, averaging, or a formal consensus meeting, but the specific method (e.g., 2+1, 3+1) is not provided.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • A MRMC comparative effectiveness study was not specifically described in the provided text. The study primarily focused on validating the standalone performance of the Viz RV/LV algorithm against a human-established ground truth. The role of the device is to "provide the user with annotated images showing ventricular measurements" and is "not intended to be used on a stand-alone basis for clinical decision-making or otherwise preclude clinical assessment of CTPA cases," suggesting it is a human-in-the-loop aid, but the study did not quantify the improvement of human readers with or without the AI assistance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

  • Yes, a standalone performance study was done. The text explicitly states, "Clinical testing was performed as a study comparing the Viz RV/LV's output to the ground truth as established by trained radiologists." This means the algorithm's raw output was directly compared to the expert ground truth, without human intervention or modification of the algorithm's output.

7. The Type of Ground Truth Used

  • Type of Ground Truth: Expert Consensus / Manual Measurements. The ground truth was "established by trained radiologists" and involved "measurements that were obtained manually."

8. The Sample Size for the Training Set

  • Training Set Sample Size: Not explicitly stated as a number of cases, but it mentions that "a larger 13 sites used as part of the training data set." The absolute number of cases from these 13 sites is not given.

9. How the Ground Truth for the Training Set was Established

  • The text does not explicitly detail how the ground truth for the training set was established. It only mentions that the "4 clinical sites used in the pivotal study were a subset of a larger 13 sites used as part of the training data set," implying that the training data also came from clinical sources. One can infer that it likely involved similar methods of expert annotation or manual measurement, given the nature of the task and the validation approach. However, specific details are absent.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

Viz.ai, Inc. % Vi Ma Regulatory Affair Specialist 201 Mission Street, 12th Floor SAN FRANCISCO CA 94105

Re: K221100

August 29, 2022

Trade/Device Name: Viz RV/LV Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: QIH Dated: August 4, 2022 Received: August 5, 2022

Dear Vi Ma:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

{1}------------------------------------------------

https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jessica Lamb, Ph.D. Assistant Director Imaging Software Team DHT8B: Division of Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below

510(k) Number (if known)

K221100

Device Name

Viz RV/LV

Indications for Use (Describe)

The Viz RV/LV Software device is designed to measure the maximal diameters of the right and left ventricles of the heart from a volumetric CTPA acquisition and report the ratio of those measurements. Viz RV/LV analyzes cases using an artificial intelligence algorithm to identify the location and measurements of the ventricles. The Viz RV/LV software provides the user with annotated images showing ventricular measurements. Its results are not intended to be used on a stand-alone basis for clinical decision-making or otherwise preclude clinical assessment of CTPA cases.

Type of Use (Select one or both, as applicable)

図 Prescription Use (Part 21 CFR 801 Subpart D) Subpart C)

□ Over-The-Counter Use (21 CFR 801

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY-K221100

Viz.ai, Inc.'s Viz RV/LV

Applicant Name:Viz.ai, Inc.
201 Mission Street, 12th Floor
San Francisco, CA 94105
  • Contact Person: Vi Ma Regulatory Affairs Specialist 201 Mission Street, 12th Floor San Francisco, CA 94105 Tel. (415) 663-6130 vi.ma@viz.ai
  • Date Prepared: August 24, 2022

Device Name and Classification

Name of Device:Viz RV/LV
Common or Usual Name:Automated Radiological Image Processing Software
Classification Panel:Radiology
Regulation No:21 C.F.R. § 892.2050
Regulatory Class:Class II
Product Code:QIH

Predicate Device

ManufacturerDevice NameApplication No.
ImbioImbio RV/LV SoftwareK203256

Device Description

The Viz RV/LV is a software-only device that uses a locked artificial intelligence machine learning (AI/ML) algorithm to measure the maximal diameters of the right and left ventricles of the heart from a computed tomography pulmonary angiogram (CTPA) and report the ratio of those measurements.

Viz RV/LV produces an Annotated Image Series (Figure 1) and an RV/LV Summary Report (Figure 2) in DICOM format. The Annotated Image Series shows an RGB overlay on each slice of the input scan:

  • . The red and blue solid lines indicate the maximum ventricular diameter for each ventricle.
  • . The dashed line indicates a diameter measured on a slice that is within 10 slices of the global maximum ventricular diameter.

{4}------------------------------------------------

  • . The interventricular septum is marked in solid green on all images where diameters are marked.
  • . The maximal diameter is presented along with solid lines on slices with global maximum diameter.

Image /page/4/Picture/2 description: The image shows a CT scan of a human chest. The scan shows the heart, lungs, and other organs in the chest. The scan also shows the measurements of the right ventricle (RV) and left ventricle (LV) of the heart. The RV measures 58.3mm, and the LV measures 36.5mm.

Figure 1: Annotated Image Series

Image /page/4/Picture/4 description: The image shows a Viz RV/LV Analysis Summary. The RV/LV Diameter ratio is 1.6, the LV is 36.5mm (slice 218), and the RV is 58.3mm (slice 212). There are two CT scan images, one showing the Maximal RV Diameter Slice and the other showing the Maximal LV Diameter Slice. The image also includes a legend indicating the Largest RV diameter, Detected Interventricular Septum, and Largest LV diameter.

Figure 2: RV/LV Summary Report

The RV/LV Summary Report summarizes the results of the ventricle analysis and shows the slices with the maximum right and left ventricular diameters. The lines measuring the maximum RV and LV

{5}------------------------------------------------

diameters are displayed over the original CTPA slice image, along with the lengths of the largest RV and LV diameters, and the RV/LV ratio.

Viz RV/LV is hosted on Viz.ai's Backend Server and analyzes applicable CTPA scans that are acquired on CT scanners and are forwarded to Viz.ai's Backend Server. The results of the analysis are exported in DICOM format are sent to a PACS destination for review by thoracic radiologists, general radiologists, pulmonologists, cardiologists, or other similar physicians to assist in the assessment of right ventricle enlargement. The basic principle of operations is outlined in Figure 3 below.

Image /page/5/Figure/2 description: The image shows a diagram of a medical imaging workflow. It starts with a CT scanner, which sends data to a Viz.ai backend server. The server uses an RV/LV measurement algorithm to analyze the data and generate a report, which is then sent to a PACS server. Finally, the report is displayed on a monitor.

Figure 3. Data and Event Workflow

Intended Use / Indications for Use

Viz.ai RV/LV software and the predicate device, Imbio RV/LV software (K203256), are Medical Image Management and Processing System software intended to calculate of the ratio of right ventricular diameter to left ventricular diameter from contrast enhanced CT images of the chest acquired using a standard CT pulmonary angiogram acquisition.

Viz RV/LV has the following indications for use:

The Viz RV/LV Software device is designed to measure the maximal diameters of the right and left ventricles of the heart from a volumetric CTPA acquisition and report the ratio of those measurements. Viz RV/LV analyzes cases using an artificial intelligence algorithm to identify the location and measurements of the ventricles. The Viz RV/LV software provides the user with annotated images showing ventricular measurements. Its results are not intended to be used on a stand-alone basis for clinical decision-making or otherwise preclude clinical assessment of CTPA cases.

The predicate device, Imbio RV/LV Software (K203256), has the following indications use:

The Imbio RV/LV Software device is designed to measure the maximal diameters of the right and left ventricles of the heart from a volumetric CTPA acquisition and report the ratio of those measurements. RV/LV analyzes cases using an artificial intelligence algorithm to identify the location and measurements of the ventricles. The RV/LV

{6}------------------------------------------------

software provides the user with annotated images showing ventricular measurements. Its results are not intended to be used on a stand-alone basis for clinical decisionmaking or otherwise preclude clinical assessment of CTPA cases.

In comparing the intended use and indications for use of Viz RV/LV and the predicate device, Viz RV/LV has the same intended use as its predicate device.

A substantial equivalence chart (Table 1) comparing the subject device, Viz RV/LV, and the predicate, the Imbio RV/LV software, is provided below:

Table 1 Technological Characteristic Comparison
Predicate DeviceSubject Device
Imbio RV/LV SoftwareViz RV/LV
Device Class22
Product CodeQIHQIH
RegulationNumber:21 C.F.R. § 892.205021 C.F.R. § 892.2050
Indications forUse:The Imbio RV/LV Software device isdesigned to measure the maximaldiameters of the right and left ventricles ofthe heart from a volumetric CTPAacquisition and report the ratio of thosemeasurements. RV/LV analyzes casesusing an artificial intelligence algorithm toidentify the location and measurements ofthe ventricles. The RV/LV softwareprovides the user with annotated imagesshowing ventricular measurements. Itsresults are not intended to be used on astand-alone basis for clinical decision-making or otherwise preclude clinicalassessment of CTPA cases.The Viz RV/LV Software device isdesigned to measure the maximaldiameters of the right and left ventricles ofthe heart from a volumetric CTPAacquisition and report the ratio of thosemeasurements. Viz RV/LV analyzes casesusing an artificial intelligence algorithm toidentify the location and measurements ofthe ventricles. The Viz RV/LV softwareprovides the user with annotated imagesshowing ventricular measurements. Itsresults are not intended to be used on astand-alone basis for clinical decision-making or otherwise preclude clinicalassessment of CTPA cases.
Input DataRequirements:Non-gated, CT Pulmonary Angiography(CTPA)Same
DICOMCompliantYesSame
Left VentricleSegmentationYesSame
Right VentricleSegmentationYesSame
DiameterMeasurementsYes - AutomatedSame
Fully automatedsegmentationYesSame
OutputsReports, DICOM Secondary CaptureSeriesSame

Table 1 Technological Characteristic Comparison

{7}------------------------------------------------

Performance Data

Clinical testing was performed as a study comparing the Viz RV/LV's output to the ground truth as established by trained radiologists. There was a high degree of agreement between the different trained radiologist as demonstrated by statical analysis. The study demonstrated that the MAE (Mean absolute error) was less than 7.2 mm between the established ground truth, which was aligned with the performance goal.

Additional statistical analysis was performed to describe the degree of agreement between the ventricle diameters measured by the RV/LV algorithm in comparison to the measurements that were obtained manually. The algorithm's ventricle diameter measurements were aligned when compared against the measurements that were obtained manually.

Stratification of device performance was divided by clinical site, gender, age, slice thickness, scanner manufacture and model.

The 4 clinical sites used in the pivotal study were a subset of a larger 13 sites used as part of the training data set. There was no overlapping data between the training sets and the pivotal study in terms of time and patient images.

Conclusions

A comparison of intended use and technological characteristics, along with clinical performance data, demonstrated that the device is as safe and effective as the previously cleared Imbio RV/LV software (K203256). Thus, Viz RV/LV is substantially equivalent to Imbio RV/LV.

N/A