K Number
K220992

Validate with FDA (Live)

Date Cleared
2022-10-19

(198 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Device Description

A Disposable Powder Free Vinyl Exam Glove (with color Black, Blue, or Purple) is a patient examination glove made from poly (vinyl chloride), non-sterile (as per 21 CFR 880.6250, Class I). This device is available in size S, M, L, and XL. The operation principle of this medical device is to provide single use barrier protection for the wearer and the device meets all the requirement specifications for barrier protection, tensile properties as defined in ASTM D5250-19, Standard Specification for Poly (vinyl chloride) Gloves for Medical Application.

AI/ML Overview

The provided text describes the acceptance criteria and the results of non-clinical performance testing for a medical device, specifically Disposable Powder Free Vinyl Exam Gloves.

Here's a breakdown of the requested information:

1. A table of acceptance criteria and the reported device performance

TestAcceptance CriteriaReported Device Performance
DimensionLength: 230 mm min.Thickness:- Palm - 0.08 mm min.- Finger - 0.08 mm min.Palm Width:- Small 85 ± 5 mm- Medium 95 ± 5 mm- Large 105 ± 5 mm- X Large 115 ± 5 mmPass
Freedom from holes (Water leak)G-I/AQL 2.5Pass
Tensile strengthBefore Aging: 11 MPa, min.After Aging: 11 MPa, min.Pass
ElongationBefore Aging: 300% min.After Aging: 300% min.Pass
Powder Residual< 2mg per glovePass
Biocompatibility - Skin IrritationNegative ResponsePass
Biocompatibility - Skin SensitizationNo contact sensitizationPass
Biocompatibility - CytotoxicityNo in vitro cytotoxicityPass

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document does not explicitly state the sample size used for each specific test or the data provenance. It only states "Pass" for each criterion, indicating that the tests were conducted.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not applicable as the acceptance criteria are based on standardized test methods and physical/chemical properties, not on expert interpretation of medical images or patient data.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not applicable as the assessment involves standardized laboratory tests with objective measurements, not expert adjudication of subjective findings.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This information is not applicable. The device is a physical product (examination gloves), not an AI-assisted diagnostic tool. Therefore, an MRMC study is irrelevant.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

This information is not applicable. The device is a physical product, not a standalone algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The "ground truth" for the performance evaluation of these gloves is based on established international and national standards for physical, chemical, and biological properties of medical gloves. Specifically, the tests reference:

  • ASTM D3767 (for Dimension)
  • 21 CFR 800.20 & ASTM D5151-19 (for Freedom from holes)
  • ASTM D412-16 & ASTM D573-04 (for Tensile strength & Elongation)
  • ASTM D6124-06 (for Powder Residual)
  • ISO 10993-10:2010 and ISO 10993-23:2021 (for Biocompatibility - Skin Irritation and Sensitization)
  • ISO 10993-5:2009 (for Biocompatibility - Cytotoxicity)

These standards define the acceptable range or condition for each characteristic.

8. The sample size for the training set

This information is not applicable. The described study is a non-clinical performance evaluation against predetermined standards, not a machine learning model requiring a training set.

9. How the ground truth for the training set was established

This information is not applicable for the same reasons as point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

October 19, 2022

Ever Global (Vietnam) Enterprise Corporation % Elizabeth Deng US Agent Elizabeth Deng 5748 Eaglewood Place Ranch Cucamonga, California 91739

Re: K220992

Trade/Device Name: Disposable Powder Free Vinyl Exam Glove, Black/Blue/Purple Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LYZ Dated: September 15, 2022 Received: September 19, 2022

Dear Elizabeth Deng:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Bifeng Qian, M.D., Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K220992

Device Name

Disposable Powder Free Vinyl Exam Glove, Black/Blue/Purple

Indications for Use (Describe)

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (6/20)

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image is a logo for Ever Global Enterprise Corp. The logo features the letters 'Eg' in yellow, with 'E' being larger than 'g'. Behind the letters is a blue hand. The words 'EVER GLOBAL' are at the top of the logo, and 'ENTERPRISE CORP.' is at the bottom, both in white text on a blue background.

LONG THANH INDUSTRIAL ZONE, TAMAN VILLAGE, DONG FAX: (0251)351 4023 024 BOOK.COM/PMGLOVESSUPERIEURVIETNAM

510(k) SUMMARY

The assigned 510(k) Number: K220992

This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21 CFR 807.92.

0.0 Summary Preparation Date: -Oct 18th, 2022

1.0 Submitter:

Submitter's name :Ever Global (Vietnam) Enterprise Corporation
Submitter's address:Long Thanh Industrial ZoneTaman Village Dong Nai Province, VN 810000
Phone number:84-61-3514022
Fax number:84-61-3514023
Name of contact person:Jerry Lin

Elizabeth Deng

2.0 US Agent:

US Representative Name: Company Address:

Telephone Number: Contact Email Address:

5748 Eaglewood Place Rancho Cucamonga, California Rancho Cucamonga, CA 91739 909 4659188 baxianunited48(@yahoo.com

3.0 Name of the Device

Proprietary/Trade name:

Common Name: Classification Name: Device Classification: Regulation Number: Product Code:

4.0 Predicate device

Device Name:

Company name: 510(K) Number: Disposable Powder Free Vinyl Exam Glove, Black/Blue/Purple Vinyl Examination Gloves Non-powdered Patient Examination Glove Class I 21 CFR 880.6250 LYZ

Disposable Powder Free Vinyl Examination Glove, Clear/Yellow Ever Global (Vietnam) Enterprise Corporation K170575

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image is a logo for Ever Global Enterprise Corp. The logo features the letters 'E', 'G', and 'C' in yellow, superimposed on a blue hand. The words 'EVER GLOBAL' are at the top of the logo, and 'ENTERPRISE CORP.' is at the bottom. The logo is framed by a blue border.

LONG THANH INDUSTRIAL ZONE, TAMAN VILLAGE, DONG NAI 810000 VIETNA AX: (0251)351 4023 024 ACEBOOK.COM/PMGLOVESSUPERIEURVIETNAM

510(k) SUMMARY

The assigned 510(k) Number: K220992

5.0 Device Description:

A Disposable Powder Free Vinyl Exam Glove (with color Black, Blue, or Purple) is a patient examination glove made from poly (vinyl chloride), non-sterile (as per 21 CFR 880.6250, Class I). This device is available in size S, M, L, and XL. The operation principle of this medical device is to provide single use barrier protection for the wearer and the device meets all the requirement specifications for barrier protection, tensile properties as defined in ASTM D5250-19, Standard Specification for Poly (vinyl chloride) Gloves for Medical Application.

6.0 Device Indications for use:

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

DeviceCharacteristicPredicate DeviceSubject DeviceComparison
Product NameDisposable Powder Free VinylExamination Glove,Clear/YellowDisposable Powder Free VinylExam Glove, Black/Blue/PurpleSame exceptcolor
510(K) No.K170525K220992n/a
Product OwnerEverGlobal(Vietnam)Enterprise CorporationEverGlobal(Vietnam)Enterprise Corporationsame
Product CodeLYZLYZsame
Regulation21 CFR 880.625021 CFR 880.6250same
ClassIIsame
Indications forUseA patient examination glove is adisposable device intended formedical purposes that is wornon the examiner's hand or fingerto prevent contaminationbetween patient and examiner.A non-powdered patientexamination glove is adisposable device intended formedical purposes that is wornon the examiner's hand or fingerto prevent contaminationbetween patient and examiner.same
Power freeYesYessame
SizeSmall/ Medium/Large/X LargeSmall/Medium/Large/X Largesame
Over-The-Counter UseYes (21 CFR 801 Subpart C)Yes (21 CFR 801 Subpart C)same
Single UseYESYESsame
Non-SterileYESYESsame
Dimensions-LengthComplies with ASTM D5250-06230 mm min.Complies with ASTM D5250-19230 mm min.same

7.0 Comparison of device technological characteristics:

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image is a logo for Ever Global Enterprise Corp. The logo features the letters 'F' and 'g' in a stylized font, with the letters overlapping each other. The letters are yellow, and they are set against a blue background that resembles a hand. The words 'EVER GLOBAL' are at the top of the logo, and the words 'ENTERPRISE CORP.' are at the bottom.

Long Thanh Industrial Zone, Taman Village, Dong NAI 810000 VIETNAM TEL: (0251) 896 6676 FAX: (0251)351 4023 024 EMAIL: INFO@EGVNCO.COM FACEBOOK.COM/PMGLOVESSUPERIEURVIETNAM

510(k) SUMMARY

The assigned 510(k) Number: K220992

DeviceCharacteristicPredicate DeviceSubject DeviceComparison
Dimensions -Palm WidthComplies with ASTM D5250-06Small    85 ± 5Medium  95 ± 5Large   105 ± 5X large   115 ± 5Complies with ASTM D5250-19Small    85 ± 5Medium  95 ± 5Large   105 ± 5X large   115 ± 5same
Dimensions -ThicknessComplies with ASTM D5250-06Palm - 0.08mm min.Finger - 0.1 mm min.Complies with ASTM D5250-19Palm - 0.08 mm min.Finger - 0.08 mm min.same, bothare thickerthanstandardrequest
PhysicalPropertiesTensile StrengthBefore Aging: 11 MPa, min.After Aging: 11 MPa, min.Elongation:Before Aging: 300% min.After Aging: 300% min.Tensile StrengthBefore Aging: 11 MPa, min.After Aging: 11 MPa, min.Elongation:Before Aging: 300% min.After Aging: 300% min.same
ResidualpowderComplies with D5250-06< 2mg per gloveComplies with ASTM D5250-19< 2mg per glovesame
Freedom fromHolesIn accordance with ASTMD5250-06 (G-1 with AQL 2.5)In accordance with ASTMD5250-19 (G-1 with AQL 2.5)same
BiocompatibilityISO 10993-10:-Skin Sensitization: PassIrritation test: PassISO 10993-10 (Blue & PurpleGloves) & ISO 10993-23 (BlackGlove):-Skin Irritation test: PassISO 10993-10-Skin Sensitization Test: PassISO 10993-5-In vitro cytotoxicity test: Passsimilar.Subjectdevice passcytotoxicitytest

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image is a logo for Ever Global Enterprise Corp. The logo features the letters 'E' and 'G' in a stylized, interconnected design, with the letters 'e' and 'c' below. The letters are in yellow and are set against a blue background that resembles a hand. The words 'EVER GLOBAL' are at the top of the logo, and 'ENTERPRISE CORP.' is at the bottom.

long thanh industrial zone, taman village, dong NAI 810000 VIETNAM TEL: (0251) 896 6676 FAX: (0251)351 4023 024 EMAIL: INFO@EGVNCO.COM FACEBOOK.COM/PMGLOVESSUPERIEURVIETNAM

510(k) SUMMARY

The assigned 510(k) Number: K220992

8.0 Assessment of Non-Clinical Performance Data:

TestTest MethodPurposeAcceptance CriteriaResults
DimensionASTM D3767Determine thegeometrical dimensionof glovesLength: 230 mm min.Thickness:Palm - 0.08 mm min.Finger - 0.08 mm min.Palm Width:Small 85 ± 5 mmMedium 95± 5 mmLarge 105 ± 5 mmX Large 115 ± 5 mmPass
Freedom from holes(Water leak)21 CFR 800.20. &ASTM D5151-19Detect the holes on thegloves.G-I/AQL 2.5Pass
Tensile strength(Before aging/Afteraging)ASTM D412-16& ASTM D573-04Evaluate the tensile(tenson) properties ofthe gloves. In addition,it also determines theinfluence of elevatedBefore Aging: 11 MPa,min.After Aging: 11 MPa,minPass
Elongation (Beforeaging/After aging)ASTM D412-16& ASTM D573-04temperature on thephysical properties ofgloves.Before Aging: 300%min.After Aging: 300% min.Pass
Powder ResidualASTM D6124-06Determine the averagepowder mass found onthe gloves< 2mg per glovePass
Biocompatibility-Skin IrritationISO 10993-10:2010 (Blue &Purple Gloves) &ISO 10993-23:2021 (BlackGlove)determine the potentialof glove to promoteskin sensitization &irritation reactions afterrepeated applicationsNegative ResponsePass
Biocompatibility-Skin SensitizationISO 10993-10:2010No contact sensitizationPass
Biocompatibility-cytotoxicityISO 10993-5:2009determine thecytotoxicity potentialof gloveNo in vitro cytotoxicityPass

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image is a logo for Ever Global Enterprise Corp. The logo features the letters 'F' and 'g' in a stylized font, with the letter 'F' stacked on top of the letter 'g'. The letters are yellow and are set against a blue background that resembles a hand. The words 'EVER GLOBAL' are written above the letters, and the words 'ENTERPRISE CORP.' are written below the letters.

Long Thanh Industrial Zone, Taman Village, Dong NAI 810000 VIETNAM TEL: (0251) 896 6676 FAX: (0251)351 4023 024 EMAIL: INFO@EGVNCO.COM FACEBOOK.COM/PMGLOVESSUPERIEURVIETNAM

510(k) SUMMARY

The assigned 510(k) Number: K220992

9.0 Assessment of Clinical Performance Data:

Clinical data is not needed for this type of device.

10.0 Conclusion:

The conclusion drawn from the nonclinical tests demonstrate that the subject device Disposable Powder Free Vinyl Exam Glove, Black/Blue/Purple is as safe, as effective, and performs as well as or better than the legally marketed device.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.