K Number
K220993

Validate with FDA (Live)

Date Cleared
2022-06-23

(80 days)

Product Code
Regulation Number
892.1550
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Ultrasound System SONIMAGE MX1 and its transducers are products designed to collect ultrasonic image data of the human body for diagnostic purposes. The system employs the ultrasonic pulse-echo method to visualize the anatomic structures, characteristics, and dynamics of the human body, and using an image display, Doppler display and/or Doppler sound, offers a procedure applied to the human body for medical diagnosis or examination.

The range of intended clinical applications is same as other conventional ultrasound imaging systems for general purpose, such as small parts, abdomen, musculoskeletal, cardiac, and peripheral vascular.

This device is intended for use in healthcare facilities, such as health clinics and hospitals.

Intended user for the device are physician, sonographer, and other trained qualified healthcare professionals.

Modes of operation include B-mode, PWD-mode, CWD-mode, Color Doppler-mode, Power Doppler-mode, and their Combined mode.

Device Description

The Ultrasound System SONIMAGE MX1 is a portable diagnostic ultrasound system for general purposes. The system provides ultrasound imaging information such as used for the purpose of diagnosing the human body, which visually represents the internal geometry, characteristics and dynamics of the human body, and transmits / receives ultrasound waves to obtain image data of the visual representation.

This system provides ultrasound images in conventional modes of B-mode, M-mode, Color Doppler-mode, Power Doppler-mode, PW Doppler-mode and CW Doppler-mode.

The optional items are available, such as a Cradle, a Three port probe unit, an Additional Battery, and a Foot Switch with dual/triple pedals.

The system can be connected to LAN through the wired Ethernet and, is also capable of wireless LAN with the OTS USB-WiFi adapter supporting security of WPA/WPA2 and WEP.

This system conforms to Real Time Display of Thermal and Mechanical Output Indices on Diagnostic Ultrasound Equipment (Track 3). Transducers have their own characteristic applications and are brought into contact with the body surface.

AI/ML Overview

The provided text is a 510(k) summary for the Konica Minolta Ultrasound System SONIMAGE MX1. It does not contain any information about acceptance criteria or a study proving that the device meets specific performance criteria through a human-in-the-loop or standalone AI study.

Instead, the document focuses on demonstrating substantial equivalence to a previously cleared predicate device (K180084). The "Performance Data" section primarily lists compliance with various medical device standards (e.g., AAMI ANSI ES60601-1, IEC 60601-1-2, IEC 60601-2-37, NEMA UD 2-2004, IEC 62304) and references relevant FDA guidance documents for diagnostic ultrasound systems, software, and cybersecurity.

The key statement regarding performance is: "Risk Analysis and verification and validation activities demonstrate that the established specifications for these devices have been met. The results of risk management did not require clinical studies to demonstrate the substantial equivalency of the proposed device." This indicates that the device's performance was evaluated through engineering verification and validation (V&V) activities against technical specifications and established standards, rather than clinical studies involving human readers or AI algorithms with specific performance metrics like sensitivity, specificity, or AUC.

Therefore, I cannot provide the requested information because it is not present in the provided document. The document describes a regulatory submission based on substantial equivalence, relying on standard compliance and V&V activities, not a performance study as typically seen for AI/ML-driven devices requiring clinical validation against defined acceptance criteria.

{0}------------------------------------------------

June 23, 2022

Image /page/0/Picture/1 description: The image contains the logos of the Department of Health and Human Services and the U.S. Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, featuring a stylized human figure. To the right is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

Konica Minolta, Inc. % Jan Maniscalco Director of QA/RA Konica Minolta Healthcare Americas, Inc. 411 Newark-Pompton Turnpike WAYNE NJ 07470

Re: K220993

Trade/Device Name: Ultrasound System SONIMAGE MX1 Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: March 31, 2022 Received: April 4, 2022

Dear Jan Maniscalco:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For

Michael D. O'Hara, Ph.D. Deputy Director DHT 8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K220993

Device Name Ultrasound System SONIMAGE MX1

Indications for Use (Describe)

The Ultrasound System SONIMAGE MX1 and its transducers are products designed to collect ultrasonic image data of the human body for diagnostic purposes. The system employs the ultrasonic pulse-echo method to visualize the anatomic structures, characteristics, and dynamics of the human body, and using an image display, Doppler display and/or Doppler sound, offers a procedure applied to the human body for medical diagnosis or examination.

The range of intended clinical applications is same as other conventional ultrasound imaging systems for general purpose, such as small parts, abdomen, musculoskeletal, cardiac, and peripheral vascular.

This device is intended for use in healthcare facilities, such as health clinics and hospitals.

Intended user for the device are physician, sonographer, and other trained qualified healthcare professionals.

Modes of operation include B-mode, PWD-mode, CWD-mode, Color Doppler-mode, Power Dopplermode, and their Combined mode.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image contains the Konica Minolta logo. The logo features a blue circle with several horizontal white lines across the middle. Below the circle is the text "KONICA MINOLTA" in a bold, sans-serif font.

510(k) Summary

K220993

Company:KONICA MINOLTA, INC.1 Sakura-machi, Hino-shi, Tokyo 191-8511, Japan
Contact:Tsutomu FukuiSenior Manager of Regulatory & QMS DivisionQuality Assurance Operations, Healthcare Business UnitTelephone: +81 42 589 8429Email: tsutomu.fukui1@konicaminolta.com
Date Prepared:June 22, 2022
Device Name:Ultrasound System SONIMAGE MX1
Common Name:Diagnostic Ultrasound System and Transducers
Regulation Name(s):Ultrasonic Pulsed Doppler Imaging System(21 CFR 892.1550)
Ultrasonic Pulsed Echo Imaging System(21 CFR 892.1560)
Diagnostic Ultrasound Transducer(21 CFR 892.1570)
Regulatory Class:Class II
Product Code(s):IYN, IYO, ITX
Predicate Device:K180084 - Ultrasound System SONIMAGE MX1
Reference Device:K182153 - Ultrasound System SONIMAGE HS1

Device Description

The Ultrasound System SONIMAGE MX1 is a portable diagnostic ultrasound system for general purposes. The system provides ultrasound imaging information such as used for the purpose of diagnosing the human body, which visually represents the internal geometry, characteristics and dynamics of the human body, and transmits / receives ultrasound waves to obtain image data of the visual representation.

  • Page 1 of 6 -

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image contains the logo for Konica Minolta. The logo features a blue sphere at the top with several horizontal white lines across it. Below the sphere, the words "KONICA MINOLTA" are written in a bold, sans-serif font.

This system provides ultrasound images in conventional modes of B-mode, M-mode, Color Doppler-mode, Power Doppler-mode, PW Doppler-mode and CW Doppler-mode.

The optional items are available, such as a Cradle, a Three port probe unit, an Additional Battery, and a Foot Switch with dual/triple pedals.

The system can be connected to LAN through the wired Ethernet and, is also capable of wireless LAN with the OTS USB-WiFi adapter supporting security of WPA/WPA2 and WEP.

This system conforms to Real Time Display of Thermal and Mechanical Output Indices on Diagnostic Ultrasound Equipment (Track 3). Transducers have their own characteristic applications and are brought into contact with the body surface.

Indications for Use

The Ultrasound System SONIMAGE MX1 and its transducers are products designed to collect ultrasonic image data of the human body for diagnostic purposes. The system employs the ultrasonic pulse-echo method to visualize the anatomic structures, characteristics, and dynamics of the human body, and using an image display, Doppler display and/or Doppler sound, offers a procedure applied to the human body for medical diagnosis or examination.

The range of intended clinical applications is same as other conventional ultrasound imaging systems for general purpose, such as small parts, abdomen, musculoskeletal, cardiac, and peripheral vascular.

This device is intended for use in healthcare facilities, such as health clinics and hospitals.

Intended user for the device are physician, sonographer, and other trained qualified healthcare professionals.

Modes of operation include B-mode, M-mode, PWD-mode, CWD-mode, Color Dopplermode, Power Doppler-mode, and their Combined mode.

  • Page 2 of 6 -

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image contains the Konica Minolta logo. The logo features a blue sphere with several horizontal white lines across the center. Below the sphere, the words "KONICA MINOLTA" are written in a bold, sans-serif font.

Comparison Table

The comparison to the predicate devices was summarized in the table blow.

Subject DevicePredicate Device
Device NameUltrasound System SONIMAGE MX1Ultrasound System SONIMAGE MX1
510(K) #K220993K180084
Indicationsfor UseThe Ultrasound System SONIMAGE MX1and its transducers are products designedto collect ultrasonic image data of thehuman body for diagnostic purposes. Thesystem employs the ultrasonic pulse-echomethod to visualize the anatomicstructures, characteristics, and dynamicsof the human body, and using an imagedisplay, Doppler display and/or Dopplersound, offers a procedure applied to thehuman body for medical diagnosis orexamination.The range of intended clinicalapplications is same as otherconventional ultrasound imaging systemsfor general purpose, such as small parts,abdomen, musculoskeletal, cardiac, andperipheral vascular.This device is intended for use inhealthcare facilities, such as health clinicsand hospitals.Intended user for the device arephysician, sonographer, and other trainedqualified healthcare professionals.The Ultrasound System SONIMAGE MX1and its transducers are productsdesigned to collect ultrasonic image dataof the human body for diagnosticpurposes. The system employs theultrasonic pulse-echo method tovisualize the anatomic structures,characteristics, and dynamics of thehuman body, and using an image display,Doppler display and/or Doppler sound,offers a procedure applied to the humanbody for medical diagnosis orexamination. The range of intendedclinical applications is same as otherconventional ultrasound imaging systemsfor general purpose, such as small parts,abdomen, musculoskeletal, andperipheral vascular.
Modes of operation include B-mode, M-mode, PWD-mode, CWD-mode, ColorDoppler-mode, Power Doppler-mode,and their Combined mode.*The transducer functions tablesinclude; B-mode, PWD-mode, ColorDoppler-mode, Power Doppler-mode,and their Combined mode.
510(k) TrackTrack 3Track 3
ScanningMethodElectronic LinearElectronic ConvexElectronic SectorElectronic LinearElectronic Convex
Modes ofoperation· B (2D)-mode· M-mode· B (2D)-mode· Color Flow-mode
Subject DevicePredicate Device
Device NameUltrasound System SONIMAGE MX1Ultrasound System SONIMAGE MX1
510(K) #K220993K180084
• w/Anatomical M mode• Color Flow-mode• Power Doppler-mode• PW Doppler-mode• CW Doppler-mode• Combination (BM, BD, Bc, BcD)• Power Doppler-mode• PW Doppler-mode• Combination (BD, Bc, BcD)
Transducer• CONVEX PROBE C5-2• CONVEX PROBE MC10-3• LINEAR PROBE L14-4• LINEAR PROBE L11-3• LINEAR PROBE L18-4• LINEAR PROBE HL18-4• SECTOR PROBE S4-2• CONVEX PROBE C5-2• CONVEX PROBE MC10-3• LINEAR PROBE L14-4• LINEAR PROBE L11-3
Function• Needle Visualization(Simple Needle Visualization)• Auto IMT• Wireless LAN• RA Workflow• Anti -Virus• DICOM Storage• DICOM Worklist• Internal Storage Encryption• Library• Shared Library• ECG Function• Vascular NAVI• Image improvement (iXRET)• Camera Function• Direct Recording• Needle Visualization(Simple Needle Visualization)• Auto IMT• Wireless LAN• RA Workflow• Anti -Virus• DICOM Storage• DICOM Worklist• Library
Optional kitBiopsy AdaptersAcoustic standoffReference Signal Unit (ECG Trigger)Biopsy AdaptersAcoustic standoff

— Page 3 of 6 —

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the Konica Minolta logo. The logo consists of a blue globe with horizontal white lines across it. Below the globe, the words "KONICA MINOLTA" are written in a sans-serif font.

The following technological differences exist between the subject and predicate devices:

  • Addition of M-mode and CW Doppler-mode (K182153). -

  • -Addition of Transducers: Linear probe L18-4 (K142197), HL18-4 (K162065) and Sector probe S4-2 (K151060).

  • ECG Function, which was already cleared on the reference device (K151060). -

  • Page 4 of 6

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Konica Minolta. The logo consists of a blue circle with white horizontal lines inside it. Below the circle, the words "KONICA MINOLTA" are written in a bold, sans-serif font.

  • -Shared Library Function: To save the still images and video/audio clips in the shared folder.
  • Vascular NAVI: To simplify vascular ultrasound by adjusted optimized parameters for blood flow evaluation.
  • iXRET (EXtended Resolution Enhancement Technology): To improve resolution while maintaining frame rate.
  • Camera Function: To simultaneously display ultrasonic images from a -USB-connected WEB Camera.
  • -Direct Recording Function: To save the still images and clip files to external storage media.

Performance Data

The Ultrasound System SONIMAGE MX1 is designed to comply with the following standards:

  • AAMI ANSI ES60601-1:2005/(R) 2012 and A1:2012, C1:2009/(R)2012 and • A2:2010/(R)2012 (Consolidated Text); Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD)
  • IEC 60601-1-2 Edition 4.0 2014-02; Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
  • IEC 60601-1-6 Edition 3.2 2020-07 CONSOLIDATED VERSION; Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
  • IEC 60601-2-37 Edition 2.1 2015; Medical electrical equipment - Part 2-37: Particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment
  • IEC 62359 Edition 2.1 2017-09 CONSOLIDATED VERSION; Ultrasonics Field characterization - Test methods for the determination of thermal and mechanical indices related to medical diagnostic ultrasonic fields
  • NEMA UD 2-2004 (R2009); Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment Revision 3
  • IEC 62304 Edition 1.1 2015-06 CONSOLIDATED VERSION; Medical device software -• Software life cycle processes

Page 5 of 6 -

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image contains the logo for Konica Minolta. The logo features a blue sphere at the top with several horizontal white lines across it. Below the sphere, the words "KONICA MINOLTA" are written in a bold, sans-serif font.

FDA guidance document "Marketing Clearance of Diagnostic Ultrasound Systems and Transducers", issued on June 27, 2019 was referenced for this submission, along with "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices Document" issued on May 11, 2005. Additionally, cybersecurity documentation, per the FDA guidance document "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices" issued on October 18, 2018 was also included in this submission, along with "Design Considerations and Premarket Submission Recommendations for Interoperable Medical Devices", issued on September 6, 2017.

Risk Analysis and verification and validation activities demonstrate that the established specifications for these devices have been met. The results of risk management did not require clinical studies to demonstrate the substantial equivalency of the proposed device.

Conclusion

The Ultrasound System SONIMAGE MX1 has the same intended use and the cardiac extended indications for use, technological characteristics, and principle of operations. The technological differences raised no new issues of safety or effectiveness as compared to its predicate device (K180084). Performance tests demonstrate that the Ultrasound System SONIMAGE MX1 performs according to specifications and functions as intended.

Therefore, as for our conclusion, the Ultrasound System SONIMAGE MX1 is substantially equivalent to the predicate device and presents no new questions of safety or effectiveness.

§ 892.1550 Ultrasonic pulsed doppler imaging system.

(a)
Identification. An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.