K Number
K221069

Validate with FDA (Live)

Date Cleared
2022-11-21

(223 days)

Product Code
Regulation Number
880.5440
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Infusion Sets for Single Use with Needles is intended to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into the vein.

Device Description

The Infusion Sets for Single Use with Needles is intended to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into the vein. It consists of basic configuration and optional components. The basic configuration consists of protective cap of spike, spike, drip chamber, flow regulator, flexible tube, conical fitting, and injection needle/intravenous needle. Air vent, fluid filter, clamp, and injection site are optional components. The device is provided sterile and single use.

AI/ML Overview

The provided document is a 510(k) summary for a medical device (Infusion Sets for Single Use with Needles) and does not describe a study involving an AI/Machine Learning device that utilizes a test set, expert readers, or ground truth in the way a diagnostic AI would.

Instead, this document focuses on demonstrating substantial equivalence to a previously cleared predicate device by comparing technical characteristics and performance standards. The "acceptance criteria" and "study" described here refer to non-clinical performance testing to ensure the device meets specified engineering and safety standards, rather than evaluating diagnostic accuracy of an AI model.

Therefore, many of the requested categories for AI/ML device studies (such as sample size for test set, number of experts for ground truth, MRMC studies, effect size of AI assistance, standalone performance, and training set details) are not applicable to this 510(k) submission.

Here's an interpretation of the document's content in a way that aligns with your request as much as possible, focusing on the quality and performance testing described:


Device: Infusion Sets for Single Use with Needles (K221069)

Type of Study: Non-Clinical Performance Testing and Biocompatibility Testing for Substantial Equivalence Determination. This is not an AI/ML diagnostic performance study.

1. Table of Acceptance Criteria and Reported Device Performance

For medical devices like infusion sets, acceptance criteria are typically defined by recognized international and national standards. The "performance" is demonstrated by showing compliance with these standards.

Acceptance Criteria Category/StandardReported Device Performance/Compliance
Biocompatibility
ISO 10993-5:2009 (Cytotoxicity)No Cytotoxicity
ISO 10993-10:2010 (Irritation and Skin Sensitization)No Intracutaneous Reactivity, No Skin Sensitization
ISO 10993-11:2017 (Systemic Toxicity)No Acute Systemic Toxicity, No Subacute Systemic Toxicity
ISO 10993-4:2017 & ASTM F756-17 (Interactions with Blood/Hemolysis)No Hemolysis, No different from control device (for Complement Activation), No thrombosis (for In vivo vein thromboresistance)
Sterilization
ISO 10993-7:2008 (Ethylene Oxide Sterilization Residuals)Complies (EO sterilized, SAL 10^-6)
ASTM F88/F88M-15, F1929-15, F1886/F1886M-16 (Seal Integrity)Complies (Tested for Sterility, Dye Penetration, Visual Inspection)
USP <151> & <85> (Pyrogen & Bacterial Endotoxins)No Pyrogen, Endotoxin Limit: 20 EU per device
Physical/Functional Performance
ISO 7864:2016 (Sterile hypodermic needles)Complies (Needle performance meets requirements despite different gauges/types than predicate)
ISO 9626:2016 (Stainless Steel Needle Tubing)Complies (Needle performance meets requirements)
ISO 80369-7:2016 (Small-bore connectors - intravascular/hypodermic)Complies
ISO 80369-20:2015 (Small-bore connectors - Common test methods)Complies
ISO 8536-4:2019 (Infusion sets for single use, gravity feed)Complies (Performance tests conducted, results meet requirements despite different capacity/length than predicate; "Infusion Set Performance" in table)
ISO 8536-14:2016 (Clamps and flow regulators)Complies (Assumed, as part of general infusion set compliance)
USP <788> (Particular Matter in Injections)Complies

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: The document does not specify exact sample sizes for each non-clinical test (e.g., number of infusion sets tested for flow rate, sterility, etc.). It generally states that "Non clinical tests were conducted to verify that the proposed device met all design specifications." Regulatory standards often dictate minimum sample sizes for such tests, but these are not enumerated in the summary.
  • Data Provenance: The testing was conducted by Sichuan Prius Biotechnology Co., Ltd. (China). The data represents laboratory testing results performed on the manufactured devices. The studies are by nature prospective, as they are specifically conducted to support the 510(k) submission for the new device.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • This concept is not applicable. For non-clinical performance and biocompatibility testing, "ground truth" is established by adherence to validated international standards and test methods (e.g., ISO, ASTM, USP). The expertise lies in the qualifications of the testing laboratories and personnel conducting these specific, standardized tests (e.g., chemists, material scientists, engineers, microbiologists). The document does not specify how many or the qualifications of the testing personnel, as it's assumed to be part of standard quality system requirements for medical device manufacturing and testing.

4. Adjudication Method for the Test Set

  • This concept is not applicable. Performance testing results are either compliant with the specified standard and acceptance criteria or they are not. There is no "adjudication" in the sense of reconciling differing expert opinions or interpretations, as would be the case in subjective diagnostic AI assessments.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not applicable. This is not an AI/ML device, nor is it a diagnostic imaging device that involves human reader interpretation. No MRMC study was performed.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not applicable. This is not an algorithm or AI device. The device itself (infusion set) functions independently of human interpretation concerning its core performance metrics (e.g., flow rate, sterility, tensile strength).

7. The type of ground truth used

  • Ground Truth: For this type of device, "ground truth" is defined by established international and national standards, regulatory requirements, and validated test methods. Compliance with these standards (e.g., ISO for biological evaluation, physical properties; ASTM for material properties, package integrity; USP for sterility, pyrogens) indicates that the device performs as intended and is safe. The document repeatedly states, "Comply with ISO X," "meets the requirements of ISO Y," or "test result can meet the requirements."

8. The sample size for the training set

  • Not applicable. There is no concept of a "training set" for this type of medical device which undergoes compliance testing rather than AI model training.

9. How the ground truth for the training set was established

  • Not applicable, as there is no training set.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 21, 2022

Sichuan Prius Biotechnology Co., Ltd. % Diana Hong General Manager Mid-Link Consulting Co., Ltd P.O.box 120-119 Shanghai, 200120 China

Re: K221069

Trade/Device Name: Infusion Sets for Single Use with Needles Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular administration set Regulatory Class: Class II Product Code: FPA Dated: October 21, 2022 Received: October 21, 2022

Dear Diana Hong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

David Walloschek

David Wolloscheck, Ph.D. For Joyce M. Whang, Ph.D. Acting Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K221069

Device Name Infusion Sets for Single Use with Needles

Indications for Use (Describe)

The Infusion Sets for Single Use with Needles is intended to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into the vein.

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K220169 - 510(k) Summary

This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section 807.92.

The assigned 510(k) Number: K220169

  • Date of Preparation: 11/21/2022 1.
    1. Sponsor Identification

Sichuan Prius Biotechnology Co., Ltd.

No.2 Prius Road, Luo Long Industrial Park Nanxi District, 644104 Yibin City, Sichuan Province, PEOPLE'S REPUBLIC OF CHINA

Establishment Registration Number: Not registered yet

Contact Person: Yan Liu Position: Management Representative Tel: +86-831-3839889 Fax: +86-831-3839887 Email: 48363603@qq.com

    1. Designated Submission Correspondent
      Ms. Diana Hong (Primary Contact Person) Ms. Tingting Su (Alternative Contact Person)

Mid-Link Consulting Co., Ltd

P.O. Box 120-119, Shanghai, 200120, China

Tel: +86-21-22815850, Fax: 360-925-3199 Email: info@mid-link.net

{4}------------------------------------------------

Identification of Proposed Device 4.

Trade Name: Infusion Sets for Single Use with Needles Common Name: Intravascular Administration Set

Regulatory Information Classification Name: Intravascular Administration Set Classification: II Product Code: FPA Regulation Number: 21 CFR 880.5440 Review Panel: General Hospital;

Indications for Use Statement:

The Infusion Sets for Single Use with Needles is intended to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into the vein.

Device Description

The Infusion Sets for Single Use with Needles is intended to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into the vein. It consists of basic configuration and optional components. The basic configuration consists of protective cap of spike, spike, drip chamber, flow regulator, flexible tube, conical fitting, and injection needle/intravenous needle. Air vent, fluid filter, clamp, and injection site are optional components. The device is provided sterile and single use.

  • న్. Identification of Predicate Device
    510(k) Number: K163160 Product Name: Sterile Single-use Infusion Set

    1. Substantially Equivalent (SE) Comparison
ITEMProposed DevicePredicate DeviceK163160Remark
ProductInfusion Sets for Single Use withNeedlesSterile Single-use Infusion Set/
Product CodeFPAFPASame
Regulation Number21 CFR 880.544021 CFR 880.5440Same
ClassClass IIClass IISame
Indications for UseThe Infusion Sets for Single Use withNeedles is intended to administerfluids from a container to a patient'svascular system through a needle orcatheter inserted into the vein.The Sterile Single-use Infusion Setis intended to administer fluidsfrom a container to a patient'svascular system through a needleor catheter inserted into the vein.Same
ConfigurationBasicConfigurationProtective cap ofspikeProtector Cap of SpikeSame
SpikeSpikeSame
Drip chamberDrip chamberSame
Flow RegulatorFlow RegulatorSame
Flexible TubeFlexible TubeSame
Conical fitting (luerlock or luer slip)Conical fitting (luer lock or luerslip)Same
Intravenous Needleorhypodermic needlesNeedleDifferent
OptionalconfigurationInjection siteInjection SiteSame
Fluid filterFluid filterSame
Air ventAir ventSame
ClampClampSame
Operation ModeFor manual use onlyFor manual use onlySame
Filter Characteristics15μm15μmSame
Capacity (ml)18ml19, 21mlDifferent
Needle GaugeIntravenous Needle: 18G, 19G, 20G,21G, 22G, 23G, 24G, 25G, 26G, 27G21GDifferent
Hypodermic Needles: 16G, 18G,
19G, 20G, 21G, 22G, 23G, 24G, 25G,26G, 27G, 28G, 29G, 30G
Tubing Diameter3.9mm3.9mmSame
Length (mm)1635mm, 1665mm1610, 1900mmDifferent
Infusion Set PerformanceComply with ISO 8536-4Comply with ISO 8536-4Same
Needle PerformanceComply with ISO 7864 and ISO 9626Comply with ISO 7864 and ISO9626Same
Patient contact Material
SpikeAcrylonitrile Butadiene Styrene(ABS)High-density Polyethylene(HDPE)Different
Drip ChamberPoly Vinyl Chloride (PVC)Acrylonitrile Butadiene Styrene(ABS)Different
Flexible TubePoly Vinyl Chloride (PVC)Poly Vinyl Chloride (PVC)Same
conical fittingPoly Vinyl Chloride (PVC)Poly Vinyl Chloride (PVC)Same
Injection sitePoly Vinyl Chloride (PVC)Acrylonitrile Butadiene Styrene (ABS)Different
Fluid filterPolyamideAcrylonitrile Butadiene Styrene (ABS)Acrylonitrile Butadiene Styrene (ABS)Different
IntravenousNeedleFlexible TubePoly Vinyl Chloride (PVC)/Different
Needle HandlePoly Vinyl Chloride (PVC)/Different
NeedleStainless Steel/Different
Conical FittingPoly Vinyl Chloride (PVC)/Same
HypodermicNeedleNeedleStainless SteelStainless SteelSame
Needle HubPolypropylene (PP)Polypropylene (PP)Same
Biocompatibility
CytotoxicityNo CytotoxicityNo CytotoxicitySimilar
Intracutaneous ReactivityNo Intracutaneous ReactivityNo Intracutaneous Reactivity
Skin SensitizationNo Skin SensitizationNo Skin Sensitization
Acute Systemic ToxicityNo Acute Systemic ToxicityNo Acute Systemic Toxicity
HemolysisNo HemolysisNo Hemolysis
PyrogenNo PyrogenNo Pyrogen
Subacute ToxicityNo Subacute Systemic Toxicity/
Complement ActivationNo different from control device/
In vivo vein thromboresistanceNo thrombosis/
Sterilization
MethodEO sterilizedEO sterilizedSame
SAL10-610-6Same
Endotoxin Limit20 EU per deviceSame

Table 1 Comparison of Technology Characteristics

{5}------------------------------------------------

{6}------------------------------------------------

Different-Configuration

The configuration for the proposed device includes basic configuration and optional configuration. The optional configuration for the proposed device can be covered by the predicate device. The only difference between the proposed device and predicate device is the needle. Both the proposed device and predicate device are a single use infusion set with a pre-attached needle. However, the pre-attached needle for the proposed device is different from predicate device. The pre-attached needle for the proposed device is available in two types, which are hypodermic needle and intravenous needle. These two type needles can be provided with/without a safe feature. While the pre-attached needle for the predicate device is a hypodermic needle. Although the pre-attached needle

{7}------------------------------------------------

for the proposed device cannot be covered by the pre-attached needle for the predicate device. The needle performance has been tested and results demonstrate that the needle meets the requirements of ISO 7864 and ISO 9626. Therefore, the differences do not raise new questions of safety and effectiveness for the proposed device.

Different-Capacity

The total capacity for the proposed device is different from predicate device. However, the performance test has been conducted on the proposed device and the test result can meet the requirements of ISO 8536-4. Therefore, this difference does not raise new questions of safety and effectiveness for the proposed device.

Different - Needle Gauge

The subject device is available in more gauges compared to the predicate device. However, the needle performance has been tested and results demonstrate that the needs the requirements of ISO 7864 and ISO 9626. Therefore, this difference does not raise new questions of safety and effectiveness for the proposed device.

Different-Length

The total length for the proposed device is different from predicate device. However, the performance test has been conducted on the proposed device and the test result meet the requirements of ISO 8536-4. Therefore, the differences do not raise new questions of safety and effectiveness for the proposed device.

Different -Patient Contact Material

The patient contact materials for proposed device are different from predicate device. However, the biocompatibility testing per ISO 10993-1 was performed and the results show there are no adverse effect. Therefore, the differences do not raise new questions of safety and effectiveness for the proposed device.

Similar- Biocompatibility

The biocompatibility test was performed on the proposed device and three additional endpoints were evaluated compared to the predicate device, which are subacute toxicity, in vivo Thromboresistance and Complement Activation. The test results for these endpoints show there are no adverse effect on the material. Therefore, the provided biocompatibility testing supports substantial equivalence to the predicate device.

7. Non-Clinical Test Conclusion

Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:

{8}------------------------------------------------

  • ISO 10993-5:2009 Biological evaluation of medical Devices-Part 5: Tests for in Vitro Cytotoxicity

  • ISO 10993-10:2010 Biological evaluation of medical devices- Part 10: Test for irritation and skin sensitization

  • ISO 10993-11:2017 Biological evaluation of medical devices- Part 11: Tests for systemic toxicity

  • ISO 10993-4:2017 Biological Evaluation of Medical Devices--Part 4: Selection of Tests for Interactions with Blood

  • ASTM F756-17 Standard Practice for Assessment of Hemolytic Properties of Materials

  • ISO 10993-7:2008 Biological Evaluation of Medical Devices-Part 7:Ethylene Oxide Sterilization Residuals

  • ASTM F88/F88M-15, Standard Test Method for Seal Strength of Flexible Barrier Materials. (Sterility)

  • ASTM F1929-15 Standard Test Method for Detecting Seal Leaks in Porous Medical Package by Dye Penetration

  • ASTM F1886/F1886M-16 Standard Test Method for Determining Integrity of Seals for Flexible Packaging by Visual Inspection

  • ISO 7864:2016 Sterile hypodermic needles for single use - Requirements and test methods

  • ISO 9626:2016, Stainless Steel Needle Tubing For The Manufacture of Medical Devices

  • ISO 80369-7:2016 Small-bore connectors for liquids and gases in healthcare applications-Part 7: Connectors for intravascular or hypodermic applications

  • ISO 80369-20:2015 Small-bore connectors for liquids and gases in healthcare applications-Part 20: Common test methods.

  • ISO 8536-4:2019 Infusion equipment for medical use- Part 4: Infusion sets for single use, gravity feed.

  • ISO 8536-14:2016 Clamps and flow regulators for transfusion equipment without fluid contact

  • USP <151>Pyrogen Test

  • USP <85> Bacterial Endotoxins Test

  • USP <788> Particular Matter in Injections.

    1. Clinical Test Conclusion

No clinical study is included in this submission.

  • Substantially Equivalent (SE) Conclusion 9.
    Based on the comparison and analysis above, the proposed device is determined to be Substantially Equivalent (SE) to the predicate device with respect to the indications for use, target populations, treatment method, and technological characteristics.

§ 880.5440 Intravascular administration set.

(a)
Identification. An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.(b)
Classification. Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.