Intended Use

The Power Wheelchair is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.

Device Description

This Power wheelchair is a motor driven, indoor and outdoor transportation vehicle, which a device for assisting action handicapped people and disabled people to move. It is suitable for disabled people with mobility difficulties and elderly people. The device consists of front wheel, drive wheel, frame, controller, motor, armrest, push-handle, backrest, seat cushion, pedal, battery box and charger. The device is powered by 2 Li-ion Battery pack (24V 10Ah, 240Wh) with 20 Km (12.5 miles) range, which can be recharged by an off-board battery charger that can be plugged into an AC socket outlet (100-240V, 50/60Hz) when the device is not in use. The patient can activate the controller handle (joystick) to control the speed and direction of the wheelchair movement. In addition, when the patient releases the joystick, the joystick will return back to the central position and the wheelchair will be automatically stopped soon due to automatic electromagnetic brake system. Once the joystick is activated again move to other position, the wheelchair will be re-energized.

AI/ML Overview

The provided text describes the regulatory clearance for a Power Wheelchair, W5907, not an AI/ML powered medical device. Therefore, the information requested about acceptance criteria and studies related to AI/ML device performance (such as sample sizes for test and training sets, expert qualifications, adjudication methods, MRMC studies, standalone performance, and ground truth establishment) is not applicable or present in the document.

The document primarily focuses on demonstrating substantial equivalence to a predicate device (K113463) through non-clinical performance and biocompatibility testing, adhering to various ISO standards relevant to wheelchairs.

However, I can extract and present the performance criteria and reported performance based on the non-clinical testing described for the Power Wheelchair, W5907 as an analogy to acceptance criteria.


Acceptance Criteria and Reported Device Performance (Non-AI/ML Device)

Since this document describes a power wheelchair and not an AI/ML device, the typical acceptance criteria for AI/ML performance metrics (like sensitivity, specificity, AUC) are not applicable. Instead, the "acceptance criteria" here are defined by compliance with established international standards for wheelchairs. The reported device performance is implicitly that the device met these standards.

1. Table of Acceptance Criteria and the Reported Device Performance

Performance Criterion (Standard Compliance)Reported Device Performance
Mechanical & Electrical Safety/Performance
Static Stability (ISO 7176-1:2014)Complied (the subject device "met all design specifications and provided support of the substantial equivalence determination")
Dynamic Stability (ISO 7176-2:2017)Complied (the subject device "met all design specifications and provided support of the substantial equivalence determination")
Effectiveness of Brakes (ISO 7176-3:2012)Complied (the subject device "met all design specifications and provided support of the substantial equivalence determination")
Energy Consumption & Range (ISO 7176-4:2008)Complied (the subject device "met all design specifications and provided support of the substantial equivalence determination")
Dimensions, Mass, Maneuvering (ISO 7176-5:2008)Complied (the subject device "met all design specifications and provided support of the substantial equivalence determination")
Max Speed, Accel., Decel. (ISO 7176-6:2018)Complied (the subject device "met all design specifications and provided support of the substantial equivalence determination")
Seating & Wheel Dimensions (ISO 7176-7:1998)Complied (the subject device "met all design specifications and provided support of the substantial equivalence determination")
Static, Impact, Fatigue Strength (ISO 7176-8:2014)Complied (the subject device "met all design specifications and provided support of the substantial equivalence determination")
Climatic Tests (ISO 7176-9:2009)Complied (the subject device "met all design specifications and provided support of the substantial equivalence determination")
Obstacle-Climbing Ability (ISO 7176-10:2008)Complied (the subject device "met all design specifications and provided support of the substantial equivalence determination")
Test Dummies (ISO 7176-11:2012)Complied (the subject device "met all design specifications and provided support of the substantial equivalence determination")
Coefficient of Friction (ISO 7176-13:1989)Complied (the subject device "met all design specifications and provided support of the substantial equivalence determination")
Power & Control Systems (ISO 7176-14:2008)Complied (the subject device "met all design specifications and provided support of the substantial equivalence determination")
Information Disclosure (ISO 7176-15:1996)Complied (the subject device "met all design specifications and provided support of the substantial equivalence determination")
Resistance to Ignition (ISO 7176-16:2012)Complied (the subject device "met all design specifications and provided support of the substantial equivalence determination")
EMC (ISO 7176-21:2009, IEC 60601-1-2:2014)Complied (the subject device "met all design specifications and provided support of the substantial equivalence determination")
Set-up Procedures (ISO 7176-22:2014)Complied (the subject device "met all design specifications and provided support of the substantial equivalence determination")
Batteries and Chargers (ISO 7176-25:2013)Complied (the subject device "met all design specifications and provided support of the substantial equivalence determination")
Biocompatibility
General Requirements (ISO 10993-1:2018)Complied ("Biocompatibility Tests are carried out in accordance with ISO 10993-1: 2018")
Cytotoxicity (ISO 10993-5:2009)Complied ("including cytotoxicity (ISO 10993-5:2009)")
Sensitization & Irritation (ISO 10993-10:2010)Complied ("sensitization (ISO 10993-10:2010) and irritation (ISO 10993-10:2010)")
Other
Risk Analysis (ISO 14971:2019)Developed in accordance with the standard.
Software ValidationSoftware validation was performed.

2. Sample size used for the test set and the data provenance: Not applicable. The "test set" in this context refers to the physical device undergoing non-clinical bench testing according to international standards (e.g., ISO, IEC). The data provenance is from laboratory testing of the manufactured device.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable, as this is not an AI/ML device requiring expert ground truth for interpretation. Compliance with standards is typically assessed by qualified testing engineers/laboratories.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable, as no human interpretation or adjudication of AI/ML output is involved.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI/ML device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is not an AI/ML device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not applicable in the context of AI/ML. For this physical device, "ground truth" is defined by the objective pass/fail criteria specified within each referenced ISO and IEC standard.

8. The sample size for the training set: Not applicable. This is not an AI/ML device.

9. How the ground truth for the training set was established: Not applicable. This is not an AI/ML device.


{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

May 12, 2022

Zhejiang Innuovo Rehabilitation Devices Co., Ltd. % Ivy Wang Technical Manager Shanghai Sungo Management Consulting Company Limited 14th Floor, 1500# Central Avenue Shanghai, Shanghai 200122 China

Re: K220740

Trade/Device Name: Power Wheelchair, W5907 Regulation Number: 21 CFR 890.3860 Regulation Name: Powered Wheelchair Regulatory Class: Class II Product Code: ITI Dated: March 14, 2022 Received: March 14, 2022

Dear Ivy Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Heather Dean, PhD Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K220740

Device Name Power Wheelchair, W5907

Indications for Use (Describe)

The Power Wheelchair is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.

Type of Use (Select one or both, as applicable)

  • Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officen Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

l. SUBMITTER

Name: Zhejiang Innuovo Rehabilitation Devices Co., Ltd.

Address: No.196 Industry Road, Hengdian Movie Zone, Dongyang, Zhejiang, China

Name of contact person: Leo Zheng

Telephone: +86 18358936043

Fax: +86-579-89327232

Date prepared: 2022-05-05

II. Device

Device trade name: Power Wheelchair, W5907 Classification name: Powered wheelchair Regulation class: 2 Regulation number: 21CFR 890.3860 Panel: Physical Medicine Product code: ITI

III. Predicate device

K113463 Power wheelchair, PL00I

SUZHOU KD Medical Appliance Co. Ltd.

IV. Device description

This Power wheelchair is a motor driven, indoor and outdoor transportation vehicle, which a device for assisting action handicapped people and disabled people to move. It is suitable for disabled people with mobility difficulties and elderly people.

The device consists of front wheel, drive wheel, frame, controller, motor, armrest, push-

handle, backrest, seat cushion, pedal, battery box and charger.

The device is powered by 2 Li-ion Battery pack (24V 10Ah, 240Wh) with 20 Km (12.5 miles) range, which can be recharged by an off-board battery charger that can be plugged into an AC socket outlet (100-240V, 50/60Hz) when the device is not in use.

{4}------------------------------------------------

The patient can activate the controller handle (joystick) to control the speed and direction of the wheelchair movement. In addition, when the patient releases the joystick, the joystick will return back to the central position and the wheelchair will be automatically stopped soon due to automatic electromagnetic brake system. Once the joystick is activated again move to other position, the wheelchair will be re-energized.

V. Indication for use

The Power Wheelchair is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.

VI. Comparison of technological characteristics with the predicate device
-------------------------------------------------------------------------------------
AttributeSubject devicePredicate deviceDiscussion/Conclusion
ManufacturerZhejiang Innuovo Rehabilitation Devices Co., Ltd.SUZHOU KD Medical Appliance Co. Ltd./
Proprietary name, modelPower Wheelchair, W5907 (Q50 R Carbon)power wheelchair, PL00I/
510(k) numberK220740K113463/
Device classification nameClass IIClass IISame
Classification regulations21 CFR 890.386021 CFR 890.3860Same
Product codeITIITISame
Similarities
Indication for useThe Power Wheelchair is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.They are motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.Same
Intended userdisabled people with mobility difficulties and elderly peopledisabled or elderly person limited to a seated positionSame
Use conditionindoor and outdoor useindoor and outdoor useSame
Number of wheels4, including two front wheels and two rear wheels4, including two pivoting casters and two rear drive wheelsSame
Function of wheelsFront wheels: driventwo pivoting casters: drivenSame
AttributeSubject devicePredicate deviceDiscussion/Conclusion
wheels suitable forrotation, acceleration,retrogradeRear wheels: drivingwheels to control thespeed and directionwheels suitable for rotation,acceleration, retrogradetwo rear drive wheels:driving wheels to controlthe speed and direction
Movement controlmethodBy Joystick controlBy Joystick controlSame
Driving systemDirect drive on the rearwheelsDirect drive on the rearwheelsSame
Brake systemAutomatic electromagneticbrake systemIntelligent regenerativeElectromagnetic brakeSame
Braking distance≤1.5 mForward: 1.5m (59") at maxspeedSame
Batteryli-ion battery pack;rechargeable, 24 VDC10Ah *2pcsLi-ion, Rechargeable; 24VDC 20AhSame
Maximum distanceof travel on the fullycharged battery20 km20 kmSame
Battery chargerOff-board chargerInput: 100-240V, 50/60Hz,1.5A,Output: 24 Vdc, 2A;Off-board, Automatic TypeInput: 110-220 V / 50-60Hz,Output: 24 Vdc, 2A;Same
Differences
Main frame materialCarbon fiber materialaluminum alloyDifferent material usedfor frame, that suchdifference will not impactthe safety andeffectiveness of thesubject device as theperformance tests areconducted according toISO 7176 series. Thecarbon fiber material islighter and easy to carry.
back cushionPolyester fabricPU foam covered by nylonfabric clothdifferent material usedfor parts in contact withuser, which suchdifferences will notimpact the safety andeffectiveness of the
AttributeSubject devicePredicate deviceDiscussion/Conclusion
seat cushionrubber patch cloth andOxford fabricPU foam covered by nylonfabric clothsubject device asbiocompatibility tests arecarried out according toISO 10993 series.Different material usedfor parts in contact withuser, which suchdifferences will notimpact the safety andeffectiveness of thesubject device asbiocompatibility tests arecarried out according toISO 10993 series.
ArmrestT700PUDifferent material usedfor parts in contact withuser, which suchdifferences will notimpact the safety andeffectiveness of thesubject device asbiocompatibility tests arecarried out according toISO 10993 series.
Overall Dimension(lengthwidthheight)923570928mm880570890mmMinor difference onwheelchair dimensionwill not cause differentperformance. All safetyand performance havebeen validated with themaximum rated weightdummy.
Stowage Dimension(lengthwidthheight)370570767mm720570400mm
Front wheelsize/type6.5" x 1.5"/PU Solid tire6" x 2" /PU Solid tireMinor difference ondimension of drivenwheel will not causedifferent performance.
Rear wheel size/type8.5"x 1.8"/ PU Solid tire8" x 2.4"/PU Solid tireMinor difference ondimension of drivingwheel will not causedifferent performance.
Max speed forwardUp to 6 km/h (1.6 m/s),adjustableUp to 6 km/h (3.75 mph),variableminor difference on max.forwarding speed will not
AttributeSubject devicePredicate deviceDiscussion/Conclusion
Max backward SpeedLess than 3 km/h (0.4m/s)2.4 mph (3.84 km/h)cause different performance. lower speed will be more safety.lower speed on max. backward speed will be more safety.
Max loading weight136kg (~300 lbs)114 kg (251 1bs)Difference on loading weight will not cause different performance. more loading weight provide more convenient and stable performance for the transportation.
Maximum safe operational incline degree8 °9 °minor difference on safe operational incline degree will not cause new safety and effectiveness concerns are raised as both the static and dynamic stability under specific inclining degree have been evaluated according to standard ISO 7176 series.
MotorBrushless DC motor; 24VDC; 250W; 2pcsBrushless DC motor; 24 VDC; 180 W; 2 pcsminor difference on motor power will not cause different performance. larger power will provide more driving force, no safety and effectiveness concerns raised.
Electronic controllerBrushless dual-drive rocker controllerBrushless dual-drive rocker controllerSimilar controller is used, both the control system, including the joystick controller, the electromagnetic brakes and the user interface are similar. The joystick controls the directions
AttributeSubject devicePredicate deviceDiscussion/Conclusion
and speed of movement, and when the joystick is released, the powered wheelchair will slow down to stop and the brakes will automatically re-engage. The controller also provides the battery status displaying and abnormal condition displaying. Both of the control systems are evaluated according to standard ISO 7176-14:2008 and software validation requirement and there are no new safety and effectiveness concerns due to the difference.
Turning Radius765 mm31.5" (800 mm)The minor difference in the turning radius is caused by different size of wheelchair and may cause a little bit inconvenience when it turns in a narrow space while the minor difference will not raise any new safety and effectiveness concerns.
Maximum obstacleclimbing40 mm1.2" (30 mm)Longer distance in the obstacle climbing will not impact the safety and effectiveness of the subject device.

{5}------------------------------------------------

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

VII. Summary of substantial equivalence discussion

The power wheelchair complied with the requirements of ISO 7176-1:2014, ISO 7176-2:2001,

ISO 7176-3:2012, ISO 7176-4:2008, ISO 7176-5:2008, ISO 7176-6:2018, ISO 7176-7:1998,

ISO 7176-8:2014, ISO 7176-9:2009, ISO 7176-10:2008, ISO 7176-11:2008, ISO

{9}------------------------------------------------

7176-13:1989, ISO 7176-14:2008, ISO 7176-15:1996, ISO 7176-16:2012, ISO 7176-21:2009, ISO 7176-22:2014, ISO 7176-25:2013, IEC 60601-1-2: 2014, ISO 10993-1:2018, ISO10993-5:2009, ISO 10993-10:2010.

The intended uses for both devices are the same. Mainframes of two devices are folded by way of front and rear close, and frame materials all meet the Tensile Strength, Yield Load, and Elongation tests. The design principles of the controller and Driving system are the same, and both meet the requirements of the ISO 7176-14:2008. Software validation is carried out on both control systems. Brake system and speed control are designed in the same way as well, and both meet the requirements of the ISO 7176-3:2012. Maximum obstacle climbing and Maximum safe operational incline are slightly different while such differences will not impact the safety and effectiveness of the subject device or raise new safety and effectiveness concerns as well as both meet the requirements of the ISO 7176-2:2001, ISO 7176-10:2008. The biocompatibility of the Predicate device and Subject device meet the requirements of the ISO 10993-5:2009 & ISO 10993-10:2010.

The flame retardant test of the seat cushion/backrest of both subject device and predicate device is carried out according to the ISO 7176-16 test. Therefore, both devices are assured to be under the same safety level.

In conclusion, the technological characteristics, features, specifications, materials, mode of operation, and intended use of the device substantially equivalent to the predicate devices quoted above. The differences between the subject device and predicate devices do not raise new issues of safety or effectiveness. The differences between the subject device and predicate devices do not raise new issues of safety or effectiveness.

VIII. Summary of non-clinical testing

A Performance testing-bench

The following performance data were provided to verify that the subject device met all design specifications and provided support of the substantial equivalence determination.

  • . Risk Analysis developed in accordance with ISO 14971:2019.
  • . Software validation
  • ISO 7176-1:2014 Wheelchairs - Part 1: Determination of static stability
  • ISO 7176-2:2017 Wheelchairs - Part 2: Determination of dynamic stability of electric

{10}------------------------------------------------

wheelchairs

  • . ISO 7176-3:2012 Wheelchairs - Part 3: Determination of effectiveness of brakes
  • . ISO 7176-4:2008 Wheelchairs - Part 4: Energy consumption of electric wheelchairs and scooters for determination of theoretical distance range
  • . ISO 7176-5:2008 Wheelchairs - Part 5: Determination of dimensions, mass and maneuvering space
  • . ISO 7176-6:2018 Wheelchairs - Part 6: Determination of maximum speed, acceleration and deceleration of electric wheelchairs
  • . ISO 7176-7:1998 Wheelchairs - Part 7: Measurement of seating and wheel dimensions
  • ISO 7176-8:2014 Wheelchairs - Part 8: Requirements and test methods for static, impact and fatigue strength
  • ISO 7176-9:2009 Wheelchairs - Part 9: Climatic tests for electric wheelchairs
  • . ISO 7176-10:2008 Wheelchairs - Part 10: Determination of obstacle-climbing ability of electrically powered wheelchairs
  • . ISO 7176-11:2012 Wheelchairs -- Part 11: Test dummies
  • . ISO 7176-13:1989 Wheelchairs - Part 13: Determination of coefficient of friction of test surfaces.
  • ISO 7176-14:2008 Wheelchairs -- Part 14: Power and control systems for electrically powered wheelchairs and scooters -- Requirements and test methods
  • . ISO 7176-15:1996 Wheelchairs - Part 15: Requirements for information disclosure, documentation and labeling.
  • ISO 7176-16:2012 Wheelchairs -- Part 16: Resistance to ignition of postural support devices
  • . ISO 7176-21:2009 Wheelchairs - Part 21: Requirements and test methods for electromagnetic compatibility of electrically powered wheelchairs and scooters
  • ISO 7176-22: 2014 Wheelchairs - Part 22: Set-up procedures
  • . ISO 7176-25:2013 Wheelchairs - Batteries and chargers for powered wheelchairs
  • . Electromagnetic Compatibility Testing in accordance with IEC 60601-1-2:2014.

{11}------------------------------------------------

A Biocompatibility of patient-contacting material

Biocompatibility Tests are carried out in accordance with ISO 10993-1: 2018, including cytotoxicity (ISO 10993-5:2009), sensitization (ISO 10993-10:2010) and irritation (ISO 10993-10:2010).

IX. Summary of clinical testing

No animal study and clinical studies are available for our device. Clinical testing was not required to demonstrate the substantial equivalence of the electric wheelchair to its predicate device.

X. Conclusions

The conclusions drawn from the nonclinical tests demonstrate that the subject device is as safe, as effective, and performs as well as the legally marketed predicate device K113463.

§ 890.3860 Powered wheelchair.

(a)
Identification. A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.(b)
Classification. Class II (performance standards).