K Number
K220988

Validate with FDA (Live)

Device Name
EksoNR
Manufacturer
Date Cleared
2022-06-09

(66 days)

Product Code
Regulation Number
890.3480
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The EksoNR™ is intended to perform ambulatory functions in rehabilitations under the supervision of a trained physical therapist for the following populations:

· Individuals with multiple sclerosis (upper extremity motor function of at least one arm).

• Individuals with acquired brain injury, including traumatic brain injury and stroke (upper extremity motor function of at least 4/5 in at least one arm).

· Individuals with spinal cord injuries at levels T4 to L5 (upper extremity motor function of at least 4/5 in both arms).

· Individuals with spinal cord injuries at levels of C7 to T3 (ASIA D with upper extremity motor function of at least 4/5 in both arms).

The therapist must complete a training program prior to use of the devices are not intended for sports or stair climbing

Device Description

The Ekso is a powered motorized orthosis. It consists of a fitted metal brace that supports the legs, feet, and torso. It is worn via straps on the body, legs, and feet. Battery powered motors drive knee and hip joints. It has an integrated solid torso containing the computer and power supply. It has a hand-held user interface to specify settings and initiate steps. The Ekso is used with a cane, crutch, or walker.

AI/ML Overview

Here's an analysis of the provided text, extracting information related to acceptance criteria and the study proving the device meets them. Please note that the document is a 510(k) summary for a medical device (exoskeleton), which focuses on demonstrating substantial equivalence to a predicate device rather than presenting a formal "acceptance criteria" table as might be seen for an AI/ML diagnostic tool. Therefore, some requested information may not be directly available or applicable in the provided context.

Overview:
The document clears the EksoNR powered lower extremity exoskeleton. The primary purpose of this 510(k) submission (K220988) is to expand the indications for use of the already cleared EksoNR (predicate device K200574) to include individuals with multiple sclerosis (MS). The device itself is essentially unchanged from the predicate.


Acceptance Criteria and Device Performance

Given this is a 510(k) for an updated indication for an existing medical device (exoskeleton) rather than an AI/ML diagnostic, the concept of "acceptance criteria" isn't framed as statistical thresholds for sensitivity/specificity. Instead, the acceptance criteria are implicitly that the device is safe and effective for the expanded population, demonstrating gait ambulation effectiveness and no new safety concerns in the specified population, thus maintaining substantial equivalence to the predicate.

Table of Acceptance Criteria (Implicit) and Reported Device Performance:

Acceptance Criteria (Implicit)Reported Device Performance
Safety: Device is safe for the expanded MS population (no new adverse events).ADEMBI MS Study: "There were no falls or other adverse events reported."Kessler MS Study: "There were no falls or other adverse events reported."Overall Conclusion: "The clinical data reported no adverse events demonstrating the device is safe on this patient population when used in accordance with existing labeling."
Effectiveness: Device effectively facilitates gait ambulation in the expanded MS population.ADEMBI MS Study (n=17-18):- TUG scores improved significantly: from 24s at baseline to 20.61s at completion of all Ekso sessions.- Average walking speed maintained: from 0.69m/s at baseline to 0.66m/s at completion of all Ekso sessions.- Average cognitive MOCA scores improved slightly.Kessler MS Study (n=8):- TUG scores improved significantly: from 16.99s at baseline to 14.15s at completion of all Ekso sessions.- Average walking speed maintained: from 10.37s at baseline to 10.63s at completion of all Ekso sessions (note: units here are stated as 's' which is unusual for speed, likely a typo and referring to time for a specific distance or a different test. See 6MWT below).- Symbol digit modalities test (cognition) improved.- 6 minute walk test (6MWT) distance increased: from 279.65m to 294.69m.- Average functional reach test distance increased.Overall Conclusion: "The supporting clinical data demonstrating the use of the product with patients with multiple sclerosis (MS), show that the device effectively facilitates gait ambulation in the expanded patient population."
Substantial Equivalence: Device remains substantially equivalent to the predicate (K200574)."This device and the previously cleared (predicate) device (K200574) are essentially the same products.""The device is essentially unchanged from the current (predicate) device.""The indications for use are identical to that of the predicate device, with the addition of the following: Individuals with multiple sclerosis...""When used as instructed, the device is as safe to use with a broader population of patients with neurological conditions to include the already cleared ABI and SCI population and this new MS population."

Study Details

  1. Sample sizes used for the test set and the data provenance:

    • Test Sets:
      • ADEMBI MS Study: 18 subjects total used the Ekso. (n=17 for TUG post-testing, n=18 for others).
      • Kessler MS Study: 9 subjects total used the Ekso. (n=8 for pre/post testing).
    • Data Provenance: Not explicitly stated regarding country of origin, but described as "Single center, prospective parallel-assignment, single-blinded, randomized controlled study" for ADEMBI MS and "Single center, randomized study" for Kessler MS. This implies prospective data collection for these specific studies. No specific mention of retrospective vs. prospective is made for the original 7 studies that supported the predicate.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This is not an AI/ML diagnostic device where ground truth is established by expert readers interpreting images or data. For an exoskeleton, the "ground truth" for effectiveness is whether the patient can ambulate and improve functional scores, as measured by standardized clinical tests (TUG, 6MWT), and the "ground truth" for safety is the absence of adverse events/falls during supervised use.
    • The device is used under the supervision of a "trained physical therapist." While not "experts establishing ground truth" in the diagnostic sense, their training and supervision are crucial for the device's safe and effective use and thus for the clinical outcomes.
  3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not applicable. This is a functional device study, not an imaging diagnostic study requiring adjudication of image interpretations for ground truth. Clinical outcomes were measured directly using standardized tests and adverse events were reported.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This is not an AI-assisted diagnostic tool. No MRMC study was performed or needed. The device (EksoNR) is the primary intervention.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. The device is a physical exoskeleton designed for human-in-the-loop use under therapist supervision. It does not operate as a standalone algorithm.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Outcomes Data: The primary ground truth is established through measured clinical outcomes using validated functional assessment tools (e.g., Timed Up and Go (TUG) test, 6-minute walk test (6MWT), cognitive assessments like MOCA/Symbol Digit Modalities Test) and direct observation/reporting of adverse events (falls). This is outcomes-based evidence of safety and functional improvement.
  7. The sample size for the training set:

    • Not applicable in the context of device approval for an exoskeleton. This is not an AI/ML algorithm that requires a "training set" in the computational sense. The device itself is "trained" during the manufacturing and design process, and the physical therapists are trained for its use.
  8. How the ground truth for the training set was established:

    • Not applicable. As above, there is no AI/ML "training set" for this device. The physical therapists are trained users, and their "ground truth" would be established through their professional education, experience, and the specific training program provided for the EksoNR device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

June 9, 2022

Ekso Bionics, Inc. Michael Glover Global Director of Clinical Experience 1414 Harbour Way South, Suite 1201 Richmond, California 94804

Re: K220988

Trade/Device Name: EksoNR Regulation Number: 21 CFR 890.3480 Regulation Name: Powered Lower Extremity Exoskeleton Regulatory Class: Class II Product Code: PHL Dated: March 31, 2022 Received: April 4, 2022

Dear Michael Glover:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Heather Dean, PhD Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K220988

Device Name EksoNR

Indications for Use (Describe)

The EksoNR™ is intended to perform ambulatory functions in rehabilitations under the supervision of a trained physical therapist for the following populations:

· Individuals with multiple sclerosis (upper extremity motor function of at least one arm).

• Individuals with acquired brain injury, including traumatic brain injury and stroke (upper extremity motor function of at least 4/5 in at least one arm).

· Individuals with spinal cord injuries at levels T4 to L5 (upper extremity motor function of at least 4/5 in both arms).

· Individuals with spinal cord injuries at levels of C7 to T3 (ASIA D with upper extremity motor function of at least 4/5 in both arms).

The therapist must complete a training program prior to use of the devices are not intended for sports or stair climbing

Type of Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary as required by 21 CFR 807.92(c)

Device nameEkso
Submitters name &contact infoEkso Bionics® Inc.1414 Harbour Way SouthSuite 1201Richmond, CA 94804
Contact Details:Scott DavisPresident and COOTel: +1 603-831-2980Email: sdavis@eksobionics.com
Ekso Bionics Tel: +1 (510) 984-1761Ekso Bionics Fax: +1 (510) 927-2647
Preparation DateMarch 29, 2022
Device Name &ClassificationTrade Name: EksoNR™Common Name: ExoskeletonClassification Name: Powered ExoskeletonDevice Classification: Class II, 21 CFR 890.3480Product Code: PHL
Legally MarketedPredicate DeviceK200574, EksoNR, Ekso Bionics, Inc.
Device DescriptionThe Ekso is a powered motorized orthosis. It consists of a fitted metalbrace that supports the legs, feet, and torso. It is worn via straps on thebody, legs, and feet. Battery powered motors drive knee and hip joints. Ithas an integrated solid torso containing the computer and power supply. Ithas a hand-held user interface to specify settings and initiate steps. TheEkso is used with a cane, crutch, or walker.
Indication for UseStatementThe EksoNR™ is intended to perform ambulatory functions in rehabilitationinstitutions under the supervision of a trained physical therapist for thefollowing populations:• Individuals with multiple sclerosis (upper extremity motor functionof at least 4/5 in at least one arm)• Individuals with acquired brain injury, including traumatic braininjury and stroke (upper extremity motor function of at least 4/5 inat least one arm).• Individuals with spinal cord injuries at levels T4 to L5 (upperextremity motor function of at least 4/5 in both arms).• Individuals with spinal cord injuries at levels of C7 to T3 (ASIA Dwith upper extremity motor function of at least 4/5 in both arms).The therapist must complete a training program prior to use of the device.The devices are not intended for sports or stair climbing.
SubstantialEquivalenceDiscussionThis device and the previously cleared (predicate) device (K200574) areessentially the same products.
Differences in Indications for UseThe purpose of this 510(k) is to update the indications for use. Theindications for use are identical to that of the predicate device, with theaddition of the following:
● Individuals with multiple sclerosis (upper extremity motor functionof at least 4/5 in at least one arm)
The change noted above expands the existing indications to includeindividuals with multiple sclerosis (MS). The intended use of the device isunchanged. For the purposes of this intended use (ambulatoryrehabilitation), patients with MS present similarly, are screened for devicesafety requirements, and are treated in the same way as patients with ABIor SCI. As such, gait ambulation effectiveness of the device is no differentwhen used on patients with MS. When used as instructed, the device is assafe to use with a broader population of patients with neurologicalconditions to include the already cleared ABI and SCI population and thisnew MS population.

{4}------------------------------------------------

Technical Characteristics

The device is essentially unchanged from the current (predicate) device. All changes since the previous submission fall below the threshold requiring a 510(k) per the FDA's October 25, 2017, " Guidance for Industry and FDA Staff: Deciding When to Submit a 510(k) for a Change to an Existing Device". Even though there is no change, for completeness, the technical characteristics of the device are compared with the current device below.

Current DeviceCurrent Device
ManufacturerEkso Bionics®, Inc.Ekso Bionics®, Inc.
Trade NameEksoNR™EksoNR™Differences
510(k) NumberK220988K200574N/A
Product CodePHLPHLSame
Regulation Number890.3480890.3480Same
Regulation NamePowered ExoskeletonPowered ExoskeletonSame
Body CoverageWorn over legs and upperbody with rigid torsoWorn over legs and upperbody with rigid torsoSame
Size of ComponentsAdjustable upper leg, lowerleg, and hip width; controlunit integrated into thetorsoAdjustable upper leg, lowerleg, and hip width; controlunit integrated into thetorsoSame
Comparison of Technical Characteristics
Current DeviceCurrent Device
ManufacturerEkso Bionics®, Inc.Ekso Bionics®, Inc.
Trade NameEksoNR™EksoNR™Differences
Mobility AidWalker, Crutches, CaneWalker, Crutches, CaneSame
Ability of User MobilitySit, stand, walk, andTurnSit, stand, walk, andTurnSame
Walking Speed~2 km/hr~2 km/hrSame
Grade of Inclination1.15 deg1.15 degSame
Type of SurfaceSmooth, cement, carpetSmooth, cement, carpetSame
Height of Patient~62″ to 74″ (1.58 m to 1.88m)~62″ to 74″ (1.58 m to 1.88m)Same
Weight of PatientUp to 220 lbs (100kg)Up to 220 lbs (100kg)Same
Control MethodHandheld interface for PT;weight shift to initiate stepsHandheld interface for PT;weight shift to initiatestepsSame
Range of MotionHips: 135° flexion to 20°extensionKnees: 130° flexion to 0°extensionAnkles: 10° flexion to 10°extensionHips: 135° flexion to 20°extensionKnees: 130° flexion to 0°extensionAnkles: 10° flexion to 10°extensionSame
Device Weight50 lbs (23 kg)50 lbs (23 kg)Same
Rechargeable BatteryRechargeable lithium ionbatteries 48.1V, 30A peakcurrent, 1 hour ofcontinuous usage perchargeRechargeable lithium ionbatteries 48.1V, 30A peakcurrent, 1 hour ofcontinuous usage perchargeSame
Battery Charge Time1 hour1 hourSame
Expected Useable Life4 years4 yearsSame
Training ProgramYesYesSame
Certification ProgramYesYesSame
User FeedbackProvides visual feedback onthe handheld controller andauditory feedbackProvides visual feedback onthe handheld controllerand auditory feedbackSame
Fall Detection andMitigationNoneNoneSame
Failsafe FeatureIn event of power failure-knees become locked andhips free (similar to typicalpassive leg braces)In event of power failure-knees become locked andhips free (similar to typicalpassive leg braces)Same
Current DeviceCurrent Device
ManufacturerEkso Bionics®, Inc.Ekso Bionics®, Inc.
Trade NameEksoNR™EksoNR™Differences
Operating Temperature10° to 95°F (-12° to -35° C)10° to 95°F (-12° to -35° C)Same
Operating HumidityNot availableNot availableSame
Electrical Safety TestingIEC 60601-1:2005 with US deviationsIEC 60601-1:2005 with US deviationsSame
Electromagnetic Compatibility TestingPassed IEC 60601-1-2: 2014Passed IEC 60601-1-2: 2014Same

Comparison of Technical Characteristics

{5}------------------------------------------------

{6}------------------------------------------------

Clinical Performance Summary

The current device utilized 7 different studies to demonstrate safety and efficacy of the current indications for use (K200574). This submission builds from those 7 studies with 2 additional clinical studies (summarized below) focusing on the specific population added by the expanded indication for use in this submission. These studies demonstrate the device effectively enables gait ambulation, and there were no reported adverse events indicating the device is safe to use with this expanded population.

{7}------------------------------------------------

StudyDescription
ADEMBI MSSingle center, prospective parallel-assignment, single-blinded, randomized controlledstudy of patients with multiple sclerosis. Only patients who were randomized toutilize Ekso were included in this submission.
Duration of Intervention• Ambulation during a mean of 18.4 sessions per patient (range 13 to 20)
18 subjects total used the Ekso• Mean time since MS diagnosis: $12.94\pm8.11$ years• 3 were diagnosed with primary progressive MS• 8 were diagnosed with secondary progressive MS• 7 were diagnosed with relapsing remitting MS
Results:For those participants who completed pre and post testing (n=17 for TUG, n=18 forothers)
• TUG scores improved significantly from 24s at baseline to 20.61s atcompletion of all Ekso sessions• Average walking speed was maintained throughout the intervention from0.69m/s at baseline to 0.66m/s at completion of all Ekso sessions• Average cognitive MOCA scores improved slightly from 25.39 at baseline to26.06 at completion of all Ekso sessions• There were no falls or other adverse events reported

{8}------------------------------------------------

Single center, randomized study of patients with multiple sclerosis. Only patients who Kessler MS were randomized to utilize Ekso were included in this submission.

Duration of Intervention

  • Ambulation during a mean of 7.78 sessions per patient (range 6 to 8) ●
    9 subjects total used the Ekso

  • 1 was diagnosed with secondary progressive MS

  • . 8 were diagnosed with relapsing remitting MS

Results:

For those participants who completed pre and post testing (n=8)

  • . TUG scores improved significantly from 16.99s at baseline to 14.15s at completion of all Ekso sessions
  • Average walking speed was maintained throughout the intervention from 10.37s at baseline to 10.63s at completion of all Ekso sessions
  • Symbol digit modalities test measuring cognition improved from baseline scores of 39.38 to post scores of 45.63
  • 6 minute walk test (6MWT) distance increased from 279.65m to 294.69m
  • Average functional reach test distance increased from baseline of 9.61cm to 10.46cm at completion of all Ekso sessions
  • There were no falls or other adverse events reported

Substantial Equivalence Conclusion

This device is substantially equivalent to the current device has technical characteristics and intended uses listed above to facilitate gait ambulation. The supporting clinical data demonstrating the use of the product with patients with multiple sclerosis (MS), show that the device effectively facilitates gait ambulation in the expanded patient population. The clinical data reported no adverse events demonstrating the device is safe on this patient population when used in accordance with existing labeling.

N/A