K Number
K220791

Validate with FDA (Live)

Manufacturer
Date Cleared
2022-05-03

(46 days)

Product Code
Regulation Number
880.6850
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Surgical and Prosthetic Trays are intended for organizing, sterilizing and storing of instruments. The Surgical and Prosthetic Trays are not intended to maintain sterility and are to be used in conjunction with a legally marketed, validated sterilization pouch.

Device Description

The proposed Surgical and Prosthetic Trays are reusable perforated containers that are intended to store and organize the instruments needed before, during and after implant surgery and the prosthetic procedure as well as hold instruments during sterilization.

The Trays are containers composed of three main components: a lid, a base and an overlay, which are all made of polyphenylsulfone (Radel R5000NT). The trays have co-molded silicone tooling supports and silicone rubber grommets to hold instruments. The base and overlay are custom printed to indicate the guided surgery or prosthetic procedure workflow and the position of the instruments within the tray for the different product lines.

The proposed Surgical and Prosthetic Trays are available in two tray sizes. The Surgical trays are available in medium size with outer dimensions of 7.3 x 5.5 x 2.4 inches and the Prosthetic trays are available in small size with outer dimensions of 5.4 x 3.9 x 2.4. Each tray size is offered in 2 inner tray configurations with varied instrument loads, see Table 1.

AI/ML Overview

The provided text describes the acceptance criteria and the studies that prove the device meets these criteria for "Surgical and Prosthetic Trays" (K220791).

Here's a breakdown of the requested information:

1. Table of Acceptance Criteria and Reported Device Performance

Test NameTest MethodologyAcceptance CriteriaReported Device Performance
Sterilization cycle validationANSI/AAMI/ISO 17665-1:2006/(R)2013, ISO 17665-2:2009Sterility assurance level (SAL) of ≤ 10-6Pass
Drying validationAAMI TIR12:2020, ANSI/AAMI/ISO 17665-2:2009No visible moisture, ≤ 3% weight gain of packaging and absorbable materialsPass
Cleaning (Manual Pre-Cleaning and automated Cleaning)AAMI TIR 30:2011, ISO 15883-1:2014No visible soil.Protein and TOC content Limit < 5 [µg/cm²] Protein and TOC content Limit < 95 [µg/device]Pass
Reprocessing of trays (cleaning and sterilization)Internal Test MethodNo signs of flush rust, rust corrosion, deformation or damagePass
Simulated use of traysInternal Test MethodNo significant wear of the holdersPass
Packaging testingISTA 2A(2011)No signs of cracks or ruptures on the surgical traysPass
CytotoxicityISO 10993-5:2009Inhibition of cell proliferation must be at or below 30% compared to untreated culturesPass

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify sample sizes for individual tests (e.g., number of trays tested for sterilization, drying, or cleaning). It refers to the "Proposed device" for sterility and drying validation, and the "Predicate device" for cleaning, reprocessing, simulated use, packaging, and cytotoxicity.

The data provenance is from non-clinical performance data, meaning these are laboratory or bench tests, not involving human subjects. The country of origin of the data is not explicitly stated, but the submission is to the U.S. FDA, and the company is Dentsply Sirona Inc., with an address in York, Pennsylvania, USA. The testing methodologies are based on international standards (ANSI/AAMI/ISO, ISO, AAMI TIR) which are widely recognized.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

This information is not provided in the document. For these types of non-clinical, performance-based tests (sterilization, drying, material properties), the "ground truth" is typically established by adhering to referenced international standards and specifications, rather than expert consensus on individual cases. The "experts" would be the personnel performing and analyzing these standardized tests, whose qualifications are usually governed by quality management systems within the testing facilities.

4. Adjudication Method for the Test Set

This information is not provided. Given the nature of the tests, which are objective measurements against predefined criteria (e.g., SAL ≤ 10-6, ≤ 3% weight gain, specific chemical limits), an adjudication method in the sense of multiple expert reviews is not typically applicable. The results are usually a direct outcome of the test protocol.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is relevant for evaluating the performance of diagnostic imaging devices or algorithms where human readers interpret medical images. The device in question is a surgical tray for sterilization and storage, which does not involve human interpretation of data in the same way.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

No, this is not applicable. The device is a physical sterile processing tray, not a software algorithm. Therefore, "standalone algorithm performance" is not a concept that applies to this device.

7. The Type of Ground Truth Used

The "ground truth" for these tests is based on objective criteria and methodologies defined by recognized international and national standards (e.g., ANSI/AAMI/ISO, ISO, AAMI TIR, ISTA). For example:

  • Sterilization: Sterility Assurance Level (SAL) ≤ 10-6.
  • Drying: No visible moisture, ≤ 3% weight gain.
  • Cleaning: Absence of visible soil and specific limits for protein and TOC content.
  • Material Biocompatibility (Cytotoxicity): Defined inhibition of cell proliferation percentage.
  • Mechanical Performance (Reprocessing, Simulated Use): Absence of visible damage, deformation, significant wear.

8. The Sample Size for the Training Set

Not applicable. The device is a physical product (surgical tray), not an AI/ML algorithm. Therefore, there is no "training set" in the context of machine learning.

9. How the Ground Truth for the Training Set Was Established

Not applicable for the same reason as point 8.

{0}------------------------------------------------

May 3, 2022

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Dentsply Sirona Courtney Clark Senior Director of Regulatory Affairs, Corporate 221 West Philadelphia Street, Suite 60W York, Pennsylvania 17401 USA

Re: K220791

Trade/Device Name: Surgical and Prosthetic Trays Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap containers, Trays, Cassettes & Other Accessories Regulatory Class: Class II Product Code: KTC Dated: March 16, 2022 Received: March 18, 2022

Dear Courtney Clark:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for

Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K220791

Device Name Surgical and Prosthetic Trays

Indications for Use (Describe)

The Surgical and Prosthetic Trays are intended for organizing, sterilizing and storing of instruments. The Surgical and Prosthetic Trays are not intended to maintain sterility and are to be used in conjunction with a legally marketed, validated sterilization pouch.

Sterilization parameters

Pre-Vacuum Steam at 132 °C (270 °F) for 4 min with a 20 minute dry time Do not exceed the worst-case validated maximum load:

Product nameArticle NumberMaximum Load (g)Vent to Volume Ratio (in -1)
OmniTaper Surgical Tray68015282513.70.033
PrimeTaper Surgical Tray68015321513.70.033
PrimeTaper Surgical Tray GS Medium68017268563.50.033
OmniTaper Surgical Tray GS68017203563.50.033
Bone Reamer Tray68015280271.90.046
Prosthetic Tray68015288271.90.046

Type of Use (Select one or both, as applicable)

| X | Prescription Use (Part 21 CFR 801 Subpart D)

__ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995,

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Submitter Information:

Dentsply Sirona Inc. 221 West Philadelphia Street Suite 60W York, PA 17401

Contact Person: Courtney Clark Telephone Number: 248-895-4379 Fax Number: 717-849-4343 Email: corporate-ra@dentsplysirona.com

Date Prepared: May 3, 2022

Device Name:

• Proprietary Name:Surgical and Prosthetic Trays
• Classification Name:Sterilization Wrap Containers, Trays, Cassettes & Other
Accessories
• CFR Number:21 CFR 880.6850
• Device Class:Class II
• Product Code:KCT

Predicate Device:

The predicate device identified is the following:

Predicate Device Name510(k)Company Name
Surgical TraysK212281Dentsply Sirona

Device Description:

The proposed Surgical and Prosthetic Trays are reusable perforated containers that are intended to store and organize the instruments needed before, during and after implant surgery and the prosthetic procedure as well as hold instruments during sterilization.

The Trays are containers composed of three main components: a lid, a base and an overlay, which are all made of polyphenylsulfone (Radel R5000NT). The trays have co-molded silicone tooling supports and silicone rubber grommets to hold instruments. The base and overlay are custom printed to indicate the guided surgery or prosthetic procedure workflow and the position of the instruments within the tray for the different product lines.

The proposed Surgical and Prosthetic Trays are available in two tray sizes. The Surgical trays are available in medium size with outer dimensions of 7.3 x 5.5 x 2.4 inches and the Prosthetic trays are available in small size with outer dimensions of 5.4 x 3.9 x 2.4. Each tray size is offered in 2 inner tray configurations with varied instrument loads, see Table 1.

{4}------------------------------------------------

Product namePartnumberTray sizeMax no ofinstrumentsMax load(g)Vent to volumeratio(in⁻¹)
PrimeTaperSurgical Tray GSMedium68017268Medium55563.50.033
OmniTaperSurgical Tray GS68017203Medium56563.50.033
Bone ReamerTray68015280Small24271.90.046
Prosthetic Tray68015288Small22271.90.046

Table 1: Surgical and Prosthetic Trays configurations

Indications for Use

The Surgical and Prosthetic Trays are intended for organizing, sterilizing and storing of instruments.

The Surgical and Prosthetic Trays are not intended to maintain sterility and are to be used in conjunction with a legally marketed, validated sterilization pouch.

Sterilization parameters

Pre-vacuum Steam at 132°C (270°F) for 4 minutes with a 20 minute dry time. Do not exceed the worst-case validated maximum load:

Product nameArticleNumberMaximumLoad (g)Vent to VolumeRatio (in-1)
OmniTaper Surgical Tray68015282513.70.033
PrimeTaper Surgical Tray68015321513.70.033
PrimeTaper Surgical Tray GSMedium68017268563.50.033
OmniTaper Surgical Tray GS68017203563.50.033
Bone Reamer Tray68015280271.90.046
Prosthetic Tray68015288271.90.046

Comparison of Technological Characteristics

An overview of the similarities and differences between the proposed and predicate device is given in Table 2. The proposed Surgical and Prosthetic Trays and the predicate device Surgical Trays (K212281), have the same intended use, same sterilization method and parameters, are materials, undergo the same manufacturing process, have a similar design, and have identical instrument holder material, design and spacing between the holders. The use of the proposed Surgical and Prosthetic Trays has been validated through performance, biocompatibility and sterilization testing or via analysis of existing data for inclusion.

{5}------------------------------------------------

Table 2: Similarities and Differences between the proposed and predicate devices

ElementsProposed DeviceSurgical and Prosthetic TraysPredicate DeviceSurgical Trays (K212281)Comparison
ManufacturerSirona Dental Systems GmbHSirona Dental Systems GmbHSame Manufacturer
Indications for useThe Surgical and Prosthetic Trays are intended for organizing,sterilizing and storing of instruments.The Surgical and Prosthetic Trays are not intended to maintainsterility and are to be used in conjunction with a legallymarketed, validated sterilization pouch.The Surgical Trays are intended for organizing,sterilizing and storing of instruments.The Surgical Trays are not intended to maintainsterility and are to be used in conjunction with alegally marketed, validated sterilization pouch.Similar
Sterilization parametersPre-vacuum Steam at 132°C (270°F) for 4 minutes with a 20minute dry time.Sterilization parameters:Pre-vacuum Steam at 132°C (270°F) for 4 minuteswith a 20 minutes dry time.The tested Surgical Tray represents the worst-casevalidated load of 513.7g.
Do not exceed theworst-casevalidatedmaximumload:ProductnameDo not exceed the following maximum load:Product name
ArticleNumber MaximumLoad (g) Vent toVolume Ratio(in-1) 68015282 513.7 0.033 68015321 513.7 0.033 68017268 563.5 0.033 68017203 563.5 0.033 68015280 271.9 0.046 68015288 271.9 0.046ArticleNumber MaxLoad(g) Vent toVolumeRatio (in-1) 68015282 513.7 0.033 68015321 513.7 0.033
OmniTaperSurgical TrayPrimeTaperSurgical TrayPrimeTaperSurgical Tray GSMediumOmniTaperSurgical Tray GSBone Reamer TrayProsthetic TrayOmniTaperSurgical TrayPrimeTaperSurgical Tray
Product codeKCTKCTSame

{6}------------------------------------------------

ElementsProposed DeviceSurgical and Prosthetic TraysPredicate DeviceSurgical Trays (K212281)Comparison
General DesignPlastic trays with locking lid. Co-molded silicone andsilicone grommet supportsPlastic tray with locking lid. Co-molded siliconeand silicone grommet supportsSame
DimensionsMedium Trays:7.3 in x 5.5 in x 2.4 inSmall Trays:5.4 in x 3.9 in x 2.4 in7.3 in x 5.5 in x 2.4 inSame for the medium trays,different for the small trays
MaterialBase - Radel 5000Lid - Radel 5000Overlay - Radel 5000Tooling support - SiliconeBase - Radel 5000Lid - Radel 5000Overlay - Radel 5000Tooling support - SiliconeSame
Air permeanceYes, allow moist heat (steam) penetration to achievesterilizationYes, allow moist heat (steam) penetration toachieve sterilizationSame
Mass ofmaximumsterilization loadOmniTaper Surgical Tray GSPrimeTaper Surgical Tray GSMediumProsthetic TrayBone Reamer Tray$563.5g$$563.5g$$271.9g$$271.9g$OmniTaper Surgical TrayPrimeTaper Surgical Tray$513.7 g$$513.7 g$Different
Vent to volumeratioOmniTaper Surgical Tray GSPrimeTaper Surgical Tray GSMediumProsthetic TrayBone Reamer Tray$0.033 in^{-1}$$0.033 in^{-1}$$0.046 in^{-1}$$0.046 in^{-1}$OmniTaper Surgical TrayPrimeTaper Surgical Tray$0.033 in^{-1}$$0.033 in^{-1}$Different
SterilityNon-sterileNon-sterileSame
SterilizationMethodMoist heat (steam)Moist heat (steam)Same
SterilizationParametersPre vacuum,At 132°C for 4 minutes with a 20 minute dry timePre vacuum,At 132°C for 4 minutes with a 20 minutes dry timeSame
Sterile barrierFDA cleared sterilization pouchFDA cleared sterilization pouchSame
ReusableYesYesSame

{7}------------------------------------------------

Non-Clinical Performance Data

The proposed Surgical and Prosthetic Trays are reusable devices provided non-sterile which need to be end user sterilized. Sterilization of the proposed trays was validated according to the FDA guidance document, "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling".

Non-clinical testing data submitted, referenced, or relied upon, including acceptance criteria and set specifications in test methodology and standards, are summarized below:

Test NameTest MethodologyProducttestedPurposeAcceptanceCriteriaResults
Sterilizationcycle validation• ANSI/AAMI/ISO17665-1:2006/(R)2013• ISO 17665-2:2009ProposeddeviceTo validate that the trayscan be sterilized viamoist-heat sterilizationas specified on labeling(132° C for 4 minutes)Sterilityassurance level(SAL) of ≤ 10-6Pass
Dryingvalidation• AAMI TIR12:2020• ANSI/AAMI/ISO17665-2:2009ProposeddeviceTo validate that the trayscan be dried as specifiedon labeling (drying timeof 20 minutes)No visiblemoisture≤ 3% weightgain ofpackaging andabsorbablematerialsPass
Cleaning(Manual Pre-Cleaning andautomatedCleaning)• AAMI TIR30:2011• ISO 15883-1:2014PredicatedeviceTo validate that the trayscan be cleaned asspecified on labelingNo visible soil.Protein and TOCcontent Limit$<5 [µg/cm²]$$<95 [µg/device]$Pass
Reprocessingof trays(cleaning andsterilization)Internal Test MethodPredicatedeviceTo confirm that the trayscan be reprocessed asspecified on labeling (upto 200 reprocessingcycles without any signsof abrasion)No signs of flushrust, rustcorrosion,deformation ordamagePass
Simulated useof traysInternal Test MethodPredicatedeviceTo confirm that the trayscan withstand simulateduse of up to 1,500repeated repositioncyclesNo significantwear of theholdersPass
PackagingstestingISTA 2A(2011)PredicatedeviceTo confirm that thepackaging protects thedevice during simulateddistributionNo signs ofcracks orruptures on thesurgical traysPass
CytotoxicityISO 10993-5:2009PredicatedeviceTo confirm that nocytotoxic substances arereleased afterreprocessing of the traysInhibition of cellproliferationmust be at orbelow 30%compared tountreatedculturesPass

Conclusion

The conclusions drawn from the non-clinical test data and evaluation support that the proposed Surgical and Prosthetic Trays are as safe, as effective and perform as well as the legally marketed predicate device Dentsply Sirona Surgical Trays (K212281).

§ 880.6850 Sterilization wrap.

(a)
Identification. A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.(b)
Classification. Class II (performance standards).