K Number
K221001

Validate with FDA (Live)

Date Cleared
2022-05-12

(38 days)

Product Code
Regulation Number
878.4810
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

IPL Home Use Hair Removal Device is an over-the-counter device intended for removal of unwanted body and/or facial hair.

Device Description

The IPL Home Use Hair Removal Device is a personal, light-based, hair reduction device intended to be sold over-the-counter directly to the end user. The device provides hair reduction using Intense Pulsed Light (IPL) technology. It works below the skin's surface and does not involve any cutting or pulling, reducing hair growth with minimal pain. The device is only powered by the external power adapter and its IPL emission is by finger switch. The device contains a Xenon lamp and a skin sensor to detect appropriate skin contact. If the device is not properly applied to the treatment area (in full contact with the skin), the device cannot be triggered a pulse emitting.

The IPL Home Use Hair Removal Device includes D-1198, D-1186, D-1189, D-1197 five models. Their intended use, performance, structure design and operation are basically identical, with the differences mainly contains PCB layout, product appearance and internal construction, but these differences do not affect or change the intended use of the device.

AI/ML Overview

This document is a 510(k) Pre-market Notification for an IPL Home Use Hair Removal Device. It focuses on demonstrating substantial equivalence to a predicate device rather than presenting a detailed clinical study with acceptance criteria and specific performance metrics for the subject device's clinical effectiveness.

Therefore, the requested information regarding acceptance criteria, device performance, sample sizes, expert ground truth, adjudication methods, MRMC studies, standalone performance, and training set details for proving clinical effectiveness cannot be fully provided from the given text. This document primarily addresses safety and technical aspects through comparisons and compliance with standards.

Here's a breakdown of what can be extracted and what is missing based on your request:

1. Table of Acceptance Criteria and Reported Device Performance:

  • Acceptance Criteria: No specific numerical acceptance criteria for clinical effectiveness (e.g., percentage hair reduction, safety endpoints with thresholds) are explicitly stated for the subject device in this document. The document refers to "similar safety and effectiveness profile" and "performance data supports that the device is safe and as effective."
  • Reported Device Performance:
    • Clinical Effectiveness: The document states the device "works below the skin's surface and does not involve any cutting or pulling, reducing hair growth with minimal pain." However, no quantitative data on hair reduction percentage, pain scores, or other effectiveness metrics are provided for the subject device. The safety is "similar to the predicate device."
    • Technical Performance (derived from comparison table in Section VI):
      • Wavelength range: 550-1100nm
      • Energy medium: Xenon Arc Flashlamp
      • Energy density: 2.0-4.3 J/cm²
      • Spot size: D-1198: 2.7cm², D-1185: 2.7cm², D-1186: 3.0cm², D-1189: 3.0cm², D-1197: 3.4cm²
      • Pulse duration: 5-12ms
      • Pulsing control: Finger switch
      • Delivery device: Direct illumination to tissue
    • Safety Performance (as per Section VII - Performance Data):
      • Biocompatibility: Materials are identical to the predicate device (K211185), thus deemed safe.
      • Electrical Safety & Eye Safety: Passed IEC 60601-1-2, IEC 60601-1, IEC 60601-1-11, IEC 60601-2-57, and IEC 62471 standards.
      • Software Verification and Validation: Passed, meeting all software requirement specifications and mitigating hazards to acceptable risk levels.

2. Sample size used for the test set and the data provenance:

  • Not provided. This document does not describe a clinical test set with specific sample sizes for evaluating hair reduction or other clinical efficacy outcomes. The safety and effectiveness claims are based on comparison to a predicate device and compliance with technical standards.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not applicable/Not provided. Since no specific clinical test set for efficacy is described, there is no mention of experts establishing ground truth for such a set.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

  • Not applicable/Not provided. No clinical test set involving human judgment and adjudication is described.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Not applicable. This device is an IPL hair removal device, not an AI-powered diagnostic or assistive tool for human readers. Therefore, an MRMC study is not relevant to this submission.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Not applicable in the context of clinical efficacy. While the device itself operates standalone for hair removal, the performance data presented primarily pertains to safety (electrical, eye, biocompatibility) and technical specifications, not a clinical "algorithm only" performance for diagnosis/treatment in the typical sense of AI devices.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

  • Not provided for clinical efficacy. For safety, the "ground truth" is established by compliance with international standards (IEC standards) and prior FDA clearance of similar materials (for biocompatibility).

8. The sample size for the training set:

  • Not applicable/Not provided. There is no mention of a "training set" in the context of an IPL hair removal device's clinical efficacy or safety. The device does contain software, and "Software Verification and Validation" was performed, but details about a "training set" for software development are not included.

9. How the ground truth for the training set was established:

  • Not applicable/Not provided. As with point 8, no training set for clinical efficacy is mentioned.

Summary of what is present:

This 510(k) submission primarily relies on demonstrating substantial equivalence to an existing legally marketed predicate device (Bosidin IPL Home Use Hair Removal Device/D-1171, K211185). The key arguments for equivalence are:

  • Identical Indications for Use: "removal of unwanted body and/or facial hair." (Section V, VI)
  • Similar Mode of Action and Operational Characteristics: Uses Intense Pulsed Light (IPL) technology. (Section VI)
  • Comparable Technical Specifications: Wavelength range, energy medium, energy density, spot size, pulse duration, pulsing control, delivery method are similar or within acceptable ranges. (Section VI table)
  • Safety Compliance: The device components are biocompatible (identical to predicate materials), and it has passed multiple electrical and eye safety standards (IEC 60601-1-2, IEC 60601-1, IEC 60601-1-11, IEC 60601-2-57, IEC 62471). Software has also undergone verification and validation. (Section VII)

The document concludes that "the IPL Home Use Hair Removal Device was found to have a safety and effectiveness profile that is similar to the predicate device" and is "substantially equivalent." This type of submission usually implies that extensive new clinical efficacy trials for the subject device were not deemed necessary by the FDA because of the strong similarity to an already cleared device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" on top and "ADMINISTRATION" on the bottom.

May 12, 2022

Shenzhen Beauty Every Moment intelligent electric Co., Ltd. % Rain Yip Registered Engineer Feiying Drug & Medical Consulting Technical Service Group Contact Address

Re: K221001

Trade/Device Name: IPL Home Use Hair Removal Device Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OHT Dated: March 21, 2022 Received: April 4, 2022

Dear Rain Yip:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Purva Pandya, D.Eng. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K221001

Device Name

IPL Home Use Hair Removal Device, models: D-1198, D-1185, D-1186, D-1189, D-1197

Indications for Use (Describe)

IPL Home Use Hair Removal Device is an over-the-counter device intended for removal of unwanted body and/or facial hair.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary K221001

''510(k) Summary'' as required by 21 CFR Part 807.92.

Date: 2022-03-21

I. Submitter

Shenzhen Beauty Every Moment intelligent electric Co.,Ltd. Rm.101,201,501, Bldg. 4, Haoer Jiashitai Technology Park, Fengtang Avenue No.164, Zhancheng Community, Fuhai Street, Bao'an District, Shenzhen, Guangdong, China Post code: 518100 Tel .: +86 755 3388 3239

LUO HUAJUN Certification engineer Tel: +86755 3388 3239 Email: 309024340@qq.com

II. Device

Name of Device: IPL Home Use Hair Removal Device Model(s): D-1198, D-1185, D-1186, D-1189, D-1197 Common or Usual Name: Light Based Over-The-Counter For Hair Removal Classification Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: OHT Regulation Number: 21 CFR 878.4810

III. Predicate Device

ManufacturerPredicate Device510(k) NumberApproval Date
ShenzhenTechnology Co., Ltd.BosidinIPL Home Use Hair Removal Device/D-1171K211185Dec. 20, 2021

IV. Device Description

The IPL Home Use Hair Removal Device is a personal, light-based, hair reduction device intended to be sold over-the-counter directly to the end user. The device provides hair reduction using Intense Pulsed Light (IPL) technology. It works below the skin's surface and does not involve any cutting or pulling, reducing hair growth with minimal pain. The device is only powered by the external power adapter and its IPL emission is by finger switch. The device contains a Xenon lamp and a skin sensor to detect appropriate skin contact. If the device is not properly applied to the treatment area (in full contact with the skin), the device cannot be triggered a pulse emitting.

{4}------------------------------------------------

The IPL Home Use Hair Removal Device includes D-1198, D-1186, D-1189, D-1197 five models. Their intended use, performance, structure design and operation are basically identical, with the differences mainly contains PCB layout, product appearance and internal construction, but these differences do not affect or change the intended use of the device.

V. Indications for Use

IPL Home Use Hair Removal Device is an over-the-counter device intended for removal of unwanted body and/or facial hair.

VI. Comparison of Technological Characteristics With the Predicate Device

The IPL Home Use Hair Removal Device has the same intended use, mode of action and similar operational characteristics as the predicate devices. Any minor differences between the subject device and the listed predicate devices do no raise any issues of safety or efficacy. Performance data supports that the device is safe and as effective as the predicate device for its intended use. Therefore, the IPL Home Use Hair Removal Device may be found substantially equivalent to its predicate device.

Comparison ElementsSubject Device K221001Predicate Device K211185
K NumberPendingK211185
Trade nameIPL Home Use Hair RemovalDevice/D-1198, D-1185, D-1186,D-1189, D-1197IPL Home Use Hair RemovalDevice/D-1171
Wavelength range550-1100nm530-1100nm
Energy mediumXenon Arc FlashlampXenon Arc Flashlamp
Energy density2.0~4.3J/cm²2.0~4.3J/cm²
D-1198: 2.7cm²D-1185: 2.7cm²Non-removable regular window:3.6cm²
Spot sizeD-1186: 3.0cm²D-1189: 3.0cm²D-1197: 3.4cm²Removable regular window:1.9cm²
Pulse duration5~12ms7.5-12ms
Pulsing controlFinger switchFinger switch
Delivery deviceDirect illumination to tissueDirect illumination tissue
Indicationforuse/Intended useIPL Home Use Hair RemovalDevice is an over-the-counterdevice intended for removal ofunwanted body and/or facial hair.IPL Home Use Hair RemovalDevice is an over-the-counterdevice intended for removal ofunwanted body and/or facial hair.
Location for useOTCOTC

IPL Home Use Hair Removal Device is compared with the following Predicate Devices in terms of intended use, design, material, specifications, and performance:

{5}------------------------------------------------

VII.Performance Data

The following performance data were provided in support of the substantial equivalence determination.

1) Biocompatibility Safety

The materials of the patient-directly contacting components of the subject device is containing PC plastic and Aluminum alloy, and both materials is identical to the material of the device IPL Home Use Hair Removal Device, which is cleared by FDA with K211185, produced by us. So these components' biocompatibility is safe.

2) Electrical Safety and Eve Safetv

Electrical safety and Eye safety testing was performed to, and passed, the following standards:

  • IEC 60601-1-2 Medical electrical equipment -Part 1-2: General requirements for basic safety and essential performance -Collateral standard: electromagnetic compatibility

  • IEC 60601-1 Medical electrical equipment -Part 1: General requirements for basic safety and essential performance

  • IEC 60601-1-11 Medical Electrical Equipment -Part 1: General Requirements for Basic Safety and Essential Performance -Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment

  • IEC 60601-2-57 Medical electrical equipment -Part 2-57: Particular requirements for the basic safety and essential performance of non-laser source equipment intended for therapeutic, diagnostic, monitoring and cosmetic/aesthetic use

3) Eye Safety

  • IEC 62471 Photobiological safety of lamps and lamp systems

4) Software Verification and Validation

Software documentation consistent with moderate level of concern was submitted in this 510(k). System validation testing presented in this 510(k) demonstrated that all software requirement specifications are met and all software hazards have been mitigated to acceptable risk levels. Summary

Based on the above performance as documented in this application, the subject device IPL Home Use Hair Removal Device was found to have a safety and effectiveness profile that is similar to the predicate device.

VIII. Conclusions

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the comparison of intended use, design, materials and performance, the subject device IPL Home Use Hair Removal Device is to be concluded substantial equivalent to its predicate device.

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.