K Number
K221061

Validate with FDA (Live)

Device Name
21HQ613D
Date Cleared
2022-06-03

(53 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

This Medical Monitor is indicated for use in displaying radiological images (including full-field digital mammography and digital breast tomosynthesis) for review, and diagnosis by trained medical practitioners.

Device Description

The Medical monitor is intended to provide high resolution color and grayscale medical imaging for PACS and Radiology system. This Medical Monitor is intended to be used by trained medical practitioners for displaying, reviewing, and analysis of medical images

AI/ML Overview

The provided document is a 510(k) summary for the LG Electronics Inc. device, 21HQ613D, which is a medical monitor. This document states that no clinical studies were considered necessary and performed to demonstrate substantial equivalence.

Therefore, the requested information regarding acceptance criteria and studies proving the device meets those criteria, specifically concerning performance metrics that would typically come from clinical trials or standalone algorithm performance studies, cannot be extracted from this document in the manner requested (e.g., sample size for test set, number of experts for ground truth, adjudication method, MRMC studies, standalone performance, specific ground truth types for training/test sets, training set sample size).

The document instead focuses on demonstrating substantial equivalence to predicate devices through non-clinical technical performance testing against established standards and guidance for display devices, and through comparison of technological characteristics.

Here is what can be extracted regarding the non-clinical acceptance criteria and reported performance:

1. Table of Acceptance Criteria and Reported Device Performance (Non-Clinical)

The document refers to technical performance items tested according to the "Display Devices for Diagnostic Radiology – Guidance for Industry and Food and Drug Administration Staff, issued on October 2, 2017." While specific numerical acceptance criteria are not explicitly stated within this document for each measurement (e.g., "MTF must be X" or "pixel defects must be less than Y"), the "SE Note" for each item indicates that the device's performance was considered "Equivalent" or "Same" to predicate devices, implying it met the expected performance targets set by the guidance for substantial equivalence.

MeasurementsDescriptionReported Device Performance (SE Note)
Spatial resolutionMeasurements of the transfer of information from the image data to the luminance fields at different spatial frequencies of interest typically done by reporting the modulation transfer function.Equivalent
Pixel defectsMeasurements (count, types (e.g., sub-pixel or entire pixel, always-on, always-off), and locations (map) of pixel defects.Equivalent
ArtifactsEvaluate for image artifacts such as ghosting and/or image sticking from displaying a fixed test pattern for a period of time.Same
Temporal responseMeasurements of the temporal behavior of the display in responding to changes in image values from frame to frame.Equivalent
LuminanceMeasurements of the maximum and minimum luminance that the device outputs as used in the application under recommended conditions and the achievable values if the device is set to expand the range to the limit.Same
Conformance to a grayscale-to-luminance functionMeasurements of the mapping between image values and the luminance output following a target model response for 256 or more levels.Equivalent
Luminance at 30° and 45° in diagonal, horizontal, and verticalMeasurements of the luminance response at off-normal viewing related to the target model for the luminance response.Equivalent
Luminance uniformity or Mura testMeasurements of the uniformity of the luminance across the display screen.Equivalent
Stability of luminance and chromaticity responseMeasurements of the change in luminance and chromaticity response with temperature and use time.Same
Spatial noiseMeasurements of the spatial noise level as represented by the noise power spectrum using an appropriate ratio of camera and display pixels.Equivalent
Reflection coefficientMeasurements of the reflection coefficients of the display device.Same
Veiling glare or small-spot contrastMeasurements of the contrast obtained for small targets.Equivalent
Color trackingChromaticity at different luminance levels of primary colors as indicated by the color coordinates.Equivalent
Gray trackingChromaticity at different luminance levels of gray shades, including the white point.Equivalent

2. Sample size used for the test set and the data provenance: Not applicable, as no clinical studies with test sets of medical images were performed. The testing was instrumental, evaluating the display's physical performance characteristics as a device.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is a medical monitor, not an AI-assisted diagnostic tool.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This device is a medical monitor, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable for clinical ground truth. The "ground truth" for the non-clinical tests would be the established engineering specifications and performance standards for display monitors used in diagnostic radiology.

8. The sample size for the training set: Not applicable, as this is a hardware device (monitor), not an AI algorithm requiring a training set.

9. How the ground truth for the training set was established: Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" stacked on top of each other.

LG Electronics Inc. % Jonghyun Kim Chief Consultant GMS Consulting 4th Floor, Digital Cube, 34, Sangamsan-ro Seoul, Mapo-gu 03909 KOREA

Re: K221061

June 3, 2022

Trade/Device Name: 21HO613D Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: PGY Dated: April 7, 2022 Received: April 11, 2022

Dear Jonghyun Kim:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For

Michael D. O'Hara, Ph.D. Deputy Director DHT 8C: Division of Radiological Imaging and Radiation Therapy Devices OHT 8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K221061

Device Name 21HQ613D

Indications for Use (Describe)

This Medical Monitor is indicated for use in displaying radiological images (including full-field digital mammography and digital breast tomosynthesis) for review, and diagnosis by trained medical practitioners.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

[As Required by 21 CFR 807.92]

1. Date Prepared [21 CFR 807.92(a)(a)]

April 07, 2022

2. Submitter's Information [21 CFR 807.92(a)(1)]

• Name of Sponsor:LG Electronics Inc.
- Address:222, LG-ro, Cheongho-ri, Jinwi-myeon, Pyeongtaek-si,Gyeonggi-do, 17709, Republic of Korea
• Contact Name:Jinhwan Jun / Chief Research Engineer
- Telephone No.:+82-31-8066-5641
- Email Address:jinhwan.jun@lge.com
• Name of Manufacturer:LG Electronics Inc.
- Address:77, Sanho-daero, Gumi-si, Gyeongsangbuk-do, 39381,

3. Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)]

Trade Name:21HQ613D
-----------------------
  • Common Name: ●
    Medical Monitor

Republic of Korea

  • Classification: ●
Classification NameMedical image management and processing system.
Classification Number21 CFR 892.2050
Product CodePGY
Device ClassII
Review PanelRadiology

{4}------------------------------------------------

4. Identification of Predicate Device(s) [21 CFR 807.92(a)(3)]

The identified predicate devices within this submission are shown as follow;

K172738

RX560

EIZO Corporation

Medical image management and processing system.

Primary Predicate Device

510(k) Number:
------------------
  • Applicant:
  • Classification Name: ●
  • Trade Name: ●

Predicate Device

  • K201777 510(k) Number: Applicant: LG Electronics Classification Name: Medical image management and processing system. Trade Name: 31HN713D

5. Description of the Device [21 CFR 807.92(a)(4)]

The Medical monitor is intended to provide high resolution color and grayscale medical imaging for PACS and Radiology system. This Medical Monitor is intended to be used by trained medical practitioners for displaying, reviewing, and analysis of medical images

6. Indications for use [21 CFR 807.92(a)(5)]

This Medical Monitor is indicated for use in displaying radiological images (including full-field digital mammoqraphy and digital breast tomosynthesis) for review, analysis, and diagnosis by trained medical practitioners.

{5}------------------------------------------------

7. Technological Characteristics (Equivalence to Predicate Device) [21 CFR 807.92(a)(6)]

The table below presents comparisons between the subject device (21HQ613D) and the legally marketed predicate devices (K172738 and K201777):

Proposed DevicePrimary Predicate Device
K NumberNot knownK172738
ManufacturerLG Electronics Inc.EIZO Corporation
Model Name21HQ613DRX560
Classification NameMedical image management andprocessing system.Medical image management andprocessing system.
Classification Number21 CFR 892.205021 CFR 892.2050
Indications for UseThis Medical Monitor is indicated for usein displaying radiological images(including full-field digital mammographyand digital breast tomosynthesis) forreview, analysis, and diagnosis by trainedmedical practitioners.This product is intended to be used indisplaying and viewing digital images,including standard and multiframe digitalmammography, for review, analysis anddiagnosis by trained medicalpractitioners. It is specially designed forbreast tomosynthesis applications.
Display TechnologyThe 21HQ613D has a TM213XDGP31module for displaying. TheTM213XDGP31 is a " TFT Liquid CrystalDisplay module with LED Backlight unitand one 30pin eDP port and 4ch-LVDSinterface. This module supports2048x2560 5M Pixel mode and candisplay color driven by 10bit drivers. TheLCD module includes built-in converterfor BacklightThis is a color LCD monitor for viewingdigital images, including standard andmultiframe digital mammography, forreview, analysis and diagnosis. The colorpanel employs in-plane switching (IPS)technology allowing wide viewing anglesand the matrix size (or resolution) is2048x2560 pixels.
Power ConsumptionMAX. 120WOff Mode ≤ 0.3W87 W / Less than 1 W
Screen size473.4 x 364.5 mm337.9 mm x 422.4 mm
LCD ScreenTFT LCDTFT LCD
Pixel Pitch0.165 x 0.165 mm0.165 x 0.165 mm
Resolution2,048 x 2,560 pixels2,048 x 2,560
Horizontal Frequency30 kHz to 135kHz31 kHz to 135 kHz
Vertical Frequency24 Hz to 61 Hz59 Hz to 61 Hz
Input video signalsDVI IN x 1,DP IN x 2DP OUT x 1DVI x 1DP IN x 1DP OUT x 1

[Table 1. Comparison of Proposed Device to Primary Predicate Device]

The comparison table shows that the subject device (21HQ613D) has the similar indications for use the primary predicate device. Although the devices have some different technological characteristics (screen size, pixel pitch, Resolution, horizontal/vertical frequency and input video signals), these differences do not make the subject device less safe and reliable, so the subject device fits for diagnostic use as the predicate device does. There are no significant differences in the technological characteristics of the subject device. All the differences between the subject and predicate device do not raise different questions of safety and effectiveness. It is substantially equivalent to a predicate device in indications for use and technology characteristics.

{6}------------------------------------------------

Proposed DevicePredicate Device
K NumberNot knownK201777
ManufacturerLG Electronics Inc.LG Electronics Inc.
Model Name21HQ613D31HN713D
Classification NameMedical image management and processing system.Medical image management and processing system.
Classification Number21 CFR 892.205021 CFR 892.2050
Indications for UseThis Medical Monitor is indicated for use in displaying radiological images (including full-field digital mammography and digital breast tomosynthesis) for review, analysis, and diagnosis by trained medical practitioners.This Medical Monitor is indicated for use in displaying radiological images (including full-field digital mammography and digital breast tomosynthesis) for review, analysis, and diagnosis by trained medical practitioners.
Display TechnologyThe 21HQ613D has a TM213XDGP31 module for displaying. The TM213XDGP31 is a " TFT Liquid Crystal Display module with LED Backlight unit and one 30pin eDP port and 4ch-LVDS interface. This module supports 2048x2560 5M Pixel mode and can display color driven by 10bit drivers. The LCD module includes built-in converter for BacklightThe 31HN713D has a LB310FTM module for displaying. The LB310FTM is a 31" TFT Liquid Crystal Display module with LED Backlight unit. This module supports 4,200 x 2,800 pixels and can display color driven by 10bit drivers.
Power ConsumptionMAX. 120WOff Mode ≤ 0.3WMAX. 150WSleep Mode ≤ 0.5WOff Mode ≤ 0.3W
Screen size473.4 x 364.5 mm676.9 x 459.7 mm
LCD ScreenTFT LCDTFT LCD
Pixel Pitch0.165 x 0.165 mm0.1554 x 0.1554 mm
Resolution2,048 x 2,560 pixels4,200 x 2,800 pixels
Horizontal Frequency30 kHz to 135kHz30 kHz to 135kHz
Vertical Frequency24 Hz to 61 Hz56 Hz to 61 Hz
Input video signalsDVI IN x 1,DP IN x 2DP OUT x 1DisplayPort x 2HDMI x 1
[Table 2. Comparison of Proposed Device to Predicate Device]
----------------------------------------------------------------------

The comparison table shows that the subject device (21HQ613D) has the similar indications for use the predicate device. Although the devices have some different technological characteristics (screen size, pixel pitch, Resolution, horizontal/vertical frequency and input video signals), these differences do not make the subject device less safe and reliable, so the subject device fits for diagnostic use as the predicate device does. There are no significant differences in the technological characteristics of the subject device. All the differences between the subject and predicate device do not raise different questions of safety and effectiveness. It is substantially equivalent to a predicate device in indications for use and technology characteristics.

{7}------------------------------------------------

8. Non-Clinical Test summary

    1. Electrical Safety and Electromagnetic Compatibility
      The test results demonstrated that the proposed device complies with the following standards:
  • · Electrical Basic Safety and Essential Performance reguirements in accordance with IEC 60601-1:2005/AMD1:2012

  • · Electromagnetic Compatibility Testing in accordance with IEC 60601-1-2 Edition 4.0:2014

  • · Medical Electrical Equipment Part 1-6: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Usability in accordance with IEC 60601-1-6:2010/A1:2013

    1. Software Validation

The 21HQ613D contains MODERATE level of concern software. The software was designed and developed according to a software development process and was verified and validated.

The software information is provided in accordance with FDA guidance: The content of premarket submissions for software contained in medical devices, on May 11, 2005.

    1. Guidance
      Display Devices for Diagnostic Radiology – Guidance for Industry and Food and Drug Administration Staff, issued on October 2, 2017
MeasurementsDescriptionSE Note
a.Spatial resolutionMeasurements of the transfer ofinformation from the image data to theluminance fields at different spatialfrequencies of interest typically done byreporting the modulation transferfunction. Non-isotropic resolutionproperties should be characterizedproperly by providing two-dimensionalmeasurements or measurements alongat least two representative axes.(Using TG18 QC Test Pattern)Equivalent
b.Pixel defectsMeasurements (count, types (e.g., sub-pixel or entire pixel, always-on, always-off), and locations (map) of pixeldefects. This is typically provided as atolerance limit. Pixel defects caninterfere with the visibility of smalldetails in medical images.Equivalent
c.ArtifactsEvaluate for image artifacts such asghosting and/or image sticking fromdisplaying a fixed test pattern for aperiod of time.(Using 5x5 mosaic pattern, 64Gray /127 Gray judgment)Same
d.Temporal responseMeasurements of the temporal behaviorof the display in responding to changesin image values from frame to frame.Since these transitions are typically notsymmetric, rise and fall time constantsEquivalent
MeasurementsDescriptionSE Note
are needed to characterize the system.Slow displays can alter details andcontrast of the image when large imagestacks are browsed or in video,panning, and zooming modes.
e.LuminanceMeasurements of the maximum andminimum luminance that the deviceoutputs as used in the application underrecommended conditions and theachievable values if the device is set toexpand the range to the limit.Same
f.Conformance to agrayscale-to-luminancefunctionMeasurements of the mapping betweenimage values and the luminance outputfollowing a target model response for256 or more levels.Equivalent
g.Luminance at 30° and 45°in diagonal, horizontal,and vertical directions atcenter and four cornersMeasurements of the luminanceresponse at off-normal viewing relatedto the target model for the luminanceresponse.Equivalent
h.Luminance uniformity orMura testMeasurements of the uniformity of theluminance across the display screen.Equivalent
i.Stability of luminance andchromaticity responsewith temperature and timeof operation (on-time)Measurements of the change inluminance and chromaticity responsewith temperature and use time.Same
j.Spatial noiseMeasurements of the spatial noise levelas represented by the noise powerspectrum using an appropriate ratio ofcamera and display pixels. Spatial noiseand resolution affect the way imagesare presented to the viewer and canalter features that are relevant to theinterpretation process of the physicianor radiologist.Equivalent
k.Reflection coefficientMeasurements of the reflectioncoefficients of the display device.Specular and diffuse reflectioncoefficients can be used as surrogatesfor the full bidirectional reflectiondistribution function.Same
l.Veiling glare or small-spotcontrastMeasurements of the contrast obtainedfor small targets.Equivalent
m.Color trackingChromaticity at different luminancelevels of primary colors as indicated bythe color coordinates in an appropriateunits system (e.g., CIE u'v') and thecolor gamut enveloped by the primarycolors.Equivalent
n.Gray trackingChromaticity at different luminancelevels of gray shades, including thewhite point, as indicated by the colorcoordinates in an appropriate unitssystemEquivalent

According to the guidance, we test following performance items:

{8}------------------------------------------------

{9}------------------------------------------------

According to the above test results, there are no significant performance differences between 21HQ613D and the predicate devices that would adversely affect the use of the product. It has substantially equivalent performance compared to the predicate devices.

Clinical Test Summary:

No clinical studies were considered necessary and performed.

9. Conclusion [21 CFR 807.92(b)(3)]

In according with the Federal Food & Drug and cosmetic Act, 21 CFR Part 807, and based on the information provided in this premarket notification LG Electronics, concludes that the 21HQ613D is substantially equivalent in safety and effectiveness to the predicate devices as described herein.

N/A